Equus 3310 User Manual [en, es, fr]

E
315mA FUSED
CATII 600V
Title Page No.
1
E
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
SAFETY PRECAUTIONS/WARNINGS . . . . . . . . . . . . . . . . .1
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
TESTING PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . .6
BATTERY AND FUSE REPLACEMENT . . . . . . . . . . . . . . .10
MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
INTRODUCTION
Congratulations.You have purchased a precision instrument
manufactured to the highest quality standards. This Digital Multimeter is a general-purpose instrument designed for use in general electronics, home electrical applications, and auto­motive electrical/electronic systems.
This meter is designed to test or measure AC voltage, DC
voltage, batteries, DC current, resistance, diodes and contin- uity.
Please take the time to read these operating instructions thor-
oughly and completely .Failure to f ollow these instructions ma y result in electrical shock, instrument damage and/or damage to the equipment under test. Always use extreme caution when working on or around electrically operated equipment.
SAFETY PRECAUTIONS/ WARNINGS
Do not operate this multimeter before reading this manual in its
entirety.The following guidelines must be followed to avoid acci­dents that can result in electric shock or personal injury.
Pay close attention to WARNINGS stamped on the front
and rear of the meter's case.These warnings, as well as all warnings and precautions used through out this manual, must be followed to avoid electric shock and/or personal injury.
The RESPONSIBLE PARTY shall be made aware that, if the
equipment is used in a manner not specified by the manu­facturer, the protection provided by the equipment may be impaired.
Before using any of the functions on this meter, verify its
proper operation on a known similar function source where the unit value is also known.Take corrective action based on the indicated results.
To prevent electrical shock and/or damage to the tester or the equipment under test, observe the following safety precautions:
DO NOT apply more than the rated voltage, as marked on the meter, between terminals or between any terminal and earth ground.
Use caution when working above 30V AC rms, 42 V peak, or 60 V DC.Such voltages pose a shock hazard.
To avoid false readings that could lead to possible electric shock or personal injury , replace the batteries as soon as the low battery indicator displays.
Always inspect the multimeter, test leads and any other accessories for damage prior to every use. If any damage is found, do not use tester until repairs are done.
Always consider electrical and electronic equipment to be ener­gized (live).Never assume any equipment is de-energized.
Never ground yourself when taking electrical measurements. Isolate yourself from ground by using dry rubber insulating mats to cover all exposed/grounded metal. Stand on rubber mats and wear dry clothing.
Never take resistance measurements on energized (live) electrical or electronic equipment.
Use one hand, instead of two, whenever possible to take measurements.If two hands must be used, use extreme cau­tion not to contact any energized conductors with your hands. Be certain test leads are dry and clean.
Do not hold the instrument when taking measurements. Place the instrument on a clean, insulating surface prior to taking any measurement.
Don't become part of the circuit. Think safety.Act safely.
If working on a vehicle, take the following added precautions:
Only work on vehicle in a well ventilated area.
Always wear safety eye protection.
Avoid moving fan blades or any potentially moving parts.
Avoid hot engine parts.
Put transmission in "park" (automatic transmission vehicles) or "neutral" (manual transmission vehicles).Set the parking brake.
Turn the ignition "off" before connecting or disconnecting any testing equipment.
Put blocks on drive wheels.
Avoid wearing loose clothing or jewelry when working on a vehicle.
Read your vehicle's service manual and follow it's safety pro­cedures.
2
E
SPECIFICATIONS
3
E
GENERAL SPECIFICATIONS AND FEATURES
3-½ digit LCD display (maximum reading of 2000)
Manual range control
Automatic negative (-) polarity indication
Automatic zero adjustment
Over range indicator (except 10 A function). Displays "1" on
LCD Low battery indicator.Displays battery symbol on LCD
Automatic power shut off (after 15 idle minutes)
Safety standard: CE EMC/LVD. The meter is up to the stan-
dards of IEC1010 Pollution Degree 2, over voltage Category
2. Operating environment:
Temperature - 32° to 104° F.(0° C to 40° C) Humidity - Less than 80% relative humidity (non-condensing) Altitude - up to 6562 ft (2000 meters)
Storage environment:
Temperature - 4° to 140° F (- 20° to 60° C) Humidity - Less than 90% relative humidity (non-condensing)
Power Supply:Two 1.5V AAA batteries
Fuse: 315mA/250V 5X20 mm (Radio Shack, GMA/217
series; #270-1046) fuse Dimensions:
Height - 5.50 in. (139 mm) Width - 3.50 in. (89 mm) Depth - 1.25 in. (32 mm)
Weight (including batteries): approximately 5.7 oz (170 g)
ELECTRICAL SPECIFICATIONS
Accuracy of specifications in the following tables are based on an operating temperature of 64°F to 82°F (18°C to 28°C) and a relative humidity of less than 75%.
