Enthermics EC1850BL Owner's Manual

Enthermics Medical Systems
An ISO 13485:2003 certified company
W164 N9221 Water St | Menomonee Falls WI 53051
Tel 262-251-8356 | 800-TO-B-WARM
www.enthermics.com
Printed in the U.S.A. Specifications are subject to change without notice.
Made in the U.S.A.
Owner’s Manual
MN-39362 • Rev 0 • 04/18
Titan Series Combination
Blanket/Fluid Warmer
EC1850BL
MN-39362 • Rev 0 • 04/18 • Titan Series EC1850BL Combination Blanket/Fluid Warmer
Table of Contents
Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Transportation Damage and Claims . . . . . . . . . . . . . . . 3
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Safety Procedures . . . . . . . . . . . . . . . . . . . . . . . . . 5
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Electrical Information . . . . . . . . . . . . . . . . . . . . . . 7
Operation Instructions
Preparation Before Operating the Warmer . . . . . . . . . .10
Using the Blanket Controller . . . . . . . . . . . . . . . . . .11
Operating the Blanket Chamber . . . . . . . . . . . . . . . .12
Using the Fluid Controller . . . . . . . . . . . . . . . . . . .13
Operating the Fluid Chamber . . . . . . . . . . . . . . . . .15
Cleaning and Preventative Maintenance . . . . . . . . . . . .16
Preventative Maintenance Checklist. . . . . . . . . . . . . .17
Dimension Drawings . . . . . . . . . . . . . . . . . . . . . . .18
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . .20
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Wire Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Authorized Representative:
MDSS GmbH Schi graben 41 30175 Hannover Germany
REPEC
3
MN-39362 • Rev 0 • 04/18 • Titan Series EC1850BL Combination Blanket/Fluid Warmer
Delivery
Receipt of Appliance
The appliance has been thoroughly tested and inspected to ensure only the highest quality appliance is provided. Upon receipt, inspect for any possible shipping damage and report it at once to the delivering carrier. See Transportation Damage and Claims section.
This appliance, complete with unattached items and accessories, may be delivered in one or more packages. Confi rm that all standard items and options have been received with each appliance as ordered. Save all the information packed with the appliance.
Register the appliance online to assure prompt service in the event of a warranty parts and labor claim.
http://www.enthermics.com/warranty-registration
The serial number is required for all inquiries.
Always include both model and serial number(s) in any correspondence regarding the appliance.
Model: ______________________________________________
Serial number: ______________________________________________
Purchased from: ______________________________________________
Date installed: ____________________ Voltage: _____________
Environmental Conditions
Transport and storage environmental conditions (not to exceed 15 days)
Ambient temperature range of -40°C to +70°C (-40°F to +159°F) Relative humidity range of 10% to 95%, non-condensing Atmospheric pressure range of 50kPa to 106kPa
Operational environmental conditions
The appliance must acclimate to the room temperature in the environment it will be placed—24 hours is recommended. The recommended environmental temperature range is 15°C to 32°C (60°F to 90°F). The recommended relative humidity is above 20%, non-condensing.
Indicates that the package contents should not be used if the package has been damaged or opened.
All Enthermics Medical Systems appliances are sold Free on Board (F.O.B.) shipping point, and when accepted by the carrier, such shipments become the property of the consignee.
Should damage occur in shipment, do not put the appliance into service until the damage has been inspected by an authorized service provider.
Should damage occur in shipment, it is a matter between the carrier and the consignee. In such cases, the carrier is assumed to be responsible for the safe delivery of the merchandise, unless negligence can be established on the part of the shipper.
1. Conduct an immediate inspection while the appliance is still in the truck or immediately a er it is moved to the receiving area. Do not wait until a er the appliance is moved to a storage area.
2. Do not sign a delivery receipt or a freight bill until a proper count has been made and inspection of all appliances are received.
3. Note all damage to packages directly on the carrier’s
delivery receipt.
4. Have the driver sign the delivery receipt. If the driverrefuses to sign, make a notation of this refusal on the receipt.
