1.2 Check Order ....................................................................................................................................................................... 1
1.3 Serial Numbers................................................................................................................................................................... 1
2.0 F
EATURES OF THE
3.0 I
NSTALLATION OF THE
3.1 Mounting the EX-5130-IS Enclosure ................................................................................................................................. 3
3.2 Wiring the EX-5130-IS to a Control Unit .......................................................................................................................... 4
4.0 O
PERATION OF THE
4.1 Start up ............................................................................................................................................................................... 5
Typical Start Up ....................................................................................................................................................................................... 5
4.2 Normal Display Mode ........................................................................................................................................................ 6
5.1 Maintenance Menu ............................................................................................................................................................. 7
5.2 Calibration of the EX-5130-IS ........................................................................................................................................... 8
5.2.1 Zero Adjust...................................................................................................................................................................................... 9
5.2.2 Gas Span ......................................................................................................................................................................................... 9
5.2.3 Exit Maintenance Menu .................................................................................................................................................................. 9
Table 2: Examples of Calibration Gas ....................................................................................................................... 8
Figure 5: EX-5130-IS Maintenance Menu Flow chart .............................................................................................. 10
: [important information about use of instrument – if not followed may have to redo some steps.]
C
W
ARNING
AUTION
: [affects equipment – if not followed may cause damage to instrument, sensor etc…]
:
[affects personnel safety – if not followed may cause bodily injury or death.]
Page 3
EX-5130-IS ENMET Corporation
1.0 Introduction
The ENMET EX-5130-IS sensor/transmitters (S/T) is a three wire 4-20mA S/T for the detection of toxic gas.
The EX-5130-IS is meant to be used in conjunction with an appropriate power supply and controller.
The ENMET EX-5130-IS sensor/transmitter has been designed to be used in a Class I, Div. 1, Groups B, C, D,
classified areas.
N
OTE
: All specifications stated in this manual may change without notice.
1.1 Unpack
Unpack the EX-5130-IS and examine it for shipping damage. If such damage is observed, notify both ENMET
customer service personnel and the commercial carrier involved immediately.
Regarding Damaged Shipments
N
OTE
: It is your responsibility to follow these instructions. If they are not followed, the carrier will
not honor any claims for damage.
This shipment was carefully inspected, verified and properly packaged at our company and delivered to the
carrier in good condition.
When it was picked up by the carrier at ENMET, it legally became your company’s property.
If your shipment arrives damaged:
•Keep the items, packing material, and carton “As Is.” Within 5 days of receipt, notify the carrier’s local
office and request immediate inspection of the carton and the contents.
•After the inspection and after you have received written acknowledgment of the damage from the carrier,
contact ENMET Customer Service for return authorization and further instructions. Have your Purchase
Order and Sales Order numbers available.
ENMET either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability
coverage, usually $100.00. Repair or replacement charges above that value are your company’s responsibility.
The shipping company may offer optional insurance coverage. ENMET only insures shipments with the
shipping company when asked to do so in writing by our customer. If you need your shipments insured, please
forward a written request to ENMET Customer Service.
Regarding Shortages
If there are any shortages or questions regarding this shipment, please notify ENMET Customer Service within 5 days
of receipt at the following address:
ENMET Corporation
680 Fairfield Court
Ann Arbor, MI 48108
734-761-1270 734-761-3220 Fax
1.2 Check Order
Check, the contents of the shipment against the purchase order. Verify that the EX-5130-IS is received as ordered.
[Each EX-5130-IS is labeled with its target gas.] If there are accessories on the order, ascertain that they are present.
Check the contents of calibration kits. Notify ENMET customer service personnel of any discrepancy immediately.
1.3 Serial Numbers
Each EX-5130-IS is serialized. These numbers are on tags on the equipment and are on record in an ENMET
database.
