Enthermics EC770s Operation Manual

#OCM EC770s 5/01
PRINTED IN U. S. A .
OPERATION AND CARE MANUAL
Corporate Headquarters:
W164 N9221 Water Street ●P.O. Box 450 Menomonee Falls, Wisconsin 53052-0450
PHONE: 262.251.8356 800.TO.B.WARM U . S . A ./CANADA FAX: 262.251.7067
www.enthermics.com 800.862.9276
U . S . A ./CANADA 800.329.8744
BLANKET
WARMING
CABINET
EC770
s
Transport and Storage Environmental Conditions (not to exceed 15 days)
Ambient temperature range of -40°F (-40°C) to +158°F (+70°C) Relative humidity range of 10% to 100%, including condensation Atmospheric pressure range of 50KPa to 106KPa
The Enthermics Medical Systems Warming Cabinet has been thoroughly tested, checked for calibration, and inspected to insure only the highest quality. When you receive your unit, check for any possible
shipping damage and report it at once to the delivering carrier.
The cabinet, complete with unattached items and accessories, may be delivered in one or more packages. Save all the information and instructions packed inside the cabinet.
Complete and return the warranty card to the factory as soon as possible to assure prompt service in the event of a warranty parts and labor claim. All claims for warranty must include
the full model number and serial number of the cabinet.
After unpacking the unit, allow it to remain under stable atmospheric conditions for several hours before connecting to the power source. Extreme changes in temperatures and/or humidity during storage and/or transportation could have allowed condensation to form in the unit.
NOTE: Pay particular attention throughout this manual to
the warning and caution symbol which means: Attention, Consult Operation
Manual before proceeding.
ELECTRICAL INFORMATION
The power specifications are located on the unit identification nameplate. This nameplete is permanently attached to the unit and must be located to determine and verify power requirements.
POWER REQUIREMENTS
120 V.A.C. — 60 Hz, 1 ph 1000 Watts, 8.0 Amps
The single compartment warming cabinet is constructed with 22 gauge stainless steel exterior casing and door, with handle and hinges designed to withstand heavy usage. The cabinet is heated by the Halo Heat@ system incorporating low watt density cable mounted around the compartment walls. This system provides evenly distributed heat without the need of a heat-circulating fan. The cabinet is controlled by an adjustable, electronic thermostat, which includes a heat indicator L.E.D., temperature set button, temperature increase and decrease button, and digital display to monitor compartment temperature. The thermostat has a temperature range of 90°F (32°C) minimum to 200°F (93°C) maximum. The chamber is equipped with one (1) white, epoxy­coated blanket support assembly with shelf, and furnished with one (1) set of 5" (127mm) heavy-duty casters — two rigid and two swivel with brake.
PREPARATION
Before operating the cabinet, clean both the interior and exterior of the unit with a damp cloth and mild soap solution. Wipe with an appropriate disinfectant. Clean and install the cabinet shelf assembly. Do not install or operate the blanket warmer in close proximity of flammable anesthetics.
BLANKET WARMING CABINET
TRANSPORT and STORAGE GENERAL INFORMATION
NEMA 5-15P 15A - 125V Plug Hospital Grade
1
Grounding reliability can only be achieved when equipment is connected to an equivalent receptacle marked "Hospital Grade."
Protective Earth Ground Symbol
Enthermics Medical Systems #OCM EC770s 5/01
ENSURE POWER SOURCE
MATCHES VOLTAGE STAMPED
ON UNIT NAMEPLATE.
Hazardous
Voltage Present
At no time should the inside or outside of the cabinet be washed down by flooding with water or liquid solution. NEVER STEAM CLEAN. Severe damage or electrical hazard could result.
Explosion risk if used in the presence of flammable anesthetics.
Danger!
UNPACKING and SET-UP
Medical Equipment classified by Underwriters Laboratories with Respect to Electrical Shock, Fire and Mechanical Hazards only, in Accordance with UL 544 and CAN/CSA C22.2 No. 601.1.
IP50
MADE IN USA
MODEL
SERIAL#
WATTS AMPS
VOLTS 1 PH Hz
LISTED
MEDICAL EQUIPMENT
C
LISTED
US
MEDICAL EQUIPMENT
1P50
SET
prg
!
ERROR CODE
INDICATOR L.E.D.
HEAT
INDICATOR L.E.D.
CHAMBER
TEMPERATURE
L.E.D.
TEMPERATURE
SET
BUTTON
POWER ON/OFF
ROCKER SWITCH
L.E.D.1
INCREASE /DECREASE
BUTTON
200
THERMOSTAT
"I" = "ON"
"" = "OFF"
OPERATIONAL PROCEDURES
• L.E.D., Light Emitting Diode, referred to is an
electronic device providing illumination.
