HotBlock™ Information and Markings 5
Unpacking and Installation 6
HotBlock™ Temperature Information 7
Potential Hazards and Maintenence 8
Care and Maintenance of Your HotBlock™
Circuitry Diagram 9
HotBlock™ Replacement Parts 10
Troubleshooting Your Block 11-13
Adaptation of EPA Methods for Use with Your HotBlock™
EPA Approval Letter 14
Method 200.2, Revision 2.8 15-16
Method 200.7 17-18
Method 200.8 19-20
Methods 245.1, 7470, and 7471 for Mercury Digestions 21-22
Method 365.2 for Phosphorus 23-24
Method 3050B 25-26
NIOSH Method 7303 27
Sample Preparation for Lead Analysis Using the GhostWipe 28-29
Method 3060 for use with the HotBlock™ and StirBase Systems 30-31
HotBlock™ Parts and Supplies
50mL HotBlock™ Accessories and Consumables 32-33
100mL HotBlock™ Accessories and Consumables 34
15mL HotBlock™ Accessories and Consumables 35
AirLite HEPA-Filtered Enclosure 36
Sample Filtration Supplies 36-37
GhostWipes for Surface Sampling 38
HotBlock™: Operation and Instruction Manual Warranty and Repair Policy
Limited Warranty
The Environmental Express HotBlockTM is warranted against defects in materials and workmanship when used in accordance
with applicable instructions, for a period of one year from the date of shipment. This warranty extends to parts, labor, and
any approved transportation charges. This warranty applies only to damage or failure caused by normal laboratory use. The
warranty is limited to product repair. If Environmental Express is unable to repair the HotBlock, the customer may, at his or
her option, receive a replacement unit or a full refund. Operating the HotBlockTM at temperatures higher than 150°C will void
the warranty.
Environmental Express makes no other warranty, expressed or implied for this product with respect to merchantability, tness
for a particular use or any other matter. Environmental Express is not liable for any consequential or compensatory damages
arising from use of, or in conjunction with this product. The maximum liability shall be the invoice price of this product.
Repair Policies
Under Warranty Repair:
If the HotBlock™ should fail to operate within the warranty
period (one year from date of shipment) Environmental
Express will repair it and ship it back to the customer at
our expense. The remainder of the warranty period will
be honored from the original ship date. Environmental
Express will bear the cost of ground transportation both
to and from the customer’s location, and bear the cost of
any parts, labor and cleanup required. However, if it is
determined that the damage to the HotBlock™ was caused
by negligence or improper use, this warranty will not apply.
The warranty is also void if the system is used beyond its
intended purpose or in the event of any unauthorized
repair. In such cases, reasonable and customary repair
charges will apply. Repair charges will be quoted prior to
work being done.
Loaner HotBlocks™ MAY be available during the repair period. There are only a limited number of these units.
A reasonable charge for “cleanup” will be charged if a loaner is issued. The customer will be responsible for all
shipping charges associated with a loaner unit.
Out of Warranty Repair:
If the HotBlock™ fails after the warranty period has lapsed,
the repair procedure is as follows:
First, notify an Environmental Express Technical Service
Representative of product’s failure and place an order for
repair. Whenever possible, our customer service technician
will walk you through possible troubleshooting scenarios
which may enable you to repair your block on site. (See the
troubleshooting section of this manual, pages 11-13).
If on site repair is not possible, the customer may return
the non-working unit to Environmental Express using
appropriate shipping containers and insurance. Repair
charges will be assessed and estimated prior to work being
done. Repair charges will include all freight costs as well as
reasonable and customary charges for parts and labor.
Call 800.343.5319 or 843.881.6560 • www.environmentalexpress.com
2345 A Charleston Regional Pkwy • Charleston, SC 29492
Product Information:
Item # Date of Purchase
HotBlock™ Serial #
Please record the serial # of your HotBlock™ here for easy reference. Your serial # is located on the back of your HotBlock™.
HotBlock™: Operation and Instruction Manual Declaration of Conformity
The manufacturer, Environmental Express, 2345A Charleston Regional Parkway,
Charleston, SC 29492 declares that the following products, HotBlock™ Catalog
Numbers, SC196, SC154, SC150, SC151, SC100, and C6002, are in conformity with:
Standard for Safety Electrical Equipment for
Measurement, Control, and Laboratory Use; Part 1
General Requirements, UL 61010-1, CAN/CSA-C22.2
No. 61010-1, 2nd Edition, Issued 12 July, 2004 with
revisions through and including 28 October, 2008;
Equipment for Measurement, Control, and Laboratory
Use Part 2-010: Particular Requirements for
Laboratory Equipment for the Heating of Materials,
IEY 61010-2-010, 2nd Edition, Issued 1 June, 2003,
Safety Requirements for Electrical Equipment for
4000040
Environmental Express, Inc. declares that all HotBlocks
conform with the essential requirements of the applicable EC directives.
Measurement, Control, and Laboratory Use - Part
2-010:Particular Requirements for Laboratory
Equipment for the Heating of Materials, CSA
C22.2.61010.2.01
HotBlock™: Operation and Instruction Manual HotBlock™ Information and Markings
Environmental Express HotBlock™
Environmental Express HotBlocks™ provide an efcient method of digesting and
storing water, wastewater, soil and sludge samples for metals analysis. These
innovative digestion systems allow samples to be digested in a corrosion-free
environment. In addition, samples are handled in a small area with minimal
radiant heat loss. Users should be aware of potential dangers from heating certain
types of compounds. Such hazards may include explosion or the release of toxic or
ammable gases.
Denitions/Markings
Each HotBlock™ displays certain markings and symbols. All personnel working
with the HotBlocks™ should have an understanding of the following symbols and
denitions:
• V = voltage
• ~ = alternating current
• Hz = frequency
• A = amperes
This symbol means Caution Hot Surface. The surface of the
HotBlock™ may be too hot to
safely touch with bare hands.
This symbol means Read
and become familiar with
instructions before operation of
instrument.
HotBlock™: Operation and Instruction Manual Unpacking and Installation
Unpacking Your HotBlock™:
1. Remove the HotBlock™ from the shipping container by lifting from the
bottom of the block. The lid should not be used for lifting.
2. Your HotBlock™ is shipped with metal screws securing the bottom panel.
The metal screws must be removed before operating your HotBlock™.
Remove the metal screws and replace them with the PVC screws and
rubber feet included with your shipment. The corrosion-resistant PVC
screws and rubber feet are designed to secure the bottom plate.
Important: Do not over tighten the PVC screws!
Installation Requirements
Locate the HotBlock™ under a fume hood with a minimum face velocity of
100fpm, and allow a minimum of 2" of space on all sides. The following
environmental conditions should be observed:
• Ambient temperature range: 5-30°C
• Ambient relative humidity: 0-90%RH
• Altitude: sea level to 2500 meters
HotBlocks™ are rated as Pollution Degree 2 and Installation Category 2.
Electrical Requirements
Required Voltage: 120 volts, ~60Hz, 15A
(all HotBlocks™ are also available in 240V with CE mark)
Power should not vary greater than±10%. Use the supplied heat-resistant power
cord or equivalent to connect to the power supply.
For safety reasons, a separate power receptacle should be provided for each unit in
the system. Do not use extension cords or outlet adaptors. Make certain that power
outlets are earth-grounded at the grounding pin.
See individual specications for each HotBlock™ model, pages 3-4.
HotBlock™: Operation and Instruction Manual HotBlock™ Temperature Information
HotBlock™ Temperature Settings:
The pre-set factory “set point” temperature of your HotBlock™ is 106°C. Factory
tests have shown that this temperature is “sea level safe.” Liquids in uncovered
polypropylene tubes should not boil at this setting. Please note that the set point
of the block is not the same as the temperature of the liquids being digested. The
block temperature should be optimized for the specic digestion. The temperature
of liquid contents of the digestion cup will vary according to:
• The material being digested
• The number of samples being digested
• The air movement of the digestion area
• The addition of a watch glass or reux cap
If watch glasses are being used, the control point temperature should be
lowered approximately 10°C to avoid boiling.
Note: The maximum sample temperature tolerance for our polypropylene digestion
vessels (SC475) is 130°C.
Remember that the temperature display (current block temperature) is not the
temperature of the sample. Sample temperature will usually be 5-15° less than
the display temperature.
