1.2 Check Order ....................................................................................................................................................................... 1
1.3 Serial Numbers................................................................................................................................................................... 1
2.0 F
EATURES OF THE
3.0 I
NSTALLATION OF THE
3.1 Mounting the EX-5155-MOS Enclosure ........................................................................................................................... 3
3.2 Wiring the EX-5155-MOS to a Control Unit ..................................................................................................................... 4
4.0 O
PERATION OF THE
4.1 Start up ............................................................................................................................................................................... 6
4.1.1 Typical Start Up .............................................................................................................................................................................. 6
4.1.2 Alternate Start Up ........................................................................................................................................................................... 6
5.1 Maintenance Menu ............................................................................................................................................................. 8
5.2 Calibration of the EX-5155-MOS ...................................................................................................................................... 9
5.2.1 Zero Adjust.................................................................................................................................................................................... 11
5.2.2 Gas Span ....................................................................................................................................................................................... 11
5.2.3 Exit Maintenance Menu ................................................................................................................................................................ 11
5.3 Heater Voltage Settings ................................................................................................................................................... 13
: [important information about use of instrument – if not followed may have to redo some steps.]
C
W
ARNING
AUTION
: [affects equipment – if not followed may cause damage to instrument, sensor etc…]
:
[affects personnel safety – if not followed may cause bodily injury or death.]
Page 3
EX-5155-MOS ENMET Corporation
1.0 Introduction
The ENMET EX-5155-MOS sensor/transmitters (S/T) is a 3-wire, 24 VDC 4-20mA S/T for the detection of toxic gas, utilizing
a Metal Oxide Semiconductor (MOS) sensor. The EX-5155-MOS is meant to be used in conjunction with an appropriate
power supply and controller. The ENMET EX-5155-MOS sensor/transmitter has been designed and approved to be used in a
Class I, Div. 1, Groups B, C, D, classified areas. The approval was issued by CSA International.
N
OTE
: All specifications stated in this manual may change without notice.
1.1 Unpack
Unpack the EX-5155-MOS and examine it for shipping damage. If such damage is observed, notify both ENMET customer
service personnel and the commercial carrier involved immediately.
Regarding Damaged Shipments
N
OTE
: It is your responsibility to follow these instructions. If they are not followed, the carrier will not honor
any claims for damage.
This shipment was carefully inspected, verified and properly packaged at our company and delivered to the carrier in
good condition.
When it was picked up by the carrier at ENMET, it legally became your company’s property.
If your shipment arrives damaged:
•Keep the items, packing material, and carton “As Is.” Within 5 days of receipt, notify the carrier’s local office and
request immediate inspection of the carton and the contents.
•After the inspection and after you have received written acknowledgment of the damage from the carrier, contact
ENMET Customer Service for return authorization and further instructions. Have your Purchase Order and Sales
Order numbers available.
ENMET either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability coverage,
usually $100.00. Repair or replacement charges above that value are your company’s responsibility.
The shipping company may offer optional insurance coverage. ENMET only insures shipments with the shipping
company when asked to do so in writing by our customer. If you need your shipments insured, please forward a written
request to ENMET Customer Service.
Regarding Shortages
If there are any shortages or questions regarding this shipment, please notify ENMET Customer Service within 5 days of
receipt at the following address:
ENMET Corporation
680 Fairfield Court
Ann Arbor, MI 48108
734-761-1270 734-761-3220 Fax
1.2 Check Order
Check, the contents of the shipment against the purchase order. Verify that the EX-5155-MOS is received as ordered. [Each
EX-5155-MOS is labeled with its target gas.] If there are accessories on the order, ascertain that they are present. Check the
contents of calibration kits. Notify ENMET customer service personnel of any discrepancy immediately.
1.3 Serial Numbers
Each EX-5155-MOS is serialized. These numbers are on tags on the equipment and are on record in an ENMET database.
