ENMET EX-5120 User Manual

Page 1
ENMET Corporation
PO Box 979 Ann Arbor, MI 48106-0979
www.enmet.com
EX-5120
I
NFRARED
0 – 100% LEL
Sensor/Transmitter
Manual
Manual Part Number 80003-092
MCN-13-011, 12/20/13
Page 2
Table of Contents
1.0 I
NTRODUCTION
1.1 Unpack ............................................................................................................................................................................... 1
1.2 Check Order ....................................................................................................................................................................... 1
1.3 Serial Numbers................................................................................................................................................................... 1
2.0 F
EATURES OF THE
3.0 I
NSTALLATION OF THE
3.1 Mounting the EX-5120 Enclosure ..................................................................................................................................... 3
3.2 Wiring the EX-5120 to a Control Unit ............................................................................................................................... 4
4.0 O
PERATION OF THE
4.1 Start up ............................................................................................................................................................................... 6
4.1.1 Typical Start Up .............................................................................................................................................................................. 6
4.2 Normal Display Mode ........................................................................................................................................................ 7
4.2.1 Alarm Conditions EX-5120............................................................................................................................................................ 7
5.0 M
AINTENANCE OF THE
5.1 Maintenance Menu ............................................................................................................................................................. 8
5.2 Calibration of the EX-5120 ................................................................................................................................................ 9
5.2.1 Zero Adjust...................................................................................................................................................................................... 9
5.2.2 Gas Span ....................................................................................................................................................................................... 10
5.2.3 Exit Maintenance Menu ................................................................................................................................................................ 10
5.3 Sensor Replacement ......................................................................................................................................................... 12
6.0 R
EPLACEMENT PART NUMBERS
7.0 WARRANTY....................................................................................................................................................... 14
....................................................................................................................................................... 1
EX-5120 ................................................................................................................................... 2
EX-5120 .............................................................................................................................. 3
EX-5120.................................................................................................................................. 6
EX-5120 ............................................................................................................................. 8
............................................................................................................................ 13
List of Figures and Tables
Figure 1: EX-5120 Features ....................................................................................................................................... 2
Figure 2: EX-5120 Mounting Dimensions .................................................................................................................. 3
Figure 3: Terminal Positions EX-5120 Sensor/Transmitter ....................................................................................... 5
Table 1: EX-5120 maintenance Menus Sequence .................................................................................................... 8
Figure 4: Calibration Adapter EX-5120 Sensor/Transmitter ...................................................................................... 9
Figure 5: EX-5120 Maintenance Menu Flow chart................................................................................................... 11
Figure 6: Sensor Replacement ................................................................................................................................ 12
Reference information:
N
OTE
: [important information about use of instrument – if not followed may have to redo some steps.]
C
W
ARNING
AUTION
: [affects equipment – if not followed may cause damage to instrument, sensor etc…]
:
[affects personnel safety – if not followed may cause bodily injury or death.]
Page 3
EX-5120 ENMET Corporation
1.0 Introduction
The ENMET EX-5120 infrared sensor/transmitter (S/T) is a 3-wire, 24Vdc, 4-20mA S/T for the detection of combustible gas. The EX-5120 is meant to be used in conjunction with an appropriate power supply and controller. The ENMET EX-5120 sensor/transmitter has been designed for use in a Class I, Div. 1, Groups B, C, D, classified areas.
N
OTE
: All specifications stated in this manual may change without notice.
1.1 Unpack
Unpack the EX-5120 and examine it for shipping damage. If such damage is observed, notify both ENMET customer service personnel and the commercial carrier involved immediately.
Regarding Damaged Shipments
N
OTE
: It is your responsibility to follow these instructions. If they are not followed, the carrier will not honor
any claims for damage.
This shipment was carefully inspected, verified and properly packaged at our company and delivered to the carrier in good condition.
When it was picked up by the carrier at ENMET, it legally became your company’s property.
