Emerson Rosemount Analytical NGA2000 CLD Instruction Manual

Page 1
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
Chemiluminescence Analyzer Module
http://www.processanalytic.com
Page 2

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and in­ternational standards. Because these instruments are sophisticated technical products, you
MUST properly install, use, and maintain them
normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
to ensure they continue to operate within their
Read all instructions
prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical representative
for clarification.
Follow all warnings, cautions, and instructions
marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of
the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruc-
tion Manual and per applicable local and national codes. Connect all products to the proper elec-
trical and pressure sources.
To ensure proper performance, use qualified personnel
to install, operate, update, program, and
maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by
Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY
. Look-alike substitutions may result
in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Teflon is a Registered Trademark of E.I. duPont de Nemours and Co., Inc. Kynar is a Registered Trademark of Atochem North America, Inc. SNOOP is a registered trademark of NUPRO Co.
Emerson Process Management
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
Page 3
Model NGA2000 CLD

TABLE OF CONTENTS

PREFACE...........................................................................................................................................P-1
Definitions ...........................................................................................................................................P-1
Safety Summary .................................................................................................................................P-2
General Precautions For Handling And Storing High Pressure Gas Cylinders .................................P-6
Documentation....................................................................................................................................P-7
Compliances .......................................................................................................................................P-7
Glossary Of Terms ...........................................................................................................................P-8
1-0 DESCRIPTION AND SPECIFICATIONS..............................................................................1-1
1-1 Overview................................................................................................................................1-1
1-2 Typical Applications...............................................................................................................1-1
1-3 Theory of Technology............................................................................................................1-1
1-4 Features.................................................................................................................................1-1
1-5 Specifications.........................................................................................................................1-4
a. General ...........................................................................................................................1-4
b. Physical...........................................................................................................................1-4
c. Sample ............................................................................................................................1-4
d. Gas Connections.............................................................................................................1-5
Instruction Manual
760000-C
November 2002
2-0 INSTALLATION ....................................................................................................................2-1
2-1 Unpacking..............................................................................................................................2-1
2-2 Assembly ...............................................................................................................................2-1
2-3 Location .................................................................................................................................2-2
2-4 Gases ....................................................................................................................................2-2
a. Overview .........................................................................................................................2-2
b. Connections ....................................................................................................................2-2
c. Specifications ..................................................................................................................2-3
2-5 Electrical Connections ...........................................................................................................2-4
3-0 OPERATION .........................................................................................................................3-1
3-1 Overview................................................................................................................................3-1
3-2 Displays & Operating Keys....................................................................................................3-1
a. Menu Lines & Softkey Functionality................................................................................3-1
b. Common Function Keys..................................................................................................3-2
c. Entering & Changing Variables.......................................................................................3-3
d. Starting a Function..........................................................................................................3-3
e. Measure Mode Display ...................................................................................................3-4
f. Main Menu ......................................................................................................................3-4
3-3 Startup & Initialization............................................................................................................3-7
a. Routine Operation ...........................................................................................................3-7
3-4 Basic Controls and Status .....................................................................................................3-8
a. Analyzer Channel Status ................................................................................................3-8
b. Single Component Display..............................................................................................3-8
c. Multi Component Display................................................................................................3-9
3-5 Basic Controls........................................................................................................................3-10
Rosemount Analytical Inc. A Division of Emerson Process Management Contents i
Page 4
Instruction Manual
760000-C November 2002
3-6 Display Controls ....................................................................................................................3-11
3-7 Analyzer And I/O, Expert Controls And Setup.......................................................................3-12
a. Range Settings................................................................................................................3-12
b. Physical Measurements ..................................................................................................3-13
c. Concentration Alarms .....................................................................................................3-13
d. Linearization Parameters ................................................................................................3-14
e. Linearization Functions ...................................................................................................3-15
f. Response Time ...............................................................................................................3-17
g. Automatic Range Change...............................................................................................3-17
h. Display Units ...................................................................................................................3-18
i. Physical Measurements & Pressure Limits ....................................................................3-18
j. Single Component Display Parameters..........................................................................3-19
3-8 Calibration Procedure............................................................................................................3-21
a. Calibration Setup.............................................................................................................3-21
b. Basic Controls Calibration...............................................................................................3-22
c. Expert Controls Calibration.............................................................................................3-23
d. Unable to Calibrate .........................................................................................................3-25
3-9 System & Network I/O Module Controls (Setup) – System SIO...........................................3-26
a. Analog Output Setup.......................................................................................................3-26
b. Serial Interface Setup .....................................................................................................3-29
c. Relay Outputs Setup.......................................................................................................3-29
3-10 System & Network I/O Module Controls (Setup) – System DIO ..........................................3-31
3-11 System Configuration and Diagnostics .................................................................................3-32
a. Diagnostic Menus............................................................................................................3-32
b. Load/Save Module Configuration....................................................................................3-34
c. Date and Time.................................................................................................................3-35
d. Security Codes................................................................................................................3-35
e. System Reset..................................................................................................................3-36
3-12 Converter Temperature Adjustment ......................................................................................3-37
3-13 Measurement of Converter Efficiency ...................................................................................3-38
a. Test Setup for Measurement of Conversion Efficiency...................................................3-38
b. Test Procedure................................................................................................................3-38
c. Subnormal Conversion Efficiency...................................................................................3-39
d. Replacement of Converter ..............................................................................................3-40
e. Capillaries .......................................................................................................................3-40
f. TEA Scrubber..................................................................................................................3-40
Model NGA2000 CLD
4-0 MAINTENANCE AND SERVICE ..........................................................................................4-1
4-1 Overview................................................................................................................................4-1
4-2 Fuses .....................................................................................................................................4-1
4-3 Fans.......................................................................................................................................4-1
4-4 Converter ...............................................................................................................................4-1
4-5 Ozonator ................................................................................................................................4-1
4-6 Printed Circuit Boards............................................................................................................4-1
4-7 Detector Disassembly............................................................................................................4-3
a. Reaction Chamber Removal...........................................................................................4-3
b. Reaction Chamber Installation........................................................................................4-3
c. Photodiode Removal.......................................................................................................4-4
d. Photodiode Installation....................................................................................................4-4
5-0 TROUBLESHOOTING ..........................................................................................................5-1
5-1 Leaks .....................................................................................................................................5-1
ii Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Page 5
Model NGA2000 CLD
6-0 REPLACEMENT PARTS ......................................................................................................6-1
6-1 Matrix .....................................................................................................................................6-1
6-2 Replacement Parts ................................................................................................................6-1
7-0 RETURN OF MATERIAL......................................................................................................7-1
7-1 Return Of Material .................................................................................................................7-1
7-2 Customer Service ..................................................................................................................7-1
7-3 Training..................................................................................................................................7-1
8-0 INDEX....................................................................................................................................8-1
Instruction Manual
760000-C
November 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii
Page 6
Instruction Manual
760000-C November 2002
Figure 1-1. Flow Diagram - CLD Analyzer Module With Bypass Flow Sensor .................................1-2
Figure 1-2. Flow Diagram - CLD Analyzer Module Without Bypass Flow Sensor ...........................1-2
Figure 1-3. CLD Analyzer Module - Top View..................................................................................1-3
Figure 2-1. Analyzer Module Installation into Instrument Platform....................................................2-1
Figure 2-2. CLD Rear Panel Connections.........................................................................................2-3
Figure 2-3. Front Panel Controls and Electrical Connections ...........................................................2-4
Figure 2-4. CLD Wiring Diagram .......................................................................................................2-5
Figure 2-5. Outline and Mounting Dimensions..................................................................................2-6
Figure 3-1. Measure Mode Display ...................................................................................................3-1
Figure 3-2. The Display Screen.........................................................................................................3-3
Figure 3-3. Changing Variables.........................................................................................................3-3
Figure 3-4. Function Confirmation Display ........................................................................................3-3
Figure 3-5. Main Menu Display..........................................................................................................3-4
Figure 3-6. Module Manufacturing Data Displays .............................................................................3-5
Figure 3-7. Main Menu Sub Menus ...................................................................................................3-6
Figure 3-8. Startup Display................................................................................................................3-7
Figure 3-9. Basic Controls Menu .......................................................................................................3-10
Figure 3-10. Display Controls Menu ..................................................................................................3-11
Figure 3-11. Analyzer and I/O Expert Controls and Setup Menu......................................................3-12
Figure 3-12. Expert Controls Menu ...................................................................................................3-12
Figure 3-13. Range Settings Menu....................................................................................................3-12
Figure 3-14. Physical Measurements Menu......................................................................................3-13
Figure 3-15. Concentration Alarm Setup Menu.................................................................................3-13
Figure 3-16. Gas Measurement Parameters Menu...........................................................................3-14
Figure 3-17. Linearization Parameters Menu ....................................................................................3-14
Figure 3-18. Linearity Coefficients Menu...........................................................................................3-14
Figure 3-19. Linearization Functions Menu .......................................................................................3-15
Figure 3-20. Polynomial Setup Menu ................................................................................................3-15
Figure 3-21. Gas Concentrations Menu ............................................................................................3-15
Figure 3-22. Midpoint Correction Setup Menu ..................................................................................3-16
Figure 3-23. Response time/delay Parameters................................................................................3-17
Figure 3-24. Automatic Range Control Menu....................................................................................3-17
Figure 3-25. Actual Switch Levels Menu ...........................................................................................3-18
Figure 3-26. Display Units Menu .......................................................................................................3-18
Figure 3-27. Pressure Limits Menu ...................................................................................................3-19
Figure 3-28. Temperature Limits Menu .............................................................................................3-19
Figure 3-29. Physical Measurements – Manufacturer’s settings Display..........................................3-19
Figure 3-30. Displayed Parameters Menu.........................................................................................3-19
Figure 3-31. Calibration Gas List Menu.............................................................................................3-21
Figure 3-32. Calibration Parameters Display ....................................................................................3-22
Figure 3-33. Zero/Span Calibration Menu .........................................................................................3-23
Figure 3-34. Analyzer Zero Display...................................................................................................3-23
Figure 3-35. Zero/Span Diagnostic Data Menu.................................................................................3-24
Figure 3-36. Calibration Factors Menu..............................................................................................3-24
Figure 3-37. Range Factors Menu.....................................................................................................3-24
Figure 3-38. Range Factors -Manufacturer’s/Stored Settings Display..............................................3-25
Figure 3-39. System & Network I/O Module Controls Menu .............................................................3-26
Figure 3-40. System SIO Module Menu ............................................................................................3-26
Figure 3-41. Analyzer Modules Menu ...............................................................................................3-26
Model NGA2000 CLD

