EmersonTM designs, manufactures, and tests its products to meet many national and international standards. Because these
instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure they continue to
operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program
when installing, using, and maintaining Emerson products. Failure to comply with instructions may result in: loss of life, personal
injury, property damage, damage to this instrument, and warranty invalidation.
•Read all instructions prior to installing, operating, and servicing the product.
•If you do not understand any of the instructions, contact your Emerson representative for clarification.
•Follow all warnings, cautions, and instructions marked on and supplied with the product, including this manual.
•To ensure proper performance, use only informed and qualified personnel to install, operate, update, program, and
maintain the product.
•Install equipment as specified in the installation instructions of the appropriate instruction manual and per applicable local
and national codes. Connect all products to the proper electrical and pressure sources.
•Use only factory documented components for repair. Tampering and unauthorized substitution and parts can affect
product performance and cause unsafe operation of your process.
WARNING!
RISK OF ELECTRICAL SHOCK
1.Disconnect the main power wire from the separate power source before servicing.
2.Do not operate or energize instrument with case open!
3.Wiring connected in this box must be rated at least 240 Vac and 14 gauge.
4.Non-metallic cable strain reliefs do not provide grounding between conduit connections! Use grounding type bushings and
jumper wires.
5.Seal unused cable conduit entries securely with non-flammable enclosures to provide transmitter's enclosure integrity in
compliance with personal safety and environmental protection requirements. Seal unused conduit openings with Type 4X or
IP66 conduit plugs to maintain the ingress protection rating (Type 4X) of transmitter's enclosure.
6.Operate only with front panel fastened and in place.
7.For safety and performance, connect and properly ground the system through a three wire power source.
8.Proper use and configuration is your responsibility.
9.Install all protective equipment covers and safety ground leads after installation. Failure to install covers and ground leads
could result in serious injury or death.
10.Provide local ground connection to the RosemountTM CX1100 in order to avoid electric shock. The ground conductor should
be equal to or greater than 16 AWG for a copper conductor in free air or 12 AWG for an enclosed copper conductor.
11.Disconnect and lock out power before connecting the power supply.
12.Install all protective covers and safety ground leads after installation. Failure to install covers and ground leads could result
in serious injury or death.
13.Disconnect and lock out power before working on any electrical components. There is voltage up to 240 Vac. Verify the
reliability and continuity of the grounding of the probe connection before powering the equipment.
14.To meet the Safety Requirements of IEC 61010-1 (EC requirement) and ensure safe operation of the equipment, make a
connection to the main electrical power supply through a circuit breaker (max 10A) which will disconnect all currentcarrying conductors during a fault situation. The circuit breaker should also include a mechanically operated isolating
switch. If it does not, then locate another means of disconnecting the supply from the equipment close by. Circuit breakers
or switches must comply with a recognized standard, such as IEC 60947.
15.Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed
by qualified people, to prevent electrical shock and personal injury.
CAUTION!
•If the ducts will be washed down during outage, make sure to power down the probes and remove them from the wash
area.
•When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson.
Unauthorized parts and procedures can affect the product's performance, place the safe operation of your product at risk,
and void your warranty. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
NOTICE
The information contained in this document is subject to change without notice.
Symbols used on equipment and in instruction manual
Earth (ground) terminal
Protective conductor terminal
Risk of electrical shock
Hot surface
Refer to reference manual.
Equipment protected throughout by double insulation
(applies to Rosemount CX1100 Remote Transmitter housing only)
NOTICE
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Contents
Contents
Chapter 1 General information .......................................................................................................1
A.4.1North America .............................................................................................................37
Reference Manuali
Contents
iiRosemount CX1100
1General information
1.1Overview
The Rosemount™ CX1100 In Situ Oxygen Transmitter provides a continuous, accurate
measurement of the oxygen remaining in flue gases coming from a combustion process.
Accurate measurements of furnace exhaust excess oxygen are critical for combustion
optimization, which result in reduced energy costs, increased safety, and lower emissions.
This product is intended for use for combustion applications only.
The Rosemount CX1100 probe contains a zirconia type oxygen sensor. At high
temperatures, zirconia produces an electrical potential that is proportional to the
difference in process oxygen concentration and ambient reference air. This voltage is
transmitted to the Rosemount CX1100 remote transmitter and output as an analog signal.
