Endress+Hauser 16D/21D Operating Instructions Manual

Page 1
Products Solutions Services
Operating Instructions
Condumax CLS15D/16D/21D
Sensors with Memosens protocol For conductive measurement of conductivity in liquids
BA01147C/07/EN/03.16 71333350
Page 2
Page 3
Table of contents
Endress+Hauser 3
Table of contents
1 Document information ......... 4
1.1 Warnings ........................... 4
1.2 Symbols ............................ 4
2 Basic safety instructions ....... 5
2.1 Requirements for personnel ........... 5
2.2 Designated use ...................... 5
2.3 Occupational safety .................. 5
2.4 Operational safety ................... 6
2.5 Product safety ....................... 6
3 Incoming acceptance and
product identification .......... 9
3.1 Incoming acceptance ................. 9
3.2 Product identification ................. 9
3.3 Scope of delivery .................... 10
3.4 Certificates and approvals ............ 11
4 Installation .................... 12
4.1 Mounting the sensor ................ 12
4.2 Post-installation check ............... 16
5 Electrical connection .......... 16
5.1 Connection conditions ............... 17
5.2 Connecting the sensor ............... 19
5.3 Ensuring the degree of protection ..... 19
5.4 Post-connection check ............... 19
6 Commissioning ................ 20
7 Maintenance .................. 20
8 Repair .......................... 21
8.1 Sealing ring replacement and
recalibration (CLS16D only) .......... 21
8.2 Return ............................ 22
8.3 Disposal ........................... 22
9 Technical data ................. 23
10 EU Declaration of
conformity ..................... 29
10.1 ATEX / NEPSI II 1G Ex ia IIC T3/T4/T6
Ga, IECEx Ex ia IIC T3/T4/T6 Ga ....... 29
10.2 ATEX/NEPSI II 3G Ex ic IIC T3/T4/T6
Gc ................................ 30
Page 4
Document information Condumax CLS15D/16D/21D
4 Endress+Hauser
1 Document information
1.1 Warnings
Structure of information Meaning
L
DANGER
Causes (/consequences)
Consequences of non-compliance (if applicable)
Corrective action
This symbol alerts you to a dangerous situation. Failure to avoid the dangerous situation will result in a fatal or serious injury.
L
WARNING
Causes (/consequences)
Consequences of non-compliance (if applicable)
Corrective action
This symbol alerts you to a dangerous situation. Failure to avoid the dangerous situation can result in a fatal or serious injury.
L
CAUTION
Causes (/consequences)
Consequences of non-compliance (if applicable)
Corrective action
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or more serious injuries.
NOTICE
Cause/situation
Consequences of non-compliance (if applicable)
Action/note
This symbol alerts you to situations which may result in damage to property.
1.2 Symbols
Symbol Meaning
Additional information, tips
Permitted or recommended
Not permitted or not recommended
Reference to device documentation
Reference to page
Reference to graphic
Result of a step
Page 5
Condumax CLS15D/16D/21D Basic safety instructions
Endress+Hauser 5
2 Basic safety instructions
2.1 Requirements for personnel
• Installation, commissioning, operation and maintenance of the measuring system may be
carried out only by specially trained technical personnel.
• The technical personnel must be authorized by the plant operator to carry out the specified
activities.
• The electrical connection may be performed only by an electrical technician.
• The technical personnel must have read and understood these Operating Instructions and
must follow the instructions contained therein.
• Measuring point faults may be repaired only by authorized and specially trained personnel.
Repairs not described in the Operating Instructions provided may only be carried out directly by the manufacturer or by the service organization.
2.2 Designated use
The conductivity sensors are designed for the conductive measurement of the conductivity of liquids.
They are used in the following fields:
Sensor Applications Hazardous areas
Condumax CLS15D Measurements in pure and ultrapure water Approved for Ex zone 0
Condumax CLS16D Measurements in pure and ultrapure water with hygienic requirements Approved for Ex zone 0
Condumax CLS21D Measurements in media with medium or high conductivity Approved for Ex zone 0
Use of the device for any purpose other than that described, poses a threat to the safety of people and of the entire measuring system and is therefore not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
2.3 Occupational safety
As the user, you are responsible for complying with the following safety conditions:
• Installation guidelines
• Local standards and regulations
• Regulations for explosion protection
Electromagnetic compatibility
• The product has been tested for electromagnetic compatibility in accordance with the
applicable European standards for industrial applications.