LIST OF ILLUSTRATIONS

iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Page 7
Model NGA2000 CLD
Figure 3-42. Analog Output Setup Menu...........................................................................................3-27
Figure 3-43. Signals Menu ...............................................................................................................3-27
Figure 3-44. Output Signal If Assigned Module Fails Menu..............................................................3-28
Figure 3-45. Special Scaling for Concentration Signal Menu............................................................3-28
Figure 3-46. Analog Output Updates per Second Menu ...................................................................3-29
Figure 3-47. Serial Interface Setup Menu .........................................................................................3-29
Figure 3-48. AK Protocol Definitions Menu .......................................................................................3-29
Figure 3-49. Relay Outputs Setup Menu...........................................................................................3-30
Figure 3-50. Choose Source Module Menu ......................................................................................3-30
Figure 3-51. Choose Signal Menu.....................................................................................................3-30
Figure 3-52. System DIO Module Menu............................................................................................3-31
Figure 3-53. System Configuration and Diagnostics Menu...............................................................3-32
Figure 3-54. Diagnostic Menu ...........................................................................................................3-32
Figure 3-55. Analyzer Diagnostics Menu ..........................................................................................3-32
Figure 3-56. Power Supply Voltages Menu.......................................................................................3-33
Figure 3-57. Primary Variable Parameters Menu..............................................................................3-33
Figure 3-58. Temperature Control Menu (Screen 1) .........................................................................3-33
Figure 3-59. Temperature Control Menu (Screen 2) .........................................................................3-33
Figure 3-60. NO/NOx Flow Balance Menu........................................................................................3-34
Figure 3-61. Calculate Factor Using Pressure Ratio Menu...............................................................3-34
Figure 3-62. Load/Save Configuration (CM/MCA) Menu ..................................................................3-34
Figure 3-63. Date and Time Menu.....................................................................................................3-35
Figure 3-64. Security Codes Menu....................................................................................................3-35
Figure 3-65. Define Basic Level Security PIN Menu .........................................................................3-35
Figure 3-66. System Reset Menu......................................................................................................3-36
Figure 3-67. Measuring Efficiency of NO
Figure 3-68. Conversion Efficiency as a Function of Converter Temperature ..................................3-42
Figure 4-1. CLD Module Assembly....................................................................................................4-2
Figure 4-2. Converter Assembly........................................................................................................4-3
Figure 4-3. Detector Assembly ..........................................................................................................4-5
Instruction Manual
760000-C
November 2002
to NO Converter.............................................................3-41
2

LIST OF TABLES

Table 2-1. Gas Specifications...................................................................................................... 2-2
Rosemount Analytical Inc. A Division of Emerson Process Management Contents v
Page 8
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
vi Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Page 9
Instruction Manual
Model NGA2000 CLD
November 2002

PREFACE

The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the NGA 2000 CLD and the System Accessories of the NGA 2000 System.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.

DEFINITIONS

The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout this publication.
760000-C
DANGER .
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure, condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Page 10
Instruction Manual
760000-C November 2002
Model NGA2000 CLD

SAFETY SUMMARY

If this equipment is used in a manner not specified in these instructions, protective systems may be impaired.

AUTHORIZED PERSONNEL

To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the this equipment should be thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE IN­STRUCTIONS.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel. For safety and proper per­formance this instrument must be connected to a properly grounded three-wire source of power.
WARNING .
TOXIC AND OXIDIZING GAS
This module generates ozone which is toxic by inhalation and is a strong irritant to throat and lungs. Ozone is also a strong oxidizing agent. Its presence is detected by a characteristic pungent odor.
The module's exhaust contains both ozone and nitrogen dioxide, both toxic by inhalation, and may contain other constituents of the sample gas which may be toxic. Such gases include various ox­ides of nitrogen, unburned hydrocarbons, carbon monoxide and other products of combustion re­actions. Carbon monoxide is highly toxic and can cause headache, nausea, loss of consciousness, and death.
Avoid inhalation of the ozone produced within the module, and avoid inhalation of the sample and exhaust products transported within the module. Avoid inhalation of the combined exhaust prod­ucts at the exhaust fitting.
Keep all tube fittings tight to avoid leaks The user is responsible for testing for leakage only at the inlet and outlet fittings on the rear panel. Periodically, the user should do an internal leak test (with a test procedure chosen by the user).
Connect rear exhaust outlet to outside vent with stainless steel or Teflon line. Check vent line and connections for leakage.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Page 11
Instruction Manual
Model NGA2000 CLD
WARNING.
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair.
WARNING.
OVERBALANCE HAZARD
This analyzer module may tip instrument over if it is pulled out too far and the Platform is not prop­erly supported.
WARNING.
INTERNAL ULTRAVIOLET LIGHT
Ultraviolet light from the ozone generator can cause permanent eye damage. Do not look directly at the ultraviolet source. Use of ultraviolet filtering glasses is recommended.
November 2002
760000-C
WARNING.
TOXIC CHEMICAL HAZARD
The ozone generator lamp contains mercury. Lamp breakage could result in mercury exposure. Mercury is highly toxic if absorbed through the skin or ingested, or if vapors are inhaled.
Handle lamp assembly with extreme care. If the lamp is broken, avoid skin contact and inhalation in the area of the lamp or the mercury spill.
Immediately clean up and dispose of the mercury spill and lamp residue as follows:
Wearing rubber gloves and goggles, collect all droplets of mercury by means of a suction pump
and aspirator bottle with a long capillary tube. (Alternately, a commercially available mercury spill clean-up kit, such as J.T. Baker product No. 4439-01, is recommended.)
Carefully sweep any remaining mercury and lamp debris into a dust pan. Carefully transfer all
mercury, lamp residue and debris into a plastic bottle which cab be tightly capped.
Label and return to hazardous material reclamation center. Do not place in the trash, incinerate
or flush down the sewer.
Cover any fine droplets of mercury in non-accessible crevices with calcium polysulfide and sul-
fur dust
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Page 12
Instruction Manual
760000-C November 2002
WARNING.
HAND INJURY HAZARD
Do not place hands or fingers in the Platform front handles when front panel is open. Dropping the front panel of the Platform while hand or fingers are inside either handle can cause serious injury.
WARNING.
POSSIBLE EXPLOSION HAZARD
Ensure that all gas connections are made as labeled and are leak free. Improper gas connections could result in explosion or death.
Model NGA2000 CLD
WARNING.
POSSIBLE EXPLOSION HAZARD
This equipment is not designed and should not be used in the analysis of flammable samples. Use of this equipment in this way could result in explosion or death.
CAUTION
PRESSURIZED GAS
This module requires periodic use of pressurized gas. See General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-6
CAUTION
EXTERNALLY RESTRICT SAMPLE FLOW TO LESS THAN 2200 CC/MIN.
No restrictor is provided in the sample inlet of this module. For those users who cannot externally limit sample flow, accessory restrictor PN 657300 is available from Rosemount Analytical.
To obtain this accessory, call 800-433-6076.
CAUTION
OVER-VOLTAGE SPIKING
If this analyzer module is used with a non-Rosemount Analytical power supply, adding Rosemount Analytical PN 903341 Current Protector in series with the 24V positive power line will prevent over­voltage spiking and resultant fuse blowing when powering up the instrument.
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Page 13
Instruction Manual
Model NGA2000 CLD
NOTICE
Software compatibility is necessary for all NGA 2000 components in your system to work to­gether. The version of your Platform's software must be equal to or greater that the version of any other module(s) for successful compatibility. If it is not, contact Rosemount Analytical at 800-433-6076 to order software upgrade kit PN 657150 for the Platform.
You can locate the version of each NGA 2000 component as follows:
Platform Controller Board Turn power ON.
The display will show "Control Module V2. ...". This is the software version.
Analyzer Module Located on the right side of the Analyzer Module case.
I/O Module Located on the backplane connector of the module. If no label is present, the module is Ver­sion 2.0.
760000-C
November 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Page 14
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981
Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when lo­cated in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suit­able hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is at­tached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125 permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precau­tions must be taken to prevent striking an arc against the cylinder.
°
F (52°C). A flame should never be
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Page 15
Instruction Manual
9
6
Model NGA2000 CLD
November 2002

DOCUMENTATION

The following NGA 2000 CLD instruction materials are available. Contact Customer Service Center or the local representative to order.
760000 Instruction Manual (this document)

COMPLIANCES

This product may carry approvals from several certifying agencies, including Factory Mutual and the Cana­dian Standards Association (which is also an OSHA accredited, Nationally Recognized Testing Labora­tory), for use in non-hazardous, indoor locations.
760000-C
Rosemount Analytical Inc. has satisfied all obligations from the European Legislation to harmonize the product requirements in Europe.
This product complies with the standard level of NAMUR EMC. Recommendation (May 1993).
This product satisfies all obligations of all relevant standards of the EMC framework in Australia and New Zealand.
FM
APPROVED
NAMUR
NRTL /C
®
97-C219
N
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-7
Page 16
Instruction Manual
760000-C November 2002
Model NGA2000 CLD

