ENMET EX-6165 User Manual

Page 1
ENMET Corporation
PO Box 979 Ann Arbor, MI 48106-0979
www.enmet.com
EX-6120 S
EX-6165 S
ERIES, COMBUSTIBLE GAS
ERIES, CARBON DIOXIDE
S
ENSOR/TRANSMITTER
I
NFRARED
M
ANUAL
(IR)
Manual Part Number 80003-123
April 2013
Page 2
Table of Contents
1.0 INTRODUCTION ....................................................................................................................................................................................... 1
1.1 U
NPACK
1.2 C
1.3 S
2.0 FEATURES OF THE EX-6120, 6165 SERIES IR....................................................................................................................................... 2
3.0 INSTALLATION OF THE EX-6120, 6165 SERIES IR ................................................................................................................................ 3
3.1 M
3.2 W
3.2 P
3.3 A
3.4 RS232 O
3.5 F
3.6 A
3.7 I
4.0 OPERATION ............................................................................................................................................................................................. 8
5.0 CALIBRATION .......................................................................................................................................................................................... 9
5.1 S
6 SENSOR REPLACEMENT ........................................................................................................................................................................ 10
6.1 I
7 FUSES ....................................................................................................................................................................................................... 13
8 SPECIFICATIONS ..................................................................................................................................................................................... 13
9 CERTIFICATION ........................................................................................................................................................................................ 14
10 EX-6120, 6165 SERIES IR ACCESSORIES AND CALIBRATION GAS .................................................................................................. 14
11 MENU MODE ........................................................................................................................................................................................... 15
APPENDIX A ................................................................................................................................................................................................ 19
A.1 A
A.2 E A.3 C
A.4 F
A.4.4 W
WARRANTY ................................................................................................................................................................................................. 22
................................................................................................................................................................................................. 1
HECK ORDER ERIAL NUMBERS
OUNTING THE
IRING THE OWER SUPPLY NALOGUE OUTPUT
OLLOWING COMPLETION OF THE INSTALLATION PPLYING POWER
NDICATORS
ENSOR CALIBRATION
5.1.1 Sensor Zero ........................................................................................................................................................................................................................ 9
5.1.2 Sensor Span ....................................................................................................................................................................................................................... 9
NFERRED SENSOR
11.2.1 E : 1 – Sensor Zero ........................................................................................................................................................................................................ 15
11.2.2 E : 2 – Sensor Span ....................................................................................................................................................................................................... 15
11.2.3 E : 3 – Sensor FSD ........................................................................................................................................................................................................ 15
11.2.4 E : 4 – Output Zero (4mA) .............................................................................................................................................................................................. 15
11.2.5 E : 5 – Output Span (20mA) ........................................................................................................................................................................................... 15
11.2.6 E : 6 – Decimal Places ................................................................................................................................................................................................... 16
11.2.7 E : 7 – Firmware Version ................................................................................................................................................................................................ 16
11.2.8 E : 8 – Sensor Selection ................................................................................................................................................................................................. 16
11.2.9 E : 9 – Engineer diagnostics .......................................................................................................................................................................................... 16
11.2.10 E : 12 – Cross reference .............................................................................................................................................................................................. 16
11.2.11 E : 13 – Relay 1 ............................................................................................................................................................................................................ 17
11.2.12 E : 14 – Relay 2 ............................................................................................................................................................................................................ 17
11.2.13 E : 15 – Alarm Level 1 .................................................................................................................................................................................................. 17
11.2.14 E : 16 – Alarm Level 2 .................................................................................................................................................................................................. 17
11.2.15 E : 19 – Positive Zero Suppression .............................................................................................................................................................................. 17
11.2.16 E : 20 – Negative Zero Suppression ............................................................................................................................................................................ 18
11.2.17 E : 27 – Hysteresis ....................................................................................................................................................................................................... 18
11.2.18 E : 28 – Firmware update ............................................................................................................................................................................................. 18
11.2.19 E : 29 – Relay 3 ............................................................................................................................................................................................................ 18
11.2.20 E : 30 – Password ........................................................................................................................................................................................................ 18
11.2.21 E : 31 – Warm-up timer ................................................................................................................................................................................................ 18
NALOG OUTPUT CALIBRATION
A.1.1 Analogue Output Zero...................................................................................................................................................................................................... 19
A.1.2 Analogue Output Span .................................................................................................................................................................................................... 19
RROR CODES
OMMUNICATIONS PROTOCOL
A.3.1 Hardware connections ..................................................................................................................................................................................................... 20
A.3.2 Control Byte Constants .................................................................................................................................................................................................... 20
RAME STRUCTURE
A.4.1 Variables .......................................................................................................................................................................................................................... 20
A.4.2 Reading a Variable .......................................................................................................................................................................................................... 21
A.4.3 Read example - read live data ......................................................................................................................................................................................... 21
RITING A VARIABLE
....................................................................................................................................................................................... 1
................................................................................................................................................................................... 1
EX-6120, 6165 S
EX-6120, 6165 S
..................................................................................................................................................................................... 5
UTPUT
................................................................................................................................................................................ 5
..................................................................................................................................................................................... 5
................................................................................................................................................................................... 6
........................................................................................................................................................................................... 6
............................................................................................................................................................................ 9
............................................................................................................................................................................... 10
.................................................................................................................................................................................... 19
............................................................................................................................................................................. 20
......................................................................................................................................................................... 21
ERIES IR ENCLOSURE
ERIES IR TO A CONTROL UNIT
............................................................................................................................................................. 19
.............................................................................................................................................................. 20
........................................................................................................................... 3
..................................................................................................................... 4
: ...................................................................................................................................... 6
Page 3
List of Figures and Tables
F
IGURE 1:
F
IGURE 2:
F
IGURE3: ACCESS TO
F
IGURE 4: ARMORED CABLE GLAND INSTALLATION
F
IGURE 5: UNPLUGGING THE DISPLAY BOARD FOR CONNECTION PURPOSES
F
IGURE 6: ANALOGUE OUTPUT
F
IGURE 7: EEX D CONDUIT INSTALLATION
F
IGURE 8: REPLACING THE DISPLAY BOARD
F
IGURE 9: APPLING TEST GAS TO
F
IGURE
T
ABLE 1: MAGNETIC KEYPAD FUNCTIONALITY
T
ABLE 2:
T
ABLE 3:
T
ABLE 4:
EX-6120, 6165 S EX-6120, 6165 S
ERIES IR FEATURES ERIES IR MOUNTING
EX-6120, 6165 S
................................................................................................................ 5
EX-6120, 6165 S
10: O
PERATION OF MAGNETIC SWITCHES
EX-6120, 6165 S EX-6120, 6165 S EX-6120, 6165 S
ERIES IR RELATIVE RESPONSE CURVES BASED ON A METHANE CALIBRATION ERIES IR MENU OPTIONS ERIES IR SENSOR SELECTIONS
..................................................................................... 2
..................................................................................... 3
ERIES IR FOR CONNECTION TO A CONTROL UNIT
.......................... 4
.................................................................................. 4
............................................. 5
................................................................................................ 5
............................................................................................. 6
ERIES
IR ................................................................... 7
..................................................................................... 8
.......................................................................................... 8
.............................................................................15
....................................................................16
.. 9
2
Page 4
ENMET Corporation
EX-6120, 6165 S
1.0 Introduction
The ENMET EX-6120, 6165 S Certified for use in Group IIC hazardous locations. Directive 94/9/EC ATEX and the equivalent rating of Class1, Division 1, Groups A, B, C and D
Harmonised Standards Used:
EN60079-0:2006 Electrical apparatus for explosive gas atmospheres – Part 0: General requirements. EN60079-1:2007 Explosive atmospheres – Part 1: Equipment protection by flameproof enclosures “d”.
