ENMET SE-5175 User Manual

Page 1
ENMET Corporation
PO Box 979 Ann Arbor, MI 48106-0979
www.enmet.com
SE-5175
Sensor/Transmitter
Manual
Manual Part Number 80003-162
MCN-13-011, 12/20/13
Page 2
Table of Contents
1.0 INTRODUCTION................................................................................................................................................................ 1
1.1 U
NPACK
.............................................................................................................................................................................. 1
1.2 C
HECK ORDER
1.3 S
ERIAL NUMBERS
.................................................................................................................................................................... 1
............................................................................................................................................................... 1
2.0 COMPONENTS OF THE SE-5175................................................................................................................................... 2
2.1 SE-5175
2.2 SE-5175 O
2.3 C
ELEMENTS
PERATIONAL FEATURES
IRCUIT BOARD FEATURES
............................................................................................................................................................ 2
..................................................................................................................................... 2
................................................................................................................................................. 3
3.0 INSTALLATION ................................................................................................................................................................. 4
3.1 M
3.2 W
OUNTING IRING THE
SE-5175 ........................................................................................................................................................... 4
SE-5175 ......................................................................................................................................................... 5
3.2.1 Power Supply .............................................................................................................................................................. 5
4.0 OPERATION ....................................................................................................................................................................... 6
4.1 S
TART UP
SE-5175 ............................................................................................................................................................. 6
4.1.1 Typical Start Up .......................................................................................................................................................... 6
4.2 N
ORMAL DISPLAY MODE
.................................................................................................................................................... 7
4.2.1 Alarm Conditions SE-5175 ........................................................................................................................................ 7
5.0 MAINTENANCE ................................................................................................................................................................. 8
5.1 M
AINTENANCE MENUS
....................................................................................................................................................... 8
5.1.1 Exit Maintenance Menu ............................................................................................................................................ 10
5.2 SE-5175 M
AINTENANCE ADJUSTMENTS
........................................................................................................................... 10
5.2.1 Zero Adjust ................................................................................................................................................................ 11
5.2.2 Gas Span ................................................................................................................................................................... 12
5.2.3 Alarm Set Points ....................................................................................................................................................... 13
5.2.4 Span Set ..................................................................................................................................................................... 13
5.4 S
ENSOR REPLACEMENT
.................................................................................................................................................... 14
6.0 REPLACEMENT PARTS ................................................................................................................................................ 15
7.0 TECHNICAL DATA AND SPECIFICATIONS ............................................................................................................. 15
8.0 WARRANTY ..................................................................................................................................................................... 16
List of Tables
T
ABLE 2: FACTORY ALARM SET POINTS
T
ABLE 3:
SE-5175 M
AINTENANCE MENUS SEQUENCE
....................................................................................................................... 7
.................................................................................................. 8
List of Illustrations
F
IGURE 1: EXTERNAL
F
IGURE 2:
F
IGURE 3: MOUNTING
F
IGURE 4: POWER TERMINAL CONNECTIONS
F
IGURE 6:
F
IGURE 7: CALIBRATION ADAPTER
F
IGURE 8:
SE-5175 C
SE-5175 M SE-5175 S
Reference Information:
N
OTE
: [important information about use of instrument]
C
W
ARNING
AUTION
: [affects equipment – if not followed may cause damage to instrument, sensor etc…]
:
[affects personnel safety – if not followed may cause bodily injury or death.]
Earth Ground
SE-5175 F
IRCUIT BOARD FEATURES
EATURES
................................................................................................................... 2
........................................................................................................... 3
SE-5175 .................................................................................................................................... 4
SE-5175 .................................................................................................. 5
AINTENANCE MENU FLOW CHART
............................................................................................... 9
............................................................................................................................. 10
ENSOR REPLACEMENT
.............................................................................................................. 14
Page 3
SE-5175 ENMET Corporation
1.0 Introduction
The SE-5175 sensor/transmitter (S/T) instrument is a 3 wire, 24 VDC, 4-20mA S/T, that measures and detects various types of toxic gases utilizing an electrochemical sensor. The SE-5175 is NOT in an enclosure rated for use in a Class I, Div 1, Groups B, C, D classified area and C
Features of the SE-5175:
continuous monitoring of the sample air continuous LCD display of gas and vapor concentrations menu driven operational and maintenance controls menu driven calibration procedure mA outputs for target gas Sensor/Transmitter are three wire device, with two of the wires forming a 4-20mA loop
N
OTE
: All specifications stated in this manual may change without notice.
1.1 Unpack
Unpack the SE-5175 and examine it for shipping damage. If such damage is observed, notify both ENMET customer service personnel and the commercial carrier involved immediately.