GLOSSARY OF TERMS

Analyzer Module
The module that contains all sensor/detector components for development of a Primary Variable signal; in­cludes all signal conditioning and temperature control circuitry.
Backplane
The interconnect circuit board which the Controller Board, Power Supply, Analyzer Module power and net­work cables, I/O Modules and Expansion Modules plug into.
Control Module
The Operator Interface plus the Controller Board.
Controller Board
The computer board that serves as the Network Manager and operates the Display and Keypad.
Distribution Assembly
The Backplane and the card cages that hold I/O and Expansion Modules.
Expansion Module
A circuit board that plugs into the Backplane from the front of the Platform and performs special features not related to I/O functions.
I/O Module
A circuit board that plugs into the Backplane from the rear of the Platform. Has a connector terminal for communication with external data acquisition devices and provides an input/output function.
Operator Interface
The Display and Keyboard.
Platform
Any workable collection of the following: Controller Board, Power Supply, Distribution Assembly, Enclosure and Operator Interface.
Power Supply
Any of a variety of components that provides conditioned power to other NGA 2000 components, from the Power Supply Board that plugs into the front of the Backplane in a stand-alone instrument to several larger ones that can power larger collections of modules and components.
Primary Variable
The measured species concentration value from an Analyzer Module.
Secondary Variable
Data placed on the network by a module regarding current status, e.g., sample flow, source voltage and other diagnostic information.
P-8 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Page 17
Instruction Manual
Model NGA2000 CLD
Softkeys
The five function softkeys located below the front panel display; they assume the function displayed directly above each on the display, a function dictated by software.
System
Any collection of Analyzer Module(s), Platform(s), I/O Module(s) and Expansion Module(s).
November 2002
760000-C
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-9
Page 18
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
P-10 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Page 19
Model NGA2000 CLD
DESCRIPTION AND SPECIFICATIONS

1-1 OVERVIEW

This manual describes the Chemiluminescence (CLD) Analyzer Module of Rosemount Analyti­cal's NGA 2000 Series of gas analysis compo­nents (See Figure 1-1, Figure 1-2 page 1-2 and Figure 1-3, page 1-3).
SECTION 1
NO tronically excited state (NO the excitation)). These revert immediately to the ground state, with emission of photons (essen­tially, red light). The reactions involved are:
Instruction Manual
760000-C
November 2002
molecules thus produced are in an elec-
2
* - the "*" refers to
2
NO + O
→ NO2* + O2
3
The CLD Analyzer Module is designed to con­tinuously determine the concentration of nitric oxide and oxides of nitrogen (NO plus nitrogen dioxide [NO
]) in a flowing gaseous mixture. The
2
concentration is expressed in parts-per-million.
The CLD Analyzer Module is designed as a slide-in module (if configured in stand-alone in­strument fashion), removable from the front of the Platform, with gas connections made from the rear. All electronics relative to sample detec­tion and conditioning are included in this mod­ule.

1-2 TYPICAL APPLICATIONS

The CLD Analyzer Module has specific applica­tions in the following areas:
Oxides of nitrogen emissions from the
combustion of fossil fuels in:
Vehicle engine exhaust Incinerators Boilers Gas appliances Turbine exhaust
Nitric acid plant emissions
NO
* NO2 + red light
2
As NO and O
mix in the reaction chamber, the
3
intensity of the emitted red light is measured by a photodiode and is proportional to the concen­tration of NO in the original gas sample.
To measure NOx (NO + NO
), any NO2 in the
2
sample is reduced to NO (at <95% efficiency) by being continuously passed through a heated bed of vitreous carbon (this occurs before the sample gas is presented to the ozone). Any NO initially present in the sample passes through this converter stage unchanged before being routed to the reaction chamber.
The photodiode generates a DC current, which is then amplified, conditioned and expressed on the network as the Primary Variable.

1-4 FEATURES

Among the features included in the CLD Ana­lyzer Module are: 1) ozonator air loss shutoff and 2) NO/NOx mode capability.
De-NO
control system
X
Nitric oxide emissions from decaying or-
ganic material (e.g., landfills).

1-3 THEORY OF TECHNOLOGY

The CLD Analyzer Module uses the chemilumi­nescence method of detection. This technology is based on the reaction of NO with ozone (O to produce NO
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
and oxygen (O
2
). Some of the
2
)
3
Page 20
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
BULKHEAD REDUCER
1/4T - 1/8T BRS
OZONE AIR
SAMPLE
BULKHEAD REDUCER 1/4T - 1/8T SS
EXHAUST
BULKHEAD
1/4T SS
BULKHEAD REDUCER
CONNECTOR 1/4T SS
1/4T - 1/8T BRS
OZONE AIR
SAMPLE
BULKHEAD REDUCER 1/4T - 1/8T SS
OZONE GENERATOR
659494
CONNECTOR,GLASS BLUE
VENT CAPILLARY
70cc/min 4 PSIG
634398
PUMP CAPILLARY
1/8 OD TEFLON
657473
SAMPLE PRESSURE SENSOR 15 PSIG 655253
KYNAR
ELBOW 1/8T SS
1/8 OD SS
TEE 1/8T
ELBOW 1/8T SS
1/8 OD SS
2 70 cc/min. 5 psig CAPILLARY: PN 659657 200 cc/min. 2 psig CAPILLARY: PN 659658
CROSS
1/8T BRS
MALE CONNECTOR
1/8T-3/8NPT B RS
FLOW BALANCE VALVE 903207
TEE ASSEMBLY
1/8T-1/4MPT
1/4T - 1/8FP T SS
TEE
CONNECTOR
1/8T-1/8NPT BRS
MALE CONNECTOR 1/8T-3/8NPT BR S
PRESSURE SENSOR 30 PSIG AIR 655254
ELBOW 1/8T-1/8MPT KYNAR
1
BACK
PRESSURE
REGULATOR
REDUCER 1/4T - 1/8T SS
ELBOW 1/8T - 1/8MPT SS
.116 ID VITON BLK
PRESSURE SWITCH 662298
BYPASS
FLOW
SENSOR
902931
ELBOW 1/8T-1/4MPT SS
AIR RESTRICTOR 430cc/min @ 12 PSIG 658157
CONNECTOR,
GLASS BLUE
SOLENOID
VALVE
NO/NOx
655263
ELBOW 1/8T-1/8MPT KYNAR
TEE 1/8T KYNAR
NOx TO NO
CONVERTER
655250
TOP
NC NO
C
TEE 1/8T KYNAR
CROSS 1/8T SS
CONNECTOR, MALE 1/8T-3/8MPT SS
1 BRASS/NEOPRENE 655269 STAINLESS STEEL/VITON 659063
Figure 1-1. Flow Diagram - CLD Analyzer Module With Bypass Flow Sensor
OZONE GENERATOR
659494
CONNECTOR,GLASS BLUE
VENT CAPILLARY
70cc/min 4 PSIG
634398
PUMP CAPILLARY
2 PSIG
SAMPLE PRESSURE SENSOR 15 PSIG 655253
TEE 1/8T KYNAR
TEE 1/8T
KYNAR
UNION 1/8T S S
UNION 1/4T- 1/8T SS
REDUCING UNION 1/8T - 1/16T S S
CROSS
1/8T BRS
MALE CONNECTOR
1/8T-3/8NPT B RS
FLOW BALANCE VALVE 903207
TEE ASSEMBLY
1/8T-1/4MPT
CONNECTOR
1/8T-1/8NPT BRS
MALE CONNE CTOR 1/8T-3/8NPT BRS
PRESSURE SWITCH 662298
PRESSURE SENSOR 30 PSIG AIR 655254
BACK
PRESSURE
REGULATOR
EXHAUST
BULKHEAD
1/4T SS
.116 ID VITON BLK
ELBOW 1/8T-1/4MP T SS
AIR RESTRICTOR 430cc/mi n @ 12 PSIG 658157
CONNECTOR,
GLASS BLUE
NO/NOx
SOLENOID
VALVE
655263
TEE 1/8T SS
NOx TO NO
CONVERTER
TOP
NC NO
C
655250
TEE 1/8T KYNAR
CROSS 1/8T SS
CONNECTOR, MALE 1/8T-3/8MPT SS
Figure 1-2. Flow Diagram - CLD Analyzer Module Without Bypass Flow Sensor
REACTION CHAMBER
2
REDUCING UNION 1/8T - 1/16T SS
1/8 OD SS
1/8 OD SS
SAMPLE CAPILLARY
OZONE
EXHAUST
SAMPLE
DETECTOR
REACTION CHAMBER
OZONE
EXHAUST
SAMPLE
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Page 21
Model NGA2000 CLD
Instruction Manual
760000-C
November 2002
COMPUTER BOARD
DRIVER BOARD
OZONE GENERATOR
SIGNAL BOARD
INLET/OUTLET
FITTINGS
FANS
NO2 TO NO CONVERTER
3-WAY SOLENOID VALVE
DETECTOR
SAMPLE CAPILLARY
FLOW SENSOR
POWER SUPPLY BOARD
FUSE
FLOW BALANCE NEEDLE VALVE
NETWORK/POWER CONNECTOR
SAMPLE BACKPRESSURE REGULATOR
Figure 1-3. CLD Analyzer Module - Top View
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
Page 22
Instruction Manual
760000-C November 2002