1
Page 4
EX-5155-MOS ENMET Corporation
PO
T 3
See Note
POT 2
Note:
3 Alarm
Power /Fault
M
S
Magnetic switch
2.0 Features of the EX-5155-MOS
See Figure 1 for location of features:
Feature Description
Display
Gain Potentiometer
(POT)
Visual Alarms
Magnetic Switches
Sensor
LCD: Indicates the level of gas detected by sensor
POT 1: Display contrast adjustment
POT 2: Does Not apply to MOS, not used Do not adjust
POT 3: Does Not apply to MOS, not used Do not adjust
POT 4: MOS Heater Voltage, Do not adjust unless advised by ENMET
LED indicators:
Power / Fault Indicator LED, Green / Red
Alarm (3) Indicator Red LED, user adjustable
M
ENU
: Advances the instrument display through menus (Zero, Span, Exit)
S
ELECT
: Selects the Zero, Span, Exit menu or sets proper calibration values for Zero or Span
For sensing gas at
PPM
or LEL levels, see Table 3 for sensor types
Magnetic switches control the instrument maintenance functions. The switch locations are indicated by M
ENU
and S
magnetic field pulse is applied by momentarily putting the end of the magnet in proximity to the switch and then removing it.
Referred to as tap. Since the magnetic field penetrates the window, the enclosure cover is not removed in order to perform
calibration.
Three alarm points are preprogrammed into the EX-5155-MOS sensor/transmitters. At each alarm point, an LED on the front
panel is activated. These internal alarm settings are independent of the 4-20mA output alarm values that can be set at a
controller.
POT 1
Indicator
Indicators
Display
See Note
POT 4
Menu
ENU
Magnetic switch
Select
Internal PCB View
Sensor
POT 2 and POT 3 are not
ELECT
used with EX-5155 MOS
Do Not Adjust
ELECT
. A
External View
Figure 1: EX-5155-MOS Features
2
Page 5
EX-5155-MOS ENMET Corporation
Mounting Holes
Sensor/Transmitter
Sensor/Transmitter
3.0 Installation of the EX-5155-MOS
C
AUTION
C
AUTION
3.1 Mounting the EX-5155-MOS Enclosure
C
AUTION
:
Area must be declassified during installation.
The ENMETEX-5155-MOS gas sensor/transmitter (S/T) is a 3-wire, 24 VDC, 4-20 mA S/T for the detection of toxic gas. The
S/T is meant to be used in conjunction with an appropriate power supply and controller. The ENMET EX-5155-MOS
sensor/transmitter has been designed and approved to be used in a Class I, Div. 1, Groups B, C, D, classified areas. The
approval was issued by CSA International. Appropriate wiring, conduit and fittings are required for proper installation in a
explosion proof rated environment.
: Since the sensor/transmitter detects gas only at the sensor location, pay attention to the possible sources of gas, the
density of the gas, locations where the gas may be confined and locations where the gas may damage or injure
property or personnel, when choosing locations of sensor/transmitters.
Also, take into consideration environmental factors when deciding on S/T location. Avoid locations where the S/T
may be damaged by liquid immersion, excessive heat or other know hazards. Also, take precautions to insure
condensation inside of the conduit does not enter the S/T.
Mount the enclosure, using the two mounting holes provided see Figure 2. Pay particular attention to the source and density of
the gas being detected when choosing the location. Mount the S/T near the ceiling for lighter than air gases /vapors and near
the floor for heavier then air gas/vapors. Contact ENMET if you have questions regarding your application.
: Before connecting S/T to controller remove the power source to controller. Failure to do so may cause damage to
sensitive components.
Optional
Conduit fitting
Typically: ½″ NPT female
Sensor
½″ NPT female
2 places, 0.313″
Me
Sel
Enclosure Top View
Enclosure Side View
Sensor
Figure 2: EX-5155-MOS Mounting Dimensions
3
Page 6
EX-5155-MOS ENMET Corporation
J4
J8
Wires to
3.2 Wiring the EX-5155-MOS to a Control Unit
C
AUTION
*Contact ENMET for Modbus address information
:
Area must be declassified during installation.