Regarding Damaged Shipments
N
OTE
: It is your responsibility to follow these instructions. If they are not followed, the carrier will not honor
any claims for damage.
This shipment was carefully inspected, verified and properly packaged at our company and delivered to the carrier in good condition.
When it was picked up by the carrier at ENMET, it legally became your company’s property.
If your shipment arrives damaged:
Keep the items, packing material, and carton “As Is.” Within 5 days of receipt, notify the carrier’s local office and
request immediate inspection of the carton and the contents.
After the inspection and after you have received written acknowledgment of the damage from the carrier, contact
ENMET Customer Service for return authorization and further instructions. Have your Purchase Order and Sales Order numbers available.
ENMET either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability coverage, usually $100.00. Repair or replacement charges above that value are your company’s responsibility.
The shipping company may offer optional insurance coverage. ENMET only insures shipments with the shipping company when asked to do so in writing by our customer. If you need your shipments insured, please forward a written request to ENMET Customer Service.
Regarding Shortages
If there are any shortages or questions regarding this shipment, please notify ENMET Customer Service within 5 days of receipt at the following address:
AN NOT
be installed in a hazardous location, as defined by the National Electric Code (NEC).
ENMET Corporation
680 Fairfield Court
Ann Arbor, MI 48108
734-761-1270 734-761-3220 Fax
1.2 Check Order
Check, the contents of the shipment against the purchase order. Verify that the SE-5175 is received as ordered. Each SE­5175 is labeled with its target gas. If there are accessories on the order, ascertain that they are present. Check the contents of
calibration kits. Notify ENMET customer service personnel of any discrepancy immediately.
1.3 Serial Numbers
Each SE-5175 is serialized. These numbers are on tags on the equipment and are on record in an ENMET database.
1
Page 4
SE-5175 ENMET Corporation
Menu
Select
Strain Relief
Visual In
dicators:
Sensor
Visual Indicator
Power/Fault
Front Cover
2.0 Components of the SE-5175
2.1 SE-5175 elements
See Figure 1 for location of elements:
Feature Description Enclosure
Front Cover
2.2 SE-5175 Operational Features
The Display Panel is attached to the enclosure cover and is released by unscrewing the 4 screws located in the corners. After releasing the screws and lifting the cover and exposing the interior of the enclosure, see section 3.0 for installation and wiring instructions. See Figure 1 for location of features.
Feature Description Display
Visual Indicators and Alarms
Membrane Switches
A fiberglass-reinforced polyester 5 x 4 x 3, with a detachable front cover. Four holes for mounting the enclosure to a vertical surface. Located at the corners of the
bottom of the enclosure. See Figure 3 Detachable front cover of SE-5175 with Display Panel. See Section 2.2 and Figure 1 There are 4 Screws that hold the front cover in place.
A single line, 8 character LCD with backlight. Indicates the level of gas detected by sensor. The numerical value of gas concentration and other information is displayed.
LED indicators:
Power / Fault Indicator LED, Green / Red Alarm (3) Indicator LED, Red
2 Pushbutton Switches on front panel control the instrument maintenance functions. The pushbutton switch locations are indicated by:
M
ENU
: Advances the instrument display through operation information and maintenance
menus
S
ELECT
: Disables audio alarm temporally and
Selects the maintenance menu operations such as, Zero, Span, Exit menu or sets proper calibration values for Zero or Span
See Section 4.0 and 5.0 for operational and maintenance flow charts.
Three alarm points are preprogrammed into the SE-5175. At each alarm point, an LED on the front panel is activated. These internal alarm settings are independent of the 4-20mA output alarm values that can be set at a controller.
Alarm1, Alarm2, Alarm3
Display
Pushbutton Switches
Retaining Screws 4 places
Figure 1: External SE-5175 Features
2
Page 5
SE-5175 ENMET Corporation
J4 Terminal
Terminal J8
To Sensor Wiring
2.3 Circuit Board Features
The Display Panel is attached to the enclosure cover and is released by unscrewing the 4 screws located in the corners. After releasing the screws and lifting the cover and exposing the interior of the enclosure, see section 3.0 for installation and wiring instructions. The Circuit Board is mounted at the back surface of the enclosure Cover.
Circuit Board Features are shown in Figure 2.
Feature Description Output Terminals
Sensor Wiring
J4 plug – Terminals to controller J8 plug – Terminals to sensor wiring
4-20mA Output
Figure 2: SE-5175 Circuit Board Features
3
Page 6
SE-5175 ENMET Corporation
Heavier than Air Gas
Sensor Location
Avoid locating anywhere near a vent or window or near an outside
Lighter than Air Gas
Sensor Location
Ammonia
Hydrogen
Same Density as Air Gas
Sensor Location
Mounting Holes
R .128, 4 places
3.0 Installation
The SE-5175 sensor/transmitter detects gas only at the sensor location.
