Endress+Hauser Liquisys M CLM223, Liquisys M CLM253 Operating Instructions Manual

Page 1
Products Solutions Services
Operating Instructions
Liquisys M CLM223/253
Transmitter for Conductivity
BA00193C/07/EN/15.15 71300836
Page 2
Page 3
Endress+Hauser 3
Table of contents
1 Document information .............. 5
1.1 Warnings ............................ 5
1.2 Symbols used .......................... 5
1.3 Symbols on the device ................... 5
1.4 Electrical symbols ...................... 6
2 Basic safety instructions ............ 7
2.1 Requirements for personnel ............... 7
2.2 Designated use ........................ 7
2.3 Occupational safety ..................... 7
2.4 Operational safety ...................... 8
2.5 Product safety ......................... 8
2.5.1 State of the art .................. 8
2.5.2 IT security ...................... 8
3 Incoming acceptance and product
identification ....................... 9
3.1 Incoming acceptance .................... 9
3.2 Scope of delivery ....................... 9
3.3 Product identification ................... 10
3.3.1 Nameplate .................... 10
3.3.2 Product identification ............ 10
3.4 Certificates and approvals ............... 10
3.4.1  mark ...................... 10
3.4.2 CSA General Purpose ............. 10
4 Installation ....................... 11
4.1 Installation at a glance .................. 11
4.1.1 Measuring system ............... 12
4.2 Installation conditions .................. 13
4.2.1 Field device .................... 13
4.2.2 Panel-mounted device ............ 14
4.3 Installation instructions ................. 15
4.3.1 Field device .................... 15
4.3.2 Panel-mounted device ............ 17
4.4 Post-installation check .................. 17
5 Electrical connection .............. 18
5.1 Wiring ............................. 18
5.1.1 Wiring diagram ................. 18
5.1.2 Measuring cables and sensor
connection .................... 21
5.2 Alarm contact ........................ 24
5.3 Post-connection check .................. 24
6 Operation options ................. 25
6.1 Quick operation guide .................. 25
6.2 Display and operating elements ........... 25
6.2.1 Display ....................... 25
6.2.2 Operating elements .............. 27
6.2.3 Key functions .................. 27
6.3 Local operation ....................... 29
6.3.1 Automatic/manual mode .......... 29
6.3.2 Operating concept ............... 30
7 Commissioning .................... 32
7.1 Function check ....................... 32
7.2 Switching on ......................... 32
7.3 Quick Setup .......................... 34
7.4 Device configuration ................... 37
7.4.1 Setup 1 (conductivity) ............ 37
7.4.2 Setup 2 (temperature) ............ 39
7.4.3 Current input .................. 42
7.4.4 Current outputs ................. 45
7.4.5 Alarm ........................ 49
7.4.6 Check ........................ 50
7.4.7 Relay configuration .............. 53
7.4.8 Temperature compensation with
table ......................... 69
7.4.9 Concentration Measurement ....... 70
7.4.10 Service ....................... 73
7.4.11 E+H Service .................... 75
7.4.12 Interfaces ..................... 76
7.4.13 Communication ................. 76
7.5 Calibration .......................... 77
8 Diagnostics and troubleshooting ... 81
8.1 Trouble shooting instructions ............. 81
8.2 System error messages .................. 81
8.3 Process-specific errors .................. 84
8.4 Device-specific errors ................... 88
9 Maintenance ...................... 90
9.1 Maintenance of the entire measuring point .. 90
9.1.1 Cleaning the transmitter .......... 90
9.1.2 Cleaning the conductivity sensors ... 91
9.1.3 Simulation of conductive sensors for
device testing .................. 91
9.1.4 Simulation of inductive sensors for
device testing .................. 92
9.1.5 Checking conductive sensors ....... 93
9.1.6 Checking inductive sensors ........ 94
9.1.7 Assembly ..................... 94
9.1.8 Connecting cables and junction
boxes ........................ 94
10 Repair ............................ 95
10.1 Spare parts .......................... 95
10.2 Disassembling the panel-mounted device .... 95
10.3 Disassembling the field device ............ 98
10.4 Replacing the central module ............ 102
10.5 Return ............................. 103
10.6 Disposal ........................... 103
Page 4
Table of contents Liquisys M CLM223/253
4 Endress+Hauser
11 Accessories ...................... 104
11.1 Sensors ............................ 104
11.1.1 Sensors with conductive
measurement of conductivity ...... 104
11.1.2 Sensors with inductive measurement
of conductivity ................ 104
11.2 Connection accessories ................. 105
11.3 Installation accessories ................ 105
11.4 Software and hardware add-ons .......... 106
11.5 Calibration solutions .................. 107
12 Technical data ................... 108
12.1 Input .............................. 108
12.2 Output ............................ 108
12.3 Power supply ........................ 112
12.4 Performance characteristics ............. 113
12.5 Environment ........................ 113
12.6 Mechanical construction ............... 114
13 Appendix ........................ 115
Index ................................. 120
Page 5
Liquisys M CLM223/253 Document information
Endress+Hauser 5
1 Document information
1.1 Warnings
Structure of information Meaning
L
DANGER
Causes (/consequences)
Consequences of non-compliance (if applicable)
‣
Corrective action
This symbol alerts you to a dangerous situation. Failure to avoid the dangerous situation will result in a fatal or serious injury.
L
WARNING
Causes (/consequences)
Consequences of non-compliance (if applicable)
‣
Corrective action
This symbol alerts you to a dangerous situation. Failure to avoid the dangerous situation can result in a fatal or serious injury.
L
CAUTION
Causes (/consequences)
Consequences of non-compliance (if applicable)
‣
Corrective action
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or more serious injuries.
NOTICE
Cause/situation
Consequences of non-compliance (if applicable)
‣
Action/note
This symbol alerts you to situations which may result in damage to property.
1.2 Symbols used
Additional information, tips Permitted or recommended Forbidden or not recommended
1.3 Symbols on the device
Symbol Meaning
Reference to device documentation
Page 6
Document information Liquisys M CLM223/253
6 Endress+Hauser
1.4 Electrical symbols
Symbol Meaning
A0027423
Direct current
A terminal at which DC is present or through which DC flows.
A0027424
Alternating current
A terminal to which alternating voltage (sine-wave) is applied or through which alternating current flows.
A0027425
Direct current or alternating current
A terminal at which direct voltage or alternating voltage is present or through which direct current or alternating current flows.
A0027426
Ground connection
A terminal which, from the user's point of view, is already grounded via a grounding system.
A0027427
Protective ground connection
A terminal which must be connected to ground prior to establishing any other connections.
A0019929
Class II equipment
Reinforced or double insulation
A0027420
Alarm relay
A0027428
Input
A0027429
Output
A0027430
DC voltage source
A0027431
Temperature sensor
Page 7
Liquisys M CLM223/253 Basic safety instructions
Endress+Hauser 7
2 Basic safety instructions
2.1 Requirements for personnel
• Installation, commissioning, operation and maintenance of the measuring system may be carried out only by specially trained technical personnel.
• The technical personnel must be authorized by the plant operator to carry out the specified activities.
• The electrical connection may be performed only by an electrical technician.
• The technical personnel must have read and understood these Operating Instructions and must follow the instructions contained therein.
• Measuring point faults may be repaired only by authorized and specially trained personnel.
Repairs not described in the Operating Instructions provided may only be carried out directly by the manufacturer or by the service organization.
2.2 Designated use
Liquisys M is a transmitter for determining the conductivity and resistivity of liquid media.
The transmitter is particularly suited for use in the following areas:
• Ultrapure water
• Water treatment
• Cooling water desalination
• Condensate treatment
• Municipal wastewater treatment plants
• Chemical industry
• Food industry
• Pharmaceutical industry
Use of the device for any purpose other than that described, poses a threat to the safety of people and of the entire measuring system and is therefore not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
2.3 Occupational safety
As the user, you are responsible for complying with the following safety conditions:
• Installation guidelines
• Local standards and regulations
Electromagnetic compatibility
• The product has been tested for electromagnetic compatibility in accordance with the applicable European standards for industrial applications.
• The electromagnetic compatibility indicated applies only to a product that has been connected in accordance with these Operating Instructions.
Page 8
Basic safety instructions Liquisys M CLM223/253
8 Endress+Hauser
2.4 Operational safety
1. Before commissioning the entire measuring point, verify that all connections are correct. Ensure that electrical cables and hose connections are undamaged.
2. Do not operate damaged products, and safeguard them to ensure that they are not operated inadvertently. Label the damaged product as defective.
3. If faults cannot be rectified: Take the products out of operation and safeguard them to ensure that they are not
operated inadvertently.
2.5 Product safety
2.5.1 State of the art
The product is designed to meet state-of-the-art safety requirements, has been tested, and left the factory in a condition in which it is safe to operate. The relevant regulations and European standards have been observed.
2.5.2 IT security
We only provide a warranty if the device is installed and used as described in the Operating Instructions. The device is equipped with security mechanisms to protect it against any inadvertent changes to the device settings.
IT security measures in line with operators' security standards and designed to provide additional protection for the device and device data transfer must be implemented by the operators themselves.
Page 9
Liquisys M CLM223/253 Incoming acceptance and product identification
Endress+Hauser 9
3 Incoming acceptance and product
identification
3.1 Incoming acceptance
1. Verify that the packaging is undamaged.  Notify your supplier of any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
2. Verify that the contents are undamaged.  Notify your supplier of any damage to the delivery contents.
Keep the damaged products until the matter has been settled.
3. Check the delivery for completeness.  Check it against the delivery papers and your order.
4. Pack the product for storage and transportation in such a way that it is protected against impact and moisture.
 The original packaging offers the best protection.
The permitted ambient conditions must be observed (see "Technical data").
If you have any questions, please contact your supplier or your local sales center.
3.2 Scope of delivery
The delivery of the field device comprises:
• 1 transmitter CLM253
• 1 plug-in screw terminal, 3-pin
• 1 cable gland Pg 7
• 1 cable gland Pg 16 reduced
• 2 cable glands Pg 13.5
• 1 set of Operating Instructions
• For versions with HART communication: 1 set of Operating Instructions: Field communication with HART
• For versions with PROFIBUS interface: 1 set of Operating Instructions: Field communication with PROFIBUS PA/DP
The delivery of the panel-mounted device comprises:
• 1 transmitter CLM223
• 1 set of plug-in screw terminals
• 2 tensioning screws
• 1 set of Operating Instructions
• For versions with HART communication: 1 set of Operating Instructions: Field communication with HART
• For versions with PROFIBUS interface: 1 set of Operating Instructions: Field communication with PROFIBUS PA/DP
Page 10
Incoming acceptance and product identification Liquisys M CLM223/253
10 Endress+Hauser
3.3 Product identification
3.3.1 Nameplate
The nameplate provides you with the following information on your device:
• Manufacturer identification
• Order code
• Extended order code
• Serial number
• Ambient and process conditions
• Input and output values
• Safety information and warnings
Compare the data on the nameplate with your order.
3.3.2 Product identification
The order code and serial number of your product can be found in the following locations:
• On the nameplate
• In the delivery papers
Obtaining information on the product
1. Go to the product page for your product on the Internet.
2. In the navigation area on the right-hand side, select "Check your device features" under "Device support".
 An additional window opens.
3. Enter the order code from the nameplate into the search field.  You will receive information on each feature (selected option) of the order code.
3.4 Certificates and approvals
3.4.1  mark
The product meets the requirements of the harmonized European standards. As such, it complies with the legal specifications of the EC directives. The manufacturer confirms successful testing of the product by affixing to it the mark.
3.4.2 CSA General Purpose
The following device versions meet the requirements of CSA and ANSI/UL for Canada and the US:
• CLM253-**2/3/7***
• CLM223-**2/3/7***
Page 11
Liquisys M CLM223/253 Installation
Endress+Hauser 11
4 Installation
4.1 Installation at a glance
Proceed as follows to completely install the measuring point:
• Install the transmitter (see the "Installation instructions" section).
• If the sensor is not yet installed in the measuring point, install it (see Technical Information of the sensor).
• Connect the sensor to the transmitter as illustrated in the "Electrical connection" section.
• Connect the transmitter as illustrated in the "Electrical connection" section.
• Commission the transmitter as explained in the "Commissioning" section.
Page 12
Installation Liquisys M CLM223/253
12 Endress+Hauser
4.1.1 Measuring system
A complete measuring system comprises:
• Transmitter Liquisys M CLM223 or CLM253
• sensor with or without an integrated temperature sensor
• If necessary, Measuring cable: CYK71 or CPK9 (conductive) or CLK6 (inductive)
Optionally:
• Eintaucharmatur, z.B. CLA111
• Extension cable, junction box VBM
• Weather protection cover CYY101 for field housing
ENDRESS+HAUSER LIQUISYS S
ENDRESS+HAUSER LIQUISYS S
1
2
3
4
5
6
7
A0024642
 1 Complete measuring systems
1 Conductive sensor CLS15 2 Liquisys M CLM253 3 Liquisys M CLM223 4 Inductive sensor CLS54 5 Inductive sensor CLS50 6 Conductive sensor CLS21 7 Immersion assembly CLA111
Page 13
Liquisys M CLM223/253 Installation
Endress+Hauser 13
4.2 Installation conditions
4.2.1 Field device
Pg 7
Pg 16
M 5
11 (0.43)
Ø 6 (0.24)
115 (4.53)
157 (6.18)
247 (9.72)
70 (2.76)
70 (2.76)
154 (6.06)
170 (6.69)
Pg 13.5
A0024637
 2 Field device, dimensions in mm (inch)
There is a hole in the perforation for the cable entry (connection of supply voltage). It serves as a pressure balance during air shipment. Make sure no moisture penetrates the inside of the housing before the cable installation. The housing is completely air­tight after cable installation.
Page 14
Installation Liquisys M CLM223/253
14 Endress+Hauser
1
2
3
4
A0024640
 3 View into the field housing
1 Removable electronics box 2 Terminals 3 Partition plate 4 Fuse
4.2.2 Panel-mounted device
90 (3.54)
139 (5.47)
96 (3.78)
6 (0.24)
92 (3.62)
149.5 (5.89)
A0024641
 4 Panel-mounted device, dimensions in mm (inch)
Page 15
Liquisys M CLM223/253 Installation
Endress+Hauser 15
4.3 Installation instructions
4.3.1 Field device
There are several ways of securing the field housing:
• Wall mounting with fixing screws
• Post mounting to cylindrical pipes
• Post mounting to a square securing mast
NOTICE
Effect of climatic conditions (rain, snow, direct sunlight etc.)
Impaired operation to complete transmitter failure
‣
When installing outside, always use the weather protection cover (accessory).
Transmitter wall mounting
1
2
154 (6.06)
Ø 6 (0.24)
A0024638
 5 Field device wall mounting
1 Fixing bore holes 2 Plastic caps
Proceed as follows to mount the transmitter on the wall:
• Create the bore holes as shown in →  5.
• Drive two fixing screws through the fixing bore holes (1) from the front.
• Mount the transmitter on the wall as shown.
• Cover the bores with plastic caps (2).
Transmitter post mounting
You require a post mounting kit to secure the field device to horizontal and vertical posts or pipes (max. Ø 60 mm (2.36")). This can be acquired as an accessory (see the "Accessories" section).
Page 16
Installation Liquisys M CLM223/253
16 Endress+Hauser
1
2
3
1
1
3
Ø max. 60 (2.36)
A0024635
 6 Field device on horizontal or vertical pipes
1 Securing screws 2 Fixing screws 3 Securing plate
Proceed as follows to mount the transmitter on a post:
1. Guide the two securing screws (1) of the mounting kit through the openings on the securing plate (3).
2. Screw the securing plate onto the transmitter using the four fixing screws (2).
3. Secure the bracket with the field device on the post or pipe using the clip.
You can also secure the field device to the Flexdip CYH112 bracket in conjunction with the weather protection cover. These can be acquired as accessories, see the "Accessories" section.
A0027433
 7 Field device on Flexdip CYH112 bracket with weather protection cover
Page 17
Liquisys M CLM223/253 Installation
Endress+Hauser 17
4.3.2 Panel-mounted device
The panel-mounted device is secured with the tensioning screws supplied →  8
The necessary installation depth is approx. 165 mm (6.50").
1
2
3
3
3
92 (3.62
+0.5 +0.02
)
92 (3.62
+0.5 +0.02
)
90 (3.54)
57 (2.24)
33 (1.30)
~ 25 (0.98)
139 (5.47)
max. 45 (1.77)
6 (0.24)
96 (3.78)
~ 165 (6.50)*
92 (3.62)
A0024639
 8 Dimensions in mm (inch)
1 Mounting plate 2 Seal 3 Tensioning screws * Necessary installation depth
4.4 Post-installation check
• After installation, check the transmitter for damage.
• Check whether the transmitter is protected against moisture and direct sunlight (e.g. by the weather protection cover).
Page 18
Electrical connection Liquisys M CLM223/253
18 Endress+Hauser
5 Electrical connection
L
WARNING
Device is live
Incorrect connection may result in injury or death.
‣
The electrical connection may be performed only by an electrical technician.
‣
The electrical technician must have read and understood these Operating Instructions and must follow the instructions contained therein.
‣
Prior to commencing connection work, ensure that no voltage is present on any cable.
5.1 Wiring
L
WARNING
Risk of electric shock!
‣
At the supply point, the power supply must be isolated from dangerous live cables by double or reinforced insulation in the case of devices with a 24 V power supply.
NOTICE
The device does not have a power switch
‣
The customer must provide a protected circuit breaker in the vicinity of the device.
‣
The circuit breaker must be a switch or power switch, and you must label it as the circuit breaker for the device.
5.1.1 Wiring diagram
The wiring diagram shows the connections of a device equipped with all the options. The connection of the sensors to the various measuring cables is explained in more detail in the "Measuring cables and sensor connection" section.
