1.3 Check Order .............................................................................................................................................................................. 1
1.4 Serial Numbers.......................................................................................................................................................................... 1
2.0 Instrument Features ....................................................................................................................................................................... 2
2.1 CP-60 Exterior Features ............................................................................................................................................................ 2
2.2 Display CP-60 Front Panel Features ......................................................................................................................................... 2
2.3 Circuit Board Features .............................................................................................................................................................. 3
3.1 Mounting of Instrument ............................................................................................................................................................ 4
3.3 Power Supply ............................................................................................................................................................................ 4
3.4.4 Digital Outputs ................................................................................................................................................................... 7
4.6 Operational Menu ..................................................................................................................................................................... 9
5.2.6 Set Output Span Range .................................................................................................................................................... 15
5.2.7 Set New Password ............................................................................................................................................................ 16
5.2.8 Exit Maintenance Menu ................................................................................................................................................... 16
6.0 Technical Data and Specifications .............................................................................................................................................. 17
7.0 Replacement Part Numbers ......................................................................................................................................................... 17
: [affects equipment – if not followed may cause damage to instrument, sensor etc…]
W
ARNING
!
:
[affects personnel safety – if not followed may cause bodily injury or death.]
Attention / Warning
Earth Ground
Page 4
CP-60ENMET Corporation
1.0 Introduction
The CP-60 control panel is capable of monitoring from 1 to 3 remote sensors. Some features of the monitoring system are as
follows:
continuous LCD
menu driven operational and maintenance controls
menu driven calibration procedure
audio and visual alarms indicate unsafe conditions
alarm relay contacts available on terminals
a fault relay and visual fault alarm
alarm acknowledgement capability including audio defeat
mA output for each channel
N
OTE
: All specifications stated in this manual may change without notice.
1.1 IEC 60601-1 Classifications
Type of protection against electrical shock: Class 1.
Degree of protection against electrical shock: No Applied Parts
- Mains power quality should be that of a typical commercial environment.
1.2 Unpack
Unpack the CP-60 and examine it for shipping damage. If such damage is observed, notify both ENMET customer service
personnel and the commercial carrier involved immediately.
Regarding Damaged Shipments
N
OTE
: It is your responsibility to follow these instructions. If they are not followed, the carrier will not honor any
claims for damage.
This shipment was carefully inspected, verified and properly packaged at our company and delivered to the carrier in good condition.
When it was picked up by the carrier at ENMET, it legally became your company’s property.
If your shipment arrives damaged:
• Keep the items, packing material, and carton “As Is.” Within 5 days of receipt, notify the carrier’s local office and request immediate
inspection of the carton and the contents.
• After the inspection and after you have received written acknowledgment of the damage from the carrier, contact ENMET Customer
Service for return authorization and further instructions. Have your Purchase Order and Sales Order numbers available.
ENMET either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability coverage, usually $100.00.
Repair or replacement charges above that value are your company’s responsibility.
The shipping company may offer optional insurance coverage. ENMET only insures shipments with the shipping company when asked
to do so in writing by our customer. If you need your shipments insured, please forward a written request to ENMET Customer Service.
Regarding Shortages
If there are any shortages or questions regarding this shipment, please notify ENMET Customer Service within 5 days of receipt at the
following address:
ENMET Corporation
680 Fairfield Court
Ann Arbor, MI 48108
734-761-1270 734-761-3220 Fax
1.3 Check Order
Check, the contents of the shipment against the purchase order. Verify that the
order, ascertain that they are present. Check the contents of calibration kits. Notify ENMET customer service personnel of any discrepancy
immediately.
CP-60
is received as ordered. If there are accessories on the
1.4 Serial Numbers
Each
CP-60
is serialized. These numbers are on tags on the equipment and are on record in an ENMET database.
1
Page 5
CP-60 ENMET Corporation
SELECT
Power LED
Fault LED
Alarm 2 LED
Display
OPTION
Defe
Press
Punch out/Hole Plug
Visual Alarm LED
Mounting Flanges
Strain Relief
2.0 Instrument Features
2.1 CP-60 Exterior Features
The exterior of the instrument is shown in Figure 1. The exterior features are as follows:
Feature Description
Enclosure
Strain Relief
Audio Alarm
Mounting Flanges
2.2 Display CP-60 Front Panel Features
The display panel, shown in Figure 1, is viewed through the clear front cover of the enclosure, and is accessed by opening the
cover. Features are as follows:
Feature Description
Display
Visual
Alarms & Indicators
Pushbutton Switches
•OPTION Switch
•SELECT Switch
•
AUDIO DEFEAT
A
LARM ACKNOWLEDGE
/
Engineered thermoplastic, approximately 11x9x6, with a clear hinged front cover.
Entrance for Power and wiring to Remote Sensor
A loud horn activated by certain alarm conditions.
Flanges with holes for mounting the enclosure to a vertical surface.
2 line, 16 character per line, LCD with backlight.
The numerical values of gas concentrations, and other information are displayed.
On either sides of the display:
A red alarm LED for each sensor transmitter installed to the instrument, Alarm 1.
The top center:
A red alarm LED for all sensor transmitters, Alarm 2.
Near the center of the panel:
A green power indicator LED
A red fault alarm LED
There are three of these, located near the center of the panel; they are yellow
rectangular membrane switches. They are:
The top left switch.
Directly to the right of the
OPTION
switch.
Switch
Directly below the
OPTION
switch.
