2.0 UPON RECEIPT.....................................................................................................................................................................................3
2.2 CHECK ORDER ....................................................................................................................................................................................... 3
2.3 SERIAL NUMBERS..................................................................................................................................................................................3
3.0 FEATURES / DEFINITION OF PARTS.........................................................................................................................................4
3.1 ELECTRONICS CONTROL UNIT ............................................................................................................................................................4
3.2.1 CO Circuit......................................................................................................................................................................................7
3.3 SAMPLE HEAD ASSEMBLY...................................................................................................................................................................8
5.0 INITIAL WARM UP AND OPERATION.....................................................................................................................................10
5.3 QUICK FIELD TESTS............................................................................................................................................................................11
5.3.1 Carbon Monoxide Detection Channel Field Test.................................................................................................................11
5.3.2 Air Quality and Pressure Adjustments....................................................................................................................................12
8.2 OPTIONS AND V ARIATIONS................................................................................................................................................................18
FIGURE 1A: SAMPLE HEAD ASSEMBLY, TYPICAL H IGH PRESSURE HOOK-UP................................................................................2
FIGURE 1: CONTROL U NIT FEATURES ....................................................................................................................................................... 5
FIGURE 2: CIRCUIT BOARD WITH TERMINAL BLOCKS..........................................................................................................................6
FIGURE 3: RELAY HOOK-UP, POWER INTERNAL AND EXTERNAL.......................................................................................................7
FIGURE 4: SAMPLE HEAD ASSEMBLY, TYPICAL H OOK-UP ..................................................................................................................8
TABLE 2: CO ALARM RELAY CONTACTS .................................................................................................................................................9
TABLE 3: REFERENCE, STATE OF OPERATION OF ISA-44RAL-OD / ISA-44RAH-OD .............................................................. 10
FIGURE 5: TEST G AS HOOK-UP.................................................................................................................................................................13
TABLE 4: REPLACEMENT PART NUMBERS .............................................................................................................................................16
TABLE 5: CALIBRATION G AS PART NUMBERS.......................................................................................................................................16
FIGURE 6: HOUSING, OXYGEN CELL AND MOS SENSOR....................................................................................................................17
Reference information:
NOTE: [important information about use of instrument – if not followed may have to redo some steps.]
CAUTION: [affects equipment – if not followed may cause damage to instrument, sensor etc…]
WARNING: [affects personnel safety – if not followed may cause bodily injury or death.]
Page 5
ISA-44RAL-OD / ISA-44RAH-ODENMET Corporation
1.0 Introduction1.0 Introduction
Sandblasting, welding and spray painting are examples of activities that create potentially hazardous environments.
Such environments contain toxic gases, dust and fumes. An air compressor system with hoods or masks supplies clean
breathing air to workers and minimizes the risk of injury or death; however, carbon monoxide (CO) can contaminate
the air in a compressor system. Carbon monoxide contamination can occur either when the intake air itself is
contaminated, or when the compressor breaks down and overheats. Because of this hazard, Federal OSHA regulations
(Sec. 1910.34) require monitoring of CO in compressed air systems. In addition, the Compressed Gas Associations
has set up guidelines governing maximum tolerable levels of CO and other contaminants (Spec. G-7.1) to ensure the
health and safety of persons who use respiratory air lines.
The ENMET ISA-44RAL-OD monitors the carbon monoxide and oxygen levels in compressed air used for human
breathing. If the CO concentration exceeds any predetermined levels, the ISA-44RAL-OD will trigger audio-visual
alarms and activate any associated equipment to alert personnel in potential danger. In addition, if the oxygen
concentration falls below a predetermined level, the instrument will again alarm in the same manner.
This instrument is designed to operate continuously; there is no on/off switch. Therefore, the ISA-44RAL-OD should
be left on 24 hours a day. The instrument should not be unplugged at night or at the end of a workshift. In addition,
air must flow over the sensors continuously whenever the unit is in operation.
1.1 Operating Principle
The ISA-44RAL-OD is an all solid-state electronic gas-detecting instrument designed to continuously monitor the
carbon monoxide and oxygen levels in compressed air. The CO sensing element is a gas-sensitive metallic oxide
semiconductor (MOS sensor). A regulator humidifier assembly collects and humidifies a small air sample from the
respiratory air line and passes it over the sensor. If the level of CO in the air passing over the sensor increases to
preset value, the ISA-44RAL-OD alarms with audio (audio alarm) and/or visual (light) signals.
The sensor for the oxygen channel is a micro-fuel cell. It employs a galvanic reaction to emit a steady current. This
current indicates the presence of oxygen to the circuitry. If the oxygen level drops, the current emitted also drops
proportionally. When the current from the cell falls below a predetermined safe level, the alarm circuitry triggers.
The sensing elements of the ISA-44RAL-OD are sensitive to large scale temperature changes. Although this is usually
not a problem with compressed air lines, if the air sample which passes over the sensors is subject to large temperature
changes, the ISA-44RAL-OD may go into false alarms.
The percent oxygen (% oxygen) meter is a linear device. All regions of the meter scale represent approximate
concentrations. It differs from the ppm carbon monoxide (ppm CO) meter, which is not linear.
The ISA-44RAL-OD operates from single-phase 115Vac and/or 12Vdc power and is equipped with relay contacts to
provide additional alarm signals. Operation on single-phase 220 Vac requires several circuitry changes. Contact
ENMET or your distributor for this information.
1
Page 6
ENMET CorporationISA-44RAL-OD / ISA-44RAH-OD
Quick Release
Oxygen Cell
Hydrocarbon
Pressure
Humidifier
Metering
Cable to
High Pressure
N
:
1.2 ISA-44RAH-OD
The ISA-44RAH-OD is designed to monitor breathing air from cylinders at pressures up to 5000 psi. The
equipment consists of an ISA-44RAL-OD with the addition of a high pressure regulator at the sample air input as
shown in Figure 1a. The high pressure regulator must be plugged into the high pressure line with an appropriate
fitting; the output is connected to the ISA-44RAH-OD by means of a quick release pneumatic fitting. The
operation of the ISA-44RAH-OD is identical with that of the ISA-44RAL-OD, and calibration is also identical; the
high pressure regulator is removed from the sample input and the calibration procedures given in this manual are
employed.