If your shipment arrives damaged:
Keep the items, packing material, and carton “As Is.” Within 5 days of receipt, notify the carrier’s local office and
request immediate inspection of the carton and the contents.
After the inspection and after you have received written acknowledgment of the damage from the carrier, contact
ENMET Customer Service for return authorization and further instructions. Have your Purchase Order and Sales Order numbers available.
ENMET either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability coverage, usually $100.00. Repair or replacement charges above that value are your company’s responsibility.
The shipping company may offer optional insurance coverage. ENMET only insures shipments with the shipping company when asked to do so in writing by our customer. If you need your shipments insured, please forward a written request to ENMET Customer Service.
Regarding Shortages
If there are any shortages or questions regarding this shipment, please notify ENMET Customer Service within 5 days of receipt at the following address:
ENMET Corporation
680 Fairfield Court
Ann Arbor, MI 48108
734-761-1270 734-761-3220 Fax
1.2 Check Order
Check, the contents of the shipment against the purchase order. Verify that the EX-5120 is received as ordered. [Each EX­5120 is labeled with its target gas.] If there are accessories on the order, ascertain that they are present. Check the contents of
calibration kits. Notify ENMET customer service personnel of any discrepancy immediately.
1.3 Serial Numbers
Each EX-5120 is serialized. These numbers are on tags on the equipment and are on record in an ENMET database.
1
Page 4
EX-5120 ENMET Corporation
3 Alarm
Power /Fault
M
S
Magnetic switch
POT 3
POT 4
See Note
POT 2
POT 1
Note:
2.0 Features of the EX-5120
See Figure 1 for location of features:
Feature Description Display
Gain Potentiometer (POT)
Visual Alarms
Magnetic Switches
Infrared Sensor
LCD: Indicates the level of gas detected by sensor POT 1: Display contrast adjustment
POT 2: POT 3: Not Used, Do not adjust POT 4:
LED indicators:
Power / Fault Indicator LED, Green / Red Alarm (3) Indicator Red LED, user adjustable
M
ENU
: Advances the instrument display through menus (Zero, Span, Exit)
S
ELECT
: Selects the Zero, Span, Exit menu or sets proper calibration values for Zero or Span
For sensing gas at LEL levels, see Section 6.0 for sensor types
Magnetic switches control the instrument maintenance functions. The switch locations are indicated by M
ENU
and S
ELECT
magnetic field pulse is applied by momentarily putting the end of the magnet in proximity to the switch and then removing it. Referred to as tap. Since the magnetic field penetrates the window, the enclosure cover is not removed in order to perform calibration.
Three alarm points are preprogrammed into the EX-5120 sensor/transmitters. At each alarm point, an LED on the front panel is activated. These internal alarm settings are independent of the 4-20mA output alarm values that can be set at a controller.
Display Contrast
Indicators
See Note
See Note
Magnetic switch
Indicator
Display
ENU
Menu
Select
Internal PCB View
Sensor
POT 2, POT 3and POT 4
ELECT
are not used, Do Not Adjust
. A
External View
Figure 1: EX-5120 Features
2
Page 5
EX-5120 ENMET Corporation
Mounting Holes
Sensor/Transmitter
Sensor/Transmitter
3.0 Installation of the EX-5120
C
AUTION
C
AUTION
3.1 Mounting the EX-5120 Enclosure
C
AUTION
:
Area must be declassified during installation.
The ENMET EX-5120 gas sensor/transmitter (S/T) is a 3-wire, 24 Vdc, 4-20 mA S/T for the detection of combustible gas. The S/T is meant to be used in conjunction with an appropriate power supply and controller. The ENMET EX-5120 sensor/transmitter is in an enclosure rated for use in a Class I, Div. 1, Groups B, C, D, classified area. Appropriate wiring, conduit and fittings are required for proper installation in a explosion proof rated environment.