The sensor is mounted at the end of the probe tube which extends into a flue gas duct or
stack. Varying lengths of the probe are available to achieve desired insertion depths. The
probe tube also houses a heating element to precisely control the temperature of the
sensor, reducing errors induced by varying process temperatures.
General information
1.2Rosemount CX1100 Oxygen Probe and
Transmitter
A complete Rosemount CX1100 In Situ Oxygen Transmitter system includes a probe and
remote transmitter. Raw signals (sourced from the heater thermocouple and oxygen
sensor) and power are transmitted between the Rosemount CX1100 probe and the
Rosemount CX1000 remote transmitter, interconnect wiring consisting of nine
conductors. The specifications of the wiring arrangement can be found in Chapter 3. If
following the specifications, 200 ft (61 m) is the maximum cable length allowed.
78.74 in. (2 m)30.5 lb. (13.8 kg)23 lb. (10.43 kg)
Rosemount CX1100
Remote Transmitter
dimensions in inches
(millimeters)Shipping weightActual weightPart number
6.10 x 6.10 x 5.15 in.
4 lb. (1.8 kg)_3 lb. (1.4 kg)24490-00
(155 x 155 x 131 mm)
AccessoryPart number
Calibration flowmeter771B635H01
Cal gas adaptor kit to mate to existing tubing arrangement1A98771H07
6Rosemount CX1100
2Install
WARNING!
ELECTRIC SHOCK
Install all protective equipment covers and safety ground leads after installation. Failure to
install covers and ground leads could result in serious injury or death.
CAUTION!
EQUIPMENT DAMAGE
Do not allow the temperature of the Rosemount™ CX1100 Probe electronics to exceed 194 °F
(90 °C) or damage to the unit may result.
2.1Probe installation
Install
A weld plate for welding to the flue gas duct can be supplied for new installations.
1.If using the standard square weld plate (Figure 2-2) or an optional flange mounting
plate, weld or bolt the plate onto the duct.
The through hole diameter in the stack or duct wall and refractory material must be
at least 2.5 in. (63.5 mm).
2.Insert probe through the opening in the mounting flange and bolt through the
probe gasket and flange.
For horizontal installations, the breather port must be oriented such that it is facing
downward.
Reference Manual7
Install
Probe InstallationFigure 2-1:
A.M6 X 1 X 25 hex head machine screw (3X)
B.Calibration gas, 3/16-in. tube fitting, 2.82 liter/min (10 SCFH), 138 kPa (20 psi)
C.M5 X 0.8 x 10 pan head machine screw, ground screw
D.1/2 NPT conduit connection (power, signal)
E.Breather port
F.Calibration gas connection
NOTICE
All dimensions are in millimeters with inches in parentheses.
Removal/InstallationTable 2-1:
Probe lengthRemoval envelope
.5 m (19.68 in.)750 mm (30 in.)
1 m (39.37 in.)1250 mm (49 in.)
2 m (78.74 in.)2250 mm (89 in.)
8Rosemount CX1100
Rosemount CX1100 Probe InstallationFigure 2-2:
NOTICE
All dimensions are in millimeters with inches in parentheses.
Install
Square weld plate, ANSI pattern part 4512C34G01
Reference Manual9
Install
Mounting FlangeTable 2-2:
ANSIDIN
Flange dia185 mm (7.28 in.)
Hold dia19.1 mm (.75 in.)
4 holes eq sp on BC120.7 mm (4.75 in.)145 mm (5.71 in.)
Drip loop and Insulation RemovalFigure 2-3:
A.
Note
Replace insulation after installing CX1100.
B.Drip loop
C.Stack duct or metal wall
D.Adapter plate
E.Insulation
NOTICE
Probe installation may be vertical or horizontal.
2.2
10Rosemount CX1100
Mount remote display
Complete the following steps to install the Rosemount CX1100 Remote Transmitter.
Install
The Rosemount CX1100 Remote Transmitter is available in a panel mounting or wall/pipe
mounting configuration. Refer to Figure 2-4 or Figure 2-5 for the panel, wall, or pipe
mounting details. You need a wrench and bolts to mount the transmitter.
