Dear EasyPlus mini SMBG System Owner,
Thank you for using the EasyPlus mini Self-Monitoring Blood Glucose (SMBG) System. We design this
system for medical care on diabetes. The test results can help you determine the effects of food, exercise and diabetes medications. We design this system to be dependable, easy-to-use, compact, and lightweight and portable to help you monitor your blood glucose on a regular basis.
Please read this manual thoroughly before you begin testing. This manual will help you get the most from your EasyPlus mini SMBG System. It provides you and your diabetes care team with important information and step-by-step direction to use the EasyPlus mini Self-Monitoring Blood Glucose System. To start testing quickly, you can also refer to the Quick Reference Guide.
If you have questions, please contact your healthcare professional to get our help. You can also visit www.epsbio.com.tw
for diabetes management tools and product demonstrations.
Thanks again for choosing the EasyPlus mini SMBG system.
Page 4
Standard Accessories
Your new EasyPlus mini Blood Glucose meter and accessories work together to measure the amount of glucose in your blood. The system includes:
Optional Accessories
• EasyPlus mini Glucose Normal Control Solution
• GlucoManager Software Set
• Wallet
• User’s Manual
• Self-Test Log Book
• Warranty Card
• EasyPlus mini Blood Glucose Meter
• Test Strip Instructions
• Lancing device
• EasyPlus mini Blood Glucose
Test Strips (25 pcs)
• Check Strip
• Code Card
• Quick Reference Guide
• 1.5V (AAA) Alkaline Batteries (2 pcs)
• Lancets (25 pcs)
Page 5
Why is it so important to test blood glucose regularly?
Testing your blood glucose regularly can make a big difference in how you manage your diabetes every day. We’ve made this SMBG system as simple as possible and help you to use the meter regularly. Your meter is easy to use, and you can adjust the lancing device for your comfort.
Do you need Help?
If you have questions, please contact your healthcare professional. You can also visit www.epsbio.com.tw for diabetes management tools and product demonstrations.
Please complete your warranty card and mail it, so you receive the best customer service possible and product update news.
Although the EasyPlus mini SMBG System is easy to use, you may need to consult your healthcare professional (this may be your doctor, pharmacist or diabetes nurse educator) for instructions on how to use the system. Only the correct use of the system will ensure accurate results.
Page 6
Important Information About Your New Meter
EasyPlus mini blood glucose meter is designed and approved for testing fresh plasma from fingertip. The
meter is for outside the body (in vitro) use. It should not be used to diagnose diabetes.
EasyPlus mini blood glucose meter only use with EasyPlus mini Blood Glucose Test Strips. Other test
strips will give inaccurate results.
Testing is not valid for neonatal blood specimen.
Do not disassemble the meter as this may cause damage to the components resulting in incorrect
reading. Disassembling the meter will also void the warranty.
Always Keep the meter clean and store it in a safe place. Protect the meter from direct sunlight to ensure
a longer lifespan.
You should not store the meter and test strips in a car, a bathroom or a refrigerator.
Keep the meter, test strips and lancing device away from children and pets. You should not test critically ill patients with home-use blood glucose meters.
Elevated levels of acetaminophen, uric acid, gentisica acid, levodopa, dopamine, ascorbic acid and methyldopa may affect results.
Incorrect results may occur when performing the test. If you believe you are not feeling well, please
contact your healthcare professional.
Page 7
Remove batteries if the meter will not be used for one month or more.
Warning for potential biohazard: Healthcare professionals using this system on multiple patients should
be aware that all products or objects that come in contact with human blood, even after cleaning, should be handled as if capable of transmitting viral disease.
Do not touch the strip with a wet hand.
Do not use expired strips (the expiration date is shown on the bottle.)
Do not bend, cut or twist the strip.