Run conduit and 16
AWG
(1.5
2
MM
) wires to the enclosure from the power supply and controller. If the EX-5155-MOS is
installed in a hazardous location as defined by the National Electrical Code, then ALL wiring must be in accordance with the
National code and any local governing codes.
Open the enclosure, and remove the 2 screws that retain the display overlay to the circuit board.
Use caution when removing the over lay. Do not damage the magnetic switches.
Remove the two overlay standoffs and remove the circuit board, exposing the terminal strips on the bottom of the circuit board.
Do not disconnect the circuit board wiring.
Connect the wires from the controller (power supply) to the supplied J4 plug then attach to J4 terminal.
Connect the wires from the sensor to the supplied J8 plug then attach to the J8 terminal.
See Figure 3 for locations
When wiring is complete, reassemble the
magnetic switches. Put the cover back on the S/T
EX-5155-MOS
(PCB)
. Use caution when installing the overlay so as not to damage the
Do Not
apply power to the S/T without the cover in place.
5
Page 8
EX-5155-MOS
EX-5155
-
MOS
W
C, F, P
R
EX
-
5155
76
-
1256
0ppm
0ppmW
0ppmF
0ppmP
ENMET Corporation
4.0 Operation of the EX-5155-MOS
It is best to have the
gas to them.
When the
instrument model number, serial number and software revision. After all of the momentary screens have been displayed, the
instrument arrives at the Main Gas Display showing the gas concentration and unit of measurement.
Depending on transmitter configuration and calibration condition, the furthest right character in the display may flash a letter
indicating the instrument status. See the Section 4.1.2 below.
EX-5155-MOS
4.1 Start up
4.1.1 Typical Start Up
When power is supplied to the
N
OTE
:
Software revision may cause variations of display output.
Example of Display Function
The instrument: Serial Number
EX-5155-MOS
transmitter is first powered up, it goes through a series of momentary screens, which identify the
transmitters powered up and operational for 24 hours before applying calibration or test
EX-5155-MOS,
the S/T will display the following sequence of information:
The instrument: Model
Note MOS is not displayed
The instrument: Software Revision
IF
the right most character is a flashing
For Toxic Gas
IF
the right most character is a flashing
S/W X.X
The instrument is in Warm-up mode
This should last about 1 minute
The Signal Output is held at 4mA during warm-up
The instrument: Normal Display Mode
Measurement of target Gas
or
See Section 4.1.2
4.1.2 Alternate Start Up
Depending on
a letter indicating the instrument status. See the table below.
Purging occurs automatically for instruments with sensors that require purging. Purge will occur for 5 minutes on instrument
start-up, followed by a 5-minute recovery period. At the end of the recovery period, the transmitter should be ready for
operation. If additional purge time is required, refer to Section 4.1.3 for more information.
Example of Display (may also display 0 LEC) Function
IF
the right most
character is a flashing C
IF
the right most
character is a flashing F
IF
the right most
character is a flashing P
IF
the right most
character is a flashing R
EX-5155-MOS
S/T configuration and calibration condition, the furthest right character in the display may flash
The last calibration of the instrument was invalid
0ppmC
0ppmR
The instrument must be recalibrated
There is a sensor fault
The sensor is being purged:
This function is required for certain sensor types
The duration of purge cycle varies with sensor type
The Signal Output is held at 4mA during purge
The instrument is in Recovery mode after completing the
purge cycle
This should take about 5 minutes
4.1.3 Purge
Purging is a function that temporarily increases the sensor heater voltage to clean off contaminants. Some sensors operate at
higher heater voltages where purging is not required. The
sensor purging requirements.
EX-5155-MOS
S/T is configured at the factory for the installed
6
Page 9
EX-5155-MOS
ENMET Corporation
Purge times vary form 5 – 25 minutes depending on the sensor installed.
