Enthermics EC770L, Designer Series, DC400, DC150 Operator's Manual

Operator’s Manual
Fluid Warmer
EC770L
MN-39935
REV.01 3/19
EN

Manufacturer’s Information

Copyright © Copyright 3/19 by Enthermics Medical Systems.
All rights reserved.
This manual or any portion thereof may not be reproduced or used in any manner whatsoever without the express written permission of Enthermics Medical Systems.
Trademarks All trademarks referenced in this documentation are the property of their
respective owners.
Manufacturer
Enthermics Medical Systems
An ISO 13485:2016 certified company
W164 N9221 Water Street
Menomonee Falls, WI 53051, USA
Original instructions The content in this manual is written in American English.
Fluid Warmer Operator’s Manual MN-39935 Rev 1 3/19

FOREWORD

Thank you for your Purchase!

This warmer has been thoroughly tested and inspected to ensure only the highest quality is provided. We supply the most durable, convenient, efficient and safe warming equipment on the market. All warmers are manufactured and fully inspected in the USA with a commitment to quality.

Register Your Warmer

Register Register your Enthermics appliance online. Registering your appliance ensures
prompt service in the event of a warranty claim. You will also receive direct notifications of software updates and additional product information.
Your personal information will not be shared with any other company.
www.enthermics.com/warranty-registration

Enthermics 24/7 Emergency Repair Service

Call Call 800-558-8744 to reach our 24-hour emergency service call center for
immediate access to local authorized service agencies outside standard business hours. The emergency service access is provided exclusively for Enthermics Medical Systems equipment and is available throughout the United States through Enthermics toll free number.
Availability Emergency service access is available seven days a week, including holidays.
Fluid Warmer Operator’s Manual MN-39935 Rev 1 3/19
OREWORD
FOREWORD
This page intentionally left blank
Fluid Warmer Operator’s Manual MN-39935 Rev 1 3/19
TABLE OF CONTENTS
Manufacturer’s Information . . . . . . . . . . . . . . . . . . . . . . . . 2
Foreword 3
Thank you for your Purchase! . . . . . . . . . . . . . . . . . . . . . . . 3
Register Your Warmer . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Enthermics 24/7 Emergency Repair Service . . . . . . . . . . . . . . . 3

Table of Contents 5

Safety 7
The Meaning of Signal Words . . . . . . . . . . . . . . . . . . . . . . . 7
Safety Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Labels 11
Label Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Specifications 13
Specification Information . . . . . . . . . . . . . . . . . . . . . . . . . 13
Dimension Drawing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Installation 15
How to Unpack the Warmer . . . . . . . . . . . . . . . . . . . . . . . . 15
How to Install the Warmer . . . . . . . . . . . . . . . . . . . . . . . . . 16
Operation 19
Preparing the Warmer for First Use. . . . . . . . . . . . . . . . . . . . 19
How to Turn On and Turn Off the Warmer . . . . . . . . . . . . . . . . 20
How to Operate the Fluid Warmer . . . . . . . . . . . . . . . . . . . . 21
How to Change the Temperature Scale. . . . . . . . . . . . . . . . . . 23
How to Change the Sound Settings . . . . . . . . . . . . . . . . . . . . 24
How to Lock and Unlock the Controller. . . . . . . . . . . . . . . . . . 25
Maintenance 27
Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
How to Clean the Warmer . . . . . . . . . . . . . . . . . . . . . . . . . 28
Troubleshooting 31
What to do if the Warmer Malfunctions . . . . . . . . . . . . . . . . . 31
How to Manually Reset the Warmer . . . . . . . . . . . . . . . . . . . 32
What to do if the Over Temperature Indicator Light Illuminates . . . 33
Error Codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
ONTENTS
C
OF
Warranty 39
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Fluid Warmer Operator’s Manual MN-39935 Rev 1 3/19
ABLE
T
TABLE OF CONTENTS
This page intentionally left blank
Fluid Warmer Operator’s Manual MN-39935 Rev 1 3/19

The Meaning of Signal Words

This manual contains signal words where needed. These signal words must be obeyed to reduce the risk of death, personal injury, or equipment damage. The meaning of these signal words is explained below.
DANGER
Danger indicates a hazardous situation which, if not avoided, will result in serious injury or death.
WARNING
Warning indicates a hazardous situation which, if not avoided, could result in serious injury or death.

SAFETY

CAUTION
Caution indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
NOTICE
NOTE: Note indicates additional information that is important to a
concept or procedure.
Notice indicates a situation which, if not avoided, could result in property damage.
Fluid Warmer Operator’s Manual MN-39935 Rev 1 3/19
AFETY
SAFETY