1-5 SPECIFICATIONS

a. General

Measurement Species:.................. NO, NOx
Ranges: ......................................... 0 to 10 ppm through 0 to 10,000 ppm NO, NOx
Repeatability:................................. ±0.5% of fullscale (at constant temperature)
Min. Detectable Level:................... 0.1 ppm
Noise: ............................................ <1% of fullscale, peak to peak
Linearity: ........................................ ±1% of fullscale
Response Time: ............................ <1 sec. for 90% of fullscale for ranges of 25 ppm or greater
....................................................... <3 sec. for 90% of fullscale for ranges of less than 25 ppm
t90 Response ................................ 1-2 sec.
Zero Drift: (at constant temperature) <±1% of fullscale/24 hours, <±2% of fullscale/week Span Drift: (at constant temperature) <±1% of fullscale/24 hours, <±3% of fullscale/week
Effect of Temperature:................... <2% of fullscale (over any 10°C ambient temperature variation
Ambient Temperature:................... 0°C to 45°C (32°F to 113°F)
Power Requirements:.................... 24V DC 120W
Model NGA2000 CLD
interval for a rate of change no greater than 10°C/hour)

b. Physical

Case Classification:....................... General purpose for installation in weather-protected area
Dimensions:................................... See Figure 2-5. Outline and Mounting Dimensions p2-6
Weight: .......................................... 8.1 kg (18 lbs.)
Mounting:....................................... Inside a Platform or custom-installed in a panel
Max. Length of LON Cable:........... 1600m (1 mile) between Analyzer Module and Platform

c. Sample

Temperature: ................................ 0°C to 45°C (32°F to 131°F)
Total Flow Rate: .......................... (Externally measured) 900 to 2200 cc/min. with backpressure
regulator pressure adjusted to 344 hPa (5 psig)
Particles: ...................................... Filtered to <2 microns
Dewpoint: ..................................... 5.5°C below ambient temperature, no entrained liquid
Materials in contact with Sample:.. Stainless steel, Teflon, glass, brass and neoprene
Optional: ..................... Stainless steel, Teflon, glass and Kynar
Ozonator Gas: ............................... Clean, dry air or oxygen; flow rate: 1 L/min. maximum; pressure:
689 hPa to 1034 hPa-gauge (10 to 15 psig); maintain a constant pressure ±34 hPa (±0.5 psig)

d. Gas Connections

Ozone Air:...................................... 1/4” O.D. tube fitting, brass
Exhaust:......................................... 1/4” O.D. tube fitting, stainless steel
Sample In: ..................................... 1/4” O.D. tube fitting, stainless steel
See the Preface section of the Platform manual for specifications regarding Platform-related components (e.g., case dimensions) and the I/O Module manual for specifications regarding I/O (e.g., relay outputs).
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Page 23
Model NGA2000 CLD
A
Instruction Manual
760000-C
November 2002
SECTION 2

INSTALLATION

WARNING.
Before starting to install this equipment, read the “Essential Instructions” on the in­side cover and the Safety Summary begin­ning on page P-2. Failure to follow the safety instructions could result in serious injury or death.

2-1 UNPACKING

If the Chemiluminescence (CLD) Analyzer Mod­ule is received as a separate unit, carefully ex­amine the shipping carton and contents for signs of damage. Immediately notify the ship­ping carrier if the carton or contents is dam­aged. Retain the carton and packing material until all components associated with the CLD Analyzer Module are operational.

2-2 ASSEMBLY

If the CLD Analyzer Module requires assembly with other components (e.g., the Platform and associated I/O Modules), do so at this time. Fol­lowing the guides on the bottom left and bottom center of the Platform, carefully slide the Ana­lyzer Module halfway into place.
CAUTION.
HAND INJURY HAZARD
Do not place hands or fingers in Platform front handles when the front panel is open. Dropping front panel while hand or fingers are inside either handle can cause serious injury.
Lift the spring-loaded pins on the front of the Analyzer Module, and carefully slide it the rest of the distance. Secure the module in position by releasing the pins, which seat in the available holes in the bottom of the case (see Figure 2-1, below). If the module and Platform are difficult to assemble, remove the module, ensure the top cover of the module is firmly seated on the hold-down screws, and repeat the assembly procedure.
Refer to Section 2-5, page 2-4 for electrical connections.
Install I/O Module(s) according to guidelines in the I/O manual. After startup and calibration have been performed, secure the front panel with the six screws provided.
NALYZER MODULE GUIDES
PIN SEATS
DISENGAGED FRO NT PANEL
Figure 2-1. Analyzer Module Installation into Instrument Platform
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Page 24
Instruction Manual
760000-C November 2002
Model NGA2000 CLD

2-3 LOCATION

Install the CLD Analyzer Module in a clean, weather-protected, vibration-free location free from extreme temperature variations and mois­ture. For best results, install the instrument near the sample stream to minimize sample transport time.
The analyzer should be installed within ±15° of horizontal.
NOTE
Unrestricted air flow to the rear of the Ana­lyzer Module is critical to its performance and reliability.
Operating ambient temperature is 0
o
(32
F to 113oF). Temperature change should
not exceed 10
o
C (18oF) per hour. The same temperature range restrictions apply to the loca­tion of the air and span gas cylinders.
o
C to 45oC

2-4 GASES

a. Overview

See Table 2-1, page 2-2. The CLD Analyzer Module requires two cylinder gases: 1) breathing grade air or oxygen and 2) a standard of accurately known composition for use as a span gas.
Each gas should be supplied from a cylin­der equipped with a clean, non-corrosive, two-stage regulator. A shutoff valve is rec­ommended.

b. Connections

(See Figure 2-2, page 2-3) Connect inlet and outlet lines for sample, air, and exhaust to appropriately labeled fittings on the rear panel, each of which is a 1/4-inch fer­rule-type compression fitting.
Stainless steel or Teflon tubing is recom­mended for the sample line. Exhaust tubing should be 1/4 inch (6.3 mm) or larger, and made of stainless steel or Teflon.
MAXIMUM NOx
LEVEL IN PARTS
PER MILLION
GAS SUPPLIED
TO REAR PANEL
AIR INLET
800 Air
2,500 Air
10,000 Oxygen
SETTING ON OZONE
PRESSURE GAUGE
(pressure values:
gauge)
690 hPa (10 psig); pro­vides flow of approxi­mately 500 cc/min. to ozone generator 690 hPa (10 psig); pro­vides flow of approxi­mately 1000 cc/min. to ozone generator. 1035 hPa (15 psig); pro­vides flow of approxi­mately 1000 cc/min to ozone generator.
SETTING ON SAMPLE
PRESSURE GAUGE
(pressure values:
gauge)
344 hPa (5 psig); pro­vides flow of approxi­mately 200 cc/min. to reaction chamber 344 hPa (5 psig); pro­vides flow of approxi­mately 70 cc/min to reaction chamber. 103 hPa (1.5 psig); pro­vides flow of approxi­mately 20 cc/min. to reaction chamber.
CAUTION.
Externally Limit Sample Flow Rate To Less Than 2200 Cc/Min.
Table 2-1. Gas Specifications
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 25
Model NGA2000 CLD
Instruction Manual
760000-C
November 2002

c. Specifications

Calibration Gases
Either U.S.P. breathing grade air or nitrogen is recommended for use as zero gas. A mix­ture of NO or NO gen is recommended as span gas. For maximum accuracy, the concentration of NO in the span gas should be about the same as that in the expected sample stream.
Ozonator Source Gas
For analyzers with ranges less than 0 to 2,500 ppm, U.S.P. breathing grade air should be used for generation of the ozone required for the chemiluminescence reac­tion. For ranges greater than 0 to 2,500 ppm, breathing grade oxygen is required.
EXHAUST
EXHAUST FAN SAMPLE IN INTAKE FAN
OZONATOR AIR
Figure 2-2. CLD Rear Panel Connections
At no time should sample flow exceed 2200 cc/min. Damage to internal compo-
in a background of nitro-
2
nents may occur if this flow level is ex­ceeded.
EXTERNALLY RESTRICT SAMPLE FLOW
TO LESS THAN 2200 CC/MIN.
No restrictor is provided in the sample inlet of this module. For those users who cannot externally limit sample flow, accessory restrictor PN 657300 is avail­able from Rosemount Analytical. To ob­tain this accessory, call 800-433-6076.
WARNING
CAUTION .
Sample
The sample must be clean and dry before entering the Analyzer Module. Sample should be filtered for particulates down to two microns, and should have a dew point
o
5
C (13oF) below coldest ambient tempera-
ture.
Pressure
Between 690 and 1035 hPa-gauge (10 and 15 psig) of ozonator air (or oxygen, if ap­propriate) should be present at the rear panel of the module. Bypass sample gas flow rate should be between 700 and 2000 cc/min. with backpressure regulator pres­sure (see Capillary Pressure in “Current Measurement Parameters” menu, which can be viewed by selecting the PARMS softkey in “Main Menu”) adjusted to 344 hPa (5 psig).
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Page 26
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
Zero/span gases should be introduced at the SAMPLE IN fitting at normal sample in­let flow rate.
WARNING
At no time should ozonator gas pressure exceed 2070 hPa-gauge (30 psig). Dam­age to internal components may occur if this pressure level is exceeded.
Leak Test
The CLD Analyzer Module is completely tested at the factory for gas leakage. The user is responsible for testing for leakage only at the inlet and outlet fittings on the rear panel. Periodically, the user should do an internal leak test (with a test procedure chosen by the user).
Contaminants
White crystal deposits on the windows of the reaction chamber, and the clogging of capillaries and vent are usually due to sam­ple contaminates such as ammonia reacting with the high ozone levels and NO compo­nents.
To eliminate the contaminates, the sampling system should be reworked or a preventive
maintenance program developed (if dropout is not excessive). Another source of crystal­line formation is contaminated air.