Page 19
Liquisys M CLM223/253 Electrical connection
Endress+Hauser 19
11
12
13
47
48
49
57
58
59
15 V
10-50 V
81
82
10-50 V
93
94
41
42
43
Lf
mA
31
32
mA
33
34
51
52
53
54
55
56
S
84
83
15
16
84
PE
83
L+
L–N
L1
~ =
PE
23 +
24 -
mA
15
16
85+
86
_
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
A0008920
 9 Electrical connection of the transmitter
A Sensor (conductive) I Alarm (current-free contact position) B Sensor (inductive) J Relay 1 (current-free contact position) C Temperature sensor K Relay 2 (current-free contact position) D Signal output 1, conductivity L Relay 3 (current-free contact position) E Signal output 2, user-definable variable M Relay 4 (current-free contact position) F Binary input 1 (hold) N Current input 4 to 20 mA G Binary input 2 (Chemoclean) O Power connection H Auxiliary voltage output
Page 20
Electrical connection Liquisys M CLM223/253
20 Endress+Hauser
Please note the following:
• The device is approved for protection class II and is generally operated without a protective ground connection.
• To guarantee measuring stability and functional safety, you must connect the outer shield of the sensor cable: – Inductive sensors: terminal "S" – Conductive sensors: PE distributor rail
This is on the cover frame in the case of panel-mounted devices, and in the connection compartment in the case of field devices. Ground this PE distributor rail or the ground terminal directly on site wherever possible.
• Circuits "E" and "H" are not galvanically isolated from each other.
Field device connection
Guide the measuring cables through the PG glands into the housing. Connect the measuring cables in accordance with the terminal assignment.
42
+++++ -----
434154 55 56
NC
121311
31
83 84 15
32163433SNC 938685 94 81 82 NC
51 52 53 57 58 59 47 48 49
Lf
Temp.
(opt. )
+15V
10mA
WH
YE
GN
REL 4SENSO R REL 3 REL 2 REL 1
13109 5-4F
2
1
11 12
13
NC
Mains Hilfs energ ie
L1+N
-
(AC) (DC)
Conta cts: max. 2A AC: 250 V/500 VA DC: 30V /60W
+ -
23 24
4..20 mA
PE
1
Binär/
Binary
Binär/
Binary
A0008915
 10 Field device connection compartment sticker
1 PE distributor rail for CD/CS device version (conductive sensors)
Panel-mounted device connection
Page 21
Liquisys M CLM223/253 Electrical connection
Endress+Hauser 21
31 +
Senso r
32 -
33 +
34 -
85 +
86 -
93 +
94 - 42
81 + 43
82 - 41
Temp. (opt. )
+15V 10mA
2
1
NC
NC
NC
13108 1-4E
S
13
12
YE
GN
Mains
Contac ts: max. 2A, AC:2 50V/500VA, DC:3 0V/60W
L1 +
AC DC
N -
REL1
REL2
REL3
REL4
4..20 mA
24
23
49
COM
COM
COM
COM
-
+
59
54
55
56
52
51
53
58
57
48
47
13108 4-4C
15
11
WH
L1 +
AC DC
N -
1
Lf
84
16
83
Binary
Binary
A0008912
 11 Panel-mounted device connection sticker
1 Ground terminal
5.1.2 Measuring cables and sensor connection
You require shielded special measuring cables to connect conductivity sensors to the transmitter. The following multi-core, pre-terminated cable types can be used:
Sensor type Cable Extension
Two-electrode sensor with or without temperature sensor Pt 100
CYK71 CPK9* (for CLS16)
VBM box + CYK71 cable
Inductive sensor CLS50, CLS52 Fixed cable on sensor VBM box + CLK6 cable
* High-temperature version without PML
Maximum cable length
Conductive measurement of conductivity
Max. 100 m (328 ft) with CYK71
Resistance measurement Max 15 m (49.2 ft) with CYK71
Inductive measurement of conductivity Max. 55 m ( 180 ft) with CLK5 (incl. sensor cable)
Structure and termination of the measuring cables
Page 22
Electrical connection Liquisys M CLM223/253
22 Endress+Hauser
83
84
11
12
13
CYK71
CLK6
15
16
83
84
11
12
13
green
green
white
white
yellow
yellow
brown (nc)
brown (nc)
black
white red
semiconductor screen
semiconductor
screen
screen
screen
screen
A0008938-EN
 12 Structure of the special measuring cables
A Semi-conductor layer sc Shield
For further information on the cables and junction boxes, please refer to the "Accessories" section.
Field device measuring cable connection
Proceed as follows to connect a conductivity sensor to the field device:
1. Open the housing cover to access the terminal block in the connection compartment.
2. Break through the perforation for a cable gland, mount a cable gland and guide the cable through this gland.
3. Connect the cable in accordance with the terminal assignment (see connection compartment sticker).
4. Tighten the cable gland.
Panel-mounted device measuring cable connection
Page 23
Liquisys M CLM223/253 Electrical connection
Endress+Hauser 23
To connect a conductivity sensor, connect the measuring cable in accordance with the terminal assignment to the terminals on the rear of the device (see connection sticker).
GN WH YE
GN WH YE
BK
BK
CYK71/CPK9
83 16 15 84
23 1
11 12 13
PE
A0008919
 13 Connection of conductive sensors
83
11
S
12
161584
CLK6
13
GN
WH
YE
RD
WH
BU
RD
RD
BU
A0008918
 14 Connection of inductive sensors
Page 24
Electrical connection Liquisys M CLM223/253
24 Endress+Hauser
5.2 Alarm contact
41
42
43
41
42
43
A
B
A0006415
 15 Recommended fail-safe switching for the alarm contact
A Normal operating status B Alarm condition
Normal operating status Alarm condition
Device in operation and no error message present (alarm LED off):
• Relay energized
• Contact 42/43 closed
Error message present (alarm LED red) or device defective or de-energized (alarm LED off):
• Relay de-energized
• Contact 41/42 closed
5.3 Post-connection check
Carry out the following checks once you have made the electrical connection:
Device state and specifications Notes
Are the devices and cables free from damage on the outside? Visual inspection
Electrical connection Notes
Are the mounted cables strain relieved?
Are the connected cables provided with strain relief?
Is the cable run correct, without loops and cross-overs?
Are the power cable and signal cables connected correctly and in accordance with the wiring diagram?
Are all the screw terminals tightened?
Are all the cable entries fitted, tightened and leak-proof?
Are the PE distributor blocks grounded (if present)? Grounding is carried out at the point of
installation.
Page 25
Liquisys M CLM223/253 Operation options
Endress+Hauser 25
6 Operation options
6.1 Quick operation guide
You have the following ways of operating the transmitter:
• On site via the key field
• Via the HART interface (optional, with corresponding order version) with: – HART handheld terminal – PC with HART modem and the Fieldcare software package
• Via PROFIBUS PA/DP (optional, with corresponding order version) by PC with a corresponding interface and the Fieldcare software package or via a programmable logic controller (PLC).
For operation via HART or PROFIBUS PA/DP, please read the relevant sections in the additional Operating Instructions:
• PROFIBUS PA/DP, field communication for Liquisys M CXM223/253, BA00209C/07/EN
• HART, field communication for Liquisys M CXM223/253, BA00208C/07/EN
The following section only explains operation via the keys.
6.2 Display and operating elements
6.2.1 Display
LED displays
A0027220
Indicates the current operating mode, "Auto" (green LED) or "Manual" (yellow LED)
A0027222
Indicates the activated relay in the "Manual" mode (red LED) The status of relays 3 and 4 is indicated on the LC display.
A0027221
Indicates the working status of relay 1 and 2 LED green: measured value within the permitted limit, relay inactive LED red: measured value outside the permitted limit, relay active
A0027218
Alarm display, e.g. in event of continuous limit value overshoot, temperature sensor failure or system error (see error list)
Page 26
Operation options Liquisys M CLM223/253
26 Endress+Hauser
LC display
15
mS/cm R232
1
2 3
4
5
6
7
8
9
10
11
12
13
12
Setpoint
A0008922-EN
 16 Transmitter LC display
1 Indicator for measuring mode (normal operation) 2 Indicator for calibration mode 3 Indicator for setup mode (configuration) 4 Indicator for "Hold" mode (current outputs remain at last current state) 5 Indicator for receipt of a message on devices with communication 6 Indicator of working status of relays 3/4: inactive, active 7 Function code 8 In measuring mode: measured variable - in setup mode: configured variable 9 In measuring mode: secondary measured value - in setup/calibr. mode: e.g. set value 10 Indicator for autom. Temperature compensation 11 Indicator for man. Temperature compensation 12 "Error": error display 13 Sensor symbol (see the "Calibration" section)
Page 27
Liquisys M CLM223/253 Operation options
Endress+Hauser 27
6.2.2 Operating elements
The display shows the current measured value and the temperature simultaneously, which means you have an overview of the most important process data at once. Help text in the configuration menu helps users configure the device parameters.
1
7
6
4
5
3
2
A0024632-EN
 17 Operating elements
1 LC display for displaying the measured values and configuration data 2 Key to switch relays in manual mode and to display the active contact 3 LED for alarm function 4 Changeover switch for automatic/manual mode 5 LEDs for limit contactor relay (switch status) 6 Main operating keys for calibration and device configuration 7 Field for user-defined information
6.2.3 Key functions
A0027235
CAL key
When you press the CAL key, the device first prompts you for the calibration access code:
• Code 22 for calibration
• Code 0 or any other code for reading the last calibration data
Use the CAL key to accept the calibration data or to switch from field to field within the calibration menu.
A0027236
ENTER key
When you press the ENTER key, the device first prompts you for the setup mode access code:
• Code 22 for setup and configuration
• Code 0 or any other code for reading all the configuration data.
The ENTER key has several functions:
• Calls up the Setup menu from the measuring mode
• Saves (confirms) data entered in the setup mode
• Moves on within function groups
Page 28
Operation options Liquisys M CLM223/253
28 Endress+Hauser
A0027240
PLUS key and MINUS key
In the Setup mode, the PLUS and MINUS keys have the following functions:
• Selection of function groups. Press the MINUS key to select the function groups in the order given in the "System configuration" section.
• Configuration of parameters and numerical values
• Operation of the relays in manual mode
In the measuring mode, the following sequence of functions is accessed by repeatedly pressing the PLUS key:
• Temperature displayed in °F
• Temperature is hidden
• Current input signal in %
• Current input signal in mA
• Uncompensated conductivity value is displayed
• Return to basic settings
In the measuring mode, the following sequence of information is displayed by repeatedly pressing the MINUS key:
• The current errors are displayed consecutively (max. 10).
• Once all the errors have been displayed, the standard measurement display appears. In the function group F, an alarm can be defined separately for each error code.
A0027241
REL key
In the manual mode, you can use the REL key to switch between the relay and the manual start of cleaning. In automatic mode, you can use the REL key to read out the switch­on points (for limit contactor) or set points (for PID controller) assigned to the relay in question. Press the PLUS key to jump to the settings of the next relay. Use the REL key to get back to the display mode (automatic return after 30 s).
A0027234
AUTO key
Use the AUTO key to switch between automatic mode and manual mode.
A0027237
Escape function
If you press the PLUS and MINUS key simultaneously, you return to the main menu, or are taken to the end of calibration if calibrating. If you press the PLUS and MINUS key again, you return to the measuring mode.
A0027238
Locking the keyboard
Press the PLUS and ENTER key simultaneously for at least 3 s to lock the keyboard against any unauthorized data entry. All the settings can continue to be read. The code prompt displays the code 9999.
A0027239
Unlocking the keyboard
Press the CAL and MINUS key simultaneously for at least 3 s to unlock the keyboard. The code prompt displays the code 0.
Page 29
Liquisys M CLM223/253 Operation options
Endress+Hauser 29
6.3 Local operation
6.3.1 Automatic/manual mode
The transmitter normally operates in automatic mode. Here, the relays are triggered by the transmitter. In the manual mode, you can trigger the relays manually using the REL key or start the cleaning function.
How to change the operating mode:
A0027242
1. The transmitter is in automatic mode. The top LED (green) next to the AUTO key is lit.
A0027243
2. Press the AUTOMATIC key.
A0027240
3. To enable the manual mode, enter the code 22 via the PLUS and MINUS keys and press ENTER to confirm. The lower LED (manual mode) is lit.
A0027241
4. Select the relay or the function. You can use the REL key to switch between the relays. The relay selected and the switch status (ON/OFF) is displayed on the second line of the display. In the manual mode, the measured value is displayed continuously (e.g. for measured value monitoring for dosing functions).
A0027240
5. Switch the relay. The relay is switched on with PLUS and switched off with MINUS. The relay remains in this switched state until it is switched again.
A0027234
6. Press the AUTOMATIC key to return to the measuring mode, i.e. to the automatic mode. All the relays are triggered again by the transmitter.
• The operating mode remains in effect even after a power failure. The relays assume the quiescent state, however.
• The manual mode has priority over all other automatic functions.
• Hardware locking is not possible in the manual mode.
• The manual settings are kept until they are actively reset.
• Error code E102 is signaled during manual operation.
Page 30
Operation options Liquisys M CLM223/253
30 Endress+Hauser
6.3.2 Operating concept
Operating modes
Code
Measuring mode: standard mode of operation, displaying current measured values
Setup mode access to all configuration settings
Calibration mode: execution of calibration routine
The function groups are selected with the PLUS or MINUS key.
A0027244-EN
 18 Description of the possible operating modes
If no key is pressed in the setup mode for approx. 15 min, the device automatically returns to the measuring mode. Any active hold (hold during setup) is canceled.
Access codes
All device access codes are fixed and cannot be altered. When the device requests the access code, it distinguishes between different codes.
• CAL key + code 22: access to Calibration and Offset menu
• ENTER key + code 22: access to the menus for the parameters which make configuration and user-specific settings possible
• PLUS + ENTER keys simultaneously (min. 3 s): lock the keyboard
• CAL + MINUS keys simultaneously (min. 3 s): unlock the keyboard
• CAL or ENTER key + any code: access to read mode, i.e. all the settings can be read but not modified. The device continues measuring in the read mode. It does not shift to the "Hold" status. The current output and the controllers remain active.
Page 31
Liquisys M CLM223/253 Operation options
Endress+Hauser 31
Menu structure
The configuration and calibration functions are arranged in function groups.
• In the setup mode, select a function group with the PLUS and MINUS keys.
• In the function group itself, switch from function to function with the ENTER key.
• Within the function, select the desired option with the PLUS and MINUS keys or edit the settings with these keys. Then confirm with the ENTER key and continue.
• Press the PLUS and MINUS keys simultaneously (Escape function) to exit programming (return to the main menu).
• Press the PLUS and MINUS keys simultaneously again to switch to the measuring mode.
If a modified setting is not confirmed by pressing ENTER, the old setting is retained.
An overview of the menu structure is provided in the Appendix to these Operating Instructions.
E E
-
1
2
3
E
E
E
E
A0027245
 19 Menu structure
1 Functions (selection of parameters, entry of numbers) 2 Function groups, scroll backwards and forwards with the PLUS and MINUS keys 3 Switch from function to function with the ENTER key
Hold function: "freeze" the outputs
In both the setup mode and during calibration, the current output can be "frozen" (factory setting), i.e. it constantly retains its current status. "HOLD" appears on the display. If the controller actuating variable (steady control 4 to 20 mA) is output via current output 2, it is set to 0/4 mA during a hold.
• Hold settings can be found in the "Service" function group.
• During a hold, all contacts assume a quiescent state.
• An active hold has priority over all other automatic functions.
• With every hold, the I-component of the controller is set to "0".
• Any alarm delay is reset to "0".
• This function can also be activated externally via the hold input (see Wiring diagram; binary input 1).
• A manual hold (field S3) remains active even after a power failure.
Page 32
Commissioning Liquisys M CLM223/253
32 Endress+Hauser
7 Commissioning
7.1 Function check
L
WARNING
Incorrect connection, incorrect supply voltage
Safety risks for staff and device malfunctions
‣
Check that all connections have been established correctly in accordance with the wiring diagram.
‣
Ensure that the supply voltage matches the voltage indicated on the nameplate.
7.2 Switching on
Familiarize yourself with the operation of the transmitter before it is first switched on. In particular please read the "Basic safety instructions" and "Operation options" sections. After power-up, the device performs a self-test and then goes to the measuring mode.
Now calibrate the sensor in accordance with the instructions in the "Calibration" section.
During initial commissioning, the sensor must be calibrated so that the measuring system can return precise measurement data.
Then perform the first configuration in accordance with the instructions in the "Quick setup" section. The values set by the user are kept even in the event of a power failure.
The following function groups are available in the transmitter (the groups that are only available in the Plus Package are marked accordingly in the functional description):
Setup mode
• SETUP 1 (A)
• SETUP 2 (B)
• CURRENT INPUT (Z)
• CURRENT OUTPUT (O)
• ALARM (F)
• CHECK (P)
• RELAY (R)
• TEMPERATURE COMPENSATION (T)
• CONCENTRATION MEASUREMENT (K)
• SERVICE (S)
• E+H SERVICE (E)
• INTERFACE (I)
Calibration mode
CALIBRATION (C)
A detailed explanation of the function groups available in the transmitter can be found in the "Device configuration" section.
s F3
Err.delay
Function display: The displayed code indicates the function position in the function group.
Additional information
2
A0025560-EN
 20 Information for the user on the display
Page 33
Liquisys M CLM223/253 Commissioning
Endress+Hauser 33
C C1 C111
C121
C131
C132 C133
A0027502
 21 Function code
To make it easier for you to select and find function groups and functions, a code for the corresponding field is displayed for each function →  20 The structure of this code is illustrated in →  21. The function groups are indicated as letters in the first column (see the names of the function groups). The functions of the individual groups are displayed incrementally by row and by column.
Factory settings
The first time the device is switched on, the factory setting is set for all the functions. The table below provides an overview of the most important settings.
All other factory settings can be found in the description of the individual function groups in the "System configuration" section (the factory setting is highlighted in bold).
Function Factory setting
Type of measurement Conductive measurement of conductivity,
Temperature measurement in °C
Type of temperature compensation Linear with reference temperature 25 °C (77 °F)
Temperature compensation Automatic (ATC on)
Limit value for controller 1 9999 mS/cm
Limit value for controller 2 9999 mS/cm
Hold Active during configuration and calibration
Measuring range 0 μS/cm to 2000 mS/cm (no measuring ranges to be
configured). The setting is flowing and depends on the sensors connected.
Current outputs 1 and 2* 4 to 20 mA
Current output 1: measured value for 4 mA signal current
0 μS/cm
Current output 1: measured value for 20 mA signal current
2000 mS/cm
Current output 2: temperature value for 4 mA signal current*
-35.0 °C (-31 °F)
Current output 2: temperature value for 20 mA signal current*
250.0 °C (482 °F)
* with appropriate version
Page 34
Commissioning Liquisys M CLM223/253
34 Endress+Hauser
7.3 Quick Setup
After power-up, you must make some settings to configure the most important functions of the transmitter which are required for correct measurement. The following section gives an example of this.
