
MU Self-Monitoring Blood Glucose System
Quick Reference Guide
(Individual Version)

2
Inserting Batteries
1. Open the battery door on the back of the meter by pushing the tab in the direction of the arrow and
pulling the door up.
2. Insert two batteries. The meter will beep to confirm the batteries are inserted correctly.
3. Put the battery door back in place and snap it closed.

3
Setting The Time and Date
Setting the current time and date in your meter is important if you use the meter memory.
3. Repeat step 2 to set the date and
time. The flashing field is the one
you are currently setting.
1. Press (Power Button)
and hold the button until
the meter turns on.
2. The last 2-digits of the year flashes
at the top of the display. Press
(Left Button) or (Right Button) to
adjust the year and press (Power
Button) to confirm the setting.

4
Running a Control Solution Test
You need the meter, a test strip, and control solution.
1. Put a test strip into
the meter in the
direction of the
arrow.
3. Place the meter on a flat
surface, like a table.
4. Remove the control solution
bottle cap and wipe the tip of
the bottle with a tissue.
5. Squeeze the bottle until a
tiny drop forms at the tip of
the bottle and place solution
on top of bottle cap.
Setting the meter in control solution mode will prevent the control solution reading from
being added to your memory and averages.
2. The meter turns on automatically and the icon of
shows itself.
Press (Left Button) to select the testing mode of
Control Solution and the icon of flashes.

5
Be aware that there are Normal and High ranges listed on your test strip bottle. The
Normal and High refer to the three different control solutions that are availabl e. Please
be sure that you are looking at the correct range.
6. Touch the drop to the blood collection
area at the tip of the test strip. Do not
put control solution on the top of test
strip. The meter starts to count down
from 5 seconds and will show the
results.
7. Do not remove the test
strip until you confirm
that the reading falls
within the range printed
on the test strip vial.
8. Push the Strip Ejector to eject
the test strip.
9. Dispose properly.
10. Please check if the result on
the display is the same as
audible result.

6
Inserting a Lancet into the Lancing Device
You must first load the lancet into the lancing device to get it ready for use.
1. Unscrew the Cap.
2. Insert the lancet into the lancing
device firmly then twist off the
protective cover.
3. Recap the front cap.
Lancets are for use only and a new, sterile lancet should be used each time you perform a test.

7
1. Select 1-2 for soft or thin skin, 3 for average, and 4-5 for thick or calloused skin.
2. Lancing device and lancets are not to be shared between users. Sharing lancing devices and
lancets may transmit blood borne pathogens, such as viral hepatitis.
6. Set the lancing device aside
until later in the test.
5. Pull on the sliding barrel of the
lancing device until it clicks and
then release.
Now the lancing device is ready.
Do not prick your finger until your
meter and strip are prepared.
4. Select the desired penetration
depth.

8
Running A Blood Glucose Test With Blood From Your Fingertips
You can also turn on the meter by inserting a test strip.
1. Wash your hands with
soap and warm water.
Rinse and dry thoroughly.
4. Place the lancing device against
the pad of your finger. Press the
trigger button to activate the
lancing device.
2. Put a test strip into the meter in the
direction of the arrow. The meter
turns on automatically. The icon
shows itself.
3. When the icons of and
flashes on the display, the testing
mode of blood drop is starting.

9
Please make sure to apply blood when the blood d rop appears on the display.
5. Gently squeeze and/or
massage your fingertip
until a round drop of blood
on your fingertip.
6. Touch the blood drop at the tip of the
transparent window of the test strip.
Do not put blood on top of the strip.
Be sure to get enough blood on the
strip’s reaction zone. Otherwise, an
inaccurate reading may result.
7. The meter starts to count down
from 5 seconds and then
displays the test result.
8. Push the Strip Ejector to eject
the test strip & dispose of both
strip & lancet properly. Remove
the test strip and the meter
turns off automatically.
Apply blood to the edge of
the test strip.
Do not apply blood on top
of the test strip.
Be sure to get enough blood
on strip to make it to the
confirmation window.
Not enough blood on
strip to get a test result.

10
4. Press the trigger button to
activate the lancing device.
5. Hold the device against
forearm and increase pressure
until the blood sample size is
sufficient.
2. Massage the puncture area of
forearm for a few seconds.
3. Press and hold the device with
a clear adjustable tip against
the forearm.
1. This graphic shows where the
meter cleared for alternative
site testing.
Running a Blood Glucose Test with Blood from Your Forearm
Please use the clear cap with the lancing device for AST testing.