Notified Body for Hazardous Area Certification:
SIRA Certification Service Rake Lane, Eccleston, Chester, CH4 9JN. Notified Body Number : 0518
Hazardous Area Certificate Number:
SIRA 08ATEX1031X II 2 G Ex d IIC T4 (Ta = -20OC + 60 OC )
N
OTE
: All specifications stated in this manual may change without notice.
1.1 Unpack
Unpack the EX-6120, 6165 S
ENMET customer service personnel and the commercial carrier involved immediately.
Regarding Damaged Shipments
N
OTE
: It is your responsibility to follow these instructions. If they are not followed, the carrier will not
honor any claims for damage.
This shipment was carefully inspected, verified and properly packaged at our company and delivered to the carrier in good condition.
When it was picked up by the carrier at ENMET, it legally became your company’s property.
If your shipment arrives damaged:
Keep the items, packing material, and carton “As Is.” Within 5 days of receipt, notify the carrier’s local
office and request immediate inspection of the carton and the contents.
After the inspection and after you have received written acknowledgment of the damage from the carrier,
contact ENMET Customer Service for return authorization and further instructions. Have your Purchase Order and Sales Order numbers available.
ENMET either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability coverage, usually $100.00. Repair or replacement charges above that value are your company’s responsibility.
The shipping company may offer optional insurance coverage. ENMET only insures shipments with the shipping company when asked to do so in writing by our customer. If you need your shipments insured, please forward a written request to ENMET Customer Service.
Regarding Shortages
If there are any shortages or questions regarding this shipment, please notify ENMET Customer Service within 5 days of receipt at the following address:
ENMET Corporation
1.2 Check Order
Check, the contents of the shipment against the purchase order. Verify that the EX-6120, 6165 S as ordered. [Each EX-6120, 6165 S ascertain that they are present. Check the contents of calibration kits. Notify ENMET customer service personnel of any discrepancy immediately.
1.3 Serial Numbers
Each EX-6120, 6165 S
ENMET database.
ERIE
s IR Fixed Gas Detectors for the detection of Flammable gases is ATEX Flameproof
Notified Body for ATEX Quality Assurance Notification:
Baseefa Rockhead Business Park Staden Lane, Buxton SK17 9RZ, UK Notified Body Number : 1180
ATEX Quality Assurance Notification Number:
2056
ERIES
IR and examine it for shipping damage. If such damage is observed, notify both
680 Fairfield Court
734-761-1270 734-761-3220 Fax
ERIES
ERIES
IR is serialized. These numbers are on tags on the equipment and are on record in an
Ann Arbor, MI 48108
ERIES
IR is labeled with its target gas.] If there are accessories on the order,
IR is received
ERIES
IR
1
Page 5
EX-6120, 6165 SERIES IR
GAS DETECTOR TYPE
GAS DETECTOR TYPE
2 x 0.27 (2 x 7) dia.
Inlet/Outlet
ENMET Corporation
2.0 Features of the EX-6120, 6165 S
The EX-6120, 6165 S
ERIES
IR is an ATEX and IECEx certified fixed gas detector for use in potentially explosive
ERIES
IR
atmospheres. The unit may be optionally fitted with a protective weather guard as shown in the photograph.
The unit comprises an instrument housing having two cable gland entries and containing the connection terminals, electronics and display window.
The housing containing the gas sensor has an M27 thread and is screwed into the bottom of the unit. The sensor housing itself is a certified component and must not be removed in service.
Behind the display window are-
See Figure 1 for location of features:
Feature Description Display Magnetic Switches
Visual Indicators Power supply Analogue output Communications Output Relay
Gland (Strain Relief)
A 4-digit LCD display Four magnetically operated switches that are activated through the display window using a magnetic
pen that is supplied with the unit. LED indicators for POWER, ALARM & KEY PRESS
8 to 24 volts dc (non-intrinsically safe) 4 to 20mA dc RS232 Relay version has 3 factory-fitted relays. Each provides a set of voltage free changeover contacts to
perform the following control functions contact outputs for Alarm Level 1, Alarm Level 2 and Fault entry threads available – 20mm, ½” or ¾” NPT
Magnetic switches control the instrument maintenance functions. The magnetically operated switches allow the unit to be calibrated while power is still applied without the need to remove the cover from the unit.
Versions without relays fitted are also available. The self-adhesive label on the outer surface of the enclosure indicates whether or not the relays are fitted as shown below:
IMPORTANT – When used in environments where the instrument is exposed to salt spray or diesel it is recommended
that the EX-6120, 6165 S attached with tamperproof screws to ensure that it is not inadvertently removed.
ERIES
IR is always fitted with its protective weather guard. The weather guard is
NO CONTROL RELAYS FITTED
Port 1 of 2
MENU UP DOWN ENTER
CONTROL RELAYS FITTED
Inlet/Outlet Port 1 of 2
Mounting Holes
Display
Menu Switches
Power & Alarm LEDs
Optional Weatherguard
Figure 1: EX-6120, 6165 S
2
ERIES
IR Features
Page 6
EX-6120, 6165 SERIES IR
Certificate Numbers
Standards
Temperature Codes
Zones
Equivalent Rating
Optional
2 x 0.27 (2 x 7) dia.
ENMET Corporation
3.0 Installation of the EX-6120, 6165 S
C
AUTION
:
Area must be declassified during installation.
The EX-6120, 6165 S
ERIES
IR gas sensor/transmitter (S/T) is a 3-wire, 24 VDC 4-20 mA S/T for the detection of
ERIES
IR
combustible gas. The S/T is meant to be used in conjunction with an appropriate power supply and controller. The EX­6120, 6165 S
ERIES
IR sensor/transmitter is in an enclosure rated for use in area listed below:
IECEx SIR08.0009X,Code EX d IIC SIRA 08atex1031x, Code EX d IIC IEC 60079-0 : 2004 (Edition 4) EN 60079-0 : 2006 IEC 60079-1 : 2007-4 (Edition 6) EN 60079-1 : 2007 T4 (Ta-20 to +60 deg C) 1 & 2 Class I, Division 1, Group A, B, D & D
Appropriate wiring, conduit and fittings are required for proper installation in an explosion proof rated environment.