Maximum voltage from V ma socket to COM socket is 500V AC/DC; from COM socket to ground is 300V AC/DC.
DC VOLTS
RESISTANCE (OHMS)
DIODE/CONTINUITY TEST
DC AMPS
4
E
Range Resolution Accuracy Overload
Protection
200mV 100µV
500V DC
2V 1mV ±(0.8% of reading or AC 20V 10mV
+ 5 digits)
500V RMS
500V 1V
Input impedance: 10Mfor all ranges.
Range Resolution Accuracy Overload
Protection
200 0.1Ω
±(1% of reading
2K 1Ω
+ 5 digits)
250V DC/AC RMS
200K 10Ω 20M 10K ±(2% of reading
+ 5 digits)
Function Range Resolution Accuracy Overload
Protection
Diode test 1mV Displays,
voltage drop
250V
approximation
DC/AC Continuity 1 50Ω or less, RMS test buzzer (beeper)
will sound
Range Resolution Accuracy Overload
Protection
2mA
1µA
20mA
10µA
±(1.5% of reading
0.315A/250V fuse
200mA
100µA
+ 5 digits)
10A 10µA ±(2% of reading Non-Protected
+ 5 digits)
Measuring time is equal to or less than 15 seconds and time interval is equal to or over 15 minutes.
BA TTER Y TEST
AC VOLTS
CONTROLS AND INDICATORS
1. DCV Function: Measures
DC volts. Four (4) ranges: 200 millivolts (mV), 2, 20, and 500 volts.
2. POINTS Function :
Measures voltage drop across ignition points.
3. OHM (Ω) Function:
Measures resistance. Four (4) ranges: 200, 2k (2000), 200k (200,000) and 20M (20,000,000) ohms.
4. DIODE Function : For testing diodes.
5. CONTINUITY Function : Tests for continuity between two points.
6. 10A Input Jack.
7. COM Input Jack: Common Input Jack.
8. Volts, mA, OHMS, BAT, DIODE and CONTINUITY Input Jack.
9. DCA Function: Measures direct current. One (1) range:
10AMPS DC from 0 to 10 amperes.
10. mA Function. Measures current in milliamps. Three ranges: 2, 20 and 200 milliamps.
5
E
Range Resolution DC Current Load
1.5V 10mA 6V
0.01V
100mA
9V 10mA
12V 100mA
Overload
Range Resolution Accuracy Protection
200V 100mV ±(1.2% of reading 500V DC or AC 500V 1V
+ 10 digits) 500V RMS
Input Impedance: Approximately 5M Frequency: 40-400Hz
1 2
3 4
5 68
315mA FUSED
CATII 600V
7
15 14 13
12 11
10 9
11. BAT. LOAD TEST Function: To test small batteries. Four (4) ranges: 1.5, 6, 9 and 12 volts.
12. Function/Range Selector Switch: Selects desired func­tion or range.
13. ACV Function: Measures AC voltages. Two ranges: 200 and 500 volts.
14. OFF Function: Turns unit "off" when function is selected.
15. Liquid Crystal Display (LCD): Displays results of tests or
measurements.
Preparation and Caution Before Use
Inspect the digital multimeter for damage to the case.Do not use if cracked, distorted, excessively dirty or any abnormal condition exists.
Inspect the test leads for damage. Check for cracked insula­tion, broken or damaged probes, loose or bent probe pins. Do not use if any abnormal condition exists.
Set the Function/Range Selector Switch to the proper range BEFORE taking measurements. If the range/function must be switched during a test, ALWAYS remove the test leads from the circuit being measured before switching settings.
To avoid possible electric shock, instrument damage, and/or equipment damage when taking voltage or current measure­ments, DO NOT exceed the maximum value of the selected range.
If the unit is used near high noise Radio Frequency (RF) gen­erating equipment (spark plug wires, ignition coils or alterna­tor), the display may become unstable or indicate large errors. If you obtain erratic readings during use, isolate the Multimeter as far away as possible from these components.
TESTING PROCEDURES
A. AC/DC VOLTA GE MEASUREMENT
6
E
WARNING
To avoid possible electric shock, instrument damage and/or equipment damage, DO NOT attempt to measure voltages ABOVE 500V AC/DC or take measurements if the voltage is unknown. 500V AC/DC between the COM and V jacks is the maximum voltage that this instrument is designed to measure. The "COM" terminal potential should not exceed 300V AC/DC measured to ground.