1. TURN POWER SWITCH ON.
The heat indicator L.E.D. will illuminate and remain
lit while the unit is calling for heat. The digital display will indicate air temperature of the heated compartment. The heat indicator L.E.D. will go out when the air temperature inside the unit reaches the temperature set-point on the electronic thermostat.
2. VIEW OR CHANGE SET-POINT.
A. Push the SET button once. The set-point value
(current set-point holding temperature) will be displayed for 5 seconds.
NOTE: Factory default
is 90°F (32°C).
B. The set-point value can be changed by pressing
and holding the SET button for at least 4 seconds. The programming mode becomes active and the LED1 indicator will blink.
Press
and hold
the UP or DOWN arrow button to change the value shown in the display. The value can be stored either by pressing the SET button or by waiting the exit time-out to expire (approximately 15 seconds). The new set-point value will flash three times to confirm. This value will be retained until it is changed again. The minimum set-point temperature is 90°F (32°C) while the maximum set-point temperature is 200°F (93°C). Typical temperature settings are around 160°F (71°C).
3. LOAD THE CABINET WITH BLANKETS.
Check that the epoxy-coated blanket support
assembly and shelf are in place. The warming cabinet is designed to elevate blanket temperatures to a level which will increase patient comfort. Do not fill cabinet any closer than 2" from ceiling of warming chamber (If sensor becomes blocked, overheating could result.) A full load of blankets will take several hours to reach optimum temperature. Make certain the cabinet door is securely closed after initial loading and following each blanket removal.
DO NOT BLOCK SENSOR BY
OVERLOADING CABINET WITH
BLANKETS. OVERHEATING
COULD OCCUR.
THERMOSTAT
and INDICATOR LIGHT SEQUENCE
Whenever the thermostat is turned “ON,” the heat indicator L.E.D. will indicate the power “ON/OFF” condition of the heating cable, and consequently, the cycling of the cabinet as it maintains the set-point cavity temperature. If the L.E.D. does not illuminate after normal start-up, the main power source and thermostat must be checked. If the warming cabinet does not hold the temperature as set, the calibration of the thermostat must be checked. If the warming cabinet fails to heat with the L.E.D. illuminated, or heats continuously with the L.E.D. of thermostat “OFF,” unplug and call for service.
THERMOSTAT ACCURACY
The electronic thermostat is a precise instrument and is designed to offer trouble-free service. If you suspect the temperature inside the warming cabinet does not match the temperature indicated on the digital display, after stabilizing, follow the instructions listed below.
1. Check to make certain the power meets the requirements of the appliance, voltage and current rating as indicated on identification tag.
2. Verify the temperature inside the warming cabinet with a quality thermal indicator.
A. With the exception of the metal shelves,
completely empty the warming cabinet.
B. Make certain the sensor, located inside the
warming cabinet at the left side of the ceiling, is completely clean.
C. Suspend the thermal indicator in the center of
the warming cabinet.
D. Allow the temperature, set on the electronic
thermostat, to stabilize for a minimum of one hour before comparing the digital display with the reading on the thermal indicator.
DO NOT OPEN THE CABINET DOOR(S) DURING THE
TEMPERATURE STABILIZATION PERIOD
.
If the reading on the thermal indicator does not match the digital display within 10°F (6°C), contact Enthermics Medical Systems for appropriate action.
2
CONTROLLER
Enthermics Medical Systems #OCM EC770s 5/01
At no time should the inside or outside of the cabinet be washed down by flooding with water or liquid solution. NEVER STEAM CLEAN. Severe damage or electrical hazard could result.
OUTSIDE DIMENSIONS
TOP VIEW WITH
27-13/16" (706mm)
C
L
OPTIONAL BASE E5086
52-5/16" (1329mm)
24-9/16" (624mm)
33-5/16" (846mm)
CARE and CLEANING
The cleanliness and appearance of this equipment will contribute considerably to its operating efficiency. Make certain the cabinet and door gaskets are kept free of any debris that may accumulate. Good equipment that is kept clean works better and lasts longer.
CLEAN THE UNIT REGULARLY.
1
. Disconnect the cabinet from the power source.
2. Remove all detachable items such as metal shelves and side guards. Clean these items separately.
3. Clean the interior metal surfaces of the cabinet with a damp cloth and any mild commercial detergent. Avoid the use of abrasive cleaning compounds.
4. Clean the exterior of the cabinet with a cleaner recommended for stainless steel. Spray the cleaner on a clean cloth and wipe with the grain.
5. If equipped with window doors, clean the glass with a standard commercial glass cleaner.
NOTE: Do not use abrasive cleansers on the exterior
of the cabinet. Use a non-abrasive cleaner or distilled white vinegar to remove water spots or other stains.
DISCONNECT POWER BEFORE CLEANING.
3
Enthermics Medical Systems #OCM EC770s 5/01
(Listed as Ordinary Equipment.)
FRONT VIEW
22" (559mm)
35" (889mm)
41-1/4" (1048mm)
Loading...
+ 7 hidden pages