1
Page 4
EX-5130-IS ENMET Corporation
POT 3
POT 4
See Note
POT 2
Note:
3 Alarm
Power /Fault
M
S
Magnetic switch
2.0 Features of the EX-5130-IS
See Figure 1 for location of features:
Feature Description
Display
Gain Potentiometer
(POT)
Visual Alarms
Magnetic Switches
Sensor
LCD: Indicates the level of gas detected by sensor
POT 1: Display contrast adjustment
POT 2:
POT 3: Not Used, Do not adjust
POT 4:
LED indicators:
Power / Fault Indicator LED, Green / Red
Alarm (3) Indicator Red LED, user adjustable
M
ENU
: Advances the instrument display through menus (Zero, Span, Exit)
S
ELECT
: Selects the Zero, Span, Exit menu or sets proper calibration values for Zero or Span
For sensing gas at
PPM
or LEL levels, see Table 3 for sensor types
Magnetic switches control the instrument maintenance functions. The switch locations are indicated by M
S
ELECT
. A magnetic field pulse is applied by momentarily putting the end of the magnet in proximity to the switch
ENU
and then removing it. Referred to as tap. Since the magnetic field penetrates the window, the enclosure cover is not
removed in order to perform calibration.
Three alarm points are preprogrammed into the EX-5130-IS sensor/transmitters. At each alarm point, an LED on the
front panel is activated. These internal alarm settings are independent of the 4-20mA output alarm values that can be
set at a controller.
POT 1
Indicator
Display
ENU
Menu
Select
See Note
See Note
Magnetic switch
Internal PCB View
Sensor
POT 2, POT 3and POT 4
ELECT
are not used, Do Not Adjust
and
Indicators
External View
Figure 1: EX-5130-IS Features
2
Page 5
EX-5130-IS ENMET Corporation
Mounting Holes
Sensor/Transmitter
Sensor/Transmitter
3.0 Installation of the EX-5130-IS
C
AUTION
C
AUTION
3.1 Mounting the EX-5130-IS Enclosure
C
AUTION
:
Area must be declassified during installation.
The ENMETEX-5130-IS gas sensor/transmitter (S/T) is a 3-wire, 24 VDC, 4-20 mA S/T for the detection of toxic
gas. The S/T is meant to be used in conjunction with an appropriate power supply and controller. The ENMET EX-5130-IS sensor/transmitter has been designed to be used in a Class I, Div. 1, Groups B, C, D, classified areas.
Appropriate wiring, conduit and fittings are required for proper installation in a explosion proof rated environment.
: Since the sensor/transmitter detects gas only at the sensor location, pay attention to the possible sources of
gas, the density of the gas, locations where the gas may be confined and locations where the gas may damage
or injure property or personnel, when choosing locations of sensor/transmitters.
Also, take into consideration environmental factors when deciding on S/T location. Avoid locations where the
S/T may be damaged by liquid immersion, excessive heat or other know hazards. Also, take precautions to
insure condensation inside of the conduit does not enter the S/T.
Mount the enclosure, using the two mounting holes provided see Figure 2. Pay particular attention to the source and
density of the gas being detected when choosing the location. Mount the S/T near the ceiling for lighter than air gases
/vapors and near the floor for heavier then air gas/vapors. Contact ENMET if you have questions regarding your
application.
: Before connecting S/T to controller remove the power source to controller. Failure to do so may cause
damage to sensitive components.
Optional
Conduit fitting
Typically: ½″ NPT female
Sensor
½″ NPT female
2 places, 0.313″
Me
Sel
Enclosure Top View
Enclosure Side View
Sensor
Figure 2: EX-5130-IS Mounting Dimensions
3
Page 6
EX-5130-IS ENMET Corporation
Sensor/Transmitter Enclosure
Display Overlay Screws
(2 places)
Display Overlay
Display Overlay
Intrinsic Safety
Printed Circuit
Board (PCB)
Magnetic Switches
Optional Relay Output Board
Relay Output Board
Bottom View
N
:
Auxiliary alarms should be powered from an
1
J4
J8
Circuit Board
To J4 Wires to
3.2 Wiring the EX-5130-IS to a Control Unit
C
AUTION
*Contact ENMET for Modbus
:
Area must be declassified during installation.
Run conduit and 16
AWG
(1.5
2
MM
) wires to the enclosure from the power supply and controller. If the EX-5130-IS is
installed in a hazardous location as defined by the National Electrical Code, then ALL wiring must be in accordance
with the National code and any local governing codes.
