ENMET AM-5150 User Manual

Page 1
ENMET Corporation
PO Box 979 Ann Arbor, MI 48106-0979
www.enmet.com
AM-5150
Ambient Air Monitor
Manual
Manual part number 80003-550
Page 2
Table of Contents
1.0 INTRODUCTION................................................................................................................................................................ 1
1.1 U
NPACK
.............................................................................................................................................................................. 1
1.2 C
HECK ORDER
1.3 S
ERIAL NUMBERS
.................................................................................................................................................................... 1
............................................................................................................................................................... 1
2.0 COMPONENTS OF THE AM-5150 .................................................................................................................................. 2
2.1 AM-5150
2.2 AM-5150 O
2.3 C
IRCUIT BOARD FEATURES
ELEMENTS
PERATIONAL FEATURES
........................................................................................................................................................... 2
................................................................................................................................................. 3
................................................................................................................................... 2
3.0 INSTALLATION OF THE AM-5150 ................................................................................................................................ 4
3.1 M
OUNTING
AM-5150 ......................................................................................................................................................... 4
3.1.1 Wiring the AM-5150 .................................................................................................................................................. 4
3.1.2 Power Supply .............................................................................................................................................................. 4
3.2 S
ENSOR LOCATION
............................................................................................................................................................. 5
3.2.1 Sensor Hook-up ........................................................................................................................................................... 6
3.3 R
ELAY CONTACTS
.............................................................................................................................................................. 7
4.0 OPERATION ....................................................................................................................................................................... 8
4.1 S
TART UP
AM-5150 ........................................................................................................................................................... 8
4.1.1 Typical Start Up .......................................................................................................................................................... 8
4.2 N
ORMAL DISPLAY MODE
.................................................................................................................................................... 9
4.2.1 Alarm Conditions AM-5150 ....................................................................................................................................... 9
5.0 MAINTENANCE ............................................................................................................................................................... 10
5.1 M
AINTENANCE MENUS
5.2 C
ALIBRATION OF THE
..................................................................................................................................................... 10
AM-5150 ....................................................................................................................................... 12
5.2.1 Exit Maintenance Menu ............................................................................................................................................ 13
5.2.2 Zero Adjust ................................................................................................................................................................ 13
5.2.3 Gas Span ................................................................................................................................................................... 14
5.2.4 Alarm Set Points ....................................................................................................................................................... 15
5.2.5 mA Span Set .............................................................................................................................................................. 15
5.3 S
ENSOR REPLACEMENT
.................................................................................................................................................... 16
5.3.1 Sensor Replacement Calibration............................................................................................................................... 17
6.0 ACCESSORY AND REPLACEMENT PARTS ............................................................................................................. 18
7.0 TECHNICAL DATA AND SPECIFICATIONS ............................................................................................................. 18
8.0 WARRANTY ..................................................................................................................................................................... 19
List of Tables
T
ABLE 1 : RELAY FAILSAFE SETTINGS
T
ABLE 2:
AM-5150 M
AINTENANCE MENUS SEQUENCE
.......................................................................................................................... 7
............................................................................................... 10
List of Illustrations
F
IGURE 1: EXTERNAL
F
IGURE 2:
F
IGURE 3: MOUNTING
F
IGURE 4: POWER TERMINAL CONNECTIONS
F
IGURE 5: INTERNAL VIEW OF SENSOR WIRING
F
IGURE 6: RELAY TERMINAL CONNECTIONS
F
IGURE 7:
F
IGURE 8: CALIBRATION ADAPTER
F
IGURE 9:
AM-5150 C
AM-5150 M AM-5150 S
Reference Information:
N
OTE
: [important information about use of instrument]
C
W
ARNING
AUTION
: [affects equipment – if not followed may cause damage to instrument, sensor etc…]
:
[affects personnel safety – if not followed may cause bodily injury or death.]
AM-5150 F
IRCUIT BOARD FEATURES
EATURES
................................................................................................................... 2
.......................................................................................................... 3
AM-5150 ................................................................................................................................... 4
AM-5150 ................................................................................................. 5
............................................................................................................. 6
AM-5150 .................................................................................................. 7
AINTENANCE MENU FLOW CHART
............................................................................................ 11
............................................................................................................................. 12
ENSOR REPLACEMENT
............................................................................................................. 16
Earth Ground
Page 3
AM-5150 ENMET Corporation
1.0 Introduction
The AM-5150 is an ambient air monitoring instrument that measures and detects gases utilizing a non-specific MOS sensor. The AM-5150 is NOT in an enclosure rated for use in a Class I, Div 1, Groups B, C, D classified area and can not be installed in a hazardous location. However, the remote sensor housing and sensor are rated for Class I, Div 1, Groups A, B, C & D hazardous locations and may be installed in such locations with appropriate wiring.
Features of the AM-5150:
continuous monitoring of the ambient air continuous LCD display of gas and vapor concentrations menu driven operational and maintenance controls menu driven calibration procedure audio and visual alarms indicate unsafe conditions alarm relay contacts available on terminals a fault relay and visual fault alarm alarm acknowledgement capability including audio defeat mA outputs for target gas
N
OTE
: All specifications stated in this manual may change without notice.
1.1 Unpack
Unpack the AM-5150 and examine it for shipping damage. If such damage is observed, notify both ENMET customer service personnel and the commercial carrier involved immediately.
Regarding Damaged Shipments
N
OTE
: It is your responsibility to follow these instructions. If they are not followed, the carrier will not honor
any claims for damage.
This shipment was carefully inspected, verified and properly packaged at our company and delivered to the carrier in good condition.
When it was picked up by the carrier at ENMET, it legally became your company’s property.
If your shipment arrives damaged:
Keep the items, packing material, and carton “As Is.” Within 5 days of receipt, notify the carrier’s local office and
request immediate inspection of the carton and the contents.
After the inspection and after you have received written acknowledgment of the damage from the carrier, contact
ENMET Customer Service for return authorization and further instructions. Have your Purchase Order and Sales Order numbers available.
ENMET either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability coverage, usually $100.00. Repair or replacement charges above that value are your company’s responsibility.