N
OTE
: Sensor Location
Gases have different densities. Some are heavier than air and concentrate at the bottom of a space. Some are lighter than air and gather at the top. Consider the density of the gas you want the sensor to detect when you install the sensor. Some examples are given below.
Hydrogen sulfide Interior wall; 18-24" from floor.
DO N
Near ceiling.
DO N
Carbon Monoxide 4-6 feet above the (generally uniform) floor.
DO N
If you have a question involving the location of a unit or sensor, please contact your distributor or ENMET personnel. A technician will analyze the question and recommend a location.
OT
locate directly above or beside gas appliances (ovens,
heaters).
doorway.
OT
locate directly above appliances where it is subject to
direct exposure to heat or steam.
OT
locate in direct air currents of windows, doors, or vents.
3.1 Mounting SE-5175
Mount the SE-5175 instrument on an appropriate vertical surface, using the mounting holes provided. Avoid areas with excessive vibration or temperature extremes. The holes in the bottom of the enclosure are 0.18 inch in diameter and form a
4.09 square. See Figure 3
Side View
Dimensions are in inches.
Figure 3: Mounting SE-5175
4
Page 7
SE-5175 ENMET Corporation
Position
Function
Position
Function
Sensor
Plug J4
Wires to
J4
J8
Wires to
2
3
1
3.2 Wiring the SE-5175
The electrical installation should conform to appropriate electrical codes, such as the National Electrical Code in the United States.
W
ARNING
C
3.2.1 Power Supply
:
The compliance of the installation to appropriate codes is not ENMET’s responsibility.
AUTION
:
Area must be declassified during installation.
Run conduit and 16
AWG
(1.5
2
MM
) wires to the enclosure from the power supply and controller. Or use a 3 wire power cord of
0.20 to 0.35” in diameter. After releasing the screws and lifting the cover and exposing the terminal strips on the bottom of the circuit board. Connect the wires from the controller (power supply) to the supplied J4 plug then attach to J4 terminal.
Upon supplying power to the SE-5175:
The green power on LED is lit. The display backlight is lit, and instrument will step through a start-up sequence: unit serial number and software revision
may be shown on the display.
The instrument may go into alarm briefly, but the sensors stabilize quickly. If the instrument persists in alarm, acknowledge the alarm by pressing the S
ELECT
button. If alarm persists longer than 30 minutes, call ENMET customer service personnel.
J4 P
LUG – TERMINAL TO CONTROLLER WIRING
1 + 24 V
2 GND 3 4 - 20 mA out
4* RS-485 D+ 5* RS-485 D–
*Contact ENMET for Modbus Address information
DC
power
Controller
Circuit Board Bottom View
J8 P
LUG – TERMINAL TO SENSOR WIRING
To J4
5 4 3 2 1
1 2 3
To J8
1 V + Red 2 Signal White 3 GND Black
Sensor
Plug J8
Figure 4: Power Terminal Connections SE-5175
5
Page 8
SE-5175 ENMET Corporation
SE-5175
W
C
SE
-
5175
84
-
20
0 ppW
0 ppm
4.0 Operation
4.1 Start Up SE-5175
When the SE-5175 is first powered up, it goes through a series of momentary screens, which identify the instrument model number, serial number and software revision. After all of the momentary screens have been displayed, the instrument arrives at the Main Gas Display showing the gas concentration and unit of measurement, the POWER green LED is on and the red alarm and fault LEDs are not lit.
Depending on transmitter configuration and calibration condition, the furthest right character in the display may flash a letter indicating the instrument status. See the Section 4.1.1 below.
4.1.1 Typical Start Up
When power is supplied to the SE-5175, the instrument will display the following sequence of information: Typical start up sequence of information displayed.
Example of Typical Start Up Display
Display Function
The instrument: Serial Number
The instrument: Model
The instrument: Software Revision
S/W 6.5F
IF the right most character is a flashing
IF the right most character is a flashing
0 ppC
N
OTE
:
Software revision may cause variations of display output.
The instrument is in Warm-up mode
This should last about 1 minute The Signal Output is held at 4mA during warm-up
The instrument has failed Calibration
The last good calibration values are retained, but the sensor may not be responsive to gas A new Calibration should be performed As Soon As Possible
The instrument: Normal Display Mode Measurement of the target gas
6
Page 9
SE-5175 ENMET Corporation
4.2 Normal Display Mode
When the SE-5175 is installed as described in section 3, and in clean air, the POWER green LED is on and the display is lit. The red alarm and fault LEDs are not lit.