11
2. Press and hold the device
with a clear adjustable tip
against the palm.
3. Press the trigger button to activate
the lancing device.
4. Hold the device against palm and
increase pressure until the blood
sample size is sufficient.
1. Massage the puncture area of
palm for a few seconds.
Running A Blood Glucose Test With Blood From Your Palm
Check with your healthcare professional before testing sites o ther that the fingertip.

12
Discarding Used Lancets
1. Unscrew and remove the
cap.
2. Without touching the used
lancet, stick the lancet tip
into its protective cover.
3. Pointing the lancing device
toward a container for sharp or
biohazard material, slide the
ejection button up to release the
covered lancet into the
container.

13
Viewing and Deleting Test Results
Each review requires the user to go back to the main screen (testing mode) by pressing the (Power Button).
3. To delete a test result, press
(Left Button) for more
than 2 seconds and display
shows “dEL”, press
(Power Button) to confirm
deletion.
4. Press (Left Button) or
(Right Button) to keep
reviewing the results.
2. Press (Left Button) for
more than 2 seconds to view
the control solution records in
order from 30 to 1. Press
(Power Button) back to the
testing mode.
1. Press (Left Button) or (Right Button)
to view averages of 7-/14-/30-/90-day test
results first and then all results.
Press
(Power Button) back to the
testing mode.

14
Running With Your Computer
To Transfer data, the meter can be turned on or turned off.
1. Use the USB cable to
connect to the meter and
your computer. The display
shows ”PC”.
2. The meter starts to transfer data
to the “ GlucoManager
TM
”
software.
3. W hen all data are uploaded
from the meter to the PC,
the meter shows “OK” and
beeps.

15
Screen Messages and Troubleshooting
Never make treatment decisions based on an error message. If you have any concerns, pleas e contact yo ur
healthcare provider.
Message
What it means? What to do?
Humidified / Used strips
The meter has detected a problem
with the test strip.
Repeat the test with a new strip.
Refer to pages 27-28 for information on
sample application.
Low powe r
The meter batteries do not have
enough power to perform a test.
Replace the new batteries.

16
Message What it means? What to do?
System error
There may be a problem with the
meter.
Replace the batteries first.
Refer to pages 15 and 40. If this error
message appears again, please contact
your local dealer.
Memory Erro
Replace the batteries first.
If ERROR 005 appears again, please
contact your local dealer.
No result in memory.
The test proceeds incompletely. The
meter was unable to recall this
result.
You can still perform a blood glucose
test and get an accurate test result.

17
Message What it means? What to do?
Volume detector error
The volume of blood or control
solution is NOT enough.
Replace with a new strip.
If Err appears again, please
contact your local dealer.
Test result is higher than 630
mg/dL (35.0 mmol/L).
Re-check your glucose level.
If the result is HI again, obtain and
follow instructions from your
healthcare professional without
delay.
The test result is lower than
20 mg/dL (1.1 mmol/L).
This may require immediate
treatment according to your
healthcare professional’s
recommendations. Although this
message could be due to a test
error, it is safer to treat first and
then do another test.

18
Message What it means? What to do?
The “HF” and thermometer icon
appears. Temperature is too high,
outside the required range of
10°C - 40°C (50°F - 104°F). This
alerts users that an incorrect
result may occur if the test
continues.
Relocate the meter to a location with
temperature between 10°C - 40°C (50°F 104°F).
The “LF” and thermometer icon
appears. Temperature is too low,
outside the required range of
10°C - 40°C (50°F - 104°F). This
alerts users that an incorrect
result may occur if the test
continues.
Relocate the meter to a location with
temperature between 10°C - 40°C (50°F 104°F).

19
Obelis s.a.
Boulevard Général Wahis 53, B-1030 Brussels, Belgium
TEL:(32) 2.732.59 54 FAX:(32) 2.732.60 03
E-mail: mail@obelis.net
Manufacturer:
EPS BIO TECHNOLOGY CORP.
No.8, R&D RD. III, Hsinchu Science Park, Hsinchu, Taiwan 30077
TEL:+886-3-6686868 FAX:+886-3-6686866
E-mail:info@epsbio.com.tw
Website:http://www.epsbio.com
P/N: 71800230BI_0021B_02