ENMET Motorized Sample Pump User Manual

Page 1
ENMET Corporation
PO Box 979 Ann Arbor, MI 48106-0979
Sample Pump Kit
Manual
80021-024 09/04/91 MCN# 145, 10/25/94 MCN# 150, 02/23/96 MCN# 199, 10/08/98 MCN# 204, 10/20/98 MCN# 254, 04/09/01 MCN# 257, 06/07/01
Page 2
1.0 INTRODUCTION........................................................................................................2
1.1 The pump assembly ..........................................................................................................................2
1.2 Intrinsic Safety: ..................................................................................................................................2
2.0 F EATURES..............................................................................................................2
3.0 OPERATION............................................................................................................3
3.1 Charging:.............................................................................................................................................3
3.2 Fittings:...............................................................................................................................................3
3.3 Mountings:.........................................................................................................................................3
3.4 Turn Pump On:..................................................................................................................................3
3.5 Use:......................................................................................................................................................3
3.6 Alarms: ................................................................................................................................................3
3.7 Shutdown:...........................................................................................................................................3
4.0 TROUBLESHOOTING.................................................................................................4
4.1 Pump won’t turn on:.........................................................................................................................4
4.2 Pump running, audio and visual alarms activated:......................................................................4
5.0 MAINTENANCE........................................................................................................4
5.1 Storage: ................................................................................................................................................4
5.2 Charging:.............................................................................................................................................4
5.3 Replacement Part Numbers:............................................................................................................4
6.0 WARRANTY............................................................................................................4
List of Illustrations
FIGURE 1: SAMPLE PUMP FEATURES...........................................................................................1
Page 3
Sample Pump Kit ENMET Corporation
Fitting: Air Input
Fitting: Air Output
Line
Air Output Line
Red LED
FRONT VIEW
Power Switch
Belt Clip
Line
Green LED
BACK VIEW
Filter
Air Input Line
HOSE ASSEMBLY
WAND ASSEMBLY
Figure 1: Sample Pump Features
1
Page 4
ENMET Corporation Sample Pump Kit
1.0 Introduction1.0 Introduction
1.1 The pump assembly
The pump assembly is a fixed flow device used to supply sample air at the correct flow rate to all ENMET portable instruments models.
1.2 Intrinsic Safety:
The pump is designed to be nonincendive as defined by the National Electrical Code.
WWARNINGARNING: Do not change electrical components. In particular, do not use a different battery
pack or charger than the one designed for use with the pump. Substitutions may affect the nonincendive design of the pump; personal injury, death, and property damage may result from an explosion.
2.0 Features2.0 Features
See Figure 1 for location of features.
Feature Description POWER Switch Green LED Red LED and Audio Alarm Air Input Line Fitting, Air Input Line
Filter Air Output Line
Fitting, Air Output Line Charge Jack Charger Wand Assembly
Hose Assembly
Sensor Cover (not shown)
Controls power from the battery to the pump motor and circuits. Indicates that power is being supplied through the switch. Indicates that air flow is insufficient, or a low battery condition. Conducts air to the pump assembly. Connects with the mating fittings on the hose assembly and the wand assembly
supplied with the pump kit. Prevents particulate matter and liquids from entering the pump assembly. Conducts air from the pump assembly to the instrument or associated sensor
shield. Connects with the mating fitting on the instrument or associated sensor shield. Connects with the mating plug on the charger. Supplies charging current to the pump assembly battery. A jointed Plexiglas tube with attached flexible tubing to conduct input air to the
pump assembly. Approximately 6 meters of flexible tubing to conduct input air to the pump
assembly. A sheet metal or plastic enclosure to partially block the holes in the instrument
sensor area. This cover differs from instrument to instrument.
2
Page 5
Sample Pump Kit ENMET Corporation
3.0 Operation3.0 Operation
3.1 Charging:
Charge the pump assembly battery by plugging the charger into an appropriate wall socket and the charger plug into the jack on the pump assembly. Charge for 14 - 16 hours
WWARNINGARNING: Do not charge in a hazardous atmosphere as defined by the national electrical code.