Procedure
1.Ensure all components are available to install the Rosemount CX1100 Remote
Transmitter.
2.Select a mounting location near or removed from the Rosemount CX1100 Probe.
Consider the temperature limitations of the Rosemount CX1100 Remote
Transmitter. Refer to product specifications when selecting the mounting location.
3.Mount at a height convenient for viewing and operating the interface.
Emerson recommends approximately 5 ft. (1.5 m).
4.The keypad window on the Rosemount CX1100 Remote Transmitter may have an
exterior protective membrane. Remove the protective membrane prior to use of the
Rosemount CX1100 Remote transmitter.
Failure to remove the protective membrane may cause the display to appear
distorted. The membrane may become difficult or impossible to remove after
extended use at elevated temperatures.
Reference Manual11
Install
Wall/Surface and Pipe MountFigure 2-4:
12Rosemount CX1100
Install
Panel MountFigure 2-5:
Reference Manual13
Install
14Rosemount CX1100
3Wire
All wiring must conform to local and national codes.
WARNING!
Before installing the equipment, read the Safety Instructions at the front of this manual. Failure
to follow safety instructions could result in serious injury or death.
NOTICE
To maintain proper earth grounding, ensure a positive connection exists between the probe
terminations housing and earth. The connecting wire must be 14 AWG minimum.
NOTICE
Line voltage, signal, and relay wiring must be rated for at least 221 °F (105 °C).
Wire
3.1
3.2
NOTICE
If metal conduit is used with the Rosemount™ CX1100 Remote Transmitter, reliably bond the
conduit to protective earth. The grounding plate inside the Rosemount CX1100 Remote
Transmitter is not bonded to PE and does not provide adequate grounding.
Installation specifications - interconnect cable
Customer supplied cable up to 200 ft. (60 m) long, rated for -40 to 194 °F (-40 to 90 °C)
and voltage 300 VACrms.
Heater power: 3 multi conductor 18ga shielded cable to connect the heater control signal
Probe sensing: 3 twisted pair conductors 22ga overall shielded cable to connect the TC,
O2, and CJC signals.
Cable may be purchased as two separate cables.
Connect the cables
Raw voltages from the oxygen sensor and heater thermocouple are transmitted from the
Rosemount CX1100 Probe to the Rosemount CX1100 Remote Transmitter. The remote
transmitter also controls power to the probe heater in order to maintain the correct sensor
temperature.
This arrangement calls for interconnect wiring consisting of nine conductors. Given the
recommended wire specifications, the maximum length for this cable is 200 ft (60 m)
(refer to Section 3.1).
Reference Manual15
Wire
NOTICE
To maintain EMC/EMI noise protection, connect the customer supplied 9 conductor cable and
cable glands properly to ground.
Procedure
1.Run the signal and power cables between the probe and the installation site for the
optional Rosemount CX1100 Remote Transmitter.
2.Remove the covers from the probe and the remote transmitter (if applicable).
3.Feed all probe wiring through the conduit port of the probe.
4.Refer to Figure 3-2. Connect probe heater power leads to probe connector.
5.Connect O2 signal, thermocouple, and cold junction (CJC) wires from probe to the
Rosemount CX1100 Remote Transmitter.
6.At the Rosemount CX1100 Remote Transmitter, connect the cable leads to the
connectors on the sensor board as indicated in Figure 3-1.
16Rosemount CX1100
Wire
Wiring DiagramFigure 3-1:
A. Heater power to probe
B. Alternate 4-20 mA out
C. Signal from probe
D. Power to remote transmitter
E. Alarm relay
Reference Manual17
Wire
F. 4-20 mA out
G. Power supply
H. Combustion sensor
I. Shield ground
J. Heater power terminates to underside of the sensor board.
K. Probe sensing connect shield to GND
L. S1 Dip switch is for factory use only and should be in the Off position.
AC power wires from power supply board to the underside of the sensor board are provided with
the sensor board.
18Rosemount CX1100
Wire
Closeup Wiring DiagramFigure 3-2:
A. M4 X 0.7 X 8 pan head screw machine screw (internal ground)
Reference Manual19
Wire
20Rosemount CX1100
4Startup
1.Apply AC line power to the Rosemount™ CX1100 Remote Transmitter.