2-5 ELECTRICAL CONNECTIONS

NOTE
Electrical installation must be in compliance with National Electrical Code (ANSI/NFPA
70) and/or any applicable national or local codes.
Refer to Figure 2-3, page 2-4. Two electrical connections are required on the Analyzer Mod­ule: POWER and NETWORK. On the Analyzer Module, two NETWORK connectors are avail­able, either of which is appropriate for:
1. Interconnection with the Backplane of the Platform. (See instruction manual for the NGA 2000 Platform).
2. "Daisy-chaining" with other NGA 2000 com­ponents.
Connect Analyzer Module POWER to 24 VDC, 120 Watt power source, either the Platform or external power source.
Metering Valve
Regulator Adjus t Slot
Flow
Sample
Balance
Regulator
24V
250 V
CLD
LON
1
LON
2
1 + 2 ­3 GND
T 6A
Network1 Connect ion
Network2 Connect ion
Power Connection
Fuse
Figure 2-3. Front Panel Controls and Electrical Connections
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 27
Model NGA2000 CLD
5
3
5
4
4
4
Instruction Manual
760000-C
November 2002
POWER SUPPLY BOARD-LON/POWER
MODULE ASSEMBLY 657510
J2
J1
LON/PWR MODULE 656761
J5
3 2 1
RED
CABLE, CONTR OLLER PWR 903034
NC
J6
1 2
1
2
POWER SUPPLY
BOARD 657520
J13
8 7
1
2
J1
10
9
3
12
7
HEATERS
J11
1
10
J3
6
1
12
RED
DETECTOR 658160
CONVERTER HEATER JACKET 655228
J3
E3
E2
E1
GRN
ORN
BLK
GRN
BLK
1
GRN
J2
2
ORN
3
WHT
WHT
THERMISTOR 655216
THERMOSTAT 657298
THERMOSTAT
CONVERTER TEMPERATURE SENSOR 65528 2
CONVERTER
655250
NC
1
J1
J4
1
J2
COMPUTER ANALYSIS BOARD 655520
1
J3
J5 J6
CABLE, INTERNAL NETWOR K 903035
HARNESS, SIGNAL POWER 655252
CABLE, FLOW SENSOR 655384
PRESSURE SWITC H (NO)
RED
WHT
BLK
FLOW
SENSOR
902931
4-15 PSIG 662298
BRN (NO)
1
HARNESS, DRIVER POWER 655243
HEATER/THERMOSTAT ASSEMBLY 655235 PHOTODIODE SOCKET ASSEMBLY 655218
NO/NOx SOLENOID 655263
WHT
WHT
SHLD
GRN
JUMPER 657162
OZONATOR
POWER SUPPLY
657716
SHIELD
HARNESS, EXPANSION I/O 655249
111
1
J7
2
SAMPLE PRESSURE SENSOR 655253
YEL (COM)
OZONATOR
659494
TRANSISTOR
E
2SD1308
C
655264
B
B = YEL C = RED E = BLU
RED
SAMPLE PRESSURE SENSOR 655254
RED
WHT
BLU
RED
YEL
BLK
ORN
ORN
BRN
WHT
BLK
RED
BLK
RED
WHT
WHT
BRN BRN
YEL
J6
4 3
J8
2 1
J9
J1
J13
BRN
1 4 5 6
GRN
WHT
RED
3
J2
2 1
3
J1
2 1
J4 2
1
2
1
3
J9
2 1
1 2 3 4
J8
5 6 7 8
4 3
J5
2 1
4 3
J12
2 1
12
10
J5
J3
SIGNAL BOARD 655580
1 2 3 4 5 6 7 8 9
SHLD*
ORN
GRN
WIRE*
YEL
WHT
BLU
BLK
*COAX CABLE
RED
J11
1 2 3 6
J2
GRN
WHT
BLK
RED
BLK
WHT
RED
RED
4 3 1
J10
J3
DRIVER BOARD
655620
3
J11
1
J2 J6
10
2 1 2
CABLE, ANALO G I/O 903033
BLK
WHT
BLK
7 6
1
8
CABLE, DIGIT AL I/O 655246
CABLE, DIAGNOSTICS 903032
24
1 Optional in Bypass Flow Sensor Configuration.
FAN 655245
FAN 655245
Figure 2-4. CLD Wiring Diagram
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Page 28
Instruction Manual
[
]
[23].9[23].9[33]
760000-C November 2002
.5
[13]
Dimensions:
INCHES
[MM]
8.4
[213]
8.2
[208]
6.1
[155]
6.0
[152]
17.5
[445]
4.3
[109]
1.3
2.5 [64]
1.5 [38]
1.1 [28]
4.1
[104]
4.1
4.1
104
Figure 2-5. Outline and Mounting Dimensions
Model NGA2000 CLD
.4
[10]
.7
[18]
[20]
2.8 [71]
.8
1.3 [33]
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 29
Model NGA2000 CLD
y
Instruction Manual
760000-C
November 2002
SECTION 3

OPERATION

3-1 OVERVIEW

Once the CLD has been correctly assembled and installed, the analyzer is ready for opera­tion.
Before operating the system, verify that the Leak Checks have been performed in accor­dance with Section 2-4, page 2-2.
For the remainder of this section, Analyzer Module interconnection with a Platform or some interfacing component is assumed. Dis­play and Keypad information refers to that which the user can expect to see and do with regard to the Front Panel of the Platform.
CLD
Sample flow: 1300 ml/min 200 1500
Sample press: 4.0 hPa 50.0 490.0
Ozonator: OFF – PRESS. SW.
Converter temp: 30.9 C 150.0 500.0
Displa

3-2 DISPLAYS & OPERATING KEYS

The LCD screen shows all measurement val­ues of the analyzer, status values and all user menu instructions. Operation is performed with five function softkeys, four arrow (cursor) softkeys and the enter softkey. The function of each softkey varies depending on the installed analyzer module, any auxiliary modules in­stalled, and the individual menu displayed.
In case of power failure, all user defined spe­cific module parameters are saved by a bat­tery powered memory.
7.50
0.00 Range: 1 10.00
Status… Main… Channel BasicCal
ppm NO
Figure 3-1. Measure Mode Display
The Function Keys, also called softkeys, are assigned values depending on the menu or
a menu or to select numeric digits for adjust-
ment. screen being displayed. The legend is dis­played above the softkeys.
The
Enter Key is used to confirm a previously
entered variable value, to start a selected function or to go to a submenu selected at a menu line as opposed to the Function Keys. As an alternate to using the Enter Key to start

a. Menu Lines & Softkey Functionality

Menu lines can be selected with the softkey or the softkey. The selected line is displayed as white lettering on a black background (highlighted). Menus can con­tain four different types of lines:
a function, the softkey can be used.
Menu Line – A line ending with three dots
The
Cursor Keys ( or ) are used to move
up or down the lines within a menu or to in­crement and decrement number variables.
(…) indicates that it leads to a submenu. The submenu can be activated by press­ing the softkey or the softkey when the line is highlighted.
Cursor Keys ( or ) are used to move
The backwards or forwards between the pages of
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-1
Page 30
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
Function Line – A line ending with an
exclamation point (!) indicates that it will start a function. The function can be acti­vated by pressing the softkey or the softkey when the line is highlighted.
Variable Line – A line ending with a colon
(:) indicates that it displays a module vari­able parameter. Some parameters can be changed and some parameters display only a status and cannot be changed. Pa­rameters that cannot be changed will be displayed below a line within the menu.
Text Line – A line without punctuation
marks only displays information.
Tag Line – At the top of each menu page
is the tag line of the current channel. To the right of the Tag is the value of the in­dicated channel.
The Function Keys (Softkeys) can some­times be assigned as Functions (exclama­tion point) or Submenus (three dots) as shown below.

b. Common Function Keys

Display
ponent display to the multi-component display. F1 in the single component dis­play.
Measure – Change from menus and
submenus to the single component dis­play of the selected channel. F1.
Status – Change to the menu “Current
measurement parameters” which displays the most important parameters and infor­mation about the status of the current channel or module. F2 if available. (See Section 3-4a, page 3-8)
– Change from the single com-
Channel – Scrolls through the channels
in the same menu. In the main menu and the single component display menu it moves between the channels of the con­nected analyzers and analyzer modules. In the submenus it moves only between the channels of the current analyzer or analyzer module. F3 if available, F4 in the single component display.
Lock – Changes to the main menu and
locks all three operation levels, if a secu­rity code is enabled in the system configu­ration (See Section 3-11d, page 3-35). F4 in the main menu.
BasicCal – Change from the single com-
ponent display to the menu “Basic Con­trols and Setup.” F5 in the single component display. (See Section 3-5, page 3-10)
MFG Data – Change from the main menu
to the menu “Manufacturing Data” which displays further submenus with informa­tion about the control module and ana­lyzer module, such as address of the manufacturer, serial number of the mod­ules and software and hardware versions. F5 in the main menu. (See Section 3-2f, page 3-4 and Figure 3-6, page 3-5)
More – Changes to an additional menu
page of the current menu. F3 or F5 if available.
ESCAPE/Back – Returns to the previous
menu. Usually F2 or F4. When changing a variable, the previous value is displayed above the Back softkey. Pressing the Back softkey restores the previous value.
INFO – Context sensitive help screens for
the current menu.
Main – Change from single component
display to the main menu. F3 in the single component display. (See Section 3-2e, page 3-4)
HOME – Change for various menus to the
main menu. F1.
3-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 31
Model NGA2000 CLD
g
Y
y
y
play
y
g