4 places
Alarm
For External Wiring
Switch
2 to 4
Figure 1: CP-60 External Features
2
Switch
2 places
Page 6
CP-60 ENMET Corporation
Relays 1, 2, 3
programmed to active
is a system relay
Digital Communication
4 –
20mA Output
Ground Screw J21
Power Input
Channel 1
Channel 2
Channel 3
4–20mA Transmitter 3
Relay Terminals
1 Gas
2 Gas
3 Gas
4 Gas
5 Gas
6 Fault
2.3 Circuit Board Features
The Display Panel is hinged on the left and is released by unscrewing the 2 philips screws located in the top and bottom right
corners. After releasing the panel, it is swung to the left, exposing the interior of the enclosure. The Circuit Board is mounted
on a plate at the back surface of the enclosure interior. Features are shown in Figure 2.
Feature Description
Relay Terminals
Output Terminals
This group of terminals is located at the left side of the Circuit Board.
For the contacts for each of four alarm relays, and for the contacts of a fault relay.
There are two for each of the 4-20 mA outputs.
(5Gas Alarms)
(1 Fault alarm)
are
channels of CP-60
Relay 4 is a spare relay,
Not programmed
Relay 5
for ch 1, 2, & 3
4–20mA Transmitter 1
4–20mA Transmitter 2
Fuse Holder
Hole Plugs
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Terminals
3 outputs
2 connectors
Terminals (3 types)
Horn Terminal
Audio Alarm
Terminal J23
Strain Relief
Figure 2: CP-60 Interior Features
3
Page 7
CP-60 ENMET Corporation
Mounting Holes
3.0 Installation
3.1 Mounting of Instrument
Mount the CP-60 instrument on an appropriate vertical surface using the mounting flanges provided. Avoid areas with
excessive vibration or temperature extremes. The holes in the flanges are 0.31 inch in diameter and form a 6 x 10.75 inch
rectangle. See Figure 3.
It is recommended to use #8 drywall anchors and screws for mounting the CP-60 to a drywall/sheetrock surface.
0.31” dia. 4 places
Dimensions are in inches.
Figure 3: CP-60 Mounting Dimensions
3.3 Power Supply
The electrical installation should conform to appropriate electrical codes, such as the National Electrical Code in the United
States.
W
ARNING
:
The compliance of the installation to appropriate codes is not ENMET’s responsibility.
The CP-60 should be powered through appropriately sized circuit breakers. See Section 6.0 Technical Data.
Upon supplying power to the CP-60:
The green power on LED is lit.
The display backlight is lit, and instrument will step through a start-up sequence: unit serial number, software revision and
gases monitored may be shown on the display.
The instrument may go into alarm briefly, but should stabilize quickly. If the instrument persists in alarm, acknowledge the
alarm by pressing the A
Sensor/Transmitter installation, call ENMET customer service personnel.
Mains power line is fused for power supply protection. Fuse is 5 x 20mm, 0.630Amp is located in FH2, see Figure 5
W
!
ARNING
UDIO DEFEAT /ALARM ACKNOWLEDGE
:
Continuous gas detection and alarm systems become inoperative upon loss of primary power.
switch. If alarm persists longer than 30 minutes, verify
4
Page 8
CP-60 ENMET Corporation
Heavier than Air Gas
Sensor Location
Bottled LP (liquefied petroleum)
Lighter than Air Gas
Sensor Location
Natural gas (methane)
Same Density as Air Gas
Sensor Location
3.4 Sensor/Transmitter Connection
Sensor/Transmitters are connected to the CP-60 control unit with two or three-conductor wiring, use the correct oiltight fitting.
Size of wire depends on the distance between the sensor/transmitter and the control unit.
See Recommended Wire Gauge Table below.
2 Wire for Sensors/Transmitter
3 Wire for Sensors/Transmitter
Position Function Position Function
1, V+
2, G
3, Sig
Power +24 VDC
Not Used
Signal/Return to Ground
1, V+
2, G
3, Sig
Power +24 VDC
Power Ground
Signal
Recommended Wire Gauge
Distance from Sensor to Control Unit
5000 feet 24 AWG
Greater then 5000 feet Contact Factory
N
OTE
: Sensor Location
Gases have different densities. Some are heavier than air and concentrate at the bottom of a space. Some are lighter than air
and gather at the top. Consider the density of the gas you want the sensor to detect when you install the sensor. Some
examples are given below.
Carbon Monoxide 4-6 feet above the (generally uniform) floor.
• DO N
OT
locate directly above or beside gas appliances (ovens,
heaters).
• Avoid locating anywhere near a vent or window or near an outside
doorway.
Near ceiling.
• DO N
OT
locate directly above appliances where it is subject to direct
exposure to heat or steam.
• DO N
OT
locate in direct air currents of windows, doors, or vents.
If you have a question involving the location of a unit or sensor, please contact your distributor or ENMET personnel. A
technician will analyze the question and recommend a location.
Interior wall; 18-24" from floor.
5
Page 9
CP-60 ENMET Corporation
J16 Connector
J18 Connector
J19 Connector
Connector 1 (J7)
Connector 1 (J4)
Power Input, Terminal J23
Ground Screw
Programmable Relays
3.4.1 Outputs
Two types of alarm outputs are available, relay contacts and 4-20mA outputs.
Channel 3
Factory Default Setting
as indicated
Transmitter 1
4-20mA Input
Channel 1
Relay 1
Channel 1
Alarm 1
Relay 2
Channel 2
Alarm 1
Relay 3
Channel 3
Alarm 1
Relay 4
Spare
Not Programed
Relay 5
Ch 1-3/ System
Alarm 2
Relay 6
Ch 1-3 / System
Fault
4-20mA Output
Channel 1 & 2
4-20mA Output
Connector RS485
Connector RS232
Connector RS485
Transmitter 2
4-20mA Input
Channel 2
Transmitter 3
4-20mA Input
Channel 3
Fuse Holder FH2
L-Line
N-Neutral
Figure 5: CP-60 Relay, Input and Output Terminals
6
Page 10
CP-60 ENMET Corporation
Posit
ion Failsafe
-
Alarm
Non-Failsafe
-
Alarm
3.4.2 Relay Contacts
Auxiliary alarms should be powered from an independent power source separate form the instrument power to avoid alarm
failure due to controller malfunction.