Adjusting the Temperature of Your HotBlock™:
1. Plug the HotBlock™ into an approved receptacle. Turn the HotBlock™ on by
pressing the button on the back of the unit. Wait until the display shows the
current block temperature (in red) and the set point temperature (in green).
2. Press and hold or tap the or key. The display will show the set point
temperature on the right in blue. The adjustment is from ambient to 150°C
in increments of 0.1°C. There is no need to press the green (advance) or ∞
button.
Safe-Sample™ Temperature Protection:
Your HotBlock™ is protected from runaway temperatures by a fail-safe alarm
system. In the unlikely event that the heating system fails to respond to the
controller, the Safe-Sample™ system will automatically shut the system off and
sound an audible alarm.
This alarm sequence occurs if the actual temperature of the block reaches a
temperature that is fteen degrees higher than the set point temperature. If this
should occur, the HotBlock™ will stop heating, preventing the loss of samples. The
HotBlock™ must be turned off, then turned back on to reset the alarm.
If the alarm sounds, see the troubleshooting section of this manual, page 11-13.
HotBlock™: Operation and Instruction Manual Potential Hazards and Maintenance
Potential Hazards:
The HotBlock™ should only be operated by properly trained personnel using
standard laboratory safety practices.
Use extreme caution when operating the HotBlock™. Plastic and graphite surfaces
of the HotBlock™ may be too hot to safely touch with bare hands.
The HotBlock™ contains electrical circuits and devices and compounds operating at
dangerous voltages. Contact with these circuits, devices and components can cause
serious injury or painful electric shock.
Proper grounding is essential to avoid a potentially serious electric shock hazard.
Ensure that there is an internal ground connection between the metal base of the
system and the 3-pin, earth-grounded receptacle.
For safety reasons a separate power outlet receptacle should be provided for each
unit in the system. Do not use extension cords or outlet adaptors. Make certain
each power outlet is earth-grounded at the grounding pin.
See individual block specications for power requirements, pages 3-4.
Application of the wrong supply voltage can create a re hazard and a potentially
serious shock hazard, and could seriously damage the HotBlock™ system. See
specications for individual HotBlocks™.
Users should be aware of potential dangers from heating certain types of
compounds. Such dangers may include explosion or the release of toxic or
ammable gases.
Always lift the HotBlock™ from the bottom of the unit.
Maintenance:
Any service inquiries should be directed to Environmental Express Technical Service
Department.
After each use, clean exterior surfaces with a damp sponge to remove acid residue.
For acid spills, sponge with a diluted solution of sodium bicarbonate followed
by distilled water. Acid that is spilled directly into the digestion wells should be
neutralized and removed.
Before using any cleaning or decontamination methods except those
recommended, check with Environmental Express to conrm the proposed method
will not damage your HotBlock™.
Avoid excessive spills, as liquid allowed to overow into the HotBlock™ casing can
severely damage electronic components.
HotBlock™: Operation and Instruction Manual Troubleshooting Your Block
HotBlock™ Troubleshooting Guide:
Please consult the following troubleshooting guide if you experience problems with your HotBlock™. See wiring schematic
(page 9) for component identication. If you are unable to resolve the problem or if replacement components are necessary,
please contact technical service at 1-800-745-8218 as component replacement varies in degree of difculty. We recommend
that only qualied personnel attempt troubleshooting electrical components.
When the HotBlock™ is initially powered on, the controller will cycle through a self-test sequence. It will then display the
current temperature and begin heating until it reaches your set point temperature, where it will hold until the unit is powered
off. The set point may be changed at any time. A change in the controller’s factory default settings or a failed component may
cause the HotBlock™ to perform unsatisfactorily or render it inoperable.
The controller digital display will
not illuminate.
There are two possible reasons that your controller will not illuminate.
1. The controller is not getting voltage or;
2. The controller itself has failed internally.
The problem can be effectively diagnosed by determining if the controller is or
is not getting voltage using the following steps:
1. Conrm that the power cord is plugged securely into the HotBlock™
receptacle and a working outlet.
2. Conrm that the switch is in the “on” position. Press button on the back of
HotBlock™.
3. Check the fuse located in the power module:
a. Locate the fuse drawing indented into the power module next to the
socket.
b. Using a small screwdriver, pry open the fuse compartment cover.
c. Examine the exposed fuse for a break in the lament and if necessary,
check for continuity using a volt-meter.
d. If the fuse is determined to be blown, replace it with the spare fuse
located in the slide-out compartment beneath the operating fuse.
Caution: This procedure is a potential electrical hazard and should
only be performed by qualied personnel.
4. Inside the HotBlock™, check voltage leading from the power module to
the controller:
a. Remove the bottom panel of the HotBlock™ by unscrewing the rubber
feet.
b. On the back of the controller, locate the black wire at terminal 98 and
white wire at terminal 99.
c. Set your volt-meter on AC voltage.
d. Touch your red lead to the exposed white wire and black lead to the
exposed black wire.
e. If your volt-meter reads 110-122V, the controller is receiving power but
has failed internally. It must be replaced (see parts list, page 10).
f. If your volt-meter registers less than 110-122V, using step d above check
the black and white wires at the terminal board and then at the power
module to determine if there is a faulty connection.
HotBlock™: Operation and Instruction Manual Troubleshooting Your Block
The audible alarm has sounded
immediately after powering on
and the HotBlock™ will not heat.
The temperature has overshot
the set point and the audible
alarm has sounded.
There are two possible causes for your HotBlock™ to sound the alarm
immediately after the controller cycles through the self test. These are:
1. Your set point has been set to a value (≥) 15° less than ambient or current
set point temperature. Turn the set point to within 15° of the actual
temperature (blue numbers)
2. The controller is faulty. Call Environmental Express at 1-800-745-8218 for
more information.
The function of the fail-safe system is to cease heating of the HotBlock™ in the
event of a set point overshoot of 15°C and to alert the technician of the incident.
The HotBlock™ can be “fooled” into fail-safe mode if the set point is manually
changed to a value ≥15° below your current temperature. However, the primary
cause for the runaway temperature is a faulty relay that has exceeded its
useful life.
You may troubleshoot the relay by following these steps:
Caution: This procedure is a potential electrical hazard and should
only be performed by qualied personnel.
1. Power off your HotBlock™.
2. Remove the bottom panel of the HotBlock™ by unscrewing the rubber feet.
3. Power the HotBlock™ on and allow it to overshoot your set point
temperature.
4. Locate the solid state relay mounted to the bottom panel.
5. Set your volt-meter to measure AC voltage.
6. Touch your red lead to terminal T1 (white wire) of the solid state relay and
touch your black lead to a ground source (e.g., the green/yellow wire from
the power module or an empty terminal on the terminal board).
7. If your volt-meter reads 110V-122V, then the relay is stuck in the “closed”
position and it must be replaced (see parts list, page 10).
The temperature controller is
performing erratically or displays
an error message.
-OR-
The temperature controller is
ashing Er.L Attn- Thermocouple
First reset default settings:
1. Hold the up and down arrow buttons for six seconds until Ai Set appears
2. Hold the down arrow until glbl set is in the window
3. Press the green advance key to enter
4. Continue pressing the green advance key until none user appears.
5. Press the down arrow key until Set1 user appears.
6. Pressing the advance key will restore default settings.
Fault
To troubleshoot the thermocouple, follow these steps:
1. Power off your HotBlock™.
2. Remove the bottom panel of your HotBlock™ by unscrewing the rubber feet.
3. Locate the thermocouple wires at terminals R1 (yellow) and S1 (red) of the
controller and remove using a small screwdriver.
4. Cut the exposed ends of the two wires.
5. Strip 1/4" of insulation from each wire and reconnect them to the
appropriate controller terminal and tighten.
6. Power on your HotBlock™.
7. If your display continues to ash Er.1 Attn- the thermocouple is faulty and
must be replaced (see parts list, page 10).
HotBlock™: Operation and Instruction Manual Troubleshooting Your Block
The HotBlock™ will not heat
beyond ambient temperature.
A HotBlock™ that will not heat beyond ambient temperature typically has a failed
relay, heater mat or controller.
Relay: To test the relay, the output voltage must be determined with a volt-meter.