C
AUTION
: Since the sensor/transmitter detects gas only at the sensor location, pay attention to the possible sources of gas,
the density of the gas, locations where the gas may be confined and locations where the gas may damage or injure property or personnel, when choosing locations of sensor/transmitters.
Take into consideration environmental factors when deciding on S/T location. Avoid locations where the S/T may be damaged by liquid immersion, excessive heat or other known hazards. Also, take precautions to insure condensation inside of the conduit does not enter the S/T.
3.1 Mounting the EX-6120, 6165 S
ERIES
IR Enclosure
Mount the enclosure, using the two mounting holes provided see Figure 2. Pay particular attention to the source and density of the gas being detected when choosing the location. Mount the S/T near the ceiling for lighter than air gases /vapors and near the floor for heavier then air gas/vapors. Contact ENMET if you have questions regarding your application.
C
AUTION
: Before connecting S/T to controller remove the power source to controller. Failure to do so may cause damage
to sensitive components.
Inlet/Outlet
Port 1 of 2
4.96 (126) Centers
4.53 (115)
Inlet/Outlet Port 1 of 2
MENU UP DOWN ENTER
190)
250)
Figure 2: EX-6120, 6165 S
Power & Alarm LEDs
Optional Weatherguard
ERIES
IR Mounting
Conduit fitting Typically: 3/4 NPT female
Mounting Holes
Approximate Measurements in Inches (mm)
Maximum Depth = 5 Inches (127mm)
Display
Menu Switches
3
Page 7
EX-6120, 6165 SERIES IR
N/O
COM
N/C
Alarm 1
N/O
COM
N/C
Alarm 2
N/O
COM
N/C
Fault
ENMET Corporation
3.2 Wiring the EX-6120, 6165 S
C
AUTION
:
Area must be declassified during installation.
If the EX-6120, 6165 S
ERIES
IR is installed in a hazardous location as defined by the National Electrical Code, then
ERIES
IR to a Control Unit
ALL wiring must be in accordance with the National code and any local governing codes. Open the enclosure, and remove the 2 screws that retain the display overlay to the circuit board. Turn cover securing stud counter-clockwise to allow cover to open.
Remove cover by rotating counter-clockwise.
Figure3: Access to EX-6120, 6165 S
ERIES
IR for Connection to a Control Unit
The cable entry threads are 20mm, ½” or ¾” NPT female. The EX-6120, 6165 S
ERIES
IR enclosure is manufactured from die cast aluminum*. Therefore, the use of glands,
conduit fittings and blanks made from brass should be avoided because if moisture is present, bi-metallic corrosion may occur due to the chemical reaction between the two materials.
Glands and fittings plated with nickel, tin or zinc will provide improved protection but in harsh environments the use of stainless steel is recommended.
Conduit
or Shielded Cable
ATEX certified Ex d armoured
cable glands
Conduit
or Armoured Cable
- +
Aout
Mount vertically
Relay 3
as shown
Figure 4: Armored Cable Gland Installation
Relays - three relays are provided:
Alarm 1 and Alarm 2 are associated with the alarm points. Alarm 3 is associated with fault conditions. All relays are factory set to de-energize state non alarm.
RS232
0V
Relay 2
Rx
Tx
- +
PSU
Relay 1
Important
Connect enclosure
earth-ground to a high
integrity earth ground
4
Page 8
EX-6120, 6165 SERIES IR
N/O
COM
N/C
Alarm 1
N/O
COM
N/C
Alarm 2
N/O
COM
N/C
Fault
0V
A+
0V
12-24V DC
ENMET Corporation
Figure 5: Unplugging the Display Board for Connection Purposes
3.2 Power Supply
The EX-6120, 6165 S
a) An independent supply powering one or more units. b) A supply that is an integral part of a control card within a monitoring panel.
ERIES
IR is powered from an 8 – 24 volt dc, 5 W maximum output supply. This may be either:-
3.3 Analogue Output
The analogue output provides a means of indicating to external equipment (e.g. data loggers, remote displays, control cards) the gas levels currently being detected by the system. The output is in the form of the industry standard 4 to 20mA current source.
The analogue output is factory set such that 4mA represents zero gas and 20mA represents the full scale gas level of the
EX-6120, 6165 S
ERIES
IR.
I (4-20mA)
PSU
Control
Unit
EX-6120,
Figure 6: Analogue Output
3.4 RS232 Output
This output is designed to be compatible with a PC fitted with an RS232 serial communications port. Suitable software is available for installation in the PC. Contact ENMET Corporation for details.
Conduit
Mount vertically
as shown
ATEX Certified compound-filled
conduit seals (stopping boxes)
RS232
0V
Rx
Tx
- +
Aout
Relay 3
Relay 2
Relay 1
- +
PSU
Conduit
Important
Connect enclosure
earth-ground to a high
integrity earth-ground
Figure 7: EEx d Conduit Installation
5
Page 9
EX-6120, 6165 SERIES IR
Colo
r Identity
Operation
3.5 Following Completion of the Installation:
a) Replace the top display board using the locating pins on the underside of the board to ensure that it makes
connection with the remainder of the electronics - see Figure 8.
b) Ensure that the front cover has not been contaminated with dirt - paying particular attention to the thread.
Replace the cover by rotating it several times clockwise until it reaches its limit then secure using the set screw – do not over tighten.
ENMET Corporation
Figure 8: Replacing the Display Board
3.6 Applying power
Caution: The cover must be in place before applying power
Once the installation is complete and the covers are secure then power can be applied. The power source should be between 8 and 24 volts dc.
When power is first applied to the EX-6120, 6165 S
ERIES
IR, an initialization procedure is performed a example
follows:
All segments on the LCD are shown 8.8:8.8 The software version is displayed 1.0.5 The manufacturer code is displayed SSCL The sensor type is displayed Ch4L Warm up count down 50
The detected gas level present is displayed and the EX-6120, 6165 S
ERIES
IR becomes operational.
3.7 Indicators
Three LED indicators are visible from the front panel of the instrument. These are:
Red Alarm Illuminates when either alarm level has been reached. Green Power Always ON when power is connected. Yellow Key press Illuminates when the magnetic switches are operated.
6
Page 10
EX-6120, 6165 SERIES IR
Calibration Adapter,
3.8 Initial Gas Testing
EX-6120, 6165 S
instrument. However, it is always advisable after installation to confirm that the instrument reads zero with no gas present and responds accurately when presented with an appropriate concentration of the target gas.