Relay are SPDT, rated at 10Amp at 110VAC or 10Amp at 30VDC for restive loads, and may be latching or non-latching as
required by the application. They are accessed on the terminals next to each relay see Figures 2 & 5. The contact positions are
noted on the circuit board next to each terminal.
These relay coils are energized when they are in the non-alarm state; the contact conditions given above are for the nonenergized state, which is identical to the alarm state.
In addition, there is a fault relay, which changes state whenever the instrument is in a fault condition. The contact positions are
noted on the circuit board next to each terminal. The coil of this relay is energized when the instrument is in the non-fault state;
the contact conditions given on the circuit board next to the terminal, are for the non-energized state, which is identical to the
fault state.
The PC Board is labeled for the relays in their un-energized state. If the relay is configured for failsafe, then this is also the alarm
condition state. Non-failsafe configured relays in the alarm state, are the reverse of the PC board labeling. Note that the
Fault(FLT) relay cannot be set to operate in a Non-Failsafe mode. Please see the Table 1 below:
Table 1 : Relay Factory Default Failsafe Settings
Factory Default Settings
Gas Alarm 1 J5 Relay 1 - NO Normally Open Normally Closed
Channel 1 J5 Relay 1 - COM Common Common
J5 Relay 1 - NC Normally Closed Normally Open
Gas Alarm 1 J6 Relay 2 - NO Normally Open Normally Closed
Channel 2 J6 Relay 2 - COM Common Common
J6 Relay 2 - NC Normally Closed Normally Open
Gas Alarm 1 J8 Relay 3 - NO Normally Open Normally Closed
Channel 3 J8 Relay 3 - COM Common Common
J8 Relay 3 - NC Normally Closed Normally Open
Gas Alarm 1 J10 Relay 4 - NO Normally Open Normally Closed
Spare J10 Relay 4 - COM Common Common
Not Programmed J10 Relay 4 - NC Normally Closed Normally Open
Gas Alarm 2 System J14 Relay 5 - NO Normally Open Normally Closed
Channel 1 – 3 J14 Relay 5 - COM Common Common
J14 Relay 5 - NC Normally Closed Normally Open
J15 Relay 6/FLT - NO Normally Open N/A
Fault J15 Relay 6/FLT - COM Common N/A
J15 Relay 6/FLT - NC Normally Closed N/A
These relay contacts can be used to operate auxiliary alarms or other functions.
3.4.3 4-20mA Outputs
Isolated 4-20 mA outputs are available for data logging or other purposes. An output is supplied for each active channel. These
outputs are available on the Connector 1(J7) for channels 1 & 2 and Connector 2(J4) for channels 3.
4mA corresponds to the sensor reading of 0 at the bottom of the range.
20mA corresponds to a full scale reading.
Wiring requirements are the same as for the relays.
3.4.4 Digital Outputs
The RS232 & RS485 connections are used for firmware updates and digital communications. The CP-60 is designed to
operate as a modbus slave. Contact the ENMET for further information on wiring for the digital outputs.
7
Page 11
CP-60 ENMET Corporation
O2 20.9
Disabld
O
%
H2S
CH4
(
)
N
:
4.0 Operation
4.1 Normal Operation Condition
With the CP-60 installed as described in Section 3, and in clean air, the POWER green LED is on, the display is lit, and the
information on the display is as shown in Figure 6, for the sensor(s) installed in the CP-60. The red alarm and fault LEDs are
not lit.
PPM
2
H2S 0
%LEL
OTE
(
Transmitter 1
Transmitter 2
Example of display with Oxygen Channel 1, Hydrogen Sulfide Channel 2 and Methane Channel 3 Disabled.
)
(
Transmitter 3
Programmed Non-active channels
will display Disabled
)
Figure 6: CP-60 Operational Display
When the CP-60 is first powered up, it goes through a series of momentary screens, which identify the instrument model number,
serial number and software revision. After all of the momentary screens have been displayed, the instrument arrives at the Main
Gas Display showing the gas concentration and unit of measurement in ppm, % or %LEL.
Depending on transmitter configuration and calibration condition, the furthest right character in the display may flash a letter
indicating the instrument status.
4.2 Alarm Conditions
There are two alarm set points available. These alarm points are normally set at established safety levels, such as the OSHA
Permissible Exposure Limit (PEL) for toxic gases or recognized standards below the Lower Explosive Limit for combustible
gases.
These alarm set points can be changed within limits; see the maintenance section of this manual for the procedure.
When the Oxygen, Toxic or combustible gas concentration reaches the alarm set point, the associated red LED is lit, the
associated relay changes state, and the audio alarm is activated
Pressing the A
about five minutes. If a second alarm condition occurs during this time the horn will re-activate. If the alarm condition(s) have
ended during this time, the horn will not re-activate.
Differential Setting
Differential is an optional alarm relay configuration where when an alarm has been triggered, the relay will remain in alarm
state until the sensor reading has moved the differential “value” in the non-alarm direction. See example below.
The Alarm 1 differential value is the delay of the CP-60 staying in alarm condition until after the measured reading has returned past the
alarm point by the differential value. Example: If the alarm point is V 19.5 and the differential is 2, the CP-60 will go into alarm at 19.5
and stay in alarm until the reading has risen above 21.5.