Page 8
Health-Related Information
If you are very dehydrated, urinating frequently, low blood pressure, shock or hyperosmolar
hyperglycemic nonketotic coma (HHNK), you may get a test result that is lower than what your blood glucose really is. If you think you are dehydrated, call your doctor right away.
If you have followed the steps in the user’s manual, but still have symptoms that don’t seem to match
your test results, or if you have questions, please talk to your doctor.
Please read your test strip instruction carefully for additional health-related information.
Warning for potential biohazard:
Healthcare professionals using this system on multiple patients should handle all products or objects in contact with human blood carefully to avoid transmitting viral disease, even after cleaning.
Page 9
Explanation of Symbols
Please read instructions thoroughly before use
Attention, read instructions for use
Lot number
Single use only
For in vitro diagnostic use only
1.5V(AAA) x 2 batteries only
Expiration date
Storage temperature range
Manufactured by
Catalog number
SN/
Serial number
Control solutions
Contains sufficient for <n> tests
Authorized representative in the European Community
This product meets the requirements of Directive 98/79/EC in vitro diagnostic medical
devices
Page 10
Table of Content
Chapter 1: Understanding Your Meter ….......................................................................................................11
The EasyPlus mini Blood Glucose Meter …...............................................................................................11
Checking Your System …...........................................................................................................................13
Coding Your Meter .....................................................................................................................................14
Setting The Time And Date ........................................................................................................................15
Using EasyPlus mini Blood Glucose Test Strips ........................................................................................17
Chapter 2: Control Solution Test ...................................................................................................................18
Why Run A Control Solution Tests ............................................................................................................18
About The Control Solutions ......................................................................................................................19
Running A Control Solution Test ................................................................................................................20
Understanding Control Test Results …......................................................................................................23
Chapter 3: Testing Your Blood Glucose .......................................................................................................25
Using The Lancing Device .........................................................................................................................25
Inserting A Lancet into the Lancing Device ................................................................................................26
Running A Blood Glucose Test With Blood From Your Fingertip ..................
............................................29
Alternative Site Testing ……...…………………………………………………………………………………….32
Discarding Used Lancets ….......................................................................................................................34
Understanding Your Test Results …..........................................................................................................35
Unusual Test Results …………………………….......................................…………………………………...35
Comparing Your Meter Result To A Lab Result ….....................................................................................36
Page 11
Chapter 4: Meter Memory, Setup …...............................................................................................................37
Memory, Storing Test Result …..................................................................................................................38
Viewing & Deleting Test Results …............................................................................................................38
Setting The Alarm …...................................................................................................................................39
Chapter 5: Maintenance And Troubleshooting ….........................................................................................42
Installing Batteries …..................................................................................................................................42
Cleaning Your Meter ……...........................................................................................................................43
Maintenance And Testing ..........................................................................................................................44
Cleaning Your Lancing Device …...............................................................................................................44
Screen Messages And Troubleshooting …................................................................................................45
Chapter 6: Technical Information …..............................................................................................................49
Specifications ….........................................................................................................................................49
Device Information ….................................................................................................................................50
Warranty ...........................................................................
.........................................................................51
Page 12
11
Chapter 1: Understanding Your Meter
The EasyPlus mini Blood Glucose Meter
The Overview of the Display:
A. Year, month and date
B. Time
C. Test result
D. Glucose unit
E. Apply blood or control solution
F. Error code
G. Code number
H. Test records
I. Transfer data to PC
J. Insert test strip
K. Thermometer
L. Batteries status
When you use Software with the meter, the symbol of “Type” will show you to start operation.
Page 13
12
The Overview of the Meter:
A. Manufacturer information
B. Smart Jog Wheel
C. Serial number
D. Display
E. Anti-Slippery Pad
F. Battery Cover
G. Strip slot
The Overview of the Test Strip:
A. Blood collection area
B. Reaction area
C. Hand hold area
D. Strip insert direction
E. Electrode
The Overview of the Jog Wheel
Page 14
13
Setting The Time and Date—First Time Use
Setting the current time and date in your meter is important if you use the meter memory or if you want to download your results to a computer.