c. Entering & Changing Variables

1. Select the variable line desired to be
changed using the softkey or the softkey. The selected line will be highlighted white on black.
2. Press the ↵ softkey and the pa-
rameter will be selected for modifi­cation.
3. The F2 softkey changes to “Back…” and the previous value of the vari­able shows above it for easy refer­ence. When the variable being changed is numeric, the F4 softkey changes to “+/-“ to allow changing of the sign from positive to negative if applicable.
4. Use the ↑ softkey or the ↓ softkey change the entire value, scroll among the available variables or change the value of a selected digit or character.
5. Use the ← softkey or the → softkey to select digits within a number. For some variables the quantity of digits or characters can be changed.
6. Press the ↵ softkey again to confirm the new value.
Tag Line
Menu Line
Function Lines
Variable Lines
CLD 7.50 ppm
Analyzer basic controls (calibration) & setup…
Anal
zer and I/O, expert controls & setup…
stem configuration and diagnostics…
S
Dis
Time & Date:
stem tag:
S
Measure
-- Main Menu --
controls…
Status… Channel Lock… MFG Data
Figure 3-2. The Display Screen

d. Starting a Function

Instruction Manual
760000-C
November 2002
Selected Line (Reverse Text)
Lines below this separator are information and cannot be chan
Function softkeys F1 – F5 Legend
CLD 7.50 ppm
Displayed concentration digits: 4 Digits after decimal point: 2
Module identification tag: CLD
Si Signal on mini-bargraph – 2: Sample press Signal on mini-bargraph – 3: Ozonator OFF – PRESS SW
Signal on mini-bargraph – 4: Converter temp
Measure
-- Measurement Display Configuration --
nal on mini-bargraph – 1: Sample flow
Back…
Figure 3-3. Changing Variables
Pressing the softkey or the softkey while a function line is highlighted will bring up a confirmation menu as shown below.
Pressing the Yes (F2) softkey will start the function immediately.
Pressing the Back… (F4) softkey will re­turn to the previous menu page.
CLD 7.50 ppm
-- Confirmation Required –
Do you really want to do this ??
Press “Yes” or “Back…”
es
Figure 3-4. Function Confirmation
ed.
Back…
Display
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-3
Page 32
Instruction Manual
A
A
760000-C November 2002
Model NGA2000 CLD

e. Measure Mode Display

The Measure Mode is the normal mode of operation. In this mode, the Display will show the current gas measurement, the component of interest, the current opera­tions of the softkeys, and several graph­ics. A bar representing the displayed concentration is shown as a percent of full-scale and up to four lines showing user selectable secondary parameters from either the Analyzer Module or any IO module bound to it. See the Platform manual for information as to how to select these. The Measure Mode display is shown in Figure 3-1, page 3-1.
If more than one Analyzer Module is con­nected to the system, an additional Run Mode display will show as many as four (five for version 2.3 and later) gas meas­urements on the display screen.

f. Main Menu

Pressing the Main… (F3) softkey or the softkey while in any single component display will bring up the Main Menu. From the Main menu it is possible to change all operating values of the analyzer to set up and control the parameters of measure­ment, calibration and data transfer. The sub menus of the Main Menu are shown in Figure 3-7, on page 3-6
From the Main menu, the MFG Data (F5) softkey will access several submenus showing the manufacturing and version data of the analyzer as shown in Figure 3-6, page 3-4.
Selection from the Main menu:
Measure (F1) – Changes to the single
component display of the current channel. See Section 3-4b, page 3-8.
Status… (F2) – Changes to the “Current
measurement parameters” menu of the current channel. See Section 3-4a, page 3-8.
Channel (F3) – Scrolls through all chan-
nels of the connected Analyzers and Ana­lyzer modules.
Lock… (F4) – Locks any operating level
by security code. See Section 3-11d, page 3-35.
MFG Data (F5) – Changes to “Module
Manufacturing Data” menu. See Figure 3-6, page 3-5.
CLD 7.50 ppm
-- Main Menu --
nalyzer basic controls (calibration) & setup…
nalyzer and I/O, expert controls & setup…
System configuration and diagnostics…
Display control s…
Time & Dat e: 10:30:05 August 10 2001 System tag: Fisher-Ro semount
Measure Status… Channel Lock… MFG Data
Figure 3-5. Main Menu Display
3-4 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 33
Model NGA2000 CLD
y
g
g
y
g
g
y
g
y
y
play
y
Instruction Manual
760000-C
November 2002
CLD 7. 50 ppm
Analyzer basic controls (calibration) & setup…
zer and I/O, expert controls & setup…
Anal
S
stem configuration and di agnostics…
Dis
controls…
Time & Date: 10:30:05 August 10 2001 S
stem tag: Fisher-Rosemount
Measure
-- Main Menu --
Status… Channel Lock… MFG Data
CLD 7. 50 ppm
Control module manufacturing data…
Anal
-- Module Manufacturin
zer module manufacturing data…
Data --
CLD 7. 50 ppm
-- Control Module Manufacturin
(C) 2001 Fisher-Rosemount MFG GmbH & Co OHG
Manufactured by:
Fisher-Rosemount GmbH & Co OHG
Industriestrasse 1
D-63594 Hasselroth / Germany
Tel: (+49) 6055 884-0
FAX: (+49) 6055 884-209
Data --
Measure
Status… Back…<<< >>>
Measure
Or… More…
Back…
CLD 7. 50 ppm
More…
Bench confi
-- Anal
zer Module Manufacturing Data --
uration code: CLA – low range
Minimum ran ge: 10 .0 ppm Maximum range: 10000 ppm Diode s/n, block s/n: 0000, 0000 Capillary: 200.0 ml/min
Flow sensor: Present
HOME
ESCAPE RESET STORE INFO
CLD 7.5 0 ppm
Revision date: May 8 2001 Revision time: 15:30:02
Phrase dictionary version: P012/01/00 Language: English
Measure
-- Control Module Version Information --
Serial number: CM1 Manufacturi ng date: 20 04 01
Hardware revision: ACU02 R 3.3.1.D April 01 2001 Software revi sion: 3.6 0 / P010
Back…
CLD 7.5 0 ppm
Measure
-- Control Module Manufacturin
(C) 2001 Fisher-Rosemount
Manufactured by:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, CA 92807-1802 / USA
Tel: (714) 986-7600
FAX: (714) 577-8739
Or… More…
Data --
Back…
CLD 7. 50 ppm
Measure
-- Control Module Manufacturin
(C) 2001 Fisher-Rosemount
Manufactured by:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, CA 92807-1802 / USA
Tel: (714) 986-7600
FAX: (714) 577-8739
Or… More…
Data --
Back…
CLD 7.5 0 ppm
-- Anal
zer Module Version Information --
Serial number: CLD1
Manufacturing date: 08 08 01
Hardware revision: 1.0
Software revi sion: 3.6 0 / P012 Revision date: August 8 2001 Revision time: 03:26:15
Measure
Back…
Figure 3-6. Module Manufacturing Data Displays
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-5
Page 34
Instruction Manual
y
y
y
play
y
p
y
play
pp
760000-C November 2002
CLD 7.5 0 ppm
Anal
zer basic controls (calibration) & setup…
zer and I/O, expert controls & setup…
Anal
S
stem configuration and diagnostics…
Dis
controls…
Time & Date: 10:30:05 August 10 2001
S
stem tag: Fisher-Rosemount
Measure
-- Main Menu --
Status… Cha nnel Lock… MFG Data
Figure 3-7. Main Menu Sub Menus
Model NGA2000 CLD
CLD 7.50 ppm
Measurement range number: 1 Range upper lim it: 10.0 ppm
Automatic range change control: Disabled
Ranges with valid calibration: None
Calibration status: READY
Span gas concentration: 10.0 ppm
Status: STANDBY
Measurement mode: NO Ozonator status: OFF – PRES. SW. Ozonator
CLD 7.50 ppm
Analyzer module controls… System & network I/O module controls…
Analyzer module setup… System & network I/O module setup…
(Note: Controls & setup are identical for MLT/TFID)
Measure
CLD 7.50
System calibration… Diagnostic menus… Load/Save configuration (CMMCA)… Date and time… Security codes… Network module management… System reset… Pump 1: Off Pump 2: Off
Measure
CLD 7.50 ppm
Measure
ower: Enabled
HOME
-- Anal
-- System Configuration and Diagnostics --
Brightness: 70%
Contrast: 23%
Switch automatically to “Measure” after: 30 s
Switch off backlight after: Never
-- Basic Controls
NO/NOx
zer and I/O, Expert Controls and Setup --
ZERO INFO
Channel
Channel
-- Dis
Back…
Back…
Controls --
Back…
SPAN
m
3-6 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 35
Model NGA2000 CLD
Instruction Manual
760000-C
November 2002