Open the enclosure, and remove the 2 screws that retain the display overlay to the circuit board.
Use caution when removing the over lay. Do not damage the magnetic switches.
Remove the two overlay standoffs and remove the circuit board, exposing the terminal strips on the bottom of the
circuit board. Do not disconnect the circuit board wiring.
Connect the wires from the controller (power supply) to the supplied J4 plug then attach to J4 terminal.
Connect the wires from the sensor to the supplied J8 plug then attach to the J8 terminal.
See Figure 3 for locations
J4 P
LUG – TERMINAL TO CONTROLLER WIRING
Position Function
1 +
24 V
DC
power
2 GND
3 4 - 20 mA out
Bottom View
4* RS-485 D+
5* RS-485 D–
1 2 3
5
4
3
2
It is recommended that the auxiliary
alarm be powered separately.
Use 14 – 20 AWG (2.5–0.5 MM2) wire.
When on power the relays are energized.
Relays are rated at 0.5 Amp continuous.
Address information
Controller
Pl u g J4
OTE
Standoffs (2 places)
Magnetic Switches
(2 places)
independent power source separate form the
instrument power to avoid alarm failure due to
controller malfunction.
All wiring must be in accordance with the
National Code and any local governing codes.
When wiring is complete, reassemble the EX-5130-IS. Use caution when installing the overlay so as not to damage
the magnetic switches. Put the cover back on the S/T Do Not apply power to the S/T without the cover in place.
Cutaway View
4
(2 places)
Page 7
EX-5130-IS ENMET Corporation
EX-5130
-
IS
W
C, F, P
R
EX
-
5130
77
-
1256
0ppm
0ppmW
4.0 Operation of the EX-5130-IS
It is best to have the EX-5130-IS transmitters powered up and operational for 24 hours before applying calibration or
test gas to them.
When the EX-5130-IS transmitter is first powered up, it goes through a series of momentary screens, which identify
the instrument model number, serial number and software revision. After all of the momentary screens have been
displayed, the instrument arrives at the Main Gas Display showing the gas concentration and unit of measurement.
Depending on transmitter configuration and calibration condition, the furthest right character in the display may flash a
letter indicating the instrument status. See the Section 4.1.2 below.
4.1 Start up
Typical Start Up
When power is supplied to the EX-5130-IS, the S/T will display the following sequence of information:
N
OTE
:
Software revision may cause variations of display output.
Example of Display Function
The instrument: Serial Number
The instrument: Model
Note Toxic is not displayed
The instrument: Software Revision
S/W X.X
IF the right most character is a flashing
For Toxic Gas
IF the right most character is a flashing
The instrument is in Warm-up mode
This should last about 1 minute
The Signal Output is held at 4mA during warm-up
The instrument: Normal Display Mode
Measurement of target Gas
or
See Section 4.1.2
5
Page 8
EX-5130-IS ENMET Corporation
4.2 Normal Display Mode
When the EX-5130-IS is installed as described in section 3, and in clean air, the POWER green LED is on, the
display is lit and the information on the display is measurement of the target gas detected by the EX-5130-IS. The red
alarm and fault LEDs are not lit.
To advance through displays of operational information tap the magnet over the M
N
OTE
:
Software revision may cause variations of display output.
See sequence of operational information below:
Display Measurement of the target gas
Tap the magnet over the M
ENU
button
0PPM
MENU
Display indicates Alarm 1 Set point
Tap the magnet over the M
ENU
button
A1: 10
MENU
Display indicates Alarm 2 Set point
Tap the magnet over the M
ENU
button
A2: 20
MENU
Display indicates Alarm 3 Set point
Tap the magnet over the M
ENU
button
A3: 50
MENU
Display indicates mA Span range
(Full Scale)
Tap the magnet over the M
ENU
button
mA: 100
MENU
Display returns to gas measurement
SELECT
SELECT
SELECT
SELECT
SELECT
ENU
button.
No Function for the
S
ELECT
button
in this mode
Operational Display Flow Chart
4.2.1 Alarm Conditions EX-5130-IS
There are three alarm set points available. These alarm set points can be changed within limits; see the maintenance
section of this manual for the procedure.