1. Press the wheel inwards until
turning the meter on. Press the wheel inwards and release to enter setting mode.
2. The last 2-digits of year flash
at the top left corner of the display.
Push the wheel up or down to
adjust the year.
3. Press inwards to confirm the setting. Push the wheel up or down to adjust the month. Push and hold the wheel to scroll the number faster.
Page 15
14
4. Press inwards and release
the wheel to set the month. The date will then flash.
Repeat step 3 to set the date.
5. Press inwards and release
the wheel to set the date.
The colon will then flash.
Push the wheel up or down to
adjust AM / PM (12-hour clock) or 24-hour clock.
6. Repeat step 3 and 4 to set the
hour and minute.
The flashing field is the one
you are setting.
Page 16
15
Checking Your System
A new check strip is included with your meter.
1. Make sure your check strip is
inserted. The meter will turn on automatically.
2. The display shows "OK".
3. Remove the check strip and
the meter will turn off automatically. Store the check strip in a safe place.
When the display shows “FAL”, re-insert the check strip. If FAL appears again, contact your local dealer for repair.
Page 17
16
Coding Your Meter
A new Code Card is inside each new box of test strips. Code your meter every time you open a new box of test strips.
1. The display will show "no
CODE" the first time batteries are installed.
2. Insert the code card into the
slot. The screen displays "OK".
3. The code number will be
shown on the display. Make sure the code number on the display matches the code number on the test strip bottle.
Do not discard the code card until you use all test strips in the same bottle.
Page 18
17
Using EasyPlus mini Blood Glucose Test Strip
Use only with EasyPlus mini Blood Glucose Meter.
Code your meter and run a control test every time you open a new box of test strips (See Chapter 2
"Control Solution Testing.")
Keep the test strips in their original bottle.
Close the bottle tightly right away, after you take a test strip out. This keeps the test strip dry.
Use the test strip within three minutes after you take it out of the bottle.
The strip is for single use only. Do not reuse it.
Record the date you open the test strip bottle. Be sure to check the expiration date on the test strip bottle.
The test strip is good for three months from the date the bottle is opened or until the expiration date on the bottle, whichever comes first.
Store the test strip bottle and your meter in a cool dry place.
Store the test strips between 35.6°F ~86°F(2°C ~30°C). Do not freeze.
Do not apply blood or control solution to the test strip before you insert it into the meter.
Do not touch the test strip with wet hands. Do not bend, cut, or twist the test strips.
Page 19
18
Chapter 2: Control Solution Testing
Why Run A Control Solution Test
Running a control test lets you know that your meter and test strips are working properly to give reliable results. You should run a control test when:
You use the EasyPlus mini Blood Glucose Meter for the first time.
You open a new bottle of test strips.
You think the meter or test strips may be working incorrectly.
You drop the meter.
You have repeated a test and the test results are still lower or higher than expected.
You are practicing the test procedure.
Professional users are instructed to follow federal, state, and local guidelines.
Page 20
19
About The Control Solution
Use only with EasyPlus mini test strips.
Write the date you opened the bottle on the bottle label. The control solution is good for three months
from the date the bottle is opened or until the expiration date on the bottle, whichever comes first.
Do not uses control solution that is past the expiration date.
The control solution can stain clothing. If you spill it, wash your clothes with soap and water.
Close the bottle tightly after every use.
Left over control solution should not be added back into the control bottle.
Store the bottle of control solution at room temperature, between 35.6°F ~86°F (2°C~30°C). Do not
freeze.
Page 21
20
Running A Control Solution Test
You need the meter, a test strip, and control solution.
2. If you miss seeing the code number, take the
test strip out and reinsert it into the meter.
After the display shows the code number, you will see the screen shown above.