User entry Range of adjustment
(factory settings in bold)
Display
1. Press the ENTER key
2. Enter the code 22 to open access to the menus. Press the ENTER key.
3. Press the MINUS key until you get to the "Service" function group.
S
SERVICE
A0008408-EN
4. Press ENTER to be able to make your settings.
5. Select your language in S1, e.g. "ENG" for English. Press ENTER to confirm your entry.
ENG = English
GER = German FRA = French ITA = Italian NEL = Dutch ESP = Spanish
S1
Language
ENG
A0008409-EN
6. Press the PLUS and MINUS key simultaneously to exit the "Service" function group.
7. Press the MINUS key until you get to the "Setup 1" function group.
A
SETUP 1
A0007824-EN
8. Press ENTER to be able to make your settings for "Setup 1".
9. In A1, select the desired mode of operation, e.g. "cond" = conductive. Press ENTER to confirm your entry.
cond = conductive
ind = inductive MOhm = resistance Conc = concentration
A1
Oper.Mode
cond
A0009002-EN
10. In A2, press ENTER to accept the factory setting. (Only if A1 = conc, otherwise continue with Step 12)
%
ppm
mg/l TDS = total dissolved solids None
A2
Conc.Unit
%
A0009003-EN
11. In A3, press ENTER to accept the standard setting.
XX.xx
X.xxx XXX.x XXXX
A3
Format
XX.xx
A0009004-EN
12. In A4, press ENTER to accept the standard setting.
auto, μS/cm, mS/cm, S/cm, μS/m, mS/m, S/ m
A4
Unit
auto
A0009005-EN
Page 35
Liquisys M CLM223/253 Commissioning
Endress+Hauser 35
User entry Range of adjustment
(factory settings in bold)
Display
13. In A5, enter the exact cell constant of the sensor. The cell constant is provided on the sensor quality certificate.
cond: 1.000 cm
-1
ind: 1.98 cm
-1
MOhm: 0.01 cm
-1
0.0025 to 99.99 cm
-1
A5
1/cm
Cellconst
1.000
A0009006-EN
14. In A6, enter the cable resistance (only applies to conductive sensors).
0 Ω
0 to 99.99 Ω
A60
@
Cable-Res
A0009007-EN
15. In A7, enter the measured value damping. Measured value damping causes the measured value to be averaged over the specified number of individual measured values (if A7 = 1, no damping takes place). Press ENTER to confirm your entry. The display returns to the initial display of the "Setup 1" function group.
1
1 to 60
A7
Damping
1
A0001960-EN
16. Press the MINUS key until you get to the "Setup 2" function group. Press ENTER to make your settings for "Setup 2".
B
SETUP 2
A0007830-EN
17. In B1, select the temperature sensor. Press ENTER to confirm your entry.
Pt100
Pt1k = Pt 1000 NTC30 Fixed
B1
ProcTemp.
Pt100
A0009010-EN
18. In B2, select the appropriate type of temperature compensation for your process, e.g. "lin" = linear. Press ENTER to confirm your entry. Detailed information is provided in the "Setup 2" section.
None
Lin = linear
NaCl = table salt (IEC
746) Pure = ultrapure water NaCl PureH = ultrapure water HCl Tab = table
B2
TempComp.
lin
A0009011-EN
19. In B3, enter the temperature coefficient α. Press ENTER to confirm your entry.
2.1 %/K
0.0 to 20.0 %/K
B3
%/K
Alpha val
2.10
A0009012-EN
20. The current temperature is displayed in B5. If necessary, adjust the temperature sensor to an external measurement. Press ENTER to confirm your entry.
Actual value displayed and entered
-35.0 to 250.0 °C
B5
¡C
RealTemp.
0.0
A0009014-EN
Page 36
Commissioning Liquisys M CLM223/253
36 Endress+Hauser
User entry Range of adjustment
(factory settings in bold)
Display
21. The difference between the measured and entered temperature is displayed. Press the ENTER key. The display returns to the initial display of the "Setup 2" function group.
0.0 °C
-5.0 to 5.0 °C
B6
¡C
TempOffs.
0.0
A0009015-EN
22. Press the MINUS key to get to the "Current output" function group. Press ENTER to make your settings for the current outputs.
O
OUTPUT
A0025026-EN
23. In O1, select your current output, e.g. "Out 1" = output 1. Press ENTER to confirm your entry.
Out 1
Out 2
O1
Sel.Out
Out1
A0025027-EN
24. In O3, select the linear characteristic. Press ENTER to confirm your entry.
Lin = linear (1)
Lin = linear (1) Tab = table
O3
Sel.Type
lin
A0025029-EN
25. In O311, select the current range for your current output, e.g. 4 to 20 mA. Press ENTER to confirm your entry.
4 to 20mA
0 to 20 mA
O311
Sel.Range
4-20
A0025030-EN
26. In O312, specify the conductivity at which the minimum current value is applied at the transmitter output, e.g. 0 μS/cm. Press ENTER to confirm your entry.
Cond/ind: 0.00 μS/cm MOhm: 0.00 kΩ·cm Conc: 0.00 % Temp: 0.00 °C
O312
mg/l
0/4 mA
0.00
A0025031-EN
27. In O313, specify the conductivity at which the maximum current value is applied at the transmitter output, e.g. 2000 mS/cm. Press ENTER to confirm your entry. The display returns to the initial display of the "Current output" function group.
Cond/ind: 2000 mS/cm MOhm: 500 kΩ·cm Conc: 99.99 % Temp: 150 °C
O313
mg/l
20 mA
10.00
A0025032-EN
28. Press PLUS and MINUS simultaneously to switch to the measurement mode.
You must perform an air set before installing the inductive sensor. See the "Calibration" section for more information.
Page 37
Liquisys M CLM223/253 Commissioning
Endress+Hauser 37
7.4 Device configuration
7.4.1 Setup 1 (conductivity)
In the SETUP 1 function group, you can change the settings for the measuring mode and the sensor.
Functions marked in italics are not supported by the basic device version.
Coding Field Range of
adjustment (factory settings in bold)
Display Info
A
SETUP 1 function group
A
SETUP 1
A0007824-EN
Configuration of basic functions
A1 Select the
operating mode auswählen
cond = conductive
ind = inductive MOhm = resistance
conc= concentration
A1
Oper.Mode
cond
A0009002-EN
Display varies depending on the device:
• Cond/resistance/conc
• Ind/conc
When the operating mode is changed, all user settings are automatically reset.
A2 Select the
concentration unit to be displayed (only for Plus Package)
%
ppm mg/l TDS = total dissolved solids None
A2
Conc.Unit
%
A0009003-EN
A2 is only active if A1 = conc
A3 Select the display
format for the concentration unit (only for Plus Package)
XX.xx
X.xxx XXX.x XXXX
A3
Format
XX.xx
A0009004-EN
A3 is only active if A1 = conc
A4 Select the unit to
be displayed
auto, μS/cm, mS/cm, S/cm, μS/ m, mS/m, S/m, kΩ·cm, MΩ·cm, kΩ·m
A4
Unit
auto
A0009005-EN
If "auto" is selected, the highest possible resolution is automatically selected. A4 is not active if A1 = conc
A5 Enter the cell
constant for the connected sensor
cond:1.000 cm
-1
ind:1.98 cm
-1
MOhm:0.01 cm
-1
0.0025 to 99.99 cm
-1
A5
1/cm
Cellconst
1.000
A0009006-EN
The exact cell constant is provided on the sensor quality certificate.
Page 38
Commissioning Liquisys M CLM223/253
38 Endress+Hauser
Coding Field Range of
adjustment (factory settings in bold)
Display Info
A6 Enter the cable
resistance
0 Ω
0 to 99.99 Ω
A60
@
Cable-Res
A0009007-EN
Only for conductive sensors. The standardized line resistance must be multiplied by the actual cable length. CYK71: 0.165 Ω/m
A7 Enter the value
for measured value damping
1
1 to 60
A7
Damping
1
A0009008-EN
Measured value damping causes averaging over the specified number of individual measured values. This is used, for example, to stabilize the display if the measurement is unstable. There is no damping if “1” is entered.
Page 39
Liquisys M CLM223/253 Commissioning
Endress+Hauser 39
7.4.2 Setup 2 (temperature)
Temperature coefficient α indicates the change in the conductivity per degree of temperature change:
k(T) k(T )
0
(1+
=
(T - T ))
0
a
.
.
A0009163
Where
κ(T) = conductivity at process temperature T
κ(T0) = conductivity at reference temperature T
0
The temperature coefficient depends both on the chemical composition of the solution and the temperature itself.
To determine the level of dependency, four different types of compensation can be selected in the transmitter:
• Linear temperature compensation
• NaCl compensation
• NaCl ultrapure water compensation (neutral compensation)
• HCl ultrapure water compensation (acid compensation)
• Temperature compensation with table
Linear temperature compensation
The change between two temperature points is taken to be constant, i.e. α = const. The α value can be edited for linear compensation. The reference temperature can be edited in the B7 field. The default setting is 25 °C.
1
a
T
k*
A0008945
 22 Linear temperature compensation
* Uncompensated conductivity
NaCl compensationIn the case of NaCl compensation (as per IEC 60746), a fixed non-linear curve specifying the relationship between the temperature coefficient and temperature is saved in the device. This curve applies to low concentrations of up to approx. 5 % NaCl.
0
48
96
144
2.1
2.3
2.5
2.7
T [°C]
a [%/K]
A0008939
 23 NaCl compensation
Page 40
Commissioning Liquisys M CLM223/253
40 Endress+Hauser
Ultrapure water compensation (for conductive sensors)
For pure and ultrapure water, algorithms are saved in the transmitter that consider the self-dissociation of ultra pure water and its strong temperature dependency. They are used for conductivity levels of approx. 100 μS/cm.
Two types of compensation are available:
• NaCl ultrapure water compensation: this is optimized for pH-neutral contamination.
• HCl ultrapure water compensation: this is optimized for measuring the acid conductivity downstream of a cation exchanger. It is also suitable for ammonia (NH3) and caustic soda (NaOH).
• The ultrapure water compensation types always refer to a reference temperature of 25 °C (77 °F).
• The lowest conductivity displayed is the theoretical limit value of ultrapure water at 25 °C (77 °F) of 0.055 μS/cm.
Temperature compensation with table
For devices with the Plus Package, it is possible to enter a table with temperature coefficients α as a function of the temperature. The following conductivity data of the medium under measurement are required to use the alpha table function for temperature compensation:
Value pairs comprising the temperature T and conductivity κ with:
• κ(T0) for the reference temperature T
0
• κ(T) for the temperatures that occur in the process
T
1
T
0
T
2
T T
k
2
k
1
k
k( )
T
0
T1T
2
a
1
a
2
a
A
B
A0008944
 24 Determination of the temperature coefficient
A Required data B Calculated α values
Use the following formula to calculate the α values for the temperatures that are relevant in your process:
a =
100%
k(T )
0
k k(T) - (T )
0
T - T
0
.
; T T¹
0
A0009162
Enter the α-T value pairs obtained in this way in the T4 and T5 fields of the ALPHA TABLE function group.
Use this function group to change the settings for temperature measurement.
Page 41
Liquisys M CLM223/253 Commissioning
Endress+Hauser 41
Functions marked in italics are not supported by the basic device version.
Coding Field Range of
adjustment (factory settings in bold)
Display Info
B
SETUP 2 function group
B
SETUP 2
A0007830-EN
Settings for temperature measurement
B1 Select the
temperature sensor
Pt100
Pt1k = Pt 1000 NTC30 Fixed
B1
ProcTemp.
Pt100
A0009010-EN
"fixed": Manual temperature compensation (MTC), no temperature measurement. Instead a fixed temperature value is defined in B4.
B2 Select the type of
temperature compensation
None
Lin = linear
NaCl = table salt (IEC 746) Pure = ultrapure water NaCl PureH = ultrapure water HCl
Tab = Table
B2
TempComp.
lin
A0009011-EN
This option does not appear for concentration measurement. "Pure" and "PureH" are only displayed for conductive devices.
B3 Enter temperature
coefficient α
2.10 %/K
0.00 to 20.00 %/K
B3
%/K
Alpha val
2.10
A0009012-EN
Only if B2 = lin. With other settings in B2, field B3 does not have any effect.
B4 Enter the process
temperature
25.0 °C
-35.0 to 250.0 °C
B4
¡C
ProcTemp.
25.0
A0009013-EN
Only if B1 = fixed. The value entered can only be in °C.
B5 Display the
temperature and adjust the temperature sensor
Actual value displayed and entered
-35.0 to 250.0 °C
B5
¡C
RealTemp.
0.0
A0009014-EN
With the value entered here, the temperature sensor can be adjusted to an external measurement. Affects B6. Omitted if B1 = fixed.
B6 Enter the
temperature differential (offset)
Current offset
-5.0 to 5.0 °C
B6
¡C
TempOffs.
0.0
A0009015-EN
The offset is the difference between the actual value entered and the measured temperature. Omitted if B1 = fixed.
B7 Enter reference
temperature
25.0 °C
-5.0 to 100 °C
B7
C
°
ref temp.
25.0
A0009016-EN
Page 42
Commissioning Liquisys M CLM223/253
42 Endress+Hauser
7.4.3 Current input
For the "Current input" function group, you require a relay card with a current input which is not available in the basic device version. With this function group, you can monitor process parameters and use them for feedforward control. For this purpose, you must connect the current output of an external measured variable (e.g. flowmeter) to the 4 to 20mA input of the transmitter. The following assignment applies:
Flow in main stream Current signal in mA Current input signal in %
Flowmeter start of measuring range 4 0
Flowmeter end of measuring range 20 100
Flow monitoring in the main stream
This arrangement is particularly practical if the sample flow through a flow assembly in an open outlet is independent of the flow in the main stream.
This permits signaling of an alarm condition in the main stream (flow too low or has completely stopped) and triggers dosing switch-off even if the medium flow is maintained due to the method of installation.
A
E
D
F
C
B
1
0
0
0
1
1
Z2 Z3
A0008923
 25 Alarm signaling and dosing switch-off by the main stream
A Flow in main stream F Flow restoration
B Relay contacts of PID controller Z2 Delay for controller switch-off, see field Z2
C Alarm relay Z3 Delay for controller switch-on, see field Z3
D Flow below switch-off limit Z 4 or flow failure 0 Off
E Flow alarm 1 On
Feedforward control to PID controller
You can optimize control on control systems with very short response times by measuring the medium flow rate in addition to the oxygen content. Then apply this flow rate value (4 to 20 mA) as feedforward control to the PID controller.
Feedforward control is a multiplying function as illustrated in the figure below (example with factory setting):
Page 43
Liquisys M CLM223/253 Commissioning
Endress+Hauser 43
1
0 20 40 60 80 100Z7
1.5
0.5
Y
X
A0008942
 26 Multiplying feedforward control
Y Gain K
infl
X Current input signal in [%]
Page 44
Commissioning Liquisys M CLM223/253
44 Endress+Hauser
Functions marked in italics are not supported by the basic device version.
Coding Field Range of
adjustment (factory settings in bold)
Display Info
Z
CURRENT INPUT function group
Z
CUR.INPUT
A0024903-EN
Current input settings
Z1 Select flow
monitoring of main stream (with controller switch­off)
Off
On
Z1
Cont.stop
Off
A0024904-EN
Flow monitoring may only be switched on if the flowmeter is connected in the main stream. If Z1 = off, fields Z2 to Z5 are not available.
Z2 Enter the delay for
controller switch­off through current input
0 s
0 to 2000 s
Z2
s
Off Delay
0
A0024905-EN
Brief flow shortfalls can be suppressed by this delay and do not result in controller switch­off.
Z3 Enter the delay for
controller switch­on through current input
0 s
0 to 2000 s
Z3
s
On Delay
0
A0024934-EN
In the case of a controller, a delay until the reception of a representative measured value is recommended after a long period without flow.
Z4 Enter the switch-off
limit value for the current input
50 %
0 to 100 %
Z4
%
A.Thresh
50
A0024935-EN
0 to 100% corresponds to 4 to 20 mA at the current input. Observe measured value assignment to the current output of the flowmeter.
Z5 Enter the switch-off
direction for the current input
Low
High
Z5
Stop Dir
Low
A0024939-EN
The controller is switched off if the value entered in Z4 is undershot or overshot.
Z6 Select feedforward
control to PID controller
Off
Lin = linear Basic
Z6
PID influ
Off
A0024940-EN
If Z6 = off, the field Z7 is not available. Z6 = basic: disturbance variable only affects the basic load (alternatively dosing in proportion to quantity, if usual PID controller not possible, e.g. defective sensor).
Z7 Enter value for
feedforward control at which gain = 1 applies
50 %
0 to 100 %
Z7
%
Kinflu=1
50
A0024941-EN
When the value is set, the controller actuating variable is the same size when feedforward control is switched on as when feedforward control is switched off.
Page 45
Liquisys M CLM223/253 Commissioning
Endress+Hauser 45
7.4.4 Current outputs
Use the "Current output" function group to configure the individual outputs. You can enter either a linear characteristic (O3 (1)) or a user-defined current output characteristic in conjunction with the Plus Package (O3 (3)). Exception: if you have chosen a "continuous controller" for current output 2, you cannot enter a user-defined current output characteristic for this current output.
In addition, you can also simulate a current output value (O3 (2)) to check the current outputs.
If a second current output is present, you can output the controller actuating variable via the current output in accordance with field R237/O2.
1500
1000
2000
500
0 4 8 12 16 20
I [mA]
D I [mA]
D Signal
[mS/cm]c
A0008943
 27 User-defined current output characteristic (example)
The current output characteristic must increase or decrease very monotonically.
The distance per mA between two table value pairs must be greater than:
• Conductivity: 0.5 % of measuring range
• Temperature: 0.25 °C
The values for the sample characteristic →  27 are entered in the following table. The distance per mA is calculated from Δ signal / Δ mA.