If the gas concentration increases above that of the alarm set point, the associated red LED is lit.
6
Page 9
EX-5130-IS ENMET Corporation
5ppm
Exit
5.0 Maintenance of the EX-5130-IS
C
AUTION
C
AUTION
5.1 Maintenance Menu
Example of Display Function
Hold the magnet over M
The Power/Fault LED will flash Green – Red to indicate the EX-5130-IS is in Maintenance Mode
Tap the magnet over the M
:
Do not open the
EX-5130-IS S/T
in a classified area.
: Do Not Attempt A Span Procedure Without Calibration Gas Applied to The Sensor; if this is done, the S/T is
forced into a calibration fault mode.
Magnetic switches control the M
the display panel, see Figure 3. The M
changes to numbers such as alarm points, calibrations gas, etc. The S
ENU
and S
ELECT
functions. The M
ENU
switch is used to display the various menu options and make incremental
ENU
ELECT
and S
ELECT
switch locations are indicated on
switch is used to select that option, set
zero or span digit. Most maintenance functions are controlled by simple taps of the supplied magnet on the transmitter
glass, below the M
To enter the maintenance menu hold the magnet over the M
ENU
and S
ELECT
boxes on the front panel.
ENU
switch for 2 to 4 seconds
Table 1 indicates the maintenance menu sequence see Figure 5 for a detailed maintenance menu flow chart.
Table 1: EX-5130-IS maintenance Menus Sequence
Normal Display Mode
Measurement of target gas
ENU
switch for 2 – 5 seconds to enter the Maintenance Menu
To exit the maintenance Menu and return to the Normal
Display Mode:
ENU
switch to advance to the Zero procedure
Zero
If intended function Tap the magnet over S
For adjusting Zero:
If intended function Tap the magnet over S
ELECT
ELECT
switch
switch
Tap the magnet over the M
ENU
switch to advance to the Span procedure
Span
For adjusting the Span:
If intended function Tap the magnet over S
Tap the magnet over the M
Alarm1
ENU
switch to advance to each Alarm set point procedures
Alarm2
For adjusting the Alarm 1, 2 and 3 set points:
If Intended function Tap the magnet over S
Alarm3
Tap the magnet over the M
Taping the M
ENU
switch without taping the S
You must Tap the S
N
OTE
: If the S/T fails to respond, the magnet may have become weak and may need to be replaced.
ENU
switch to advance the mA Span set point procedure
mA Span
ELECT
switch in order to change the desired operation.
For adjusting the mA Span set point:
If intended function Tap the magnet over S
ELECT
switch will allow you to cycle through the menu options.
ELECT
ELECT
ELECT
switch
switch
switch
7
Page 10
EX-5130-IS ENMET Corporation
Output, for Gas
Input, for Calibration
5.2 Calibration of the EX-5130-IS
Calibration is the process of setting the instrument up to read accurately when exposed to a target gas. The Zero
function sets the clean air reference point and the Span function sets the sensitivity of the instrument.
Initial Calibration: Wait 24 hours after initially supplying power to the EX-5130-IS sensor/transmitter (S/T) before
initial calibration. The S/T has been pre-calibrated at the factory, and initial field calibration should result in only fine
tuning to circuit, as well as a way to check that installation is successful. It is not necessary to open the enclosure to
make adjustment. The calibration functions are operated with magnets from outside the enclosure through the
and
SELECT
switches. Do Not open the S/T unless the area is de-classified.
MENU
Calibration Zero and Span functions are two separate procedures. They operate independently of each other. It is
recommended that the Zero procedure be done prior to the Span procedure. ENMET Corporation recommends at least
quarterly calibration of the EX-5130-IS transmitters.
Calibration equipment is available from ENMET Corporation to calibrate the EX-5130-IS sensor/transmitters. A
calibration adapter will have a fitting for the gas cylinder on one side, and a cover to go over the sensor housing on the
other.
Generally, a cylinder of 20.9% Oxygen is used to provide a fresh air reference or Zero point for the calibration.
Another cylinder is used to provide the Span reference point for calibration. Depending on the instrument calibration,
the Span gas may be the same gas that the instrument is calibrated to display, or it may be another gas, which ENMET
has found to have a similar response.