1. Put a test strip into the meter in the direction of
the arrow. The meter turns on. Make sure the code number on the display matches the code number on the test strip bottle.
Page 22
21
3. Place the meter on a flat surface, like a table.
4. Remove the control solution bottle cap and
wipe the tip of the bottle with a tissue.
5. Squeeze the bottle until a tiny drop forms at
the tip of the bottle.
Touch the drop to the Blood collection area at
the end of the test strip.
Do not put control solution on top of the test strip.
When you hear the beeper, you have enough control solution in the test strip. The meter starts to count down from 5 seconds and will show the results.
Page 23
22
6. A result appears on the display.
Don’t remove the test strip yet. Check if the reading falls within the range printed on the test strip bottle.
7. Remove the test strip and throw it away
after you have compared the reading to the range printed on the test strip bottle.
Page 24
23
Understanding Control Test Results
The label on your test strip bottle shows the acceptable ranges for the Control Solutions. The result you get should be inside this range. Make sure you compare the result to the correct level of control.
When the control result is inside the range on the test strip bottle, your test strips and your meter is working properly.
If your control result is not inside the acceptable range (printed on your test strip bottle), here are some things you can do to solve the problem:
Page 25
24
Troubleshooting Checks Action
9 Was the test strip exposing to open air for a long
period of time?
Repeat the control test with properly stored strips.
9 Was the test strip bottle capped tightly?
This will humidify strips inside. Replace the test strips.
9 Did the code number on the meter display match
the code number on the test strip?
If they do not match, insert the correct code card in the meter and try again.
9 Was the meter functioning well?
You can use the check strip to verify the meter’s functions.
9 Is the control solution expired or contaminated?
Replace with new control solution to check the performance of SMBG system.
9 Were test strips and control solutions stored in
cool, dry places?
Repeat the control test with properly stored strips or control solutions.
9 Did you follow the testing steps properly?
Read Chapter 2 "Control Solution Testing" and test again. If you still have problems, please contact your local dealer for repair.
Page 26
25
Chapter 3: Testing Your Blood Glucose
Using The Lancing Device
The best depth setting is the lowest number that draws enough blood for a test. Try different settings to find the one that’s right for you.
Please do not share your lancing device with anyone. And always use a new, sterile lancet. Lancets are for one time use only.
Used Test strips and Lancets are considered bio-hazardous waste in accordance with local regulations and should be handled as if capable of transmitting infection. The users may discuss methods for disposing of used test strips and lancets with their doctor.
Page 27
26
Inserting A Lancet Into The Lancing Device
You must first load the lancet into the lancing device to get it ready for use.
1. Unscrew the Cap.
2. Insert the lancet into the lancing device firmly
then twist off the protective cover.
Page 28
27
3. Recap the front cap.
4. Select the desired penetration depth.
Select 1-3 for soft or thin skin. 4-5 for average. 6-8 for thick or calloused skin.
Page 29
28
5. Pull on the sliding barrel of the lancing device until it clicks and then release. Now the lancing device is ready.
Do not prick your finger until your meter and strip are prepared.
6. Set the lancing device aside until later in the test.
Page 30
29
Running A Blood Glucose Test With Blood From Your Fingertip
1. Wash your hands with soap
and warm water.
Rinse and dry thoroughly.
2. Put a test strip into the meter
in the direction of the arrow. The meter turns on. Make sure the code number on the display matches the code number on the test strip bottle.
3. If you miss seeing the code
number, take the test strip out and reinsert it into the meter. After the display shows the code number, you will see the screen shown above.
Page 31
30
4. Swab the blood collection area on your finger
with an alcohol swab. Please wait at least 5 seconds until the intended area is dry and clean before sampling the blood.
5. When the blood drop flashes on the display,
obtain a drop of blood from your fingertip.
6. Place the lancing device against the pad of
your finger. Press the trigger button to activate the lancing device.
Page 32
31
7. Gently squeeze your finger to assist the flow
of blood. This helps you get a blood drop.