Current output 1 Current output 2
Value pair [mS/cm]
[%] [°C]
Current [mA]
Distance per mA
[mS/cm] [%] [°C]
Current [mA]
Distance per mA
1 500 4
2 1000 16 41.66
3 2000 20 250
Page 46
Commissioning Liquisys M CLM223/253
46 Endress+Hauser
First enter the desired current output configuration into the following blank table with a pencil. Calculate the resulting signal distance per mA to observe the necessary minimum slope. Then enter the values in the device.
Current output 1 Current output 2
Value pair [mS/cm]
[%] [°C]
Current [mA]
Distance per mA
[mS/cm] [%] [°C]
Current [mA]
Distance per mA
1
2
3
4
5
6
7
8
9
10
Functions marked in italics are not supported by the basic device version.
Coding Field Range of
adjustment (factory settings in bold)
Display Info
O
CURRENT OUTPUT function group
O
OUTPUT
A0025026-EN
Configuration of the current output (does not apply for PROFIBUS).
O1 Select current
output
Out 1
Out 2
O1
Sel.Out
Out1
A0025027-EN
A characteristic can be selected for every output.
O2 Select measured
variable for 2nd current output
°C
mS/cm, MΩ, %
Contr
O2
Parameter
¡C
A0025028-EN
R237=curr (current output 2) can only be selected if O2=Contr (controller) is selected (relay card required).
O3 (1) Enter the
characteristic type
Lin = linear (1)
Sim = simulation (2)
Tab = table (3)
O3
Sel.Type
lin
A0025029-EN
The characteristic can have a positive or negative slope for the measured value output. In the case of actuating variable output (O2 = Contr), an increasing current corresponds to an increasing actuating variable.
Page 47
Liquisys M CLM223/253 Commissioning
Endress+Hauser 47
Coding Field Range of
adjustment (factory settings in bold)
Display Info
O311 Select current
range
4 to 20mA
0 to 20 mA
O311
Sel.Range
4-20
A0025030-EN
O312 0/4 mA value:
Enter the associated measured value
Cond/ind: 0.00
μS/cm
MOhm:
0.00kΩ*cm
Conc: 0.00 % Temp: 0.00 °C
µS/cm
O312
0/4 mA
0.00
A0025034-EN
Here you can enter the measured value at which the min. current value (0/4 mA) is applied at the transmitter output (not for controller). (Turndown see Technical data.)
O313 20 mA value:
Enter the associated measured value
Cond/ind: 2000
μS/cm MOhm: 500 kΩ*cm
Conc: 99.99 % Temp: 150 °C
mS/cm
O313
20 mA
2000
A0025035-EN
Here you can enter the measured value at which the max. current value (20 mA) is applied at the transmitter output (not for controller). (Turndown see Technical data.)
O3 (2) Simulate current
output
Lin = linear (1)
Sim = simulation (2)
Tab = table (3)
O3
Sel.Type
sim
A0025039-EN
Simulation is not ended until O3(1) or O3(3) is selected. For further characteristics, see O3 (1), O3 (3).
O321 Enter simulation
value
Current value
0.00 to 22.00 mA
O321
mA
Simulat.
10.20
A0025040-EN
Entering a current value results in this value being directly output at the current output.
O3 (3) Enter current
output table
Lin = linear (1) Sim = simulation (2)
Tab = table (3)
O3
Sel.Type
table
A0025041-EN
Only for Plus Package Values can also be subsequently added or modified. The values entered are automatically sorted by increasing current value. For further characteristics, see O3 (1), O3 (2).
O331 Select table
option
Read
Edit
O331
Sel.Table
read
A0025042-EN
O332 Enter number of
table value pairs11 to 10
O332
No.Elem.
1
A0025043-EN
Enter the number of pairs of x and y values (measured value and current value) here.
Page 48
Commissioning Liquisys M CLM223/253
48 Endress+Hauser
Coding Field Range of
adjustment (factory settings in bold)
Display Info
O333 Select table value
pair
1
1 to no. elem. Assign
O333
Sel.Elem.
1
A0025044-EN
The system runs through the O333 to O335 function chain as often as indicated in O332. "Assign" appears as the last step. The display goes to O336 after confirmation.
O334 Enter x value Cond/ind: 0.00
μS/cm
MOhm:
0.00kΩ*cm
Conc: 0.00 % Temp: 0.00 °C
O334
µS/cm
Meas.val.
0.00
A0025045-EN
x value = measured value specified by user.
O335 Enter y value 0.00 mA
0.00 to 20.00 mA
O335
mA
mA value
0.00
A0025048-EN
y value = user-defined current value pertaining to O334. Return to O333 until all values are entered.
O336 Message as to
whether table status is OK
Yes
No
O336
Status ok
yes
A0025049-EN
Back to O3. If status = no, correct the table (all settings made up until now remain intact) or go back to the measuring mode (table is deleted).
Page 49
Liquisys M CLM223/253 Commissioning
Endress+Hauser 49
7.4.5 Alarm
You can use the "Alarm" function group to define various alarms and configure output contacts.
Each individual error can be defined to be effective or not (at the contact or as an error current).
Functions marked in italics are not supported by the basic device version.
Coding Field Range of
adjustment (factory settings in bold)
Display Info
F
ALARM function group
F
ALARM
A0025141-EN
Alarm function settings.
F1 Select contact
type
Latch = latching contact
Momen = momentary contact
F1
Cont.Type
Latch
A0025142-EN
The option selected only applies for the fault-signaling contact, not for the error current.
F2 Select the time
unit for the alarm delay
s
min
F2
Time Unit
s
A0025143-EN
F3 Enter alarm
delay
0 s (min)
0 to 2000 s (min)
F3
s
Err.Delay
0
A0025144-EN
Depending on the option selected in F2, the alarm delay can be entered in s or min.
F4 Select error
current
22 mA
2.4 mA
F4
Err.Curr
22mA
A0025145-EN
If "0-20 mA" was selected in O311, "2.4 mA" may not be used.
F5 Select the error
number
1
1 to 255
F5
Sel.error
1
A0025146-EN
Here you can select all the errors which should trigger an alarm. The errors are selected by the error numbers. Please refer to the table in the "System error messages" section for the meaning of the individual error numbers. The factory settings remain in effect for all errors that are not edited.
F6 Set alarm
contact to be effective for the selected error
Yes
No
F6
Rel.Assg
yes
A0025147-EN
If “no” is selected, all the other alarm settings are deactivated (e.g. alarm delay). The settings themselves are maintained. This setting only applies to the error currently selected in F5.
Page 50
Commissioning Liquisys M CLM223/253
50 Endress+Hauser
Coding Field Range of
adjustment (factory settings in bold)
Display Info
F7 Set error
current to be effective for the selected error
No
Yes
F7
Curr.Assg
no
A0025148-EN
The option selected in F4 is effective or not effective in the event of an error. This setting only applies to the error currently selected in F5.
F8 Automatic
cleaning function start
No
Yes
F8
CleanTrig
no
A0025149-EN
This field is not available for certain errors, see the "Troubleshooting and fault elimination" section.
F9 Select return to
menu or next error
Next = next error number ←R
F9
Select
next
A0025150-EN
If ←R is selected, you return to F. If Next is selected, you go to F5.
7.4.6 Check
The CHECK function group is only available for devices with a Plus Package.
In the CHECK function group, you can select different monitoring functions for the measurement.
Polarization detection (field P1)
In the case of conductive sensors, the effects of polarization in the boundary layer between the sensor and medium restrict the measuring range. The transmitter can detect polarization effects by using a smart signal analysis process. Error message E071 is displayed.
Alarm threshold monitoring (fields P2 to P5)
You can use this function to monitor the measured value for permissible upper and lower limits and trigger an alarm (error messages E154, E155).
PCS alarm (process check system), (fields P6 to P9)
AC (alternation check; sensor activity monitoring): The function AC (field P6) is used to
check measuring signals for deviations. An alarm (E152) is triggered if the measuring signal remains constant within the space of one hour. The reason for such sensor behavior can be contamination, cable open circuit or similar.
CC (controller check): You can monitor the controller activity with the CC function . This function is primarily suited to batch operation and single-sided limit switches. A malfunction of the controller is detected and reported thanks to freely adjustable monitoring times (E156, E157).
Page 51
Liquisys M CLM223/253 Commissioning
Endress+Hauser 51
A
t
A0025196
 28 PCS alarm (live check)
A Constant measuring signal = alarm triggered after PCS alarm time has elapsed
Any PCS alarm pending is automatically deleted as soon as the sensor signal changes.
Page 52
Commissioning Liquisys M CLM223/253
52 Endress+Hauser
You can use the "Check" function group to monitor the measured value for permissible upper and lower limits and trigger an alarm.
Functions marked in italics are not supported by the basic device version.
Coding Field Range of
adjustment (factory settings in bold)
Display Info
P
CHECK function group
P
CHECK
A0009045-EN
Settings for sensor and process monitoring
P1 Switch
polarization detection on or off (only for conductive sensors)
Off
On
P1
Pol.Detec
Off
A0009046-EN
Polarization only occurs with conductive sensors. Any polarization that may occur is detected but is not compensated for. (Error no.: E071)
P2 Select alarm
threshold monitoring
Off
Low High LoHi = low and high Lo! Hi! LoHi!
P2
A.Tresh
Off
A0009048-EN
Alarm possible with or without controller switch-off. xxxx = without controller switch-off xxxx! = with controller switch-off (Error no.: E154, E155)
P3 Enter alarm delay 0 min (s)
0 to 2000 min (s)
P3
s
Err.Delay
0
A0009049-EN
Depending on the option selected in F2, the alarm delay can be entered in s or min. This delay must first elapse before undershooting/ overshooting in accordance with fields P4/P5 results in an alarm.
P4 Enter lower alarm
threshold
0 µS/cm
0 to 9999 mS/cm
P4
µS/cm
LowAlarm
0.00
A0009050-EN
P5 Enter upper alarm
threshold
9999 µS/cm
0 to 9999 mS/cm
P5
mS/cm
HighAlarm
9999
A0009051-EN
P6 Select process
monitoring
Off
AC CC AC+CC AC! CC! AC+CC!
P6
ProcMonit
Off
A0009052-EN
AC = sensor activity monitoring (E152) CC = controller monitoring (E156, E157) Alarm possible with or without simultaneous controller switch-off. xxxx = without controller switch-off xxxx! = with controller switch-off
Page 53
Liquisys M CLM223/253 Commissioning
Endress+Hauser 53
Coding Field Range of
adjustment (factory settings in bold)
Display Info
P7 Enter maximum
permissible duration for lower CC setpoint limit violation (field P9)
60 min
0 to 2000 min
P7
min
Tmax Low
60
A0009053-EN
Only if P6=CC or AC CC.
P8 Enter maximum
permissible duration for upper CC setpoint limit violation (field P9)
120 min
0 to 2000 min
P8
min
Tmax High
120
A0009054-EN
Only if P6=CC or AC CC.
P9 Enter CC set point
(for P7/P8)
1000 µS/cm
0 to 9999 mS/cm
P9
µS/cm
Setpoint
1000
A0009055-EN
Set value is an absolute value. This function is primarily suited to batch operation and single-sided limit switches.
7.4.7 Relay configuration
For the "RELAY" function group, you require a relay card which is not available in the basic device version.
The following relay contacts can be selected and configured as desired (max. four contacts, depending on options installed):
• Limit contactor for conductivity measured value: R2 (1)
• Limit contactor for temperature: R2 (2)
• PID controller: R2 (3)
• Timer for cleaning function: R2 (4)
• Chemoclean function: R2 (5)
• USP/EP: R2 (6) and R2 (7) (for Plus Package, only conductive sensors)
Limit contactor for conductivity measured value and temperature
The transmitter has different ways of assigning a relay contact. Switch-on and switch-off points and pickup and dropout delays can be assigned to the limit contactor. In addition, you can configure an alarm threshold to output an error message and to start a cleaning function in conjunction with this.
These functions can be used both for the primary value and for temperature measurement.
Please refer to the switch states in →  29 for a clear illustration of the relay contact states.
• When the measured values increase (maximum function), the relay contact is closed as of t2 after the switch-on point (t1) has been exceeded and the pickup delay has elapsed (t2-t1). The alarm contact switches if the alarm threshold (t3) is reached and the alarm delay (t4-t3) has also elapsed (errors E067 to E070).
• When the measured values decrease, the alarm contact is reset when the value falls below the alarm threshold (t5) again, as is the relay contact (t7) after the dropout delay (t7-t6).
• If the pickup and dropout delays are set to 0 s, the switch-on and switch-off points are also switch points of the contacts.
The same settings can also be made for a minimum function in the same way as for the maximum function.
Page 54
Commissioning Liquisys M CLM223/253
54 Endress+Hauser
t
1
t
1
t
2
t
2
t
3
t
3
t
4
t
4
t
5
t
5
t
6
t
6
t
7
t
7
t
t
B
1
2 3
3 2
1
4
5
6
7
4
5
6
7
A
A0025215
 29 Illustration of the alarm and limit value functions
A Switch-on point > switch-off point: Max. function B Switch-on point < switch-off point: Min. function 1 Alarm threshold 2 Switch-on point 3 Switch-off value 4 Contact ON 5 Alarm ON 6 Alarm OFF 7 Contact OFF
Page 55
Liquisys M CLM223/253 Commissioning
Endress+Hauser 55
P(ID) controller
You can define various controller functions for the transmitter. On the basis of the PID controller, P, PI, PD and PID controllers can be implemented. For an optimum control system, use the controller that best suits your application.
• P controller Used for simple linear control purposes with small system deviations. Where major changes are to be controlled, overshooting may occur. In addition, a lasting control deviation must be expected.
• PI controller Is used for control systems where overshooting is to be avoided and no lasting control deviation should occur.
• PD controller Is used for processes that require quick changes and where peaks must be corrected.
• PID controller Is used for processes where a P, PI or PD controller does not control sufficiently.
Configuration options of the P(ID) controller
The following configuration options are available for a PID controller:
• Change control gain Kp (P influence)
• Set integral action time Tn (I influence)
• Set derivative action time Tv (D influence)
Basic load dosing (basic)
You can set a constant dosage amount (field R2311) with basic load dosing (field R231).
PID control plus basic load dosing
If you have selected this function (PID + Basic) in field R231, the PID-controlled dosage amount does not drop below the basic load value entered in field R2311.
100 %
100 %
A
B
C
A0025221
 30 Control characteristic of PID controller with basic load dosing
A PID plus basic load B Basic load C PID
Commissioning
If you do not yet have any experience for setting the control parameters, set the values that yield the greatest possible stability in the control circuit. Proceed as follows to optimize the control circuit further:
• Increase the control gain Kp until the controlled variable just starts to overshoot.
• Reduce Kp slightly again and then reduce the integral action time Tn so that the shortest possible correction time without overshooting is achieved.
• To reduce the response time of the controller, also set the derivative action time Tv.
Control and fine optimization of the set parameters with a recorder
Page 56
Commissioning Liquisys M CLM223/253
56 Endress+Hauser
A
B
A
B
A
B
A
B
A
B
1
2
3
4
5
A0025218
 31 Optimization of settings Tn and K
p
A Actual value B Time 1 Tn too small
2 Tn too large 3 Kp too large 4 Kp too small 5 Optimum setting
Actuating signal outputs via contacts (R237 to R2310)
Each control contact outputs a cyclical signal whose intensity corresponds to the controller's actuating value. A distinction is made according to the type of signal cycle:
• Pulse length modulation The bigger the calculated manipulated variable is, the longer the contact affected remains picked up. The period T can be set to be between 0.5 and 99 s (field R238). Outputs with pulse length modulation are used to activate solenoid valves.
• Pulse frequency modulation The bigger the calculated manipulated variable is, the higher the switching frequency of the contact affected. The maximum switching frequency 1/T can be set to be between 60 and 180 min-1 (field R239). The on-time ton is constant. It depends on the set maximum frequency and is approx. 0.5 s for 60 min-1 and approx. 170 ms for 180 min-1. 170 ms. Outputs with pulse frequency modulation are used to activate directly controlled solenoid dosing pumps.
Page 57
Liquisys M CLM223/253 Commissioning
Endress+Hauser 57
A
1
0
T
t [s]
A
t [s]
t
1
t
2
t
1
1
0
T
1
T
2
A0025219
 32 Signal of a pulse-length modulated controller contact (left) and of a pulse frequency-modulated controller
contact (right)
Contact: 1 = on, 0 = off T Period Time (s): t1 = ton t2 = t
off
T1 T2 Impulse period length ( 1/T1 and 1/T2)
Constant controller
The controller can also control the second analog current output (if provided). This is configured in fields R237 and O2.
Control characteristic for direct and inverse control action
You can choose between two control characteristics in the R236 field:
• Direct control action = maximum function
• Inverse control action = minimum function
X
W
X
W
– X
W
– X
W
100 %
100 %
Y
Y
A
B
A0025222
 33 Control characteristic of a proportional controller with direct and inverse control action
A Direct = maximum function B Inverse = minimum function XW Control deviation Y Current output signal = controller actuating variable
Timer for cleaning function
This function includes a simple cleaning option. You can set the time interval after which cleaning should start. Therefore you can only select a constant interval sequence.
Other cleaning functions are available for selection in conjunction with the Chemoclean function (device version with four contacts required, see the "Chemoclean function" section).
The timer and Chemoclean do not work independently of one another. While one of the two functions is active, the other cannot be started.
Page 58
Commissioning Liquisys M CLM223/253
58 Endress+Hauser
A
B
1
0
t
1
t
2
1
0
t
3
t
t
t
4
t
0
A0025223
 34 Connection between cleaning time, pause time and hold dwell period
A Wiper and/or spray cleaning system B Hold function 0 Inactive 1 active t0 Normal operation t1 cleaning start t2-t1Cleaning time t3-t2Clean hold dwell period (0 to 999 s) t4-t3Pause time between two cleaning intervals (1 to 7200 min)
Chemoclean function
Just like the timer function, Chemoclean can also be used to start a cleaning cycle. However, Chemoclean also gives you the added option of defining various cleaning and rinsing intervals and of dosing cleaning agent.
Therefore, it is possible to clean irregularly with different repeat cycles and to separately set the cleaning times with post rinse times.
Please note the following:
• To use the Chemoclean function the transmitter has to be equipped with a designated relay board (see product structure or chapter "accessories").
• The timer and Chemoclean are mutually dependent. While one of the two functions is active, the other cannot be started.