: These internal sensor/transmitter alarms are independent of the 4-20mA Controller alarm point settings.
8
Page 11
EX-5130-IS ENMET Corporation
5.2.1 Zero Adjust
A ZERO function should be performed only when the EX-5130-IS sensor/transmitter is exposed to fresh air. If the air
at the sensor is in question, use a cylinder of 20.9% oxygen to provide a clean air reference. Attach the cylinder to the
calibration adapter.
Enter the maintenance menu by placing the magnet over M
ENU
switch for 2 to 4 seconds. See Figure 5,EX-5130-IS
Maintenance Menu flow chart.
The second menu available is the Zero.
Tap the S
ELECT
switch to perform a Zero.
If the Zero is successful, Cal OK appears on the display and in 1 – 2 seconds, display will change to Span.
If you wish to Span the sensor Tap the S
If you wish to Exit the maintenance menu, Tap M
ELECT
switch you are now ready to apply gas. Proceed to gas span step 2
ENU
switch until Exit is displayed, then tap S
ELECT
switch to
return to the instrument Normal Gas Display
If the Zero is Not successful, sensor is outside of safe parameters to be zeroed, the display will read Bad Zero.
Repeat Section 5.2.1 Zero Adjust making sure to use a cylinder of 20.9% Oxygen.
5.2.2 Gas Span
It is recommended that the Zero Function be performed first.
Enter the maintenance menu. See Figure 5,EX-5130-IS Maintenance Menu flow chart.
1. Tap the M
2. Tap the S
concentration and a signal level.
N
OTE
:
The M
The S
3. Attach the associated calibration gas cylinder to the regulator and calibration cover. See to Figure3.
4. Open the valve to apply the calibration gas to the sensor.
5. Watch for the signal level to stabilize. Refer to Table 4 for typical response times.
6. Once the signal level has stabilized, the EX-5130-IS will automatically lock in the calibration data and:
If the Span is successful, Cal OK appears on the display momentarily, then advances to Alarm 1. Remove
calibration gas. To exit maintenance menu tap the M
ENU
switch once to show Span on the display.
ELECT
switch to perform a Span procedure. The display will alternate between the calibration gas
You can change the Calibration Gas Level. H
ENU
switch changes digit indicated by underscore cursor
ELECT
switch locks underscored digit and moves to next digit
OLD
the magnet over the S
ENU
switch until Exit appears, then tap the S
ELECT
switch for 2 – 4 seconds
ELECT
switch.
If the sensor is outside of acceptable parameters, Bad Span is displayed momentarily, then returns to Span.
Remove calibration gas. Tap the M
ENU
switch until Exit appears, then tap the S
ELECT
switch. Check span gas
and repeat calibration in 30 – 60 minutes.
If the sensor did not respond to gas, Same mV is displayed momentarily, then returns to Span.
Remove calibration gas, tap the M
ENU
switch until Exit appears, then tap the S
ELECT
switch and try
calibration again in 30-60 minutes.
If the sensor will not calibrate See Section 5.4.
N
OTE
: Some software revisions require the S
ELECT
switch be tapped to accept the signal.
7. Calibration is complete.
5.2.3 Exit Maintenance Menu
Exit maintenance, by tapping on the M
to the instrument Normal Gas Display.
ENU
switch until Exit appears on the display. Tap the S
9
ELECT
switch to return
Page 12
EX-5130-IS ENMET Corporation
H
M
S
OR
Zero
Cal OK
If
the Zero signal is within Preset Specs the EX
-
5130-IS will
Bad ZERO
PV:
0
Apply
PV:
0
20
Cal OK
Same mV
Span
OR
OR
100
To
Select
Menu
Alarm3
50
SetTD
sec
0
Alarm1
SetTD
sec
20
Normal Display Mode
5ppm
M
ENU
Exit
M
ENU
Tap the magnet over the M
OLD
the magnet over the
ENU
to enter the Maintenance Menus
S
ELECT
Tap the
Display Mode.