The shipping company may offer optional insurance coverage. ENMET only insures shipments with the shipping company when asked to do so in writing by our customer. If you need your shipments insured, please forward a written request to ENMET Customer Service.
Regarding Shortages
If there are any shortages or questions regarding this shipment, please notify ENMET Customer Service within 5 days of receipt at the following address:
ENMET Corporation
680 Fairfield Court
Ann Arbor, MI 48108
734-761-1270 734-761-3220 Fax
1.2 Check Order
Check, the contents of the shipment against the purchase order. Verify that the AM-5150 is received as ordered. Each AM­5150 is labeled with its target gas. If there are accessories on the order, ascertain that they are present. Check the contents of
calibration kits. Notify ENMET customer service personnel of any discrepancy immediately.
1.3 Serial Numbers
Each AM-5150 is serialized. These numbers are on tags on the equipment and are on record in an ENMET database.
1
Page 4
AM-5150 ENMET Corporation
Power Wiring
Visual Indicators:
Pushbutton/Membrane
Visual Indicator
Power/Fault
Front Cover
Sensor/Remote Sensor
Wiring Access
Menu
Select
2.0 Components of the AM-5150
2.1 AM-5150 elements
See Figure 1 for location of elements:
Feature Description Enclosure
Front Cover
2.2 AM-5150 Operational Features
The Display Panel is attached by a cable and is released by unscrewing the 4 screws located in the corners. After releasing the panel, it is swung upward, exposing the interior of the enclosure. See Figure 1 for location of features.
Feature Description Display
Audio Alarm(Horn) Visual:
Indicators and Alarms
Membrane Switches
A polycarbonate box, approximately 7 x 5 x 3, with a detachable front cover. 4 holes for mounting the enclosure to a vertical surface. Located at the corners of the bottom
of the enclosure, directly beneath the 4 front cover retaining screws. See Figure 3 Detachable front cover of AM-5150 with Display Panel. See Section 2.2 and Figure 1 There are 4 Screws that hold the front cover in place.
A single line, 8 character LCD with backlight. Indicates the level of gas detected by sensor. The numerical value of gas concentration and other information is displayed.
Audio alarm (105 dB at 30cm/12in). The audio alarm is activated when the unit is in alarm.
LED indicators:
Power / Fault Indicator LED, Green / Red Alarm (3) Indicator LED, Red
2 Pushbutton Switches on front panel control the instrument maintenance functions. The pushbutton switch locations are indicated by:
M
ENU
: Advances the instrument display through operation information and maintenance
menus
S
ELECT
: Disables audio alarm temporally and
Selects the maintenance menu operations such as, Zero, Span, Exit menu or sets proper calibration values for Zero or Span
See Section 4.0 and 5.0 for operational and maintenance flow charts.
Three alarm points are preprogrammed into the AM-5150. At each alarm point, an LED on the front panel is activated. There are 4, 10 Amp relay contacts at each alarm point, plus a fault relay. See Section 3.2 for wiring information.
Alarm1, Alarm2, Alarm3
Switches
Audio Alarm
Retaining Screws 4 places
Strain Relief
Figure 1: External AM-5150 Features
2
Page 5
AM-5150 ENMET Corporation
Relay Terminals
Relays
Terminal J12
4-20mA Output
RS-485 Input/Output
Sensor Heat
er Voltage
Sensor
Connection
POT 3
2.3 Circuit Board Features
The Display Panel is attached by a cable and is released by unscrewing the 4 screws located in the corners. After releasing the panel, it is swung upward, exposing the interior of the enclosure. The Circuit Board is mounted at the back surface of the enclosure interior. Features are shown in Figure 2.
Feature Description Relay Terminals:
J14, J15, J16, J17
Terminal J12 Sensor Terminal J8 Data Terminal J19 Potentiometer 4 Potentiometer 2 & 3
POT 2
This group of terminals is located on the Circuit Board. For the contacts for each of three alarm relays, and for the contacts of a fault relay.
See Section 3.3 For VDC back-up power in and the 4-20 mA output. See Section 3.1.2 Sensor connection, See Section 3.2 RS-485 input/output Sensor heater voltage adjustment, See Section 3.2.1 Not used in AM-5150 Do Not Adjust
Terminal J19
DC Power In
Terminal J8
Adjustment POT 4
Figure 2: AM-5150 Circuit Board Features
J14, J15, J16, J17
K1, K2, K3, K4
3
Page 6
AM-5150 ENMET Corporation
3.0 Installation of the AM-5150
The AM-5150 is supplied with a strain relief for a power line cord. Use this fitting or connect a conduit fitting when supplying power to the unit.
N
OTE
: This control panel is NOT rated for hazardous locations. The control panel must be located in a NON-Hazardous area.
3.1 Mounting AM-5150
Mount the AM-5150 instrument on an appropriate vertical surface, leaving room for lid to be opened, using the mounting holes provided. Avoid areas with excessive vibration or temperature extremes. The holes in the bottom of the enclosure are 0.18 inch in diameter and form a 6.44 x 4.47 rectangle. See Figure 3
It is recommended to use #8 drywall anchors and screws for mounting the AM-5150 to a drywall/sheetrock surface.
Cover Inside View
Opened Upward
Attached to Base
Access for Sensor / Remote Sensor Wiring
Right Side View
Dimensions are in inches.
Figure 3: Mounting AM-5150
3.1.1 Wiring the AM-5150
The electrical installation should conform to appropriate electrical codes, such as the National Electrical Code in the United States.
W
ARNING
:
The compliance of the installation to appropriate codes is not ENMET’s responsibility.
The AM-5150 should be powered through circuit breakers provided for this purpose.
3.1.2 Power Supply
The input power can vary from 100 to 240 VAC, 50/60 Hz. Power should be connected to the Power Input Terminal TB1 and the Ground screw. See Figure 4 for location.
For DC wiring 24VDC may be wired to J12, (J12-1)position 1 + with ground connected to (J12-2)position 2. Upon supplying power to the AM-5150:
The green power on LED is lit. The display backlight is lit, and instrument will step through a start-up sequence: unit serial number and software revision
may be shown on the display.