: Since the sensor/transmitter detects gas only at the sensor location, pay attention to the possible sources of gas, the
density of the gas, locations where the gas may be confined and locations where the gas may damage or injure property or personnel, when choosing locations of sensor/transmitters. Also, take into consideration environmental factors when deciding on S/T location. Avoid locations where the S/T may be damaged by liquid immersion, excessive heat or other known hazards. Also, take precautions to insure condensation inside of the conduit does not enter the S/T.
Mount the enclosure, using the two mounting holes provided see Figure 2. Pay particular attention to the source and density of the gas being detected when choosing the location. Contact ENMET if you have questions regarding your application.
: Before connecting S/T to controller remove the power source to controller. Failure to do so may cause damage to
sensitive components.
Optional Conduit fitting Typically: ½ NPT female
Sensor
½″ NPT female
2 places, 0.313
Me
Sel
Enclosure Top View
Enclosure Side View
Sensor
Figure 2: EX-5120 Mounting Dimensions
3
Page 6
EX-5120 ENMET Corporation
J4
J8
Wires to
s
Wires to
Relay Output Board
Optional Relay Output Board
3.2 Wiring the EX-5120 to a Control Unit
C
AUTION
*Contact ENMET for Modbus
:
Area must be declassified during installation.
Run conduit and 16
AWG
(1.5
2
MM
) wires to the enclosure from the power supply and controller. If the EX-5120 is installed in
a hazardous location as defined by the National Electrical Code, then ALL wiring must be in accordance with the National code and any local governing codes.
Open the enclosure, and remove the 2 screws that retain the display overlay to the circuit board. Use caution when removing the over lay. Do not damage the magnetic switches. Remove the two overlay standoffs and remove the circuit board, exposing the terminal strips on the bottom of the circuit board.
Do not disconnect the circuit board wiring. Connect the wires from the controller (power supply) to the supplied J4 plug then attach to J4 terminal. Connect the wires from the sensor to the supplied J8 plug then attach to the J8 terminal. See Figure 3 for locations
J4 P
LUG – TERMINAL TO CONTROLLER WIRING
Position Function
1 +
24 V
DC
power
2 – GND
3 s
4 - 20 mA out
To J4
s
4* RS-485 D+ 5* RS-485 D–
Controller
Plu g J4
Address information
5 4 3 2 1
It is recommended that the auxiliary
alarm be powered separately.
Use 14 – 20 When on power the relays are energized. Relays are rated at 0.5 Amp continuous.
AWG
(2.5 – 0.5
MM
2
) wire.
1 2 3
Circuit Board Bottom View
J8 P
LUG – TERMINAL TO SENSOR WIRING
Position Function Sensor
1 + Power Red 2 s 3 – GND Black
Sensor
Signal White
Plu g J8
To J8
Bottom View
N
OTE
: Auxiliary alarms should be powered from an independent power source separate form the instrument power to avoid alarm failure due to controller malfunction. All wiring must be in accordance with the National Code and any local governing codes.
4
Page 7
EX-5120 ENMET Corporation
Sensor/Transmitter
Printed Circuit Board
Display Overlay Screws
(2 places)
Display Overlay Standoffs
Magnetic
Switches
Magnetic Switches
Printed Circuit Board (PCB)
Optional
Display Overlay
(2 places)
Display
Display Overlay
(2 places)
J4 and J8 Terminals are located on the
bottom side of PCB
Relay Output Circuit Board
Enclosure Cutaway View
Display Overlay
(2 places)
Figure 3: Terminal Positions EX-5120 Sensor/Transmitter
When wiring is complete, reassemble the switches. Put the cover back on the S/T
(PCB)
EX-5120
Do Not
. Use caution when installing the overlay so as not to damage the magnetic
apply power to the S/T without the cover in place.
5
Page 8
EX-5120
EX-5120
W
EX
-
5120
78
-
1256
0%
ENMET Corporation
4.0 Operation of the EX-5120
It is best to have the or test gas to them.