To advance through displays of operational information press the M
See sequence of operational information below: Example of Typical Operational Display
Display Measurement of the target gas Press M
ENU
button
0 ppm
ENU
button.
M
ENU
Display indicates Alarm 1 Set point Press M
Display indicates Alarm 2 Set point Press M
Display indicates Alarm 3 Set point Press M
Display indicates mA Span range (Full Scale)
Press M
Display returns to operational measurment
ENU
ENU
ENU
ENU
button
button
button
button
A1: 05
A2: 10
A3: 20
mA: 50
M
M
M
M
ENU
ENU
ENU
ENU
Operational Display Flow Chart
4.2.1 Alarm Conditions SE-5175
There are three alarm set points. The factory settings of these alarm set points are shown in Table 2.
Table 2: Factory Alarm Set Points
Gas Alarm 1 Alarm 2 Alarm 3
Ammonia 25 ppm 50 ppm 75 ppm Arsine 0.05 ppm 0.1 ppm 0.4 ppm Carbon Monoxide 35 ppm 50 ppm 200 ppm Chlorine 0.5 ppm 1 ppm 5 ppm Ethylene Oxide 3 ppm 5 ppm 9 ppm Hydrogen 200 ppm 500 ppm 1000 ppm Hydrogen Chloride 5 ppm 10 ppm 20 ppm Hydrogen Fluoride 3 ppm 6 ppm 9 ppm Hydrogen Sulfide 10 ppm 20 ppm 50 ppm Nitric Oxide 25ppm 50 ppm 75 ppm Nitrogen Dioxide 3 ppm 5 ppm 10 ppm Oxygen 17% V 19.5%V 23.5V Ozone 0.1 ppm 0.5 ppm 0.75 ppm Silane 5 ppm 10 ppm 20 ppm Sulfur Dioxide 2 ppm 5 ppm 10 ppm
These alarm set points can be changed within limits; see the maintenance section of this manual for the procedure. If the target gas concentration increases above that of the alarm set point, the associated red LED is lit, the associated relay
changes state, and the audio alarm is activated. Oxygen is the exception with descending and rising alarm points.
7
Page 10
SE-5175 ENMET Corporation
5ppm
Exit
5.0 Maintenance
The SE-5175 maintenance menus that are accessed by pressing the M code is set at the factory and may be changed by following the access code menu explained in section 5.5.
5.1 Maintenance Menus
C
AUTION
: Do Not Attempt A Span Procedure Without Calibration Gas Applied to The Sensor; if this is done, the instrument is
forced into a calibration fault mode.
Pushbutton switches control the M display panel, see Figure 3. The M numbers such as alarm points, calibrations gas, etc. The S
To enter the maintenance menu press and hold the M
ENU
and S
ELECT
functions. The M
ENU
button is used to display the various menu options and make incremental changes to
ELECT
ENU
button for 2 to 4 seconds
Table 3 indicates the maintenance menu sequence see Figure 6 for a detailed maintenance menu flow chart.
Table 3: SE-5175 Maintenance Menus Sequence
Example of Display Function
Normal Display Mode
Press and hold the M
ENU
button for 2 – 4 seconds to enter the Maintenance Menu
The Power/Fault LED will flash Green – Red to indicate the SE-5175 is in Maintenance Mode
Press the M
ENU
button to advance to the Zero procedure
Zero
Not Available for Oxygen units
Measurement of CO
To exit the maintenance Menu and return to the Normal Display Mode:
If intended function Press S
For adjusting Zero: If intended function Press S
ENU
button and entering a valid access code. The access
ENU
and S
ELECT
button locations are indicated on the
button is used to select that option, set zero or span digit.
ELECT
button
ELECT
button
Press the M
ENU
Span
Press the M
ENU
Alarm1
Press the M
ENU
mA Span
Pressing the M You must Press the S
ENU
button without pressing the S
button to advance to the Span procedure
For adjusting the Span: If intended function Press S
button to advance to each Alarm set point procedures
For adjusting the Alarm 1, 2 and 3 set points: If Intended function Press S
Alarm2
Alarm3
button to advance the mA Span set point procedure
For adjusting the mA Span set point: If intended function Press S
ELECT
button will allow you to cycle through the menu options.
ELECT
button in order to initiate the desired operation.