• For the Chemoclean function, the relays 3 (water) and 4 (cleaner) are used.
• If the cleaning is prematurely aborted, a post rinse time always follows.
• If the setting is "Economy", cleaning only takes place with water.
Page 59
Liquisys M CLM223/253 Commissioning
Endress+Hauser 59
A
B
C
t
0
t
1
t
2
t
3
t
4
t
5
0
0
0
1
1
1
A0025216
 35 Sequence of a cleaning cycle
A Hold function B Water valve is triggered C Cleaning valve is triggered 0 Contact off 1 Contact on t0 Normal operation t1 cleaning start t2-t1Pre-rinse time t3-t2Cleaning time t4-t3Post rinse time t5-t4Hold dwell period
Limit values for pharmaceutical water according to United States Pharmacopoeia (USP) and European Pharmacopoeia (EP) (only for conductive sensors)
In the case of conductive sensors, the transmitter has functions for monitoring "Water for Injection" (WFI), "Highly Purified Water" (HPW) and "Purified Water" (PW) according to the United States Pharmacopoeia (USP) Part 645 and European Pharmacopoeia (EP) standards.
USP function: The temperature-dependent limit values in the following table apply for "Water for Injection" (WFI) according to USP and EP and for "Highly Purified Water" (HPW) according to EP. The table is programmed into the transmitter.
Temperature [°C] Conductivity [µS/cm] Temperature [°C] Conductivity [µS/cm]
0 0.6 55 2.1
5 0.8 60 2.2
10 0.9 65 2.4
15 1.0 70 2.5
20 1.1 75 2.7
25 1.3 80 2.7
30 1.4 85 2.7
35 1.5 90 2.7
40 1.7 95 2.9
45 1.8 100 3.1
50 1.9
Page 60
Commissioning Liquisys M CLM223/253
60 Endress+Hauser
The measurement is performed in the following steps:
• The transmitter determines the uncompensated conductivity and the water temperature.
• The transmitter rounds the temperature off to the nearest 5 °C and compares the measured conductivity with the associated value in the table.
• If the measured value is larger than the value in the table, an alarm is triggered (E151).
EP-PW function: The following table lists the temperature-dependent limit values for "Purified Water" (PW) in accordance with EP; this table is also programmed into the transmitter.
Temperature [°C] Conductivity [µS/cm] Temperature [°C] Conductivity [µS/cm]
0 2.4 60 8.1
10 3.6 70 9.1
20 4.3 75 9.7
25 5.1 80 9.7
30 5.4 90 9.7
40 6.5 100 10.2
50 7.1
The measurement is performed in the following steps:
• The transmitter determines the uncompensated conductivity and the water temperature.
• If the temperature is between two table entries, the limit value for the conductivity is determined by interpolation of the two neighboring points.
• If the measured value is larger than the limit value, an alarm is triggered (E151).
Early warning: A USP early warning is also available. This is activated at a configurable switch-on point (e.g. 80 % of USP/EP limit value). This means that the user is alerted to be able to regenerate the plant in time. Der Voralarm wird in Feld R262 bzw. R272 eingestellt.
Please note the following:
• The device must be equipped with a relay card and the Plus Package to be able to use the USP and EP function.
• To output an alarm, enable the fault-signaling contact or the error current in field F5 ­F7 (error code E151 and E153).
• The switch-off point of the pre-alarm is 1 % below of the switch-on point (R262 / R272), related to the main limit value.
• The transmitter also uses the uncompensated values for the USP and EP function when it displays temperature-compensated values.
• The limit value of 100 °C (212 °F) is used for temperatures over 100 °C (212 °F).
Page 61
Liquisys M CLM223/253 Commissioning
Endress+Hauser 61
Functions marked in italics are not supported by the basic device version.
Coding Field Range of
adjustment (factory settings in bold)
Display Info
R
RELAY
R
RELAIS
A0009058-EN
Relay contact settings
R1 Select the contact
to be configured
Rel1
Rel2 Rel3 Rel4
R1
Sel.Relay
Rel1
A0009059-EN
Rel3 (water) and Rel4 (cleaner) are only available with the relevant version of the transmitter. If Chemoclean is used as the cleaning method, Rel4 is not available.
R2 (1) Configure limit
contactor for conductivity, resistance or concentration measurement
LC PV = limit contactor conductivity (1)
LC °C = limit contactor T (2) PID controller (3) Timer (4)
Clean= Chemoclean (5) USP (6) EP PV(7)
R2
Sel.Type
LC PV
A0009060-EN
PV = process value If Rel4 is selected in the R1 field, Clean = Chemoclean cannot be selected. By confirming with ENTER, another relay function already switched on is switched off and its settings are reset to the factory settings.
R211 Switch function
of R2 (1) off or on
Off
On
R211
Function
off
A0009067-EN
All the settings are retained.
R212 Enter the switch-
on point of the contact
Cond/ind: 9999
mS/cm MOhm: 200 MΩ·cm Conc: 9999 %
mS/cm
R212
On value
9999
A0009068-EN
Never set the switch-on point and the switch-off point to the same value! (Only the operating mode selected in A1 is displayed.)
R213 Enter the switch-
off point of the contact
Cond/ind: 9999
mS/cm MOhm: 200 MΩ·cm Conc: 9999 %
mS/cm R213
Off value
9999
A0009069-EN
Entering a switch-off point selects either a Max contact (switch-off point < switch-on point) or a Min contact (switch-off point > switch-on point), thereby implementing a hysteresis that is constantly required (see "Illustration of the alarm and limit functions" figure).
R214 Enter pickup
delay
0 s
0 to 2000 s
s R214
On Delay
0
A0009070-EN
Page 62
Commissioning Liquisys M CLM223/253
62 Endress+Hauser
Coding Field Range of
adjustment (factory settings in bold)
Display Info
R215 Enter dropout
delay
0 s
0 to 2000 s
s R215
Off Delay
0
A0009071-EN
R216 Enter alarm
threshold
Cond/ind: 9999
mS/cm MOhm: 200 MΩ·cm Conc: 9999 %
mS/cm
R216
A.Thresh
9999
A0009072-EN
If the alarm threshold is undershot/overshot, this triggers an alarm with the error message (E067 to E070) and an error current at the transmitter (note alarm delay in field F3). If defined as a Min contact, the alarm threshold must be < switch-off point.
R217 Display status for
limit contactor
MAX
MIN
R217
LC State
MAX
A0009073-EN
Display only
R2 (2) Configure limit
contactor for temperature measurement
LC PV = limit contactor conductivity (1)
LC °C = limit contactor T (2)
PID controller (3) Timer (4)
Clean= Chemoclean (5) USP (6) EP PV(7)
R2
Sel.Type
LC ¡C
A0009061-EN
By confirming with ENTER, another relay function already switched on is switched off and its settings are reset to the factory settings.
R221 Switch function
of R2 (2) off or on
Off
On
R221
Function
off
A0009074-EN
R222 Enter switch-on
temperature
250.0 °C
-35.0 to 250.0 °C
R222
C
°
On value
250.0
A0009075-EN
Never set the switch-on point and the switch-off point to the same value!
R223 Enter switch-off
temperature
250.0 °C
-35.0 to 250.0 °C
R223
C
°
Off value
250.0
A0009076-EN
Entering a switch-off point selects either a Max contact (switch-off point < switch-on point) or a Min contact (switch-off point > switch-on point), thereby implementing a hysteresis that is constantly required (see "Illustration of the alarm and limit functions" figure).
Page 63
Liquisys M CLM223/253 Commissioning
Endress+Hauser 63
Coding Field Range of
adjustment (factory settings in bold)
Display Info
R224 Enter pickup
delay
0 s
0 to 2000 s
R224
s
On Delay
0
A0009077-EN
R225 Enter dropout
delay
0 s
0 to 2000 s
R225
s
Off Delay
0
A0009078-EN
R226 Enter alarm
threshold (as absolute value)
250.0 °C
-35.0 to 250 °C
R226
C
°
A.Thresh
250.0
A0009079-EN
If the alarm threshold is undershot/overshot, this triggers an alarm with the error message (E067 to E070) and an error current at the transmitter (note alarm delay in field F3). If defined as a Min contact, the alarm threshold must be < switch-off point.
R227 Display status for
limit contactor
MAX
MIN
R227
LC State
MAX
A0009080-EN
Display only
R2 (3) Configure P(ID)
controller
LC PV = limit contactor conductivity (1) LC °C = limit contactor T (2)
PID controller (3)
Timer (4)
Clean= Chemoclean (5) USP (6) EP PV(7)
R2
Sel.Type
PID
A0009062-EN
By confirming with ENTER, another relay function already switched on is switched off and its settings are reset to the factory settings.
R231 Switch function
of R2 (3) off or on
Off
On Basic PID+B
R231
Function
off
A0009081-EN
On = PID controller Basic = basic load dosing PID+B = PID controller + basic load dosing
R232 Enter set point Cond/ind: 0.00
μS/cm MOhm: 0.00 kΩ·cm Conc: 0.00 %
µS/cm R232
Setpoint
0.00
A0009084-EN
The set point is the value to be maintained by the control system. Using this control process, this value is restored when an upwards or downwards deviation occurs.
Page 64
Commissioning Liquisys M CLM223/253
64 Endress+Hauser
Coding Field Range of
adjustment (factory settings in bold)
Display Info
R233 Enter control
gain K
P
1.00
0.01 to 20.00
R233
Kp
1.00
A0009085-EN
See the "P(ID) controller" section.
R234 Enter integral
action time T
n
(0.0 = no I­component)
0.0 min
0.0 to 999.9 min
R234
min
Time Tn
0.0
A0009086-EN
See the "P(ID) controller" section. With every hold, the I­component is set to zero. Although the hold can be deactivated in field S2, this does not apply for Chemoclean and timer!
R235 Enter derivative
action time T
v
(0.0 = no D­component)
0.0 min
0.0 to 999.9 min
R235
min
Time Tv
0.0
A0009087-EN
See the "P(ID) controller" section.
R236 Select controller
characteristic
Dir = direct
Inv = inverse
R236
Direction
dir
A0009088-EN
The setting is required depending on the control deviation (upward or downward deviation, see the "P(ID) controller" section).
R237 Select pulse
length or pulse frequency
Len = pulse length Freq = pulse frequency
Curr = current output 2
R237
Oper.Mode
len
A0009089-EN
Pulse length e.g. for solenoid valve, pulse frequency e.g. for solenoid dosing pump, see the "Actuating signal outputs" section. Curr = current output 2 can only be selected if O2 = Contr.
R238 Enter pulse
interval
10.0 s
0.5 to 999.9 s
R238
s
PulsePer.
10.0
A0009090-EN
This field only appears if pulse length is selected in R237. If pulse frequency is selected, R238 is skipped and entries continue with R239.
R239 Enter maximum
pulse frequency of the adjuster
120 min
-1
60 to 180 min
-1
R239
1/min
Max.PFreq
120
A0009091-EN
This field only appears if pulse frequency is selected in R237. If pulse length is selected, R239 is skipped and entries continue with R2310.
R2310 Enter minimum
switch-on time t
ON
0.3 s
0.1 to 5.0 s
R2310
s
Min.PTime
0.3
A0009082-EN
This field only appears if pulse length is selected in R237.
Page 65
Liquisys M CLM223/253 Commissioning
Endress+Hauser 65
Coding Field Range of
adjustment (factory settings in bold)
Display Info
R2311 Enter basic load 0 %
0 to 40 %
R2311
%
BasicLoad
0
A0009083-EN
When you select the basic load, you enter the desired dosing quantity.
100% basic load would correspond to:
• Constantly on if R237 = len
• Fmax if R237 = freq (field R239)
• 20 mA if R237 = curr
R2 (4) Configure
cleaning function (timer)
LC PV = limit contactor conductivity (1) LC °C = limit contactor T (2) PID controller (3)
Timer (4)
Clean= Chemoclean (5) USP (6) EP PV(7)
R2
Sel.Type
Timer
A0009063-EN
Only one cleaning agent (generally water) is used for the cleaning. By confirming with ENTER, another relay function already switched on is switched off and its settings are reset to the factory settings.
R241 Switch function
of R2 (4) off or on
Off
On
R241
Function
off
A0009092-EN
R242 Enter rinsing/
cleaning time
30 s
0 to 999 s
R242
s
RinseTime
30
A0009093-EN
Settings for hold and relay are active for this time.
R243 Enter pause time 360 min
1 to 7200 min
R243
min
PauseTime
360
A0009094-EN
The pause time is the time between two cleaning cycles (see the "Timer for cleaning function" section).
R244 Enter minimum
pause time
120 min
1 to R243
R244
min
Min.Pause
120
A0009095-EN
The minimum pause time prevents constant cleaning if a cleaning trigger is pending.
R2 (5) Configure
cleaning with Chemoclean (for version with four contacts, Chemoclean option and contacts 3 and 4 assigned)
LC PV = limit contactor conductivity (1) LC °C = limit contactor T (2) PID controller (3) Timer (4)
Clean= Chemoclean (5)
USP (6) EP PV(7)
R2
Sel.Type
Clean
A0009064-EN
See the "Chemoclean function" section. By confirming with ENTER, another relay function already switched on is switched off and its settings are reset to the factory settings.
Page 66
Commissioning Liquisys M CLM223/253
66 Endress+Hauser
Coding Field Range of
adjustment (factory settings in bold)
Display Info
R251 Switch function
of R2 (5) off or on
Off
On
R251
Function
off
A0009096-EN
R252 Select type of
start pulse
Int = internal (time­controlled)
Ext = external (digital input 2) I+ext = internal + external I+stp = internal, suppressed by external
R252
CleanTrig
int
A0009097-EN
The cycle for the "int" function is started when the pause time ends (R257). No real time clock is available. External suppression is required for irregular time intervals (e.g. weekends).
R253 Enter pre-rinse
time
20 s
0 to 999 s
R253
s
PreRinse
20
A0009098-EN
Rinsing is performed with water.
R254 Enter cleaning
time
10 s
0 to 999 s
R254
s
CleanTime
10
A0009099-EN
Cleaning is performed with cleaning agent and water.
R255 Enter post rinse
time
20 s
0 to 999 s
R255
s
PostRinse
20
A0009100-EN
Rinsing is performed with water.
R256 Enter number of
repeat cycles
0
0 to 5
R256
Rep.Rate
0
A0009101-EN
R253 to R255 is repeated.
R257 Enter pause time 360 min
1 to 7200 min
R257
min
PauseTime
360
A0009102-EN
The pause time is the time between two cleaning cycles (see the "Chemoclean function" section).
R258 Enter minimum
pause time
120 min
1 to R257
R258
min
Min.Pause
120
A0009103-EN
The minimum pause time prevents constant cleaning if an external cleaning start is pending.
Page 67
Liquisys M CLM223/253 Commissioning
Endress+Hauser 67
Coding Field Range of
adjustment (factory settings in bold)
Display Info
R259 Enter number of
cleaning cycles without cleaning agent (economy function)
0
0 to 9
R259
EconomyCl
0
A0009104-EN
After cleaning with cleaner, up to 9 cleaning sessions can be carried out with water only until the next cleaning session with cleaner takes place.
R2 (6) Configure USP
contact (only for Plus Package with relay card)
LC PV = limit contactor conductivity (1) LC °C = limit contactor T (2) PID controller (3) Timer (4) Clean= Chemoclean (5)
USP (6)
EP PV(7)
R2
Sel.Type
USP
A0009065-EN
The USP contact can be configured as an early warning alarm, i.e. it alerts the user before the actual limit value is reached. Error no. E151 appears in the event of an alarm. WFI according to USP; HPW according to EP
R261 Switch function
of R2 (6) off or on
Off
On
R261
Function
Off
A0009105-EN
R262 Early warning
threshold: enter switch-on point
80.0 %
0.0 to 100.0 %
R262
%
On Value
80.0
A0009106-EN
The early warning triggers the contact. When the alarm value is reached (100%), the alarm relay is also triggered. Example: Given 15 °C and 1.0 μS/cm, a USP early warning alarm is triggered at 0.8 μS/cm when the setting is 80%.
R264 Early warning
threshold: enter pickup delay
0 s
0 to 2000 s
R264
s
On Delay
0
A0009107-EN
R265 Early warning
threshold: enter dropout delay
0 s
0 to 2000 s
R265
s
Off Delay
0
A0009108-EN
R2 (7) Configure EP PV
contact (only for Plus Package with relay card)
LC PV = limit contactor conductivity (1) LC °C = limit contactor T (2) PID controller (3) Timer (4) Clean= Chemoclean (5) USP (6)
EP PV(7)
R2
Sel.Type
EP PW
A0009066-EN
The EP PV contact can be configured as an early warning alarm, i.e. it alerts the user before the actual limit value is reached. Error no. E151 appears in the event of an alarm. PV according to EP
Page 68
Commissioning Liquisys M CLM223/253
68 Endress+Hauser
Coding Field Range of
adjustment (factory settings in bold)
Display Info
R271 Switch function
of R2 (7) off or on
Off
On
R271
Function
Off
A0009109-EN
R272 Early warning
threshold: enter switch-on point
80.0 %
0.0 to 100.0 %
R272
%
On Value
80.0
A0009110-EN
The early warning triggers the contact. When the alarm value is reached (100%), the alarm relay is also triggered. Example: Given 15 °C and 1.0 μS/cm, an EP PV early warning alarm is triggered at 0.8 μS/cm when the setting is 80%.
R274 Early warning
threshold: enter pickup delay
0 s
0 to 2000 s
R274
s
On Delay
0
A0009111-EN
R275 Early warning
threshold: enter dropout delay
0 s
0 to 2000 s
R275
s
Off Delay
0
A0009112-EN
Page 69
Liquisys M CLM223/253 Commissioning
Endress+Hauser 69
7.4.8 Temperature compensation with table
The "ALPHA TABLE" function group is not available in the basic device version. You require the Plus Package to access this function group. With this function group, you can perform temperature compensation using a table (field B2).
Enter the α-T value pairs (see SETUP 2) in the T4 and T5 fields.
Functions marked in italics are not supported by the basic device version.
Coding Field Range of
adjustment (factory settings in bold)
Display Info
T
ALPHA TABLE function group
T
ALPHA TAB
A0009123-EN
Settings for temperature compensation.