• The electromagnetic compatibility indicated applies only to a product that has been
connected in accordance with these Operating Instructions.
Page 6
Basic safety instructions Condumax CLS15D/16D/21D
6 Endress+Hauser
2.4 Operational safety
1. Before commissioning the entire measuring point, verify that all connections are correct. Ensure that electrical cables and hose connections are undamaged.
2. Do not operate damaged products, and safeguard them to ensure that they are not operated inadvertently. Label the damaged product as defective.
3. If faults cannot be rectified: Take the products out of operation and safeguard them to ensure that they are not
operated inadvertently.
2.5 Product safety
2.5.1 State of the art
The product is designed to meet state-of-the-art safety requirements, has been tested, and left the factory in a condition in which it is safe to operate. The relevant regulations and European standards have been observed.
2.5.2 Electrical equipment in hazardous areas
ATEX / NEPSI II 1G Ex ia IIC T3/T4/T6 Ga, IECEx Ex ia IIC T3/T4/T6 Ga
• The Memosens inductive sensor-cable connection system is suitable for use in hazardous areas in accordance with EC type examination certificate BVS 04 ATEX E 121 X. The corresponding EC declaration of conformity is part of this document.
• The certified conductivity sensors CLS15D/CLS16D/CLS21D may only be connected via the measuring cable CYK10-G/I*** to the certified intrinsically safe digital sensor output circuits of measuring device Liquiline M CM42-KE/F/G/I/J********** in accordance with EC type-examination certificate TÜV 13 ATEX 7459 X.
• The electrical connection must be made according to the wiring diagram of the transmitter.
• Metallic process connection parts must be mounted at the mounting location electrostatically conductive (< 1 MΩ).
• The CLS15D-type sensors with non-metal process connections and the CLS21D-type sensors may only be employed for measurement in liquids with a minimum conductivity of 10 nS/cm.
• The CLS15D-type sensors with non-metal process connections may not be operated under process conditions in which electrostatic charging of the sensor, and particularly of the electrically insulated outer electrode, is likely to occur.
• Measuring cable CYK10-G/I*** and its terminal head must be protected against electrostatic charging if it is run through Zone 0.
• The maximum permitted cable length is 100 m.
• Ex versions of digital sensors with Memosens technology are indicated by an orange-red ring.
• Full compliance with regulations for electrical systems in hazardous areas (e.g. EN/IEC 60079-14) is mandatory when using the devices and sensors.
Page 7
Condumax CLS15D/16D/21D Basic safety instructions
Endress+Hauser 7
Temperature classes
Name Type Medium temp. Ta for temperature class (Tn) Cat.
Condumax CLS15D - A ** G -20 °C ≤ Ta ≤ +135 °C (T3)
-20 °C ≤ Ta ≤ +120 °C (T4)
-20 °C ≤ Ta ≤ +70 °C (T6)
II 1G
Condumax CLS15D - B/L ** G -20 °C ≤ Ta ≤ +135 °C (T3)
-20 °C ≤ Ta ≤ +100 °C (T4)
-20 °C ≤ Ta ≤ +50 °C (T6)
II 1G
Condumax CLS16D - ** ** G -5 °C ≤ Ta ≤ +135 °C (T3)
-5 °C ≤ Ta ≤ +115 °C (T4)
-5 °C ≤ Ta ≤ +65 °C (T6)
II 1G
Condumax CLS21D - * ** G -20 °C ≤ Ta ≤ +135 °C (T3)
-20 °C ≤ Ta ≤ +115 °C (T4)
-20 °C ≤ Ta ≤ +65 °C (T6)
II 1G
If the specified medium temperatures are complied with, temperatures that are not permitted for the respective temperature class will not occur on the equipment.
ATEX/NEPSI II 3G Ex ic IIC T3/T4/T6 Gc
• The Memosens inductive sensor-cable connection system is suitable for use in hazardous areas, Zone 2. The corresponding EU declaration of conformity is part of this document.
• The certified conductivity sensors CLS15D / CLS16D / CLS21D may only be connected via the measuring cable CYK10-V*** to the certified intrinsically safe digital sensor output circuits of measuring device Liquiline M CM42-KV*********.