8
ELECT
ELECT
ELECT
button
button
button
Page 11
SE-5175 ENMET Corporation
Apply
Same mV
H
S
0
005
To
If
display Bad ZERO
If
Normal Gas Display
5ppm
M
ENU
Press and H
OLD
the M
ENU
button for 2 – 4 seconds to enter the Maintenance Menus
Zero
Span
Exit
M
ENU
Press the M
M
ENU
M
ENU
S
ELECT
S
ELECT
S
ELECT
Press the S
ENU
button to cycle through Maintenance Menus
PV: 0
Cal OK
OR
Bad ZERO
PV: 0
ELECT
button to return to the Normal Gas Display. See Section 5.1.1
S
ELECT
Press the S
the Zero signal is within Preset Specs the SE-5175 will display
Cal OK, See Section 5.2.1
the Zero signal is not within Preset Specs the SE-5175 will
S
ELECT
You can Press and change the Calibration Gas Level
20
S
ELECT
See Section 5.2.2
Cal Gas until signal value becomes stable
Cal OK
OR
(about 1 to 4 minutes) See Figure 7
If cal is good display will indicate OK If cal is not within preset “range” display will
OR
Bad Sens
indicate Same or Bad Sens
ELECT
button to initiate Zero adjustment
OLD
the
ELECT
button to
Alarm1
M
ENU
S
ELECT
S
ELECT
Alarm2
M
ENU
S
ELECT
Alarm3
M
ENU
S
ELECT
mA Span
M
ENU
return to Normal Gas Display: Press M Then press S
ENU
button until EXIT is displayed
ELECT
button
0005
0010
0010
0020
0050
F
IGURE 6:
To change Alarm set points: Press Menu button until Alarm to be changed is displayed Press Select button to display the set point
The M
ENU
button changes digit indicated by underscore cursor The S
ELECT
button locks underscored digit and moves to
next digit
If change is not within range display returns to first digit If change is within range display moves to next menu
See Section 5.2.3
To change mA Span set point: Press Menu button until mA Span is displayed Press Select button to display the set point
The M
ENU
button changes digit indicated by underscore cursor
The S
ELECT
button locks underscored digit and moves to next digit
See Section 5.2.4
SE-5175 Maintenance Menu Flow Chart
9
Page 12
SE-5175 ENMET Corporation
Select
Menu
5.1.1 Exit Maintenance Menu
Exit maintenance, when the Exit appears on the display. Press the S Display.
Example of Exit menu:
Exit
M
ENU
Press the M
Press the S
ENU
button to cycle through Maintenance Menus
ELECT
button to return to the Normal Gas Display.
S
ELECT
ELECT
button to return to the instrument Normal Gas
5.2 SE-5175 Maintenance Adjustments
Calibration of the SE-5175
Calibration is the process of setting the instrument up to read accurately when exposed to the target gas. The Zero function sets the clean air reference point and the Span function sets the sensitivity of the instrument.
Initial Calibration: Wait 3 – 4 hours after initially supplying power to the SE-5175 instrument before initial calibration. The SE-5175 has been precalibrated at the factory, and initial field calibration should result in only fine tuning to circuit, as well as
a way to check that installation is successful. It is not necessary to open the enclosure to make adjustment. The calibration functions are operated with pushbuttons from outside the enclosure through the
Calibration Zero and Span functions are two separate procedures. They operate independently of each other. It is recommended that the Zero procedure be done prior to the Span procedure.
N
OTE
: Zero function is not used for Oxygen transmitters.
ENMET Corporation recommends at least quarterly calibration of the SE-5175 instrument. Calibration equipment is available from ENMET Corporation to calibrate the SE-5175 instrument.
Calibration adapter, a length of tubing with a regulator for the gas cylinder on one end, and a fitting to connect to the SE- 5175 sensor/transmitter on the other.
Gas cylinder, Zero gas 20.9% oxygen or Span gas, typical 50% of instrument full scale.
Generally, a cylinder of 20.9% Oxygen is used to provide a Zero point or fresh air reference for the calibration.
MENU
and
SELECT
switches.
Sampling/Calibration Adapter
Output Port Fitting
Figure 7: Calibration Adapter
10
Input Port Fitting
Page 13
SE-5175 ENMET Corporation
Zero
If
If
PV:
0
H
M
5.2.1 Zero Adjust
N
OTE
: Zero function is not used for Oxygen transmitters, proceed to Gas Span.
The ZERO function must be performed by exposing the SE-5175 instrument to clean fresh air. If the air at the sensor is in question, use a cylinder of 20.9% oxygen to provide a clean air reference. See Figure 7
Enter the maintenance menu by pressing and holding M Menu flow chart.
After entering the maintenance menu, Press the M Press the S
ELECT
button to perform a Zero.
The display will alternate between Zero and PV: To abort Zero function press and hold M Abort? will appear, press S Press the S
ELECT
button to initiate a Zero adjustment.
ELECT
button to return to Zero.