Orifice
Filter
Regulator
OTE
The high-pressure regulator
is not included in the
ISA-44RAL-OD,
only in the ISA-44RAH-OD
Figure 1a: Sample Head Assembly, Typical High Pressure Hook-up
Fittings
Regulator
Water Level mark)
and
MOS Sensor
(keep filled to
enclosure
03042-109 Angle Style
Sensor Cable
03042-109 Straight Style
Sensor Cable
2
Page 7
ISA-44RAL-OD / ISA-44RAH-ODENMET Corporation
2.0 Upon Receipt2.0 Upon Receipt
2.1 Unpack
Unpack the ISA-44RAL(RAH)OD and examine it for shipping damage. If such damage is observed, notify both
ENMET customer service personnel and the commercial carrier involved immediately.
Regarding Damaged Shipments
NOTE: It is your responsibility to follow these instructions. If they are not followed, the carrier will
not honor any claims for damage.
q This shipment was carefully inspected, verified and properly packaged at our company and delivered to the carrier
in good condition.
q When it was picked up by the carrier at ENMET, it legally became your company’s property.
q If your shipment arrives damaged:
•Keep the items, packing material, and carton “As Is.” Within 5 days of receipt, notify the carrier’s local office
and request immediate inspection of the carton and the contents.
•After the inspection and after you have received written acknowledgment of the damage from the carrier,
contact ENMET Customer Service for return authorization and further instructions. Have your Purchase Order
and Sales Order numbers available.
q ENMET either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability
coverage, usually $100.00. Repair or replacement charges above that value are your company’s responsibility.
q The shipping company may offer optional insurance coverage. ENMET only insures shipments with the shipping
company when asked to do so in writing by our customer. If you need your shipments insured, please forward a
written request to ENMET Customer Service.
Regarding Shortages
If there are any shortages or questions regarding this shipment, please notify ENMET Customer Service within 5 days
of receipt at the following address:
ENMET Corporation
680 Fairfield Court
Ann Arbor, MI 48108
734-761-1270734-761-3220 Fax
2.2 Check Order
Check, the contents of the shipment against the purchase order. Verify that the ISA-44RAL-OD / ISA-44RAH-OD is
received as ordered. If there are accessories on the order, ascertain that they are present. Check the contents of
calibration kits. Notify ENMET customer service personnel of any discrepancy immediately.
2.3 Serial Numbers
Each ISA-44RAL-OD / ISA-44RAH-OD is serialized. These numbers are on tags on the equipment and are on record
in an ENMET database.
3
Page 8
ENMET CorporationISA-44RAL-OD / ISA-44RAH-OD
3.0 Features / Definition of Parts3.0 Features / Definition of Parts
3.1 Electronics Control Unit
3.1.1 Carbon Monoxide Detection Channel
Refer to Figure 1
FEATUREDESCRIPTION
POWER LIGHT (G REEN)Light on: Channel operating
Light off: Alarm condition or power interrupted.
ALARM LIGHTSVisual alarms (steady lights).
AMBER: unit in low level alarm.
RED: unit in high level alarm.
AUDIO ALARMAudio alarm: activates when high alarm level is exceeded.
METERA CO concentration meter. The scale is marked in parts per million (ppm) CO.
This is a non-linear device. Do Not Infer Exact Readings From Unmarked
Regions Of The Meter Scale.
OPERATION SWITCH *A rotary switch.
Operate: normal MOS sensor operation
Purge: Purges (cleans) MOS sensor by heating it to very high temperatures.
Sensor inoperative as a CO detector.
Horn Off : Deactivates audio alarm during calibration and maintenance.
CABLEOiltight. Connects the sample head assembly to the control unit.
3.1.2 Oxygen Detection Channel
Refer to Figure 1
FEATUREDESCRIPTION
POWER LIGHT (GREEN )Light on: channel operating.
Light off: alarm condition or power interrupted.
ALARM LIGHT (AMBER)Visual alarm (steady light).
Light on: channel in alarm.
AUDIO ALARMAudio alarm; activates when oxygen channel goes into alarm (amber light).
METEROxygen concentration meter. The scale is marked in % oxygen. The oxygen
meter scale is linear - the points on the meter accurately correspond with the
oxygen concentration in the monitored air
OXYGEN GAIN A DJUSTRotary component which adjusts the meter reading during calibration, testing or
sensor replacement.
OPERATION SWITCH *Same as for CO channel.
CABLESame as for CO channel.
* Older equipment has separate PURGE and HORN OFF switches .
4
Page 9
ISA-44RAL-OD / ISA-44RAH-ODENMET Corporation
Oxygen Power
Oxygen Alarm
Oxygen Gain
CO Low Alarm
CO Alarm High
CO Power
Cable
Light (Amber)
Light (Amber)
Light (Green)
Light (Red)
Light (Green)
CO Meter
Audio Alarm
Adjustment
Oxygen Meter
Operation Switch
Figure 1: Control Unit Features
5
Page 10
ENMET CorporationISA-44RAL-OD / ISA-44RAH-OD
Relay
Relay
Relay
TP4
Relay
Relay
115 VAC
Terminal
Case
Ground
+
–
–
+
DC Ground
Purge
High Level
Low Level
Meter
Purge Adjust
Heater
O2 Alarm Set
Meter Full Scale Adjust
Null Adjust
COM
COM
–
+
Green
White
Black
3.2 Internal Electronic Circuitry
Figure 2 shows the circuit board and terminal blocks housed inside the hinged oiltight control unit. Specific relays and
adjustments are defined:
INTERNAL RELAY CONTACTS: Relays can be used to activate an external remote alarm system when a hazardous CO level
is detected. These are double-pole relays (Figure 2 upper right) with terminals "normally open"(N.O.), "normally
closed"(N.C.), and "common"(COM.) (See Figure 3 and Section 4.0 for relay contact hook-up). There is one relay for
each alarm level (2 alarm levels for CO, 1 for oxygen).