Caution: This procedure is a potential electrical hazard and should only
be performed by qualied personnel.
To measure the relay voltage, follow these steps:
1. Remove the bottom panel of your HotBlock™ by unscrewing the rubber feet.
2. Locate the solid state relay mounted to the bottom panel.
3. Set your volt-meter to measure AC voltage.
4. Touch your red lead to terminal T1(white wire) of the solid state relay and
touch your black lead to a ground source (i.e., the green/yellow wire from the
power module or an empty terminal on the terminal board).
5. If your volt-meter does not read 110V-122V, then the relay has stuck in the
open position and it must be replaced (see parts list, page 10).
Heater Mat: To test the heater mat, the resistance in ohms (Ω) must be determined
with a volt-meter. It is recommended that your heater mat be replaced if it
measures 25 ohms (Ω) or greater. It is also recommended that the thermocouple
and insulation be replaced as well, both are inexpensive parts that are not easily
accessible otherwise.
To measure your heater mat resistance, follow these steps:
1. Power off your HotBlock™ and remove the bottom panel of your HotBlock™ by
unscrewing the rubber feet.
2. Locate and disconnect the white wire connected to terminal T1 of the relay
and an identical wire on terminal #2 of the terminal board (note: terminal #2
of the terminal board contains 3 white wires. To ensure you have the correct
wire, trace it back and ensure it originates from the graphite portion of your
HotBlock™.
3. Set your volt-meter to measure ohms (Ω)
4. Touch the red lead to one of these wires and touch the black lead to the
remaining wire.
5. If your reading is “OL” (over limit) or a value greater than 25 ohms, then the
heater mat has failed and it must be replaced (see parts list, page 10).
Controller: To test the controller, the output voltage must be determined with a
volt-meter.
Caution: This procedure is a potential electrical hazard and should only
be performed by qualied personnel.
To measure voltage from the controller, follow these steps:
1. Remove the bottom panel of your HotBlock™ by unscrewing the rubber feet.
2. Locate the solid state relay mounted to the bottom panel.
3. Set your volt-meter to measure DC voltage.
4. Touch the red lead to terminal A1 (red) and the black lead to terminal A2 (blue)
of the solid state relay.
5. Your volt-meter should read 3V-16V.
6. If your volt-meter does not read 3V-16V, perform steps 3 and 4 on the red and
blue (5 and 6) wires at the terminal board and controller to determine if there
is a faulty or loose connection.
7. If you do not get a reading of 3V-16V at terminals Y2 (red) and W2 (blue) of the
controller then the controller has failed internally and it must be replaced (see
parts list, page 10).
HotBlock™: Operation and Instruction Manual EPA Method 200.2, Revision 2.8
Adaptation for EPA Method 200.2, Revision 2.8 for use with the
Environmental Express HotBlock™ Digestion SystemRevised 05.05
1.0 Scope and Application
1.1 The following procedures have been written as an aid to EPA digestion
procedure 200.2 for use with the Environmental Express HotBlock™.
EPA Method 200.2 is for the determination of total recoverable analytes
in groundwater, surface water, drinking water, wastewater and (with
the exception of silica) in solid samples such as sediment, sludge
and soil. Use EPA Method 200.2 for reference while following the
sample preparation steps outlined below. A complete list of elements
appropriate for analysis is included in EPA Method 200.2. Analysis can be
performed by ame atomic absorption, GF/AA, ICP and ICP-MS.
2.0 Apparatus and Materials
2.1 HotBlock™ for Metals Digestions — Model Number SC100, SC150* or
SC154
2.2 Polypropylene Digestion Vessels — Catalog # SC475 (or SC490 for use
with HotBlock™ SC150)
2.3 Ribbed Watch Glasses, Catalog # SC505 or Reux Caps, Catalog # SC506
(for use with SC490 vials for HotBlock™ SC150, use Ribbed Watch Glasses,
Catalog # SC610 )
2.4 FilterMate™ — Catalog # SC0401 (or appropriate FilterMate) for sample
ltration if necessary
*Note: For all procedures, when using the SC150 block with the SC490 digestion vials with
a 100mL sample, double the volume of all reagents and acids added.
3.0 Procedure, Aqueous Sample Preparation
3.1 Mix sample thoroughly to achieve homogeneity. For each digestion
procedure, transfer a 50mL sample (or appropriate volume for your lab)
into the SC475 polypropylene vessel. For best results, weigh the sample
directly in the vessel on a tared balance.
3.2 Add 1mL (1:1) HNO3 and 0.5mL (1:1) HCl, then swirl. Heat the sample
in the HotBlock™ at 95°C for 2.5 hours without boiling. Place a ribbed
watch glass (SC505) on top of the digestion vessel.
Note: When using a watch glass, adjust the control point temperature of the HotBlock™
so that a 50mL, 5% acid solution is heated to 85°C BEFORE placing the watch glass over
the sample. Laboratory tests have proven that the addition of the ribbed watch glass will
add approximately 10°C to the sample temperature, bringing the sample temperature
up to 95°C. Refer to page 7 of the operations manual for instructions on HotBlock™
temperature adjustment.
3.3 Using the polycarbonate transfer racks, remove samples from the
HotBlock™ and allow them to cool.
3.4 After cooling, dilute samples to 50mL with DI Water.
3.5 If necessary, lter with SC0401 (or appropriate FilterMate) to remove
insoluble material.
Note: The ltration step should be performed slowly with little pressure placed on the
plunger. If excessive back pressure occurs, stop ltration and allow sediments to “settle
out”. Applying pressure to the plunger may cause sample “blow through” allowing
sediment to pass through the lter into the digestate.
HotBlock™: Operation and Instruction Manual EPA Method 200.2, Revision 2.8
4.0 Procedure, Solids Sample Preparation
4.1 Sieve a dried sample using a 5-mesh polypropylene sieve and grind in a
mortar and pestle. Weigh a representative sample of 0.5 ±0.01g into a
digestion vessel.
4.2 Add 2mL of (1+1) HNO3 and 5mL of (1+4) HCl+DI. Cover with a ribbed
watch glass (SC505) or reux cap (SC506),and place in the HotBlock™
at 95°C. Refer to page 7 of the operations manual for instructions on
HotBlock™ temperature adjustment.
4.3 Heat sample for 30 minutes.
4.4 (Optional) Although step 3 is the nal heating step for EPA Method 200.2,
if the sample is suspected of having a high concentration of organic
compounds, it is recommend to complete this step:
Add 2mL of 30% H2O2 to the well-cooled sample. Allow the exothermic
reaction to occur (approximately 10 minutes) and place the sample
back in the HotBlock™ at a temperature of 10° less than the original set
point for an additional 30 minutes. The reaction with the H2O2 raises the
sample temperature. Boiling should not occur if the temperature of the
HotBlock™ is lowered.
Note: H2O2 helps aid in the breakdown of high organic compounds in the sample thus
creating a more complete digestion.
4.5 Using the polycarbonate transfer racks, remove samples from the
HotBlock™ and allow them to cool completely.
4.6 Bring sample volume to 50mL with DI water.
4.7 If necessary, lter with SC0401 (or appropriate FilterMate) to remove
insoluble material.
Note: The ltration step should be performed slowly with little pressure placed on the
plunger. If excessive back pressure occurs, stop ltration and allow sediments to “settle
out”. Applying pressure to the plunger may cause sample “blow through” allowing
sediment to pass through the lter into the digestate.
All QC samples, concentration limitations, elemental lists and reagent
specications are addressed in depth in EPA Method 200.2. Safety concerns are
also part of the full method. Follow the instructions listed in EPA Method 200.2.
These steps should only be used as a guide to help improve the performance of
your HotBlock™.
HotBlock™: Operation and Instruction Manual EPA Method 200.7, Revision 4.4
Adaptation for EPA Method 200.7, Revision 4.4 for use with the
Environmental Express HotBlock™ Digestion System Revised 05.05
1.0 Scope and Application
1.1 The following procedures have been written as an aid to EPA digestion
procedure 200.7 for use with the Environmental Express HotBlock™.
EPA Method 200.7 is for the determination of total recoverable analytes
in groundwater, surface water, drinking water, wastewater, and (with
the exception of silica) in solid samples such as sediment, sludge
and soil. Use EPA Method 200.7 for reference while following the
sample preparation steps outlined below. A complete list of elements
appropriate for analysis is included in EPA Method 200.7. Analysis can be
performed by ICP.
