1.2 Check Order .....................................................................................................................................................1
1.3 Serial Numbers.................................................................................................................................................1
2.0 O
ZONE GENERATOR FEATURES SECTION .................................................................................................. 2
Checking Output with Ozone Detector Tubes:................................................................................................ 3
Tests of Pump Performance ............................................................................................................................3
Check for Leakage: .........................................................................................................................................4
Correction of Pump Leakage ..........................................................................................................................4
Verification of Ozone Output.................................................................................................................................5
FIGURE 3: GAS DETECTOR TUBE PUMP........................................................................................................... 4
FIGURE 4: CUTAWAY VIEW OF PUMP ............................................................................................................... 4
F
IGURE 5: OZONE GENERATOR WITH GAS BAG ............................................................................................... 5
FIGURE 6: GAS BAG WITH DETECTOR TUBE .................................................................................................... 5
Reference information:
N
OTE: [important information about use of instrument]
AUTION: [affects equipment]
C
WARNING: [affects personnel safety]
OZONE GENERATORENMET Corporation
1.0 Introduction
NOTE: All specifications stated in this manual may change without notice.
1.1 Unpack
Unpack the OZONE GENERATOR and examine it for shipping damage. If such damage is observed, notify both
ENMET customer service personnel and the commercial carrier involved immediately.
Regarding Damaged Shipments
NOTE: It is your responsibility to follow these instructions. If they are not followed, the carrier will
not honor any claims for damage.
This shipment was carefully inspected, verified and properly packaged at our company and delivered to the
carrier in good condition.
When it was picked up by the carrier at ENMET, it legally became your company’s property.
If your shipment arrives damaged:
•Keep the items, packing material, and carton “As Is.” Within 5 days of receipt, notify the carrier’s local
office and request immediate inspection of the carton and the contents.
•After the inspection and after you have received written acknowledgment of the damage from the carrier,
contact ENMET Customer Service for return authorization and further instructions. Have your Purchase
Order and Sales Order numbers available.
ENMET either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability
coverage, usually $100.00. Repair or replacement charges above that value are your company’s responsibility.
The shipping company may offer optional insurance coverage. ENMET only insures shipments with the
shipping company when asked to do so in writing by our customer. If you need your shipments insured, please
forward a written request to ENMET Customer Service.
Regarding Shortages
If there are any shortages or questions regarding this shipment, please notify ENMET Customer Service within 5 days
of receipt at the following address:
ENMET Corporation
680 Fairfield Court
Ann Arbor, MI 48108
734-761-1270 734-761-3220 Fax
1.2 Check Order
Check, the contents of the shipment against the purchase order. Verify that the OZONE GENERATOR is received as
ordered. If there are accessories on the order, ascertain that they are present. Check the contents of calibration kits.
Notify ENMET customer service personnel of any discrepancy immediately.
1.3 Serial Numbers
Each OZONE GENERATOR is serialized. These numbers are on tags on the equipment and are on record in an ENMET
database.
1
ENMET CorporationOZONE GENERATOR
2.0 OZONE GENERATOR Features Section
See Figure 1 for location of features
FeatureDescription
Flowmeter
Air Inlet Port
Cal Gas Outlet Port
Power Switch
Operational Voltage:
This Flowmeter allows the user to verify and monitor the flow rate of the Ozone
Generator. Proper flow rate, approximately 0.5 lpm(liter per minute).
Fitting for optional connection of inlet air.
Normally ambient room air is used.
Fitting for connection of tubing to the Ozone sensing unit or gas bag.
Toggle switch
110 V
220 V
AC 50/60 Hz
AC 50/60 Hz (optional, must specify when ordering)
Air Inlet Port
Power Switch
Blue Light
Ozone Output Adjustment
Glass Tube
Flowmeter
Calibration Gas
Outlet Port
Figure 1: Ozone Generator Features
2
OZONE GENERATORENMET Corporation
Operation
OZONE GENERATOR CALIBRATION KIT: part numbers
04056-0800-2(110
04056-0800-3(220VAC).
WARNING: This device is capable of generating ozone at concentrations above its Permissible Exposure Limit (PEL).
This instrument should be used in a well ventilated area.
The Ozone generator is preset at the factory to produce an output of 0.3ppm of Ozone. See the following procedures
for verification and or adjustment.
Place the instrument in an upright position.
Plug the instrument into an appropriate V
The factory preset flow rate is 0.5 lpm and ozone out put is 0.3 ppm.
Allow the generator to operate for at least 60 minutes for stabilization.
Review and understand the operation of the detector tube pump, prior to beginning.