3-3 STARTUP & INITIALIZATION

Apply power to the CLD Analyzer Module. If it is associated with a Platform, do this by plug­ging in the Platform to a power source. The Platform has no ON/OFF power button. Once power has been supplied to the Platform, the CLD Analyzer Module will be energized.
After switching on the CLD, the analyzer will begin its booting procedure which is apparent on the LCD screen. The first part of the ini­tialization procedure is a self check of the software and analyzer components. Various displays will show the status of the initializa­tion including revision notes, “Initializing net­work interface,” “Searching for nodes,” “Scanning Module 2: CLD, 12% Complete,” and “Calculating bindings.”
If the user's system contains only one Ana­lyzer Module, all system components, the Controller Board and the network "self-install" (bind together) during initial startup. If the sys­tem contains more than one Analyzer Module, the startup sequence will interrogate the net­work to locate and identify all components on the network. The user will have to bind appro­priate combinations of components after the startup sequence. See the Platform manual for instructions on binding combinations of modules.
Pressing the F1 softkey during initializing will reset the LCD brightness and contrast to fac­tory settings (See Section 3-6, page 3-11). Pressing the F3 softkey will abort the network initializing, aborting any connection to other analyzers. In that case, only the menus of the local analyzer will be available.
ozonator air pressure and sample flow rate are within specifications (see page 1-4). Cali­brate and adjust converter efficiency, and be­gin routine operation as the following sections indicate.
(C) 2001 FISHER-ROSEMOUNT Analytical
NGA-2000 Control-Module Rev. 3.6.0 /P010
Language: P012/01/00
Initializing Network
Initializing network interface
LCDReset
Abort
Figure 3-8. Startup Display

a. Routine Operation

The CLD Analyzer Module is designed to continuously analyze the sample stream. Normally, it is never powered off except for servicing or for a prolonged shutdown.
Maximum permissible interval between calibrations depends on the analytical ac­curacy required, and therefore cannot be specified. Initially, the instrument should be calibrated at least once every eight hours. This practice should continue until experience indicates that some other in­terval is more appropriate.
At the end of the initializing routine the “measure” screen will display as shown on the next page. This screen is the access to all other channels, menus and submenus. The actual display may differ from that shown de­pending on any custom configuration as de­scribed in Sections 3-7h, page 3-18 and Section 3-7j, page 3-19.
After the warm-up period (about one hour for the CLD Analyzer Module), all modules are completely functional. Establish that correct
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-7
Page 36
Instruction Manual
p
p
y
y
A
y
760000-C November 2002
Model NGA2000 CLD

3-4 BASIC CONTROLS AND STATUS

The following sections describe the basic con­trol of the analyzer and the viewing of channel parameters. Examples of stepping through the menus are shown so that the user can be­come familiar with the operation, keeping in mind that displays and menu choices may be different depending on actual analyzer con­figuration and any customization of the menus.

a. Analyzer Channel Status

The analyzer status is displayed in the Current measurement parameters dis­play.
To access the Current measurement pa­rameters display:
Main Menu [Measure (F1)]
Measurement Mode [STATUS(F2)]
Current measurement parameters
CLD 7.50 ppm
Analyzer gas measured: NO
Measurement range number: 1 Range change control: Local
Linearization mode: Disabled
Analyzer operational state: STANDBY Analyzer alarm state: NORMAL
Alarms reported: FAILURE
Current total variable updates per second: 24
HOME
-- Current measurement
ESCAPE MORE INFO
arameters --
From the Current measurement parame­ters menu an additional sub menu is available with additional parameters.
Press MORE (F3) to access the next menu.
CLD 7.50 ppm
Response time: 3.0 s Bypass flow: 1300 ml/min. Capillary pressure: 4.0 hPa Detector pressure: 38.8 c
-- Current measurement
arameters --
HOME
ESCAPE MORE INFO
Press ESCAPE (F2) or MORE (F4) to re­turn to the first menu or HOME (F1) to re­turn to the Main menu display.

b. Single Component Display

The Measure menu that displays after startup is the Single Component display of the analyzer. If other analyzer modules are connected to the Platform, it is possi­ble to display them using the following steps to change the channel of the single component display:
From the Measure Mode display, press Channel (F4) to change to the Single Component Display of any other installed Analyzer Modules.
Example: Changing from CLD (channel 1) to CO
(channel 2). Continue pressing
2
Channel (F4) to display the desired chan­nel depending on the installed analyzer configuration, ultimately returning to the first channel.
CLD
7.50
0.00 Range: 1 10.00
Sample flow: 1300 ml/min 200 1500
Sample press: 4.0 hPa 50.0 490.0
Ozonator: OFF – PRESS. SW.
Converter temp: 30.9 C 150.0 500.0
Displa
MLT/CH1/R2
Temperature: 37.0 C 0.0 100.0
Maintenance Requests: No
ny alarms: No
Operation: Ready
Displa
CLD
Sample flow: 1300 ml/min 200 1500
Sample press: 4.0 hPa 50.0 490.0
Ozonator: OFF – PRESS. SW.
Converter temp: 30.9 C 150.0 500.0
Status… Main… Channel BasicCal
2.50
0.00 Range: 2 5.00
Status… Main… Channel BasicCal
7.50
0.00 Range: 1 10.00
ppm NO
% CO2
ppm NO
Displa
3-8 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Status… Main… Channel BasicCal
Page 37
Model NGA2000 CLD
y
A
[2]
[2]
y
Instruction Manual
760000-C
November 2002

c. Multi Component Display

If other analyzer modules are connected to the Platform, it is possible to display up to five using the following steps to change from the single component display to the multi component display as follows:
From the single channel display (Meas­ure) press Display (F1) to change to the Multi Component Display of all other in­stalled Analyzer Modules.
Changing to the multi component display can be done from each single component display.
CLD
7.50
0.00 Range: 1 10. 00
Sample flow: 1300 ml/min 200 1500
Sample press: 4.0 hPa 50.0 490.0
Ozonator: OFF – PRESS. SW.
Converter temp: 30.9 C 150.0 500.0
Displa
Status… Main… Channel BasicCal
ppm NO
MLT/CH1
% CO2
MLT/CH2
ppm CO
>
7.50
2.50
95.00
Select
Status… Tags Off LCDReset
CLD
0.00 [1] 10.00
0.000
0.00
ppm NO
5.00
250.00
Each bargraph shows the start and end of the range for the respective channel. The number in parentheses indicates the number of the selected range for that channel. (F.S. = full scale)
(Display may look different depending on installed analyzers.)
Use the F3 softkey (Tags Off) to turn the analyzer tags on or off.
To select a single channel display in the multi channel display, enable the select symbol (>) by pressing the F1 softkey or the softkey.
Then use the or softkey to select the line for the desired channel. When the desired channel is marked, select it for single component display by pressing the F1 softkey.
MLT/CH1
2.50
% CO2
0.00 Range: 2 5.00
Temperature: 37.0 C 0.0 100.0
Maintenance-Request No
ny-alarms: No
Operation: Read y
Displa
Status… Main… Chann el BasicCal
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-9
Page 38
Instruction Manual
760000-C November 2002
Model NGA2000 CLD

3-5 BASIC CONTROLS

To access the Basic Controls menu:
Measurement Mode [BasicCal (F5)]
Or
Main Menu
Analyzer basic controls (calibration) &
setup…
CLD 7.50 ppm
Measurement range number: 1 Range upper limit: 10.0 ppm
Automatic range change control: Disabled
Ranges with valid calibration: None
Calibration status: READY
Span gas concentrati on: 10.0 ppm
Status: STANDBY
Measurement mode: NO Ozonator status: OFF – PRES. SW. Ozonator power: Enabled
HOME
-- Basic Controls --
NO/NOx
ZERO INFO
Figure 3-9. Basic Controls Menu
The Basic Controls menu is used to set the range, initiate and exit a quick start and exit the sleep mode.

Measurement range number:

SPAN
See Section 3-7a, page 3-12 to change the begin and end concentration values for each of the four ranges.

Automatic range change control:

When enabled, the CLD analyzer module will change ranges automatically when the pre­sent range is exceeded.

Span gas concentration:

Allows setting of the calibration span gas con­centration. The default is the range upper limit.

Ozonator Power:

Used to turn off (Disable) the ozonator. For applications where measurements are made infrequently but the analyzer is to remain on to avoid the warm-up period, the CLD would be placed in Standby mode and the Ozonator could be Disabled to conserve the life of the lamp and converter.

NO/NOx (F2)

Press the F2 softkey to toggle the measure­ment mode between NO and NOx.
To select one of the four ranges of the CLD analyzer, Move the cursor to the “Measure­ment range number:” line and press the softkey. Change the range number using the and softkeys and then press the softkey again to save the selection.

ZERO (F3) / SPAN (F4)

To ZERO (F3) or SPAN (F4) the analyzer, flow the appropriate gas, select the correct range and press the appropriate softkey. Do a zero before span!
3-10 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 39
Model NGA2000 CLD
Instruction Manual
760000-C
November 2002

3-6 DISPLAY CONTROLS

Main Menu
Display control…
CLD 7.50 ppm
Brightness: 70 % Contrast: 23 %
Switch automatically to “Measure” after: 30 s
Switch off backlight after: Never
Measure
-- Display Controls --
Figure 3-10. Display Controls Menu
The Display Controls menu is used to adjust the display parameters.

