Enthermics EC1540 User manual

OCM/EC1540 11/06
PRINTED IN U. S .A .
OPERATION AND CARE MANUAL
Corporate Headquarters:
W164 N9221 Water Street ●P.O. Box 443 Menomonee Falls, Wisconsin 53052-0443
PHONE: 262.251.8356 800.TO.B.WARM U.S.A./CANADA FAX: 262.251.7067
www.enthermics.com 800.
862.9276 U.S.A./CANADA 800.329.8744
Blanket
Warming
Cabinet
EC1540
120V
SHOWN WITH OPTIONAL
FULL-PERIMETER RUBBER
BUMPER ASSEMBLY
1
ENTHERMICS MEDICAL SYSTEMS
OCM/EC1540 11/06
UNPACKING and SET-UP
Transport and Storage Environmental Conditions (not to exceed 15 days)
• Ambient temperature range of -40 to +159°F (-40° to +70°C)
• Relative humidity range of 10% to 100%, including condensation
• Atmospheric pressure range of 50KPa to 106KPa
BLANKET WARMING CABINET
DELIVERY
The Enthermics Medical Systems Blanket
Warming Cabinet has been thoroughly tested and
inspected to insure only the highest quality unit is
provided. Upon receipt, check for any possible
shipping damage and report it at once to the
delivering carrier. See Transportation Damage and
Claims section located in this manual.
This appliance, complete with unattached items
and accessories, may have been delivered in one or
more packages. Check to ensure that all standard
items and options have been received with each
model as ordered.
Save all the information and instructions packed
with the appliance. Complete and return the
warranty card to the factory as soon as possible to
assure prompt service in the event of a warranty
parts and labor claim.
This manual must be read and understood by all
people using or installing the equipment model.
Contact the Enthermics service department if you
have any questions concerning installation,
operation, or maintenance.
NOTE: All claims for warranty must include the
full model number and serial number of
the unit.
UNPACKING
1. Carefully remove the
appliance from the
carton or crate.
NOTE: Do not discard the
carton and other
packaging material
until you have
inspected the unit
for hidden damage
and tested it for
proper operation.
2. Read all instructions in this manual carefully
before initiating the installation of this appliance.
DO NOT DISCARD THIS MANUAL.
This manual is considered to be part of the
appliance and is to be provided to the owner
or manager of the business or to the person
responsible for training operators. Additional
manuals are available from the Enthermics
service department.
3. Remove all protective plastic film, packaging
materials, and accessories from the appliance
before connecting electrical power.
TRANSPORT AND STORAGE
2
ENTHERMICS MEDICAL SYSTEMS
OCM/EC1540 11/06
Knowledge of proper procedures is essential to the
safe operation of electrically and/or gas energized
equipment. In accordance with generally accepted
product safety labeling guidelines for potential
hazards, the following signal words and symbols
may be used throughout this manual.
Used to indicate the presence of a hazard that will cause severe personal injury, death, or substantial property damage if the warning included with this symbol is ignored.
Used to indicate the presence of a hazard that can cause personal injury, possible death, or major property damage if the warning included with this symbol is ignored.
Used to indicate the presence of a hazard that can or will cause minor or moderate personal injury or property damage if the warning included with this symbol is ignored.
Used to indicate the presence of a hazard that can or will cause minor personal injury, property damage, or a potential unsafe practice if the warning included with this symbol is ignored.
Used to notify personnel of
installation, operation, or
maintenance information that is
important but not hazard related.
1. Enthermics blanket warmers are intended for
warming cotton blankets ONLY. No other use
for this device is authorized or recommended.
2. This device is intended for use in commercial
establishments where all operators are familiar
with the purpose, limitations, and associated
hazards of this device. Operating instructions
and warnings must be read and understood by
all operators and users.
3. Any troubleshooting guides, component views, and
parts lists included in this manual are for general
reference only and are intended for use by qualified
technical personnel.
4. This manual should be considered a permanent
part of this device. This manual and all supplied
instructions, diagrams, schematics, parts lists,
notices, and labels must remain with the device
if the item is sold or moved to another location.