NOTE:Relay contacts are shown in the de-energized state (power off).
NOTE:If you want the relays to activate auxiliary equipment when the power to the control unit is interrupted, you
must connect the relays to an external power source. See Figure 3a.
POTENTIOMETERS: The unit has 9 potentiometers. These adjust critical circuit resistances and are essential to calibration
procedures.
CO Meter
CO Meter
MOS Sensor
+12 VDC
O2 Meter
O2 Meter
DC Ground
TB2
Fuse
COM
Alarm Set
Adjust
Green
Alarm Set
DC-Gnd
Adjust
Amber
TB3
Red
Low LevelLights
TP5
N.O.
K 3
High Level
N.C.
K 2
N.O.
K 1
TP2
TP1
N.C.
TB1
Low Level Set
TP3
White
Power
Green
Cover
O2 Cell
Amber
COM
N.O.
O2 Relay
N.C.
Black
Figure 2: Circuit Board with Terminal Blocks
6
Page 11
ISA-44RAL-OD / ISA-44RAH-ODENMET Corporation
Auxiliary Horn
Install a Separate
Neutral
Install a Separate
an Auxiliary Alarm
110 VAC
Neutral
3.2.1 CO Circuit
Refer to Figure 2
PURGE ADJUST
POTENTIOMETER
HEATER A DJUST
POTENTIOMETER
Adjusts purge voltage. Operation switch must be in purge position to adjust
purge voltage.
For adjusting the MOS sensor heater operating voltage. DO NOT adjust this
potentiometer unless a voltage reading indicates that an adjustment is necessary.
(May need adjusting after replacing sensor cable assembly) . Voltage: .86 Vdc
+ .03 Vdc
METER ADJUST
To adjust and set the meter for appropriate gas response during calibration.
POTENTIOMETER
LOW LEVEL ALARM S ET
For adjusting the trip level of the CO low level alarm light.
ADJUST POTENTIOMETER
HIGH LEVEL ALARM SET
For adjusting the trip level of the CO high level alarm light and the audio alarm.
ADJUST POTENTIOMETER
3.2.2 Oxygen Circuit
Refer to Figure 2.
OXYGEN ALARM SETFor adjusting the trip level of the audio alarm and the amber oxygen alarm light.
NULL ADJUSTFor adjusting oxygen circuit voltage during calibration.
METER FULL SCALE ADJUSTFor adjusting full scale meter reading during calibration.
LOW LEVEL S ETFor adjusting oxygen circuit voltage during calibration.
Figure 3: Relay Hook-up, Power Internal and External
7
Page 12
ENMET CorporationISA-44RAL-OD / ISA-44RAH-OD
Locate the Sample
Oxygen Cell
Hydrocarbon
Pressure
Quick Release
Fittings
Humidifier
Metering
Top Mount to
Locate near compressor
Plant Air
Cut-Off Valve
Cable to
3.3 Sample Head Assembly
The sample head connects to a compressed air line. This unit continuously monitors the respiratory lines for carbon
monoxide and oxygen. Figure 4 illustrates this unit. Assembly parts are listed and defined below:
PRESSURE REGULATOR:
QUICK RELEASE PNEUMATIC FITTING :
HUMIDIFIER :
METALLIC OXIDE SENSOR
(MOS Sensor):
METERING ORIFICE :
ELECTROCHEMICAL SENSOR:
Reduces the pressure of sample air from the compressed air line before passing over
the sensor. It includes a meter to measure the pressure in pounds per square inch (psi)
at the regulator exit.
For quick connect/disconnect; fits into the port of an air line or filter unit and an
ENMET calibration gas fixture. Air inlet fitting is a Milton 727, which is compatible
with Hansen 1000 series. This fitting also connects with the exit fitting of the high
pressure regulator on an ISA-44RAH-OD. If you substitute other fittings, the
calibration fixture must also be changed.
This adds moisture to the air sample that passes over the sensor, which is a
requirement for correct sensor operation.
Solid-state gas sensing element, contained in the sensor housing. When the amount
of CO in the air passing over the sensor increases to a preset level, the equipment
alarms and relays activate.
This replaceable assembly has a tiny orifice (hole) to ensure constant flow of air
between the humidifier and pressure regulator.
Oxygen sensing element, contained in the sensor housing.
Hydrocarbon FilterTo enhance low level calibrations by removal of hydrocarbon vapors from the air
sample.
Orifice
(optional)
Avoid Water Trap
Filter
holding tank so sample air
reaches the sensor before
the same respiratory air
reaches the workers.
Point after Purification
Regulator Panel.
50 – 250 PSIG
Regulator
(keep filled to
Water Level mark)
enclosure
and
MOS Sensor
Figure 4: Sample Head Assembly, Typical Hook-up
8
Page 13
ISA-44RAL-OD / ISA-44RAH-ODENMET Corporation
4.0 Installation4.0 Installation
After becoming familiar with the main parts and their functions, follow the installation procedure as outlined below:
1. Choose an appropriate location for mounting the electronics control unit. Do not locate the sampling head assemble
in an area where quick temperature changes are likely to occur, like near an outside door. The oxygen cell is
sensitive to temperature and the circuitry for this cell cannot track rapid changes in temperature. A rapid change in
temperature may cause nuisance alarms, especially for those units calibrated to alarm at a percent near the normal
fresh air volume (20.9%).
2. The three terminal block positions shown in Table 1 below (TB3-4, TB3-5, TB3-6) are attached to relay K3 (see
Figure 2) and are available as a switch to activate any remote alarms or equipment when the oxygen alarm triggers.
This relay has a rating of 2 amps continuos, 5 amps surge at 110Vac. When power is interrupted, this relay switches
to the same position as in an oxygen-deficient situation.