T1 Select table
option
Read
Edit
T1
Sel.Table
read
A0009125-EN
T2 Enter number of
table value pairs11 to 10
T2
No. Elem.
1
A0009126-EN
In the α-table, you can enter a maximum of 10 value pairs. These pairs are saved under numbers 1 to 10 and can be changed individually or consecutively.
T3 Select table
value pair
1
1 to number of table value pairs Assign
T3
Sel.Elem.
1
A0009127-EN
The system automatically runs through the T3 to T5 function chain as often as corresponds to the value indicated in T2. "Assign" appears as the last value. The display goes to T6 once the process is finished.
T4 Enter
temperature value
0.0 °C
-35.0 to 250.0 °C
T4
C
°
Temp.val.
0.0
A0009128-EN
The temperature values must be at least 1 K apart. Factory setting for the temperature value in the table value pairs:
0.0 °C; 10.0 °C; 20.0 °C; 30.0 °C ...
T5 Enter
temperature coefficient α
2.10 %/K
0.00 to 20.00 %/K
T5
%/K
alpha val
2.10
A0009129-EN
T6 Message as to
whether table status is OK
Yes
No
T6
Status ok
yes
A0009130-EN
Display only If status = "no", you must set correct values in the table (all settings made up until now remain intact) or go back to the measuring mode (then the table is not valid).
Page 70
Commissioning Liquisys M CLM223/253
70 Endress+Hauser
7.4.9 Concentration Measurement
The CONCENTRATION function group is only available for devices with a Plus Package.
The transmitter can convert from conductivity values to concentration values. For this purpose, the operating mode is first set to concentration measurement (see field A1).
Afterwards, in the measuring device you must specify the basic data on the basis of which the concentration is to be calculated. You require the conductivity characteristic curves of the medium for this.
In the case of conductive sensors, the effects of polarization in the boundary layer between the sensor and medium restrict the measuring range. The transmitter can detect polarization effects by using a smart signal analysis process. Error message E071 is displayed.
You can either refer to your data sheets for these characteristic curves or determine the characteristic curves yourself.
1. Create samples of the medium in the concentrations that occur in the process.
2. Then measure the uncompensated conductivity of these samples at temperatures that also occur in the process.
• For variable process temperatures:
If variable process temperatures are to be taken into consideration, you must measure the conductivity for at least two temperatures (ideally the minimum and maximum process temperature) for the samples created. The temperature values of the various samples must be identical in each case. The temperatures must be at least 0.5 °C apart. At least two samples of different concentrations, taken at two different temperatures in each case, are required as the transmitter needs at least four points in the table (this must include the minimum and maximum concentration values).
• For constant process temperatures:
Measure the differently concentrated samples at this temperature. At least two samples are required.
At the end you should have measuring data which resemble those shown in the following figures.
k
k
C CC
min
C
min
T
1
T
1
T
2
T
2
T
3
T
3
C
max
C
max
2
2
1
1
A0008926
 36 Example of measured data for variable temperatures
κ Conductivity C Concentration T Temperature 1 Measuring point 2 Measuring range
Page 71
Liquisys M CLM223/253 Commissioning
Endress+Hauser 71
k
C
1
T
C
min
C
max
k
C
1
T
C
min
C
max
A0008925
 37 Example of measured data for constant temperatures
κ Conductivity C Concentration T Constant temperature 1 Measuring range
The characteristic curves received from the measuring points must increase or decrease very monotonically in the range of the process conditions, i.e. neither maximum points, nor minimum points, nor ranges with a constant behavior can occur. The curve profiles opposite are therefore not permitted.
k
C
C
k
A0008927
 38 Impermissible curve profiles
κ Conductivity C Concentration
Value entry
Enter the three characteristic values for each measured sample in the fields K6 to K8 (value triplets of uncompensated conductivity, temperature and concentration).
• Variable process temperature: Enter at least the four value triplets needed.
• Constant process temperature: Enter at least the two value triplets needed.
Please note the following: If the measured values of the process are outside the range of your sample values, this considerably reduces the level of accuracy and the error message E078 or E079 will be displayed. Therefore, take the limit values of your process into consideration when determining the characteristic curves. Wird bei aufsteigender If you enter an additional value triplet of 0 μS/cm and 0% for every temperature used when the characteristic curve is increasing, you can work from the start of the measuring range with sufficient accuracy and without an error message. Enter the values in order of increasing concentration (see the following example).
mS/cm % °C
240 96 60
380 96 90
220 97 60
340 97 90
120 99 60
200 99 90
Page 72
Commissioning Liquisys M CLM223/253
72 Endress+Hauser
As the temperature is already processed in the concentration tables, the settings for temperature compensation in the Setup 2 menu (fields B2 and B3) are ineffectual for concentration measurement.
Functions marked in italics are not supported by the basic device version.
Coding Field Range of
adjustment (factory settings in bold)
Display Info
K
CONCENTRATION function group
K
CONCENTRA
A0009113-EN
Four different concentration fields can be entered in this function group.
K1 Select the active
concentration curve on which the calculation of the display value is based
1
1 to 4
K1
act.curve
1
A0009114-EN
The curves are independent of one another. Four different curves can be defined.
K2 Select the curve to be
edited
1
1 to 4
K2
editCurve
1
A0009115-EN
If a curve is being edited, another curve should be used to calculate the current display values. Example: If curve 2 is being edited, either curve 1, curve 3 or curve 4 should be active (see K1).
K3 Select table option Read
Edit
K3
Table
read
A0009116-EN
This option is valid for all concentration curves.
K4 Enter number of
reference triplets
1
1 to 10
K4
No. Elem.
1
A0009117-EN
Each triplet consists of three numeric values.
K5 Select triplet 1
1 to number of triplets in K4 Assign
K5
Sel.Elem.
1
A0009118-EN
Any triplet can be edited. "Assign" takes the user to K9
K6 Enter uncompensated
conductivity value
0.0 mS/cm
0.0 to 9999 mS/cm
K6
mS/cm
conduct.
0.0
A0009119-EN
The system automatically runs through the K5 to K6 function chain as often as corresponds to the value indicated in K4. The display then goes to K9.
Page 73
Liquisys M CLM223/253 Commissioning
Endress+Hauser 73
Coding Field Range of
adjustment (factory settings in bold)
Display Info
K7 Enter concentration
value for K6
0.00 %
0.00 to 99.99 %
K7
%
concentr.
0.00
A0009120-EN
Engineering unit as selected in A2. Format as selected in A3.
K8 Enter temperature value
for K6
0.0 °C
-35.0 to 250.0 °C
K8
C
°
Temp.val.
0.0
A0009121-EN
K9 Message as to whether
table status is OK
Yes
No
K9
Status ok
yes
A0009122-EN
Read only. If "No", then set correct values in the table (all settings made up until now remain intact) or go back to the measuring mode (then the table is not valid).
7.4.10 Service
Coding Field Range of
adjustment (factory settings in bold)
Display Info
S
SERVICE function group
S
SERVICE
A0008408-EN
Service function settings.
S1 Select language ENG = English
GER = German FRA = French ITA = Italian NL = Dutch ESP = Spanish
S1
Language
ENG
A0008409-EN
The option selected only applies for the fault­signaling contact, not for the error current.
S2 Configure a hold S+C = hold during
configuration and calibration
Cal = hold during calibration Setup = hold during configuration None = no hold
S2
Auto HOLD
S+C
A0008413-EN
S = setup C = calibration
S3 Manual hold Off
On
S3
Man. HOLD
off
A0008414-EN
The setting is retained even in the event of a power failure.
Page 74
Commissioning Liquisys M CLM223/253
74 Endress+Hauser
Coding Field Range of
adjustment (factory settings in bold)
Display Info
S4 Enter hold dwell
period
10 s
0 to 999 s
S4
s
Cont.Time
10
A0008415-EN
S5 Enter SW
upgrade release code (Plus Package)
0000
0000 to 9999
S5
0000
PlusCode
A0008416-EN
The code can be found on the nameplate. If an incorrect code is entered, you are taken back to the measurement menu. The number is edited with the PLUS or MINUS key and confirmed with the ENTER key. "1" is displayed if the code is active.
S6 Enter SW
upgrade release code for Chemoclean
0000
0000 to 9999
S6
CleanCode
0000
A0008417-EN
The code can be found on the nameplate. If an incorrect code is entered, you are taken back to the measurement menu. The number is edited with the PLUS or MINUS key and confirmed with the ENTER key. "1" is displayed if the code is active.
S7 Order number is
displayed
S7
PR0005
order
A0008418-EN
If the device is upgraded, the order code is changed automatically.
S8 Serial number is
displayed
S8
12345678
SerNo
A0008420-EN
S9 Reset the device
to the basic settings
No
Sens = sensor data Facty = factory settings
S9
S.Default
no
A0008421-EN
Sens = last calibration is deleted and is reset to factory setting. Facty = all data (apart from A1 and S1) are deleted and reset to the factory setting!
S10 Perform device
test
No
Displ = display test
S10
Test
no
A0008410-EN
Page 75
Liquisys M CLM223/253 Commissioning
Endress+Hauser 75
7.4.11 E+H Service
Coding Field Range of adjustment
(factory settings in bold)
Display Info
E
E+H SERVICE function group
E
E+H SERV
A0007857-EN
Information on the device version
E1 Select module Contr = controller
(central module) (1)
Trans = transmitter (2) Main = power unit (3) Rel = relay module (4) Sens = sensor (5)
E1
Select
Contr
A0007858-EN
E111 E121 E131 E141 E151
Software version is displayed
E111
SW-Vers.
xx.xx
A0007859-EN
If E1 = contr: device software If E1 = trans, main, rel: module firmware If E1 = sens: sensor software
E112 E122 E132 E142 E152
Hardware version is displayed
E112
HW-Vers.
xx.xx
A0007861-EN
Info display
E113 E123 E133 E143 E153
Serial number is displayed
E113
12345678
SerNo
A0007860-EN
Info display
E114 E124 E134 E144 E154
Module ID is displayed
E114
Modul-ID
LSG
A0007862-EN
Info display
Page 76
Commissioning Liquisys M CLM223/253
76 Endress+Hauser
7.4.12 Interfaces
Coding Field Range of
adjustment (factory settings in bold)
Display Info
I
INTERFACE function group
I
INTERFACE
A0007863-EN
Communication settings (only for HART or PROFIBUS device version).
I1 Enter bus
address
Address HART: 0 to 15 or PROFIBUS: 0 to
126
I1
Address
126126
A0007864-EN
Each address may only be assigned once in a network. If a device address ≠ 0 is selected for a HART device, the current output is automatically set to 4 mA and the device prepares for multi-drop operation.
I2 The tag name is
displayed
I2
@@@@@@@@
Tag
A0007865-EN
7.4.13 Communication
For devices with a communication interface, please also refer to the separate Operating Instructions BA00208C/07/EN (HART®) or BA00209C/07/DE (PROFIBUS®).
Page 77
Liquisys M CLM223/253 Commissioning
Endress+Hauser 77
7.5 Calibration
Use the CAL key to access the calibration function group.
Use this function group to calibrate and adjust the transmitter. The calibration can be performed in two different ways:
• By measuring in a calibration solution of known conductivity.
• By entering the exact cell constant of the conductivity sensor.
Please note the following:
• During initial commissioning of inductive sensors, an air set is absolutely essential for residual coupling compensation (from field C111) so that the measuring system can return precise measurement data.
• If the calibration procedure is aborted by pressing the PLUS and MINUS keys at the same time (return to C114, C126 or C136) or if the calibration is faulty, then the previous calibration data are reinstated. A calibration error is indicated by “ERR” and the sensor symbol flashes on the display. Repeat calibration!
• For each calibration, the device automatically switches to hold (factory setting).
• On completion of the calibration, the device returns to the measurement mode. The "hold" symbol appears on the display during the hold dwell period (field S4).
• Only items C121 to C126 are relevant for conductive sensors.
Coding Field Range of
adjustment (factory settings in bold)
Display Info
C
CALIBRATION function group:
C
CALIBRAT
A0009141-EN
Airs and InstF do not apply for conductive measurement.
C1(1) Calibration of
inductive sensors with annular opening
Airs = air set (1)
Cellc = cell constant (2) InstF = installation factor (3)
C1
Calibrat
AirS
A0009142-EN
When commissioning inductive sensors, an air set is mandatory. The sensor air set must be performed in air and in a dry state.
Take the inductive sensor out of the liquid and dry it fully.
A0025597
C111 Residual coupling
start calibration (air set)
Current measured value
C111
µS/cm
AirSet
0.0
A0009145-EN
Press CAL to start the calibration.
Page 78
Commissioning Liquisys M CLM223/253
78 Endress+Hauser
Coding Field Range of
adjustment (factory settings in bold)
Display Info
C112 The residual
coupling is displayed (air set)
-80.0 to 80.0 μS/cm
C112
µS/cm
AirSetVal
5.3
A0009146-EN
Residual coupling of measuring system (sensor and transmitter).
C113 Calibration status
is displayed
o.k.
E xxx
C113
Status
o.k.
A0009147-EN
If the calibration status is not o.k., a reason for the error is provided on the second line of the display.
C114 Store calibration
result?
Yes
No New
C114
Store
yes
A0009148-EN
If C113 = E xxx, then only No or New. If New, return to C. If Yes/No, return to "Measurement".
C1(2) Cell constant
calibration
Airs = air set (1)
Cellc = cell constant (2)
InstF = installation factor (3)
C1
Calibrat
Cellc
A0009143-EN
The sensor should be immersed in such a way that there is sufficient distance to the vessel wall (the installation factor has no influence if a > 15 mm).
Immerse the sensor in the calibration solution.
The following section describes calibration with the temperature-compensated conductivity value of the reference solution. If calibration is to be performed with the uncompensated conductivity value, you must set the temperature coefficient α to zero.
A0025598
C121 Enter calibration
temperature (MTC)
25 °C
-35.0 to 250.0 °C
Only available if B1 = fixed.
C122 Enter the α value
of the calibration solution
2.10 %/K
0.00 to 20.00 %/K
%/K C122
alpha val
2.10
A0009150-EN
The value is given in the Technical Information for all E+H calibration solutions. You can also use the printed-on table to calculate the value. Set α to 0 for calibration with uncompensated values.
C123 Enter the correct
conductivity value of the calibration solution
Current measured value
0.0 μS/cm to 9999 mS/cm
C123
mS/cm
Real. val
10.30
A0009151-EN
A value close to the subsequent operating range should be selected.
Page 79
Liquisys M CLM223/253 Commissioning
Endress+Hauser 79
Coding Field Range of
adjustment (factory settings in bold)
Display Info
C124 The calculated cell
constant is displayed
0.0025 to 99.99 cm
-1
C124
1/cm
Cellconst
1.000
A0009152-EN
The calculated cell constant is displayed and accepted in A5.
C125 Calibration status
is displayed
o.k.
E xxx
C125
Status
o.k.
A0009153-EN
If the calibration status is not o.k., a reason for the error is provided on the second line of the display.
C126 Store calibration
result?
Yes
No New
C126
Store
yes
A0009154-EN
If C125 = E xxx, then only No or New. If New, return to C. If Yes/No, return to "Measurement".
C1(3) Calibration with
sensor adjustment for inductive sensors (only for Plus Package)
Airs = air set (1) Cellc = cell constant (2)
InstF = installation factor (3)
C1
Calibrat
InstF
A0009144-EN
Sensor adjustment with compensation for wall effects. In the case of inductive sensors, the measured value is influenced by the distance between the sensor and the pipe wall and by the material of the pipe (conductive or insulating). The installation factor indicates these dependencies. See the Technical Information of the sensor used for more information
The inductive sensor is mounted at the place of installation.
A0025599
C131 Enter the process
temperature (MTC)
25 °C
-35.0 to 250.0 °C
C131
C
°
MTC temp.
25.0
A0009155-EN
Only available if B1 = fixed.
C132 Enter the α value of
the calibration solution
2.10 %/K
0.00 to 20.00 %/K
C132
%/K
alpha val
2.10
A0009156-EN
The value is given in the TI for all E+H calibration solutions. You can also use the printed-on table to calculate the value. Set α to 0 for calibration with uncompensated values.
C133 Enter the correct
conductivity value of the calibration solution
Current measured value
0.0 μS/cm to 9999 mS/cm
C133
mS/cm
Real val.
10.30
A0009157-EN
A value close to the subsequent operating range should be selected.
Page 80
Commissioning Liquisys M CLM223/253
80 Endress+Hauser
Coding Field Range of
adjustment (factory settings in bold)
Display Info
C134 Calculated
installation factor is displayed
1
0.10 to 5.00
C134
InstFact
1
A0009158-EN
The installation factor indicates the dependency between the measured value and the distance between the sensor and pipe wall. See the Technical Information of the sensor used for more information.
C135 Calibration status is
displayed
o.k.
E xxx
C135
Status
o.k.
A0009159-EN
If the calibration status is not o.k., a reason for the error is provided on the second line of the display.
C136 Store calibration
result?
Yes
No New
C136
Store
yes
A0009160-EN
If C135 = E xxx, then only No or New. If New, return to C. If Yes/No, return to "Measurement".
Page 81
Liquisys M CLM223/253 Diagnostics and troubleshooting
Endress+Hauser 81
8 Diagnostics and troubleshooting
8.1 Trouble shooting instructions
The transmitter continuously monitors its functions itself. If an error occurs which the device recognizes, this is indicated on the display. The error number is shown below the display of the main measured value. If more than one error occurs, you can call these up with the MINUS key.
Refer to the "System error messages" table for the possible error numbers and remedial measures.
Should a malfunction occur without any transmitter error message, please refer to the "Process-specific errors" or the "Device-specific errors" tables to localize and rectify the error. These tables provide you with additional information on any spare parts required.
8.2 System error messages
You can display and select the error messages with the MINUS key.
Error No.
Display Tests/remedial action Alarm
contact
Error current
Autom. cleaning start
PROFIBU S Status
Facty Facty Facty PV
1)
User User User Temp
E001 EEPROM memory
error
• Switch off device and switch it on again.
• Load software compatible with hardware
• Load measurement­parameter specific device software.
• If the error persists, send in the device for repair to your local sales center or replace the device.
Yes No X 0C
X 0C
E002 Device not calibrated,
calibration data invalid, no user data or user data invalid (EEPROM error), device software not suitable for hardware (controller)
Yes No X 0C
X 0C
E003 Download error Invalid configuration.