LARM ACKN/AUDIO DEFEAT
button can temporally disable the Audio Alarm. The horn will be disabled for
4.3 Alarm Latching
An instrument is shipped with the alarms in the latching mode. The alarms may be independently configured in the non-latching
mode or differential setting by use of the maintenance menu. See Section 5.2.2, for setting alarm 1 and alarm 2.
Standard Setting
I
N THE LATCHING MODE
contacts do not revert to the non-alarm state, until the ALARM ACKN/AUDIO DEFEAT switch is pressed. An alarm can also be
acknowledged by pressing the switch during the alarm condition; then at the cessation of the alarm condition, alarm indications cease and
alarm relays revert to the non-alarm state. After an alarm is acknowledged, alarms in the latching configuration are re-armed to latch at the
next alarm condition.
I
N THE NON-LATCHING MODE
relay contacts revert to the non-alarm state.
: at the cessation of the condition which causes an alarm, the alarm indications do not cease, and the alarm relay
: at the cessation of the condition that causes an alarm, the alarm indications automatically cease, and the alarm
4.4 Audio Defeat
Pressing the
60 not on the sensor/transmitter. Relays and alarm LEDs continue to function, in the alarm state, during an alarm condition.
As long as the alarm condition persists, the audio alarm will “chirp” every 20 seconds. After the alarm condition clears, the
audio will continue to “chirp” until the audio switch is pressed.
If after 15 minutes the alarm condition continues the audio alarm will reactivate at full intensity.
If any other alarm condition occurs while the audio alarm has been silenced it will force the audio alarm to reactivate immediately
AUDIO DEFEAT /ALARM ACKNOWLEDGE
switch during an alarm temporarily silences the audio alarm on the CP-
.
8
Page 12
CP-60 ENMET Corporation
O
S
ALARM1 Delays
4.5 Display
In clean air a display is shown in Figure 4. This position of the display is termed the "operational display". As explained below,
the display can be changed to furnish other information by using the OPTION and SELECT switches.
4.6 Operational Menu
The operational menu allows the user to:
View alarm set point concentration values
View alarm latching configurations
Enter the maintenance menu with the proper Password.
The operational menu is accessed with the OPTION and SELECT switches. The operational menu flow chart is shown in
Figure 7,
Pressing the OPTION switch is indicated with a "O"
Pressing the SELECT switch is indicated with a "S".
If the instrument is left at any location in the operational or maintenance menus, other than the operational display, with no
action taken for a period of 45 seconds, it returns to the operational display.
CH-1 CH-3
CH-2
S
No
Function
O
Relays 123456
6=ON
O
CH1 SCALE
0.0 – 30.0%
(O2)
O
vL19.5 A1 vL19.5
vD 19.5
O
vL17 A2 v17
vD17
No
S
Function
S
for each active channel
CH2 SCALE (O2)
O
0.00 – 30.0%
S
O
CH3 SCALE (O2)
0.0 – 30.0 %
S
No
S
Function
No
S
Function
Displays are examples:
6 – Indicates relay is engaged
Displays are examples of Alarms
Λ
- Indicates alarm triggered on increasing
value of reading
v - Indicates alarm triggered on decreasing
value of reading
Displays are examples of Alarms
L – Indicates alarm is in latching mode. Note:
Factory setting for alarms is latching
(no L present) – Indicates alarm is in non-
latching mode.
Displays are examples of Alarms
D – Indicates alarm is in Differential Setting.
(no D present) – Indicates alarm is in Standard
Setting.
O
(Seconds)
Alternating
5 5
5 5
No
S
Function
O
30 mASPAN 30.0
30
No
S
Function
O
O = Press Option switch
S = Press Select switch
Enter Maint Menu
Enter Password
_
See Maintenance Menu Diagram
Figure 7: CP-60 Operation Menu Flow Chart
4.7 Fault Indications
Fault indication occurs:
When a transmitter is not connected to an active channel of the CP-60.
When a connected Sensor/Transmitter reading falls out side of its operation parameters (See sensor/transmitter manual).
9
Page 13
CP-60 ENMET Corporation
!
O
S
From Operationa l Menu
O
S
O
S
O
S
O
S
Alarm 2
O
O
S
O
O
S
Ch 1, 2, 3
O
O
S
O
S
mA Output Span
S
O
S
Changes digit indicated by underscore cursor
Locks underscored digit and moves cursor to next digit
O
Valid
Invalid
O
S
5.0 Maintenance
The CP-60 has no specific preventative maintenance requirements. Entering the maintenance menu, as outlined in Section 5.2
may change instrument configurations.
5.1 Cleaning Instructions
C
AUTION
5.2 Maintenance Menu CP-60
To/Fro m
Operat ional
Dis play
: Never spray a cleaning solution on the surfaces of the CP-60.
Clean the exterior of the CP-60 enclosure with a mild soap solution on a clean, damp cloth. Do not soak the cloth with solution
so that moisture drips onto, or lingers on, external surfaces.
Under no circumstances should organic solvents such as paint thinner be used to clean instrument surfaces.
The maintenance menu diagram is shown in Figure 8 Maintenance Menu Flow Chart. From the operational display, press
the OPTION switch until; "EnterMAINTENANCE Menu" is displayed.
(6)
Press OPTION
Enter Maint Menu
Enter Password
O = Press Option
S = Press Select
0000
Note: When in maintenance
mode the “_” indicates
MAINTENANCE MENU
Scale mA Xmtrs
Scale mA Xmtrs
Select (Gas)
Press OPTION until gas to be Set-Up is displayed
See Section 5.2.1 for transmitter set-up instructions.
the location of the cursor,
the “I” block indicates a
non-used location.