The probe takes approximately 45 minutes to warm up to the 1357 °F (736 °C)
setpoint. The 4-20 mA signal remains at a default value of 3.5 mA, and the O
reading remains at 0% through the warmup period.
2.After warmup, the probe begins reading oxygen, and the 4-20 mA output is based
on the default range of 0-10% O2.
If there is an error condition at startup, an alarm message is displayed on the
Rosemount CX1100 Remote Transmitter.
Startup
2
Reference Manual21
Startup
22Rosemount CX1100
Calibration and maintenance
5Calibration and maintenance
The Rosemount™ CX1100 In Situ Oxygen Transmitter can be calibrated in the installed
position without removing the instrument from the process duct and also while the
combustion process is online.Gas is applied to the sensor through the calibration gas
fitting. New Rosemount CX1100 In Situ Oxygen Transmitter systems are factory calibrated
and are generally acceptable for initial startup and operation. High accuracy can be gained
by calibrating a system during normal operating conditions. Emerson ™recommends
calibrating on a semi-annual to annual basis for most applications; however actual
calibration frequency may vary per process unit.
Calibrations are conducted using a two point calibration. Factory calibration uses 0.4% O
and 8.0% O2 gases with a balance of nitrogen and is therefore recommended. Calibration
gases with other oxygen concentrations are acceptable. If using different calibration gas
values, additional configuration through the Rosemount CX1100 Remote Transmitter is
required. Emerson does not recommend pure nitrogen as a calibration gas.
5.1Procedure
The calibration of the Rosemount CX1100 Probe is initiated from the Rosemount CX1100
Remote Transmitter. The Rosemount CX1100 Remote Transmitter display prompts you
with the calibration instructions.
1.Press the x1 or x2 menu button on the Rosemount CX1100 Remote Transmitter
display to enter the probe.
2.Navigate, using Up and Down buttons to Setup; press Enter.
3.Navigate to Cal Gas values; press Enter.
4.Navigate to Cal Gas 1 and press Enter twice to highlight the value setting.
5.Use the Up and Down buttons to set Cal Gas 1 value and Left and Right buttons
to move to decimals setting.
6.After setting the correct value, press Enter and press Left to go to the Calibration
Setup for Cal Gas 2, Gas time, and Purge time settings.
2
7.After setting the values (0.4% O2, 8.0% O2, and 300s are the defaults), press Left to
go up to the Calibration submenu.
8.Navigate using the Up and Down buttons to Calibration and press Enter to start.
9.Follow the instructions on the display, manually applying calibration gases (0.4% and
8.0% O2 at 5 SCFH are recommended) to the calibration port.
10.After the calibration is finished, press Left and use the Up and Down buttons the
check the current and previous 10 calibrations log.
After completing the procedure, the Rosemount CX1100 software calculates new
calibration values and determines whether they meet an acceptance criteria. If successful,
the new calibration values automatically replace the previous values. In the event the
Reference Manual23
Calibration and maintenance
calibration values do not meet the accepted performance criteria, the existing calibration
values remain in effect, and Emerson recommends replacing the Rosemount CX1100
Probe.
EQUIPMENT DAMAGE
Make sure that the calibration gas cap is replaced tightly after calibration is complete. Many
combustion processes operate at a slight negative pressure (draft pressure) and can draw
ambient air down the cal gas lines and into the sensing cell, causing a false elevated oxygen
reading. The same phenomenon is possible if the calibration gas hoses become degraded or
loose.
CAUTION!
24Rosemount CX1100
5.2Calibration record
For Rosemount CX1100 In-Situ O2 Probe
Probe serial number:
Probe tag number:
Probe location:
Date placed into service:
Calibration and maintenance
NOTICE
Response
valve of the second calibration gas bottle.
Response
concentration after closing the valve of the second calibration gas bottle.