Repeat download.
Yes No No 0C
0C
E004 Device software
version not compatible with module hardware version
Load software compatible with hardware Load measurement­parameter specific device software.
Yes No No 0C
0C
E007 Transmitter
malfunction, device software not compatible with transmitter version
Yes No X 0C
X 0C
E008 Sensor or sensor
connection faulty
Check sensor and sensor connection
Yes No Yes 0C
0C
E010 Temperature sensor
defective, not connected or short­circuited (measurement continues with 25 °C)
Check temperature sensor and connections; check the measuring device and measuring cable with a temperature simulator if necessary. Check that correct option is selected in field B1
Yes No No 80
0C
Page 82
Diagnostics and troubleshooting Liquisys M CLM223/253
82 Endress+Hauser
Error No.
Display Tests/remedial action Alarm
contact
Error current
Autom. cleaning start
PROFIBU S Status
Facty Facty Facty PV
1)
User User User Temp
E025 Limit value for air set
offset exceeded
Perform air set again (in air) or replace sensor. Dry sensor.
No No No 80
80
E036 Calibration range of
sensor exceeded
Clean sensor and recalibrate; if necessary, check sensor and connections.
No No No 80
80
E037 Below calibration
range of sensor
No No No 80
80
E045 Calibration aborted Repeat the calibration No No X 80
X 80
E049 Calibration range of
installation factor exceeded
Check pipe diameter, clean sensor and perform calibration again.
No No X 80
X 80
E050 Below calibration
range of installation factor
No No X 80
X 80
E055 Below main parameter
measuring range
Check measurement, control and connections
Yes No No 44
80
E057 Main parameter
measuring range exceeded
Yes No No 44
80
E059 Below temperature
measuring range
Yes No No 80
44
E061 Temperature
measuring range exceeded
Yes No No 80
44
E063 Below current output
range 1
Check measured value and current assignment
Yes No No 80
80
E064 Current output range 1
exceeded
Yes No No 80
80
E065 Below current output
range 2
Yes No No 80
80
E066 Current output range 2
exceeded
Yes No No 80
80
E067 Set point exceeded
limit contactor 1
Check configuration Yes No No 80
80
E068 Set point exceeded
limit contactor 2
Yes No No 80
80
E069 Set point exceeded
limit contactor 3
Yes No No 80
80
E070 Set point exceeded
limit contactor 4
Yes No No 80
80
E071 Incorrect
measurement / polarization
Clean sensor; check table; select suitable sensor.
Yes No No 44
80
Page 83
Liquisys M CLM223/253 Diagnostics and troubleshooting
Endress+Hauser 83
Error No.
Display Tests/remedial action Alarm
contact
Error current
Autom. cleaning start
PROFIBU S Status
Facty Facty Facty PV
1)
User User User Temp
E077 Temperature outside α
value table range
Clean sensor; check tables Yes No No 44
80
E078 Temperature outside
concentration table
Yes No No 44
80
E079 Conductivity outside
concentration table
Yes No No 44
80
E080 Current output 1 range
too small
Increase range in "Current outputs" menu.
Yes No X 80
X 80
E081 Current output 2 range
too small
Yes No X 80
X 80
E085 Incorrect setting for
error current
If the current range "0 to 20 mA" was selected in field O311, the error current may not be set to "2.4 mA".
Yes No No 80
80
E100 Current simulation
active
Yes No X 80
X 80
E101 Service function active Switch off service function
or switch device off and then on again.
No No X 80
X 80
E102 Manual mode active No No X 80
X 80
E106 Download active Wait for download to
finish.
No No X 80
X 80
E116 Download error Repeat download. Yes No X 0C
X 0C
E150 Distance between
temp. values in α value table too small or not monotonously increasing
Enter correct α value table (temperatures must be entered at intervals of at least 1 K)
No No No 44
80
E151 USP-EP error Yes No No 44
80
E152 PCS alarm Check sensor and
connection.
Yes No No 44
44
E153 USP-EP temperature
error
Yes No No 80
44
E154 Below lower alarm
threshold for period exceeding alarm delay
Perform manual comparison measurement if necessary. Service sensor and recalibrate.
Yes No No X
X
E155 Above upper alarm
threshold for period exceeding alarm delay
Yes No No X
X
E156 Actual value
undershoots alarm threshold for longer
Yes No No X
Page 84
Diagnostics and troubleshooting Liquisys M CLM223/253
84 Endress+Hauser
Error No.
Display Tests/remedial action Alarm
contact
Error current
Autom. cleaning start
PROFIBU S Status
Facty Facty Facty PV
1)
User User User Temp
than the set permissible maximum period
X
E157 Actual value exceeds
alarm threshold for longer than the set permissible maximum period
Yes No No X
X
E162 Dosage stop Check settings in the
CURRENT INPUT and CHECK function groups.
Yes No No X
X
E171 Flow in main stream
too low or zero
Restore flow. Yes No No X
X
E172 Switch-off limit for
current input exceeded
Check process variables at sending measuring device. Change range assignment if necessary.
Yes No No X
X
E173 Current input < 4 mA Yes No No X
X
E174 Current input > 20 mA Yes No No X
X
1) PV = process variable, primary value
8.3 Process-specific errors
Use the following table to localize and rectify any errors occurring.
Error Possible cause Tests/remedial action Tools, spare parts
Incorrect reading compared to comparison measurement
Device calibrated incorrectly
Calibrate the device according to Kap. the "Calibration" section
Calibration solution or cell certificate
Sensor fouled Clean sensor See the "Cleaning of
conductivity sensors" section
Incorrect temperature measurement
Check temperature measured value in measuring device and reference device
Temperature measuring device, precision thermometer
Temperature compensation incorrect
Check compensation method (none / ATC / MTC) and compensation type (linear/ substance/user table)
Please note: the transmitter has separate calibration and operating temperature coefficients
Reference device is calibrated incorrectly
Calibrate reference device or use verified device
Calibration solution, Operating Instructions of reference device
Incorrect ATC setting in reference device
Compensation method and compensation type must be identical on both devices.
Operating Instructions of reference device
Polarization errors Use suitable sensor:
• Use larger cell constant
• Use graphite instead of stainless steel (check resistance)
Measuring range tables e.g. in SI "Conductivity" or technical data of conductivity sensors
Page 85
Liquisys M CLM223/253 Diagnostics and troubleshooting
Endress+Hauser 85
Error Possible cause Tests/remedial action Tools, spare parts
Incorrect line resistance in field A6
Enter correct value CYK71: 165 Ω/km
Implausible measured values in general:
• Continuous measured value overflow
• Measured value constantly 000
• Measured value too low
• Measured value too high
• Measured value frozen
• Current output value not as expected
Short-circuit/moisture in sensor
Check sensor See the "Checking
inductive conductivity sensors" section.
Short-circuit in cable or socket
Check cable and socket See the "Connecting lines
and junction boxes" section.
Disconnection in sensor Check sensor See the "Checking
inductive conductivity sensors" section.
Disconnection in cable or socket
Check cable and socket See the "Connecting lines
and junction boxes" section.
Incorrect cell constant setting
Check cell constant Sensor nameplate or
certificate
Incorrect output assignment
Check assignment of measured value to current signal
Incorrect output function Check preset value (0-20 / 4 -20
mA) and curve shape (linear / table)
Air pockets in assembly Check assembly and installation
position
Grounding short at or in device
Measure in insulated device Plastic vessel, calibration
solutions
Transmitter module defective
Check with new module See the "Spare parts"
section
Device in impermissible operating condition (does not respond when key pressed)
Switch off device and switch it on again
EMC problem: if this persists, check the grounding, shields and line routing or have checked by E+H Service.
Incorrect conductivity measured value in the process
No/incorrect temperature compensation
ATC: select type of compensation; if linear, set suitable coefficients. MTC: set process temperature.
Incorrect temperature measurement
Check temperature measured value.
Reference device, thermometer
Bubbles in medium Suppress formation of bubbles by:
• Gas bubble trap
• Creating counterpressure (orifice plate)
• Measurement in bypass
Polarization effects (only with conductive sensors)
Use suitable sensor:
• Use larger cell constant
• Use graphite instead of stainless steel (check resistance)
Measuring range tables e.g. in SI "Conductivity" or technical data of conductivity sensors
Flow rate too high (can lead to bubble formation)
Reduce flow rate or select less turbulent mounting location.
Voltage potential in medium (only for conductive)
Ground medium close to sensor. Problem mainly occurs in
plastic lines and tanks
Sensor fouling or buildup on sensor
Clean the sensor (see the "Cleaning the conductivity sensors" section).
For heavily contaminated media: Use spray cleaning
Incorrect line resistance in field A6
Enter correct value CYK71: 165 Ω/km
Page 86
Diagnostics and troubleshooting Liquisys M CLM223/253
86 Endress+Hauser
Error Possible cause Tests/remedial action Tools, spare parts
Incorrect temperature value
Incorrect sensor connection
Check connections using wiring diagram. Three-wire connection always required.
Wiring diagram in "Electrical connection" section
Measuring cable defective
Check cable for interruptions/ short-circuit/shunt.
Ohmmeter
Incorrect sensor type Set type of temperature sensor at
the device (field B1).
Fluctuations in measured value
Interference on measuring cable
Connect cable shield as per wiring diagram
See the "Electrical connection" section
Interference on signal output cable
Check cable routing, possibly route cable separately
Route signal output and measuring input lines separately
Interference potential in medium
Eliminate source of interference or ground medium as close as possible to sensor.
Electromagnetic interference on signal lines with conductive sensors
Use shielded cables and ground the cable shield
Controller or timer cannot be activated
No relay module available
Install LSR1-2 or LSR1-4 module
Controller/limit contact does not work
Controller switched off Activate controller See fields R2xx
Controller in operating mode "Manual off"
Select "Auto" or "Manual on" mode Keyboard, REL key
Pickup delay setting too long
Switch off or shorten pickup delay time
See fields R2xx
"Hold" function active “Auto hold” for calibration,
“Hold” input activated; “Hold” active via keyboard
See fields S2 to S5
Controller/limit contact works continuously
Controller in operating mode "Manual on"
Select "Auto" or "Manual off" mode Keyboard, REL and AUTO
keys
Dropout delay setting too long
Shorten dropout delay time See fields R2xx
Control loop interruption Check measured value, current
output value, actuators, chemical supply
No conductivity current output signal
Cable disconnected or short-circuited
Disconnect cable and measure directly at device
mA meter 0–20 mA
Output defective See the "Device-specific errors"
section
Fixed conductivity current output signal
Current simulation active Switch off simulation. See field O3
Impermissible operating state of processor system
Disconnect line voltage for approx. 10 seconds
Possibly EMC problem: if problem persists, check grounding and wire routing.
Incorrect current output signal
Incorrect current assignment
Check current assignment: 0–20 mA or 4–20 mA?
Field O311
Total load in the current loop too high (> 500 Ω)
Disconnect output and measure directly at device
mA meter for 0–20 mA DC
EMC (interference coupling)
Disconnect both output cables and measure directly at device
Use shielded cables, ground shields at both ends, where necessary route cable in another cable duct
Page 87
Liquisys M CLM223/253 Diagnostics and troubleshooting
Endress+Hauser 87
Error Possible cause Tests/remedial action Tools, spare parts
Current output table is not accepted
Value interval too small Select practical intervals
No temperature output signal
Device does not have a second current output
Check version using nameplate, if necessary replace the LSCH-x1 module
LSCH-x2 module, see the "Spare parts" section
Device with PROFIBUS-PAPA device has no current output!
Chemoclean function not available
No relay module (LSR1­x) installed or only LSR1-2 available Additional function not enabled
Install LSR1-4 module. Chemoclean is enabled using the release code supplied by the manufacturer in the Chemoclean retrofit kit. To check the version, see the nameplate
LSR1-4 module, see the "Spare parts" section
Plus Package functions not available
Plus Package not enabled (enable by entering a code which depends on the serial number and which is supplied by E+H when a Plus Package is ordered)
• For Plus Package retrofit: code is supplied by E+H → enter this code.
• After replacing a defective LSCH/LSCP module: first enter device serial number manually (see nameplate), then enter the existing code number.
For a detailed description, see the "Replacement of central module" section.
No HART communication
No HART central module Verify using nameplate: HART = -
xxx5xx and -xxx6xx
Upgrade to LSCH-H1 / ­H2
No or incorrect DD (device description)
For further information, see BA00208C/07/EN, "HART field communication with Liquisys CxM223/253”
HART interface missing
Current output < 4 mA
Load too small (must be > 230 Ω)
HART receiver (e.g. FXA
191) not connected via load but via power supply
Incorrect device address (addr. = 0 for single operation, addr. > 0 for multidrop operation)
Line capacitance too high
Interference on line
Several devices set to same address
Assign addresses correctly No communication
possible if several devices set to the same address
No PROFIBUS communication
No PA/DP central module
Verify using the nameplate: PA =
-xxx3xx /DP = xxx4xx
Upgrade to LSCP module, see the "Spare parts" section
Incorrect device software version (without PROFIBUS)
For further information, see BA00209C/07/EN "PROFIBUS PA/DP - Field communication for Liquisys CxM223/253".
Information about PROFIBUS configuration is provided in Technical Information TI00260F, while detailed information about instrumentation and accessories is provided in Operating Instructions BA00198F
With Commuwin (CW) II: CW II version and device software version incompatible
No or incorrect DD/DLL
Incorrect baud rate setting for segment coupler in DPV-1 server
Page 88
Diagnostics and troubleshooting Liquisys M CLM223/253
88 Endress+Hauser
Error Possible cause Tests/remedial action Tools, spare parts
Bus user (master) has wrong address or address assigned twice
Bus user (slave) has wrong address
Bus line not terminated
Line problems (too long, cross-section too small, not shielded, shield not grounded, wires not twisted)
Bus voltage too low (Bus voltage typ. 24 V DC for non-Ex)
The voltage at the device's PA/DP connector must be at least 9 V
8.4 Device-specific errors
The following table helps you during the diagnosis and points to any spare parts required.
Depending on the degree of difficulty and the measuring equipment present, diagnosis is carried out by:
• Trained operator personnel
• The user's trained electrical technicians
• Company responsible for system installation/operation
• Endress+Hauser Service
Information on the exact spare part designations and on how to install these parts can be found in the "Spare parts" section.
Error Possible cause Tests/remedial action Execution, tools, spare
parts
Device cannot be operated, display value 9999
Operation locked Press CAL and MINUS keys
simultaneously.
See the "Key functions" section
Display dark, no light-emitting diode active
No line voltage Check whether line voltage is
present
Electrical technician / e.g. multimeter
Supply voltage wrong/too low
Compare actual line voltage and nameplate data
User (data for energy supply company or multimeter)
Connection faulty • Terminal not tightened
• Insulation jammed
• Wrong terminals used
Electrical technician
Device fuse defective Compare line voltage and the
nameplate data and replace fuse
Electrical technician/suitable fuse; see exploded drawing in the the "Spare parts" section
Power unit defective Replace power unit, note
version
Onsite diagnosis by Endress +Hauser Service, test module necessary
Central module defective Replace central module, note
version
Onsite diagnosis by Endress +Hauser Service, test module necessary
Field device: ribbon cable loose or defective
Check ribbon cable, replace if necessary
See the "Spare parts" section
Display dark, light­emitting diode active
Central module defective (module: LSCH/LSCP)
Replace central module, note version
Onsite diagnosis by Endress +Hauser Service, test module necessary
Page 89
Liquisys M CLM223/253 Diagnostics and troubleshooting
Endress+Hauser 89
Error Possible cause Tests/remedial action Execution, tools, spare
parts
Values appear on display but:
• Display does not change and / or
• Device cannot be operated
Device or module in device not correctly mounted
Panel-mounted device: reinstall insert. Field device: remount display module
Perform with the aid of the installation drawings in the the "Spare parts" section
Impermissible operating system condition
Disconnect line voltage for approx. 10 seconds
Possibly Possible EMC problem: if this persists, check the installation or have checked by Endress+Hauser Service
Device gets hot Voltage wrong/too high Compare line voltage and
nameplate data
User, electrical technician
Power unit defective Replace power unit Diagnosis only by Endress
+Hauser Service
Conductivity/MΩ measured value and/or temperature measured value incorrect
Transmitter module defective (module: MKIC), please first carry out tests and take measures as per the "Process-specific errors" section
Measuring input test:
• Connect a resistor instead of the conductivity sensor
• 100 Ω resistance at terminals 11/ 12 + 13 = display 0 °C
If the test is negative: replace module (note version). Perform with the aid of the exploded drawings in the "Spare parts" section.
Current output, incorrect current value
Incorrect adjustment Check with integrated current
simulation, connect mA meter directly to current output.
If simulation value incorrect: adjustment in factory or new LSCH module required. If simulation value correct: check current loop for load and shunts.
Load too large
Shunt / short to ground in current loop
Incorrect mode of operation
Check whether 0–20 mA or 4– 20 mA is selected.
No current output signal
Current output stage defective (only for LSCH module; LSCP has no current output)
Check with integrated current simulation, connect mA meter directly to current output
If test negative: Replace central module (note version)
Additional relays not working
Field device: ribbon cable loose or defective
Check ribbon cable seating, replace cable if required.
See the "Spare parts" section
Only 2 additional relays can be triggered
LSR1-2 relay module with 2 relays is installed
Upgrade to LSR1-4 with 4 relays.
User or Endress+Hauser Service
Additional functions (Plus Package) missing
No or incorrect release code used
If retrofitting: check whether the correct serial number was quoted when ordering the Plus Package.
Handled by Endress+Hauser Sales
Incorrect device serial number saved in LSCH/ LSCP module
Check whether serial number on the nameplate matches SNR in LSCH/ LSCP (field S 8).
The serial number of the device is definitive for the Plus Package.
Additional functions (Plus Package and/or Chemoclean) missing after LSCH/LSCP module replacement
Replacement modules LSCH or LSCP have the device serial number 0000 when they leave the factory. The Plus Package or Chemoclean are not enabled on leaving the factory.
In the case of LSCH/LSCP with SNR 0000, a device serial number can be entered once in fields E115 to E117. Then enter the release codes for the Plus Package and/or Chemoclean if necessary.
For a detailed description, see the "Replacement of central module" section.