When the instrument model number, serial number and software revision. After all of the momentary screens have been displayed, the instrument arrives at the Main Gas Display showing the gas concentration and unit of measurement.
Depending on transmitter configuration and calibration condition, the furthest right character in the display may flash a letter indicating the instrument status.
EX-5120
4.1 Start up
4.1.1 Typical Start Up
When power is supplied to the
N
OTE
:
Software revision may cause variations of display output.
Example of Display Function
The instrument: Serial Number
The instrument: Software Revision
S/W X.X
EX-5120
transmitter is first powered up, it goes through a series of momentary screens, which identify the
transmitters powered up, operational and stable for several minutes before applying calibration
Consult ENMET Distributor or ENMET Corp
EX-5120,
the S/T will display the following sequence of information:
The instrument: Model
.
IF
the right most character is a flashing
0%W
The instrument is in Warm-up mode
This should last about 1 minute
The Signal Output is held at 4mA during warm-up
The instrument: Normal Display Mode Measurement of target Gas
6
Page 9
EX-5120
0%
A1:
10
M
ENU
A2:
20
M
ENU
A3:
50
M
ENU
mA:
100
ENMET Corporation
4.2 Normal Display Mode
When the the information on the display is measurement of the target gas detected by the not lit.
To advance through displays of operational information tap the magnet over the
N
OTE
See sequence of operational information below:
Display Measurement of the target gas Tap the magnet over the M
Display indicates Alarm 1 Set point Tap the magnet over the M
Display indicates Alarm 2 Set point Tap the magnet over the M
Display indicates Alarm 3 Set point Tap the magnet over the M
Display indicates mA Span range (Full Scale)
Tap the magnet over the M Display returns to gas measurement
EX-5120
:
Software revision may cause variations of display output.
is installed as described in section 3, and in clean air, the
ENU
button
M
ENU
ENU
button
ENU
button
ENU
button
ENU
button
M
ENU
S
S
S
S
S
ELECT
ELECT
ELECT
ELECT
ELECT
POWER
EX-5120.
M
ENU
green LED is on, the display is lit and
The red alarm and fault LEDs are
button.
No Function for the
S
ELECT
button
in this mode
Operational Display Flow Chart
4.2.1 Alarm Conditions EX-5120
There are three alarm set points available. These alarm set points can be changed within limits; see the maintenance section of this manual for the procedure.
If the gas concentration increases above that of the alarm set point, the associated red LED is lit.
7
Page 10
EX-5120
5%
Exit
ENMET Corporation
5.0 Maintenance of the EX-5120
C
AUTION
C
AUTION
5.1 Maintenance Menu
:
Do not open the
: Do Not Attempt A Span Procedure Without Calibration Gas Applied to The Sensor
EX-5120
S/T
in a classified area.
; if this is done, the S/T is forced
into a calibration fault mode.
Magnetic switches control the display panel, see
Figure 3.
numbers such as alarm points, calibrations gas, etc. The
MENU
The
and
S
ELECT
functions. The
M
ENU
switch is used to display the various menu options and make incremental changes to
S
ELECT
M
ENU
and
S
ELECT
switch locations are indicated on the
switch is used to select that option, set zero or span digit.
Most maintenance functions are controlled by simple taps of the supplied magnet on the transmitter glass, below the
S
ELECT
boxes on the front panel.
To enter the maintenance menu hold the magnet over the
Table 1
indicates the maintenance menu sequence see
M
ENU
Figure 5
switch for 2 to 4 seconds for a detailed maintenance menu flow chart.