Reference Manual25
initial
: The time (sec) in which the oxygen reading returns to process oxygen
final
: The time (sec) to observe the initial change in oxygen reading after closing the
Calibration and maintenance
26Rosemount CX1100
1 Diagnostics
2 Setup
1 Active Alarms
2 Status
3 Maintenance
1 Revision
2 NV Memory Status
3 Reset
1 Version
2 Checksum
1 Language
2 Display
3 Security
1 Main Display
2 Revert Time
3 Set Contrast
4 Flash Alarm
1 Line 1
2 Line 2
3 Line 3
1 Center
1 Center
1 Left
2 Right
1 Password
2 Enable Password
1 Restart Counter
2 Reset Restart Count
3 Reset Main Board
4 Factory Defaults
5 Offline
Rosemount™ CX1100 In Situ Oxygen Transmitter menu trees
6Rosemount™ CX1100 In Situ Oxygen
Transmitter menu trees
6.1System menu tree
System Menu TreeFigure 6-1:
Reference Manual27
1 Process Variables
1 Process Variables
2 Temperatures
3 Raw Values
4 Analog Output
5 Maximums
6 Heater
1 O
2
Temp
2 CJC Temp
3 Board Temp
4 OP Mode
1 O
2
2 O
2
Temp
3 Cell Imp
4 O
2
Cell
1 O
2
Cell
2 Cell Imp
3 T/C Voltage
4 Heater
5 Update Cell Imp
1 O
2
AO
2 O
2
AO%
3 O
2
LRV
4 O
2
URV
1 Temperature
2 Voltage
3 Ramp Rate
1 O
2
Temp Max
2 CJC Temp Max
3 Board Temp Max
1 Heater Voltage Max
2 Cell Voltage Max
1 Duty Cycle
2 O
2
Temp SP
3 O
2
Temp
4 Heater Ramp Rate
1 Heater Ramp Rate Max
Rosemount™ CX1100 In Situ Oxygen Transmitter menu trees
6.2Probe menu tree
Probe Menu TreeFigure 6-2:
28Rosemount CX1100
2 Diagnostics
3 Setup
1 Active Alarms
2 Status
3 Acknowledge Alarms
4 Maintenance
1 Current
2 Previous
1 Failed
2 Maintenance
3 Advisory
1 Previous Alarms
1 Ack Previous Alarms
2 Ack Cal Failed
1 Revision
2 Analog Output
3 Operating Status
4 I/O States
5 NV Memory Status
6 Reset
1 Version
2 Checksum
1 Trim mA Output 1
1 Restart Counter
2 Reset Restart Count
3 Reset Sensor Board
1 Sensor
2 Analog Output
3 Relay
4 Calibration
5 Communication
1 T90 Filter
2 Low O
2
Alm SP
1 O
2
LRV
2 O
2
URV
3 AO Range
4 Alarm Level
1 Relay 1
2 Relay 2
1 AO Track
2 Cal Gas 1
3 Cal Gas 2
4 Gas Time
5 Purge Time
1 No Response
2 Bad Response
Rosemount™ CX1100 In Situ Oxygen Transmitter menu trees
Reference Manual29
Rosemount™ CX1100 In Situ Oxygen Transmitter menu trees
30Rosemount CX1100
7Parameters
7.1System and probe parameters
System ParametersTable 7-1:
Parameter labelDescription
Active AlarmsAlarms
StatusAlarm status bits
VersionMain board software version number
ChecksumMain board software checksum
Restart CounterMain board restarted counter
OfflineThe number of times the sensor board is discon-
nected from the main board since last power up
LanguageSupported languages
Line 1 | CenterMain display Line 1 data
Line 2 | CenterMain display Line 2 data
Line 3 | LeftMain display Line 3 left hand side data
Line 3 | RightMain display Line 3 right hand side data
Revert timeThe time set for the display to revert back to the
main display after the period of time with no
button press (2-9999 min)
ContrastDisplay contrast (110 - 200)
Flash AlarmTurn off/on the main display flashing when
T90 FilterAnalog output T90 time. It represents the time
to take a step change in oxygen to reach 90% of
the final value at the filter output. (0-300 sec)
Low O2 Alm SPLow O2 alarm threshold (0.0 - 50.0%)
32Rosemount CX1100
Parameters
Probe Parameters (continued)Table 7-2:
Parameter labelDescription
O2 LRVPrimary variable, O2%, lower range value. (0.0%
only)
O2 URVPrimary variable, O2% upper range value (1.0% -
50.0%)
AO RangeAnalog ouput polarity
Alarm LevelO2 alarm level
Relay 1Alarm Relay 1 mode
Relay 2Alarm Relay 2 mode
AO TrackAnalog output track O2 sensor measurement
during a calibration
Cal Gas 1Test Gas 1 value. This is the actual value of the
gas being applied during the Test Gas 1 phase
of a calibration. (0.05 - 50.0%)
Cal Gas 2Test Gas 2 value. this is the actual value of the
gas being applied during the Test Gas 2 phase
of calibration. (0.05 - 50.0%)
Gas TimeTest gas application time. This is the length of
time test gases are applied to the O2 probe during low or high test gas phase of a calibration.