2.0 Apparatus and Materials
2.1 HotBlock™ for Metals Digestions — Model Number SC100, SC150*, or
SC154
2.2 Polypropylene Digestion Vessels — Catalog # SC475 (or SC490 for use
with HotBlock™ SC150)
2.3 Ribbed Watch Glasses, Catalog # SC505 or Reux Caps, Catalog # SC506
(for use with SC490 vials for HotBlock™ SC150, use Ribbed Watch Glasses,
Catalog # SC610 )
2.4 FilterMate™ — Catalog # SC0401 (or appropriate FilterMate) for sample
ltration if necessary
*Note: For all procedures, when using the SC150 block with the SC490 digestion vials with
a 100mL sample, double the volume of all reagents and acids added.
3.1 Mix sample thoroughly to achieve homogeneity. For each digestion
procedure, transfer a 50mL sample (or appropriate volume for your lab)
into the SC475 polypropylene vessel. For best results, weigh the sample
directly in the vessel on a tared balance.
3.2 Add 1mL 1:1 HNO3 and 0.5mL 1:1 HCl and swirl. Heat in the HotBlock™
at a sample temperature of 85°C and evaporate to about 10mL taking
care that the cup does not go dry. This will take approximately 2 hours.
3.3 Cover the vessel with a disposable watch glass (catalog # SC505) and
reux an additional 30 minutes.
Note: When using a watch glass, adjust the set point temperature of the HotBlock™ so
that a 50mL, 5% acid solution is heated to 85°C BEFORE placing the watch glass over the
sample. Laboratory tests have proven that the addition of the ribbed watch glass will
add approximately 10°C to the sample temperature, bringing the sample temperature
up to 95°C. Refer to page 7 of the operations manual for instructions on HotBlock™
temperature adjustment.
3.4 Using the polycarbonate transfer racks, remove samples from the
HotBlock™ and allow them to cool. Bring samples up to volume with DI
water.
HotBlock™: Operation and Instruction Manual EPA Method 200.7, Revision 4.4
3.5 If necessary, lter with SC0401 (or appropriate FilterMate) to remove
insoluble material.
Note: The ltration step should be performed slowly with little pressure placed on the
plunger. If excessive back pressure occurs stop ltration and allow sediments to “settle
out”. Applying pressure to the plunger may cause sample “blow through” allowing
sediment to pass through the lter into the digestate.
4.1 For each digestion procedure, transfer 0.5g ± 0.01g of a dried sieved
sample into the SC475 polypropylene vessel.
4.2 Add 2.0mL (1:1) HNO3 and 5mL (1:4) HCl and swirl. Cover the vessel with
a disposable watch glass (catalog # SC505) and reux an additional 30
minutes at a sample temperature of 85°C.
Note: When using a watch glass, adjust the set point temperature of the HotBlock™ so
that a 50mL, 5% acid solution is heated to 85°C BEFORE placing the watch glass over the
sample. Laboratory tests have proven that the addition of the ribbed watch glass will add
approximately 10°C to the sample temperature, bringing the sample temperature up to
95°C. Refer to page 7 of the operations manual for instructions on HotBlock™ temperature
adjustment.
4.3 Allow to cool and bring to 50mL volume with DI water.
4.4 Filter with SC0401 (or appropriate FilterMate) to remove insoluble
material.
All QC samples, concentration limitations, elemental lists and reagent specications
are addressed in depth in EPA Method 200.7. Safety concerns are also part of the
full method. Follow the instructions listed in EPA Method 200.7. These steps should
only be used as a guide to help improve the performance of your HotBlock™.
HotBlock™: Operation and Instruction Manual EPA Method 200.8
Adaptation for EPA Method 200.8, for use with the
Environmental Express HotBlock™ Digestion System Revised 05.05
1.0 Scope and Application
1.1 The following procedures have been written as an aid to EPA digestion
procedure 200.8 for use with the Environmental Express HotBlock™.
EPA Method 200.8 is for the determination of total recoverable analytes
in groundwater, surface water, drinking water, wastewater, and (with
the exception of silica) in solid samples such as sediment, sludge
and soil. Use EPA Method 200.8 for reference while following the
sample preparation steps outlined below. A complete list of elements
appropriate for analysis is included in EPA Method 200.8. Analysis can be
performed by ICP-MS.
2.0 Apparatus and Materials
2.1 HotBlock™ for Metals Digestions — Model Numbers SC100, SC150*, or
SC154
2.2 Polypropylene Digestion Vessels — Catalog # SC475 (or SC490 for use
with HotBlock™ SC150)
2.3 Ribbed Watch Glasses, Catalog # SC505 or Reux Caps, Catalog # SC506
(for use with SC490 vials for HotBlock™ SC150, use Ribbed Watch Glasses,
Catalog # SC610 )
2.4 FilterMate™ — Catalog # SC0401 (or appropriate FilterMate) for sample
ltration if necessary
*Note: For all procedures, when using the SC150 block with the SC490 digestion vials with
a 100mL sample, double the volume of all reagents and acids added.
3.1 For the determination of dissolved analytes in ground and surface
waters, pipet a 20mL or greater aliquot of ltered, acid-preserved sample
into the SC475 digestion vessel.
3.2 Add an appropriate volume of (1+1) HNO3 to adjust the acid
concentration of the aliquot to approximate a 1% (v/v) nitric acid
solution. If the direct addition procedure is being used, add internal
standards, cap and mix.
Note: If a precipitate is formed during acidication, transport or storage the sample
aliquot must be treated using procedures for Total Recoverable Analytes.
Note: This section applies to water samples containing turbidity of greater than 1 NTU.
4.1 For each digestion procedure, transfer 50mL of well-mixed, unltered,
acid-preserved sample into the SC475 polypropylene vessel.
4.2 Add 1.0mL (1+1) HNO3 and 0.5mL of (1+1) HCl and swirl. Heat in
the HotBlock™ at a sample temperature of 85°C. The HotBlock™ set
temperature should be approximately 105°C. The temperature of a
reference blank should be tested to ensure correct temperature.
HotBlock™: Operation and Instruction Manual EPA Method 200.8
4.3 Reduce the volume of the sample aliquot to about 10mL at this
temperature. This should take about 2.5 hours.
4.4 Place a ribbed watch glass (SC505) over the digestion vessel to reduce
additional evaporation.
Note: When using a watch glass, adjust the set point temperature of the HotBlock™ so
that a 50mL, 5% acid solution is heated to 85°C BEFORE placing the watch glass over the
sample. Laboratory tests have proven that the addition of the ribbed watch glass will add
approximately 10°C to the sample temperature, bringing the sample temperature up to
95°C. Refer to page 7 of the operations manual for instructions on HotBlock™ temperature
adjustment.
4.5 Gently reux the sample for 30 minutes.
4.6 Using the polycarbonate transfer racks, remove samples from the
HotBlock™ and allow them to cool.
4.7 Add reagent water to bring the sample to the 50mL mark on the
digestion vessel. Cap and mix.
4.8 If necessary, lter with SC0401 (or appropriate FilterMate) to remove
insoluble material.
Note: The ltration step should be performed slowly with little pressure placed on the
plunger. If excessive back pressure occurs, stop ltration and allow sediments to “settle
out.” Applying pressure to the plunger may cause sample “blow through” allowing
sediment to pass through the lter into the digestate.
4.9 Prior to analysis, adjust the chloride concentration by pipetting 20mL
of the prepared solution into another SC475 digestion vessel and bring
up to 50mL volume with reagent water. If the dissolved solids in this
solution are >0.2%, additional dilution may be necessary to prevent
clogging of the extraction and/or skimmer cones. Note the dilution factor
for concentration calculations. If the direct addition procedure is being
used, add internal standards, cap and mix.
5.0 Procedure, Solid Sample Preparation
5.1 For each digestion procedure, transfer 0.5g ± 0.01g of a dried sieved
sample into the SC475 polypropylene vessel.
5.2 Add 2.0mL (1:1) HNO3 and 5mL (1:4) HCl and swirl. Cover the vessel with
a disposable watch glass (catalog # SC505) and reux an additional 30
minutes at a sample temperature of 85°C.