Table 1: EX-5120 maintenance Menus Sequence
Example of Display Function
Hold the magnet over
M
ENU
switch for 2 – 5 seconds to enter the Maintenance Menu
The Power/Fault LED will flash Green – Red to indicate the
Tap the magnet over the
M
ENU
switch to advance to the Zero procedure
Zero
Normal Display Mode
Measurement of target gas
EX-5120
is in Maintenance Mode
To exit the maintenance Menu and return to the Normal Display Mode:
If intended function Tap the magnet over
For adjusting Zero: If intended function Tap the magnet over
S
ELECT
S
ELECT
switch
switch
M
ENU
and
Tap the magnet over the
Span
M
ENU
switch to advance to the Span procedure
For adjusting the Span: If intended function Tap the magnet over
Tap the magnet over the
Alarm1
M
ENU
switch to advance to each Alarm set point procedures
Alarm2
For adjusting the Alarm 1, 2 and 3 set points: If Intended function Tap the magnet over
Alarm3
Tap the magnet over the
mA Span
Taping the You must Tap the
N
OTE
M
ENU
switch without taping the
S
ELECT
:
If the S/T fails to respond, the magnet may have become weak and may need to be replaced.
M
ENU
switch to advance the mA Span set point procedure
For adjusting the mA Span set point: If intended function Tap the magnet over
S
ELECT
switch will allow you to cycle through the menu options.
switch in order to change the desired operation.
8
S
ELECT
S
ELECT
S
ELECT
switch
switch
switch
Page 11
EX-5120
Output, for Gas
I
nput, for Calibration
ENMET Corporation
5.2 Calibration of the EX-5120
Calibration is the process of setting the instrument up to read accurately when exposed to a target gas. The Zero function sets the clean air reference point and the Span function sets the sensitivity of the instrument.
Initial Calibration:
before initial calibration. The S/T has been precalibrated at the factory, and initial field calibration should result in only fine tuning to circuit, as well as a way to check that installation is successful. It is not necessary to open the enclosure to make adjustment. The calibration functions are operated with magnets from outside the enclosure through the switches. Do Not open the S/T unless the area is de-classified.
Calibration Zero and Span functions are two separate procedures. They operate independently of each other. It is recommended that the Zero procedure be done prior to the Span procedure. quarterly calibration of the
Calibration equipment is available from adapter will have a fitting for the gas cylinder on one side, and a cover to go over the sensor housing on the other.
Generally, a cylinder of 20.9% Oxygen is used to provide a fresh air reference or Zero point for the calibration. Another cylinder is used to provide the Span reference point for calibration. Depending on the instrument calibration, the Span gas may be the same gas that the instrument is calibrated to display, or it may be another gas, which response.
Wait several minutes for stabilization after supplying power to the
EX-5120
transmitters.
ENMET
ENMET
Corporation to calibrate the
Corporation recommends at least
EX-5120
EX-5120
sensor/transmitter (S/T)
MENU
and
sensor/transmitters. A calibration
ENMET
has found to have a similar
SELECT
Select
Menu
Calibration Cover
or Optional Gas Sampling
Regulator
Gas Cylinder
Figure 4: Calibration Adapter EX-5120 Sensor/Transmitter
5.2.1 Zero Adjust
A ZERO function should be performed only when the is in question, use a cylinder of 20.9% oxygen to provide a clean air reference. Attach the cylinder to the calibration adapter.
Enter the maintenance menu by placing the magnet over Maintenance Menu flow chart.
The second menu available is the Zero. Tap the
S
ELECT
switch to perform a Zero.
If the Zero is successful
, Cal OK appears on the display and in 1 – 2 seconds, display will change to Span.
If you wish to Span the sensor Tap the
S
ELECT
If you wish to Exit the maintenance menu, Tap instrument Normal Gas Display
If the Zero is Not successful,
sensor is outside of safe parameters to be zeroed, the display will read Bad Zero. Repeat
Section 5.2.1 Zero Adjust making sure to use a cylinder of 20.9% Oxygen.
EX-5120
M
sensor/transmitter is exposed to fresh air. If the air at the sensor
ENU
switch for 2 to 4 seconds. See
switch you are now ready to apply gas.