(60 - 1200 sec)
Purge TimeTest gas purge time. This is the length of time
before the output will be returned to the process reading after a calibration. (60 - 1200 sec)
Imp Check HoldTime to hold analog output during impedance
check. (0 - 60 sec) (Factory mode only)
Remote CJCUse local or remote CJC measurement for cor-
rection. (No/Yes) (Factory mode only)
Raw Counts | Cell VoltageO2 sensor voltage raw counts. (Factory mode
only)
Raw Counts | Cell ImpCell impedance/sensor resistance raw counts.
(Factory mode only)
Raw Counts | T/C VoltageO2 T/C raw counts. (Factory mode only)
Raw Counts | Heater VoltageHeater voltage raw counts. (Factory mode only)
Cal StateThis represents the state of the calibration cycle
it is in.
Time RemainTime remaining in the present calibration cycle
state.
SlopeCurrent calibration slope. This is the slope value
that was calculated as a result of the last successful calibration. (34.5 to 57.5 mV/Dec)
Reference Manual33
Parameters
Probe Parameters (continued)Table 7-2:
Parameter labelDescription
ConstantCurrent calibration constant. This is the con-
stant value that was calculated as a result of the
last successful calibration. (-20.0 to 20.0 mV)
ImpedanceCell impedance. This is the sensor resistance
that was calculated as a result of the last successful calibration.
TimeTime stamp of the last successful calibration
Calibration Log | SlopePrevious calibration slope. There are ten calibra-
tion results. 1 is the most recent, and 10 is the
least recent calibration slope.
Calibration Log | ConstantPrevious calibration constant. there are ten cali-
bration results. 1 is the most recent, and 10 is
the least recent calibration constant.
Calibration Log | ImpedancePrevious cell impedance. This is the sensor re-
sistance that was calculated as a result of previous successful calibration. There are ten calibration results. Index 1 is the most recent, and
Index 10 is the least recent sensor resistance
measured.
Calibration Log | Log TimeTime stamp of the previous successful calibra-
tion. There are ten calibration time stamps. In-dex 1 is the most recent, and Index 10 is the
least recent time stamp.
Cal ResultCalibration result
7.2Rosemount™ CX1100 In Situ Oxygen
Transmitter system alarms
System AlarmsTable 7-3:
Alarm labelHelp messages and recommended actions
NV Memory FailChecksum error was detected in the nonvolatile memory
SB DisconnectCommunication failure detected between the sensor
34Rosemount CX1100
configuration data when the unit was turned on. Default
values have been loaded. Check to see that your configurations have not been changed. Cycle the power to clear
alarm.
board and the main board, indicating the sensor board has
been disconnected from the main board.Check the connecting cable pins.
Parameters
System Alarms (continued)Table 7-3:
Alarm labelHelp messages and recommended actions
SB MismatchThe sensor type does not match, indicating wrong sensor
board is connected.
SB VersionThe sensor board software version does not match the ex-
pected version, indicating the sensor board software needs
to be upgraded. Place the jumper on J1 of the sensor board
and cycle the power to start the sensor board software upgrade.
SB Bootload ModeThe Bootload Mode jumper at J1 of the sensor board is on
at power up. This mode should only be used to upgrade
the sensor board software. Remove the jumper from J1 of
the sensor board and cycle the power to clear the alarm.
Probe AlarmsTable 7-4:
Alarm labelHelp messages and recommended actions
O2 T/C ReversedThe sensor heater thermocouple voltage is reading a nega-
tive voltage, indicating the thermocouple wire connections may be reversed. Check wiring.