MAINTENANCE MENU
Set Alarm1
MAINTENANCE MENU
Set Alarm2
MAINTENANCE MENU
Set Alarm Delays
MAINTENANCE MENU
Configure Alarms
MAINTENANCE MENU
Relay Failsafes
MAINTENANCE MENU
mA Output Span
MAINTENANCE MENU
Set New Password
Alarm 1
Sele ct: XX
Select: XX
Alarm Delay
Sele ct: XX
R1 L
Relay Fai lsafes
R:1 Failsa fe ON
Sele ct: (Gas)
Password
XXXX
Press OPTION until the gas alarm to be Set is displayed
See Section 5.2.2 for setting alarms instructions.
Press
PTION until the gas alarm to be Set is displayed
See Section 5.2.2 for setting alarms instructions.
Press
PTION until the gas alarm delay to be Set is displayed
See Section 5.2.3 for setting alarms instructions.
Pressing
See Section 5.2.4 for configuring relay instructions.
Pressing OPTION changes Failsafe setting from ON to OFF
See Section 5.2.5 for configuring relay failsafe instructions.
Press OPTION until the gas span to be Set is displayed
See Section 5.2.6 for output span instructions.
PTION changes letter indicated by underscore cursor
Changes digit indicated by underscore cursor
Locks underscored digit and move s curso r to next digit
MAINTENANCE MENU
Exit maint menu
Press OPTION to return to top of maintenance menu.
Press SELECT to return to operational menu.
See Section 5.3.8 for instructions on how to exit Maintenance Menus.
Figure 8: CP-60 Maintenance Menu Flow Chart
10
Page 14
CP-60 ENMET Corporation
Example:
O
S
S
Calibration
Select O2
MAINTENANCE MENU
Scale
mA
Xmtrs
Set Up
CH1:
O2 mA XMTER
4mA
= 0.0
S
each digit
CH1:
O2 mA XMTER
20mA
= 30.0
S
each digit
5.2.1 Set 4 –20mA Transmitter Scale
This section of the maintenance menu is for external sensor/transmitters. This function is normally performed at the factory
and is not usually required in the field unless a new transmitter with a different range is installed.
After entering a valid password into maintenance menu, the Scale mA Xmtrs section is the second menu section, if it is
installed, enter by pressing the S
Press the SELECT switch "mA Xmter Scale: Select XX" is displayed. XX = the gas to be set up.
Press the OPTION switch, if needed, to change to the gas to be set up.
Press the SELECT switch, “Ch#: mAXmter: 4mA: 0000” is displayed
Press the SELECT switch, that moves the cursor one digit to the right when the last digit is accepted the display move to
the full Scale mA Xmtrs menu
Press the SELECT switch, “Ch#: mAXmter: 20mA: 0000” is displayed
Press the SELECT switch, that moves the cursor one digit to the right when the last digit is accepted the display will return
to the Scale mA Xmtrs menu
Repeat these steps for each 4 –20mA transmitter.
Press the OPTION switch, to continue on to the next section of the Maintenance Menu.
Oxygen Sensor/Transmitter 0 – 30%
ELECT
switch
Press O
O
PTION
is displayed
until the gas to be
O = Press Option
S = Press Select
11
Page 15
CP-60 ENMET Corporation
5.2.2 Set Alarm Points
Factory alarm set points are discussed in Section 4.2, See Table 1. To change the alarm points, you must enter the maintenance
menu.
N
OTE
: Changing the alarm points on the CP-60 will NOT change the alarm points on the sensor transmitter.
Entrance to the maintenance menu is guarded with a four-digit Password. The factory default setting of the password is 1270.
When a valid numerical password is inserted, the user is allowed to enter the maintenance menu.
In the "Enter Maint Menu" position
Press the
will be the first digit of the password.
In theT000 position, the underline cursor is under the left digit.
Press the
Press the
Continue this process until the four-digit password is complete. When a valid password is inserted in this manner, the display is
transferred to the "Calibration" portion of the menu. If an invalid password is inserted you are returned to the Enter Maint
Menu display.
After entering a valid password:
Press the
Press the
Press the
Λ
for ascending trigger point or V for descending trigger point indicator.
Press the
Press the
Press the
Press the
IF: STD is selected, "ALARM 1 V L " is displayed.
The next character is the latching indicator L or NOL press the
The next characters are the alarm 1 value, press the
When the last digit is accepted display returns to the Maintenance Menu "Set Alarm1" position.
IF: DIFF is selected, "ALARM 1
The next characters are the alarm 1 value, press the
Move the cursor to the first digit and, Press the
right.
"ALARM 1
Press the
Note: The Alarm 1 differential value is the delay of the CP-60 staying in alarm condition until after the measured reading has
returned past the alarm point by the differential value. Example: If the alarm set point is V 19.5 and the differential is 2, the
CP-60 will go into alarm at 19.5 and stay in alarm until the reading has risen above 21.5.
Repeat the above procedure for each sensor alarm 1 to be changed.
Press the
Repeat as for alarm 1 using the STD section. Alarm Diff is not available for Alarm 2.
Press
Display
SELECT
OPTION
SELECT
OPTION
SELECT
SELECT
OPTION
SELECT
OPTION
SELECT
OPTION
OPTION
OPTION
switch "Enter Password T 0" is displayed. Press
SELECT
switch once, to move cursor to next digit, this
switch to change the left digit; select the correct digit.
switch, which locks the digit in place and moves the cursor one digit to the right.
switch until; “Maintenance Menu Set Alarm1” appears on display.
switch, "ALARM1 Select: O2" example is displayed.
switch; "ALARM 1 V " is displayed, with the indicator flashing,
switch to toggle between Λ and V; select the correct indicator.
switch to lock in the correct indicator. "ALARM 1
STD
" is displayed
switch to toggle between STD and DIFF; select the correct indicator.
switch to lock in the correct indicator.