VAC)
AC outlet and turn the instrument on.
Gas Bag
Valve: Inlet
From Ozone Generator
Valve: Outlet
To Sample Draw System
Ozone Monitor
Figure 2: OZONE GENERATOR Calibration Connection
Detector Tube Pump Operation
Checking Output with Ozone Detector Tubes:
Each box of detector tubes includes instructions for use of the pump and tubes, and for reading the tube after sampling.
AUTION: Do not use any other manufacturer’s tubes with this pump, or another manufacturer’s pump with these tubes.
C
Doing so may result in erroneous reading.
Tests of Pump Performance
Before making gas concentration determinations, the pump should be checked for proper performance using the
leakage test described in the next section.
To ensure accuracy and reliability these tests should be performed whenever:
The pump is used after an extended idle period
In routine use should be repeated after every 10 – 12 samples.
3
ENMET CorporationOZONE GENERATOR
Check for Leakage:
a) Insert an unused sealed detector tube in the pump inlet; do not break the tips.
b) Line up the guide marks (red dots) on the pump shaft and stopper, and pull the handle back all the way.
It locks automatically.
c) Wait exactly two minutes; then release the pump handle with a 0ne-quarter turn.
d) When the handle is released, the piston should spring back all the way to 0cc.
If it does not, the pump is not leak-tight, and the amount of leakage is indicated by the position of the handle as
it comes to rest.
If there is more than 5cc of leakage in two minutes, the accuracy of any concentration determination will be
subject to error. Do not use the pump until this is corrected.
a) Insert tubeb) Pull handle all the wayc) unlock handle
Figure 3: Gas Detector Tube Pump
Correction of Pump Leakage
If excessive leakage is found, it usually takes place either at the pump inlet, or between the piston & cylinder walls.
The latter source of leakage is much less likely than the former.
Leakage between piston & cylinder can usually be eliminated simply by cleaning and re-lubricating.
Leakage at the inlet may result from a poor seal between the detector tube and the rubber inlet tip, or between the
flange of the rubber inlet tip and the pump body.
To check for leakage at the tip:
Carefully reposition the sealed detector tube or replace it with another one, and repeat steps a – c leak check.
If the leakage persists, it is probably at the flange-body interface, and simply tightening down the inlet clamping
may suffice to eliminate it.
If tightening the clamp does not work, remove the rubber inlet and examine it for cracks and/or foreign matter on
its sealing surface.
If it is clean and undamaged, replace it in the pump inlet, and repeat the leak test.
If necessary, replace a worn rubber inlet with a fresh one.
Figure 4: Cutaway View of Pump
4
OZONE GENERATORENMET Corporation
Pump Maintenance
In normal use, the pump should be taken apart, cleaned and re-lubricated after 100 to 200 samples have been drawn.
To disassemble:
Unscrew the cylinder and cover
Slowly pull the pump and pressure-relief valve from the front and of the pump
Clean all the bearing surfaces carefully using non-abrasive wipers and a good degreasing solvent such as
trichloroethylene or perchloroethylene.
Lubricate all surfaces that make metal-to-metal contact with the special silicone grease supplied with the pump kit.
Reassemble the pump carefully, making sure that abrasive particles that could score the walls do not get inside the pump.
Verification of Ozone Output
Verify, and if necessary, adjust flow rate to 0.5 lpm. After the generator has been on for 60 minutes or longer, open
the valves on the gas bag. Next flatten and roll it up to empty as much of the air from the bag as possible.
Close one valve and attach the bag to the output port on the generator.
Allow the bag to fill partially with ozone, remove it from the generator and empty the bag again, as described above.
Reattach the bag to the generator and partially fill. Open the second valve on the bag, insert the end of the detector
tube into the valve on the bag and draw a sample.
Valve: Inlet
Valve: Outlet
Figure 5: OZONE GENERATOR with Gas Bag
Ozone in from Generator
Gas Bag
Figure 6: Gas Bag with Detector Tube
When sampling cycle is complete, check tube for Ozone concentration.
If within 10% of desired output, generator is working properly.
If not within 10% of desired output, see Ozone Output Adjustment section for Ozone Generator.
5
ENMET CorporationOZONE GENERATOR
Ozone Output Adjustment:
If the output is not the required level:
Verify that the flow meter is set to 0.5 lpm.
When the Ozone output is close to the desired concentration, adjustment should be made by a change in the position of
the glass tube. This can be made by single turns of the knurled adjustment collar. Pushing the glass tube in decreases
Ozone output and pulling the glass tube out increases the Ozone output.