1. Plug the BLACK test lead into "COM" jack of the multimeter; plug the RED test lead into the "V" jack.
2. Set the meter’s Function/Range Selector Switch to the
appropriate ACV or DCV position as desired (see Controls and Indicators, Items 1 and 13).
3. Place the RED test lead onto the positive (+) side of the item being tested and the BLACK test lead onto the negative (-) (across the source/load) side of the item.BE CAREFUL not to touch any energized conductors with any part of your body.
4. Read the results on the display.
B. RESISTANCE MEASUREMENT Ω(OHMS)
1. Plug the RED test lead into the "" jack of the multimeter; plug the BLACK test lead into the "COM" jack.
2. Set the meter’s Function/Range Selector Switch to the OHM
"Ω" range function (see Controls and Indicators, Item 3).
NOTE:T o obtain accurate readings , disconnect at least one side of the item under test from the circuit or circuit board before measuring resistance.
3. Place the RED test lead onto one side of the item being test­ed and the BLACK test lead onto the other side of the item. (Polarity does not matter when checking resistance).
4. Read the results on the display.
C. DIODE TEST
7
WARNING
Resistance measurements must be made on "de-ener­gized" (dead) circuits ONLY.Impressing a voltage across the multimeter's terminals while set to any resistance range may result in electric shock, instrument damage and/or damage to equipment under test. MAKE SURE equipment is completely de-energized before taking any resistance measurements.
E
WARNING
To avoid electrical shock and/or damage to the multimeter, ensure the power is removed from the circuit before any DIODE testing procedure is conducted.Test diodes on de­energized (dead) circuits only, never on live circuits.
NOTE: A diode is a semiconductor device that lets cur-
rent flow in one direction only .If the diode to be tested is part of a circuit (with other electronic components), you must isolate it from the other components by discon­necting at least one side of it from the circuit before testing. A good diode will show a low voltage drop across its junction (0.5-0.8 volts for a silicon diode or about 0.3V for a germanium diode) when the leads are connected in one polarity and a very high resistance (or open circuit) when the leads are reversed (con­nected in the opposite polarity).
1. Plug the RED test lead into the jack of the multimeter; plug the BLACK test lead into the "COM" jack.
2. Set the meter’s Function/Range Selector Switch to the position (see Controls and Indicators, Item 4).
3. Place the RED test lead onto one side of the diode being tested and the BLACK test lead onto the other side.
4. Read the results on the display.
5. Reverse the test leads and again read the results on the dis-
play. Compare the two readings. One reading should indi­cate a voltage drop value;the other reading should indicate an overrange (1) condition.
See note above
.
D. CONTINUITY TEST
1. Plug the RED test lead into the jack of the multimeter; plug the BLACK test lead into the "COM" jack.
2. Set the meter’s Function/Range Selector Switch to the
position (see Controls and Indicators, Item 5).
3. Place the RED Test Lead to one end of the wire or device being tested for continuity and the BLACK Test Lead to the opposite end.
4. Listen to the sound of the beeper and confirm the results by reading the display.
NOTE: The beeper will sound only if the continuity of the item under test (resistance between the two test leads) measures less than approximately 50 ohms.
8
E
WARNING
To avoid electric shock, shut off the power to the test arti­cle before testing it for continuity.
E. DC CURRENT MEASUREMENT (AMPS)
1. Plug the RED test lead into the "mA" or the "DC10A" jack of
the multimeter, as applicable; plug the BLACK test lead to the "COM" jack.
2. Set the meter's Function/Range Selector Switch to the appropriate Amps range position as desired (see Controls and Indicators, Items 9 and 10).
To measure from 0 to 200mA, set the Selector Switch to the desired "DCmA" position.
To measure from 200mA to 10 Amps DC, set the Selector Switch to the "DC10A" position.
3. Disconnect the battery, or shut off the power to the circuit being tested.
NOTE: To measure current on a particular circuit, you must open up the circuit and connect the test leads in series with the circuit before a reading can be obtained.
4. Disconnect one end of the wire or device, from the circuit where current will be measured.
5. Place the RED test lead on the disconnected wire and place the BLACK test lead at the location from which the wire was disconnected (series connection).
6. Reconnect the battery, or apply power to the circuit being tested.
7. Read the results on the display.
CAUTION: After the test is completed, shut the power off to the circuit before removing the test leads and before reconnecting any disconnected wires or devices.