• The electrical connection must be made according to the wiring diagram of the transmitter.
• Metallic process connection parts must be mounted at the mounting location electrostatically conductive (< 1 MΩ).
• The CLS15D-type sensors with non-metal process connections and the CLS21D-type sensors may only be employed for measurement in liquids with a minimum conductivity of 10 nS/cm.
• The CLS15D-type sensors with non-metal process connections may not be operated under process conditions in which electrostatic charging of the sensor, particularly of the electrically insulated outer electrode, is likely to occur.
• The maximum permitted cable length is 100 m.
• Full compliance with regulations for electrical systems in hazardous areas (EN/IEC 60079-14) is mandatory when using the devices and sensors.
Temperature classes
Name Type Medium temp. Ta for temperature class (Tn) Cat.
Condumax CLS15D - A ** V -20 °C ≤ Ta ≤ +135 °C (T3)
-20 °C ≤ Ta ≤ +120 °C (T4)
-20 °C ≤ Ta ≤ +70 °C (T6)
II 3G
Condumax CLS15D - B/L ** V -20 °C ≤ Ta ≤ +135 °C (T3)
-20 °C ≤ Ta ≤ +100 °C (T4)
-20 °C ≤ Ta ≤ +50 °C (T6)
II 3G
Page 8
Basic safety instructions Condumax CLS15D/16D/21D
8 Endress+Hauser
Name Type Medium temp. Ta for temperature class (Tn) Cat.
Condumax CLS16D - ** ** V -5 °C ≤ Ta ≤ +135 °C (T3)
-5 °C ≤ Ta ≤ +115 °C (T4)
-5 °C ≤ Ta ≤ +65 °C (T6)
II 3G
Condumax CLS21D - * ** V -20 °C ≤ Ta ≤ +135 °C (T3)
-20 °C ≤ Ta ≤ +115 °C (T4)
-20 °C ≤ Ta ≤ +65 °C (T6)
II 3G
If the specified medium temperatures are complied with, temperatures that are not permitted for the respective temperature class will not occur on the equipment.
FM/CSA IS/NI Cl.1 Div.1&2 Gr. A-D
Observe the documentation and the control drawings of the transmitter.
Page 9
Condumax CLS15D/16D/21D Incoming acceptance and product identification
Endress+Hauser 9
3 Incoming acceptance and product identification
3.1 Incoming acceptance
1. Verify that the packaging is undamaged.  Notify your supplier of any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
2. Verify that the contents are undamaged.  Notify your supplier of any damage to the delivery contents.
Keep the damaged products until the matter has been settled.
3. Check the delivery for completeness.  Check it against the delivery papers and your order.
4. Pack the product for storage and transportation in such a way that it is protected against impact and moisture.
The original packaging offers the best protection.
The permitted ambient conditions must be observed (see "Technical data").
If you have any questions, please contact your supplier or your local sales center.
3.2 Product identification
3.2.1 Type code for versions with explosion protection
Name Type Version
Condumax CLS15D
CLS16D CLS21D
-
-
-
* ** *
** ** **
G G G
Process connections, materials not Ex-relevant
For use in hazardous areas, ATEX/NEPSI II 1G Ex ia IIC T3/T4/T6 Ga, IECEx Ex ia IIC T3/T4/T6 Ga
Name Type Version
Condumax CLS15D
CLS16D CLS21D
-
-
-
* ** *
** ** **
O O O
Process connections, materials not Ex-relevant
For use in hazardous areas, FM/CSA IS/NI Cl I Div.1&2 Gr. A-D
Name Type Version
Condumax CLS15D
CLS16D CLS21D
-
-
-
* ** *
** ** **
V V V
Process connections, materials not Ex-relevant
For use in hazardous areas, ATEX/NEPSI II 3G Ex ic IIC T3/T4/T6 Gc
Page 10
Incoming acceptance and product identification Condumax CLS15D/16D/21D
10 Endress+Hauser
3.2.2 Nameplate
The nameplate provides you with the following information on your device:
• Manufacturer identification
• Extended order code
• Serial number
• Safety information and warnings
• Cell constant (nominal value)
• Protection class
• Ex labeling on hazardous area versions
Compare the data on the nameplate with your order.