The instrument may go into alarm briefly, but the sensors stabilize quickly. If the instrument persists in alarm, acknowledge the alarm by pressing the S
W
ARNING
:
Continuous gas detection and alarm systems (110VAC/220VAC / 24VDC/12VDC powered) become inoperative upon loss of
primary power. Contact factory for specifications and pricing of backup battery systems
ELECT
button. If alarm persists longer than 30 minutes, call ENMET customer service personnel.
.
4
Page 7
AM-5150 ENMET Corporation
Heavier than Air Gas
Sensor Location
Lighter than Air Gas
Sensor Location
Same Density as Air Gas
Sensor Location
AC Power Supply Terminal: TB1
Label on PCB Function 110VAC
TB1 ACL Line
Ground Screw
220VAC
Optional TB1 ACL Line
Ground Screw
C
AUTION
TB1 ACN Neutral
AC GND
TB1 ACN Neutral
AC GND
: 110/220 VAC when applied
Cover Inside View
Opened Upward
Sensor Heater Voltage Label
DC Power Supply/4-20mA
Terminal: J12
Position Function
1 + 24 V 2 GND 3 4 - 20 mA out
DC
power
Sensor Terminal: J8
Position Function Color
1 Heater Orange 2 Signal Blue 3 Ground 2 wires, Brown & Yellow
Sensor Heater Voltage Adjustment POT 4
Figure 4: Power Terminal Connections AM-5150
3.2 Sensor Location
Gases have different densities. Some are heavier than air and concentrate at the bottom of a space. Some are lighter than air and gather at the top. Consider the density of the gas you want the sensor to detect when you install the sensor. Some examples are given below.
If you have a question involving the location of a unit or sensor, please contact your distributor or ENMET personnel. A technician will analyze the question and recommend a location.
Bottled LP (liquefied petroleum) Interior wall; 18-24" from floor. Propane Butane Gasoline Trichloroethylene Vaporized hydrocarbons Hydrogen sulfide
Natural gas (methane) Near ceiling. Ammonia Hydrogen
Carbon Monoxide 4-6 feet above the (generally uniform) floor.
DO N  Avoid locating anywhere near a vent or window or near an outside
DO N
DO N
OT
locate directly above or beside gas appliances (ovens,
heaters).
doorway.
OT
locate directly above appliances where it is subject to
direct exposure to heat or steam.
OT
locate in direct air currents of windows, doors, or vents.
5
Page 8
AM-5150 ENMET Corporation
Sensor Type
Distance from Sensor to Control Unit
Recommended Wire Gauge
Wiring for
4-wire Terminal
Sensors
N
Figure 5: Internal View of Sensor Wiring
Optional
Sensor Wiring Terminal*
Optional
3.2.1 Sensor Hook-up
The MOS sensor is connected to the AM-5150 control unit with four-conductor wiring, use the correct oil tight fitting. Two conductors supply heater voltage and heater ground for the sensor. The third and fourth conductors are signal and signal ground wires. Size of the wire depends on the distance between the particular sensor and the control unit.
Recommended Wire Gauge
High Voltage: 812, 813, 826, etc.
Low Voltage: 109
Splash Shield
Splash Shield
500 feet
501 – 800 feet 14 AWG Longer Distances Contact ENMET Corp
50 feet
Longer Distances Contact ENMET Corp
16 AWG
16 AWG
Position Function Wire Color
1 Heater Orange 2 Signal Blue 3 Heater Ground Brown 4 Signal Ground Yellow
Wiring for 3-wire Terminal Sensors
Position Function Wire Color
1 Heater Orange 2 Signal Blue 3 Heater Ground Brown
Wire Nut Signal Ground Yellow
The 4th (yellow) wire is not connected to the terminal block, use supplied wire nut
Wire length between the AM-5150 and the sensor greater then 100ft will require that the sensor heater voltage be reset. After you mount and install the AM-5150 and Sensor, you must verify the sensor heater voltage. Use position 2 and 3 to measure heater voltage.
Locate the sensor heater voltage table label inside the instrument, see Figure 4. Measure the sensor heater voltage at the sensor, see Figure 5 and adjust the heater adjustment POT 4 until required voltage is reached, see Figure 4.
OTE
: The yellow signal wire is not connected to the
terminal block, use supplied wire nut.
6
Page 9
AM-5150 ENMET Corporation
Alarm
Position
Relay Terminals
Relays
Sensor Terminal J8
3.3 Relay Contacts
Relay contacts are available for each alarm; these are SPDT, rated at 10Amp at 110VAC, and may be latching or non-latching as required by the application.
They are accessed on the terminals next to each relay see Figure 6. The contact positions are noted on the circuit board next to each terminal.
The following table is for the relays in their un-energized state. This is also the alarm condition state. Non-failsafe configured relays in the alarm state, are the reverse of the PC board labeling. Note that the Fault(FLT) relay cannot be set to operate in a Non-Failsafe mode. Please see Table 1 below:
T
ABLE 1 : RELAY FAILSAFE SETTINGS
J14 (K1)Relay 1 - NO Normally Open
Alarm 1
J14 (K1)Relay 1 - NC Normally Closed J14 (K1)Relay 1 - COM Common
J15 (K2)Relay 2 - NO Normally Open
Alarm 2
J15 (K2)Relay 2 - NC Normally Closed J15 (K2)Relay 2 - COM Common
J16 (K3)Relay 3 - NO Normally Open
Alarm 3
J16 (K3)Relay 3 - NC Normally Closed J16 (K3)Relay 3 - COM Common
J17 (K4)Relay 4 - NO Normally Open
Fault Alarm
J17 (K4)Relay 4 - NC Normally Closed J17 (K4)Relay 4 - COM Common
These relay contacts can be used to operate auxiliary alarms or other functions. The relay contacts are DRY, power must be supplied. It is recommended that power for auxiliary equipment be supplied from an independent power source separate form the AM-5150. Use the existing hole in the enclosure for wire to enter and exit and use appropriate cable fittings. Wiring should be grouped together, VAC wires should be separated for VDC wires.