Touch the drop to the tip of the transparent
window of the test strip.
Do not put blood on top of the strip.
Be sure to get enough blood on the strip’s reaction zone. Otherwise
,
an inaccurate reading may result.
8. The meter will beep when enough blood has
entered the strip’s reaction zone. The result will appear on the display after 5 seconds.
Page 33
32
1. This graphic shows where
the meter cleared for
alternative site testing.
2. Massage the puncture area of
forearm or palm for a few seconds.
3. Press and hold the device with
one clear adjustable tip against
the forearm.
Alternative Site Testing
Check with your health care professional before test other sites that not fingertip.Please do NOT use the first d rop of blood sample.
Page 34
33
4. If you choose the palm,
repeat step 1-3 to obtain a
drop of blood sample.
5. Press the trigger button to
activate the lancing device.
6. Hold the device against
forearm or palm and increase pressure until the blood sample size is sufficient
Page 35
34
Discarding Used Lancets
1. Unscrew and remove the adjustable cap.
Without touching the used lancet, stick the lancet tip into its protective cover. Grip the lancet carrier firmly and pull the lancet safely out.
2. Discard the used disposable lancet into an
appropriate sharps or biohazard container.
Page 36
35
Understanding Your Test Results
The EasyPlus mini Blood Glucose test strips are plasma referenced and calibrated for easier comparison to
lab results. The normal fasting blood glucose range for an adult without diabetes is 70~120 mg/dL (3.9~6.7 mmol/L).
Two hours after meals, the blood glucose range for an adult without diabetes is less than 140 mg/dL
(7.8 mmol/L).
For people with diabetes: please consult your doctor for the blood glucose range appropriate for
you.
Unusual Test Results
If your test result doesn’t match the way you feel, please follow these steps:
1. Run a control test, Chapter 2 "Control Solution Testing."
2. Repeat a blood glucose test, Chapter 3 "Testing Your Blood glucose."
3. If your test results still don’t reflect the way you feel, call your doctor immediately.
1. Extremely high humidity may affect the test results. A relative humidity greater than 90% may cause inaccurate results.
2. A red blood cell count (Hematocrit) that is either very high (above 55%) or very low (below 30%) may not provide accurate results.
3. Some studies have shown that electromagnetic fields may affect results. Do not test nea
r
an operatingmicrowave oven.
Page 37
36
Comparing Your Meter Result To A Lab Result
A common question is how the blood glucose results on your meter compare to the lab results. Your blood glucose can change quickly, especially after eating, taking medication, or exercising. If you test yourself in the morning, then go to the doctor’s office for a blood glucose test. The results will probably not match, even if you are fasting. This is typically not a problem with your meter, it just means that time has elapsed and your blood glucose has changed.
If you want to compare your meter result to the lab result, you must be fasting. Bring your meter to the doctor’s office, and test yourself by fingertip within five minutes of having blood drawn from your arm by a healthcare professional. Keep in mind that the lab could use different technology than EasyPlus mini blood glucose meter, and that blood glucose meters for self testing generally read somewhat lower or higher than the lab result.
For accuracy and precision data and for important information on limitations, see the instructions that come with your test strips.
Page 38
37
Chapter 4: Meter Memory, Setup
Memory, Storing Test Results
Your meter stores a maximum of 480 test results with the time and date of the test. You can review them at any time. When the memory is full, the oldest result is dropped as the newest is added, so it is very important to have the correct time and date set in the meter.
1. Do not change your therapy based on one individual result in memory.
2. The memory is not lost when you replace th e battery. You do need to check that the time and date are still correct. See Section "Setting the time and date" in Chapter 1.
3. Once 480 results are in memory, adding a new result causes the oldest one to be deleted.
Page 39
38
Viewing & Deleting Test Results
1. Press inwards and hold the
wheel until turning the meter on.