DIFF BAND 0.0
OPTION
V DIFF 19.5
OPTION
" is displayed, Factory default setting.
OPTION
SELECT
switch to select each digit of the value
switch to select each digit of the value
switch to lock in the correct character and move the cursor to the
" is displayed, to set alarm 1 differential. With
switch to toggle the latching mode.
SELECT
switch move cursor to left.
switch to select each digit of the value.
switch to move to alarm 2, "Set ALARM2" is displayed.
switch until “Exit maint menu” appears, then press
SELECT
switch to return the instrument to the Operational
Note: When in maintenance mode the “_”
indicates the location of the cursor, the
“I” block indicates a non-used location.
12
Page 16
CP-60 ENMET Corporation
S
MAINTENANCE MENU
Set Alarm1
Alarm
1 -
Select: O2
Alarm
1
S
S
Alarm 1 DIFF
BAND
0
00
Alarm
1
S
Changes character indicated by
Alarm
1
S
Alarm
1
Alarm
1
DIFF
1
9.5
Locks underscored character and
S
O
S
O
S
Alarm 2 -
Select:
XX
MAINTENANCE MENU
Set Alarm2
Changes character indicated by underscore cursor
Locks underscored character and moves cursor to next
O
Press
O
until the channel to be
Set
is displayed
Alarm 2
V L 0
000
Example: Set Alarms Flow Chart
Displays shown are factory default settings.
Λ -
Indicates alarm triggered on increasing value of reading
v
-
Indicates alarm triggered on decreasing value of reading
L-
Indicates alarm is set for latching
NO
L-
Indicates alarm is set for non-latching
STD – Indicates alarm in standard setting, can be set in latched or non-latched mode
DIFF – Indicates alarm in differential setting, instrument will stay in alarm beyond the alarm set point by the differential value
S
O
= Press Option
S = Press Select
O
v T
v
STD
DIFF
TD
V
TL
v L–19.5
O
Toggles status between Λ increasing trigger and v decreasing trigger
S
Locks selection
O
Toggles status between Standard and D
S
Locks selection
v
Toggles status between
O
Latching and Non Latching
S
Locks selection
Changes character indicated
O
by underscore cursor
Locks underscored character and
S
moves cursor to next digit
IFF
Changes character indicated by
O
underscore cursor
S
moves cursor to next digit
O
underscore cursor
S
Locks underscored character and
moves cursor to next digit
PTION
13
Note: When in maintenance mode the “_”
indicates the location of the cursor, the
“I” block indicates a non-used location.
Page 17
CP-60ENMET Corporation
Example:
Example:
Channel 1
Channel 2
Channel 3
Channel 4
MAINTENANCE MENU
Set Alarm Delays
Alarm Delay -
Set
Select:
XX
Alarm Delay -
MAINTENANCE MENU
Configure Alarms
Press Option changes letter indicated by underscore cursor
R1
2 3
R2
5.2.3 Set Alarm “Delay On”
The alarms may be set to delay activation by 1 second increments, from 1 to 30 seconds.
W
ARNING
!
Factory set default value is 5 seconds.
After entering a valid password:
Press the OPTION switch until; “Maintenance Menu Set Alarm Delay” appears on display.
Press the SELECT switch, "ALARM Delay Select: O2" is displayed.
Press the SELECT switch; "ALARM Delay = T005" is displayed, with the underscore cursor under the left position.
Press the SELECT switch to move the cursor and the OPTION switch to lock in the correct digit and move the cursor one
digit to the right. Press the SELECT switch when the last digit is accepted display returns to the "Set Alarm Delay"
position.
Press the OPTION switch to continue on to the Set New Password section
Note:
:
Extended alarm delays may cause operational failures, injury or death.
When in maintenance mode the “_” indicates the location of the cursor, the “
Set Alarm Delay Flow Chart
S
Press Option until the channel to be
O
I
” block indicates a non-used location.
is displayed
O
0000
O
Changes digit indicated by underscore cursor
S
Locks underscored digit and moves cursor to next digit
5.2.4 Relay Configuration
S
To change a relay configuration you must enter the maintenance menu. Press the
displayed then press
SELECT
switch for the Enter Password menu. Enter the valid password as described below.
In the "Enter Maint Menu" position
Press the
SELECT
switch "Enter Password T 0" is displayed. Press
SELECT
this will be the first digit of the password.
In theT000 position, the underline cursor is under the left digit.
Press the
Press the
OPTION
SELECT
switch to change the left digit; select the correct digit.
switch, which locks the digit in place and moves the cursor one digit to the right.
Continue this process until the four-digit password is complete. When a valid password is inserted in this manner, the
display is transferred to the "Calibration" portion of the menu. If an invalid password is inserted you are returned to the
Enter Maint Menu display.
After entering a valid password:
Press the
OPTION
switch until “Configure Alarms” is displayed
L = Low Alarm = Alarm 1, H = High Alarm = Alarm 2, B = Both Alarms, = No Relay linked to channel
Set Relay Configuration Flow Chart
S
Ch 1,2,3
O
L
S
O
Ch 1
L
O
Changes digit indicated by underscore cursor
S
Locks underscored digit and moves cursor to next digit
OPTION
switch until “Enter Maint Menu” is
switch once, to move cursor to next digit,
O
= Press Option
S =
Press Select
The
Table 2
below shows the default relay links for 3 channel system.