5. If the driver refuses to allow inspection, write the following on the delivery receipt: Driver refuses to
allow inspection of containers for visible damage.
6. Contact the carrier’s offi ce immediately upon fi nding damage and request an inspection. Mail a written confi rmation to the carrier’s offi ce with the time, date, and the person called.
7. Save any packages and packing material for further inspection by the carrier.
8. Promptly fi le a written claim with the carrier and attach copies of all supporting paperwork.
Enthermics will continue our policy of assisting our customers in collecting claims which have been properly fi led and actively pursued. Enthermics cannot, however, fi le any damage claims, assume the responsibility of any claims, or accept deductions in payment for such claims.
Transportation Damage and Claims
4
MN-39362 • Rev 0 • 04/18 • Titan Series EC1850BL Combination Blanket/Fluid Warmer
Unpacking
1. Remove the appliance from the carton or crate.
NOTE: Do not discard the carton and other
packaging material until the appliance has been inspected for hidden damage and tested for proper operation.
Do not discard this manual. This manual
is considered to be part of the appliance and is to be provided to the owner or manager of the business or to the person responsible for training operators. Additional manuals are available from the service department.
2. Read all instructions in this manual carefully before initiating the installation of this appliance, using the appliance or performing routine maintenance.
Following procedures other than those indicated in this
manual to use and clean the appliance is considered inappropriate and may cause damage, injury or fatal accidents, in addition to invalidating the warranty and relieving the manufacturer of all liability.
3. Before connecting appliance to electrical power
a) Remove all protective plastic fi lm and packaging
material from the outside of the appliance
b) Remove all packaging material from the inside of
the appliance.
c) Remove packaging material from accessories and
store accessories in a convenient place for future use.
4. Slowly roll the appliance across high thresholds.
CAUTION
The door may swing during transport. Only transport the appliance when the door is closed and secure.
CAUTION
Appliance and accessories may be heavy. To prevent serious injury, always use a su icient number of trained and experienced workers when moving or leveling appliance and handling accessories.
CAUTION
To prevent injury or property damage: Always apply both caster brakes on mobile carts, appliances, or accessories when stationary. Appliances on casters can move or roll on uneven floors.
Attention:
Due to the size and weight of the warming cabinet, additional moving equipment (fork li, dolly, etc.) will be needed to li and move the warming cabinet and packaging materials. Make appropriate plans when moving the warming cabinet.
Dimensions
(HxWxD)
With casters: 74.8" x 30.0" x 28.0" (1900mm x 762mm x 712mm)
Dimensions (HxWxD)
With casters: 74.8" x 30.0" x 28.0" (1900mm x 762mm x 712mm)
Capacity
Fluid compartment: 30 1-liter bottles or
24 1-liter bags
Blanket compartment: 30-40 blankets (16.4 
3
)
Weight**
Net: 457 lbs (207 kg) Ship: 568 lbs (258 kg)
Shipping dimensions: 38" L x 38" W x 90" H
** Domestic ground shipping information. Contact the manufacturer for export weight and dimensions.
5
MN-39362 • Rev 0 • 04/18 • Titan Series EC1850BL Combination Blanket/Fluid Warmer
Safety Procedures
DANGER
Indicates a hazardous situation that, if not avoided, will result in death or serious injury.
The blanket chamber is intended for warming dry,
cotton blankets or towels only. The blanket chamber is designed to elevate blanket temperatures to a level which will increase patient comfort.
The fl uid chamber is only intended for warming
medical solutions for irrigation and injection prior to use. Refer to the labeling of the manufacturer of the products to be warmed regarding the recommended
temperature and the duration of warming. No other use
for this appliance is authorized or recommended.
This warmer is intended for use in commercial
establishments where all operators are familiar with the purpose, limitations, and associated hazards of this appliance. The warmer can be used wherever there is appropriate space and electrical source including patient support areas, ER, ICU, PACU, surgical suites, patient rooms, and nursing stations. Do not use the warmer in the presence of fl ammable anesthetic mixtures (with air, oxygen, or nitrous oxide).