Note: When using a watch glass, adjust the set point temperature of the HotBlock™ so
that a 50mL, 5% acid solution is heated to 85°C BEFORE placing the watch glass over the
sample. Laboratory tests have proven that the addition of the ribbed watch glass will add
approximately 10°C to the sample temperature, bringing the sample temperature up to
95°C. Refer to page 7 of the operations manual for instructions on HotBlock™ temperature
adjustment.
5.3 Allow to cool and bring to 50mL volume with DI water.
5.4 Filter with SC0401 (or appropriate FilterMate) to remove insoluble
material.
All QC samples, concentration limitations, elemental lists and reagent specications
are addressed in depth in EPA Method 200.8. Safety concerns are also part of the
full method. Follow the instructions listed in EPA Method 200.8. These steps should
only be used as a guide to help improve the performance of your HotBlock™.
HotBlock™: Operation and Instruction Manual Adaptation of EPA Methods 245.1, 7470, 7471
Adaptation of EPA Methods for Mercury Digestions 245.1,
7470, 7471, for use with the Environmental Express
HotBlock™ Digestion System Revised 05.05
1.0 Scope and Application
1.1 The following procedures have been written as an aid to EPA Digestion
Methods 245.1, 7470 and 7471 for use with the Environmental Express
HotBlock™. EPA Methods 245.1 and 7470 are for the determination of
mercury in aqueous samples and TCLP extracts. EPA Method 7471 is for
the determination of mercury in soil, oil and sediment. Use the full EPA
Methods for reference while following the sample preparation steps
outlined below.
2.0 Apparatus and Materials
2.1 HotBlock™ for Metals Digestions — Model Numbers SC100, SC150*, or
SC154
2. Polypropylene Digestion Vessels — Catalog # SC475 (or SC490 for use
with HotBlock™ SC150)
3. Ribbed Watch Glasses, Catalog # SC505 or Reux Caps, Catalog # SC506
(for use with SC490 vials for HotBlock™ SC150, use Ribbed Watch Glasses,
Catalog # SC610 )
*Note: For all procedures, when using the SC150 block with the SC490 digestion vials with
a 100mL sample, double the volume of all reagents and acids added.
3.0 Procedure, Aqueous Sample Preparation
3.1 Add 20mL of well-mixed sample to a SC475 digestion vessel or 2mL of
TCLP extract and 18mL water.
Note: This is a 10x dilution for instruments that can achieve the detection limit including
the dilution, otherwise digest 20mL of TCLP extract.
3.2 Add 20mL of standard solutions or appropriate amount of standard
spiking solution to give desired concentration when diluted to 20mL.
3.3 Add 0.5mL of concentrated HNO3.
3.4 Add 1.0mL of concentrated H2SO4.
3.5 Mix thoroughly. Add 3mL of 5% KMnO4 and let stand for 15 minutes.
If sample does not maintain purple or brown color, add additional
portions to all samples, blanks and standards until the color persists for
at least 15 minutes.
3.6 Add 1.6mL of 5% persulfate solution. Place watch glass or reux cap
on top of digestion vessel to allow pressure to vent while minimizing
evaporative loss.
3.7 Digest the sample for 2 hours at 95°C.
3.8 Remove samples and let cool to room temperature.
3.9 Add 1.2mL of sodium chloride/hydroxyl amine hydrochloride solution.
HotBlock™: Operation and Instruction Manual Adaptation of EPA Methods 245.1, 7470, 7471
4.0 Procedure, Soil, Oil and Sediments, 7471A:
4.1 Weigh 0.25-0.30g of homogenized sample into a tube. For best results,
weigh the sample directly in the vessel on a tared balance.
4.2 Add 5mL DI water and 5mL aqua regia.
4.3 Heat for 2 minutes at 95 ±3°C.
4.4 Allow samples to cool and add 25mL DI water.
4.5 Add 7.5mL of KMnO4 and let stand for 15 minutes. If sample does not
maintain purple or brown color, add additional portions to all samples,
blanks and standards until the color persists for at least 15 minutes. Heat
samples at 95 ±3°C for 30 minutes.
Note: When using a watch glass, adjust the set point temperature of the HotBlock™ so
that a 50mL sample is heated to 85°C BEFORE placing the watch glass over the sample.
Laboratory tests have proven that the addition of the ribbed watch glass will add 10°C to
the sample temperature, bringing the sample temperature up to 95°C. Refer to page 7 of
the operations manual for instructions on HotBlock™ temperature adjustment.
4.6 Let samples cool to room temperature and add 3mL of Sodium
Choloride-Hydroxylamine Sulfate solution.
All QC samples, concentration limitations, elemental lists and reagent specications
are addressed in depth in EPA Methods 245.1, 7470 and 7471. Safety concerns are
also part of the EPA Methods. These steps should only be used as a guide to help
improve the performance of your HotBlock™.
HotBlock™: Operation and Instruction Manual Adaptation of EPA Method 365.2
Adaptation of EPA Method 365.1, Determination of
Phosphorous by Semi-Automated Colorimetry, for use with
the Environmental Express HotBlock™ Digestion System Revised 05.05
1.0 Scope and Application
1.1 The following procedures have been written as an aid to EPA Method
3651 for use with the Environmental Express HotBlock™. EPA Method
365.1 is for the determination of phosphorous in drinking water, surface
water and saline water and in domestic and industrial wastes. Use EPA
Method 365.1 for reference while following the sample preparation steps
outlined below. Range is 0.01 to 0.5 mg/L.
2.0 Apparatus and Materials
2.1 HotBlock™ for metals digestions — Model Numbers SC100, SC150*, or
SC154
2.2 Polypropylene Digestion Vessels — Catalog # SC475 (or SC490 for use
with HotBlock™ SC150)
2.3 FilterMate™ — Catalog # SC0401 (or appropriate FilterMate) for sample
ltration if necessary
* Note: For all procedures, when using the SC150 block with the SC490 digestion vials
with a 100mL sample, double the volume of all reagents and acids added.
3.0 Procedure, Aqueous Sample Preparation
3.1 For each digestion procedure, transfer 50mL of sample (or appropriate
volume for your lab) into the SC475 polypropylene vessel.
3.2 Add 1mL of H2SO4 solution and swirl. (11 N H2SO4 — Slowly add 310mL
concentrated H2SO4 to 600mL distilled water. When cooled dilute to 1
liter.)
3.3 Add 0.4g of ammonium persulfate.
3.4 Heat the sample in the HotBlock™ at approximately 100°C for 40
minutes. The sample should slightly boil.
Note: Adjust the HotBlock™ temperature so that the sample obtains a temperature
of 100°C. Please refer to page 7 of the HotBlock™ Operation Manual for temperature
adjustment and principles of operation.
3.5 Using the polycarbonate transfer racks, remove samples from the
HotBlock™ and allow them to cool.
3.6 After cooling, dilute to 50mL with DI Water.
3.7 If necessary, lter with SC0401 (or appropriate FilterMate) to remove
insoluble material.
Note: The ltration step should be performed slowly with little pressure placed on the
plunger. If excessive back pressure occurs, stop ltration and allow sediments to “settle
out.” Applying pressure to the plunger may cause sample “blow through” allowing
sediment to pass through the lter into the digestate.
All QC samples, concentration limitations, elemental lists and reagent specications
are addressed in depth in EPA Method 365.1. Safety concerns are also part of the
full method. Follow the instructions listed in EPA Method 365.1. These steps should
only be used as a guide to help improve the performance of your HotBlock™.
HotBlock™: Operation and Instruction Manual Adaptation of EPA Method 3050B
Adaptation of EPA Method 3050B for use with the
Environmental Express HotBlock™™ Digestion System Revised 05.05
1.0 Scope and Application
1.1 The following procedures have been written as an aid to EPA Method
3050B for use with the Environmental Express HotBlock™. EPA Method
3050B is for the preparation of sediment, sludge, and soil samples
for analysis. Use EPA Method 3050B for reference while following the
sample preparation steps outlined below. A complete list of elements
appropriate for analysis is included in EPA Method 3050B. Analysis can
be performed by ame atomic absorption, GF/AA, ICP and ICP-MS.