An auto detect sequence is initiated. After 15 seconds, the SE-5175 will monitor the zero reading for stability.
If the reading stabilizes, within the pre-programmed perimeters, an automatic zero adjustment will be made.
Cal OK appears on the display and in 1 – 2 seconds, display will change to Span. If you wish to Span the sensor press the S
ELECT
If you wish to Exit the maintenance menu, press M the instrument Normal Gas Display
If the reading does not stabilize, within 255 seconds, the procedure will be aborted. Sensor is outside of safe parameters to
be zeroed, the display will read Bad Zero. Repeat Section 5.2.2 Zero Adjust making sure to use a Zero gas of 20.9% Oxygen. ENMET part number 03296-209.
Example of Zero adjustment display:
S
ELECT
ENU
button for 2 to 4 seconds. See Figure 6, SE-5175 Maintenance
ENU
button until the Zero menu is displayed.
ENU
button for 3 – 4 seconds,
button you are now ready to apply gas. Proceed to gas span step 2
ENU
button until Exit is displayed, then press S
Note: You can Press and
M
ENU
When Abort? appears press S
OLD
S
ELECT
the
ENU
button to abort Zero
return to Zero menu
ELECT
ELECT
button to return to
button to
Abort?
S
ELECT
Press the S
ELECT
button to force Zero adjustment
M
ENU
Zero
Cal OK
OR
Bad ZERO
the Zero signal is within Preset Specs the SE-5175 will
display Cal OK momentarily then advance to Span menu
the Zero signal is not within Preset Specs the SE-5175
will display Bad Zero and return to Zero menu
11
Page 14
SE-5175 ENMET Corporation
Apply
cal is not within preset “range” display will indicate Bad Sens
H
S
button to advance to
20
Same mV
5.2.2 Gas Span
It is recommended that the Zero Function be performed first. Do not perform a calibration unless span gas is applied to sensor. Calibration can be aborted by pressing and holing the M
button for 3 – 4 seconds. Enter the maintenance menu. See Figure 6, SE-5175 Maintenance Menu flow chart.
1. Press the M
2. Press the S
The display will alternate between the calibration gas concentration: example (Cal 20) and a signal level (PV). To Abort calibration press and Hold M Span. To change calibration gas level to be used, press and Hold S digit and select button to move to next digit.
3. Attach the associated calibration gas cylinder to the calibration adapter. See Figure 7 on calibration adapter.
4. Open the valve to apply the calibration gas to the sensor. An auto detect sequence is initiated after 30 seconds, the SE-5175 will monitor the cal reading for stability.
5. Watch for the signal level to stabilize. 1 – 4 minutes.
6. Once the signal level has stabilized,
If the Span is successful, “Cal OK” appears momentarily, then will advance to Alarm1 menu. If the sensor is outside of acceptable parameters, “Bad Span” is displayed. If the sensor did not respond, an incompatible span gas was applied and the sensor did not respond at all,
“Same mV” is displayed then will return to Span. If calibration is not successful, it is suggested that calibration be attempted again in 30-60 minutes.
7. Remove the calibration gas.
8. Calibration is complete.
Note: The instrument will return to operation mode in 3 – 5 seconds.
9. Press the M
Example of Calibration Display:
Span
ENU
button until Span display.
ELECT
button to perform a Span procedure.
ENU
button for 3 – 4 seconds, Abort? will appear, press S
If the sensor will not calibrate See Section 5.4.
ENU
button to advance to next desired menu
N
OTE
:
To abort calibration or change calibration gas level.
M
You can Press and H
ENU
Abort?
You can Press and
ELECT
Calibration Gas Level
-Use the M
-Use the S
S
ELECT
S
ELECT
S
PV: 0
ELECT
button for 3 – 4 seconds, use menu button to change
OLD
the M
ENU
When Abort? appears press S
ENU
button to change digits
ELECT
S
ELECT
OLD
the
ELECT
button to move to next digit
ELECT
button to return to
button to abort Calibration
ELECT
Alarm1 menu
button to change the
ENU
M
ENU
PV: 0
20
Cal OK
Bad Sens
OR
When cal signal is stable SE-5175 will automatically update: If cal is good display will indicate OK or Same and advance to
If
To exit press M
Cal Gas until signal value becomes stable
(about 1 to 4 minutes) See Figure 7
Alarm1
or Same mV The SE-5175 will return to the Span Menu
ENU
button until Exit appears and press S
ELECT
12
Page 15
SE-5175 ENMET Corporation
5.2.3 Alarm Set Points
The SE-5175 alarm set points can be changed within limits. See Table 4 for factory set alarm points. To change any of the three alarm points: Enter the maintenance menu as shown in Figure 6 SE-5175 Maintenance Menu flow chart.