3.2.3 Product identification
Product page
www.endress.com/cls15d www.endress.com/cls16d www.endress.com/cls21d
Interpreting the order code
The order code and serial number of your product can be found in the following locations:
• On the nameplate
• In the delivery papers
Obtaining information on the product
1. Go to the product page for your product on the Internet.
2. At the bottom of the page, select the "Online Tools" link followed by "Check your device features".
An additional window opens.
3. Enter the order code from the nameplate into the search field, and then select "Show details".
You will receive information on each feature (selected option) of the order code.
Manufacturer's address
Endress+Hauser Conducta GmbH+Co. KG Dieselstraße 24 D-70839 Gerlingen
3.3 Scope of delivery
The scope of delivery includes:
• Sensor in the version ordered
• Operating Instructions
Page 11
Condumax CLS15D/16D/21D Incoming acceptance and product identification
Endress+Hauser 11
3.4 Certificates and approvals
3.4.1
mark
Declaration of conformity
The product meets the requirements of the harmonized European standards. As such, it complies with the legal specifications of the EU directives. The manufacturer confirms successful testing of the product by affixing to it the mark.
3.4.2 Hazardous area approvals
ATEX / NEPSI II 1G Ex ia IIC T3/T4/T6 Ga, IECEx Ex ia IIC T3/T4/T6 Ga FM/CSA IS/NI Cl. I Div.1&2 Gr. A-D in conjunction with Liquiline M CM42 transmitter ATEX/NEPSI II 3G Ex ic IIC T3/T4/T6 Gc for use in Zone 2 with Liquiline M CM42-KV***
transmitter
ATEX and FM/CSA versions of digital sensors with Memosens technology are indicated by a red-orange ring in the plug-in head.
3.4.3 EHEDG (CLS16D only)
Validated as follows:
• Cleanability in accordance with EHEDG, Document 2
• Sterilizability in accordance with EHEDG, Document 5
• Bacteria-tightness in accordance with EHEDG, Document 7
3.4.4 FDA (CLS16D only)
All materials in contact with the product are listed by the FDA.
3.4.5 Manufacturer inspection certificate
Stating the individual cell constant
3.4.6 Biological reactivity test (USP class VI, CLS16D only)
Biological reactivity test certificate according to USP (United States Pharmacopeia)part <87> and part <88> class VI with batch traceability of materials in contact with the medium
3.4.7 Inspection certificate in accordance with EN 10204 3.1
A test certificate 3.1 in accordance with EN10204 is supplied depending on the version (→ Product Configurator on the product page).
3.4.8 ASME (CLS16D only)
Manufactured in accordance with ASME criteria (American Society of Mechanical Engineers)
3.4.9 Certification body DEKRA EXAM GmbH
Bochum
Page 12
Installation Condumax CLS15D/16D/21D
12 Endress+Hauser
4 Installation
4.1 Mounting the sensor
4.1.1 CLS15D
The sensors are installed directly via the process connection thread NPT ½" or ¾" or clamp 1 ½". As an option, the sensor can also be installed using a commercially available T­piece or cross fitting or using a flow assembly.
NOTICE
Incorrect mounting or disassembly
The Memosens head could become loose and fall off, resulting in total sensor failure
Only mount the sensor via the process connection.
To do so, use a suitable tool, such as an open-ended wrench.
Page 13
Condumax CLS15D/16D/21D Installation
Endress+Hauser 13
2
1
3
35 .38(1 )
A0019015
 1 With NPT ½" thread in T-piece or cross
fitting
1 T-piece or cross fitting (DN 32, 40 or 50) 2 Glue-in VC threaded coupling (NPT ½" for
DN 20)
3 Glue-in adapter coupling (for DN 32, 40, 50)
NPT ½”
160 (6.30)
Ø 29 (1.14)
Ø 6 (0.24)
30 (1.18)
25 (0.98)
Ø 6 (0.24)
38.5
(1.52)
1
2
3
A0019014
 2 With NPT ½" thread in flow assembly
71042405, dimensions in mm (inch)
1 Sensor holder NPT ½" 2 Inlet 3 Outlet
Ensure that the electrodes are fully immersed in the medium during measurement. The immersion depth must be at least 35 mm (1.38"). If the sensor is being used in the ultrapure water range, you must work under air-evacuated conditions. Otherwise the CO
2
in the air can dissolve in the water and its (weak) dissociation can increase the conductivity by up to 3 µS/cm.