J14, J15, J16, J17
K1, K2, K3, K4
Figure 6: Relay Terminal Connections AM-5150
7
Page 10
AM-5150 ENMET Corporation
AM-5150
W
C
AM
-
5150
311
-
20
0ppC
0ppm
0LEL
0ppm
0ppP
0ppm
0ppR
4.0 Operation
When the AM-5150 is installed as described in Section 3, and in clean air, the POWER green LED is on, the display is lit and the information on the display is measurement of the target detected by the AM-5150. The red alarm and fault LEDs are not lit.
4.1 Start Up AM-5150
When the AM-5150 is first powered up, it goes through a series of momentary screens, which identify the instrument model number, serial number and software revision. After all of the momentary screens have been displayed, the instrument arrives at the Main Gas Display showing the gas concentration and unit of measurement, ppm or %LEL.
Depending on transmitter configuration and calibration condition, the furthest right character in the display may flash a letter indicating the instrument status. See the Section 4.1.1 below
4.1.1 Typical Start Up
When power is supplied to the AM-5150, the instrument will display the following sequence of information: Typical start up sequence of information displayed.
Example of Typical Start Up Display Function
Example for reference only
S/W 6.6F
Example for reference only
IF the right most character is a flashing
0ppW
The instrument: Model
The instrument: Serial Number
The instrument: Software Revision
The instrument is in Warm-up mode
This should last about 1 minute The Signal Output is held at 4mA during warm-up
IF the right most character is a flashing
OR
The instrument has failed Calibration
The last good calibration values are retained, but the sensor may not be responsive to gas A new Calibration should be performed As Soon As Possible
The instrument: Normal Display Mode Measurement of the target gas
The instrument: in Purge Mode Display
Optional feature, not required in all instruments
The instrument: in Recovery Mode Display
Optional feature, associated with purge mode
N
OTE
:
Software revision may cause variations of display output.
Prolonged storage of the AM-5150-MOS may result in some contamination of the sensor. Following start up, if the display reads above 000 a sensor purge function should be preformed.
Sensor Purge: A sensor purge is performed to clean the MOS sensor (Metal Oxide Sensor) of molecules that may have settled on the sensor.
N
OTE
:
Not all instruments are equipped with or require the Purge feature. To determine if the purge feature is operational.
Press and hold the S If the AM-5150-MOS is equipped with the purge feature, the display will alternate between 0ppm and 0ppP and the green Power/Fault LED will flash. In approximately 2 minutes the purge function will be completed and the display will, alternate between 0ppm and 0ppR indicating the instrument is in recovery mode. Recovery mode lasts approximately 5 minutes.
When the purge cycle is complete, the display will become stable at 0ppm and the green Power/Fault LED will stop flashing.
ELECT
switch, 3 to 5 seconds, if purge feature is operational 0ppP appears in the display.
8
Page 11
AM-5150 ENMET Corporation
4.2 Normal Display Mode
When the AM-5150 is installed as described in section 3, and in clean air, the POWER green LED is on, the display is lit and the information on the display in measurement units of ppm or %LEL detected by the AM-5150. The red alarm and fault LEDs are not lit.
To advance through displays of operational information press the M
N
OTE
:
Software revision may cause variations of display output.
See sequence of operational information below:
Display Measurement of the target gas Press M
ENU
button
Display indicates Alarm 1 Set point Press M
ENU
button
0ppm
M
ENU
A1: 05
M
ENU
ENU
S
button.
ELECT
S
ELECT
Display indicates Alarm 2 Set point Press M
ENU
button
Display indicates Alarm 3 Set point Press M
ENU
button
Display indicates mA Span range (Full Scale)
Press M
ENU
button
A2: 10
M
ENU
A3: 20
M
ENU
mA: 50
M
ENU
S
ELECT
S
ELECT
S
ELECT
Press S
ELECT
button
To temporally disable audio alarm, see Section 4.2.1
Display returns to gas measurent
Operational Display Flow Chart
4.2.1 Alarm Conditions AM-5150
There are three alarm set points available. These alarm points are normally set at established safety levels, such as the OSHA Permissible Exposure Limit (PEL) for toxic gases or recognized standards below the Lower Explosive Limit for combustible gases.
These alarm set points can be changed within limits; see the maintenance section of this manual for the procedure. If the gas concentration increases above that of the alarm set point, the associated red LED is lit, the associated relay changes
state, and the audio alarm is activated. Pressing the S
second alarm condition occurs during this time the horn will re-activate. If the alarm condition(s) have ended during this time the horn will not re-activate.
ELECT
button can temporally disable the Audio Alarm. The horn will be disabled for about five minutes. If a
9
Page 12
AM-5150 ENMET Corporation
5ppm
Exit
5.0 Maintenance
The AM-5150 maintenance menus that are accessed by pressing the M maintenance menu section.
5.1 Maintenance Menus
Pushbutton switches control the M display panel, see Figure 1. The M numbers such as alarm points, calibrations gas, etc. The S
To enter the maintenance menu press and hold the M Table 2 indicates the maintenance menu sequence see Figure 7 for a detailed maintenance menu flow chart.
C
AUTION
: Do Not Attempt A Span Procedure Without Calibration Gas Applied to The Sensor; if this is done, the instrument is
forced into a calibration fault mode.
N
OTE
:
Software revision may cause variations of display output.
Example of Display Function
Normal Display Mode Measurement of CO
T
ABLE 2:
ENU
and S
ELECT
functions. The M
ENU
button is used to display the various menu options and make incremental changes to
ELECT
ENU
button for 2 to 4 seconds
AM-5150 M
AINTENANCE MENUS SEQUENCE
ENU
button and S
ENU
and S
ELECT
button is used to select that option, set zero or span digit.
ELECT
button as described in the
button locations are indicated on the
Press and hold the M The Power/Fault LED will flash Green – Red to indicate the AM-5150 is in Maintenance Mode
Press the M
Press the M
Press the M
ENU
ENU
ENU
Alarm1
ENU
button for 2 – 4 seconds to enter the Maintenance Menu
To exit the maintenance Menu and return to the Normal Display Mode:
If intended function Press S
button to advance to the Zero procedure
Zero
button to advance to the Span procedure
Span
button to advance to each Alarm set point procedures
Alarm2
For adjusting Zero: If intended function Press S
For adjusting the Span: If intended function Press S
For adjusting the Alarm 1, 2 and 3 set points: If Intended function Press S
Alarm3
Press the M
Pressing the M You must Press the S
ENU
button to advance the mA Span set point procedure
ENU
button without pressing the S
ELECT
button in order to initiate the desired operation.
mA Span
ELECT
button will allow you to cycle through the menu options.