Operating instructions and warnings must be read and
understood by all operators and users.
Any troubleshooting guides, component views, and
parts lists included in this manual are for general reference only and are intended for use by qualifi ed and trained technicians.
This manual should be considered a permanent
part of this appliance. This manual and all supplied instructions, diagrams, schematics, parts lists, notices, and labels must remain with the appliance if the item is
sold or moved to another location.
NOTICE: Indicates information considered important,
but not hazard-related (e.g., messages relating to property damage).
Knowledge of proper procedures is essential to the
safe operation of electrically and/or gas energized equipment. The following signal words and symbols may be used throughout this manual.
Indicates that referral to operating instructions is a mandatory action. If not followed the operator could su er personalinjury.
Indicates that referral to operating instructions is recommended to understand operation of equipment.
WARNING
Indicates a hazardous situation that, if not avoided, could result in death or serious injury.
CAUTION
Indicates a hazardous situation that, if not avoided, could result in minor or moderate injury.
NOTICE: For equipment delivered for use in any
location regulated by the following directive: 2012/95/EC WEEE
Do not dispose of electrical or electronic
equipment with other municipal waste.
6
MN-39362 • Rev 0 • 04/18 • Titan Series EC1850BL Combination Blanket/Fluid Warmer
Installation
WARNING
To prevent serious personal injury, death, or property damage:
Do not steam clean, hose down or flood the interior or exterior with water or liquid solution of any kind. Do not use water jet toclean. Failure to observe this precaution will void the warranty.
DANGER
To prevent serious personal injury, death, or property damage:
Do not use this warming appliance in the presence of flammable anesthetic mixtures (with air or with oxygen or nitrous oxide).
Not category AP or APG equipment
Specifications:
• Dual-chamber warming appliance
• Stainless steel exterior and interior
• Doors are fully gasketed with a magnetic closure
• WarmSafe™ heating system incorporates a multiple zone warming technology (Patent No: US 8,217,316; US 8,581,152) that heats where and when it is needed. All chamber surface temperatures are monitored, providing an effi cient balance of heat, low energy consumption and minimal heat loss.
• Casters are standard
• Fluid chamber capacity: 30 1-liter bottles or 24 1-liter bags
• Blanket chamber capacity: 30 blankets (16.4 
3
)
Control:
• Adjustable temperature range » Blanket chamber has an
adjustable temperature range of 32°C - 71°C (90°F - 160°F)
» Fluid chamber can be
programmed to warm either irrigation fl uids (IRR) or injection fl uids (INJ), with separate temperature ranges provided depending on the choice selected:
- IRR temperature range:
32°C to 66°C (90°F to 150°F),
- INJ temperature range:
32°C to 40°C (90°F to 104°F).
• Accurate temperature
- IRR +0/-1.67°C (-3°F)
- INJ +0/-1.10°C (-2°F)
• Operate in Celsius or Fahrenheit
• LED display
• Built-in speaker for audible feedback
• Integrated controller lock-out
Additional features:
• Safety: The controller stores all settings and resumes operation using these settings in the event of a power failure. When the power is restore, the controller sounds an alarm. The on/off status indicator fl ashes.
• Safety: A warming shut-off system, separate from the controller, prevents overheating.
• Convenience: Access point at rear panel allows the addition of data logging or temperature management hardware.
Clearance requirements:
4" (102mm) from rear
1" (25mm) from top
2" (51mm) from sides
1/2" (13mm) from bottom
Specications and Features
7
MN-39362 • Rev 0 • 04/18 • Titan Series EC1850BL Combination Blanket/Fluid Warmer
Electrical Information
Locate the Rating Tag
Verify the power requirements for the appliance. The power specifi cation is located on the appliance identifi cation rating tag. This tag is permanently attached to the appliance.
Hazardous
Voltage Present
CAUTION
Power source must match voltage identified on appliance rating tag. The rating tag provides essential technical information required for any appliance installation, maintenance or repairs. Do not remove, damage or modify the rating tag.