NV Memory FailChecksum error was detected in the nonvolatile memory
configuration data when the unit was turned on. Default
values have been loaded. Check to see that your configurations have not been changed. Cycle the power to clear
alarm.
Board Temp HighThe electronics board tempreature reading is above the
threshold. The board may not function correctly.
Factory ModeThe Factory Mode switch is set to On at power up. This
mode should only be uses for factory setup. Turn off the
Factory Mode switch and cycle the power to clear alarm.
Program IntegritySoftware checksum error detected when the unit was
turned on. Please replace sensor board.
O2 T/C ShortedThis diagnostic is only intended to detect a copper short of
the thermocouple connections at the electronics connector. The cell heater T/C voltage is reading close to zero, indicating the thermocouple wires may be shorted.
O2 Sensor OpenThe cell impedance voltage is reading out of normal range,
indicating the sensor wires may be disconnected or the
sensor junction may be open. Check wiring.
O2 T/C OpenThe sensor heater thermocouple voltage is out of range,
indicating the T/C wires may be disconnected, or the T/C
junction may be open. Check wiring.
O2 Temp LowThe sensor heater temperature is below the low tempera-
ture threshold.
Reference Manual35
Parameters
Probe Alarms (continued)Table 7-4:
Alarm labelHelp messages and recommended actions
O2 Temp HighThe sensor heater temperature is above the high tempera-
ture threshold.
RTD OpenThe CJC temperature is out of range, indicating the RTD
may be open. Check RTD wiring.
Cal FailedCalibration has failed. Check the Cal Result for more de-
tails. Acknowledge Calibration Failed or recalibrate
to clear alarm.
Low O
2
The O2 reading is below the threshold. The threshold is defined by the "Low O2 Alm SP".
36Rosemount CX1100
Rosemount™ CX1100 product certifications
Appendix A
Rosemount™ CX1100 product certifications
A.1European Directive information
A copy of the EC Declaration of Conformity can be found at the end of the Quick Start
Guide. The most recent revision of the EC Declaration of Conformity can be found at
Emerson.com/Rosemount.
A.2Ordinary location certification
As standard, the transmitter has been examined and tested to determine that the design
meets the basic electrical, mechanical, and fire protection requirements by a nationally
recognized test laboratory (NRTL) as accredited by the Federal Occupational Safety and
Health Administration (OSHA).
A.3Installing equipment in North America
The US National Electrical Code (NEC) and the Canadian Electrical Code (CEC) permit the
use of Division marked equipment in Zones and Zone marked equipment in Divisions. The
marking must be suitable for the area classification, gas, and temperature class. This
information is clearly defined in the respective codes.
A.4
A.4.1North America
Rosemount CX1100 In-Situ Oxygen Transmitter
CSA
Certificate: 70172073
Standards: CAN/CSA C22.2 No. 61010-1-12, CAN/CSA C22.2 No. 61010-2-010:15, UL
61010-1 (3rd Edition), UL 61010-2-10 (3rd Edition), UL 50E (2012), C22.2 No. 94.2-07, IEC
60529:2013 (Edition 2.2)
Markings: Type 4X, IP66
Reference Manual37
Rosemount™ CX1100 product certifications
Condition of acceptability:
1.The measuring of net O2 range is limited 0-23%.
2.The equipment shall be installed in accordance with manufacturer's specification by
qualified personnel.
3.This equipment is for permanently connection to power source with approved
power cord at end installation in accordance with local codes.
4.The Probe unit of the equipment shall be connected to PE separately at end
installation.
5.End installation to provide the means of disconnection from power sources.
6.End installation to comply with the requirement of IP66 and 4X.
38Rosemount CX1100
Rosemount™ CX1100 product certifications
Reference Manual39
Rosemount™ CX1100 product certifications
40Rosemount CX1100
Rosemount™ CX1100 product certifications
Reference Manual41
GLOBAL HEADQUARTERS
Emerson Automation Solutions
6021 Innovation Blvd
Shakopee, MN 55379, USA
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a
mark of one of the Emerson family of companies. All other marks are the property of their
respective owners.