ELECT
switch to return to the Normal
See Section 5.2.3
ENU
switch to cycle through Maintenance Menus
switch for 2 – 4 seconds
M
ENU
M
ENU
M
ENU
Alarm2
M
ENU
S
ELECT
S
ELECT
Bad Sens
S
ELECT
S
ELECT
S
ELECT
S
ELECT
Λ 10
Λ
0
Λ
S
ELECT
Tap the S
display Cal OK, See Section 5.2.1
If the Zero signal is not within Preset Specs the EX-5130-IS will
display Bad ZERO
N
OTE
: Some software revisions require the S
tapped to accept the signal.
N
OTE
: You can change the Calibration Gas Level.
S
ELECT
The M
ENU
The S
ELECT
ELECT
switch to initiate Zero adjustment
ELECT
H
OLD
the magnet over the S
switch changes digit indicated by underscore cursor
switch locks underscored digit and moves to next digit
ELECT
switch be
switch
See Section 5.2.2
Calibration Gas until signal value becomes
stable (about 1 to 4 minutes) See Figure 4If cal is good display will indicate OK or Same
If cal is not within preset “range” display will
indicate Bad Sens
To change Alarm set points:
Tap Menu switch until Alarm to be changed is displayed
Tap Select switch to display the set point
The M
ENU
cursor
The S
moves to next digit
If change is not within range display returns to first digit
If change is within range display moves to Set Time Delay
Use M
delay. Between 0 and 5 seconds is allowed
If change is within range display moves to next menu
Λ - Indicates increasing alarm
V - Indicates decreasing alarm
switch: changes digit indicated by underscore
ELECT
switch: locks in the underscored digit and
ENU
and S
ELECT
switches as above to change time
M
ENU
mA Span
M
ENU
return to Normal Gas Display:
Tap M
ENU
Then tap S
switch until EXIT is displayed
ELECT
SetTDsec
0
S
switch
ELECT
To change mA Span set point:
Tap the M
Tap the S
The M
The S
ENU
switch until mA Span is displayed
ELECT
switch to display the set point
ENU
switch changes digit indicated by underscore cursor
ELECT
switch locks underscored digit and moves to next digit
Figure 5: EX-5130-IS Maintenance Menu Flow chart
10
Page 13
EX-5130-IS ENMET Corporation
Set Screw
Sensor /P C Board
5.3 Sensor Replacement
C
AUTION
Sensors should be replaced when they can no longer be calibrated. Replacement sensor part numbers are listed in
Section 6.0 of this manual. If you do not know the proper part number for your sensor, be sure to have the EX-5130-IS serial number available when contacting your Distributor or ENMET Corporation Technical Support.
To replace a sensor, it is not necessary to open the transmitter housing.
Remove the set screw from sensor housing base.
Unscrew the sensor housing cover and remove Sensor/PC Board Assembly.
Plug in new Sensor/PC Board Assembly.
Reassemble the sensor housing.
After the new sensor has been installed, it is suggested to allow the sensor to stabilize for 24 hours.
A Factory calibration must be performed.
After entering the Maintenance menu, advance to the Zero menu. Then while viewing the Zero menu, hold the magnet
over the M
After 2-4 seconds, an F will appear on the far right hand side of the display. The F indicates that the instrument is in
Factory mode.
Perform the calibration Zero and Span procedures as outlined in Section 5.2. Be sure that the F is present when
selecting the Zero and Span functions.
The Factory calibration sets a calibration window for future standard instrument calibrations.
Only perform a factory calibration when installing a new sensor!!
: Area must be declassified during sensor replacement.
N
OTE
: Sensor and PC Board are attached and should not be separated from each other.
ENU
switch for 2-4 seconds.
(0.050 inch, 1.27mm Hex Key)
Sensor Housing Base
Assembly
Sensor Housing Cover
Figure 6: Sensor Replacement
5.4 Intrinsic Barrier Replacement
C
AUTION
: Area must be declassified during sensor replacement.