2.0 Apparatus and Materials
2.1 HotBlock™ for Metals Digestions — Model Numbers SC100, SC150*, or
SC154
2.2 Polypropylene Digestion Vessels — Catalog # SC475 (or SC490 for use
with HotBlock™ SC150)
2.3 Ribbed Watch Glasses, Catalog # SC505 or Reux Caps, Catalog # SC506
(for use with SC490 vials for HotBlock™ SC150, use Ribbed Watch Glasses,
Catalog # SC610 )
2.4 FilterMate™ — Catalog # SC0401 (or appropriate FilterMate) for sample
ltration if necessary
* Note: When using the SC150 block with the SC490 digestion vials with a 100mL sample,
double the volume of all reagents and acids added.
3.0 Procedure, Sample Preparation
3.1 Mix sample thoroughly to achieve homogeneity and sieve if appropriate.
For each digestion procedure, weigh to the nearest 0.01 gram and
transfer a 0.5-1-gram aliquot (or appropriate amount for your lab) to the
SC475 polypropylene digestion vessel. For best results, weigh the sample
directly in the vessel on a tared balance.
3.2 Add 5mL (1:1) HNO3 + DI Water and swirl. Cover with a ribbed watch
glass or reux cap and heat the sample in the HotBlock™ at 95±5°C for
15 minutes without boiling.
Note: If using the watch glass, adjust the HotBlock™ temperature so that a 50mL, 5% acid
solution is heated to 85°C. Laboratory tests have proven that the addition of the ribbed
watch glass will add approximately 10°C to the sample temperature. Refer to page 7 of
the operations manual for instructions on HotBlock™ temperature adjustment.
3.3 Allow the sample to cool, then add 2.5mL concentrated HNO3 and reux
at 95°C for 30 minutes. Repeat this step until NO brown fumes are given
off by the sample.
3.4 Heat sample with the ribbed watch glass to a volume of about 5mL or for
2 hours at 95±5°C . Do not allow the sample to boil or go dry. Neither
of these aspects should occur if the temperature of the HotBlock™ is
adjusted correctly.
3.5 Using the polycarbonate transfer racks, remove samples from the
HotBlock™ and cool completely (it is very important that samples are
cooled completely).
HotBlock™: Operation and Instruction Manual Adaptation of EPA Method 3050B
3.6 Add 1mL DI Water and 1.5mL of 30% H2O2 slowly. Allow an exothermic
reaction to occur. Wait 5-10 minutes and place samples back in the
HotBlock™ with the ribbed watch glasses in place. If effervescence starts
to occur lift the samples out of the HotBlock™ and allow the reaction to
continue. Do not let the samples foam out of the vessel. (Reducing the
HotBlock™ display temperature by 10°C should reduce the effervescence
while maintaining the sample temperature).
3.7 Continue to add 0.5mL of H2O2 (no more than 5mL total) until the sample
remains unchanged in color (no longer than 30 minutes). Continue
heating for a total of 2 hours or until volume has been reduced to
approximately 5mL.
3.8 For the analysis of samples for FLAA or ICP-AES, add 5mL concentrated
HCl to each sample and cover with a ribbed watch glass and reux at
95°C for 15 minutes. GFAA and ICP-MS digestions, skip this step.
3.9 After cooling, dilute to 50mL with DI Water.
3.10 If necessary, lter with SC0401 (or appropriate FilterMate) to remove
insoluble material.
All QC samples, concentration limitations, elemental lists and reagent
specications are addressed in depth in EPA Method 3050B. Safety concerns are
also part of the full method. Follow the instructions listed in EPA Method 3050B.
These steps should only be used as a guide to help improve the performance of
your HotBlock™.
Adaptation of NIOSH Method 7303 for use with the
Environmental Express HotBlock™™ Digestion System Revised 05.05
1.0 Scope and Application
1.1 The following procedures have been written as an aid to NIOSH Method
7303 for use with the Environmental Express HotBlock™. Method 7303 is
for the preparation of cellulose ester membrane lters commonly used in
air sampling devices. Use EPA Method 7303 for reference while following
the sample preparation steps outlined below. A complete list of elements
appropriate for analysis is included in Method 7303. Analysis can be
performed by ICP or AA.
2.0 Apparatus and Materials
2.1 HotBlock™ for Metals Digestions — Model Numbers SC100, SC150*, or
SC154
2.2 Polypropylene Digestion Vessels — Catalog # SC475 (or SC490 for use
with HotBlock™ SC150)
2.3 Ribbed Watch Glasses, Catalog # SC505 or Reux Caps, Catalog # SC506
(for use with SC490 vials for HotBlock™ SC150, use Ribbed Watch Glasses,
Catalog # SC610 )
2.4 FilterMate™ — Catalog # SC0401 (or appropriate FilterMate) for sample
ltration if necessary
HotBlock™: Operation and Instruction Manual NIOSH Method 7303
* Note: When using the SC150 block with the SC490 digestion vials with a 100mL sample,
double the volume of all reagents and acids added.
3.0 Procedure, Filter Preparation
3.1 Remove lter from cassette holder and fold into quarters taking care not
to lose any sample.
3.2 Place in SC475 digestion vessel and add 1.25mL HCl.
3.3 Cover with plastic watch glass or reux cap. Place in HotBlock™ and heat
at a sample temperature of 95°C for 15 minutes.
Note: If using the watch glass, adjust the HotBlock™ temperature so that a 50mL, 5% acid
solution is heated to 85°C. Laboratory tests have proven that the addition of the ribbed
watch glass will add approximately 10°C to the sample temperature. Refer to page 7 of
the operations manual for instructions on HotBlock™ temperature adjustment.
3.4 Remove samples from the HotBlock™ and cool for 5 minutes.
3.5 Remove watch glass and add 1.25mL HNO3. Replace watch glass and
return to HotBlock™ at sample temperature of 95°C for 15 minutes.
3.6 Remove the sample from the HotBlock™ and cool for at least 5 minutes.
Rinse watch glass into vessel. Discard watch glass.
3.7 Dilute to 25mL nal volume with distilled, deionized type II water.
All QC samples, concentration limitations, elemental lists and reagent specications
are addressed in depth in Method 7303. Safety concerns are also part of the full
method. Follow the instructions listed in Method 7303. These steps should only be
used as a guide to help improve the performance of your HotBlock™.
Sample Preparation for Lead Analysis with the GhostWipe
Revised 05.05
Methodology Note:
The following procedure has been written as an aid for use with the Environmental
Express HotBlock™™ and strictly follows HUD (Housing and Urban Development)
guidelines for Lead in Dust Wipes Appendix A-5.0 which references NIOSH 7082.
1.0 Scope and Application
1.1 This application deals with the preparation of samples using lead wipes,
specically the GhostWipe from Environmental Express. The sample is
heated in the presence of Nitric and Hydrochloric acids to dissolve the
wipe and all lead compounds into solution. Analysis can be performed
by ame atomic absorption (FLAA) or inductively coupled plasma
technique (ICP).
2.0 Apparatus and Materials
2.1 HotBlock™ for Metals Digestions — Model Numbers SC100, SC150*, or
SC154
4.1 Transfer the wipe to the SC475 digestion cup. It is recommended that
the wipe is sent to the sampling area with a cup and wipe so no transfer
steps are required. The wipe must be transported in a rigid-walled
container according to the sampling procedure ASTM E1728.
4.2 Add 10mL DI Water, 2mL HNO3 and 2mL HCl. Cover with a ribbed watch
glass.
4.3 Heat the sample for 45 minutes at 95°C.
Note: When using a watch glass, adjust the set point temperature of the HotBlock™ so
that a 50mL solution is heated to 85°C BEFORE placing the watch glass over the sample.
Laboratory tests have proven that the addition of the ribbed watch glass will add
approximately 10°C to the sample temperature, bringing the sample temperature up to
95°C. Refer to page 7 of the operations manual for instructions on HotBlock™ temperature
adjustment.
4.4 Using the polycarbonate transfer racks, remove samples from the
HotBlock™ and allow them to cool for approximately 10 minutes. Bring
the sample to 50mL volume. If excess amounts of undigested material
remain, lter the sample using a 2.0µm Teon® FilterMate.