Some sensors require additional purge time after start up. Particularly if the sensor is new or the system powered down for an
extended period of time.
To initiate a purge:
See
Figure 5
Turn the power off and back on. This will initialize a purge cycle.
for the Operational portion of the Maintenance Menu Flow Chart.
4.2 Normal Display Mode
When the
and the information on the display is measurement of the target gas detected by the
LEDs are not lit.
To advance through displays of operational information tap the magnet over the
N
OTE
See sequence of operational information below:
Display Measurement of the target gas
Tap the magnet over the
Display indicates Alarm 1 Set point
Tap the magnet over the
Display indicates Alarm 2 Set point
Tap the magnet over the
Display indicates Alarm 3 Set point
Tap the magnet over the
Display indicates mA Span range
(Full Scale)
Tap the magnet over the
Display returns to gas measurement
EX-5155-MOS
:
Software revision may cause variations of display output.
is installed as described in section 3, and in clean air, the
M
ENU
button
0PPM
MENU
M
ENU
button
A1: 10
MENU
M
ENU
button
A2: 20
MENU
M
ENU
button
A3: 50
MENU
mA: 100
M
ENU
button
MENU
SELECT
SELECT
SELECT
SELECT
SELECT
POWER
EX-5155-MOS.
M
ENU
button.
No Function for the
green LED is on, the display is lit
The red alarm and fault
S
ELECT
button
in this mode
Operational Display Flow Chart
4.2.1 Alarm Conditions EX-5155-MOS
There are three alarm set points available. These alarm set points can be changed within limits; see the maintenance section of
this manual for the procedure.
If the gas concentration increases above that of the alarm set point, the associated red LED is lit.
7
Page 10
EX-5155-MOS
5ppm
Exit
ENMET Corporation
5.0 Maintenance of the EX-5155-MOS
C
AUTION
C
AUTION
5.1 Maintenance Menu
:
Do not open the
: Do Not Attempt A Span Procedure Without Calibration Gas Applied to The Sensor
EX-5155-MOS
S/T
in a classified area.
; if this is done, the S/T is forced
into a calibration fault mode.
Magnetic switches control the
display panel, see
Figure 3.
numbers such as alarm points, calibrations gas, etc. The
MENU
The
and
S
ELECT
functions. The
M
ENU
switch is used to display the various menu options and make incremental changes to
S
ELECT
M
ENU
and
S
ELECT
switch locations are indicated on the
switch is used to select that option, set zero or span digit.
Most maintenance functions are controlled by simple taps of the supplied magnet on the transmitter glass, below the
S
ELECT
boxes on the front panel.
To enter the maintenance menu hold the magnet over the
Table 1
indicates the maintenance menu sequence see
M
ENU
Figure 5
switch for 2 to 4 seconds
for a detailed maintenance menu flow chart.
Table 1: EX-5155-MOS maintenance Menus Sequence
Example of Display Function
Hold the magnet over
M
ENU
switch for 2 – 5 seconds to enter the Maintenance Menu
The Power/Fault LED will flash Green – Red to indicate the
Tap the magnet over the
M
ENU
switch to advance to the Zero procedure
Zero
Normal Display Mode
Measurement of target gas
EX-5155-MOS
is in Maintenance Mode
To exit the maintenance Menu and return to the Normal
Display Mode:
If intended function Tap the magnet over
For adjusting Zero:
If intended function Tap the magnet over
S
ELECT
S
ELECT
switch
switch
M
ENU
and
Tap the magnet over the
M
ENU
switch to advance to the Span procedure
Span
For adjusting the Span:
If intended function Tap the magnet over
Tap the magnet over the
Alarm1
M
ENU
switch to advance to each Alarm set point procedures
Alarm2
For adjusting the Alarm 1, 2 and 3 set points:
If Intended function Tap the magnet over
Alarm3
Tap the magnet over the
Taping the
You must Tap the
N
OTE
M
ENU
switch without taping the
S
ELECT
:
If the S/T fails to respond, the magnet may have become weak and may need to be replaced.