Property damage, injury, or death may result if an explosion occurs.
3.2 Fittings:
Attach the air output line fitting to the mating fitting on the instrument or associated sensor cover. Attach the input device selected, wand assembly or hose assembly to the air input line fitting.
3.3 Mountings:
The pump assembly is equipped with a belt clip, used to attach the pump to a belt, pocket or harness strap. To mount the pump on an instrument:
Remove the belt clip. Note: The screws also hold the internal circuit board in place.
Replace the screws.
Using the four pieces of adhesive backed Velcro supplied with the kit , two with hooks and two with loops.
Remove the protective paper, stick these to the pump and instrument, and use them to mount the pump on the
instrument.
3.4 Turn Pump On:
Supply electrical power to the pump motor by putting the POWER Switch in the on position (S /N below 1000) or by pushing the POWER Switch. The Red LED and associated AUDIO ALARM may activate momentarily; this is acceptable.
Observe that the green ON LED is lit, and listen for the noise made by the pump. Units with S/N 1000 and above are very quiet.
3.5 Use:
Use the pump assembly as required to supply appropriate sample air to the instrument.
3.6 Alarms:
Audio alarm and red LED indicate either air flow interruption or low battery voltage.
CAUTION: If at anytime during use of the pump the alarms are activated, discontinue use of the pump. If in a confined
space or other potentially dangerous location leave immediately
3.7 Shutdown:
When use is no longer required, turn the pump off, remove the sensor cover from the instrument, and disassemble the pump kit as required for storage.
CAUTION: Failure to remove sensor cover from instrument will cause the gas detector to be ineffectual for ambient air
monitoring.
3
Page 6
ENMET Corporation Sample Pump Kit
4.0 Troubleshooting4.0 Troubleshooting
4.1 Pump won’t turn on:
§ Put pump on charge (12 – 16 hours).
§ Check output of charger.
4.2 Pump running, audio and visual alarms activated:
§ Check tubes for obstructions, fluid or particulate.
§ Check input filter.
§ Charge pump (12 – 16 hours).
§ Check output of charger.
If unable to achieve proper operation of pump contact ENMET or your local distributor.
5.0 Maintenance5.0 Maintenance
5.1 Storage:
Store the pump and associated equipment in a clean, dry location.
5.2 Charging:
Do not charge the battery continuously. If the pump is not used for a period of time, exercise the battery by running the pump until the battery reaches "Low Batt" alarm then recharging the battery for 14 – 16 hours.
5.3 Replacement Part Numbers:
ENMET Part Number Description
73081-101 Input Filter 67019-102 Battery Pack 67019-001 Battery Pack; for S/N 1000 and above 67051-025 Charger, 110VAC 67051-027 Charger, 220VAC 03700-018 Wand assembly 73073-001 Flexible tubing, sold by the foot 73073-017 Fitting, quick release, male 04805-006 Pump, for S/N 1879 and below 04805-008 Pump, for S/N 1880 and above
6.0 Warranty6.0 Warranty
ENMET warrants new pump assemblies to be free from defects in workmanship and material under normal use for a period of nine months from date of shipment from ENMET. The warranty covers both parts and labor. Equipment believed to be defective should be returned to ENMET within the warranty period (transportation prepaid) for inspection. If the evaluation by ENMET confirms that the product is defective, it will be repaired or replaced at no charge, within the stated limitations, and returned prepaid to any location in the United States. ENMET shall not be liable for any loss or damage caused by the improper use of the product. The purchaser indemnifies and saves harmless the company with respect to any loss or damages that may arise through the use by the purchaser or others of this equipment.
Material shipped to ENMET for warranty evaluation must be packed so it is not damaged in shipping. Material damaged in shipment is not covered by warranty.
This warranty is expressly given in lieu of all other warranties, either expressed or implied, including that of merchantability, and all other obligations or liabilities of ENMET which may arise in connection with this equipment. ENMET neither assumes nor authorizes any representative or other person to assume for it any obligation or liability other than that which is set forth herein.
4
Loading...