DANGER
To prevent serious personal injury, death, or property damage:
Do not use this warming appliance in the presence of flammable anesthetic mixtures (with air or with oxygen or nitrous oxide).
Not category AP or APG equipment
Safety Class I
Equipment
Protective Earth Ground Symbol
Medical Equipment classifi ed by Underwriters Laboratories with Respect to Electrical Shock, Fire and Mechanical Hazards only, in Accordance with UL 61010–1 and CAN/CSA C22.2 No. 61010–1.
E471516
Grounding reliability can only be achieved when appliance is
connected to an equivalent receptacle marked “Hospital Grade.”
Wire diagram is located under top lid of appliance.
Power Requirements
Fluid compartment: 30 1-liter bottles or
24 1-liter bags
Blanket compartment: 30-40 blankets (16.4 
3
)
Net: 457 lbs (207 kg) Ship: 568 lbs (258 kg)
Shipping dimensions: 38" L x 38" W x 90" H
120 VAC — 50/60 Hz, 1 ph
1.3 kW, 11.0 Amps
Safety Class I Equipment No Applied Parts Mode of Operation: Continuous
NEMA 5-20P
20A - 125V Plug Hospital Grade
IP-X0
E471516
220 VAC — 50/60Hz, 1 ph
1.2 kW, 5.5 Amps
Type B Equipment
CEE 7/7*
220-230V Plug
IP-X0
230 VAC — 50/60 Hz, 1 ph
1.3 kW, 5.6 Amps
Type B Equipment
BS 1363 Plug* (UK only)
CEE 7/7* 220-230V Plug
IP-X0
EC1850BL
*Other international plugs are available, contact the manufacturer for more information.
WARNING
To prevent serious injury, death, or property damage:
All electrical connections must be made by a qualified and trained service technician in accordance with applicable electrical codes.
This appliance must be adequately grounded in accordance with local electrical codes or, in the absence of local codes, with the current edition of the National Electrical Code ANSI/NFPA No.
70. In Canada, all electrical connections are to be made in accordance with CSA C22.1, Canadian Electrical Code Part 1 or local codes.
CE-approved appliances include an equipotential-bonding terminal marked with the symbol shown on the le . Provisions for earthing are to be made in accordance with IEC:2010 60335-1 section 27 or local codes.
8
MN-39362 • Rev 0 • 04/18 • Titan Series EC1850BL Combination Blanket/Fluid Warmer
Electrical Information
The appliance requires special precautions regarding EMC (Electromagnetic Compatibility) and needs to be installed and put into service according to the EMC information provided in the accompanying documents.
Portable and mobile RF communications equipment can aff ect medical electrical equipment.
A risk of increased emissions or decreased immunity
may result if the power cord attached is altered or a manufacturer supplied power cable is not used.
The appliance should not be used adjacent to or stacked with other equipment.
Observe to verify normal operation if it is necessary to use adjacent to or stacked with other equipment.
The essential performance of the appliance is to not exceed an internal temperature of 180°F / 82°C (+10%) for blanket warmers or 150°F / 66°C (+10%) for fl uid warmers.
The appliances are intended for use in the electromagnetic environment specifi ed below. The customer or the end user of this appliance should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions; CISPR 11 Group 1 The appliance uses RF energy only for internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions; CISPR 11 Class B The appliance is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions; IEC 61000-3-2 Class A
Voltage fl uctuations/Flicker emissions; IEC 61000-3-3
Complies
The appliance is intended for use in the electromagnetic environment specifi ed below. The customer or the end user of this appliance should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electromagnetic discharge (ESD)
IEC 61000-4-2
±6 kV contact ±8 kV air
±6 kV contact ±8 kV air
Floors should be wood, concrete or ceramic tile. If fl oors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply lines; ±1 kV for input/output lines
+2 kV for power supply lines
Main power quality should be that of a typical commercial or hospital environment. The appliance does not have any input/output lines.