M
ENU
switch until Exit is displayed, then tap
Figure 5, EX-5120
Proceed to gas span step 2
S
ELECT
switch to return to the
9
Page 12
EX-5120
H
M
S
Zero
If
Select
5.2.2 Gas Span
It is recommended that the Zero Function be performed first.
Enter the maintenance menu. See
1.
2.
Tap the Tap the
M
ENU
switch once to show Span on the display.
S
ELECT
switch to perform a Span procedure. The display will alternate between the calibration gas concentration
Figure 5, EX-5120
and a signal level.
N
OTE
:
You can change the Calibration Gas Level. H
The M
ENU
The S
3.
Attach the associated calibration gas cylinder to the regulator and calibration cover. See to
4.
Open the valve to apply the calibration gas to the sensor.
5.
Watch for the signal level to stabilize.
6.
Once the signal level has stabilized, the
If the Span is successful, Cal OK appears on the display momentarily, then advances to Alarm 1. Remove calibration gas. To exit maintenance menu tap the
If the sensor is outside of acceptable parameters, Bad Span is displayed momentarily, then returns to Span. Remove calibration gas. Tap the
switch changes digit indicated by underscore cursor
ELECT
switch locks underscored digit and moves to next digit
EX-5120
M
ENU
M
ENU
repeat calibration.
N
OTE
:
Some software revisions require the
7.
Calibration is complete.
5.2.3 Exit Maintenance Menu
Exit maintenance, by tapping on the instrument Normal Gas Display.
Normal Display Mode
OLD
5%
M
ENU
the magnet over the
to enter the Maintenance Menus
M
ENU
switch until Exit appears on the display. Tap the
Maintenance Menu flow chart.
OLD
the magnet over the S
ELECT
will automatically lock in the calibration data and:
switch until Exit appears, then tap the
switch until Exit appears, then tap the
S
ELECT
switch be tapped to accept the signal.
ENU
switch for 2 – 4 seconds
ENMET Corporation
switch for 2 – 4 seconds
S
ELECT
S
ELECT
Figure 4
switch.
switch. Check span gas and
S
ELECT
.
switch to return to the
Exit
M
ENU
S
Tap the magnet over the M
S
M
ENU
Tap the
ELECT
Display Mode.
ENU
ELECT
PV: 0
Cal OK
OR
Bad ZERO
ELECT
switch to return to the Normal
See Section 5.2.3
switch to cycle through Maintenance Menus
S
ELECT
Tap the S
the Zero signal is within Preset Specs the EX-5120 will display
Cal OK, See Section 5.2.1
If the Zero signal is not within Preset Specs the EX-5120 will
display Bad ZERO
N
OTE
:
Some software revisions require the
ELECT
switch to initiate Zero adjustment
tapped to accept the signal.
Menu
S
ELECT
switch be
10
Page 13
EX-5120
Apply
Same mV
Span
mA Span
To
SetTD
sec
0
10
20
S
PV: 0
ELECT
50
M
ENU
S
ELECT
N
OTE
: You can change the Calibration Gas Level.