2. Push the wheel up or down to
view the previous results in order from the latest to the
earliest.
3. To delete a test result, press
inwards and hold the wheel for more than 3 seconds. The display will show "dEL".
4. Continue pressing the wheel
without releasing, the currently viewed result will be erased.
5. Press inwards and release
to review the average results.
6. Push the wheel up or down
to view averages for 7, 14, 30 or 90 days.
7. Press inwards to close the
memory mode.
Page 40
39
Setting The Alarm
The meter can set 4 alarm, that your control your test time as your requirements.
he
1. Press inwards and hold the
wheel until turning the meter
on.
2. Press inwards twice and the
display will show above.
3. Push the wheel down to
choose 4 alarm set. Press inwards to confirm the setting.
4. Push the wheel up or down to
choose “On” and press inwards to confirm the setting. The colon will then flash. Push the wheel up or down to adjust the time. (AM / PM, 24-hour clock) Press inwards to confirm the setting.
Page 41
40
5. If the time sets at 12:00 AM,
push up or down for twelve times to set at "12". Press inwards to confirm.
6. The minute flashes. Push the wheel up or down to
adjust the minutes, and press inwards to confirm the setting.
7. And then can set up other
alarms fitting in with your demand. Please repeat the step 2 to 5, and finally press inwards to confirm the setting.
8. If you don’t want to turn on
the alarm at all, turn every alarm "OFF" at the upper left corner of the display. Press inwards to confirm the setting.
Page 42
41
9. If you want to set one or more
alarms, repeat step 1-6.
10. If you turn the alarm OFF,
press inwards to confirm the setting and then enter the next setting mode.
11. If you don’t want to turn
any alarms On, make sure all OFF at the upper left corner and then press inwards to enter the setting mode of testing your blood glucose.
1. The alarm setting is still working, when the meter turns off.
2. Press the wheel inwards to stop the alarm.
Page 43
42
Chapter 5: Maintenance and Troubleshooting
Installing Batteries
The meter uses two alkaline 1.5V (AAA) batteries. Batteries will normally last for over 2000 tests. Other types of 1.5V (AAA) batteries are also acceptable, yet the capacity of test times may differ. Install the batteries when you first use the meter or replace with new batteries when the "LP" message and the low battery symbol appear on the display.
The display will show "no CODE" the first time batteries are installed. Please code your meter correctly.
1. The meter won’t delete earlier reco rd s after you replace batteries.
2. You should reset the time and date again after you replace the batteries.
3. 1.5V (AAA) x 2 batteries is available in shops. You may take the old batteries with you for replacement.
4. Remove batteries when you will not be using the meter for one month or more.
Low battery symbol
Page 44
43
Cleaning Your Meter
Caring for your EasyPlus mini SMBG system does not require special cleaning. Please keep the meter free of dirt, dust, bloodstain, and water stains. Follow these guidelines carefully to help you get the best performance possible:
Do Not:
Get any moisture in the test strip slot.
Spray any cleaning solution directly onto the meter.
Put the meter under water or liquid.
Pour liquid into the meter.
Do:
Make sure the meter is turned off.
Gently wipe the meter’s surface with a soft
cloth slightly dampened.
Page 45
44
Maintenance And Testing
Your meter needs little or no maintenance with normal use. It automatically tests its own systems every time you turn it on and lets you know if something is wrong. (See "Screen Messages" and what to do about them.)
To make sure the display is working properly, turn off the meter. Press and hold the Jog Wheel to see the complete display. All the indicators should be clear and look exactly like the picture to the left. If not, please contact your healthcare professional
Cleaning Your Lancing Device
To clean the lancing device, wipe them with a soft cloth dampened with water and mild detergent.
DO NOT place the entire device under water.
To disinfect the cap after cleaning by placing it in 70%~75% rubbing alcohol for 10 minutes at least
once a week. Allow the cap to air-dry after disinfecting.