For adjusting the mA Span set point: If intended function Press S
ELECT
ELECT
ELECT
ELECT
ELECT
button
button
button
button
button
10
Page 13
AM-5150 ENMET Corporation
20
Apply
Same mV
H
S
5
To
Zero
Cal OK
If
display Bad ZERO
If
5150 will display
PV:
0
N
OTE
:
Software revision may cause variations of display output.
Normal Gas Display
5ppm
M
ENU
Press and H
OLD
the M
ENU
button for 2 – 4 seconds to enter the Maintenance Menus
Exit
M
M
Span
M
ENU
ENU
ENU
S
ELECT
Press the M
S
ELECT
OR
S
ELECT
Press the S
ENU
button to cycle through Maintenance Menus
ELECT
button to return to the Normal Gas Display. See Section 5.2.1
S
ELECT
Press the S
the Zero signal is within Preset Specs the AM-
Cal OK, See Section 5.2.2
Bad ZERO
PV: 0
the Zero signal is not within Preset Specs the AM-5150 will
S
ELECT
You can Press and change the Calibration Gas Level
See Section 5.2.3
S
ELECT
Cal Gas until signal value becomes stable
Cal OK
OR
(about 1 to 4 minutes) See Figure 8 Press S
ELECT
If cal is good display will indicate OK or Same
OR
Bad Sens
If cal is not within preset “range” display will
indicate Bad Sens
ELECT
button to initiate Zero adjustment
OLD
the
to enter the cal signal:
ELECT
button to
Alarm1
M
ENU
S
ELECT
S
ELECT
Alarm2
M
ENU
S
ELECT
Alarm3
M
ENU
S
ELECT
mA Span
M
ENU
return to Normal Gas Display: Press M Then press S
ENU
button until EXIT is displayed
ELECT
button
5
10
10
20
0050
F
IGURE 7:
To change Alarm set points: Press Menu button until Alarm to be changed is displayed Press Select button to display the set point
The M
ENU
button changes digit indicated by underscore cursor The S
ELECT
button locks underscored digit and moves to
next digit
If change is not within range display returns to first digit If change is within range display moves to next menu
See Section 5.2.4
To change mA Span set point: Press Menu button until mA Span is displayed Press Select button to display the set point
The M
ENU
button changes digit indicated by underscore cursor
The S
ELECT
button locks underscored digit and moves to next digit
See Section 5.2.5
AM-5150 Maintenance Menu Flow Chart
11
Page 14
AM-5150 ENMET Corporation
Menu
Select
Sensor
Enclosure
Humidifier
0ppm
0ppR
0ppm
0ppP
5.2 Calibration of the AM-5150
Sensor Purge: Prior to calibration some MOS(Metal Oxide Sensor) sensor should be purged. A sensor purge is performed to clean the MOS sensor of molecules that may have settled on the sensor. Press and hold the S 5150-MOS is equipped with the purge feature, the display will alternate between 0ppm and 0ppP and the green Power/Fault LED will flash. Example of purge display
In approximately 2 minutes the purge function will be completed and the display will, alternate between 0ppm and 0ppR indicating the instrument is in recovery mode. Recovery mode lasts approximately 5 minutes.
Example of purge recovery display When the purge cycle is complete, the display will become stable at 0ppm and the green Power/Fault LED will stop flashing. Calibration is the process of setting the instrument up to read accurately when exposed to the target gas. The Zero function sets the clean air reference point and the Span function sets the sensitivity of the instrument.
Initial Calibration: Wait at least 3 – 4 hours after initially supplying power to the AM-5150 instrument before initial calibration, overnight is preferred. The AM-5150 has been pre-calibrated at the factory, and initial field calibration should result in only fine-tuning to circuit, as well as a way to check that installation is successful. It is not necessary to open the enclosure to make adjustment. The calibration functions are operated with pushbuttons from outside the enclosure through the
MENU
and
SELECT
Calibration Zero and Span functions are two separate procedures. They operate independently of each other. It is recommended that the Zero procedure be done prior to the Span procedure.
ENMET Corporation recommends at least quarterly calibration of the AM-5150 instrument. Calibration equipment is available from ENMET Corporation to calibrate the AM-5150 instrument.
Calibration adapter, a length of tubing with a regulator for the gas cylinder on one end, and a calibration cup to connect to
the sensor of the AM-5150 on the other.
Gas cylinder, Zero gas 20.9% oxygen or Span gas, factory determined, varies by intent.
Generally, a cylinder of 20.9% Oxygen is used to provide a Zero point or fresh air reference for the calibration.
N
OTE
:
Software revision may cause variations of display output.
ELECT
switches.
switch, 3 to 5 seconds, if the purge feature is operational 0ppP appears in the display. If the AM-
To Cal Cup
Top View of Humidifier
Sensor
Calibration Cup
Figure 8: Calibration Adapter
12
To Regulator
Regulator
Gas Cylinder
Page 15
AM-5150 ENMET Corporation
If
If
PV:
0
H
M
Abort
?
5.2.1 Exit Maintenance Menu
Exit maintenance, by pressing the Exit appears on the display. Press the S Display.
Example of Exit menu:
S
ELECT
Exit
M
ENU
Press the M
Press the S
ENU
button to cycle through Maintenance Menus
ELECT
button to return to the Normal Gas Display.
ELECT
button to return to the instrument Normal Gas
5.2.2 Zero Adjust
The ZERO function must be performed by exposing the AM-5150 instrument to clean fresh air. If the air at the sensor is in question, use a cylinder of 20.9% oxygen to provide a clean air reference.
Enter the maintenance menu by pressing and holding M Menu flow chart.
After entering the maintenance menu, Press the M Press the S
ELECT
button to perform a Zero.