Intrinsic Safety Barrier
Figure 7: Intrinsic Barrier Replacement
11
Page 14
EX-5130-IS ENMET Corporation
6.0 Replacement Part Numbers
ENMET replacement part numbers:
Description Part Number
For EX-5130-IS p/n
Sensor: Oxygen
Carbon Monoxide
Hydrogen Sulfide
Chlorine
Bromine
Consult ENMET Distributor or ENMET Corp for additional sensors
Regulator Assembly, for 34/58/103l cylinders
Regulator Assembly, for 17/34l cylinders
Calibration Gas, Consult ENMET Distributor or ENMET Corp.
67027-xxxx
02506-002
02506-004
7.0 WARRANTY
ENMET warrants new instruments to be free from defects in workmanship and material under normal use for a period
of one year from date of shipment from ENMET. The warranty covers both parts and labor excluding instrument
calibration and expendable parts such as calibration gas, filters, batteries, etc... Equipment believed to be defective
should be returned to ENMET within the warranty period (transportation prepaid) for inspection. If the evaluation by
ENMET confirms that the product is defective, it will be repaired or replaced at no charge, within the stated
limitations, and returned prepaid to any location in the United States by the most economical means, e.g. Surface
UPS/FedEx Ground. If an expedient means of transportation is requested during the warranty period, the customer is
responsible for the difference between the most economical means and the expedient mode. ENMET shall not be
liable for any loss or damage caused by the improper use of the product. The purchaser indemnifies and saves
harmless the company with respect to any loss or damages that may arise through the use by the purchaser or others of
this equipment.
This warranty is expressly given in lieu of all other warranties, either expressed or implied, including that of
merchantability, and all other obligations or liabilities of ENMET which may arise in connection with this equipment.
ENMET neither assumes nor authorizes any representative or other person to assume for it any obligation or liability
other than that which is set forth herein.
NOTE: When returning an instrument to the factory for service:
Be sure to include paperwork.
A purchase order, return address and telephone number will assist in the expedient repair and return of your unit.
Include any specific instructions.
For warranty service, include date of purchase
If you require an estimate, please contact ENMET Corporation.
There are Return for Repair Instructions and Form on the last pages of this manual. This Form can be copied or used
as needed.
Manual Part Number
80003-093
March 2009
MCN-13-011, 12/20/13
12
Page 15
PO Box 979
680 Fairfield Court
Ann Arbor, Michigan 48106-0979
734.761.1270 Fax 734.761.3220
Returning an Instrument for Repair
ENMET instruments may be returned to the factory or any one of our Field Service Centers for regular
repair service or calibration. The ENMET Repair Department and Field Service Centers also perform
warranty service work.
When returning an instrument to the factory or service center for service, paperwork must be included
which contains the following information:
A purchase order number or reference number.
A contact name with return address, telephone and fax numbers
Specific instructions regarding desired service or description
of the problems being encountered.
Date of original purchase and copy of packing slip or invoice
for warranty consideration.
If a price estimate is required, please note it accordingly and be
sure to include a fax number.
Providing the above information assists in the expedient repair and return of your unit.
Failure to provide this information can result in processing delays.
ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the
closest tenth of an hour. All instruments sent to ENMET are subject to a minimum evaluation fee, even
if returned unrepaired. Unclaimed instruments that ENMET has received without appropriate
paperwork or attempts to advise repair costs that have been unanswered, after a period of 60 days,
may be disposed of or returned unrepaired COD with the evaluation fee.
Service centers may have different rates or terms. Be sure to contact them for this information.
Repaired instruments are returned by UPS/FedEx Ground and are not insured unless otherwise specified.
If expedited shipping methods or insurance is required, it must be stated in your paperwork.
Note: Warranty of customer installed components.
If a component is purchased and installed in the field, and fails within the warranty term, it can
be returned to ENMET and will be replaced, free of charge, per ENMET’s returned goods
procedure.
If the entire instrument is returned to ENMET Corporation with the defective item installed, the
item will be replaced at no cost, but the instrument will be subject to labor charges at half of the
standard rate.
Page 16
Mailing Address:
Shipping Address:
Repair Return Form
ENMET Corporation
PO Box 979
Ann Arbor, Michigan 48106
Contact Name: __________________________ Your Phone: _______________________
Your PO/Reference Number: _______________ Your FAX: _______________________
ENMET Corporation
Attn: Repair Department
680 Fairfield Court
Ann Arbor, Michigan 48108