Ethicon CONTOUR CS40B, CONTOUR CS40G Optimized Device Performance Manual

CONTOUR® Curved Cutter Stapler
Optimized Device Performance Guide
Retaining pin
Cartridge anvil
Cartridge module
Firing trigger
Push ro d Release button
Closure trigger
Retaining pin actuator
Handle
Proper reload selection
Tissue thickness should be carefully evaluated prior to using the instrument. Refer to the Instructions for Use for proper reload selection.
Device Staple Cut line Staple Max
code Tissue rows length line length irings Qty/box
CS40B Regular 4 40mm 51mm 6 3
CS40G Thick 4 40mm 51mm 6 3
Reload Closed Open Wire
code Tissue staple height stap le height d iameter Q ty/box
CR40B ■ Regular 1.5mm 3.5mm 0.20mm 6
CR40G ■ Thick 2.0mm 4.7mm 0.23mm 6
3. Either manually advance the retaining pin using the actuator button on top of the handle to capture the tissue
within the jaw opening (A), or automatically advance the retaining pin by partially squeezing the closure trigger (B). In the intermediate position, the retaining pin is fully seated in the anvil capturing the tissue, and the jaws are partially open, allowing for desired placement of tissue. Note that there is an audible click halfway through the closure stroke indicating that the device is in the intermediate position.
Device preparation and placement
1. Remove the staple retainer from the instrument
by pushing on the pad labeled “Push to Remove.”
2. Position tissue to be cut and stapled between the anvil and the cartridge. If the tissue doesn’t it comfortably
within the jaws, you should consider two irings.
Device is reloadable to complete a single transection with multiple irings when needed. Any tissue covering the hole in the anvil will be pierced by the retaining pin. The knife extends to the retaining pin and the staple line extends beyond the pin supporting multiple irings.
A B
Caution: To prevent premature iring, do not grasp the iring trigger.
4. Squeeze the closure trigger until latched, listening
for an audible click. After closing, holding the jaws in place for 15 seconds prior to iring may result in better tissue compression and staple formation.
To reposition tissue, open the jaws by squeezing the handle and closure trigger together slightly, and push the release button.
Single iring
Multiple irings
Device iring
5. Completely ire the instrument by squeezing the
iring trigger until it touches the closure trigger (plastic to plastic).
6. For a controlled opening, simultaneously squeeze
the closure trigger and push the release button with the thumb.
7. Retract the retaining pin manually. Pull back retaining pin actuator to ensure that the retaining pin
has been fully retracted.
8. To remove the reload, grasp the top and follow the
curvature of the device. Be sure to grab the body of the cartridge, not the black sleeve of the device.
9. Thoroughly clean the instrument by vertically
submerging completely the anvil and cartridge channel in a sterile solution and swish vigorously. After swishing, visually inspect the anvil and cartridge channel and remove all residual staples and/or foreign matter from instrument prior to reloading.
Reloading the device
CAUTION: To prevent bunching, stretching or overloading of tissue in the jaws of the device, multiple irings may be used to complete a single transection. If the
closure trigger is di icult to lock, reposition the instrument and take a smaller amount of tissue. Attempting to force the trigger to complete the closing stroke may result in poor staple formation, and this may result in leakage, lack of hemostasis or disruption of the staple line. In addition, instrument damage or failure may result.
Reloading the device (continued)
10. Insert the new reload by using the tracks on each
side as guides to align reload with jaws of the instrument. Following the curvature of the device, insert the new reload into the metal housing and snap into position.
For complete product information, including full steps for use, indications, contraindications, warnings and precautions, please see Instructions for Use.
©2018 Ethicon US, LLC . All Rights Reserved. 02939 9180607
11. Remove the staple retainer and check that the
reload is held irmly within the jaws. The loaded device cannot be used until the staple retainer is removed.
If the staple retainer has already been removed from the cartridge, the cartridge cannot be reloaded into the device, regardless of whether it has been spent or not.
The instrument may be reloaded ive times for a maximum of six irings per instrument during a single procedure. Some situations may require multiple irings to complete a single transection.
CAUTION: The staple retainer should not be removed until the cartridge is correctly loaded into the device.
Loading...