The display will alternate between Zero and PV: To abort Zero function press and hold M Abort? will appear, press S Press the S
ELECT
button to initiate a Zero adjustment.
ELECT
button to return to Zero.
An auto detect sequence is initiated. After 15 seconds, the AM-5150 will monitor the zero reading for stability.
If the reading stabilizes, within the pre-programmed perimeters, an automatic zero adjustment will be made.
Cal OK appears on the display and in 1 – 2 seconds, display will change to Span. If you wish to Span the sensor press the S
ELECT
If you wish to Exit the maintenance menu, press M the instrument Normal Gas Display
ENU
button for 2 to 4 seconds. See Figure 7, AM-5150 Maintenance
ENU
button until the Zero menu is displayed.
ENU
button for 3 – 4 seconds,
button you are now ready to apply gas. Proceed to gas span step 2
ENU
button until Exit is displayed, then press S
ELECT
button to return to
If the reading does not stabilize, within 255 seconds, the procedure will be aborted. Sensor is outside of safe parameters to
be zeroed, the display will read Bad Zero. Repeat Section 5.2.2 Zero Adjust making sure to use a Zero gas of 20.9% Oxygen. ENMET part number 03296-209.
Note: You can Press and
M
ENU
When Abort? appears press S
OLD
S
ELECT
the
ENU
button to abort Zero
ELECT
button to
return to Zero menu
Example of Zero adjustment display:
Zero
M
ENU
N
OTE
:
Software revision may cause variations of display output.
S
ELECT
Zero
Cal OK
OR
Bad ZERO
S
ELECT
Press the S
the Zero signal is within Preset Specs the AM-5150 will
display Cal OK momentarily then advance to Span menu
the Zero signal is not within Preset Specs the AM-5150
will display Bad Zero and return to Zero menu
ELECT
button to force Zero adjustment
13
Page 16
AM-5150 ENMET Corporation
Abort?
Apply
cal is not within preset “range” display will indicate Bad Sens
5.2.3 Gas Span
It is recommended that the Zero Function be performed first. Do not perform a calibration unless span gas is applied to sensor. Calibration can be aborted by pressing and holding the M
button for 3 – 4 seconds. Enter the maintenance menu. See Figure 7, AM-5150 Maintenance Menu flow chart.
1. Press the M
2. Press the S
The display will alternate between the calibration gas concentration (Example: Cal 20) and a signal level (PV).
To Abort calibration press and Hold M to Span.
N
OTE
:
To abort calibration. Press and H
M
ENU
ENU
button until Span display.
ELECT
button to perform a Span procedure.
When “Abort?” appears press S
ENU
button for 3 – 4 seconds, “Abort?” will appear, press S
OLD
the M
ENU
button to abort Calibration
ELECT
button to advance to desired menu
ELECT
button to return
S
ELECT
ENU
To change calibration gas level to be used, press and Hold S
ELECT
button for 3 – 4 seconds, use menu button to change
digit and select button to move to next digit.
N
OTE
:
To change calibration gas level. Press and H
S
ELECT
-Use the M
-Use the S
ENU
button to change digits
ELECT
button to move to next digit
OLD
the S
ELECT
button to change the Calibration Gas Level
20
3. Attach the associated calibration gas cylinder to the calibration adapter. See Figure 8 calibration adapter.
4. Open the valve to apply the calibration gas to the sensor. An auto detect sequence is initiated after 30 seconds, the AM-5150 will monitor the cal reading for stability.
5. Watch for the signal level to stabilize. This should take about 1 – 4 minutes.
6. Once the signal level has stabilized,
If the Span is successful, Cal OK appears momentarily, then will advance to Alarm1 menu.
To exit cal, press M
ENU
button until Exit appears and press S
ELECT
button
If the sensor is outside of acceptable parameters, Bad Span is displayed. If the sensor did not respond, an incompatible span gas was applied and the sensor did not respond at all,
Same mV is displayed then will return to Span. If calibration is not successful, it is suggested that calibration be attempted again in 30-60 minutes.
If the sensor will not calibrate See Section 5.3 for sensor replacement.
7. Remove the calibration gas.
8. Press the M
N
OTE
:
Software revision may cause variations of display output.
ENU
button to advance to Exit menu or desired menu.
Example of Calibration Display:
S
Span
ELECT
PV: 0
S
ELECT
M
ENU
PV: 0
20
Cal OK
Bad Sens
OR
Same mV
When cal signal is stable AM-5150 will automatically update: If cal is good display will indicate OK or Same and advance to
If
To exit press M
Cal Gas until signal value becomes stable
(about 1 to 4 minutes) See Figure 8 Calibration Adapter
Alarm1
or Same mV The AM-5150 will return to the Span Menu
ENU
button until Exit appears and press S
ELECT
14
Page 17
AM-5150 ENMET Corporation
Alarm1
Alarm2
Alarm3
20
mA Span
0200
5.2.4 Alarm Set Points
The AM-5150 has three alarm set points set at the factory. These alarm points are normally set at established safety levels. Alarm set points can be changed within limits.
To change any of the three alarm points: Enter the maintenance menu as shown in Figure 7 AM-5150 Maintenance Menu flow chart.
1. Press the M
2. Press the S
3. Press the M
4. Press the S When last digit is entered the AM-5150 will advance to the next menu
5. Press the M
N
OTE
: Alarms 2 and 3 can not be set below the Alarm 1 setting.
Example of Alarm Set Point menus:
M
M
ENU
ENU
ENU
button until to display Alarm1 is displayed.
ELECT
button to initiate alarm set point change
ENU
button to change the digit indicated by the underscore cursor
ELECT
button to move the cursor to the next digit
ENU
button to advance to the next menu
S
S
ELECT
S
ELECT
ELECT
5
5
10
10
To change Alarm set points: Press Menu button until Alarm to be changed is displayed Press Select button to display the set point
The M
ENU
cursor The S
ELECT
next digit
If change is not within range display returns to previous digit If change is within range display moves to next menu
button changes digit indicated by underscore
button locks underscored digit and moves to
M
ENU
N
OTE
:
Software revision may cause variations of display output.
