1.3 Check Order ................................................................................................................................................................. 1
1.4 Serial Numbers............................................................................................................................................................. 1
2.0 Instrument Features .......................................................................................................................................................... 2
2.1 ISA-60M Exterior Features .......................................................................................................................................... 2
2.2 Display ISA-60M Front Panel Features ....................................................................................................................... 2
2.3 MRI-5175 Features ...................................................................................................................................................... 2
2.3 Circuit Board Features ................................................................................................................................................. 4
3.2.4 Digital Outputs ...................................................................................................................................................... 8
3.3 Power Supply ............................................................................................................................................................... 9
4.1 Normal Operation Condition ......................................................................................................................................10
4.2 Alarm Set Points .........................................................................................................................................................10
4.7 Operational Menu .......................................................................................................................................................11
5.2 Maintenance Menu ISA-60M .....................................................................................................................................12
5.2.1 Set 4 –20mA Transmitter Scale ...........................................................................................................................13
5.2.3 Standard Calibration ............................................................................................................................................13
5.2.6 Set Alarm Points ..................................................................................................................................................15
5.2.7 Set Alarm “Delay On” .........................................................................................................................................17
5.2.10 Set New Password ..............................................................................................................................................18
5.2.11 Exit Maintenance Menu .....................................................................................................................................19
6.1.1 Standard Calibration ............................................................................................................................................20
7.0 Technical Data and Specifications ..................................................................................................................................22
8.0 Replacement Part Numbers .............................................................................................................................................23
Table 5: Fault LED Code Sequence ............................................................................................................................ 14
Table 3: Set up values ................................................................................................................................................. 14
[affects personnel safety – if not followed may cause bodily injury or death.]
Attention / Warning
!
Earth Ground
ISA-60MENMET Corporation
1.0 Introduction
The ISA-60M W/MRI-5175 is an ambient air monitoring system that measures oxygen concentrations in magnetic resonance
imaging environments. The ISA-60M control panel is capable of monitoring from 1 to 3 MRI-5175 remote sensors. Note: The
ISA-60M and MRI-5175 are produced as an aligned set and should be installed as such. Some features of the monitoring
system are as follows:
continuous monitoring of the sample air
continuous LCD display oxygen concentrations, at ISA-60M
menu driven operational and maintenance controls
menu driven calibration procedure
audio and visual alarms indicate unsafe conditions
alarm relay contacts available on terminals
a fault relay and visual fault alarm
alarm acknowledgement capability including audio defeat, ISA-60M control only
mA output for each channel
internal non-magnetic metallic coating of MRI-5175 to minimize RFI/EMI
N
OTE
: All specifications stated in this manual may change without notice.
1.1 IEC 60601-1 Classifications
Type of protection against electrical shock: Class 1.
Degree of protection against electrical shock: No Applied Parts
- Equipment not suitable for use in presence of flammable anesthetic mixture with air or with oxygen or nitrous oxide.
- Mains power quality should be that of a typical commercial or hospital environment.
- Power frequency magnetic fields should be at levels characteristic of a typical location in a commercial or hospital
environment.
1.2 Unpack
Unpack the
the commercial carrier involved immediately.
ISA-60M
and examine it for shipping damage. If such damage is observed, notify both ENMET customer service personnel and
Regarding Damaged Shipments
N
OTE
: It is your responsibility to follow these instructions. If they are not followed, the carrier will not honor any
claims for damage.
This shipment was carefully inspected, verified and properly packaged at our company and delivered to the carrier in good condition.
When it was picked up by the carrier at ENMET, it legally became your company’s property.
If your shipment arrives damaged:
•Keep the items, packing material, and carton “As Is.” Within 5 days of receipt, notify the carrier’s local office and request
immediate inspection of the carton and the contents.
•After the inspection and after you have received written acknowledgment of the damage from the carrier, contact ENMET
Customer Service for return authorization and further instructions. Have your Purchase Order and Sales Order numbers available.
ENMET either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability coverage, usually $100.00.
Repair or replacement charges above that value are your company’s responsibility.
The shipping company may offer optional insurance coverage. ENMET only insures shipments with the shipping company when
asked to do so in writing by our customer. If you need your shipments insured, please forward a written request to ENMET Customer
Service.