M
ENU
switch to advance the mA Span set point procedure
For adjusting the mA Span set point:
mA Span
If intended function Tap the magnet over
S
ELECT
switch will allow you to cycle through the menu options.
switch in order to change the desired operation.
S
ELECT
S
ELECT
S
ELECT
switch
switch
switch
8
Page 11
EX-5155-MOS
Output, for Gas
Input, for Calibration
ENMET Corporation
5.2 Calibration of the EX-5155-MOS
Calibration is the process of setting the instrument up to read accurately when exposed to a target gas. The Zero function sets
the clean air reference point and the Span function sets the sensitivity of the instrument.
Initial Calibration:
calibration. The S/T has been pre-calibrated at the factory, and initial field calibration should result in only fine tuning to
circuit, as well as a way to check that installation is successful. It is not necessary to open the enclosure to make adjustment.
The calibration functions are operated with magnets from outside the enclosure through the
Not open the S/T unless the area is de-classified.
Calibration Zero and Span functions are two separate procedures. They operate independently of each other. It is
recommended that the Zero procedure be done prior to the Span procedure.
quarterly calibration of the
Calibration equipment is available from
calibration adapter will have a fitting for the gas cylinder on one side, and a cover to go over the sensor housing on the other.
Generally, a cylinder of 20.9% Oxygen is used to provide a fresh air reference or Zero point for the calibration. Another
cylinder is used to provide the Span reference point for calibration. Depending on the instrument calibration, the Span gas may
be the same gas that the instrument is calibrated to display, or it may be another gas, which
response.
Sensors require a humidified calibration gas sample. Fill the humidifier bowl, half way up with clean fresh water
prior to attaching the Zero or Span gases. Be careful not to let the humidifier bowl tip, allowing water to enter the gas delivery
tubing.
See
Table 3
Wait 24 hours after initially supplying power to the
A ZERO function should be performed only when the
sensor is in question, use a cylinder of 20.9% oxygen to provide a clean air reference. Attach the cylinder to the calibration
adapter, fill the humidifier bowl halfway with water and allow gas to flow over the sensor for 3 – 4 minutes.
Enter the maintenance menu by placing the magnet over
Maintenance Menu flow chart.
The second menu available is the Zero.
Tap the
S
ELECT
switch to perform a Zero.
If the Zero is successful
, Cal OK appears on the display and in 1 – 2 seconds, display will change to Span.
If you wish to Span the sensor Tap the
S
ELECT
If you wish to Exit the maintenance menu, Tap
instrument Normal Gas Display
If the Zero is Not successful,
sensor is outside of safe parameters to be zeroed, the display will read Bad Zero. Repeat
Section 5.2.1 Zero Adjust making sure to use a cylinder of 20.9% Oxygen.
EX-5155-MOS
M
ENU
switch for 2 to 4 seconds. See
sensor/transmitter is exposed to fresh air. If the air at the
switch you are now ready to apply gas.
M
ENU
switch until Exit is displayed, then tap
Figure 5, EX-5155-MOS
Proceed to gas span step 2
S
ELECT
switch to return to the
5.2.2 Gas Span
It is recommended that the Zero Function be performed first.
Enter the maintenance menu. See
1.
2.
Tap the
Tap the
M
ENU
switch once to show Span on the display.
S
ELECT
switch to perform a Span procedure. The display will alternate between the calibration gas concentration
Figure 5, EX-5155-MOS
and a signal level.
N
OTE
:
You can change the Calibration Gas Level. H
The M
ENU
The S
3.
Attach the associated calibration gas cylinder to the regulator and calibration cover. See to
4.
Open the valve to apply the calibration gas to the sensor.
5.
Watch for the signal level to stabilize. Refer to
6.
Once the signal level has stabilized, the
If the Span is successful, Cal OK appears on the display momentarily, then advances to Alarm 1. Remove calibration gas.