Relay 1 Low Alarm
Relay 2 Low Alarm
Relay 3 Low Alarm
Relay 4 Spare Spare Spare
Relay 5 High Alarm High Alarm High Alarm
Relays can be linked to specific alarms.
N
OTE
: Each operating channel must be linked to at least 1 relay.
Table 2: Default Relay Links
14
Channel 4 Not
Available in CP-60
Default
programming of
circuit only
Page 18
CP-60ENMET Corporation
Example:
MAINTENANCE MENU
Relay
Failsafes
Relay
Failsafes
R: 1 Failsafe ON
Relay
Failsafes
R: 2 Failsafe
MAINTENANCE MENU
mA Output Span
mA OUTPUT SPAN
Select: CO
mA OUTPUT SPAN
_
50 CO
5.2.5 Failsafe Configuration
The CP-60 is factory set in a failsafe configuration. This means that if power is disconnected to the unit, or the relay fails
to properly engage, it fails in such a way that it is in the alarm position. ENMET recommends leaving the relay in a failsafe
configuration. However, relay 1 – 5 can be re-configured by using the following procedure.
To change a relay failsafe configuration you must enter the maintenance menu. Press the
Menu” is displayed then press
SELECT
switch for the Enter Password menu. Enter the valid password as described below.
In the "Enter Maint Menu" position
Press the
SELECT
switch "Enter Password T 0" is displayed. Press
SELECT
switch once, to move cursor to next digit,
this will be the first digit of the password.
In theT000 position, the underline cursor is under the left digit.
Press the
Press the
OPTION
SELECT
switch to change the left digit; select the correct digit.
switch, which locks the digit in place and moves the cursor one digit to the right.
Continue this process until the four-digit password is complete. When a valid password is inserted in this manner, the
display is transferred to the "Calibration" portion of the menu. If an invalid password is inserted you are returned to the
Enter Maint Menu display.
After entering a valid password:
Press the
Note:
When in maintenance mode the “_” indicates the location of the cursor, the “I” block indicates a non-used location.
OPTION
switch until “Relay Failsafes” is displayed
Set Relay Failsafe Configuration Flow Chart
S
O
Pressing
O
PTION
changes Failsafe setting from ON to OFF
O
O
OFF
O
Changes
S
Continue pressing
Failsafe setting from ON to OFF
S
ELECT
to cycle through all 5 relays
S
S = Press Select
OPTION
switch until “Enter Maint
= Press Option
5.2.6 Set Output Span Range
To change 4-20 mA output range. This range is set at the factory and should not be changed, contact ENMET for information.
Press the
Press
Operational Display
Example: Set Output Span Flow Chart
O
OPTION
OPTION
switch to continue to next section of maintenance menu.
switch until “Exit maint menu” appears and then press
S
O
Press
PTION
until the channel to be
O
S
O
Changes
S
Locks underscored digit and moves cursor to next digit
digit indicated by underscore cursor
S
SELECT
Set
switch to return the instrument to the
is displayed.
O = Press Option
S = Press Select
15
Page 19
CP-60 ENMET Corporation
Example:
Example:
MAINTENANCE MENU
Set New Password
Password
0000
S
5.2.7 Set New Password
To change the password, you must enter the maintenance menu. Press the
then press
SELECT
switch for the Enter Password menu. Enter the valid password as described in Section 5.2.1.
To set a new password, after inserting a valid password,
Press the
Press the
Use the
Press the
OPTION
SELECT
OPTION
SELECT
switch until; "Set New Password” is displayed.
switch; "Password T1270" is displayed, with the underscore cursor under the left digit.
switch to change the left digit, when the desired digit is displayed.
switch to lock the digit in place and move the cursor one digit to the right.
When all four digits of the new password have been selected, "Set New Password" is displayed.
Record the new password; without it, the maintenance menu cannot be reentered once you exit the Maintenance Menu. If the
password is lost, call ENMET customer service personnel.
From the "Password XXXX" position,
Press the
Press the
Note:
SELECT
OPTION
When in maintenance mode the “_” indicates the location of the cursor, the “I” block indicates a non-used location.
switch to return to Set New Password section.
switch; to continue to "exit MAINTENANCE Menu"
Set Password Flow Chart
S
T
O
Changes digit indicated by underscore cursor
S
Locks underscored digit and moves cursor to next digit
O
OPTION
switch until “Enter Maint Menu” is displayed
O = Press Option
S = Press Select
5.2.8 Exit Maintenance Menu
From the "exit MAINTENANCE Menu" position
Press the
Press the
SELECT
OPTION
switch to resume the operational display.
switch to reenter the maintenance menu at the "Calibration" position.
Exit Maintenance Menu Flow Chart
MAINTENANCE MENU
Exit maint menu
CH-1 CH-3
CH-2
MAINTENANCE MENU
Calibration
O
5.3 Channel Activation/Deactivation
Each channel of the CP-60 can be deactivated and activated as needed.
C
AUTION
: Deactivated channels will not display or respond to connected sensor transmitters.
A fault indication occurs when a transmitter is not connected to an active channel of the CP-60.
To Activate or deactivate a channel of the CP-60:
Enter the maintenance menu as described in Section 5.2 and press the
Press and Hold the
Press the
Press the
Press the
Press the
SELECT
OPTION
SELECT
SELECT
OPTION
switch until “ADV maint menu” is displayed.
switch to locate the channel to be changed.
switch to change the status of that channel.
switch to accept the change in status. “ADV maint menu Exit” is displayed.
switch to return to the operational menu.
OPTION
switch until Exit maint menu appears.