Page 14
Installation Condumax CLS15D/16D/21D
14 Endress+Hauser
4.1.2 CLS16D
The sensors are installed directly via the process connection. Take the flow direction into consideration when installing in pipes.
A0019016
 3 Permitted flow direction
A0019017
 4 Inadmissible flow direction
Ensure that the electrodes are fully immersed in the medium during measurement. If the sensor is being used in the ultrapure water range, you must work under air-evacuated conditions. Otherwise the CO2 in the air can dissolve in the water and its (weak) dissociation can increase the conductivity by up to 3 µS/cm.
4.1.3 CLS21D Clamp connection
Both sheet-metal brackets and solid brackets can be used to secure the sensor. Sheet­metal brackets have a lower dimensional stability, uneven bearing surfaces causing point loads, and sometimes sharp edges that can damage the clamp. We recommend you only use solid brackets due to their higher dimensional stability. Solid brackets can be used over the entire pressure/temperature range (see pressure-temperature ratings).
Page 15
Condumax CLS15D/16D/21D Installation
Endress+Hauser 15
The sensors are installed directly via the process connection. As an option, the sensor can also be installed via a flow assembly.
NOTICE
Incorrect mounting or disassembly
The Memosens head could become loose and fall off, resulting in total sensor failure
Only mount the sensor via the process connection.
To do so, use a suitable tool, such as an open-ended wrench.
A0019019
 5 Installation in flow assembly CLA751
A0019020
 6 Installation in flow assembly CLA752
Page 16
Electrical connection Condumax CLS15D/16D/21D
16 Endress+Hauser
The Dipfit CLA111 immersion assembly is available to install sensors with a G1 thread in vessels.
A B
D
A0024145
 7 Installation in Dipfit CLA111 immersion assembly, fastening versions A, B and D
Ensure that the electrodes are fully immersed in the medium during measurement.
4.2 Post-installation check
• Are the sensor and cable undamaged?
• Is the sensor installed in the process connection and is not suspended from the cable?
5 Electrical connection
L
WARNING
Device is live
Incorrect connection may result in injury or death
The electrical connection may be performed only by an electrical technician.
The electrical technician must have read and understood these Operating Instructions and must follow the instructions contained therein.
Prior to commencing connection work, ensure that no voltage is present on any cable.
Page 17
Condumax CLS15D/16D/21D Electrical connection
Endress+Hauser 17
5.1 Connection conditions
5.1.1 Connection at a glance
Sensors for zone 0
IIC 2G
CPS11D CPS12D CPS71D CPS72D CPS91D CPS92D CPS16D CPS76D CPS96D CPS171D
IIC 1G
IIC 3G
EX
Zone 1
Zone 0
Zone 2
CYK10 Ex ia Ga
CPS41D CPS42D
CYK10 Ex ia Ga
EX
CLS50D
CLS15D
CLS16D
CLS21D
COS22D
COS51D
CPS471D CPS491D
CPS441D
CLS82D
CPS341D
A0031174
 8 Electrical connection in hazardous environment
Page 18
Electrical connection Condumax CLS15D/16D/21D
18 Endress+Hauser
Sensors for zone 2
CPS11D CPS12D CPS71D CPS72D CPS91D CPS92D CPS16D CPS76D CPS96D CPS171D
EX
CYK10 Ex ic
CPS41D CPS42D
CYK10 Ex ic
EX
CLS50D
CLS15D
CLS16D
CLS21D
COS22D
COS51D
CPS471D CPS491D
CPS441D
CLS82D
CPS341D
II 3G
Zone 2
A0031184
 9 Electrical connection in hazardous environment
Page 19
Condumax CLS15D/16D/21D Electrical connection
Endress+Hauser 19
5.2 Connecting the sensor
The electrical connection of the sensor to the transmitter is performed using the measuring cable CYK10.
GN/YE
YE
GN
BN
WH
GND
+
Com A
Com B
GY
A0024019
 10 Measuring cable CYK10
NOTICE
Mechanical twist protection for CLS15D and CLS21D
If too much force is applied to the Memosens head, this can shear the connections and therefore destroy the sensor
There is no need to exert excessive force when connecting the sensor to the cable coupling. Always connect the sensor gently.
If the Memosens coupling clearly will not close, check the coupling for dirt or mechanical damage and make sure you are turning in the right direction. Pay attention to the lock symbol on the coupling.