The EX-6120, 6165 S to stabilize for the period of five to ten minutes before attempting to check the zero setting and gas response.
Refer to Figure 9 showing the arrangement for applying the test gas to the EX-6120, 6165 S
1. Before application of the test gas, check that the EX-6120, 6165 S
present in the atmosphere. If necessary, carry out adjustment of the zero setting as described in section 5.1.1
2. Apply the test gas and allow a sufficient time for the sample to reach the gas detector and for the sensor
response to stabilize.
Carrying out the above procedure, and comparing the results with previous readings, reference the certificate of calibration, will confirm that the EX-6120, 6165 S There is therefore no requirement for any further maintenance other than to clean the display window as required.
In the unlikely event that the response has changed from the previous readings, recalibrate the unit. See Section 5.1.1 and 5.1.2.
ERIES
IR units are factory calibrated as detailed on the calibration certificate supplied with the
ERIES
IR display becomes operational within 60 seconds however; the sensor should be allowed
ERIES
ERIES
IR reads zero with no known gas
ERIES
IR is functioning correctly, both physically and electrically.
IR.
ENMET Corporation
part number 03700-058
Figure 9: Appling Test Gas to EX-6120, 6165 S
3.8.1 Suggested calibration Test gas levels.
Hydrocarbon sensor:
Zero contaminant free air. Span 50% of the target gas(for 0 – 100% LEL range).
Carbon Dioxide sensor:
Zero Nitrogen. Span Approximately 50% of Range.
Sampling Tube
With Weatherguard
Calibration Adaptor with Weatherguard, part number 04546-012
Flow Regulator
Certified Test Gas Cylinder
ERIES
IR
7
Page 11
EX-6120, 6165 SERIES IR
GAS DETECTOR
Button
Function
ENMET Corporation
4.0 Operation
The
EX-6120, 6165 S
To ‘press’ the switch requires movement of the stylus pen over the appropriate button. The key press LED will light to indicate that the button press is being detected. In some instances it may be necessary to hold the pen over the button for sufficient time to allow the microcontroller to determine the key that has been pressed.
Description
EX-6120, 6165 S
ERIES
The menu system featured within the EX-6120, 6165 S performed.
ERIES
IR
incorporates magnetically operated switches to allow non-intrusive operation of the menu system.
IR
EX-6120, 6165 S
Figure 10: Operation of Magnetic Switches
ERIES
IR allows all calibration and configuration activities to be
ERIES
IR
GAS DETECTOR FITTED WITH
PROTECTIVE WEATHERGUARD
MENU UP DOWN ENTER
Table 1: Magnetic Keypad Functionality
MENU Open / Close Menu
UP Next / Increase
DOWN Previous / Decrease
ENTER Accept selection
This section of the manual discusses how the zero and calibration menu options can be accessed. Additional menu option associated with the operational parameter may be changed via the menu mode section 11.
8
Page 12
EX-6120, 6165 SERIES IR
ENMET Corporation
5.0 Calibration
MENU UP DOWN ENTER
1 2 3 4
5.1 Sensor Calibration
This feature allows the sensor to be calibrated. Ensure that the correct sensor type is selected in the configuration prior to calibration. Refer to section for details of the menu system operation.
5.1.1 Sensor Zero
Tap the magnet over the menu switch, the display will change to Pass, Tap the magnet again on menu display will show E 1.
Tap the magnet over the enter switch the display will then alternate E: 01 Wait approximately 1 minute and Tap enter.
If the zero is accepted the display will show ---- Tap the magnet over the menu switch to exit. If the zero calibration was not within acceptable range the display will show F2. A recalibration is required; you
may need a cylinder of 20.9 air.
Note: 1. Carbon Dioxide sensors, except for high % range, should be zeroed in air due to the background levels of CO2
present. These sensors are best zeroed while being exposed to 100% nitrogen.
2. Where a purging gas has to be applied, use a flow rate of between 0.5 and 1.0 L/min. Allow sufficient time for the sensor to respond.
5.1.2 Sensor Span
Tap the magnet over the menu switch, the display will change to Pass, Tap the magnet again on menu display will show E 1.
Tap the magnet over the up switch, the display will change to E 2. Tap the magnet over the enter switch the display will flash E: 2 and XX (XX = the calibration gas level).
Caution:
Table 2: EX-6120, 6165 S
Be sure that the calibration gas you use matches the type and concentration of calibration gas display on the EX-
6120, 6165 S
ERIES
IR
If the calibration gas you have does not match the displayed value, place the magnet over the up or down switch to adjust the value.
Tap the magnet over the enter switch, the display will flash E: 2 Apply the calibration gas. The display will flash E: 2 and a number (sensor response to gas) when the number stops
changing, Place the magnet over the enter switch. The display will show ---- if the calibration was successful. Tap the menu switch to exit and remove the calibration gas.
ERIES
IR Relative response curves based on a Methane Calibration
See Accessories Section 10 for ENMET Gas Cylinder part numbers.
9
Page 13
EX-6120, 6165 SERIES IR
Step1
Step 2
6 Sensor Replacement
W
ARNING
6.1 Inferred Sensor
:
Power must be removed from the EX-6120, 6165 S
to do so may cause damage to equipment, bodily injury or death.
Remove the set screw counter-clockwise.
ENMET Corporation
ERIES
IR before this or any internal procedure. Failure
Set Screw
Turn cover securing stud counter-clockwise to allow cover to open.
10
Page 14
EX-6120, 6165 SERIES IR
Step 4
Step 3
Remove the sensor by pulling downwards.
ENMET Corporation
Carefully locate the pins in the sockets then push the new sensor upwards.
11
Page 15
EX-6120, 6165 SERIES IR
Step 5
Replace the sensor housing by rotating in a clockwise direction until fully tightened.
ENMET Corporation
Step 6
Rotate the set screw in a clockwise direction until fully tightened.
Set Screw
12
Page 16
EX-6120, 6165 SERIES IR
:
7 Fuses
A 1Amp anti-surge fuse, housed within a surface mounted holder, is located within the EX-6120, 6165 S overloading of the electronic circuitry in the event of an internal fault.
In the event of the internal fuse blowing, access can be gained using the following procedure:
1. Switch OFF the supply to the EX-6120, 6165 S
ERIES
IR.
2. Release the grub screw located near the lip of the enclosure cover.
N
OTE
: It is not necessary to remove the grub screw to release the enclosure front cover.
3. Remove the enclosure front cover by rotating it several times in a counter-clockwise direction.
4. Unplug the display board from within the enclosure.
5. Remove the 3 screws located around the edge of the next board.
6. Unplug this board.
7. The surface mounted fuse holder can now be located on the board below. Replace only with a fuse of the same type and
rating:­1Amp Anti-surge ‘Nano fuse’
Assembly is the reverse of the above procedure.