5.2.5 mA Span Set
The AM-5150 4-20mA span range is set at the factory, normally to the full scale of the measurement and can be changed within limits.
To change the span range: Enter the maintenance menu as shown in Figure 7 AM-5150 Maintenance Menu flow chart.
1. Press the M
2. Press the S
3. Press the M
4. Press the S When last digit is entered the AM-5150 will advance to the next menu
5. Press the M
Example of mA Span menu:
M
ENU
Default mA Span
4mA 20mA 0 ppm 200 ppm
ENU
button until to display Span is displayed.
ELECT
button to initiate the mA Span menu
ENU
button to change the digit indicated by the underscore cursor
ELECT
button to move the cursor to the next digit
ENU
button to advance to the next menu
S
ELECT
To change mA Span set points: Press Menu button until mA Span is displayed Press Select button to display the set point
The M
ENU
button changes digit indicated by underscore cursor
The S
ELECT
button locks underscored digit and moves to next digit
15
Page 18
AM-5150 ENMET Corporation
Sensor
Spacer Ring
5.3 Sensor Replacement
The MOS sensor is durable, it can be purged of contaminants by operating in PURGE for a sufficient length of time and at regular intervals.
Gross contamination usually occurs during unavoidable misuse. Close exposure to an open gas flame, submersion of the sensor in a liquid, or continuous exposure to heavy concentrations of industrial vapors will grossly contaminate a sensor. A grossly contaminated sensor causes a continuous alarm.
If a sensor fails to calibrate, replace it.
P
W
ARNING
ROCEDURE
1. Obtain a new sensor assembly. Make sure the sensor type is identical to your original sensor
Sensor marking is on side of sensor see Figure 9 (Example: 813-4 the first 3 digits are sensor ID the 4 digit indicates a 4-wire sensor).
2. Disconnect the instrument for power.
3. Disconnect the orange, brown, blue and yellow sensor wires.
4. Unscrew the assembly from the sensor enclosure.
5. Remove the bad sensor.
6. Replace with the new sensor and reconnect the wires (See Section 3.2.1).
N
is not connected to the terminal block, use supplied wire nut.
7. Reconnect the instrument to power and verify sensor heater voltage. (See Section 3.2.1)
8. Recalibrate the instrument (See Section 5.3.1).
:
:
Power must be removed from the AM-5150 before this or any internal procedure. Failure to do so may cause
damage to equipment, bodily injury or death.
OTE
: The user must perform the four color-coded wiring attachments when replacing the sensor. If the yellow signal wire
*Replace facing direction as shown
Optional Splash Shield
Standard Sensor Shield
View of 4-wire Terminal
View of 3-wire Terminal
Figure 9: AM-5150 Sensor Replacement
After the new sensor assembly has been installed, it is suggested to allow the sensor to stabilize for at least 3 – 4 hours, preferably over night.
16
Page 19
AM-5150 ENMET Corporation
0ppm
0ppR
0ppm
0ppP
5.3.1 Sensor Replacement Calibration
Sensor Purge: Prior to calibration some MOS(Metal Oxide Sensor) sensor should be purged. A sensor purge is preformed to clean the MOS sensor of molecules that may have settled on the sensor. Press and hold the S 5150-MOS is equipped with the purge feature, the display will alternate between 0ppm and 0ppP and the green Power/Fault LED will flash. Example of purge display
In approximately 2 minutes the purge function will be completed and the display will, alternate between 0ppm and 0ppR indicating the instrument is in recovery mode. Recovery mode lasts approximately 5 minutes.
Example of purge recovery display When the purge cycle is complete, the display will become stable at 0ppm and the green Power/Fault LED will stop flashing. Calibration is the process of setting the instrument up to read accurately when exposed to the target gas. The Zero function sets the clean air reference point and the Span function sets the sensitivity of the instrument.
Initial Calibration: Wait at least 3 – 4 hours after initially supplying power to the AM-5150 instrument before initial calibration, overnight is preferred. The AM-5150 has been pre-calibrated at the factory, and initial field calibration should result in only fine-tuning to circuit, as well as a way to check that installation is successful. It is not necessary to open the enclosure to make adjustment. The calibration functions are operated with pushbuttons from outside the enclosure through the
MENU
and
SELECT
Calibration Zero and Span functions are two separate procedures. They operate independently of each other. It is recommended that the Zero procedure be done prior to the Span procedure.
ENMET Corporation recommends at least quarterly calibration of the AM-5150 instrument. Calibration equipment is available from ENMET Corporation to calibrate the AM-5150 instrument.
Calibration adapter, a length of tubing with a regulator for the gas cylinder on one end, and a calibration cup to connect to
the sensor of the AM-5150 on the other.
Gas cylinder, Zero gas 20.9% oxygen or Span gas, factory determined, varies by intent.
Generally, a cylinder of 20.9% Oxygen is used to provide a Zero point or fresh air reference for the calibration.
N
OTE
:
Software revision may cause variations of display output.
After entering the Maintenance menu, press and hold the M After 2-4 seconds, an F will appear on the far right hand side of the display. The F indicates that the instrument is in Factory
mode. Perform the calibration Zero and Span procedures as outlined in Section 5.2. Be sure that the F is present when selecting the
Zero and Span functions. The Factory calibration sets a calibration window for future standard instrument calibrations.
ELECT
switches.
switch, 3 to 5 seconds, if the purge feature is operational 0ppP appears in the display. If the AM-
ENU
button for 2-4 seconds while viewing the Zero menu.
17
Page 20
AM-5150 ENMET Corporation
Description of Part
Part Number
Electrical Power
Storage and Tr
ansport
Operation
Mechanical
Outputs
6.0 Accessory and Replacement Parts
ENMET accessory part numbers:
Sensor Marking, 109-4 03034-109 Sensor Marking, 812-4 03034-812 Sensor Marking, 813-4 03034-813 Sensor Marking, 018-4 03034-018 Sensor Marking, 019-4 03034-019 Sensor Marking, 821-4 03034-821 Sensor Marking, 830-4 03034-830 Sensor Marking, 832-4 03034-832
Zero Gas, 20.9% O2, 17 liter 03296-209 Span Gas Cylinder of Calibration Gas
N
OTE
: The gases should be identical, or equivalent (correlation gas) to those used to initially calibrate the unit, unless you want
to recalibrate to a new gas or different concentration. In addition, calibration gases must be in a background of air; do not use gases with an inert gas background (such as nitrogen or argon).