H
OLD
S
ELECT
The M
ENU
The S
ELECT
See Section 5.2.2
the magnet over the S
switch changes digit indicated by underscore cursor
switch locks underscored digit and moves to next digit
ENMET Corporation
ELECT
switch
Alarm1
M
ENU
Alarm2
M
ENU
Alarm3
S
ELECT
S
ELECT
S
ELECT
Cal OK
OR
OR
Bad Sens
Λ
SetTDsec
0
Λ
SetTDsec
0
Λ 50
Calibration Gas until signal value becomes
stable (about 1 to 4 minutes)
See
Figure 4
If cal is good display will indicate OK or Same If cal is not within preset “range” display will
indicate Bad Sens
To change Alarm set points: Tap Menu switch until Alarm to be changed is displayed Tap Select switch to display the set point
The M
ENU
switch: changes digit indicated by underscore cursor The S
ELECT
switch: locks in the underscored digit and
moves to next digit
If change is not within range display returns to first digit If change is within range display moves to Set Time Delay
Use M
ENU
and S
ELECT
switches as above to change time
delay. Between 0 and 5 seconds is allowed If change is within range display moves to next menu
Λ - Indicates increasing alarm V - Indicates decreasing alarm
M
ENU
S
ELECT
M
ENU
return to Normal Gas Display:
Tap M
ENU
switch until EXIT is displayed
Then tap S
ELECT
switch
100
To change mA Span set point: Tap the M Tap the S
The M The S
ENU
switch until mA Span is displayed
ELECT
switch to display the set point
ENU
switch changes digit indicated by underscore cursor
ELECT
switch locks underscored digit and moves to next digit
Figure 5: EX-5120 Maintenance Menu Flow chart
11
Page 14
EX-5120
Set Screw
Sensor /P C Board
ENMET Corporation
5.3 Sensor Replacement
C
AUTION
Sensors should be replaced when they can no longer be calibrated. Replacement sensor part numbers are listed in Section 6.0 of this manual. If you do not know the proper part number for your sensor, be sure to have the available when contacting your Distributor or
To replace a sensor, it is not necessary to open the transmitter housing. Remove the set screw from sensor housing base. Unscrew the sensor housing cover and remove Sensor/PC Board Assembly.
Plug in new Sensor/PC Board Assembly. Reassemble the sensor housing. After the new sensor has been installed, it is suggested to allow the sensor to stabilize for several minutes.
A Factory calibration must be performed.
After entering the Maintenance menu, advance to the Zero menu. Then while viewing the Zero menu, hold the magnet over the
M
After 2-4 seconds, an F will appear on the far right hand side of the display. The F indicates that the instrument is in Factory mode.
Perform the calibration Zero and Span procedures as outlined in Section 5.2. Be sure that the F is present when selecting the Zero and Span functions.
The Factory calibration sets a calibration window for future standard instrument calibrations. Only perform a factory calibration when installing a new sensor!!
: Area must be declassified during sensor replacement.
ENMET
N
OTE
:
Sensor and PC Board are attached and should not be separated from each other.
ENU
switch for 2-4 seconds.
Corporation Technical Support.
EX-5120
serial number
Sensor Housing Cover
(0.050 inch, 1.27mm Hex Key)
Sensor Housing Base
Assembly
Figure 6: Sensor Replacement
12
Page 15
EX-5120
6.0 Replacement Part Numbers
ENMET
replacement part numbers:
Description Part Number
Sensor, Infrared Methane, 0 – 100% LEL 03056-002 Sensor, Infrared Butane, 0 – 100% LEL 03056-012 Sensor, Infrared Propane, 0 – 100% LEL 03056-008
Consult ENMET Distributor or ENMET Corp for additional ranges Regulator Assembly, for 17l cylinders
Magnet 50030-001 Calibration/Sampling Adapter 03700-034
Calibration Gas, 17l cylinders, 50% LEL Methane Calibration Gas, 17l cylinders, 50% LEL Butane Calibration Gas, 17l cylinders, 50% LEL Propane Consult ENMET Distributor or ENMET Corp for additional gases
02506-004
03220-050 03282-050 03221-050
ENMET Corporation
13
Page 16
EX-5120
ENMET Corporation
7.0 WARRANTY
ENMET
year from date of shipment from expendable parts such as calibration gas, filters, batteries, etc... Equipment believed to be defective should be returned to
ENMET
product is defective, it will be repaired or replaced at no charge, within the stated limitations, and returned prepaid to any location in the United States by the most economical means, e.g. Surface UPS/FedEx Ground. If an expedient means of transportation is requested during the warranty period, the customer is responsible for the difference between the most economical means and the expedient mode. the product. The purchaser indemnifies and saves harmless the company with respect to any loss or damages that may arise through the use by the purchaser or others of this equipment.