8 Specifications
Instrument Body – Aluminium Pressure Die Casting
Materials :
Cable entries :
Weights :
Gas type : Input voltage : Input power : Internal fuse : Analogue
output
Relays
RS232 output : Sensor type :
Measurement range
Response time : Measurement
resolution IP rating : Display : Keypad : Software : Operating
temperature Humidity range : Operating
pressure Performance
standards
Sensor Housing – Stainless Steel Grade 303 Weather guard – Stainless Steel Grade 304 & Nylon 66
2 x 20 mm, or 2 x ½ NPT, or 2 x ¼ NPT
EX-6120, 6165 S Magnetic Pen – 60 grams Weather guard – 200 grams
Infrared – (Note: IR Sensors have not response to Hydrogen) 8 to 24 volts dc 5 Watts maximum 1 Amp antisurge 'Nanofuse' 4 to 20mA (10 bit resolution)
Relay 1 – Alarm 1 Relay 2 – Alarm 2 Relay 3 – Fault Contact Rating 300Vac, 3 Amps (Non-inductive load)
Communications with PC at 19200 baud NDIR Infrared Hydrocarbon Sensor (Methane Calibration)
0 – 100% LEL (5% vol. CH4) or 0 – 100% volume CH4
:
Carbon Dioxide Sensor
Ranges form 0-1000ppm up to 0-100% Volume CO2 Typically T90 < 30 seconds (CH4)
:
1% LEL or 1% volume (CH4) Enclosure IP66, Sensor IP65
4 Digit, 7 segment liquid crystal 4-Button magnetically operated Software configuration provided via LCD display and multifunction keypad
:
- 20 to +60 O C 0 to 95% RH non-condensing
:
Atmospheric + or - 10%
:
EN 60079-29-1:2007
ERIES
IR (excluding weather guard) – 2.0 Kg ( 3.3 lb)
ENMET Corporation
ERIES
IR to prevent
13
Page 17
EX-6120, 6165 SERIES IR
Description of Accessory
Part Number
ENMET Corporation
9 Certification
The EX-6120, 6165 S explosive atmospheres and is marked as follows:-
** Denotes additional characters that define the gas detector configuration e.g. Enclosure type and sensor arrangement. YYYY – Denotes year of manufacture.
Special Conditions of Safe Use
Only Loctite 243 or a compound of equal strength shall be reapplied to the threads of the sensor after replacing the sensing element
Electrical data
Power supply - 8Vdc to 24Vdc, 5W Relay contact rating – 300Vac, 3 Amps
Installation instructions
The cable entry devices and blanking elements of unused apertures shall be of a certified flameproof type, suitable for the conditions of use and correctly installed.
With the use of conduit, a suitable certified sealing device such as a sealing fitting with compound shall be provided immediately at the entrance to the flameproof enclosure.
ERIES
IR(FGD10 Series) of Gas Detectors are ATEX and IECEx certified for use in potentially
WARNING – DO NOT OPEN WHEN AN EXPLOSIVE GAS
STATUS SCIENTIFIC CONTROLS LTD. MANSFIELD NG18 5ER.UK.
II 2 G Ex d IIC 1180 SIRA 08ATEX1031X
SERIAL No. XXXXXX CIRCUIT RATING: 24VDC, 5W YYYY RELAY CONTACT RATING: 300VAC. 3A MAX.
ATMOSPHERE IS PRESENT
GAS DETECTOR TYPE FGD10 **
IECEx SIR 08.0009X
T4 (Ta -20OC to +60OC)
10 EX-6120, 6165 S
EX-6120, 6165 S
Sensor Calibration Adaptor Regulator Assembly 02506-004 Calibration Gas, 50%LEL Methane Calibration Gas, 2000ppm CO2 Weather Guard 04546-012 Maintenance Tool, Sensor 73413-210
ERIES
IR partial list of typical accessories and calibration gas ENMET part number
(for sensors without weather guard)
ERIES
IR Accessories and Calibration Gas
Contact ENMET Corp
03700-058
(for 0 – 100% LEL CH4)
(for 0 – 500ppm CO2)
03220-050
03220-2000
14
Page 18
EX-6120, 6165 SERIES IR
ENMET Corporation
11 Menu Mode
The EX-6120, 6165 S compromise the use of the equipment. Menu Mode should only be entered by qualified personal.
To enter the menu mode Sequence Tap:
Menu, Up, Down, Enter, Up, Enter, Up, Menu
If no inputs (taps) the EX-6120, 6165 S Several calibration modes exist in the EX-6120, 6165 S
To select a calibration mode follow this procedure:
Press the MENU button, E: 1 appears on the display.
Press UP or DOWN until the required menu option is displayed.
Press ENTER to select the calibration mode.
To exit the menu press MENU.
While the instrument is in menu mode – any data displayed on the screen will alternate between the cal number and the reading.
The following features are available via the ‘EX-6120, 6165 S
11.2.1 E : 1 – Sensor Zero
Refer to section 5.1.1 for sensor calibration details.
11.2.2 E : 2 – Sensor Span
Refer to section 5.1.2 for sensor calibration details.
11.2.3 E : 3 – Sensor FSD
This menu option allows the full-scale gas level to be selected. From the menu system select menu option: E: 3 and press ENTER.
Using the INCREASE and DECREASE buttons; adjust the FSD to the required level.
Press ENTER, ‘- - - - ‘will be displayed to indicate that the new setting has been accepted.
Note: Pressing MENU instead of ENTER will exit the menu without changing the FSD.
Press MENU to close the menu system.
11.2.4 E : 4 – Output Zero (4mA)
Refer to section Appendix A for output calibration details.
11.2.5 E : 5 – Output Span (20mA)
Refer to section Appendix A for output calibration details.
ERIES
IR uses a password system to restrict the end user from carrying out certain changes that may
ERIES
IR returns to operational mode.
ERIES
IR and these are accessible via the instruments simple menu system.
ERIES
IR’ menu system:
Table 3: EX-6120, 6165 S
ERIES
IR Menu Options
Menu Option Section
E : 1 – Sensor Zero 11.2.1 E : 2 – Sensor Span 11.2.2 E : 3 – Sensor FSD 11.2.3 E : 4 – Output Zero (4mA) 11.2.4 E : 5 – Output Span (20mA) 11.2.5 E : 6 – Decimal Places 11.2.6 E : 7 – Firmware Version 11.2.7 E : 8 – Restore 11.2.8 E : 9 – Engineer Diagnostics 11.2.9
E : 12 – Cross Reference 11.2.10
E : 13 – Relay 1 11.2.11 E : 14 – Relay 2 11.2.12 E : 15 – Alarm Level 1 11.2.13 E : 16 – Alarm Level 2 11.2.14 E : 19 – Positive Zero Suppression 11.2.15 E : 20 – Negative Zero Suppression 11.2.16 E : 27 – Hysteresis 11.2.17
E : 28 – Firmware Update 11.2.18
E : 29 – Relay 3 11.2.19 E : 30 – Password 11.2.20 E : 31 – Warm-Up Timer 11.2.21
Contact ENMET Corporation before entering this menu option. Do Not enter this menu option.