3. The six terminal block positions shown in Table 2 below (TB1-5 - TB1-10) are attached to the two CO alarm relays, K-
1 and K-2 (see Figure 2). These positions are available as switches to activate any remote alarms or equipment when
either the CO low alarm or CO high alarm triggers. These relays have the rating of 2 Amps continuous, 5 Amps surge
at 110Vac. When power is interrupted, these relays switch to the same position as in a gas alarm situation.
NOTE:K-1 is the relay for the CO low alarm.
K-2 is the relay for the CO high alarm.
Table 2: CO Alarm Relay Contacts
PositionFunctionRelay
TB1-5CommonK-2 CO low alarm
TB1-6Normally OpenK-2 CO Low alarm
TB1-7Normally ClosedK-2 CO Low alarm
TB1-8CommonK-1 CO High alarm
TB1-9Normally OpenK-1 CO High alarm
TB1-10Normally ClosedK-1 CO High alarm
4.Run the relay contact leads through the oiltight fitting for the power supply leads. If you want to run the relay
contact leads through another part of the enclosure, drill a hole in the side or bottom of the enclosure and use a
NEMA 4X oiltight fitting.
NOTE: The ISA-RAL-OD is completely oiltight. Use the correct oiltight fitting for the cord or conduit when supplying
power to the unit.
NOTE: The ISA-44RAL-OD is engineered for respiratory air lines having 50 - 250 PSI. If your air line has a higher
pressure, you must use the ISA-44RAH-OD (see Section 1.2).
NOTE: If used in conjunction with an ENMET air filtration panel, the ISA-44RAL-OD can be mounted on the air filtration
panel mounting plate. Attach the RAL-OD assembly to the quick disconnect fitting on the air filtration unit.
9
Page 14
ENMET CorporationISA-44RAL-OD / ISA-44RAH-OD
5.Apply 115 Vac and/or 12 Vdc power to the appropriate terminals (refer to Figure 2). As an emergency back-up
power source, both ac and dc power can be applied at the same time. Current will flow from the ac source. DC
current will flow only when the ac power is interrupted.
NOTE: 220 Vac power requires a change in the transformer hook-up. This c hange is best done at ENMET or an ENMET
service center.
6.Attach the sample head of the ISA-44RAL-OD to the respiratory air line so that the sample air reaches the sensors
before the same respiratory air reaches the workers. If you have any problems involving the location of the unit,
contact your distributor or ENMET personnel. A technician will analyze the problem and provide recommendations
for a location.
CAUTION: Do not use a rubber hose to connect regulator to air line. This will cause inaccurate, upscale meter readings.
CAUTION: Do not locate the sample point on a low spot in the line. Water condensation may clog or damage the
regulator or sensor assembly. If you must locate the sample point at a low spot, install a water trap between
the air line and the sample head assembly.
NOTE:Because the ISA-44RAL-OD will be operating continuously, make sure that air is flowing over the sensors
continuously. Sensors exposed to stagnant or non-moving air may overheat or become contaminated.
NOTE:If press ure is applied to the regulator input, and air is not bubbling through the humidifier, the metering orifice
is probably plugged; you should replace it (Section 7.1).
7.Add tap water to the humidifier up to the Water Level mark.
8.Adjust the pressure regulator to align the gauge needle with the black dots marked"...SET" or between 7 – 12 PSI on
the dial.
9.Proceed to Section 5.0.
5.0 Initial Warm Up and Operation5.0 Initial Warm Up and Operation
5.1 Carbon Monoxide Detection Channel
Mount and install the ISA-44RAL-OD as outlined in Section 4.0.
NOTE:The unit has no ON/OFF switch. This safety monitoring device is designed to be powered and on at all times.
1. Rotate the operation switch to PURGE ON. This will silence the audio alarm, both alarm lights may go on soon; don't
worry, this is normal. They will turn off after awhile.
2.While operation switch is in PURGE ON position, measure purge voltage across TB2-6 and TB2-7 . It should be 1.6
Vdc +/- .03 Vdc. If not 1.6 Vdc +/- .03 Vdc, adjust Purge Adjust Potentiometer (see Figure 2) until desired reading is
shown. Turn operation switch to HORN OFF position and measure sensor heater voltage across TB2-6 and TB2-7. It
should now read 0.86 Vdc +/ - .03 Vdc. If not 0.86 Vdc +/- .03 Vdc, adjust Heater Adjust Potentiometer (see Figure
2) until desired reading is shown. Turn operation switch back to PURGE ON position.
3.Let clean air pass over the sensors for at least one hour. Clean Air Must Pass Over The Sensor . If you suspect
contaminated air or dirty conditions in your compressor, use bottled clean air. DO NOT operate the unit without air
passing over the sensor. If you do, you may contaminate or burn out the sensor.
4.The sensor is purged when the amber alarm light goes out, and the meter has stabilized.
5.Turn operation switch to HORN OFF position. Some meter drift is normal. The unit may show both green and red
lights momentarily at this point. This is also normal. After the sensor becomes stable, the red light will deactivate
and the green light will remain on, which is the normal operating state.
6. Turn the operation switch to OPERATE position.
Table 3: Reference, State of Operation of ISA-44RAL-OD / ISA-44RAH-OD
Normal operating state
Alarm state
Sensor fault or disconnected
• No alarms
• Green light on
• Green light off
• Amber light (low level alarm) or
• Amber light, Red light and Audio Alarm (high level alarm)
• Appropriate relay contacts in alarm position
• Green and red light on together.
10
Page 15
ISA-44RAL-OD / ISA-44RAH-ODENMET Corporation
5.2 Oxygen Detection Channel
After installation is completed, the oxygen meter must be adjusted to a clean air reference point of 20.9% oxygen by
volume. The equipment necessary for this is included in a calibration kit for ISA-44RAL-OD / RAH-OD units, ENMET
part number 03430-001.
1.Turn operation switch to HORN OFF position.
2.Connect the calibration fixture to the bottled clean air (20.9% oxygen) and attach the calibration fixture to the quickrelease fitting on the low pressure regulator. If you have an ISA-44RAH-OD unit, first disconnect the high-pressure
regulator attached to the incoming air line until you are finished.