1. Press the M
2. Press the S
3. Press the M
4. Press the S
When last digit is entered the SE-5175 will advance to the next menu
5. Press the M
Example of Alarm Set Point menus:
Alarm1
M
Alarm2
M
Alarm3
ENU
ENU
ENU
button until to display Alarm1 is displayed.
ELECT
button to initiate alarm set point change
ENU
button to change the digit indicated by the underscore cursor
ELECT
button to move the cursor to the next digit
ENU
button to advance to the next menu
S
S
S
ELECT
ELECT
ELECT
0005
0005
0010
0010
To change Alarm set points: Press Menu button until Alarm to be changed is displayed Press Select button to display the set point
The M
ENU
cursor The S
ELECT
next digit
If change is not within range display returns to first digit If change is within range display moves to next menu
0020
button changes digit indicated by underscore
button locks underscored digit and moves to
M
ENU
5.2.4 Span Set
The SE-5175 4-20mA span range can be changed within limits. See Table 4 for factory set range. To change the span range: Enter the maintenance menu as shown in Figure 6 SE-5175 Maintenance Menu flow chart.
1. Press the M
2. Press the S
3. Press the M
4. Press the S
When last digit is entered the SE-5175 will advance to the next menu
5. Press the M
Example of mA Span menu:
mA Span
M
ENU
Default mA Span
4mA 20mA 0 ppm 50 ppm
ENU
button until to display Span is displayed.
ELECT
button to initiate the mA Span menu
ENU
button to change the digit indicated by the underscore cursor
ELECT
button to move the cursor to the next digit
ENU
button to advance to the next menu
S
ELECT
0050
To change mA Span set points: Press Menu button until mA Span is displayed Press Select button to display the set point
The M
ENU
button changes digit indicated by underscore cursor
The S
ELECT
button locks underscored digit and moves to next digit
13
Page 16
SE-5175 ENMET Corporation
5.4 Sensor Replacement
W
ARNING
:
Power must be removed from the SE-5175 before this or any internal procedure. Failure to do so may cause
damage to equipment, bodily injury or death.
Sensors should be replaced when they can no longer be calibrated. Replacement sensor part numbers are listed in Section 6.0 of this manual. If you do not know the proper part number for your sensor, have the SE-5175 serial number available when contacting your Distributor or ENMET Corporation Technical Support.
1. Remove, the set screw from SE-5175 sensor head, and remove sensor cap. See Figure 8
N
OTE
: Set screw must be loosened before attempting to rotate or remove the sensor housing cap.
Hex Key size is 0.050 inch (1.27mm).
2. Remove, the sensor assembly, see Figure 8.
Sensor assembly is made up of the sensor attached to the sensor PCB with spring loaded contact pins.
3. Insert, the new sensor assembly.
4. Replace, sensor cap and replace the set screw.
5. Re-supply power to the SE-5175
Sensor Assembly
Figure 8: SE-5175 Sensor Replacement
After the new sensor assembly has been installed, it is suggested to allow the sensor to stabilize for 3 – 4 hours.
A Factory calibration must be performed. When entering the Maintenance menu EXIT will be displayed. Press and hold the M
M
ENU
switch to change display to Zero or Span.
An F will appear on the far right hand side of the display. The F indicates that the instrument is in Factory mode. Perform the calibration Zero and Span procedures as outlined in Section 5.2. Be sure that the F is present when selecting the
Zero and Span functions. The Factory calibration sets a calibration window for future standard instrument calibrations.
ENU
button for 2-4 seconds. Press the
14
Page 17
SE-5175 ENMET Corporation
Electrical Power
Storage and Transport
Operation
Mechanical
Outputs
6.0 Replacement Parts
ENMET replacement part numbers:
Description of Part Part Number
Sensor – Contact ENMET 67027-xxxx Calibration Adapter 03700-034 Calibration Regulator, 58/103 liter 02506-002 Calibration Regulator, 17/34 liter 02506-004
Zero Gas, 20.9% O2, 17 liter 03296-209 Calibration gas – Contact ENMET
7.0 Technical Data and Specifications
24 VDC
0.6A, 24VDC
Temperature: preferred Relative Humidity Atmospheric Pressure
Temperature: Relative Humidity Atmospheric Pressure
Dimensions: Weight: Material:
Strain relief:
N
OTE
: All specifications stated in this manual may change without notice.