15
Page 19
EX-6120, 6165 SERIES IR
OPTION
TYPE
RANGE
COMMENTS
ENMET Corporation
11.2.6 E : 6 – Decimal Places
The number of decimal places that are used to display the detected gas levels can be changed via this menu option.
From the menu system select menu option: E: 6 and press ENTER.
Using the NEXT and PREVIOUS buttons, select the required number of decimal places.
Press ENTER.
N
OTE
: Pressing MENU instead of ENTER will exit the menu without changing the number of decimal places to
which the gas level will be displayed.
Press MENU to close the menu system.
Note that when the sensor type is first selected, the optimum number of decimal places for a given sensor type and range is automatically selected. Increasing the number of decimal places to which detected gas levels are displayed does not increase the sensitivity or accuracy of the instrument and in some cases may degrade the instruments perceived stability.
11.2.7 E : 7 – Firmware Version
The Firmware version is displayed in the form X.Xyy, where X is numerical and y is a letter.
Press MENU to close the menu system.
11.2.8 E : 8 – Sensor Selection
This menu option allows the sensor type to be selected. The available sensor options are listed below:
Table 4: EX-6120, 6165 S
CH4L Infrared 0-100%LEL Methane
CH4H Infrared 0-100% v/v Methane
HC Infrared 0-100%LEL General hydro carbons
HHC Infrared 0-100% v/v General hydro carbons CO2P Infrared 0-5000 ppm Carbon Dioxide CO2L Infrared 0-5% v/v Carbon Dioxide CO2H Infrared 0-100% v/v Carbon Dioxide
The sensor type is selected as follows:
From the menu system select menu option: E: 8 and press ENTER.
Using the NEXT and PREVIOUS buttons, select the required sensor.
Press ENTER.
N
OTE
: Pressing ENTER while in menu option E: 8 will reset the factory defaults for the displayed sensor. A
calibration must be performed even if the sensor type was not changed.
Press MENU to close the menu system.
ERIES
IR Sensor Selections
N
OTE
: It is important that the sensor selected via this menu option corresponds to the sensor that is installed. Incorrect
settings here may result in the inability of the instrument to detect gas.
11.2.9 E : 9 – Engineer diagnostics
This feature is a view-only feature. No configuration changes are possible from within this menu. The information is for use of ENMET Corporation personnel.
From the menu system select menu option: E: 9 and press ENTER.
Using the NEXT button, display the required setting. The displayed values are as follows:
E : 90 Gas level E : 93 Active sensor A to D counts
Press MENU to close the menu system.
11.2.10 E : 12 – Cross reference
Contact ENMET Corporation before entering this menu option.
This option is used to allow the user to calibrate the sensor with a commonly available gas (e.g. methane or propane) but use the unit to detect a different gas (e.g. methanol or acetone etc.). This is achieved by adjusting the cross-reference factor according to the difference in signal that is detected for the calibration gas compared to the target gas.
Select menu mode E:12 and press ENTER.
Using the UP and DOWN buttons set the required cross-reference factor as required.
Press ENTER to store the new value, Pressing MENU instead of ENTER aborts the feature.
Press MENU to return the instrument to its standard mode of operation.
N
OTE
: the default setting of 1.00 is used – no cross sensitivity to target gas.
Contact ENMET Corporation for details.
16
Page 20
EX-6120, 6165 SERIES IR
ENMET Corporation
11.2.11 E : 13 – Relay 1
The unit is fitted with a relay that is operated in conjunction with the alarm level. The user can select if the relay is normally Energized, E’ or normally de-energized, ‘d’ when the unit is not in an alarm condition.
From the menu system select menu option: E: 13 and press ENTER.
The display will show the following:
E:r Normally energized, rising alarm d:r Normally de-energized, rising alarm E:F Normally energized, falling alarm d:F Normally de-energized, falling alarm
The mode of operation can be changed by pressing the UP button.
Press ENTER to accept the new relay mode of operation.
N
OTE
: Pressing MENU instead of ENTER leaves the unit without change.
Press MENU to return the instrument to its standard mode of operation.
11.2.12 E : 14 – Relay 2
The unit is fitted with a relay that is operated in conjunction with the alarm level. The user can select if the relay is normally Energized, ‘E’ or normally de-energized, ‘d’ when the unit is not in an alarm condition.
From the menu system select menu option: E: 14 and press ENTER.
The display will show the following:
E:r Normally energized, rising alarm d:r Normally de-energized, rising alarm E:F Normally energized, falling alarm d:F Normally de-energized, falling alarm
The mode of operation can be changed by pressing the UP button.
Press ENTER to accept the new relay mode of operation.
N
OTE
: Pressing MENU instead of ENTER leaves the unit without change.
Press MENU to return the instrument to its standard mode of operation.
11.2.13 E : 15 – Alarm Level 1
The gas level that will cause Relay 1 to change state is set via this menu option.
From the menu system select menu option: E: 15 and press ENTER.
Using the UP and DOWN buttons, adjust the displayed reading so that it matches the desired alarm set point.
Press ENTER to accept the new alarm level.
N
OTE
: Pressing MENU instead of ENTER aborts the feature.
Press MENU to return the instrument to its standard mode of operation.
11.2.14 E : 16 – Alarm Level 2
The gas level that will cause Relay 2 to change state is set via this menu option.
From the menu system select menu option: E: 16 and press ENTER.
Using the UP and DOWN buttons, adjust the displayed reading so that it matches the desired alarm set point.
Press ENTER to accept the new alarm level.
N
OTE
: Pressing MENU instead of ENTER aborts the feature.
Press MENU to return the instrument to its standard mode of operation.
11.2.15 E : 19 – Positive Zero Suppression
This option is used to allow the user to suppress small amounts of positive sensor zero drift. The setting can be set between 0 and 10% of the sensor range as set by the FSD value.
Press MENU to open the menu system.
Using the NEXT and PREVIOUS buttons, select menu option: E:19
Press ENTER.
Using the INCREASE and DECREASE buttons, set the required zero suppression value.
Press ENTER to store the new value.
Note: Pressing the MENU button rather than the ENTER button exits without any change.
Press MENU to close the menu system.
17
Page 21
EX-6120, 6165 SERIES IR
ENMET Corporation
11.2.16 E : 20 – Negative Zero Suppression
This option is used to allow the user to suppress small amounts of negative sensor zero drift. The setting can be set between 0 and 10% of the sensor range as set by the FSD value.
Press MENU to open the menu system.
Using the NEXT and PREVIOUS buttons, select menu option: E:20
Press ENTER.