Use another Memosens cable if necessary.
5.3 Ensuring the degree of protection
Only the mechanical and electrical connections which are described in these instructions and which are necessary for the required, designated use, may be carried out on the device delivered.
Exercise care when carrying out the work.
Otherwise, the individual types of protection (Ingress Protection (IP), electrical safety, EMC interference immunity) agreed for this product can no longer be guaranteed due, for example, to covers being left off or cable (ends) which are loose or insufficiently secured.
5.4 Post-connection check
Device condition and specifications Notes
Are the outside of the sensor, assembly, cable undamaged? Visual inspection
Electrical connection Notes
Are the installed cables strain-relieved and not twisted?
Is a sufficient length of the cable cores stripped, and is it positioned in the terminal correctly?
Check the fit (by pulling gently)
Are all the screws terminals properly tightened? Tighten
Page 20
Commissioning Condumax CLS15D/16D/21D
20 Endress+Hauser
Device condition and specifications Notes
Are all cable entries mounted, tightened and leak-tight? For lateral cable entries, make sure the cables
loop downwards to allow water to drip off
Are all cable entries installed downwards or mounted laterally?
6 Commissioning
Before first commissioning, check if:
• the sensor is correctly installed
• the electrical connection is correct.
If using an assembly with automatic cleaning, check that the cleaning medium (e.g. water or air) is connected correctly.
L
WARNING
Escaping process medium
Risk of injury from high pressure, high temperatures or chemical hazards
Before applying compressed air to an assembly with cleaning facility, make sure the connections are correctly fitted.
Do not install the assembly in the process if you cannot make the correct connection reliably.
7 Maintenance
L
CAUTION
Corrosive chemicals
Danger of chemical burns to the eyes and skin. Danger of damage to clothing and equipment
It is absolutely essential to protect the eyes and hands properly when working with acids, bases and organic solvents!
Wear protective goggles and safety gloves.
Clean away splashes on clothes and other objects to prevent any damage.
Pay particular attention to the information provided in the safety data sheets for the chemicals used.
L
WARNING
Hydrofluoric acid and mineral acids
Risk of serious or fatal injury from caustic burns
Wear protective goggles to protect your eyes.
Wear protective gloves and appropriate protective clothing.
Avoid all contact with the eyes, mouth and skin.
If using hydrofluoric acid, only use plastic vessels.
Page 21
Condumax CLS15D/16D/21D Repair
Endress+Hauser 21
L
WARNING
Thiocarbamide
Harmful if swallowed. Limited evidence of carcinogenicity. Possible risk of harm to the unborn child. Dangerous for the environment with long-term effects.
Wear protective goggles, protective gloves and appropriate protective clothing.
Avoid all contact with the eyes, mouth and skin.
Avoid releases into the environment.
Clean away fouling on the sensor as follows depending on the type of fouling:
1. Oily and greasy films: Clean with grease remover, e.g. alcohol, as well as hot water and (alkaline) agents
containing surfactants (e.g. dishwashing detergent).
2. Lime, cyanide and metal hydroxide buildup and low solubility organic buildup: Dissolve buildup with diluted hydrochloric acid (3 %) and then rinse thoroughly with
plenty of clear water.
3. Sulfidic buildup (from flue gas desulfurization or sewage treatment plants): Use a mixture of hydrochloric acid (3 %) and thiocarbamide (commercially available)
and then rinse thoroughly with plenty of clear water.
4. Buildup containing proteins (e.g. food industry): Use a mixture of hydrochloric acid (0.5 %) and pepsin (commercially available) and then
rinse thoroughly with plenty of clear water.
5. Readily soluble biological buildup: Rinse with pressurized water.
After cleaning or regeneration, you must rinse the sensor thoroughly with water and then recalibrate it.
8 Repair
8.1 Sealing ring replacement and recalibration (CLS16D only)
Intact seals are a prerequisite for safe and reliable measurements. The seal should be replaced at regular intervals to guarantee maximum sensor operational safety and hygiene.