NOTE: If the reading obtained is a negative number,
reverse the test leads.
9
WARNING
To prevent electrical shock when performing current measurements, follow all steps as indicated below DO NOT skip any steps or take any short cuts.
The DC10A range is not fused. To avoid current hazard and/or damage to the tester, DO NOT try to take meas­urements on circuits that have more than 10 amps. DO NOT take more than 15 seconds to take the reading. A waiting period of AT LEAST 15 MINUTES is necessary between every 15 second testing period.
E
F . BATTER Y TEST
1. Plug the RED test lead into the "BAT." jack of the multime- ter; plug the BLACK test lead into the "COM" jack.
2. Determine the voltage rating of the battery to be tested.
3. Set the meter’s Function/Range Selector Switch to the
desired "Battery Test" range (see Controls and Indicators, Item 12).
4. Place the RED test lead on the positive post of the battery under test; place the BLACK lead on the negative post.
5. Read the results on the display.
NOTE: Battery life is directly proportional to the current
draw (load) of the device that the battery is powering.To determine battery condition, consult the manufacturer of the device for minimum operating Voltage requirements.
BATTERY AND FUSE REPLACEMENT
1. Turn the Digital Multimeter “ OFF”and remove the test leads.
2. Remove the two screws on the back of the meter and sep-
arate the case.
3. Replace the fuse or batteries as necessar y:
For battery replacement: Remove the batteries from the battery compartment and replace only with two AAA
(1-½ volt) alkaline batteries.
For fuse replacement: Remove the fuse from
the fuse holder and replace with a 0.315A/250V - UL
Listed Bussmann, GMA Type (Radio Shack
GMA/270 series; #270-1046 ) fuse.
NOTE: Use a 0.315A/250V, 5x20mm type fuse ONLY -
Bussmann, GMA Type (Radio Shack #270-1046 or similar). Using an incorrect fuse may result in serious injury and/or damage to the unit.
4. Reassemble the case and secure with the two screws.
10
E
WARNING
When replacing the battery or the fuse, remove only the rear panel. Do not remove or disassemble the circuit board or the front panel, these items are not serviceable and if disassembled there is the possibility of loose metal parts shorting the circuit board and causing an electrocu­tion danger to the user.
MAINTENANCE
1. No periodic maintenance is required other than the replace-
ment of the battery, the fuse , and visual inspection of the meter.
2. Keep the meter clean and dry.DO NOT use solv ent to clean, use a damp (not wet) cloth and fully dry after cleaning.
3. The only replaceable parts are the 1.5 AAA batteries,
0.315A/250V fuse (see page 10 for battery and fuse replacement) and the Test Leads (for Test Leads call serv­ice department).
SERVICE PROCEDURES
The Manufacturer warrants to the original purchaser that this unit is free of defects in materials and workmanship under normal use and maintenance for a period of one (1) year from the date of orig­inal purchase. If the unit fails within the one (1) year period, it will be repaired or replaced, at the Manufacturer's option, at no charge, when returned prepaid to the Technical Service Center with Proof of Purchase.The sales receipt may be used f or this pur­pose. Installation labor is not covered under this warranty.
All replacement parts, whether new or re-manufactured, assume as their warranty period for only the remaining time of this war­ranty .This warranty does not apply to damage caused by improp­er use, accident, abuse, improper voltage, service, fire, flood, lightning, or other acts of God, or if the product was altered or repaired by anyone other than the Manufacturer's Technical Service Center. Consequential and incidental damages are not recoverable under this warranty. Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you.
This warranty gives you specific legal rights, and you may also have other rights, which vary from state to state. No portion of this warranty may be copied or duplicated without the e xpressed written permission from the Manufacturer.
Obtaining Warranty Service:
Products requiring service should be returned as follows:
1. Call the Technical Service Center to obtain a Return Reference Number:
USA & Canada = 1-800-544-4124 Other = 714-241-6805
2. Package the product carefully to prevent shipping damage
3. Include your name, return address, and a day contact phone
4. Enclose a copy of the dated sales receipt
5. Describe the problem
11
E
6. Ship prepaid to: Technical Service Center, 17291 Mt. Herrmann Street, Fountain Valley, CA 92708 U.S.A.
Phone:1-800-544-4124 or 714-241-6805 Fax:714-432-7910 Fax:714-432-7910 Web: www
.iEQUUS.com
Email: service@iEQUUS.com
12
E
13
14
Loading...
+ 33 hidden pages