Using the INCREASE and DECREASE buttons, set the required zero suppression value.
Press ENTER to store the new value.
Note: Pressing the MENU button rather than the ENTER button exits without any change.
Press MENU to close the menu system.
11.2.17 E : 27 – Hysteresis
This option sets the alarm hysteresis. The relays will chatter if this level is set to 100. 95% is the usual setting for pellistor sensors.
11.2.18 E : 28 – Firmware update
Do Not Enter this Menu
11.2.19 E : 29 – Relay 3
The unit is fitted with a relay that is operated in conjunction with a fault condition. The user can select if the relay is normally Energized, ‘E’ or normally de-energized, ‘d’ when the unit is not in a fault condition.
From the menu system select menu option: E: 29 and press ENTER.
The display will show the following:
E:r Normally energized, rising alarm d:r Normally de-energized, rising alarm E:F Normally energized, falling alarm d:F Normally de-energized, falling alarm
The mode of operation can be changed by pressing the UP button.
Press ENTER to accept the new relay mode of operation.
N
OTE
: Pressing MENU instead of ENTER leaves the unit without change.
Press MENU to return the instrument to its standard mode of operation.
11.2.20 E : 30 – Password
Use this menu option to enable / disable the password feature. Place the EX-6120, 6165 S
Press the MENU to open the menu system.
Using the NEXT and PREVIOUS buttons, select menu option: E:30
Press ENTER.
The display shows either On or OFF
Press the UP button to select the desired setting.
Press ENTER to accept the setting.
N
OTE
: Pressing MENU instead of ENTER leaves the unit without change.
Press MENU to return the instrument to its standard mode of operation.
ERIES
IR in the password menu as follows:
Note: If the password is in operation then the user will be prompted with PASS when ever the menu key is pressed. Pressing
the MENU key again will result in the restricted user access, i.e. only the zero and span options will be available. Entering the correct password will give access to the full menu facility.
11.2.21 E : 31 – Warm-up timer
The sensor goes through a stabilization period when power is first applied. This option allows the user to adjust the time that is applied before readings are displayed / transmitted.
From the menu system select menu option: E: 31 and press ENTER.
Using the INCREASE and DECREASE buttons, set the required warm-up time.
Press ENTER to store the new value.
Note: Pressing the MENU button rather than the ENTER button exits without any change.
Normally warm time is 45 seconds.
18
Page 22
EX-6120, 6165 SERIES IR
+
4mA –
20mA
+
40mV –
200mV
Appendix A
A.1 Analog Output Calibration
The analogue output is that of a current source. In order to calibrate the output it is necessary to monitor the output signal. This can be performed in one of two ways:
Current measurement:
Connect an ammeter (or multimeter set to measure current in the mA range) in series with the analogue output.
FGD10 Analogue Output (4 – 20mA)
– +
mA
Advantage: Direct measurement of output. Disadvantage: The analogue output has to be disconnected to allow the connection of the meter.
Voltage measurement:
Permanently connect a 10R resistor in series with the analogue output. When calibration is required, connect a voltmeter (or multimeter set to measure voltage in the mV range) across the 10R resistor.
FGD10 Analogue Output (4 – 20mA)
– +
10R
mV
Advantage: No need to disturb wiring between EX-6120, 6165 S Disadvantage: Measurement accuracy dependent upon resistor tolerance.
A.1.1 Analogue Output Zero
Monitor the current sourced from the analogue output of the EX-6120, 6165 S
From the menu system select menu option: E: 4 and press ENTER.
Using the INCREASE and DECREASE buttons; adjust the output to the required level (4mA or 40mV).
Press ENTER.
N
OTE
: Pressing the MENU button rather than the ENTER button exits the zero feature without performing the
calibration.
Press the MENU button to close the menu system.
N
OTE
: The ZERO factor will be displayed momentarily on exit.
A.1.2 Analogue Output Span
Always zero the analogue output prior to performing a span operation.
From the menu system select menu option: E: 5 and press ENTER.
Using the INCREASE and DECREASE buttons; adjust the output to the required level.
Press ENTER button.
N
OTE
Press the MENU button to close the menu system.
: Pressing the MENU button rather than the ENTER button exits the span feature without performing the calibration.
N
OTE
: The ANALOGUE OUTPUT SPAN factor will be displayed momentarily on exit.
A.2 Error Codes
Not applicable to the Pellistor Catalytic Sensor variant
Control
Unit
Control
Unit
Meter range:
0.004A – 0.020A
Meter range:
0.04V – 0.20V
ERIES
IR and control unit.
ERIES
IR using a suitable method.
ENMET Corporation
19
Page 23
EX-6120, 6165 SERIES IR
EX-6120, 6165
EX-6120, 6165
ENMET Corporation
A.3 Communications Protocol
The communications protocol used by the EX-6120, 6165 S connected via an RS232 connection at 19200 baud, 8 data bits 1 stop bit, no parity. This point-to-point, P2P, protocol is a frame-based protocol.
A.3.1 Hardware connections
ERIES
IR is used for communications between devices
PC 9-way ‘D’ connector
2 Tx 3 Rx 5 0V
S
ERIES
IR
connection
PC 25-way ‘D’ connector
3 Tx 2 Rx 7 0V
S
ERIES
IR
connection
A.3.2 Control Byte Constants
The following control byte constants are used in the P2P protocol.
Read, RD = 0x13 (00010011)
Data Link Escape, DLE = 0x10 (00010000)
Write, WR = 0x15 (00010101)
Acknowledge, ACK = 0x16 (00010110)
Negative Acknowledge, NAK = 0x19 (00011001)
Single Data Frame, DAT = 0x1A (00011010)
End of Frame, EOF = 0x1F (00011111)
Write Password 1, WP1 = 0xE5 (11100101) Write Password 2, WP2 = 0xA2 (10100010)
A.4 Frame Structure
The start of a frame is indicated by a DLE byte followed by the type of frame to follow (RD, WR, ACK, NAK, DAT). The end of frame is indicated by a DLE byte followed by an EOF byte.
Note: Each of the constants has bit 4 set and so is slip-resistant (i.e. if shifted this bit will be out of position). The values have a Hamming Distance of 2 (each code is at least 2 bits different from every other code).
Any DLE bytes that occur between a frame’s start and end are prefixed with another DLE (byte-stuffing). Following the EOF is a 16-bit checksum of the entire frame, each byte is added to produce the checksum.
A.4.1 Variables
Each piece of accessible data on a device is referred to as a Variable. Each variable is referenced by a Variable ID. A variable ID may be any number of bytes long.
The available Variables and their corresponding Variable IDs depend on the type of device, but here are a few examples for the EX-6120, 6165 S
The structure of the data returned in each variable usually depends both on the type of device and the version of firmware running on the device.
Refer to device documentation for more information.