3.Open the small regulator on the calibration fixture to release clean air into the sampling head assembly. Keep
opening the small regulator until the needle on the pressure guage rises to the "...SET" mark or between 7 – 12 PSI.
4.After exposing the oxygen cell to clean air for 3-5 minutes (use bottled 20.9% oxygen if necessary), dial down the
oxygen Gain Potentiometer on the front panel of the unit so that the meter needle passes the alarm point. At that
point, the amber light should go on. If the alarm does not go on at the specified oxygen alarm point, recalibrate the
oxygen channel (Section 6.2). In addition, while the unit is in alarm, turn the horn switch on briefly to verify that the
audio alarm works.
5.Dial the oxygen Gain Potentiometer up again and leave the meter at 20.9%. Before closing the small regulator on the
calibration fixture and detaching it from both the low pressure regulator and the clean air cylinder, make sure the
pressure is equalized (regulator for sample head reads zero). If you do not wait until the pressure is equalized, the
pressure will cause a backflow of water from the humidifier, which will damage the sensor.
6.The initial adjustment is complete. Turn the audio alarm back on and reattach the high pressure regulator if you have
one.
5.3 Quick Field Tests
5.3.1 Carbon Monoxide Detection Channel Field Test
To Verify The Instrument's Response Capability: Use a cylinder of high-level CO/air mixture (for example, 200 ppm
CO for a 20 ppm CO calibrated unit) with a quick-disconnect gas fixture. See Figure 5.
1.Turn operation switch to HORN OFF position.
2.Remove the compressed air line from the sample head assembly. Insert the test gas.
3.Open the small regulator on the calibration fixture to release clean air into the sampling head assembly. Keep
opening the small regulator until the needle on the pressure gauge rises to the "...SET" mark or between 7 – 12 PSI.
4.Meter should move upscale in less than 20 seconds and read full-scale withi n one minute.
Whenever you use a test gas containing a level of gas greater than the calibration level, the meter swings upscale, green
power light goes out, red alarm light goes on, and relay contacts "close" to trigger any remote alarms being used.
This Test Is To Verify Response Capability; It Does Not Imply Correct Calibration. If The Test Does Not Bring
The Unit Into Alarm, Contact ENMET Personnel Or An Authorized Service Center.
11
Page 16
ENMET CorporationISA-44RAL-OD / ISA-44RAH-OD
5.3.2 Air Quality and Pressure Adjustments
For the OXYGEN DETECTION CHANNEL: Atmospheric conditions and conditions inside your air compressor are
constantly changing. In order to maintain a clean oxygen reading of 20.9% oxygen the oxygen Gain Potentiometer must
be adjusted periodically. Check the oxygen meter daily. If it continually shows a percentage other than 20.9% for 15
minutes in the presence of compressor air, double check that the compressor air is really clean by exposing the cell to
bottled clean air from a pressurized cylinder. During exposure to bottled air, if the meter still does not read 20.9%
(especially if it is higher), adjust the oxygen Gain to reference clean air.
The electrochemical oxygen cell current output is directly proportional to the partial pressure of the oxygen in the
atmosphere surrounding the cell. The partial pressure of oxygen is a function of the percent by volume of oxygen in the
air and the air pressure.
Air pressure varies as natural weather systems move through the area, causing changes in barometric pressure as well as
oxygen content. The barometric pressure changes cause a relatively small change in oxygen content indication; for
example, if the gain is set to give an oxygen content meter indication of 20.9% oxygen when the barometric pressure is
29.9 inches of mercury, normal variations of barometric pressure cause variations in meter readings of + 0.25%, from
20.75% to 21.15%.
If the gain is set during a normal high pressure weather cycle, the variation is 0.50% downscale; conversely, if the gain is
set during a normal low, the variation is upscale.
This response to atmospheric pressure is not distressing when understood. The variation of the alarm point by + 0.25% is
not significant when the liberal safety factor between the alarm point, usually 19.5%, and the point at which oxygen
deficiency first emphatically affects human performance, approximately 16%, is taken into consideration.
PRECAUTIONS
DO NOT: expose the oxygen cell to temperatures below 32 F (0 C) or above 130 F (55 C).
NEVER : adjust the oxygen cell when the air passing over it may be oxygen deficient. Always adjust the cell in the
presence of CLEAN AIR.
AVOID: gross exposure to toxic gases. Pure gases, liquid hydrocarbons (butane, propane, gasoline etc.), lacquer fumes,
and cigarette smoke should be kept away from the oxygen cell.
If you wish to recalibrate to a different alarm point, consult an authorized ENMET service center or ENMET personnel
directly.
5.4 Humidifier
Periodically check the humidifier water level, and add water when required to maintain proper water level. Operating
the equipment with a dry air sample and no water in the humidifier bowl renders the carbon monoxide sensor
ineffectual.
12
Page 17
ISA-44RAL-OD / ISA-44RAH-ODENMET Corporation
ENMET
Calibration
Adapter
Pneumatic
6.0 Calibration6.0 Calibration
6.1 Carbon Monoxide Detection Channel Calibration
The following procedure is for checking the calibration level and for recalibrating the Carbon Monoxide Channel of the
ISA-44RAL-OD. Refer to Figure 5 and Figure 2.
Perform A Complete Recalibration Periodically (at least monthly) and:
• When you replace the sensor.
• When the sensor encounters severe conditions (i.e., gross exposure to smoke or liquid hydrocarbon).
MATERIALS
•Cylinders of calibration gas of known CO concentration in air (i.e. 20 ppm and 50 ppm CO in air.).
DO NOT use CO in an inert carrier gas, such as nitrogen or argon.
• Calibration gas fixture (with quick-disconnect fitting – ENMET part number 03605-001)
• Small screwdriver (for potentiometer adjustment)
• dc voltmeter
Quick Release
Fittings
Test/Calibration
Gas
Figure 5: Test Gas Hook-up
PROCEDURE
1.Turn operation switch to HORN OFF. Be sure humidifier bottle is filled with tap water to Water Level mark.
