CelCulture CO2 Incubator ESCO
CCL-170B-8
CelCulture Control System
DECON Button & DECON LED
To activate 90oC moist heat decontamination procedure
During decontamination procedure, the yellow DECON LED will light up
MENU Button
When the main LCD shows normal display – MENU button will activate the
menu options
Within the menu display – MENU button will bring up the previous menu level
SET Button
Within the menu display – SET button will confirm a selection or value
UP/DOWN Buttons
Within the menu display – UP and DOWN buttons will scroll the display up and
down
When an input is required – UP and DOWN buttons will increase and decrease
a value
MUTE Button & ALARM LED
MUTE button will mute the audible alarm for a period of time
The red ALARM LED will light up whenever an alarm condition is triggered
![](/html/3c/3c9d/3c9d4c6971cdc2f98cfd9d07f2a234483c4827e5ec31873bcb2ac5e0e82731ca/bg2.png)
LABORATÓRIO DE ENSAIOS TECNOLÓGICOS EM ÁREAS
LIMPAS
ESTRADA NACIONAL 10, 2695-066 BOBADELA
2 HM
HEAT LED
HEAT LED will light up whenever a heating process is activated
INJECT LED
INJECT LED will light up whenever a gas injection process is activated
oC/%RH Display
When oC displayed, the display shows the temperature in the chamber
When %RH displayed, the display shows the relative humidity in the chamber
(optional)
%CO2/%O2 Display
When %CO2 displayed, the display shows CO2 concentration level in the
chamber
When %O2 displayed, the display shows O2 concentration level in the chamber
(suppressed O2 model)
Problems / Corrective Actions
Unit will not turn on
Possible Cause
Is the cord set connected to the power inlet at rear of cabinet as defined
in the installation section of the manual?
Is there power at the electrical outlet which the unit is plugged into?
Use voltmeter or test pen to check power on the electrical socket.
Is the unit’s power switch turned on?
Check whether power cord has been connected properly into the
wall/building socket and also into the incubator.
Check whether the power cord is giving power, measure the AC voltage
between the live and the neutral terminal of the cord by using a