ERIES
IR:
General Configuration 0x00 Live Data 0x01 Zero Sensor 0x02 Span Sensor 0x03 Version Information 0x04 Firmware Update 0x05
20
Page 24
EX-6120, 6165 SERIES IR
A.4.2 Reading a Variable
Send a read frame with the Variable ID to be read:
DLE RD var-id DLE EOF Csum hi Csum lo
Byte stuffing
Device response on success, where requested variable data < 255 bytes:
DLE DAT data-len data DLE EOF Csum hi Csum lo
Byte stuffing
Device response on failure:
DLE NAK reason
Where ‘reason’ is a single byte failure code, the meaning of which depends on the device type, i.e. Reason = 1, NotReadable, Reason = 2, ReadOutOfRange
A.4.3 Read example - read live data
The following example is based on V4.x.xE where the data is either byte or integer types only. Send the following bytes:
DLE, RD, Variable ID, DLE, EOF, Checksum High byte, Checksum low byte i.e.
0x10, 0x13, 0x01, 0x10, 0x1F, 0x00, 0x53 Device response on success:
DLE, DAT, Data length, Data, DLE, EOF, Checksum High byte, Checksum low byte, i.e.
0x10 DLE 0x1A DAT 0x0D Data length 0x03 Version 0x01 Display mode 0x07 Sensor Type 0x00, 0x00 Gas reading (signed integer x 100) = 0 0xAE, 0x0B Current sensor temperature (signed integer x 100) = 29.9 0x92, 0x08 Active sensor signal (unsigned integer) = 2194 0xD7, 0x04 Reference sensor signal (unsigned integer) = 1239 0x10 DLE 0x1F EOF 0x02 Checksum high byte 0x9F Checksum low byte
A.4.4 Writing a Variable
Send a write frame with the Variable ID to be written:
DLE WR WP1 WP2 var-id DLE EOF Csum hi Csum lo
Byte stuffing
Device response on success:
DLE ACK
Where data to write is < 255 bytes, send a DAT frame:
DLE DAT data-len data DLE EOF Csum hi Csum lo
Byte stuffing
Device response on write success:
DLE ACK
Device response on write failure:
DLE NAK reason
Where ‘reason’ is a single byte failure code, the meaning of which depends on the device type, i.e. Reason = 1, NotWritable Reason = 2, WriteOutOfRange Reason = 3, BadDataLength Reason = 4, IncorrectVersion
ENMET Corporation
21
Page 25
EX-6120, 6165 SERIES IR
ENMET Corporation
WARRANTY
ENMET warrants new instruments to be free from defects in workmanship and material under normal use for a period of one year from date of shipment from ENMET. The warranty covers both parts and labor excluding instrument calibration and expendable parts such as calibration gas, filters, batteries, etc... Equipment believed to be defective should be returned to ENMET within the warranty period (transportation prepaid) for inspection. If the evaluation by ENMET confirms that the product is defective, it will be repaired or replaced at no charge, within the stated limitations, and returned prepaid to any location in the United States by the most economical means, e.g. Surface UPS/FedEx Ground. If an expedient means of transportation is requested during the warranty period, the customer is responsible for the difference between the most economical means and the expedient mode. ENMET shall not be liable for any loss or damage caused by the improper use of the product. The purchaser indemnifies and saves harmless the company with respect to any loss or damages that may arise through the use by the purchaser or others of this equipment.
This warranty is expressly given in lieu of all other warranties, either expressed or implied, including that of merchantability, and all other obligations or liabilities of ENMET which may arise in connection with this equipment. ENMET neither assumes nor authorizes any representative or other person to assume for it any obligation or liability other than that which is set forth herein.
NOTE: When returning an instrument to the factory for service:
Be sure to include paperwork. A purchase order, return address and telephone number will assist in the expedient repair and return of your unit. Include any specific instructions. For warranty service, include date of purchase If you require an estimate, please contact ENMET Corporation.
There are Return for Repair Instructions and Form on the last pages of this manual. This Form can be copied or used as needed.
Manual Part Number 80003-123
April, 2013
Notes:
22
Page 26
PO Box 979 680 Fairfield Court Ann Arbor, Michigan 48106-0979
734.761.1270 Fax 734.761.3220
Returning an Instrument for Repair
ENMET instruments may be returned to the factory or any one of our Field Service Centers for regular repair service or calibration. The ENMET Repair Department and Field Service Centers also perform warranty service work.
When returning an instrument to the factory or service center for service, paperwork must be included which contains the following information:
A purchase order number or reference number.  A contact name with return address, telephone and fax numbers  Specific instructions regarding desired service or description
of the problems being encountered.
Date of original purchase and copy of packing slip or invoice
for warranty consideration.
If a price estimate is required, please note it accordingly and be
sure to include a fax number.
Providing the above information assists in the expedient repair and return of your unit.
Failure to provide this information can result in processing delays.
ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour. All instruments sent to ENMET are subject to a minimum evaluation fee, even if returned unrepaired. Unclaimed instruments that ENMET has received without appropriate paperwork or attempts to advise repair costs that have been unanswered, after a period of 60 days, may be disposed of or returned unrepaired COD with the evaluation fee.
Service centers may have different rates or terms. Be sure to contact them for this information.
Repaired instruments are returned by UPS/FedEx Ground and are not insured unless otherwise specified. If expedited shipping methods or insurance is required, it must be stated in your paperwork.
Note: Warranty of customer installed components.
If a component is purchased and installed in the field, and fails within the warranty term, it can be
returned to ENMET and will be replaced, free of charge, per ENMETs returned goods procedure.
If the entire instrument is returned to ENMET Corporation with the defective item installed, the
item will be replaced at no cost, but the instrument will be subject to labor charges at half of the standard rate.
Page 27
Mailing A
ddress:
Shipping Address:
Repair Return Form
ENMET Corporation PO Box 979 Ann Arbor, Michigan 48106
Phone Number: 734.761.1270 FAX Number: 734.761.3220
Your Mailing Address:
ENMET Corporation Attn: Repair Department 680 Fairfield Court Ann Arbor, Michigan 48108
Your Shipping Address:
Contact Name: __________________________ Your Phone: _______________________ Your PO/Reference Number: _______________ Your FAX: _______________________
Payment Terms: K COD
(Check one) K VISA / MasterCard______________________ ________ ________
Card number Expiration Card Code
K American Express______________________ ________ ________
Card number Expiration Card Code
Name as it appears on the credit card___________________________________
Return Shipping Method:
K UPS: K Ground K 3 Day Select K Next Day Air K ND Air Saver K 2-Day Air K UPS Account number: ________________________
K Federal Express: K Ground K Express Saver K P-1 K Standard K 2-Day Air K FedEx Account number: ________________________
Would you like ENMET to insure the return shipment? K No K Yes Insurance Amount: $_________________
Loading...