To exit maintenance menu tap the
switch changes digit indicated by underscore cursor
ELECT
switch locks underscored digit and moves to next digit
EX-5155-MOS
M
ENU
OLD
Table 4
switch until Exit appears, then tap the
Maintenance Menu flow chart.
the magnet over the S
ELECT
switch for 2 – 4 seconds
Figure 3
for typical response times.
will automatically lock in the calibration data and:
S
ELECT
switch.
.
If the sensor is outside of acceptable parameters, Bad Span is displayed momentarily, then returns to Span.
Remove calibration gas. Tap the
repeat calibration in 30 – 60 minutes.
If the sensor did not respond to gas, Same mV is displayed momentarily, then returns to Span.
Remove calibration gas, tap the
again in 30-60 minutes.
If the sensor will not calibrate See Section 5.4.
N
OTE
:
Some software revisions require the
7.
Calibration is complete.
MOS PPM
5.2.3 Exit Maintenance Menu
Exit maintenance, by tapping on the
instrument Normal Gas Display.
M
ENU
switch until Exit appears, then tap the
M
ENU
switch until Exit appears, then tap the
S
ELECT
switch be tapped to accept the signal.
Table 2: Calibration Time
3 – 4
minutes
M
ENU
switch until Exit appears on the display. Tap the
11
S
ELECT
S
switch. Check span gas and
ELECT
switch and try calibration
S
ELECT
switch to return to the
Page 14
EX-5155-MOS
H
S
If
MOS will
Apply
Bad Sens
Select
SetTD
sec
0
Alarm1
20
Normal Display Mode
5ppm
MENU
OLD the magnet over the MENU switch for 2 – 4 seconds
to enter the Maintenance Menus
ENMET Corporation
Exit
SELECT
Tap the
ELECT switch to return to the Normal
Display Mode. See Section 5.2.3
MENU
Tap the magnet over the MENU switch to cycle through Maintenance Menus
Zero
MENU
SELECT
PV: 0
Cal OK
OR
Bad ZERO
SELECT
Tap the SELECT switch to initiate Zero adjustment
the Zero signal is within Preset Specs the EX-5155-MOS will
display Cal OK, See Section 5.2.1
If the Zero signal is not within Preset Specs the EX-5155-
display Bad ZERO
NOTE: Some software revisions require the SELECT switch be
tapped to accept the signal.
Span
MENU
SELECT
PV: 0
20
SELECT
Cal OK
NOTE: You can change the Calibration Gas Level.
SELECT
The MENU switch changes digit indicated by underscore cursor
The SELECT switch locks underscored digit and moves to next digit
HOLD the magnet over the SELECT switch
See Section 5.2.2
Calibration Gas until signal value becomes
stable (about 1 to 4 minutes) See Figure 4
OR
Same mV
OR
If cal is good display will indicate OK or Same
If cal is not within preset “range” display will
indicate Bad Sens
NOTE: Some software revisions require the SELECT switch
be tapped to accept the signal.
Menu
SELECT
MENU
Alarm2
SELECT
MENU
Alarm3
SELECT
MENU
Figure 5: Continued on next page
Λ 10
Λ
SetTDsec
0
Λ 50
SetTDsec
0
To change Alarm set points:
Tap Menu switch until Alarm to be changed is displayed
Tap Select switch to display the set point
The M
ENU
switch: changes digit indicated by underscore
cursor
The S
ELECT
switch: locks in the underscored digit and
moves to next digit
If change is not within range display returns to first digit
If change is within range display moves to Set Time Delay
Use M
ENU
and S
ELECT
switches as above to change time
delay. Between 0 and 5 seconds is allowed
If change is within range display moves to next menu
Λ - Indicates increasing alarm
V - Indicates decreasing alarm
12
Page 15
EX-5155-MOS
100
To
Set Screw
Sensor
Spacer
ENMET Corporation
mA Span
SELECT
MENU
return to Normal Gas Display:
To change mA Span set point:
Tap the MENU switch until mA Span is displayed
Tap the SELECT switch to display the set point
The MENU switch changes digit indicated by underscore cursor
The SELECT switch locks underscored digit and moves to next digit
Tap MENU switch until EXIT is displayed
Then tap SELECT switch
Figure 5: EX-5155-MOS Maintenance Menu Flow chart
5.3 Heater Voltage Settings
Heater Voltages are necessary for MOS sensors. They are preset at the factory and should not require field adjustment. Do not
adjust these voltages unless specifically instructed to do so by
C
AUTION
:
Improper adjustment of heater voltages can damage sensors voiding any warranties and also alter the operating
characteristics of the sensor in such a way that the
ENMET
Corporation Technical Support Staff.
