ERICHSEN 243/I, 243 Series, 243/II, 243/II/4, 243/III Technical Description And Operating Instructions

...
testing equipment for quality management
Flowcups Models 243
A Simple Method
to Measure Viscosity
ASTM D 1200
SNV 37 110
DIN 53 211
ISO 2431
DIN EN ISO 2431
Technical Description and Operating Instructions
VISCOSITY CUPS, Models 243
Purpose and Application
Viscosity Cups remain the most important quick and simple method to measure the consistency of fluids in the laboratory and in production.
Viscosity Cups are, however, only suitable for lower viscosities and not for highly viscous substances such as pastes, etc.. The most accurate results are achieved with fluids within the range complying with Newton’s Law.
Design Details
Viscosity Cups consist of a hollow cylinder which terminate in a nozzle in the bottom. Excess fluid is collected in a gutter channel.
In general, the Viscosity Cups are made of anodised aluminium and the nozzles of stainless steel. Only the low price model 243/I is made of Phenol resin (but still with stainless steel nozzle).
Measuring Principle
The viscosity is established in terms of the time in seconds in which a precisely controlled volume of fluid flows out through the nozzle of the viscosity cup.
Test Procedure
The general test process is described as follows:
Level stand using spirit level Place cup in the stand or in the temperature
control jacket
Close nozzle Fill with fluid Wipe off surplus material across the edge
of the cup with a glass plate, and close cup with glass plate
Open nozzle Pull off glass plate horizontally and
simultaneously set stop watch in motion
At the first break off in the fluid flow stop
stop watch and note time
Further details are described in the various standards. It is to be noted that the Cups are only calibrated for the specified ranges and for the specified flow times.
In addition, it is particularly important to follow the instructions concerning temperatures carefully. Only a very small difference could lead to considerable changes and hence wrong measuring results. Of particular importance is to understand that it is not sufficient just to make sure that the fluid is at the right temperature: the Viscosity Cup must also be at the correct temperature, if necessary by surrounding it with a temperature jacket so that the intended temperature is maintained. It is also essential to check the temperature of the jet of fluid coming out of the Cup, since it is at this point that the actual temperature of the fluid being measured can be established. Comparable, reproducible results are only possible if the temperature is kept fully under control.
Care and Maintenance
After making a measurement the flow cup - particularly the nozzle - must be carefully cleaned without, however, using any hard or sharp implements.
Important: If remnants of fluid (paint) remain in the nozzle,
the effective size of the nozzle will be altered and the cup can no longer give accurate measuring results.
Reference Class:
All versions of Models 243/III and 243/VII as well as Model 243/II with nozzle dia. 4 mm are supplied with a Manufacturer’s Certificate M accordance with DIN 55 350-18 that includes among others the following information:
Flow time with percentage deviation from setting value, type of calibration oil, product identification, test equipments used with calibration status, date, name of inspector.
In the medium range of application the flow time is determined by means of a retraceably calibrated test oil. The maximum deviation from the setting value should not exceed 3 %.