NOTE:
SAFETY PROCEDURES AND PRECAUTIONS
3
ENTHERMICS MEDICAL SYSTEMS
OCM/EC1540 11/06
ELECTRICAL INFORMATION
The power specifications are located on the unit identification nameplate. This nameplate is
permanently attached to the unit and must be located to verify power requirements.
PREPARATION
Before operating the cabinet, clean both the interior and exterior of the unit with a damp cloth and mild
soap solution. Wipe with an appropriate disinfectant. Clean and install the cabinet basket assembly.
Hazardous
Voltage Present
AT NO TIME SHOULD THE INTERIOR OR EXTERIOR BE STEAM CLEANED, HOSED DOWN, OR FLOODED WITH WATER OR LIQUID SOLUTION OF ANY KIND. DO NOT USE WATER JET TO CLEAN.
SEVERE DAMAGE OR ELECTRICAL HAZARD
COULD RESULT.
WARRANTY BECOMES VOID IF APPLIANCE IS FLOODED.
POWER REQUIREMENTS - EC1540
120 V.A.C. 50/60 Hz, 1 ph 2000 Watts, 16.0 Amps
Safety Class I Equipment
Medical Equipment classified by Underwriters Laboratories with Respect to Electrical Shock, Fire and Mechanical Hazards only, in Accordance with UL 2601-1 and CAN/CSA C22.2 No. 601.1.
UL File No.
E201645
Grounding reliability can only be achieved when equipment is
connected to an equivalent receptacle marked "Hospital Grade."
NEMA 5-20P 20A - 125V Plug Hospital Grade
Protective Earth Ground Symbol
S
GENERAL INFORMATION
4
ENTHERMICS MEDICAL SYSTEMS
OCM/EC1540 11/06
This warming cabinet is designed to elevate blanket
temperatures to a level which will increase patient comfort.
The dual-chambered warming cabinet is constructed with
20 gauge stainless steel exterior casing and door with handle and
hinges designed to withstand heavy usage. Doors with windows
allow observation of inventory with the doors closed. The
cabinet is warmed using the patented Halo Heat® low-heat-
density electrothermal cable array. The electrothermal cable is
positioned in the floor and two sides of the warming cabinet,
providing even heating of the interior chamber. The interior
chamber temperature is regulated by an electronic control. The
design and operational characteristics of the cabinet eliminate the
need for a heat circulating fan.
Each chamber has its own electronic control consisting of
a 4 digit L.E.D. display, ON/OFF key, INCREASE and
DECREASE keys, TIMER key, integrated LOCK feature and a
series of prompt sequence indicators. Each electronic control has
an adjustable temperature range of 90° to 200°F (32° to 94°C).
A warming shut-off system, separate from the electronic control,
prevents overheating.
The TIMER key allows the user to program the control
to automatically turn on and turn off once during a 24 hour
period at selected times. The electronic control can easily be
set to operate in Fahrenheit or Celsius. After a power failure,
the cabinet will remember its programming and begin to
operate as before. The ON/OFF indicator will blink to indicate
a failure occurred; pressing the ON/OFF key once will
eliminate this blinking.
Each chamber is equipped with one (1) white, epoxy-coated
blanket support assembly and one (1) center shelf. The cabinet
is furnished with one (1) set of 5" (127mm) heavy duty casters,
two with locking brake.
AT NO TIME SHOULD THE INTERIOR OR EXTERIOR BE STEAM CLEANED, HOSED DOWN, OR FLOODED WITH WATER OR LIQUID SOLUTION OF ANY KIND. DO NOT USE WATER JET TO CLEAN.
SEVERE DAMAGE OR ELECTRICAL HAZARD
COULD RESULT.
WARRANTY BECOMES VOID IF APPLIANCE IS FLOODED.
DO NOT use this warming cabinet in the
presence of a flammable anethestic mixture
(with air or with oxygen or nitrous oxide).
THIS COULD RISK AN EXPLOSION!