Regarding Shortages
If there are any shortages or questions regarding this shipment, please notify ENMET Customer Service within 5 days of receipt at the
following address:
ENMET Corporation
680 Fairfield Court
Ann Arbor, MI 48108
734-761-1270 734-761-3220 Fax
1.3 Check Order
Check, the contents of the shipment against the purchase order. Verify that the
the order, ascertain that they are present. Check the contents of calibration kits. Notify ENMET customer service personnel of any
discrepancy immediately.
ISA-60M
is received as ordered. If there are accessories on
1.4 Serial Numbers
Each
ISA-60M
and
MRI-5175
is serialized. These numbers are on tags on the equipment and are on record in an ENMET database.
1
ISA-60M ENMET Corporation
2.0 Instrument Features
2.1 ISA-60M Exterior Features
The exterior of the instrument is shown in Figure 1. The exterior features are as follows:
Feature Description
Enclosure
Strain Relief
Audio Alarm
Mounting Flanges
2.2 Display ISA-60M Front Panel Features
The display panel, shown in Figure 1, is viewed through the clear front cover of the enclosure, and is accessed by opening the
cover. Features are as follows:
Feature Description
Display
Visual
Alarms & Indicators
Pushbutton Switches
•OPTION Switch
•SELECT Switch
•
AUDIO DEFEAT
A
LARM ACKNOWLEDGE
Switch
/
Engineered thermoplastic, approximately 11x9x6, with a clear hinged front cover.
Entrance for wiring to Remote Sensor
A loud horn activated by certain alarm conditions.
Flanges with holes for mounting the enclosure to a vertical surface.
2 line, 16 character per line, LCD with backlight.
The numerical values of gas concentrations, and other information are displayed.
On either sides of the display:
A red alarm LED for each sensor transmitter installed to the instrument, Alarm 1.
The top center:
A red alarm LED for all sensor transmitters, Alarm 2.
Near the center of the panel:
A green power indicator LED
A red fault alarm LED
There are three of these, located near the center of the panel; they are yellow
rectangular membrane switches. They are:
The top left switch.
Directly to the right of the
OPTION
switch.
Directly below the
OPTION
switch.
2.3 MRI-5175 Features
The exterior and panel of the sensor/transmitter is shown in Figure 1. Internal features are shown in Figure 4.
Feature Description
Enclosure
Strain Relief
Mounting Flanges
MENU Switch
Fiberglass-reinforced polyester with non-magnetic RFI/EMI shielding and hardware
Nickel plated brass, entrance for wiring to ISA-60M control unit
Flanges with holes for mounting the enclosure to a vertical surface.
The left most pushbutton switch
Menu
SELECT Switch
The right most pushbutton switch
Select
Power/Fault LED
Visual Alarm LED
Sensor
Audio Alarm
Red/Green LED, Green indicating power on and red indicating a fault condition
3 Red LED, Triggered at level of alarm
Long life electrochemical oxygen sensor
A loud horn activated by certain alarm conditions.
(Optional)
2
ISA-60M ENMET Corporation
O
O
SELECT
OPTION
Alarm Defeat /
Visual Alarm
Mounting Flanges
Visual Alarm
5175
remote sensors are
Punch out/Hole Plug
Audio Alarm
Visual O2
MENU
SELECT
Sensor
Mounting Flanges
Menu
Select
SN:
-XXXX
88
Display
O
2
(
If additional MRI-
connected)
2
%
2
%
Switch
Alarm 2
Switch
Alarm Acknowledge
Switch
For External Wiring
2 places
2 places
Strain Relief
Power/Fault LED
Strain Relief
Alarm LED
Power LED
Front Cover Latch
Fault LED
Audio Alarm
(Nickel Plated Brass)
Power/Fault XXX XXX XXX
Switch
4 places
%O2 by vol.