Practical repair intervals can only be determined by the user as they depend greatly on the operating conditions, such as:
• Type and temperature of the product
• Type and temperature of the cleaning agent
• Number of cleanings
• Number of sterilizations
• Operating environment
Page 22
Repair Condumax CLS15D/16D/21D
22 Endress+Hauser
Recommended intervals for seal replacement (reference values)
Application Window
Media with temperatures from 50 to 100 °C (122 to 212 °F) Approx. 18 months
Media with temperatures < 50 °C (122 °F) Approx. 36 months
Sterilization cycles, max. 150 °C (302 °F), 45 min. Approx. 400 cycles
To ensure your sensor is operational again after being exposed to very high loads, you can have it regenerated in the factory. In the factory, the sensor is fitted with new seals and recalibrated.
Please contact your sales office for information on replacing the seal and recalibration in the factory.
8.2 Return
The product must be returned if repairs or a factory calibration are required, or if the wrong product was ordered or delivered. As an ISO-certified company and also due to legal regulations, Endress+Hauser is obliged to follow certain procedures when handling any returned products that have been in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures and conditions at www.endress.com/support/return-material.
8.3 Disposal
The device contains electronic components and must therefore be disposed of in accordance with regulations on the disposal of electronic waste.
Observe the local regulations.
Page 23
Condumax CLS15D/16D/21D Technical data
Endress+Hauser 23
9 Technical data
9.1 Input
9.1.1 Measured values
• Conductivity
• Temperature
9.1.2 Measuring ranges
Conductivity (in relation to water at 25 °C (77 °F))
CLS15D-A 0.04 to 20 μS/cm CLS15D-B/L 0.10 to 200 μS/cm CLS16D 0.04 to 500 μS/cm CLS21D 10 μS/cm to 20 mS/cm
Temperature
CLS15D -20 to 100 °C (-4 to 212 °F) CLS16D -5 to 100 °C (23 to 212 °F) CLS21D -20 to 100 °C (-4 to 212 °F)
9.1.3 Cell constant
CLS15D-A
k = 0.01 cm
–1
CLS15D-B/L k = 0.1 cm
–1
CLS16D k = 0.1 cm
–1
CLS21D k = 1.0 cm–1, nominal
9.1.4 Temperature compensation
NTC 30K
Page 24
Technical data Condumax CLS15D/16D/21D
24 Endress+Hauser
9.2 Performance characteristics
9.2.1 Uncertainty of measurement CLS15D
Each individual sensor is factory-measured in a solution of approx. 5 μS/cm for cell constant
0.01 cm-1 or approx. 50 μS/cm for cell constant 0.1 cm-1 using a reference measuring system traceable to NIST or PTB. The exact cell constant is entered into the quality certificate supplied. The uncertainty of measurement in determining the cell constant is 1.0 %.
CLS16D
Each individual sensor is factory-measured in a solution of approx. 5 μS/cm using a reference measuring system traceable to NIST or PTB. The exact cell constant is entered into the quality certificate supplied. The uncertainty of measurement in determining the cell constant is
1.0 %.
CLS21D
Each individual sensor is factory-measured in a solution of approx. 5 mS/cmusing a reference measuring system traceable to NIST or PTB. The exact cell constant is entered into the quality certificate supplied. The uncertainty of measurement in determining the cell constant is
1.0 %.
9.2.2 Response time
Conductivity
t95 ≤ 3 s
Temperature
CLS15D-A t90 ≤ 39 s CLS15D-B/L t90 ≤ 17 s CLS16D t90 ≤ 13 s CLS21D t90 ≤ 296 s
9.2.3 Maximum measured error
CLS15D
2% of reading
CLS16D 2 % of reading up to 200 μS/cm
3 % of reading from 200 to 500 μS/cm
CLS21D 5% of reading
9.2.4 Repeatability
0.2% of reading
Page 25
Condumax CLS15D/16D/21D Technical data
Endress+Hauser 25
9.3 Environment
9.3.1 Ambient temperature range
-20 to +60 °C (-4 to 140 °F)
9.3.2 Storage temperature
-25 to +80 °C (-10 to +180 °F)
9.3.3 Degree of protection
IP 68 / NEMA type 6P (1 m water column, 25 °C, 24 h)
9.4 Process
9.4.1 Process temperature
CLS15D
Normal operation -20 to 120 °C (-4 to 248 °F) Sterilization (max. 1 h)
1)
Max. 140 °C (284 °F)
CLS16D
Normal operation -5 to 120 °C (23 to 248 °F) Sterilization (max. 45 min) Max. 150 °C (302 °F) at 6 bar (87 psi) absolute
CLS21D -20 to +135 °C (-4 to 275 °F) at 3.5 bar (50 psi)
absolute
1) Threaded versions: max. 30 minutes
The maximum temperature for communication with the transmitter is 130 °C (266°F) for the Memosens versions.