2.Turn operation switch to PURGE ON. Measure purge voltage across TB-6(+) and TB2-7(-). It should be 1.6 Vdc +/.03 Vdc. If not 1.6 Vdc +/- .03 Vdc, adjust Purge Adjust Potentiometer (see Figure 2) until desired reading is shown.
Turn operation switch to HORN OFF and measure sensor heater voltage across TB2-6 and TB2-7. It should now read
0.86 Vdc +/ - .03 Vdc. If it is not 0.86 Vdc +/- .03 Vdc, adjust Heater Adjust Potentiometer (see Figure 2) until
desired reading is shown.
3.Turn operation switch back to PURGE ON. Purge at least one hour, up to overnight.
REMEMBER : If you suspect contaminated air in your compressor system, use clean bottled air to supply the sensors with a
continuous airflow.
4.After purging, turn operation switch to OPERATE. Allow 30 minutes to stabilize.
5.Remove the compressed air line and insert the calibration apparatus (see Figure 5).
13
Page 18
ENMET CorporationISA-44RAL-OD / ISA-44RAH-OD
6.Turn operation switch to HORN OFF. Open the regulator on the top of the high-level calibration gas cylinder. This
brings the gas into the sampling assembly.
7.Regulate the flow of calibration gas according to the "...SET" mark or between 7 – 12 PSI on the pressure regulator
dial.
8.Let the high-level calibration gas flow into the system for 5 minutes. This assures a full and balanced response from
the sensor. During this time, the meter should have risen above the low-level alarm point and the amber light should
have gone on.
9.With the gas still flowing, adjust the Meter Adjust Potentiometer with a small sc rewdriver (adjust this pot CCW to
increase, CW to decrease meter reading). You want to have the same meter reading as the calibration gas
concentration. This sets the alarm point. Refer to Figure 2.
Note:Do not adjust for instantaneous alarm when test gas is first released. If you do, the calibration will be
inaccurate, and the sensor will be overly sensitive.
10. Adjust the High Level Alarm Set Adjust Potentiometer CCW until the red alarm light is just barely triggered. At this
point, verify the audio alarm is operational by turning the operation switch to OPERATE momentarily.
NOTE:If red alarm light is on prior to the potentiometer adjustment, adjust this pot. cw until the light deactivates,
then readjust CCW until the light just barely activates.
11. Shut off the calibration gas flow (at the regulator on the gas cylinder).
12. Allow pressure to equalize – wait until the pressure meter on the humidifier-regulator assembly reads zero. Remove
calibration fixture, and reconnect humidifier-regulator assembly to compressed air line. Allow air to flow over
sensor for 10 minutes to clear the sensor.
13. In a similar manner, apply the low level calibration gas for five minutes
14. While the gas is still flowing, adjust the Low Alarm Set Adjust potentiometer CCW until the amber alarm light just
activates.
NOTE: If amber light is on prior to pot. adjustment, adjust this pot. CW until light is off, then readjust CCW until light just
activates.
IMPORTANT: The meter may not display the exact low level alarm.
DO NOT change the low level alarm meter reading unless grossly off (four meter needle widths or more). If far off,
recalibrate the high alarm, meter reading (step 9).
15. Repeat step 11.
16. Repeat step 12.
17. Reconnect the sample head to the air line.
18. The flow of clean air should clear the unit from its gas alarm state within 3-5 minutes. After 5 minutes, turn the
operation switch to OPERATE.
19. Clean compressed air lines will show a meter reading from 0 – 1/3 full scale. Green power light will indicate a clean
condition.
20. Carbon Monoxide Detection Channel Calibration is complete.
Verify the operation switch is in the OPERATE position.
14
Page 19
ISA-44RAL-OD / ISA-44RAH-ODENMET Corporation
6.2 Oxygen Circuit Calibration
Refer to Figure 2
This procedure is given in reference to an S-2 cell, which is ENMET P/N 67013-008 as furnished as a replacement cell.
MATERIALS
§ digital voltmeter
§ small screwdriver
PROCEDURE
1.Turn operation switch to HORN OFF position.
2.Null Adjust
a.Disconnect the oxygen cell from the circuit by disconnecting leads at TB3-8 (+) (terminal block 3-position 8)
and/or TB3-9 (-). Removing one lead is sufficient to break the circuit. See Figure 2.
b. Connect negative (-) voltmeter lead to TP-1 (test point one) and positive (+) voltmeter lead to TP-2 in the oxygen
circuit. These are in the lower right hand corner of the printed circuit board.
c.Adjust the Null Adjust potentiometer to read 0.00 Vdc between TP-1 and TP-2.
3.Low Level Set:
a.Reconnect the oxygen cell to TB3-9(-) and TB3-8(+).
b. Make sure the oxygen cell is properly connected by measuring the voltage between TB3-9 and TB3-8. Voltage
should read between .043 and .09 Vdc. Allow oxygen cell to stabilize for 15 minutes.
c.Connect negative lead of voltmeter to TP-1 and positive lead to TP-3 in the oxygen circuit.
d. Adjust the Low Level Set potentiometer to read .766 Vdc.
4.Full Scale Adjust:
a.Leave negative lead of voltmeter at TP-1 and connect positive lead of voltmeter to TP-2 in oxygen circuit.
b. Adjust oxygen gain potentiometer (on front panel of unit) so the vol tmeter reads 1.24 Vdc.
c.Adjust Meter Full Scale Adjust potentiometer on circuit board so the meter on the front panel reads full scale
(26%). Adjust this potentiometer cw to increase, ccw to decrease meter reading.
d. Verify that the oxygen alarm comes on at the predetermined alarm point (19.5%) on the meter. If it does not,
repeat steps a. and b. in this section. If the oxygen alarm still does not come on, repeat the entire recalibration
procedure checking terminal connections, testpoints and voltage settings very carefully. Contact your local
Authorized ENMET Service Center if problems persist.
e.Adjust and leave the oxygen gain potentiometer at a meter reading of 20.9% (setpoint).
5.The oxygen cell calibration is now complete. Turn the operation switch to OPERATE.