Safety Precautions

Before you begin Read and understand all instructions in this manual.
Electrical precautions Obey these electrical precautions when using the warmer:
Connect the warmer to a properly grounded outlet. Do not use the warmer if it is
not properly grounded. Consult an electrician if there is any doubt that the outlet used is properly grounded.
Keep the cord away from hot surfaces.
Do not attempt to service the warmer or its cord and plug, when plugged in.
Do not operate the warmer if it has a damaged cord or plug.Do not immerse the cord or plug in water.
Do not let the cord hang over the edge of a table or counter.
Do not use an extension cord.
Usage precautions Obey these usage precautions when using the warmer:
Only use this warmer for its intended use of warming.
Do not use this warmer for warming blood or blood products.
Do not cover or block any of the openings of this warmer.
Do not use this warmer in a wet location.
Only clean the warmer when the power cord is unplugged.
Do not use corrosive chemicals when cleaning the warmer.
Do not use the warmer cavity for storage.
Do not remove the top cover or exterior side panels. There are no user-
serviceable components inside.
Operator training All personnel using the warmer must have proper operator training. Before
using the warmer:
Read and understand the operating instructions contained in all the
documentation delivered with the warmer.
Know the location and proper use of all controls.
Keep this manual and all supplied instructions, diagrams, schematics, parts lists,
notices, and labels with the warmer if the warmer is sold or moved to another location.
Fluid Warmer Operator’s Manual MN-39935 Rev 1 3/19
SAFETY
Operator qualifications
Only trained personnel with the following operator qualifications are permitted to use the warmer:
Have received proper instruction on how to use the warmer.
Are familiar with the purpose, limitations, and associated hazards of the warmer.
The warmer must not be used by:
People impaired by drugs or alcohol.
Condition of warmer Only use the warmer when:
All controls operate correctly.
The warmer is installed correctly.
The warmer is clean.
The warmer labels are legible.
Servicing the warmer Obey precautions in the manual, on tags, and on labels attached to or shipped
with the warmer.
Only trained personnel are permitted to service or repair the warmer. Repairs
that are not performed by a trained technician, or the use of non-factory parts, will void the warranty and relieve all liability.
Any troubleshooting guides and components views included with this manual
are for reference only and are intended for use by qualified and trained service technicians.
To prevent serious injury, death or property damage, have the warmer inspected
and serviced at least every twelve (12) months by a trained technician.
Contact Enthermics for the authorized service partner in your area.
Fluid Warmer Operator’s Manual MN-39935 Rev 1 3/19
SAFETY
This page intentionally left blank
10
Fluid Warmer Operator’s Manual MN-39935 Rev 1 3/19

Label Locations

EC-PHD-006516
a
c
d
b

LABELS

ABELS
Fluid Warmer Operator’s Manual MN-39935 Rev 1 3/19
11
LABELS
Security Seal
Security Seal
Security Seal
Security Seal
Security Seal
Security Seal
Security Seal
Security Seal
Security seal
Electrical warning label
Enthermics contact information
Quality inspected
IMPORTANT
Do not load baskets beyond recommended capacity. Overloading may cause uneven temperature of product and damage to basket and basket rail supports.
12
Fluid Warmer Operator’s Manual MN-39935 Rev 1 3/19

Specification Information

E471516
Protective Earth Ground Symbol
Model EC770L

SPECIFICATIONS

Weight
Storage cavity capacity
Temperature range
Clearance requirements
Transportation and storage conditions
Operating conditions The warmer must acclimate to the room temperature it will be placed—24 hours
Ambient temperature range of -40°F to +159°F (-40°C to +70°C).
Relative humidity range of 10% to 95%, non-condensing.
Atmospheric pressure range of 7.25 psi to 15.37 psi (50kPa to 106kPa).
is recommended.
The recommended environmental temperature range is 60°F to 90°F
(15°C to 32°C).
The recommended relative humidity is above 20%, non-condensing.
Irrigation fluids (IRR): 98°F to 150°F (37°C to 66°C)
72 1-liter bags, 24 bags per basket
54 1-liter bottles, 18 1-liter bottles per basket
Injection fluids (INJ): 98°F to 104°F (37°C to 40°C)
Net: 173 lb (78 kg)
Ship: 258 lb (117 kg)
3” (76mm) from rear
1” (25mm) from top
1” (25mm) from sides
Standards for electrical equipment
Fluid Warmer Operator’s Manual MN-39935 Rev 1 3/19
Medical equipment classified by Underwriters Laboratories with respect
to electrical shock, fire, and mechanical hazards only, in accordance with UL 60601-1 and CAN/CSA C22.2 No. 601.1.
Grounding reliability can only be achieved when the appliance is
connected to an equivalent receptacle marked “Hospital Grade.”
IP-XO ordinary
PECIFICATIONS
13
SPECIFICATIONS
EC-DIM-006501
11.0" (280mm)
52.8" (1341mm)
35.0" (890mm)
28.0" (731mm)
11.0"
(280mm)
2.20" (55.6mm)
OFF
ON
30.10" (764mm)
25.10" (638mm)
5.5" (140mm)
18.9" (480mm)
Cavity
25.2" (639mm)
26.9" (684mm)
Cavity
35.1" (892mm)
40.8" (1036mm)
54.0" (1372mm)
33.876" (860.5mm)
22.3" (566mm)
22. 1" (561mm)
7.7" (195mm)
17.1" (434mm)
CORD LENGTH (est.)
8' (2,438mm)
26.6" (676mm)
Cavity

Dimension Drawing

EC770L
14
Fluid Warmer Operator’s Manual MN-39935 Rev 1 3/19
Loading...