4.5 Cap and mix well.
According to AIHA Policies 2001- Revision Nov. 2000, one LCS, one matrix spike
and one duplicate must be run every 20 samples. “The LCS shall be a solid matrix
material with an established concentration obtained from a source independent
of the instrument calibration and traceable to NIST or other similar reference
material. Liquid spikes may not be used for preparing LCSs”. Environmental Express
does offer NIST traceable spiked Ghost Wipes. The catalog number is SC4252. The
range is 175-200µg of Lead per wipe.
HotBlock™: Operation and Instruction Manual Adaptation of EPA Method 3060
Adaptation of EPA Method 3060 for use with the HotBlock™™
and StirBase Systems Revised 05.05
1.0 Scope and Application
1.1 The following procedures have been written as an aid to EPA Method
3060A for use with the Environmental Express HotBlock™ and StirBase.
EPA Method 3060A is an alkaline digestion procedure for extracting
hexavalent chromium [Cr(6+)] from soluble, adsorbed, and precipitated
forms of chromium compounds in soils, sludges, sediments, and similar
waste materials. Use EPA Method 3060A for reference while following the
sample preparation steps outlined below. The quantication of Cr6+ in
Method 3060A digests should be performed using a suitable technique
with appropriate accuracy and precision.
2.0 Apparatus and Materials
2.1 HotBlock™ for sample digestion – Model Numbers SC100, SC154, SC150,
or SC151
2.2 StirBase stirring device – Catalog # SC160
2.3 Polypropylene Digestion Vessels – Catalog # SC475 (or SC490 for use with
the SC150 or SC151 HotBlocks™)
2.4 Ribbed Watch Glasses – Catalog # SC505 (or SC610 for use with the SC150
or SC151 HotBlocks™)
2.5 Reux Caps – Catalog # SC506 (used as an alternative to the SC505 only)
2.6 Stir Bars – Catalog # SC168
2.7 FilterMate – Catalog # SC0407 (or appropriate FilterMate) for sample
ltration if necessary
2.8 FlipMate – Catalog # SC0301 (or appropriate FlipMate) for sample
ltration if necessary
3.0 Procedure, Solid Sample Preparation
3.1 Place 2.5 ±0.10 g of the eld-moist sample into a clean and labeled
digestion vessel. The sample should have been mixed thoroughly before
the aliquot is removed.
3.2 For the specic sample aliquot being spiked, the spike material should
be added directly to the sample aliquot at this point.
3.3 Add 50mL±1mL of digestion solution to each sample using a graduated
cylinder, and also add approximately 400 mg of magnesium chloride
and 0.5mL of 1.0M phosphate buffer. For analytical techniques that can
correct for oxidation/reduction of Cr, the addition of Mg(2+) is optional.
Cover all samples with watch glasses or reux caps.
3.4 Stir the samples continuously (unheated) for at least ve minutes using
the StirBase.
3.5 Heat the samples to 90 - 95ºC, then maintain the samples at 90 - 95ºC for
at least 60 minutes with continuous stirring.
3.6 Gradually cool, with continued agitation, each vessel to room
temperature.
HotBlock™: Operation and Instruction Manual Adaptation of EPA Method 3060
3.7 Filter each sample using a 0.45µm FlipMate (SC0601).
3.8 Adjust the pH of the digestate accordingly to the method being used for
analysis. If a precipitate forms, the sample should be ltered as in 3.7.
3.9 Adjust the sample volume to 100mL with reagent water. Mix well. The
sample digestates are new ready to be analyzed.
All QC samples, concentration limitations, interferences, and reagent specications
are addressed in depth in EPA method 3060A. Safety concerns are also part of
the full method. Follow the instructions listed in EPA method 3060A. These
steps should only be used as a guide to help improve the performance of your
HotBlock™ and StirBase.
HotBlock™: Operation and Instruction Manual 50mL HotBlock™ Accessories and Consumables
Accessories and Consumables for 50mL HotBlocks™
Polycarbonate Transfer Racks
Each sample transfer rack for the HotBlock™ is designed to hold and handle
eighteen SC475 50mL digestion vessels. Racks are made of polycarbonate for heat
and acid resistance and t over the HotBlock™ to facilitate handling of samples
and improve temperature and evaporation uniformity.
Polycarbonate Transfer Rack Each Catalog # SC200
Plastic Storage Racks
24-cup capacity plastic racks neatly hold the SC475 digestion cup. Made of
polypropylene, racks accommodate any tube up to 30mm and are stackable for
efcient storage.
Plastic Storage Rack Each Catalog # SC446
Urethane Foam Storage Racks
Low-cost storage system uses high density urethane foam for semi-permanent
storage of digested samples. Each rack holds 25 capped SC475 digestion cups.
Racks with cups may be stacked to save space. Overall size of rack is 10" x 10" x 1".
Urethane Foam Storage Racks Pack of 5 Catalog # SC700
Temperature Probe
The temperature probe accessory allows the technician to monitor actual sample
temperature in 50mL vessels. The digital thermometer is adjustable in Teon®
holder to allow placement of thermometer tip just above bottom surface of
digestion vessel. Thermometer has LCD display and can be set in either Fahrenheit
or Celsius.
Description Quantity Catalog #
Temperature Probe (thermometer to 150ºC)
Fits SC475 Each SC980
High Range Temperature Probe (thermometer to 250ºC)
Fits cups SC432, SC545, and SC475 Each SC983
Digital Digestion Timer
Use of this digital timer with the HotBlock™ allows the analyst to turn the
HotBlock™ on and off automatically. Digestions that are begun late in the work
day may be completed and cooled while unattended. The timer plugs directly into
standard wall outlet.
HotBlock™: Operation and Instruction Manual 50mL HotBlock™ Accessories and Consumables
Disposable Polypropylene Digestion Vessels
These carefully engineered digestion vessels provide premium performance in your
HotBlock™™ or AutoBlock™. Molded of claried homopolymer polypropylene,
these vessels provide a higher working temperature and greater chemical
resistance than commonly used co-polymer polypropylene.
Each vessel is graduated to 50mL and has a total capacity of 68mL. Molded-in
graduations are extremely accurate, allowing analysts to easily reproduce volumes
to within 0.5%. Using the vessels’ graduations for sample measurement reduces
the need for graduated cylinders and volumetric asks while eliminating sample
transfers. The result is a “one-cup” system that greatly reduces labor and costs
associated with metals preparation.
The 33mm threaded cap is lined with a 0.040" polyethylene-faced foam liner
for a leak-proof seal. Only the metals-free polyethylene liner comes into contact
with the sample. SC475 vessels come in cases of 500 and are packaged to ship as
a standard size UPS package, keeping shipping costs at a minimum. Vessels come
standard with green caps. Vessels with white caps are also available (SC480-W).
• Certied for low metals background
• Compact packaging reduces shipping costs
• Lined closure system ensures leak-free performance with no contamination
• Accurate graduations eliminate the need for graduated cylinders
Specialty Digestion Vessels are also available for a variety of specialty applications.
Visit our web site at www.envexp.com or call us for more information.
Disposable 50mL Digestion Cup Pack of 500 Catalog # SC475
Ribbed Watch Glasses
44mm ribbed watch glasses t all 50mL digestion vessels. Disposable
polypropylene construction is naturally free of contaminants. Ribbed design allows
partial reux during digestion. Watch glass has a center stem to aid in handling.
Ribbed Watch Glasses, 44mm Pack of 1000 Catalog # SC505
Reux Cap
Special cap has a 0.5" hole in its center to allow the addition of reagents while cap
remains in place on the SC475 digestion vessel. The disposable reux cap does not
have to be removed during the digestion process. The cap is not ribbed, but ts
over the top of the SC475 digestion vessel to maintain a centered position. Tests
have shown equal reux between SC506 reux cap and SC505 watch glass.
HotBlock™: Operation and Instruction Manual 100mL HotBlock™ Accessories and Consumables
Accessories and Consumables for 100mL HotBlocks™
Polycarbonate Transfer Racks
This set of two polycarbonate racks was designed to t the SC150 Hot Block (see
photo page 4). A 15-place rack paired with a 10-place rack give you 25 sample
capacity. Racks t the SC150 HotBlock™ securely for safe, accurate digestions.
Polycarbonate Tramsfer Racks for SC150 HotBlock™ Set of 2 Catalog # SC225
Wire Storage Rack for SC490s
Powder-coated wire rack shown above is for storing SC490 cups. Rack has an
18-cup capacity.
