ENMET ISA-RAL-M User Manual

ENMETENMET Corporation
PO Box 979 Ann Arbor, MI 48106-0979
ISA-RAL-M
Operation and Maintenance
80002-030
04/10/95 MCN 149, 10/31/95 MCN 162, 03/25/97 MCN 164, 04/02/97 MCN 179, 10/06/97 MCN 204, 10/20/98 MCN 205, 01/19/99 MCN 214, 03/16/99 MCN 220, 09/17/99 MCN 241, 09/22/00 MCN-304, 10/17/03
Manual
Table of Contents
1.0 INTRODUCTION............................................................................................................................................................................1
1.1 OPERATION PRINCIPLE.........................................................................................................................................................................1
1.2 ISA-RAH-M HIGH PRESSURE MONITOR..........................................................................................................................................1
1.3 HYDROCARBON FILTER........................................................................................................................................................................1
2.0 UPON RECEIPT.....................................................................................................................................................................................1
3.0 FEATURES, DEFINITION OF PARTS ..........................................................................................................................................2
3.1 ELECTRONICS CONTROL UNIT ............................................................................................................................................................2
3.2 INTERNAL ELECTRONIC CIRCUITRY...................................................................................................................................................3
3.3 SAMPLE HEAD ASSEMBLY...................................................................................................................................................................4
4.0 INSTALLATION....................................................................................................................................................................................5
5.0 INITIAL WARM-UP AND OPERATION......................................................................................................................................7
5.1 QUICK FIELD TESTS..............................................................................................................................................................................8
6.0 CALIBRATION......................................................................................................................................................................................9
7.0 MAINTENANCE...................................................................................................................................................................................10
7.1 ROUTINE MAINTENANCE ...................................................................................................................................................................10
7.2 SENSOR CABLE REPLACEMENT.........................................................................................................................................................10
7.3 HYDROCARBON FILTER REPLACEMENT ..........................................................................................................................................10
7.4 REPLACEMENT PARTS.........................................................................................................................................................................11
8.0 SPECIFICATIONS ..............................................................................................................................................................................12
8.1 TYPICAL CALIBRATIONS .....................................................................................................................................................................12
8.2 OPTIONS AND V ARIATIONS................................................................................................................................................................12
9.0 WARRANTY ..........................................................................................................................................................................................13
9.1REPAIR OF CERTAIN UNITS.................................................................................................................................................................13
List of Illustrations
Figure 1: Electronics Control Unit................................................................................................................................................................2
Figure 2: High Pressure Regulator................................................................................................................................................................2
Figure 3: Utilizing the Relay Contacts.........................................................................................................................................................3
Figure 5: Internal Electronic Circuitry .........................................................................................................................................................3
Figure 4: Sample Head Assembly ................................................................................................................................................................. 4
Figure 6: Connecting to Compressed Air Line ........................................................................................................................................... 6
Figure 7: Calibration Adapter and Gas ........................................................................................................................................................8
ISA-RAL-M, ISA-RAH-M ENMET Corporation
1.01.0 IntroductionIntroduction
Sandblasting, welding, and spray painting are examples of activities that create potentially hazardous environments. Such environments contain toxic gases, dust and fumes. An air compressor system with hoods or masks supplies clean breathing air to workers and minimizes the risk of injury or death; however, carbon monoxide (CO) can contaminate the air in a compressor system. Carbon monoxide contamination can occur either when the intake air itself is contaminated, or when the compressor breaks down and overheats. Because of this hazard, Federal OSHA regulations (Sec. 1910.134) require monitoring of CO in compressed air systems. In addition, the Compressed Gas Association has set up guidelines governing maximum tolerable levels of CO and other contaminants (Spec. G-7.1) to ensure the health and safety of persons who use respiratory air lines.
The ENMET ISA-RAL-M (See Figure 1) monitors the carbon monoxide in compressed air used for human breathing. If the CO concentration exceeds any predetermined levels, the ISA-RAL-M will triggers audio-visual alarms and activates any associated equipment to alert personnel in potential danger.
NOTE: All specifications stated in this manual may change without notice.
1.1 Operation Principle
The ISA-RAL-M is an all-solid state electronic gas-detecting instrument designed to continuously monitor the carbon monoxide level in compressed air. The sensing element is a gas-sensitive metallic-oxide semiconductor (MOS). A regulator-humidifier assembly collects and humidifies a small air sample from the respiratory air line and passes it over the sensor. When the level of CO in the air passing over the sensor increases to preset values, the instrument alarms with audio (horn) and visual (light) signals.
The ISA-RAL-M operates from single-phase 110 Vac, single-phase 220 Vac and/or 12 Vdc power and is equipped with relay contacts to provide additional alarm signals.
1.2 ISA-RAH-M High Pressure Monitor
The ISA-RAH-M is designed to monitor breathing air from cylinders at pressures up to 5000 psi. The equipment consists of an ISA-RAL-M with the addition of a high-pressure regulator at the sample air input as shown in Figure 2. The high pressure regulator must be plumbed into the high pressure line with an appropriate fitting; the output is connected to the ISA-RAL-M by means of a quick release pneumatic fitting. The operation of the ISA-RAH-M is identical with that of the ISA-RAL-M, and calibration is also identical; the high pressure regulator is removed from the sample input and the calibration procedures given in this manual are employed.