16
Page 20
CP-60 ENMET Corporation
Electrical Power
Storage
and Transport
Operation
Mechanical
Outputs
6.0 Technical Data and Specifications
The technical data and specifications of the CP-60 instrument.
15 Amp fused branch circuit
Strain relief: 5 – 12 mm OD
Digital: RS-232 – Modbus
Audio: 90 db at 2 ft
N
OTE
: All specifications stated in this manual may change without notice.
100 – 240 VAC
0.9 A
50/60 Hz
Board Mounted Fuse FH2, 0.630A, 5 x 20mm
Temperature:
preferred
Relative Humidity
Ambient Pressure
Temperature:
Relative Humidity
Ambient Pressure
Dimensions:
Weight:
Material:
Relays: SPDT
Analog: 4-20 mA x 3
-20° to +60°C (-4° to +140°F)
0° to +20°C (32° to 68°F)
0 – 100% RH, non-condensing
20 to 36 inHg (68 to 133 kPa)
0° to +40°C (32° to +104°F)
0 – 100% RH, non-condensing
20 to 36 inHg (68 to 133 kPa)
11 x 9 x 6 inches (28 x 23 x 16 cm)
8 lbs (3.6 kg)
Engineered thermoplastic with hinged front cover
Non-magnetic hardware
Resistive Load Inductive Load
10A at 110 VAC 7.5A at 110 VAC
10A at 30 VDC 5A at 30 VDC
RS-485 – Modbus
7.0 Replacement Part Numbers
ENMET part numbers for replacement parts:
Part number Description
64002-630 Fuse, 0.630 Amp 5x20mm
65057-011 Terminal plug, 3 position
65057-012 Terminal plug, 4 position
65057-010 Terminal plug, 2 position
17
Page 21
CP-60 ENMET Corporation
8.0 WARRANTY
ENMET warrants new instruments to be free from defects in workmanship and material under normal use for a period of one
year from date of shipment from ENMET. The warranty covers both parts and labor excluding instrument calibration and
expendable parts such as calibration gas, filters, batteries, etc ... Equipment believed to be defective should be returned to
ENMET within the warranty period (transportation prepaid) for inspection. If the evaluation by ENMET confirms that the
product is defective, it will be repaired or replaced at no charge, within the stated limitations, and returned prepaid to any
location in the United States by the most economical means, e.g. Surface UPS/FedEx Ground. If an expedient means of
transportation is requested during the warranty period, the customer is responsible for the difference between the most
economical means and the expedient mode. ENMET shall not be liable for any loss or damage caused by the improper use of
the product. The purchaser indemnifies and saves harmless the company with respect to any loss or damages that may arise
through the use by the purchaser or others of this equipment.
This warranty is expressly given in lieu of all other warranties, either expressed or implied, including that of merchantability,
and all other obligations or liabilities of ENMET which may arise in connection with this equipment. ENMET neither assumes
nor authorizes any representative or other person to assume for it any obligation or liability other than that which is set forth
herein.
NOTE: When returning an instrument to the factory for service:
Be sure to include paperwork.
A purchase order, return address and telephone number will assist in the expedient repair and return of your unit.
Include any specific instructions.
For warranty service, include date of purchase
If you require an estimate, please contact ENMET Corporation.
There is Return for Repair Instructions and Form on the last pages of this manual. This form can be copied or used as needed.
Manual Part Number
80003-501
July 2007
MCN-390, 11/14/07
MCN-401, 05/09/08
MCN-423, 07/15/09
Notes:
18
Page 22
PO Box 979
680 Fairfield Court
Ann Arbor, Michigan 48106-0979
734.761.1270 Fax 734.761.3220
Returning an Instrument for Repair
ENMET instruments may be returned to the factory or any one of our Field Service Centers for regular repair
service or calibration. The ENMET Repair Department and Field Service Centers also perform warranty
service work.
When returning an instrument to the factory or service center for service, paperwork must be included which
contains the following information:
A purchase order number or reference number.
A contact name with return address, telephone and fax numbers
Specific instructions regarding desired service or description
of the problems being encountered.
Date of original purchase and copy of packing slip or invoice
for warranty consideration.
If a price estimate is required, please note it accordingly and be
sure to include a fax number.
Providing the above information assists in the expedient repair and return of your unit.
Failure to provide this information can result in processing delays.
ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest
tenth of an hour. All instruments sent to ENMET are subject to a minimum evaluation fee, even if returned
unrepaired. Unclaimed instruments that ENMET has received without appropriate paperwork or attempts to
advise repair costs that have been unanswered, after a period of 60 days, may be disposed of or returned
unrepaired COD with the evaluation fee.
Service centers may have different rates or terms. Be sure to contact them for this information.
Repaired instruments are returned by UPS/FedEx Ground and are not insured unless otherwise
specified. If expedited shipping methods or insurance is required, it must be stated in your paperwork.
Note: Warranty of customer installed components.
If a component is purchased and installed in the field, and fails within the warranty term, it can be
returned to ENMET and will be replaced, free of charge, per ENMET’s returned goods procedure.
If the entire instrument is returned to ENMET Corporation with the defective item installed, the item will
be replaced at no cost, but the instrument will be subject to labor charges at half of the standard rate.
Page 23
Mailing Address:
Shipping Address:
Your Mailing Address:
Your Shipping Address:
Repair Return Form
ENMET Corporation
PO Box 979
Ann Arbor, Michigan 48106
Contact Name: __________________________ Your Phone: _______________________
Your PO/Reference Number: _______________ Your FAX: _______________________
ENMET Corporation
Attn: Repair Department
680 Fairfield Court
Ann Arbor, Michigan 48108