EX-5155-MOS
may not respond to it’s target gas.
5.4 Sensor Replacement
C
AUTION
Sensors should be replaced when they can no longer be calibrated. Replacement sensor part numbers are listed in Section 6.0
of this manual. If you do not know the proper part number for your sensor, be sure to have the
available when contacting your Distributor or
To replace a sensor, it is not necessary to open the transmitter housing.
Remove the set screw from sensor housing base.
Unscrew the sensor housing cover and remove spacer.
Unplug the sensor form PC Board.
Plug new sensor into PC Board and replace spacer.
Reassemble the sensor housing.
After the new sensor has been installed, it is suggested to allow the sensor to stabilize for 24 hours.
A Factory calibration must be performed.
After entering the Maintenance menu, advance to the Zero menu. Then while viewing the Zero menu, hold the magnet over the
M
After 2-4 seconds, an F will appear on the far right hand side of the display. The F indicates that the instrument is in Factory
mode.
Perform the calibration Zero and Span procedures as outlined in Section 5.2. Be sure that the F is present when selecting the
Zero and Span functions.
The Factory calibration sets a calibration window for future standard instrument calibrations.
Only perform a factory calibration when installing a new sensor!!
: Area must be declassified during sensor replacement.
ENU
switch for 2-4 seconds.
ENMET
(0.050 inch, 1.27mm Hex Key)
Corporation Technical Support.
Note the orientation of spacer
Replace spacer with grooved edge toward sensor housing cover
EX-5155-MOS
serial number
.
.
Sensor Housing Base
Note the orientation of spacer for replacement
Grooved edge of spacer toward Sensor Housing Cover
year from date of shipment from
expendable parts such as calibration gas, filters, batteries, etc... Equipment believed to be defective should be returned to
ENMET
product is defective, it will be repaired or replaced at no charge, within the stated limitations, and returned prepaid to any
location in the United States by the most economical means, e.g. Surface UPS/FedEx Ground. If an expedient means of
transportation is requested during the warranty period, the customer is responsible for the difference between the most
economical means and the expedient mode.
the product. The purchaser indemnifies and saves harmless the company with respect to any loss or damages that may arise
through the use by the purchaser or others of this equipment.
This warranty is expressly given in lieu of all other warranties, either expressed or implied, including that of merchantability,
and all other obligations or liabilities of
nor authorizes any representative or other person to assume for it any obligation or liability other than that which is set forth
herein.
NOTE: When returning an instrument to the factory for service:
There are Return for Repair Instructions and Form on the last pages of this manual. This Form can be copied or used as needed.
warrants new instruments to be free from defects in workmanship and material under normal use for a period of one
ENMET
within the warranty period (transportation prepaid) for inspection. If the evaluation by
Be sure to include paperwork.
A purchase order, return address and telephone number will assist in the expedient repair and return of your unit.
Include any specific instructions.
For warranty service, include date of purchase
If you require an estimate, please contact
. The warranty covers both parts and labor excluding instrument calibration and
ENMET
ENMET
ENMET
ENMET
shall not be liable for any loss or damage caused by the improper use of
which may arise in connection with this equipment.