1.3 Hydrocarbon Filter
There are two versions of the ISA-RAL-M: P/N 04526-001 is supplied with a hydrocarbon filter P/N 04526-003 is supplied without a hydrocarbon filter
2.0 Upon Receipt2.0 Upon Receipt
1. Unpack and inspect your new ISA-RAL-M for damage, which might have occurred during shipment.
2. Make sure the instrument is calibrated according to your order.
3. Do not use damaged or incorrect equipment. If there is a problem, contact your distributor or ENMET personnel
immediately.
4. Read this manual thoroughly to assure proper handling, operation, and maintenance of your ISA-RAL-M.
1
ENMET Corporation ISA-RAL-M, ISA-RAH-M
Alarm Light
Switch Indicator Light 4 holes, .31 Dia.
3.0 Features, Definition of Parts3.0 Features, Definition of Parts
3.1 Electronics Control Unit
Figure 1 shows the control unit which contains the electronics circuitry and controls. Main features are listed below.
Power Light (Green)
Alarm Light (Amber)
Alarm Light (Red)
Audio Alarm Audio alarm (2900 Hz,95 dB at 2 ft.). The audio alarm is on when the unit is in alarm. Meter
Operation Switch A rotary switch, for OPERATE (normal operation), HORN OFF (disengage audio
Switch Indicator Light (Red)
Sensor Cable Oiltight cable. Connects the sample head assembly to the control unit
When this light is on, the unit is operating and is not in alarm. When this light is off, the unit is in alarm OR power to the unit has been interrupted.
Visual alarm (steady light). When this light is on, power to the unit is on and the unit is in low alarm.
Visual alarm (steady light). When this light is on, power to the unit is on and the unit is in high alarm.
A carbon monoxide (CO) concentration meter. The scale is marked in parts per million (ppm) CO. This is a non-linear device. Do not infer exact readings from unmarked regions of the meter scale.
alarm) or PURGE ON (sensor temperature control). PURGE ON to clean (purge) sensor surface of adsorbed contaminants (sensor hot).
Indicates OPERATION SWITCH is not in OPERATE position.
(length 8-10 feet). See Caution in section 4.0 Installation step 5.
(Red)
Alarm Light (Amber)
Power Light (Green)
Figure 1: Electronics Control Unit
(Red)
Operation Switch
Meter
Audio Alarm
Sensor Cable
Figure 2: High Pressure Regulator
2
ISA-RAL-M, ISA-RAH-M ENMET Corporation
Normally Common
110 Vac
Manual On/Off
Suggested wiring configuration for Utilizing Relay Contacts
Low Level Set
High Level Set
Meter Adjust
110 / 115
12 VDC
Internal Relay
TB3
TB2
TB1
K1
K2
3.2 Internal Electronic Circuitry
Figure 5 shows the circuit board and terminal blocks housed inside the hinged oiltight control unit. Specific relays and adjustments are defined:
Internal Relay Contacts Relays can be used to activate an external remote alarm signal when a
hazardous CO level is detected, or when the ac or dc power is interrupted. These are double-pole relays with terminals normally open, normally closed and common (see Figure 3 for relay contact hook-up). There is one relay for each alarm level.
Potentiometers: The unit has five potentiometers. Only 3 are used by the customer. These vary
critical circuit resistances and are essential to calibration procedures. These potentiometers are described below.
Meter Adjust To adjust and set the meter for appropriate gas response during calibration. Low Level Set To adjust the trip level of the amber gas alarm light and the audio alarm. High Level Set To adjust the trip level of the red gas alarm light and the audio alarm.
(manual on/off switch is suggested; user supplied component)
Contact
Closed Contact
Switch
Horn
Black
White
Power supply Outside
of Control Unit
Non-Latching Relay contacts:
Identified below is the non-powered (power to unit is off), alarm positions of the non-latching relay contacts.
Contacts
Terminal Block 2 see Figure 5
Relay 1 High Alarm Relay 2 Low Alarm
NOTE: N.C.= normally closed
N.O.= normally open
Red Lamp ----
INPUT
Jumper Holder
Green Lamp
N.O.
COM.
Relay 1
Battery
Purge
Amber Lamp
N.C.
COM.
Relay 2
GND
Backup
Sensor
GND
N.O.
Signal
Meter+
N.C.
Heater
Meter–
GND
Potentiometer
Jumper 2
GND
GND
Jumper 3
Jumper 5
RV38
TP1
TP2
RV35
RV32
110 VAC
Jumper 4
Jumper 1
TB4
RV33
110 VAC
Potentiometer
RV34
Potentiometer
GND
OR
210 / 230* *See Section 4.0
Figure 3: Utilizing the Relay Contacts
Figure 5: Internal Electronic Circuitry
3
Loading...
+ 15 hidden pages