This warranty is expressly given in lieu of all other warranties, either expressed or implied, including that of merchantability, and all other obligations or liabilities of nor authorizes any representative or other person to assume for it any obligation or liability other than that which is set forth herein.
NOTE: When returning an instrument to the factory for service:
    
There are Return for Repair Instructions and Form on the last pages of this manual. This Form can be copied or used as needed.
Manual Part Number 80003-092
October 2008 MCN-426, 10/13/10 MCN-13-011, 12/20/13
warrants new instruments to be free from defects in workmanship and material under normal use for a period of one
ENMET
within the warranty period (transportation prepaid) for inspection. If the evaluation by
Be sure to include paperwork. A purchase order, return address and telephone number will assist in the expedient repair and return of your unit. Include any specific instructions. For warranty service, include date of purchase If you require an estimate, please contact
. The warranty covers both parts and labor excluding instrument calibration and
ENMET
ENMET
ENMET
ENMET
shall not be liable for any loss or damage caused by the improper use of
which may arise in connection with this equipment.
Corporation.
confirms that the
ENMET
neither assumes
14
Page 17
PO Box 979 680 Fairfield Court Ann Arbor, Michigan 48106-0979
734.761.1270 Fax 734.761.3220
Returning an Instrument for Repair
ENMET instruments may be returned to the factory or any one of our Field Service Centers for regular repair service or calibration. The ENMET Repair Department and Field Service Centers also perform warranty service work.
When returning an instrument to the factory or service center for service, paperwork must be included which contains the following information:
A purchase order number or reference number. A contact name with return address, telephone and fax numbers Specific instructions regarding desired service or description
of the problems being encountered.
Date of original purchase and copy of packing slip or invoice
for warranty consideration.
If a price estimate is required, please note it accordingly and be
sure to include a fax number.
Providing the above information assists in the expedient repair and return of your unit.
Failure to provide this information can result in processing delays.
ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour. All instruments sent to ENMET are subject to a minimum evaluation fee, even if returned unrepaired. Unclaimed instruments that ENMET has received without appropriate paperwork or attempts to advise repair costs that have been unanswered, after a period of 60 days, may be disposed of or returned unrepaired COD with the evaluation fee.
Service centers may have different rates or terms. Be sure to contact them for this information.
Repaired instruments are returned by UPS/FedEx Ground and are not insured unless otherwise specified. If expedited shipping methods or insurance is required, it must be stated in your paperwork.
Note: Warranty of customer installed components. If a component is purchased and installed in the field, and fails within the warranty term, it can be
returned to ENMET and will be replaced, free of charge, per ENMETs returned goods procedure.
If the entire instrument is returned to ENMET Corporation with the defective item installed, the item will
be replaced at no cost, but the instrument will be subject to labor charges at half of the standard rate.
Page 18
Mailing Address:
Shipping Addr
ess:
Repair Return Form
ENMET Corporation PO Box 979 Ann Arbor, Michigan 48106
ENMET Corporation Attn: Repair Department 680 Fairfield Court
Ann Arbor, Michigan 48108 Phone Number: 734.761.1270 FAX Number: 734.761.3220
Your Mailing Address:
Contact Name: __________________________ Your Phone: _______________________
Your Shipping Address:
Your PO/Reference Number: _______________ Your FAX: _______________________ Payment Terms: K COD
(Check one) K VISA / MasterCard______________________ ________ ________
Card number Expiration Card Code
K American Express______________________ ________ ________
Card number Expiration Card Code Name as it appears on the credit card___________________________________________
Return Shipping Method:
K UPS: K Ground K 3 Day Select K Next Day Air K ND Air Saver K 2-Day Air K UPS Account number: _________________________
K Federal Express: K Ground K Express Saver K P-1 K Standard K 2-Day Air K FedEx Account number: ________________________
Would you like ENMET to insure the return shipment?
K No K Yes Insurance Amount: $_________________
Loading...