C
AUTION: Small changes to the position (length of exposed lamp) will greatly affect the output of the generator.
Loosen the retaining knurled collar on the ozone generator light (blue light).
If the level of output is low, pull the glass tube out slightly.
See section on verification of Ozone output.
If it is too high, push the glass tube in slightly.
See section on verification of Ozone output.
Tighten the retaining ring on the generator light.
Allow the generator to stabilize for at least 60 minutes and repeat the verification steps.
When output level is correct apply Ozone to monitor.
N
OTE: A sample draw system is used with the ENMET Ozone monitors. The sample bag must be in place during
monitor calibration.
(See Figure 2: Ozone Generator out to sample bag, sample bag out to sample draw system.)
Replacement Part Numbers
ENMET replacement part numbers:
DescriptionPart Number
Ozone Generator04052-009
Tubing (per inch)73073-000
Pump Kit, for Detector Tube90000-111
Detector Tube, Ozone (box of 10)90182-200
Gas Bag, (2 liter, 2 valve)04052-014
6
OZONE GENERATORENMET Corporation
WARRANTY
ENMET warrants new instruments to be free from defects in workmanship and material under normal use for a period
of one year from date of shipment from ENMET. The warranty covers both parts and labor excluding instrument
calibration and expendable parts such as calibration gas, filters, batteries, etc... Equipment believed to be defective
should be returned to ENMET within the warranty period (transportation prepaid) for inspection. If the evaluation by
ENMET confirms that the product is defective, it will be repaired or replaced at no charge, within the stated
limitations, and returned prepaid to any location in the United States by the most economical means, e.g. Surface
UPS/RPS. If an expedient means of transportation is requested during the warranty period, the customer is responsible
for the difference between the most economical means and the expedient mode. ENMET shall not be liable for any
loss or damage caused by the improper use of the product. The purchaser indemnifies and saves harmless the
company with respect to any loss or damages that may arise through the use by the purchaser or others of this
equipment.
This warranty is expressly given in lieu of all other warranties, either expressed or implied, including that of
merchantability, and all other obligations or liabilities of ENMET which may arise in connection with this equipment.
ENMET neither assumes nor authorizes any representative or other person to assume for it any obligation or liability
other than that which is set forth herein.
NOTE: When returning an instrument to the factory for service:
Be sure to include paperwork.
A purchase order, return address and telephone number will assist in the expedient repair and return of your unit.
Include any specific instructions.
For warranty service, include date of purchase
If you require an estimate, please contact ENMET Corporation.
There are Return for Repair Instructions and Form, on the last pages of this manual. This Form can be copied or used
as needed.
7
ENMET CorporationOZONE GENERATOR
Notes:
8
PO Box 979
680 Fairfield Court
Ann Arbor, Michigan 48106-0979
734.761.1270 Fax 734.761.3220
Returning an Instrument for Repair
ENMET instruments may be returned to the factory or any one of our Field Service Centers for regular
repair service or calibration. The ENMET Repair Department and Field Service Centers also perform
warranty service work.
When returning an instrument to the factory or service center for service, paperwork must be included
which contains the following information:
¾ A purchase order number or reference number.
¾ A contact name with return address, telephone and fax numbers
¾ Specific instructions regarding desired service or description
of the problems being encountered.
¾Date of original purchase and copy of packing slip or invoice
for warranty consideration.
¾If a price estimate is required, please note it accordingly and be
sure to include a fax number.
Providing the above information assists in the expedient repair and return of your unit.
Failure to provide this information can result in processing delays.
ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the
closest tenth of an hour. All instruments sent to ENMET are subject to a minimum $30 evaluation fee,
even if returned unrepaired. Unclaimed instruments that ENMET has received without appropriate
paperwork or attempts to advise repair costs that have been unanswered, after a period of 60 days,
may be disposed of or returned unrepaired COD with the evaluation fee.
Service centers may have different rates or terms. Be sure to contact them for this information.
Repaired instruments are returned by UPS/FedEx Ground and are not insured unless otherwise
specified. If expedited shipping methods or insurance is required, it must be stated in your
paperwork.
Note: Warranty of customer installed components.
If a component is purchased and installed in the field, and fails within the warranty term, it can
be returned to ENMET and will be replaced, free of charge, per ENMET’s returned goods
procedure.
If the entire instrument is returned to ENMET Corporation with the defective item installed, the
item will be replaced at no cost, but the instrument will be subject to labor charges at half of the
standard rate.
Repair Return Form
Mailing Address:
ENMET Corporation
PO Box 979
Ann Arbor, Michigan 48106