15
Page 20
ENMET CorporationISA-44RAL-OD / ISA-44RAH-OD
7.0 Routine Maintenance7.0 Routine Maintenance
Since the ISA-44RAL-OD/RAH-OD is a safety instrument, it requires periodic testing and recalibration. Refer to
Operation and Calibration sections of this manual for testing and recalibration procedures.
AT REGULAR INTERVALS:
1.Rotate operation switch to PURGE ON. This interval varies with the amount of contaminants present in the air
system, but purge for at least 10 minutes. Rotate operation switch to OPERATE when purging is complete.
2.Check the unit for m echanical damage or obvious malfunctions (i.e. burned-out lamps).
3.Check the oxygen channel response periodically.
4.Check the water level and verify air bubbles.
5.If the unit will not operate, check the fuse (See Figure 2). If this fuse is cracked or broken, replace it (ENMET part
no. 64004-001). If the instrument remains inoperative after this fuse is replaced, send the instrument in for repair.
7.1 Replacement Parts
Table 4: Replacement Part Numbers
DescriptionENMET Part #
Metering Orifice Assembly (s/n below 2311)03415-000
Orifice03415-001
Pressure Regulator73089-000
Humidifier RAL Regulator Assembly03406-000
Male Coupler Plug (quick-release fitting)73061-000
Gauge (0 - 30 psi)73089-003
Red, Lamp/LED-for units with s/n 2271 and below52006-003 *
Green, Lamp/LED-for units with s/n 2271 and below52006-001 *
Amber, Lamp/LED-for units with s/n 2271 and below52006-002 *
Lamp-for units with s/n 2283 and above63001-002
Lens, Red-for units with s/n 2283 and above62012-001
Lens, Green-for units with s/n 2283 and above62012-002
Lens, Amber-for units with s/n 2283 and above62012-003
5 wire RAL-OD Cable Assembly (without oxygen cell)03042-109
CO Meter03411-000
Oxygen Meter03411-005
Oxygen Cell, S-267013-008
Hydrocarbon Filter with hardware73089-060
Fuse 1 Amp64004-001
* Replaces 12RC lamps will appear different
7.2 Calibration Equipment
Table 5: Calibration Gas Part Numbers
DescriptionENMET Part Number
10 ppm CO in air03219-010
20 ppm CO in air03219-020
50 ppm CO in air03219-050
200 ppm CO in air03219-200
20.9% oxygen in nitrogen03296-209
17.0% oxygen in nitrogen03296-170
Calibration Gas Fixture03605-001
Carry Case73083-000
16
Page 21
ISA-44RAL-OD / ISA-44RAH-ODENMET Corporation
Enclosure Cover
Center Pin (neg.)
Outside Pin (pos.)
MOS Sensor
O2 Cell
and Clip
MOS Sensor & clip
O2 Cell w/clip & PCB
7.3 Oxygen Cell Replacement
The oxygen fuel cell has a six-month warranty. The average lifetime of the cell is twelve months. Replace the cell when
you can no longer adjust the oxygen gain potentiometer, outside on the front panel, so that the oxygen channel meter
reads 21%. An oxygen circuit that is in constant alarm is a good indication that the oxygen fuel cell is either expired or
grossly contaminated. See Figure 6
MATERIALS
§screwdriver
PROCEDURE
1.Obtain a new oxygen cell, ENMET P/N 67013-008.
2.Do not open the sealed package as soon as you receive it. Wait until you are ready to use the cell. The cell is
packaged in nitrogen and begins its useful lifespan as soon as it comes into contact with oxygen.
3.Remove the two screws from the sensor housing cover.
4.Unpackage the new oxygen cell. Remove the "shorting clip" before installation. Your unit goes into alarm when
you disconnect the oxygen cell, if it isn't already in alarm. The absence of a signal from the oxygen cell makes the
circuitry react as if there is not enough oxygen in the air. Turn the operation switch to HORN OFF before
disconnecting the oxygen cell.
5.Carefully unplug the old oxygen cell from the printed circuit board. Install the new cell (it may be necessary to clip
off approximate 1/2 of the gold pins on the back of the new oxygen cell to make sure that the pins on the cell do not
puncture the gasket on the sensor housing). See Figure 6. The alarm should shut off immediately after the new
oxygen cell is plugged in. If it doesn't, you may have plugged in the oxygen cell backwards.
6.Replace the encl osure cover and screws.
7.Wait at least four hours for the oxygen cell to stabilize. Having been packaged in nitrogen, the cell must adjust to an
environment with oxygen in it. The meter reading will be much higher than normal until the cell stabilizes. Don't
bother adjusting the meter now; it will drift.
8.After at least four hours, verify the alarm point by adjusting the oxygen gain potentiometer (front panel of control
unit) down to trigger the alarm.
NOTE:If, after following these steps, you are unable to trigger an oxygen alarm by adjusting the oxygen gain, check
the voltage between the oxygen cell pins. It should be between .043 to .09 Vdc. If not, you have a defective
oxygen cell. If the oxygen cell voltage is good and the alarm still does not trigger, contact your area's
authorized ENMET service center or ENMET personnel directly.
9.Leave the meter set at 20.9%.
10. This procedure is now complete. Turn operation switch to OPERATE.
Top View (cover removed)
Oxygen Cell (Bottom View)
Side View(cover removed)
Figure 6: Housing, Oxygen Cell and MOS Sensor
and Clip
17
Cable to Instrument
P C Board
Side View (cut away)
Page 22
ENMET CorporationISA-44RAL-OD / ISA-44RAH-OD
7.4 Hydrocarbon Filter Replacement
The charcoal absorption filter for hydrocarbons will need periodic replacement. The length of time between
replacement depends on the instrument usage and VOC (Volatile Organic Compounds) level in the breathing air supply.