Figure 1: ISA-60M & MRI-5175 External Features
3
Switch
(Optional)
ISA-60M ENMET Corporation
Power Input
Audio
Relay Terminals
Aux Terminals
Primary Detector
4 –
20mA Output
Digital Communication
Hole Plug
s
2.3 Circuit Board Features
The Display Panel is hinged on the left and is released by unscrewing the 2 philips screws located in the top and bottom right
corners. After releasing the panel, it is swung to the left, exposing the interior of the enclosure. The Circuit Board is mounted
on a plate at the back surface of the enclosure interior. Features are shown in Figure 2.
Feature Description
Relay Terminals
Output Terminals
(6 places)
This group of terminals is located at the left side of the Circuit Board.
For the contacts for each of four alarm relays, and for the contacts of a fault relay.
There are two for each of the 4-20 mA outputs.
Terminals
4 outputs
2 connectors
Channel 1
Optional (2 places)
Fuse Holder
(2 places)
Ground Screw J21
Terminal J23
Terminals (3 types)
Horn Terminal
Alarm
Strain Relief
Figure 2: ISA-60M Interior Features
4
ISA-60M ENMET Corporation
Mounting Holes
Mounting Holes R 0.28” dia.
Menu
Select
SN:
-XXXX
3.0 Installation
3.1 Mounting of Instrument
Mount the ISA-60M and MRI-5175 instrument on an appropriate vertical surface using the mounting flanges provided. Avoid
areas with excessive vibration or temperature extremes. See Figure 3.
It is recommended to use #8 drywall anchors and screws for mounting the ISA-60M and MRI-5175 to a drywall/sheetrock
surface.
Non-magnetic mounting hardware is required for control panels located in the magnet room.
!
Typical location of Strain Relief
Front View
4 places
Side View
0.31” dia. 4 places
Dimensions are in inches.
Figure 3: ISA-60M C
ONTROL &
MRI-5175 R
EMOTE SENSOR
Mounting Dimensions
N
OTE
: Remote Sensor, Each must be connected to a 4-20mA input inside the ISA-60M. UL listed type CM, PLTC or TC cable
shall be used for wiring between the MRI-5175 and the ISA-60M control panel. Three conductor, 18 gauge
for runs up to 1000 feet. For 500 feet or shorter wiring runs, 20 gauge
shielded
wiring is acceptable. The supplied strain
shielded
reliefs are rated for cable with an OD between 0.20 – 0.35 inches.
Refer to Section 7 for Technical Instillation Data.
wiring
5
ISA-60M ENMET Corporation
Channel 1, J16
Channel 1, J4
*
Channel 2,
J18
Channel 2,
J4
Channel 3, J19
Channel 3, J4
D–
J4
Shielding wire to Ground S
crew
3.2 Wiring ISA-60M to MRI-5175
Wiring for the MRI-5175 alarm, relay contacts and 4-20mA outputs may be routed through one of the two enclosure punchouts on the side or bottom of the enclosure. See Figure 1. Follow state and local guidelines when selecting appropriate strain
relief, conduit and wiring.
Remote Sensor: Each must be connected to a 4-20mA input inside the ISA-60M. UL listed type CM, PLTC or TC cable shall
be used for wiring between the MRI-5175 and the ISA-60M control panel. Three conductor, 18 gauge
runs up to 1000 feet. For 500 feet or shorter
shielded
wiring runs, 20 gauge wiring is acceptable. The supplied strain reliefs
are rated for cable with an OD between 0.20 – 0.35 inches. Refer to Section 7.0 for typical wiring information.
C
AUTION
!
:
Be careful not to scratch or otherwise damage the internal metallic coating of the MRI-5175 during
installation.
Table 1: Wiring ISA-60M to MRI-5175
ISA-60M
D+
mA
GND
24V
+24V (1)
GND (2)
mA (3)
+24V
GND
mA
Power
Ground
Signal
Power
Ground
Signal
*
(Optional)
MRI-5175 Lid, interior view
+24V
GND
mA
Power
Ground
Signal
*
*Shielding wire to Ground Screw of MRI-5175 Ground Screw
shielded
MRI-5175
+24V
GND
mA
+24V
GND
mA
+24V
GND
mA
wiring for
Shield Wire
Ground Screw
MRI-5175 enclosure, interior view
Figure 4: MRI-5175 Ground for Shielded Cable
of MRI-5175 enclosure
6
Loading...
+ 19 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.