Wire Storage Rack Each Catalog # SC7150
Temperature Probe
This temperature probe ts SC490 cups and allows the analyst to monitor actual
temperature in a 100mL sample. The digital thermometer is adjustable in a Teon®
holder to allow placement of the thermometer tip in desired position in cup.
Thermometer has LCD display and may be read in either Fahrenheit or Celsius.
Temperature Probe (thermometer to 150ºC) for SC490 Each Catalog # SC981
Disposable Polypropylene Digestion Vessels for use with SC150 HotBlock™
This 125mL, wide-mouth digestion vessel is supplied with a 48mm screw cap. The
cap is lined with coextruded polyethylene foam for leak-proof performance. Made
of homopolymer polypropylene, the SC490 digestion vessel is designed for use in
the SC150 HotBlock™. Vessels are graduated at 10mL, 25mL and 50mL, 75mL and
100mL.
Disposable 125mL Digestion Cup Pack of 225 Catalog # SC490
Ribbed Watch Glass
52mm ribbed watch glass is designed for use with SC490, 125mL digestion vessel.
The disposable polypropylene watch glass has a stem in its center to aid in
handling.
Ribbed Watch Glass, 52mm Pack of 500 Catalog # SC610
HotBlock™: Operation and Instruction Manual 15mL HotBlock™ Accessories and Consumables
Accessories and Consumables for 15mL HotBlocks™
Polycarbonate Transfer Rack
These polycarbonate racks t the SC196 HotBlock™. The racks hold our SC415
digestion vessels securely, assisting the analyst in performing safe, accurate
digestions. Three 32-place racks give you 96-sample capacity.
Polycarbonate Tramsfer Racks for SC196 HotBlock™ Set of 3 Catalog # SC296
Temperature Probe
This temperature probe ts SC415 cups and allows the analyst to monitor actual
temperature in a 15mL sample. The digital thermometer is adjustable in a Teon®
holder to allow placement of the thermometer tip in desired position in cup.
Thermometer has LCD display and may be read in either Fahrenheit or Celsius.
Temperature Probe (thermometer to 150ºC) for SC415 Each Catalog # SC985
Polypropylene Digestion Vessels for use with SC196 HotBlock™
These screw cap digestion vessels for 15mL samples, have a total volume of 18mL.
Molded-in graduations are in 5mL increments up to 15mL.
Disposable 15mL Digestion Cups with Caps Pack of 1,000 Catalog # SC415
Ribbed Watch Glass
19.5mm ribbed watch glass is designed for use with SC415, 15mL digestion
vessel. The disposable polypropylene watch glass has a stem in its center to aid in
handling.
Ribbed Watch Glass, 19.5mm Pack of 1000 Catalog # SC417
HotBlock™: Operation and Instruction Manual AirLite HEPA-Filtered Enclosure
AirLite™ HEPA-Filtered Enclosure
SC801:
Size: 25.5"h X 32"w X 24"d
Door Opening Size: 32"w x 17"h
Materials: Acrylic walls and door,
Kydex® top and bottom plates
Fan: 117CFM, PVC Coated
Electrical: 120VAC, 1 amp,
3-Prong Plug
This bench-top enclosure for the
HotBlock™ provides HEPA-ltered air
for cleaner, more uniform digestions.
The AirLite uses a quiet 117 CFM
internal PVC coated fan that is easily
accessible by the technician. The
cabinet of the AirLite is constructed
from PET-G for exceptional clarity. The
door of the hood opens a full 180°
to allow unrestricted access to the
chamber. The body of the enclosure
rests in a thermoformed Kydex® bottom that traps spills and can easily be removed
for cleaning.
The AirLite is designed to enhance the use of the HotBlock™, and can reduce
sample contamination during the digestion process. The HEPA lter can be
changed in minutes. The AirLite can be used anywhere a clean environment is
required. It can be exhausted with standard 4" exible duct to the outside or into
existing ductwork. Spare 117CFM fan included.
Description Catalog #
AirLite Enclosure SC801
Replacement HEPA Filter SC840
Replacement Exhaust Fan SC830
FlipMate™ Filtration Assemblies
The FlipMate offers a quick and effective method of ltering up to 125mL of
sample. The device accepts two threaded digestion cups (cups are sold separately)
- one containing the sample to be ltered and an empty cup to receive the ltrate.
The assembly is then “ipped” over and vacuum is applied. The sample is pulled
through the lter into the empty cup. The ltered sample can then be capped and
stored for later analysis. FlipMate is available with a variety of lters installed.
Description Sample Size Filter Material Porosity Quantity Catalog #
FlipMate 50 50mL PES Filter with Prelter 0.45µm 100 SC0601
FlipMate 50 50mL Glass Fiber 1.5µm 100 SC0602
FlipMate 100 100mL PES Filter with Prelter 0.45µm 50 SC0301
FlipMate 100 100mL Glass Fiber 1.5µm 50 SC0302
FlipMate 100 100mL PES with Prelter 0.2µm 50 SC0303
HotBlock™: Operation and Instruction Manual Sample Filtration Supplies
FilterMate™ Filtration Systems
Dirty or particulate-laden samples may require that the digested sample be ltered
prior to analysis. This is especially true with sludge, wastewater and soil samples.
FilterMate Systems provide a convenient and inexpensive method of ltration
that requires no transfer steps. The FilterMate ts directly into the digestion cup
and is pushed through the liquid being ltered to the bottom of the cup with a
detachable plunger. After the ltration is complete, the plunger is removed and
discarded while the lter assembly remains in the bottom of the cup.
This arrangement does not interfere with storage or use of the cup in an
autosampler. FilterMates are available with different choices of lters.
Membrane Type
Membrane Characteristics Sample Lot
Applications Size Porosity Certied Quantity Catalog #
Membrane Type: Teon-faced Polypropylene
Characteristics: Highest dirt capacity
Applications: Most laboratory applications 50mL 2µm No 100 SC0401
Membrane Type: Teon w/additional Teon Prelter
Characteristics: Not for particulate-laden samples
Applications: Dissolved metals only 50mL 0.45µm No 100 SC0407
Membrane Type: Teon
Characteristics: High dirt capacity
Applications: Suitable for trace level analysis 50mL 2µm Yes 100 SC0408
Membrane Type: Polyethylene Frit
Characteristics: For use with SC415 15mL 2µm No 100 SC0421
Cup Holder for FilterMate Filtrations
This machined HDPE xture adds stability when performing FilterMate
ltrations. Digestion cup is placed into center hole of FilterMate Holder
prior to beginning ltration. The holder secures the digestion cup as the
FilterMate is depressed through the liquid sample. SC0510
HotBlock™: Operation and Instruction Manual GhostWipes for Surface Sampling
Take a sample with
1.
GhostWipe.
Digest as usual in acid
2.
solution.
GhostWipe™ For Surface Sampling,
Including Lead and Beryllium Testing
This revolutionary product eases sample preparation and analysis of surface lead
or surface dust. The GhostWipe is a sturdy wiping material moistened with DI water
that holds together even on the roughest wiping surfaces. In the lab, the GhostWipe
readily and completely dissolves during the digestion process. This feature provides
more complete dispersion of analytes and more uniform recoveries.
In addition, there is no
messy brous material
to clog sample uptake
capillary or nebulizer.
GhostWipes meet all
ASTM Designation E
1792 specications for
sampling materials
for lead in surface
dust. Wipes are 15cm x
15cm and are supplied
in individual sealed
packets.
Ghost wipes totally
3.
dissolve during digestion
for easy analysis.
Description Quantity Catalog #
GhostWipes 500 SC4210
GhostWipes 1,000 SC4250
Dust Sampling Template, 10cm x 10cm 250 C1010
Dust Sampling Template, 12" x 12" 100 C1012
Lead Wipe Standard in Screw-Cap, Rigid-walled Cup (SC475)
The lead wipe standard is a Ghost Wipe spiked with in-situ dust. Using an NISTtraceable solid sample, known amounts of lead are placed on a wipe and each
sample is individually labeled with the exact amount of lead the wipe contains.
Each spike sample contains 175-250 micrograms of lead per wipe and comes with
a certicate of analysis showing the Sample ID# and the starting concentration of
the dust.
Lead Wipe Standard with Cup Box of 10 Catalog # SC4252