-20° to +60°C (-4° to +140°F) 0° to +20°C (32° to 68°F) 10-99% RH, non-condensing 20 to 36 inHg (68 to 133 kPa)
0° to +40°C (32° to +104°F) 10-99% RH, non-condensing 20 to 36 inHg (68 to 133 kPa)
3-6.5mm OD, 1 supplied
15
Page 18
SE-5175 ENMET Corporation
8.0 WARRANTY
ENMET warrants new instruments to be free from defects in workmanship and material under normal use for a period of one year from date of shipment from ENMET. The warranty covers both parts and labor excluding instrument calibration and expendable parts such as calibration gas, filters, batteries, etc... Equipment believed to be defective should be returned to ENMET within the warranty period (transportation prepaid) for inspection. If the evaluation by ENMET confirms that the product is defective, it will be repaired or replaced at no charge, within the stated limitations, and returned prepaid to any location in the United States by the most economical means, e.g. Surface UPS/FedEx Ground. If an expedient means of transportation is requested during the warranty period, the customer is responsible for the difference between the most economical means and the expedient mode. ENMET shall not be liable for any loss or damage caused by the improper use of the product. The purchaser indemnifies and saves harmless the company with respect to any loss or damages that may arise through the use by the purchaser or others of this equipment.
This warranty is expressly given in lieu of all other warranties, either expressed or implied, including that of merchantability, and all other obligations or liabilities of ENMET which may arise in connection with this equipment. ENMET neither assumes nor authorizes any representative or other person to assume for it any obligation or liability other than that which is set forth herein.
NOTE: When returning an instrument to the factory for service:
Be sure to include paperwork. A purchase order, return address and telephone number will assist in the expedient repair and return of your unit. Include any specific instructions. For warranty service, include date of purchase If you require an estimate, please contact ENMET Corporation.
There are Return for Repair Instructions and Form on the last pages of this manual. This Form can be copied or used as needed.
Manual Part Number 80003-162
May 2007 MCN-433, 10/14/10 MCN-13-011, 12/20/13
Notes:
16
Page 19
PO Box 979 680 Fairfield Court Ann Arbor, Michigan 48106-0979
734.761.1270 Fax 734.761.3220
Returning an Instrument for Repair
ENMET instruments may be returned to the factory or any one of our Field Service Centers for regular repair service or calibration. The ENMET Repair Department and Field Service Centers also perform warranty service work.
When returning an instrument to the factory or service center for service, paperwork must be included which contains the following information:
A purchase order number or reference number.  A contact name with return address, telephone and fax numbers  Specific instructions regarding desired service or description
of the problems being encountered.
Date of original purchase and copy of packing slip or invoice
for warranty consideration.
If a price estimate is required, please note it accordingly and be
sure to include a fax number.
Providing the above information assists in the expedient repair and return of your unit.
Failure to provide this information can result in processing delays.
ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour. All instruments sent to ENMET are subject to a minimum evaluation fee, even if returned unrepaired. Unclaimed instruments that ENMET has received without appropriate paperwork or attempts to advise repair costs that have been unanswered, after a period of 60 days, may be disposed of or returned unrepaired COD with the evaluation fee.
Service centers may have different rates or terms. Be sure to contact them for this information.
Repaired instruments are returned by UPS/FedEx Ground and are not insured unless otherwise specified. If expedited shipping methods or insurance is required, it must be stated in your paperwork.
Note: Warranty of customer installed components.
If a component is purchased and installed in the field, and fails within the warranty term, it can be
returned to ENMET and will be replaced, free of charge, per ENMETs returned goods procedure.
If the entire instrument is returned to ENMET Corporation with the defective item installed, the item will
be replaced at no cost, but the instrument will be subject to labor charges at half of the standard rate.
Page 20
Mailing Address:
Shipping Address:
Repair Return Form
ENMET Corporation PO Box 979 Ann Arbor, Michigan 48106
Phone Number: 734.761.1270 FAX Number: 734.761.3220
Your Mailing Address:
Contact Name: __________________________ Your Phone: _______________________ Your PO/Reference Number: _______________ Your FAX: _______________________
ENMET Corporation Attn: Repair Department 680 Fairfield Court Ann Arbor, Michigan 48108
Your Shipping Address:
Payment Terms: K COD
(Check one) K VISA / MasterCard______________________ ________ ________
Card number Expiration Card Code
K American Express______________________ ________ ________
Card number Expiration Card Code
Name as it appears on the credit card___________________________________
Return Shipping Method:
K UPS: K Ground K 3 Day Select K Next Day Air K ND Air Saver K 2-Day Air K UPS Account number: ________________________
K Federal Express: K Ground K Express Saver K P-1 K Standard K 2-Day Air K FedEx Account number: ________________________
Would you like ENMET to insure the return shipment? K No K Yes Insurance Amount: $_________________
Loading...