No HART or PROFIBUS PA/DP interface function
Incorrect central module HART: LSCH-H1 or H2 module,
PROFIBUS-PA: LSCP-PA module, PROFIBUS-DP: LSCP-DP module, see field E112.
Replace central module; User or Endress+Hauser Service.
Page 90
Maintenance Liquisys M CLM223/253
90 Endress+Hauser
Error Possible cause Tests/remedial action Execution, tools, spare
parts
Wrong software SW version, see field E111.
Bus problem Remove some devices and
repeat the test.
Contact Endress+Hauser Service.
9 Maintenance
L
WARNING
Process pressure and temperature, contamination, electrical voltage
Risk of serious or fatal injury
‣
If the sensor has to be removed during maintenance work, avoid hazards posed by pressure, temperature and contamination.
‣
Make sure the device is de-energized before you open it.
‣
Power can be supplied to switching contacts from separate circuits. De-energize these circuits before working on the terminals.
Take all the necessary precautions in time to ensure the operational safety and reliability of the entire measuring point.
The maintenance of the measuring point comprises:
• Calibration
• Cleaning the controller, assembly and sensor
• Checking the cables and connections
When performing any work on the device, bear in mind any potential impact this may have on the process control system or on the process itself.
NOTICE
Electrostatic discharge (ESD)
Risk of damaging the electronic components
‣
Take personal protective measures to avoid ESD, such as discharging beforehand at PE or permanent grounding with a wrist strap.
‣
For your own safety, only use genuine spare parts. With genuine parts, the function, accuracy and reliability are also ensured after maintenance work.
9.1 Maintenance of the entire measuring point
9.1.1 Cleaning the transmitter
Clean the front of the housing using commercially available cleaning agents only.
The front of the housing is resistant to the following in accordance with DIN 42 115:
• Ethanol (for a short time)
• Diluted acids (max. 2% HCl)
• Diluted alkaline solutions (max. 3% NaOH)
• Soap-based household cleaning agents
When performing any work on the device, bear in mind any potential impact this may have on the process control system or on the process itself.
Page 91
Liquisys M CLM223/253 Maintenance
Endress+Hauser 91
NOTICE
Prohibited cleaning agents
Damage to the housing surface or housing seal
‣
Never use concentrated mineral acids or alkaline solutions for cleaning.
‣
Never use organic cleaners such as benzyl alcohol, methanol, methylene chloride, xylene or concentrated glycerol cleaner.
‣
Never use high-pressure steam for cleaning purposes.
9.1.2 Cleaning the conductivity sensors
L
CAUTION
Risk of injury from cleaning agents, damage to clothing and equipment
‣
Wear protective goggles and safety gloves.
‣
Clean away splashes on clothes and other objects.
‣
Pay particular attention to the information provided in the safety data sheets for the chemicals used.
Clean away fouling on the sensor as follows depending on the type of fouling:
• Oily and greasy films: Clean with grease remover, e.g. alcohol, acetone, possibly hot water and dishwashing detergent.
• Lime and metal hydroxide buildup: Dissolve buildup with diluted hydrochloric acid (3 %) and then rinse thoroughly with plenty of clear water.
• Sulfidic buildup (from flue gas desulfurization or sewage treatment plants): Use a mixture of hydrochloric acid (3 %) and thiocarbamide (commercially available) and then rinse thoroughly with plenty of clear water.
• Buildup containing proteins (e.g. food industry): Use a mixture of hydrochloric acid (0.5 %) and pepsin (commercially available) and then rinse thoroughly with plenty of clear water.
9.1.3 Simulation of conductive sensors for device testing
You can test the transmitter for conductive conductivity by replacing the measurement section and temperature sensor with resistors. The accuracy of the simulation will depend on the accuracy of the resistors.
Temperature
If a temperature offset is not set at the transmitter, the temperature values of the following table apply:
Pt 100 equivalent resistors
Temperature Resistance value
-20 °C (-4 °F) 92.13 Ω
-10 °C (14 °F) 96.07 Ω
0 °C (32 °F) 100.00 Ω
10 °C (50 °F) 103.90 Ω
20 °C (68 °F) 107.79 Ω
25 °C (77 °F) 109.73 Ω
50 °C (122 °F) 119.40 Ω
80 °C (176 °F) 130.89 Ω
100 °C (212 °F) 138.50 Ω
200 °C (392 °F) 175.84 Ω
Page 92
Maintenance Liquisys M CLM223/253
92 Endress+Hauser
With the Pt 1000 type of temperature sensor, all the resistance values are increased by a factor of 10.
Attach the temperature equivalent resistor in a three-wire system.
Conductivity
If the cell constant k is set to the value in column 2 of the following table, then the conductivity values of this table apply.
Otherwise the following applies: Conductivity [mS/cm] = k[cm-1] ⋅ 1 / R[kΩ]
Resistance R Cell constant k Display for conductivity Display for MΩ
10 Ω 1 cm
-1
100 mS/cm
10 cm
-1
1000 mS/cm
100 Ω 0.1 cm
-1
1 mS/cm 1 kΩ ⋅ cm
1 cm
-1
10 mS/cm
10 cm
-1
100 mS/cm
1000 Ω 0.1 cm
-1
0.1 mS/cm 10 kΩ ⋅ cm
1 cm
-1
1 mS/cm
10 cm
-1
10 mS/cm
10 kΩ 0.01 cm
-1
1 μS/cm 1 MΩ ⋅ cm
0.1 cm
-1
10 μS/cm 100 kΩ ⋅ cm
1 cm
-1
100 μS/cm
10 cm
-1
1 mS/cm
100 kΩ 0.01 cm
-1
0.1 μS/cm 10 MΩ ⋅ cm
0.1 cm
-1
1 μS/cm 1 MΩ ⋅ cm
1 cm
-1
10 μS/cm
1 MΩ 0.01 cm
-1
0.01 μS/cm 100 MΩ ⋅ cm
0.1 cm
-1
0.1 μS/cm 10 MΩ ⋅ cm
1 cm
-1
1 μS/cm
10 MΩ 0.01 cm
-1
0.001 μS/cm
0.1 cm
-1
0.01 μS/cm 100 MΩ ⋅ cm
The MΩ measurement is normally used for pure and ultrapure water and therefore only makes sense for cell constants k= 0.01 cm-1 or k = 0.1 cm-1.
9.1.4 Simulation of inductive sensors for device testing
The inductive sensor cannot be simulated.
However, the overall system comprising the transmitter and inductive sensor can be tested using equivalent resistors. Note the cell constant k (e.g. k
nominal
= 1.98 cm-1 for CLS50,
k
nominal
= 5.9 cm-1 for CLS52, k
nominal
= 6.3 cm-1 for CLS54).
For accurate simulation, the actual cell constant used (visible in field C124) must be used to calculate the display value.
The formula for calculation also depends on the type of sensor:
• CLS50 and CLS52: conductivity reading [mS/cm] = k[cm-1] ⋅1/R[kΩ]
• CLS54: conductivity reading [mS/cm] = k[cm-1] ⋅1/R[kΩ] ⋅ 1.21
Page 93
Liquisys M CLM223/253 Maintenance
Endress+Hauser 93
Values for simulation with CLS50 at 25 °C (77 °F):
Simulation resistance R Default cell constant k Conductivity reading
2 Ω 1.98 cm
-1
990 mS/cm
10 Ω 1.98 cm
-1
198 mS/cm
100 Ω 1.98 cm
-1
19.8 mS/cm
1 kΩ 1.98 cm
-1
1.98 mS/cm
Simulation with CLS54 at 25 °C (77 °F):
Simulation resistance R Default cell constant k Conductivity reading
10 Ω 6.3 cm
-1
520 mS/cm
26 Ω 6.3 cm
-1
200 mS/cm
100 Ω 6.3 cm
-1
52 mS/cm
260 Ω 6.3 cm
-1
20 mS/cm
2.6 kΩ 6.3 cm
-1
2 mS/cm
26 kΩ 6.3 cm
-1
200 μS/cm
52 kΩ 6.3 cm
-1
100 μS/cm
Conductivity simulation:
Pull a cable through the sensor opening and then connect it to a decade resistor, for instance.
9.1.5 Checking conductive sensors
• Measuring surface connection: The measuring surfaces are directly connected to the connections of the sensor connector. Check with ohmmeter for < 1 Ω.
• Measuring surface shunt: A shunt is not permitted between the measuring surfaces. Check with ohmmeter for > 20 MΩ.
• Temperature sensor shunt: A shunt is not permitted between the measuring surfaces and the temperature sensor. Check with ohmmeter for > 20 MΩ.
• Temperature sensor: See the sensor nameplate for information on the type of the temperature sensor being used. The sensor can be checked at the sensor connector with an ohmmeter: – Pt 100 at 25 °C (77 °F) = 109.79 Ω – Pt 1000 at 25 °C (77 °F) = 1097.9 Ω – NTC 30 k at 25 °C (77 °F) = 30 kΩ
• Terminal: For sensors with a terminal connection (CLS12/13) check the assignment of the terminals to ensure they have not been inadvertently interchanged. Check the tightness of the terminal screws.
Page 94
Maintenance Liquisys M CLM223/253
94 Endress+Hauser
9.1.6 Checking inductive sensors
The following applies to the CLS50, CLS52 and CLS54 sensors.
The sensor cables must be disconnected at the device or junction box for all tests described here!
• Testing the transmitting and receiving coils: Perform the measurement at the white and red coaxial cables, between the inner conductor and shield in each case. – Ohmic resistance
CLS50/52: approx. 0.5 to 2 Ω CLS54: approx. 1 to 3 Ω
– Inductance approx. 180 to 500 mH (for 2 kHz, series circuit as equivalent circuit
diagram) CLS50: approx. 250 to 450 mH CLS52/54: approx. 180 to 550 mH
• Testing the coil shunt: A shunt is not permitted between the two sensor coils (from red coax to white coax). The measured resistance must be > 20 MΩ. Test with the ohmmeter from the red coax cable to the white coax cable.
• Temperature sensor test: Use the table in the "Simulation of conductive sensors for device testing" section to check the Pt 100 / Pt 1000 in the sensor. Measure between the green and white wires and between the green and yellow wires. The resistance values must be identical.
• Temperature sensor shunt test: Shunts are not permitted between the temperature sensor and the coils. Check with ohmmeter for > 20 MΩ Measure between the temperature sensor wires (green + white + yellow) and the coils (red and white coax).
9.1.7 Assembly
Refer to the assembly operating manual for information on servicing and troubleshooting the assembly. The assembly operating manual describes the procedure for mounting and disassembling the assembly, replacing the sensors and seals, and contains information on the material resistance properties, as well as on spare parts and accessories.
9.1.8 Connecting cables and junction boxes
• Use the methods described in chapters "Simulation of conductive/inductive sensors for device test" to perform a quick functional check from the conductivity sensor (sensor connector) to the measuring instrument via an extension. The easiest way to connect resistor decades is with the "Conductivity testing adapter" service kit, order number: 51500629
• Check junction boxes for: – Moisture (influence at low conductivity or for MΩ measurement, where necessary dry
box, replace seals, insert dehydrating bag) – Correct connection of all lines – Connection of the outer shields – Tightness of the terminal screws
Page 95
Liquisys M CLM223/253 Repair
Endress+Hauser 95
10 Repair
10.1 Spare parts
Please order spare parts from your local Sales Office. For this purpose, use the order numbers listed in the "Spare part kits" section.
For safety, you should always provide the following additional data when ordering spare parts:
• Device order code
• Serial number
• Software version, if possible
You can take the order code and serial number from the nameplate.
The software version is provided in the device software provided that the device processor system is still working.
For more detailed information on spare parts kits, please refer to the "Spare Part Finding Tool" on the Internet:
www.endress.com/spareparts_consumables
10.2 Disassembling the panel-mounted device
Please note the effects on the process if the device is taken out of service!
See the exploded drawing for the item numbers.
1. Disconnect the terminal block (item 420 b) from the rear of the device to de-energize the device.
2. Then disconnect the terminal blocks (item 420 a and 430 where applicable) from the rear of the device. Now you can disassemble the device.
3. Press in the latches of the end frame (item 340) and remove the frame from the rear.
4. Release the special screw (item 400) by turning it counter-clockwise.
5. Remove the entire electronics block from the housing. The modules are only mechanically connected and can be easily separated:
6. Simply remove the processor/display module towards the front.
7. Pull out the brackets of the rear plate (item 320) slightly.
8. Now you can remove the side modules.
9. Remove the conductivity transmitter (item 240) as follows:
10. Using a fine wire cutter, nip off the heads of the synthetic distance holders.
11. Then remove the module from above.
Assembly is the reverse of the disassembly sequence. Tighten the special screw finger­tight without using a tool.
Page 96
Repair Liquisys M CLM223/253
96 Endress+Hauser
230V
100V
115V
400
345
340
420a
420b
320
240
15/20
40 ... 91
330
15
A
B
290/300
430
310
A0008913
 39 Exploded drawing of panel-mounted device
Page 97
Liquisys M CLM223/253 Repair
Endress+Hauser 97
The exploded drawing contains the components and spare parts of the panel-mounted device. You can take the spare parts and the corresponding order number from the following section using the item numbers.
Item Kit description Name Function/contents Order no.
15 Power unit (main module) LSGA 100/115/230 V AC 51500317
20 Power unit (main module) LSGD 24 V AC + DC 51500318
40 Central module (contr.),
conductive
LSCH-S1 1 current output 51501210
50 Central module (contr.),
conductive
LSCH-S2 2 current outputs 51501212
60 Central module (contr.),
conductive
LSCH-H1 1 current output + HART 51501213
70 Central module (contr.),
conductive
LSCH-H2 2 current outputs + HART 51501214
80 Central module (contr.),
conductive
LSCP-PA PROFIBUS PA/no current output 51501215
90 Central module (contr.),
conductive
LSCP-DP PROFIBUS DP/no current output 51502502
90 Kit for CLM2x3
conductivity central module, conductive, PROFIBUS DP
LSCP-DP PROFIBUS DP central module
Relay module + 2 relays Current inp. + DP terminals Hardware version 2.10 and higher
71134726
41 Central module (contr.),
inductive
LSCH-S1 1 current output 51501216
51 Central module (contr.),
inductive
LSCH-S2 2 current outputs 51501218
61 Central module (contr.),
inductive
LSCH-H1 1 current output + HART 51501219
71 Central module (contr.),
inductive
LSCH-H2 2 current outputs + HART 51501220
81 Central module (contr.),
inductive
LSCP-PA PROFIBUS PA/no current output 51501221
91 Central module (contr.),
inductive
LSCP-DP PROFIBUS DP/no current output 51502501
91 Kit for CLM2x3
conductivity central module, inductive, PROFIBUS DP
LSCP-DP PROFIBUS DP central module
Relay module + 2 relays Current inp. + DP terminals Hardware version 2.10 and higher
71134727
240 Conductivity transmitter
(hazardous area version)
MKIC Conductivity + temperature input 71161137
Conductivity transmitter MKIC Conductivity + temperature input 71161133
290 Relay module LSR1-2 2 relays 51500320
290 Relay module LSR2-2i 2 relays + 4-20 mA current input 51504304
290 Kit for Cxm2x3 relay
module PROFIBUS DP
LSR2-DP Relay module + 2 relays
Curr. input + DP terminals Hardware version 2.10 and higher
71134732
300 Relay module LSR1-4 4 relays 51500321
300 Relay module LSR2-4i 4 relays + 4-20 mA current input 51504305
310 Side panel Kit with 10 parts 51502124
310, 320,
340, 400
Housing mechanical parts Rear plate, side panel, end frame,
special screw
51501076
Page 98
Repair Liquisys M CLM223/253
98 Endress+Hauser
Item Kit description Name Function/contents Order no.
330, 400 Housing module Housing with front membrane, sensory
tappets, seal, special screw, tensioning dogs, connection plates and nameplates
51501075
340 End frame
PROFIBUS-DP
Rear frame for PROFIBUS DP, with D­submin plug connector
51502513
345 Grounding terminal strip PE and shielding terminals 51501086
420a,
420b
Terminal strip set Complete terminal strip set, standard +
HART
51501203
420a,
420b
Terminal strip set Complete terminal strip set
PROFIBUS PA
51502126
420a,
420b
Terminal strip set Complete terminal strip set
PROFIBUS DP
51502493
430 Terminal strip Terminal strip for relay module 51501078
A Fuse Part of power unit, item 15
B Choice of line voltage Position of jumper on power unit, item
15 depending on line voltage
10.3 Disassembling the field device
Please note the effects on the process if the device is taken out of service!
See the exploded drawing for the item numbers.
You require the following tools to disassemble the field device:
• Standard set of screwdrivers
• Torx screwdriver, size TX 20
Proceed as follows to disassemble the field device:
1. Open and remove the cover of the connection compartment (item 420).
2. Disconnect the supply terminal (item 470) to de-energize the device.
3. Open the display cover (item 410) and release the ribbon cables (item 310/320) on the central module side (item 40 to 91) .
4. To remove the central module (item 40) release the screw in the display cover (item 450 b).
5. Proceed as follows to remove the electronics box (item 330):
6. Turn the screws in the housing base (item 450 a) by two revolutions to release them.
7. Then push the entire box backwards and remove it from above while making sure that the module locks do not open.
8. Release the ribbon cables (item 310/320).
9. Bend the module locks out and remove the modules.
10. To remove the docking module (item 340) remove the screws in the housing base (item 450 c) and remove the entire assembly from above.
11. To remove the conductivity transmitter (item 240) nip off the heads of the synthetic distance sleeves using a fine wire cutter.
12. Then remove the module from above.
Page 99
Liquisys M CLM223/253 Repair
Endress+Hauser 99
To assemble, carefully push the modules into the guide rails of the electronics box and let them engage in the side box noses.
It is not possible to mount the modules incorrectly. Modules inserted in the electronics box incorrectly are not operable since the ribbon cables cannot be connected.
Make sure the cover seals are intact to guarantee IP 65 ingress protection.
Page 100
Repair Liquisys M CLM223/253
100 Endress+Hauser
100 V
Pos. 10
Netzteil 100 – 230 V AC
230 V
115 V
Sicherung
15
C
D
460a
420
40 ... 91
310/320
240
15/20
340
450c
470
480
290/300
400
430
450a
330
410
460b
370
A
B
450b
403
A0008916
 40 Field device exploded drawing
Loading...