9.4.2 Process pressure
CLS15D
13 bar (188 psi) absolute, at 20 °C (68 °F) 2 bar (29 psi) absolute, at 120 °C (248 °F)
CLS16D 13 bar (188 psi) absolute, at 20 °C (68 °F)
9 bar (130 psi) absolute, at 120 °C (248 °F)
0.1 bar (1.5 psi) absolute (negative pressure), at 20 °C (68 °F)
CLS21D 17 bar (246 psi) absolute, at 20 °C (68 °F)
Page 26
Technical data Condumax CLS15D/16D/21D
26 Endress+Hauser
9.4.3 Temperature/pressure ratings
CLS15D
A
14010060
68
20
-4
180 220 260
248
284
14
43
72
188
[° ]F
–20
1
[bar]
3
5
7
9
11
13
0 20 40 60 80 100 120 140
[°C]
212
p (abs.)
T
A0031428-EN
 11 Mechanical pressure-temperature resistance
A Can be sterilized for a short time (1 hour)
CLS16D
6
–5
20 60 100 120 150
23
60
68
120
180
240
248
300
20 60 100 120 150
302
A
[° ]F
[bar]
[°C]
p (abs.)
T
14
43
72
101
130
159
188
1
3
5
7
9
11
13
87
1.5 0.1
A0031431-EN
 12 Mechanical pressure-temperature resistance
A Can be sterilized for a short time (45 min.)
Page 27
Condumax CLS15D/16D/21D Technical data
Endress+Hauser 27
CLS21D
5
9
13
17
300 60 90 120
[°C]
-4
1
10060 140 180 220
72
130
188
-20
14
[°F]
20
68 275
246
[psi]
[bar]
3.5
51
135
p (abs.)
T
20
A0031435-EN
 13 Mechanical pressure-temperature resistance
9.5 Mechanical construction
9.5.1 Weight
CLS15D and CLS21D
Approx. 0.3 kg (0.66 lbs) depending on version
CLS16D
Approx. 0.13 to 0.75 kg (0.29 to 1.65 lbs) depending on version
9.5.2 Materials
CLS15D
Electrodes Polished, stainless steel 1.4435 (AISI 316L) Sensor shaft Polyethersulfone (PES-GF20) O-ring, in contact with medium
(only Clamp version)
EPDM
CLS16D
Electrodes Electropolished, stainless steel 1.4435 (AISI 316L) Seal Gasket seal ISOLAST (FFKM), FDA-compliant
CLS21D
Electrodes Graphite Sensor shaft Polyethersulfone (PES-GF20) Thermal conductivity socket for
temperature probe
Titanium 3.7035
Page 28
Technical data Condumax CLS15D/16D/21D
28 Endress+Hauser
9.5.3 Process connection
CLS15D
Thread NPT ½" and ¾" Clamp 1½" as per ISO 2852
CLS16D
Clamp 1", 1½", 2" as per ISO 2852 (also suitable for TRI-CLAMP, DIN 32676) Tuchenhagen VARIVENT N DN 50 to 125 NEUMO BioControl D50
CLS21D
Thread G1 NPT 1" thread Clamp 2" as per ISO 2852 Sanitary connection DN 25 as per DIN 11851
9.5.4 Surface roughness (CLS15D, CLS16D only)
CLS15D
Ra ≤ 0.8 μm
CLS16D
Ra ≤ 0.8 μm, electropolished Ra ≤ 0.38 μm, electropolished, optional
Page 29
Condumax CLS15D/16D/21D EU Declaration of conformity
Endress+Hauser 29
10 EU Declaration of conformity
10.1 ATEX / NEPSI II 1G Ex ia IIC T3/T4/T6 Ga, IECEx Ex ia IIC T3/T4/T6 Ga
Page 30
EU Declaration of conformity Condumax CLS15D/16D/21D
30 Endress+Hauser
10.2 ATEX/NEPSI II 3G Ex ic IIC T3/T4/T6 Gc
Page 31
Page 32
www.addresses.endress.com
*71333350*
71333350