Page 46
45
Screen Messages And Troubleshooting
Never make treatment decisions based on an error message. If you have any concerns, please contact your healthcare professional.
Code transmission failure
Action: Re-insert the code
card. If "FAL" appears consistently, contact local dealer for repair.
Humidified / Used strips
Action: Replace with a new
stri
p
.
Low power
Action: Replace with new
batteries.
Page 47
46
System fault
Action: Replace the batteries first. If 001
ERROR appears again, contact your local dealer for repair.
System fault
Action: Replace the batteries first. If Err
appears again, contact your local dealer for repair.
Page 48
47
The "Ht" and thermometer icon appears. Temperature is too high, outside the required range of 50°F ~104°F (10°C ~40°C). This alerts users that an incorrect result may occur if the test continues.
Action: Relocate the meter to a location with
temperature between 50°F ~104°F (10°C ~40°C).
The "Lt" and thermometer icon appears. Temperature is too low, outside the required range of 50°F ~104°F (10°C ~40°C). This alerts users that an incorrect result may occur if the test continues.
Action: Relocate the meter to a location with
temperature between 50°F ~104°F (10°C ~40°C).
Page 49
48
Test result is higher than 630 mg/dL.
Action: Test again. If the result
is still the same, please contact your healthcare professional.
The test result is lower than 20 mg/dL.
Action: Test again. If the result
is still the same, please contact your healthcare professional.
First time installation of batteries and the meter has not yet been coded.
Action: Code your meter.
Page 50
49
Chapter 6: Technical Information
Specifications
Brand name EasyPlus mini Blood Glucose Meter
Range 20~630 mg/dL
Response time 5 seconds
Memory sets 480 test results
Temp. 50°F ~104°F (10°C ~40°C)
Operating condition
Relative Humidity R.H. ≦ 90%
0.6 μL
Blood sample
Fresh blood from fingertip, palm or arm
Hematocrit (Hct) 30~55%
Power 1.5V (AAA) Alkaline Battery
Battery life Over 1000 tests
Display dimension 51 × 33 (mm)
Device dimension H × W × D (mm) 88 × 54 × 22 (mm)
Weight 58 grams w/o batteries
Principles Electrochemical biosensor technology
Page 51
50
Device Information
EasyPlus mini SMBG System, EasyPlus mini Blood Glucose Test Strips, EasyPlus mini Blood Glucose Meter. GlucoManager Software Set
Manufacturer:
EPS BIO TECHNOLOGY CORP.
2F, No.49-2, Lane 2, Guang-Fu Rd., Sec. 2, Hsinchu City, Taiwan 300
TEL:+886-3-5752522 FAX:+886-3-5752552 E-mail:[email protected]
Website:http://www.epsbio.com.tw
Obelis s.a.
34, Av. de Tervuren, bte 44, B-1040 Brussels, Belgium TEL:(32) 2.732.59 54 FAX:(32) 2.732.60 03
Warranty
EPS warrants the original purchaser for a period of 5 years from the date of purchase. This means during the
Page 52
51
warranty period if your Self-Monitoring Blood Glucose System does not work for any reason (other than obvious abuse), EPS will replace it with a new system or an equivalent product free of charge.
Lancing Device
Meets the requirements of MDD 93/42/EEC
Manufacturer:
STAT MEDICAL DEVICES INC.
1835 N.E. 146th Street, North Miami, FL33181, USA
TEL:(001) 305-945-0011 FAX:(001) 305-949-7370 E-mail:[email protected]
Lancet
Meets the requirements of MDD 93/42/EEC
Manufacturer:
SAE HAN MEDICAL CORP.
# 700-113 PUB GOT-DONG, IL SAN-GU, GOYANG-CITY, KYUNGGI-DO, KOREA
TEL:82-31-923-4330 FAX:82-31-923-4331 E-mail:[email protected]
P/N. 718029400 Rev.A01
Loading...