Surge IEC 61000-4-5
±1 kV diff erential mode; ±2 kV common mode
±1 kV diff erential mode; ±2 kV common mode
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
<5 % UT (>95 % dip in UT) for 0.5 cycle
40 % UT (60 % dip in UT) for 5 cycles
70 % UT (30 % dip in UT) for 25 cycles
<5 % UT (>95 % dip in UT) for 5 sec
<5 % UT (>95 % dip in UT) for 0.5 cycle
40 % UT (60 % dip in UT) for 5 cycles
70 % UT (30 % dip in UT) for 25 cycles
<5 % UT (>95 % dip in UT) for 5 sec
Mains power quality should be that of a typical commercial or hospital environment. If the user of the appliance requires continued operation during power mains interruptions, it is recommended that the appliance be powered from an uninterrupted power supply or a battery.
Power frequency (50/60 Hz) magnetic fi eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fi elds should be at
levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration
Electromagnetic Emissions
Electromagnetic Immunity
9
MN-39362 • Rev 0 • 04/18 • Titan Series EC1850BL Combination Blanket/Fluid Warmer
Electrical Information
The appliance is intended for use in the electromagnetic environment specifi ed below. The customer or the end user of this appliance should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 V/m 150 kHz to 80 MHz
3 V/m 80 MHz to 2.5 GHz
3 V/m
3 V/m
Portable and mobile RF communications equipment should be used no closer to any part of the appliance, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance d = [3.5/3] √P d = [3.5/3] √P 80 MHz to 800 MHz d = [7/3] √P 800 MHz to 2.5 GHz where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fi xed RF transmitters, as determined by an electromagnetic site survey,
a
should be less than the
compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE: 1. At 80 MHz and 800 MHz, the higher frequency range applies. 2. These guidelines may not apply in all situations.
Electromagnetic propagation is aff ected by absorption and reflection from structures, objects and people.
a. Field strengths from fi xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fi xed RF transmitters, an electromagnetic site survey should be considered. If the measured fi eld strength in the location in which the appliance is
used exceeds the applicable RF compliance level above, the appliance should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the appliance.
b. Over the frequency range 150 kHz to 80 MHz, fi eld strengths should be less than [VI] V/m.
The appliance is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the appliance can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the appliance as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.117 0.117 0.233
0.1 0.369 0.369 0.738 1 1.167 1.167 2.333
10 3.689 3.689 7.379
100 11.667 11.667 23.333
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE: 1. At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 2. These guidelines may not apply
in all situations. Electromagnetic propagation is aff ected by absorption and reflection from structures, objects and people.
Electromagnetic Emissions
Electromagnetic Immunity Recommended Separation Distance Between Portable And Mobile RF Communications Equipment And This Appliance
10
MN-39362 • Rev 0 • 04/18 • Titan Series EC1850BL Combination Blanket/Fluid Warmer
Operation Instructions
NOTICE: A temporary odor may be noticeable upon initial
start-up of the warmer. Contact manufacturer if the odor persists a er a day or more of
continuous use.
Cleaning Agents
Always use the proper cleaning agent at the manufacturer’s recommended strength. Contact a local cleaning supplier for product recommendations.
Preparation Before Operating the Warmer
1. Remove and wash any detachable items such as the insert assembly with hot, soapy water. Dry the items with a clean, lint-free cloth.
2. Clean the interior and exterior of the warmer with a mild soap and water solution. Apply the solution with a clean, damp cloth. Do not use commercial or household cleaners that contain ammonia.
NOTE: Make sure to wipe the control panel, door vents,
door handle or door button, and door gaskets.
3. Remove all detergent residue from the interior and exterior of the warmer with a clean, damp cloth.
4. Dry the interior and exterior of the warmer with a clean, lint-free cloth. Leave the door open until the interior is completely dry.
5. Sanitize the interior of the warmer with a sanitizing solution. This solution must be approved for use on stainless steel surfaces.
6. Clean the warmer glass with glass cleaner or distilledvinegar.
7. Install the insert assembly.
Loading...