Contact ENMET for part number of target gas for each instrument. See note below.
7.0 Technical Data and Specifications
15 Amp fused branch circuit 100-240 VAC
0.45A, 50/60 Hz
0.6A, 24VDC Temperature:
preferred Relative Humidity Atmospheric Pressure
Temperature: Relative Humidity Atmospheric Pressure
Dimensions: Weight: Material: Strain relief:
Relays:
Analog: Digital: Audio:
N
OTE
: All specifications stated in this manual may change without notice.
-20° to +60°C (-4° to +140°F) 0° to +20°C (32° to 68°F) 10-99% RH, non-condensing 20 to 36 inHg (68 to 133 kPa)
-15° to +40°C (5° to +104°F) 10-99% RH, non-condensing 20 to 36 inHg (68 to 133 kPa)
7.1 x 5.1 x 3 in(180x130x75mm) 2 lbs (0.9 kg) Polycarbonate 3-6.5mm OD
SPDT Resistive Load Inductive Load 10A at 110 VAC 7.5A at 110 VAC 10A at 30 VDC 5A at 30 V 4-20mA RS-485-modbus 105 dB at 30cm/12in
DC
18
Page 21
AM-5150 ENMET Corporation
8.0 WARRANTY
ENMET warrants new instruments to be free from defects in workmanship and material under normal use for a period of one year from date of shipment from ENMET. The warranty covers both parts and labor excluding instrument calibration and expendable parts such as calibration gas, filters, batteries, etc... Equipment believed to be defective should be returned to ENMET within the warranty period (transportation prepaid) for inspection. If the evaluation by ENMET confirms that the product is defective, it will be repaired or replaced at no charge, within the stated limitations, and returned prepaid to any location in the United States by the most economical means, e.g. Surface UPS/FedEx Ground. If an expedient means of transportation is requested during the warranty period, the customer is responsible for the difference between the most economical means and the expedient mode. ENMET shall not be liable for any loss or damage caused by the improper use of the product. The purchaser indemnifies and saves harmless the company with respect to any loss or damages that may arise through the use by the purchaser or others of this equipment.
This warranty is expressly given in lieu of all other warranties, either expressed or implied, including that of merchantability, and all other obligations or liabilities of ENMET that may arise in connection with this equipment. ENMET neither assumes nor authorizes any representative or other person to assume for it any obligation or liability other than that, which is set forth herein.
NOTE: When returning an instrument to the factory for service:
Be sure to include paperwork. A purchase order, return address and telephone number will assist in the expedient repair and return of your unit. Include any specific instructions. For warranty service, include date of purchase If you require an estimate, please contact ENMET Corporation.
There are Return for Repair Instructions and Form on the last pages of this manual. This Form can be copied or used as needed.
Manual part number 80003-550
November 2006 MCN-367, 05/17/07
MCN-372, 06/18/07 MCN-385, 10/22/07 MCN-409, 12/04/08 MCN-413, 03/17/09 MCN-446, 03/24/11 MCN-453, 07/19/11
19
Page 22
PO Box 979 680 Fairfield Court Ann Arbor, Michigan 48106-0979
734.761.1270 Fax 734.761.3220
Returning an Instrument for Repair
ENMET instruments may be returned to the factory or any one of our Field Service Centers for regular repair service or calibration. The ENMET Repair Department and Field Service Centers also perform warranty service work.
When returning an instrument to the factory or service center for service, paperwork must be included which contains the following information:
A purchase order number or reference number.  A contact name with return address, telephone and fax numbers  Specific instructions regarding desired service or description
of the problems being encountered.
Date of original purchase and copy of packing slip or invoice
for warranty consideration.
If a price estimate is required, please note it accordingly and be
sure to include a fax number.
Providing the above information assists in the expedient repair and return of your unit.
Failure to provide this information can result in processing delays.
ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour. All instruments sent to ENMET are subject to a minimum evaluation fee, even if returned unrepaired. Unclaimed instruments that ENMET has received without appropriate paperwork or attempts to advise repair costs that have been unanswered, after a period of 60 days, may be disposed of or returned unrepaired COD with the evaluation fee.
Service centers may have different rates or terms. Be sure to contact them for this information.
Repaired instruments are returned by UPS/FedEx Ground and are not insured unless otherwise specified. If expedited shipping methods or insurance is required, it must be stated in your paperwork.
Note: Warranty of customer installed components.
If a component is purchased and installed in the field, and fails within the warranty term, it can be
returned to ENMET and will be replaced, free of charge, per ENMETs returned goods procedure.
If the entire instrument is returned to ENMET Corporation with the defective item installed, the item will
be replaced at no cost, but the instrument will be subject to labor charges at half of the standard rate.
Page 23
Mailing Address:
Shipping Address:
Repair Return Form
ENMET Corporation PO Box 979 Ann Arbor, Michigan 48106
Phone Number: 734.761.1270 FAX Number: 734.761.3220
Your Mailing Address:
Contact Name: __________________________ Your Phone: _______________________ Your PO/Reference Number: _______________ Your FAX: _______________________
ENMET Corporation Attn: Repair Department 680 Fairfield Court Ann Arbor, Michigan 48108
Your Shipping Address:
Payment Terms: K COD
(Check one) K VISA / MasterCard______________________ ________ ________
Card number Expiration Card Code
K American Express______________________ ________ ________
Card number Expiration Card Code
Name as it appears on the credit card___________________________________
Return Shipping Method:
K UPS: K Ground K 3 Day Select K Next Day Air K ND Air Saver K 2-Day Air K UPS Account number: ________________________
K Federal Express: K Ground K Express Saver K P-1 K Standard K 2-Day Air K FedEx Account number: ________________________
Would you like ENMET to insure the return shipment? K No K Yes Insurance Amount: $_________________
Loading...