ESS Xenon XL-A100 User manual

operation manual
service manual
high-intensity light source
with automatic intensity-control
Xenon XL-M100
high-intensity light source
with manual intensity-control
Endoscopy Support Services, Inc.
3 Fallsview Lane Brewster, NY 10509
914-277-1700 Fax 914-277-7300
ess@endoscopy.com http://www.endoscopy.com
operation manual Xenon XL-A100 / Xenon XL-M100

table of contents page 1

1
safety reference / place the equipment _______________________________________________________2
1.1 safety reference ..........................................................................................................................................2
1.2 place the equipment....................................................................................................................................2
2 general advises / signs and symbols ________________________________________________________3
3 description of the equipment _______________________________________________________________5
4 operating elements _______________________________________________________________________6
4.1 operating elements on the frontpanel..........................................................................................................6
4.2 operating elements on the rear ...................................................................................................................7
5 connecting and operating the equipment_____________________________________________________8
5.1 operating the equipment .............................................................................................................................9
5.1.1 operating the equipment without video input ..........................................................................................9
5.1.2 operating the equipment with video input (only Xenon XL-A100) ........................................................... 9
6 service manual / maintenance of the equipment ______________________________________________11
6.1 possible causes / remedy..........................................................................................................................11
6.2 selection of the line voltage setting ...........................................................................................................11
6.3 exchanging the mains fuses......................................................................................................................12
6.4 replacing the lamp.....................................................................................................................................12
6.5 cleaning / desinfecting ..............................................................................................................................13
7 technical data___________________________________________________________________________14
8 spare parts _____________________________________________________________________________14
9 table ‘technical service-information’ ________________________________________________________15
file: Xenon XL-A100 - Xenon XL-M100.doc, date: 23. September 2003
operation manual Xenon XL-A100 / Xenon XL-M100
safety reference / place the equipment page 2

1 safety reference / place the equipment

1.1 safety reference
normal use
The equipment may only be used with accessories, wearing parts and disposable items, which have been designated by E.S.S., Inc as suitable for the instrument or the safety use of which is proven.
user qualification
The equipment may only be used by persons, who have a corresponding specialised qualification and who have been instructed in use of the equipment. It is the user’s responsibility to make sure, the equipment is safe and operates properly before using the equipment.
1.2 place the equipment
unpacking / items included
Carefully unpack the equipment and accessories and remove it from their packing. Check for missing items and evidence of shipping damage. File any complaints with the manufacture or supplier immediately. Retain the original packing materials for later use. These can come in handy, when the equipment must be transported. Please verify immediately after having unpacked the equipment, whether the delivery is complete. The standard extent of delivery includes of the following:
control unit
power supply cord
manual
safety precautions at the site of installation
Always place the equipment on a solid base. Make sure, that air circulation is sufficient. Never cover the louver type slots of the unit. The equipment may be used only in rooms having electrical installations conforming to applicable national, state and local electronical codes. The unit must be joined to the central potential equalisation of the operating theatre or of the equipment trolley by means of a grounding cable. The device must be connect to line voltage using the delivered protectively earthed power supply cord. The equipment may not be used in areas, where there are dangerous flammable gases. The equipment may only be connect to devices, which also comply to the IEC601-1. Never look direct in the lightbeam of a light source.
Comment:
operation manual Xenon XL-A100 / Xenon XL-M100
description of the equipment page 3
2 general advises / signs and symbols
Thank you for your expression of confidence in the E.S.S., Inc brand name. Like all of our other products, this product is the result of years of experience and great care in engineering and manufacture. This manual is destined to learn you understanding the function and the operation of your equip­ment. Before you switch on the equipment for the first time, please thoroughly read this manual and pay special attention to all safety instructions, so that endangering for the user and the patient is pre­cluded. Please always store this manual with the equipment.
data of the equipment
The type label (rear of unit) contains technical data, type and serial num­ber of your unit. Please always indicate these data when ordering spare parts or in case of any question.
Please enter here the technical data of your device! Î
warranty
1 year according to our warranty conditions. Opening the equipment or performance of any repairs or modifications of the equipment by unau­thorised persons shall relive E.S.S., Inc of any liability for its performance. Any such opening, re­pair or modification performed during the warranty period shall void all warranty. Wear parts are not included in the warranty. The firm of E.S.S., Inc shall be liable for failure or deterioration in the safe operation, operational reliability and performance of this equipment only subject to the conditions, that all assembly op­erations, system expansions, readjustments or repairs to same have been performed by a person or persons duly authorised by E.S.S., Inc, that all electrical installations at the location of us meet applicable national and local electrical codes and that the instrument has been used in accordance with its operating instructions at all times.
Serial No.: ___________________ Type: _______________________ Date: _______________________ Class: ______________________ Hz:_________________________ Amp.:_______________________ Volt: ________________________
operation manual Xenon XL-A100 / Xenon XL-M100
description of the equipment page 4
rights
All rights on this user manual, especially the rights of duplication and publication and the rights on translation are reserved. No part of this user manual may be reproduced by any means (by photo­static copy, microfilm or other methods) without preceding written consent of the E.S.S., Inc or be reproduced, multiplied or published by means of electronic data processing. The information given in this user manual can be changed or extended without notice and do not represent any liability. Errors and technical changes excepted. © E.S.S., Inc November 1998
symbols
attention, important note!
safety note!
service
signs
please read the enclosed instructions
unit model BF
beware of dangerous electrical voltage
connection for ground potential
alternating voltage
~
Loading...
+ 11 hidden pages