Corporation.
confirms that the
ENMET
neither assumes
14
Page 17
EX-5155-MOS
ENMET Corporation
Addendum: Ammonia Calibration
Calibration Procedure for EX-5155-MOSAmmonia
The
EX-5155-MOS
All
MOS
sensors require at least 15% RH for proper operation. RH of 50% or greater is ideal for
common technique for humidifying dry gas from a cylinder is to bubble it through a humidifier assembly. However, ammonia
is readily absorbed in water.
ENMET
water, but not bubble the gas sample. Instead the gas sample is passed over the water. It is still able to pick up enough
humidity to properly calibrate the sensor, but not enough to lose all of the ammonia calibration gas.
For 0 – 500 ppm range Ammonia transmitters, the proper calibration gas cylinder is 300 ppm NH3 in air. While flowing the
300 ppm ammonia through the calibration adapter humidifier assembly,
going to the sensor
Place the
the
Do Not
Open the valve on the calibration adapter regulator, and attach the calibration cup to the sensor. Verify that the gas sample is
Apply the calibration gas for four minutes.
At the end of four minutes, use the magnet and
The display will alternate between calibration gas and concentration.
Tap the
The display will read CAL OK if the calibration was successful.
Fresh clean water must be used for each calibration.
Corporation has developed a special calibration adapter for ammonia that allows the humidifier bowl to be 1/3 full of
EX-5155-MOS
EX-5155-MOS
start the Span sequence.
passing over the water by looking for gentle ripples on the water surface and not bubbling. Flow rate is preset by the
regulator and needs no adjustment.
S
ELECT
Ammonia sensor requires special calibration techniques that are not common with other gases.
MOS
sensor calibration. The
approximately 200 ppm is the actual concentration
, the difference being absorbed by the water surface.
unit into calibration mode reference
will be at SPAN.
TAP
the
switch again to exit Span operation.
S
Section 5.0
ELECT
switch to enter the Span operation.
. Following the Zero Calibration
Notes:
15
Page 18
PO Box 979
680 Fairfield Court
Ann Arbor, Michigan 48106-0979
734.761.1270 Fax 734.761.3220
Returning an Instrument for Repair
ENMET instruments may be returned to the factory or any one of our Field Service Centers for regular repair
service or calibration. The ENMET Repair Department and Field Service Centers also perform warranty
service work.
When returning an instrument to the factory or service center for service, paperwork must be included which
contains the following information:
A purchase order number or reference number.
A contact name with return address, telephone and fax numbers
Specific instructions regarding desired service or description
of the problems being encountered.
Date of original purchase and copy of packing slip or invoice
for warranty consideration.
If a price estimate is required, please note it accordingly and be
sure to include a fax number.
Providing the above information assists in the expedient repair and return of your unit.
Failure to provide this information can result in processing delays.
ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest
tenth of an hour. All instruments sent to ENMET are subject to a minimum evaluation fee, even if returned
unrepaired. Unclaimed instruments that ENMET has received without appropriate paperwork or attempts to
advise repair costs that have been unanswered, after a period of 60 days, may be disposed of or returned
unrepaired COD with the evaluation fee.
Service centers may have different rates or terms. Be sure to contact them for this information.
Repaired instruments are returned by UPS/FedEx Ground and are not insured unless otherwise specified. If
expedited shipping methods or insurance is required, it must be stated in your paperwork.
Note: Warranty of customer installed components.
If a component is purchased and installed in the field, and fails within the warranty term, it can be
returned to ENMET and will be replaced, free of charge, per ENMET’s returned goods procedure.
If the entire instrument is returned to ENMET Corporation with the defective item installed, the item will
be replaced at no cost, but the instrument will be subject to labor charges at half of the standard rate.
Page 19
Mailing Address:
Shipping Address:
Repair Return Form
ENMET Corporation
PO Box 979
Ann Arbor, Michigan 48106
Contact Name: __________________________ Your Phone: _______________________
Your PO/Reference Number: _______________ Your FAX: _______________________
ENMET Corporation
Attn: Repair Department
680 Fairfield Court
Ann Arbor, Michigan 48108