For ISA-RAL monitors that are in continuos air streams, 24 hours, 7 days a week 365 days a year it is recommended
that the filter be replaced when any of these conditions occur:
• At least annually
• Any time fluid is found inside the filter
• Any time the filter is damaged
• If any odor is detected in the breathing mask
8.0 Specifications8.0 Specifications
Specifications are subject change without notice
MAXIMUM POWER♦ 117 Vac, 60 Hz, 25 watts
♦ 12 Vdc, 25 watts
RELAY CURRENT2 amp steady, 5 amp surge @ 110Vac
MOS SENSOR LIFEUp to 3 years in clean air (no oil)
RESPONSE TIME
MAXIMUM AIR
LINE PRESSURE250 PSIG (ISA-44 RAL-OD)
SAMPLE FLOW RATE
OXYGEN MICRO-FUEL CELL
OUTPUT
Up to 5 minutes for low-level carbon monoxide gas
calibrations
Approximately .4 SCFH (Standard Cubic Foot Hour)
linearity error (0 – 100 % oxygen)less than 0.5%
at 0% oxygenless than 2.0 ua
nominal (in air @ 25 degrees C)175 ua
variation from nominal+20%
RESPONSE TIME
@ 25 DEGREES C (TYPICAL) 90%30 seconds
OPERATING TEMPERATURE RANGE32 - 125 degrees F
EXPECTED LIFEup to 12 - 14 months in air
ENMET WARRANTY6 months
Carbon Monoxide10 or 20 ppm20 or 50 ppm
Oxygen Deficiency19.5%
8.2 Options and Variations
OPTIONS
♦ 220 Vac power input connection
VARIATIONS
a) Available in an explosion-proof instrument housing, Models ISA-44-RALE-OD and ISA-44AHE-OD. With this, the
equipment is suitable for operation in Class I, Division I, Group C and D atmosphere, as defined in the National
Electrical Code. Units for Group B are available upon special order. DO NOT use the standard ISA-44RAL-OD
where you are required by the National Electrical Code or local codes to use an explosion-proof variety.
b) ENMET Filtration Panels filter oil, water and solid particles from compressed air.
• AFS-50:equipped to distribute air to four respirators.
Ask for product bulletin on air purification and monitoring systems.
18
Page 23
ISA-44RAL-OD / ISA-44RAH-ODENMET Corporation
9.0 Warranty9.0 Warranty
ENMET warrants new instruments to be free from defects in workmanship and material under normal use for a period of
one year from date of shipment from ENMET. The warranty covers both parts and labor; however, oxygen cells are
limited to a warranty period of six (6) months from date of shipment from ENMET. Equipment believed to be defective
should be returned to ENMET within the warranty period (transportation prepaid) for inspection. If the evaluation by
ENMET confirms that the product is defective, it will be repaired or replaced at no charge, within the stated limitations,
and returned prepaid to any location in the United States. ENMET shall not be liable for any loss or damage caused by
the improper use of the product. The purchaser indemnifies and saves harmless the company with respect to any loss or
damages that may arise through the use by the purchaser or others of this equipment.
Material shipped to ENMET for warranty evaluation must be packed so it is not damaged in shipping. This applies
particularly to oxygen cells, which must be packed so that the connector pins cannot damage the membranes of adjacent
cells, and cells cannot be damaged by the movement of other cells or items in the shipment. We strongly suggest that
oxygen cell pins be protected with a small piece of styrofoam, that individual cells be placed in small plastic bags, and
that adequate packing material be used to constrain the movement of the cells and other items during shipment. Material
damaged in shipment is not covered by warranty.
This warranty is expressly given in lieu of all other warranties, either expressed or implied, including that of
merchantability, and all other obligations or liabilities of ENMET which may arise in connection with this equipment.
ENMET neither assumes nor authorizes any representative or other person to assume for it any obligation or liability
other than that which is set forth herein.
There are Return for Repair Instructions and Form on the last pages of this manual. This Form can be copied or used
as needed.
19
Page 24
ENMET CorporationISA-44RAL-OD / ISA-44RAH-OD
Notes:
20
Page 25
PO Box 979
680 Fairfield Court
Ann Arbor, Michigan 48106-0979
734.761.1270 Fax 734.761.3220
Returning an Instrument for Repair
ENMET instruments may be returned to the factory or any one of our Field Service Centers for
regular repair service or calibration. The ENMET Repair Department and Field Service Centers
also perform warranty service work.
When returning an instrument to the factory or service center for service, paperwork must be
included which contains the following information:
Ø A purchase order number or reference number.
Ø A contact name with return address, telephone and fax numbers
Ø Specific instructions regarding desired service or description
of the problems being encountered.
ØDate of original purchase and copy of packing slip or invoice
for warranty consideration.
ØIf a price estimate is required, please note it accordingly and be
sure to include a fax number .
Providing the above information assists in the expedient repair and return of your unit.
Failure to provide this information can result in processing delays.
ENMET charges a one hour minimum billing for all approved repairs with additional time billed
to the closest tenth of an hour. All instruments sent to ENMET are subject to a minimum $30
evaluation fee, even if returned unrepaired. Unclaimed instruments that ENMET has received
without appropriate paperwork or attempts to advise repair costs that have been unanswered,
after a period of 60 days, may be disposed of or returned unrepaired COD with the evaluation
fee.
Service centers may have different rates or terms. Be sure to contact them for this information.
Repaired instruments are returned by UPS/FedEx Ground and are not insured unless
otherwise specified. If expedited shipping methods or insurance is required , it must be
stated in your paperwork.
Note: Warranty of customer installed components.
If a component is purchased and installed in the field, and fails within the warranty term,
it can be returned to ENMET and will be replaced, free of charge, per ENMET’s returned
goods procedure.
If the entire instrument is returned to ENMET Corporation with the defective item
installed, the item will be replaced at no cost, but the instrument will be subject to labor
charges at half of the standard rate.
Page 26
Page 27
Repair Return Form
Mailing Address:
ENMET Corporation
PO Box 979
Ann Arbor, Michigan 48106
q UPS: q Groundq 3 Day Selectq Next Day Air q ND Air Saverq 2-Day Air
q Federal Express:q Groundq Express Saverq P-1 q Standard q 2-Day Air
q FedEx Account number: ________________________
Would you like ENMET to insure the return shipment?
q Noq YesInsurance Amount: $_________________
Page 28