Brightness and Contrast:

These controls can be adjusted to accommo­date the ambient lighting conditions. The range of values are 20-100% for brightness and 1-45% for contrast.
These values can be reset to the defaults from the Multi Channel display screen (Sec­tion 3-4c, page 3-9) by pressing LCDReset (F1) function softkey and from the Start up display (Figure 3-8, page 3-7) by pressing LCDReset (F5).
Switch automatically to “Measure” after:
This variable line allows setting of the delay time before any selected menu switches back to the Measure screen. The selectable values are:
10 sec 30 sec
1 min
5 min 10 min 30 min
Never

Switch off backlight after:

This variable line allows setting of the delay time before switching off the backlight of the display screen. The selectable values are:
10 sec 30 sec
1 min
5 min 10 min 30 min
Never
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-11
Page 40
Instruction Manual
g
p
y
760000-C November 2002
Model NGA2000 CLD
3-7 ANALYZER AND I/O, EXPERT CONTROLS
AND SETUP
Main Menu
Analyzer and I/O expert controls & setup…
CLD 7.50 ppm
Analyzer module controls… System & network I/O module controls…
Analyzer module setup… System & network I/O module setup…
(Note: Controls & setup are identical for MLT/TFID)
zer and I/O, Expert Controls and Setup --
-- Anal
Measure
Channel
Back…
Figure 3-11. Analyzer and I/O Expert Controls
and Setup Menu
The Expert Controls and Setup menus pro­vide for the configuration of system and net­work I/O (SIO & DIO), and for the configuration of various functions on the CLD analyzer module:
softkeys and then press the softkey again to save the selection.
CLD 7.50 ppm
Measurement range number: 1 Range upper limit: 10.0 ppm Range settings… Range and functional control: Local Measurement mode: NO Ozonator status: OFF – PRES. SW. Ozonator power: Enabled Zero/Span calibration… Ranges with valid calibration: None Physical measurements…
HOME
-- Ex
ert Controls --
CAL
NO/NOx
Figure 3-12. Expert Controls Menu

Range and functional control:

Local = Range control by the menus
Inputs I/O module = Range control by remote communication
Program I/O module = Range control by the I/O module

a. Range Settings

INFOESCAPE
NOTE
Whenever the “Channel” tag appears above the F3 softkey, pressing F3 will switch to any other installed analyzer modules, one after the other and eventu­ally back to the CLD module. When activat­ing any other installed module, the menus will be different depending on that module. See each module’s manual for a descrip­tion of those menus.
In the “Analyzer and I/O expert controls and setup” menu, select “Analyzer module con­trol…” . The “Expert Controls” menu will dis­play as shown in Figure 3-12.
NOTE
Zero/Span calibration… is described in Section 3-8c, page 3-23.

Measurement range number:

To select one of the four ranges of the CLD analyzer, Move the cursor to the “Range number:” line and press the softkey. Change the range number using the and
Used to change the upper and lower limit for each of the four ranges. select Range settings… from the Expert Controls menu.
Main Menu
Analyzer and I/O expert controls &
setup…
Analyzer module controls…
Range settings…
CLD 7.50 ppm
Minimum range: 10.0 ppm Maximum range: 10000 ppm Range 1 lower limit: 0.0 ppm Range 1 upper limit: 10.0 ppm Range 2 lower limit: 0.0 ppm Range 2 upper limit: 25.0 ppm Range 3 lower limit: 0.0 ppm Range 3 upper limit: 100.0 ppm Range 4 lower limit: 0.0 ppm Range 4 upper limit: 250.0 ppm
HOME
ESCAPE
-- Ran
e Settings --
INFO
Figure 3-13. Range Settings Menu
3-12 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 41
Model NGA2000 CLD
y
p
Instruction Manual
760000-C
November 2002
To change any of the limits, move the cursor to the desired line and press the softkey. Change the limit value using the and softkeys and then press the softkey again to save the selection.
Press the ESCAPE (F2) softkey before pressing the softkey to restore the pre­vious value.
The Range Settings menu is also acces­sible from Gas Measurement Parameters menu (Figure 3-16).
The Analyzer Module will not allow the user to increase the upper limit of a range beyond the maximum range software set­ting. To change the Maximum Range value, select “Range x upper limit”, and use the arrow softkeys to scroll the indi­cated value. The same applies for the “Range x lower limit” value.

b. Physical Measurements

Use the Physical Measurements menu to view various conditions of the CLD mod­ule and to adjust the bypass flow limits.
Main Menu
Analyzer and I/O expert controls &
Analyzer module controls…
Physical measurements…
CLD 7.50 ppm
Bypass sample flow: 1300 ml/min Flow lower limit: 200 ml/min Flow upper limit: 1500 ml/min Sample pressure: 3.4 hPa Ozone supply pressure: 2.7 hPa Sensor block temperature: 29.6 C
sical Measurements --
-- Ph
HOME
Figure 3-14. Physical Measurements
setup…
NO/NOx
Menu
The bypass flow path is used to increase the response time of the detector. Upper and lower limits can be set to trigger a di­agnostic alarm.
To change any of the limits, move the cursor to the desired line and press the softkey. Change the limit value using the and softkeys and then press the softkey again to save the selection.
Press the Back (F2) softkey before press­ing the ↵ softkey to restore the previous value.
The normal value for bypass flow is 700­2000 cc/min.

c. Concentration Alarms

Use the Concentration Alarm Setup menu to establish or change the alarms for the sample gas concentration measurement. (Calibration parameters are described in Section 3-8a on 3-21)
Main Menu
Analyzer and I/O expert controls &
setup…
Analyzer module setup…
Concentration alarms…
CLD 7.50 ppm
Alarm generation is: On Level for Low-Low alarm: 0.000
Level for Low alarm: 0 .000 Level for High alarm: 9.000 Level for High-High alarm: 10.000 Alarm delay: 1.0 s Low-Low alarm: On Low alarm: On High alarm: On High-High alarm: On
HOME
-- Concentration Alarm Setu
ESCAPE
ACKN
--
Figure 3-15. Concentration Alarm
Setup Menu
INFOESCAPE
To change the alarm limits, move the cur­sor to the desired line and press the softkey. Change the limit value using the and softkeys and then press the softkey again to save the selection.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-13
Page 42
Instruction Manual
p
760000-C November 2002
Model NGA2000 CLD
Press the Back (F2) softkey before press­ing the ↵ softkey to restore the previous value.
Alarm delay:
Set the alarm delay for the desired time delay after the concentration value ex­ceeds the limit before the alarm is acti­vated.
Alarm generation is:
This setting establishes the generation of alarms as “Off,” “On,” or “On (Hold Alarm).” The “On (Hold Alarm)” setting provides that the alarm will remain active even after the gas concentration returns below the appropriate level until the ACKN softkey is pressed or it is reset from the remote I/O.
ACKN
Press the F3 softkey to acknowledge and reset any alarm.

d. Linearization Parameters

Linearization parameters allows the es­tablishment of linearizer coefficients and the assignment of coefficient sets to indi­vidual ranges.
Main Menu
Analyzer and I/O expert controls &
Analyzer module setup…
Gas measurement parameters…
CLD 7.50 ppm
Linearization parameters…
Response time/delay parameters… Range settings… Automatic range change parameters… Units…
-- Gas measurement
Linearization functions…
HOME
ESCAPE
Figure 3-16. Gas Measurement
Parameters Menu
setup…
arameters --
INFO
From the Gas measurement parameters menu, select Linearization parameters…
CLD 7.50 ppm
Range 1 linearizer: Disabled If enabled, uses curve no.: 1 Range 2 linearizer: Disabled If enabled, uses curve no.: 2 Range 3 linearizer: Disabled If enabled, uses curve no.: 3 Range 4 linearizer: Disabled If enabled, uses curve no.: 4
Set coefficients…
HOME
-- Linearization Parameters --
ESCAPE
INFO
Figure 3-17. Linearization Parameters
Menu
The linear polynomials act over a range (not the same as the measurement range). The system uses the linearizer polynomial appropriate for the measure­ment range chosen. This is the polyno­mial with the next higher range. It is, however, possible to specify that the ana­lyzer use a wider range polynomial.
Note that the use of different polynomials on different ranges will give different read­ings on a new range.
Coefficients may be edited for custom curves.
CLD 7.50 ppm
A0 coefficient: 0.000000 A1 coefficient: 1.000000 A2 coefficient: 0.000000 A3 coefficient: 0.000000 A4 coefficient: 0.000000 Curve upper limit: 10.0 ppm Curve over-range: 5.0 % Curve under-range: 5.0 %
Status: Disabled
HOME
Linearity coefficients
ESCAPE
Curve 1
NEXT LAST
INFO
Figure 3-18. Linearity Coefficients
Menu
Edit the polynomial coefficients as de­sired. Make sure that the curve upper limit is correct. This is the limit of the range that this polynomial will support.
The last line (Status) selects whether the curve is in use. Use the NEXT (F3), LAST (F4), and BACK (F4) softkeys to access all four curves.
3-14 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 43
Model NGA2000 CLD
y
A
A
Instruction Manual
760000-C
November 2002

e. Linearization Functions

The linearization functions allow the pri­mary variable output to be linearized by either a polynomial of up to 20 set points, or by midpoint piecewise correction with up to three midpoints.
Analyzer and I/O expert controls &
Analyzer module setup…
Gas measurement parameters…
Linearization functions…
CLD 7.50 ppm
Polynomial setup… Midpoint correction setup…
Use the polynomial setup to generate a linearizing polynomial from up to 20 gases. With more than 6 gases it will produce a fourth order polynomial linearizer. Use the midpoint correction for a piecewise-linear final correction, to bring up to three points precisely onto the curve.
HOME
-- Linearization Functions --
ESCAPE
Figure 3-19. Linearization Functions
Polynomial Set Up
Use the polynomial set up to generate a linearizing polynomial from up to 20 gases. With more than 6 gases, it will produce a fourth order polynomial lin­earizer.
Analyzer and I/O expert controls &
Analyzer module setup…
Gas measurement parameters…
Linearization functions…
Polynomial set up…
Main Menu
setup…
Menu
Main Menu
setup…
INFO
CLD 7.50 ppm
Range to be li nearized: 1 Current span gas: 10.0 ppm Calculated polynomial order: 4
Gas value shown is: ppm Gas concentrations…
Analyzer function: READY
HOME
ESCAPE
-- Pol
nomial Setup --
CALC
INFO
Figure 3-20. Polynomial Setup Menu
Setup:
Select the range to linearize.
Make sure that the span gas value is cor­rect and set into the “Correct span gas” line.
Choose whether to define the gas con­centrations as absolute values (ppm) or as a percent of the span gas (Percent of span gas) in the “Gas values shown as” line.
Percent would be used if the span gas is being diluted with a mixing device.
From the “Polynomial set up” menu (Figure 3-20, page 3-15) select the “Gas concentrations…” submenu to enter up to 20 points for each range.
CLD 7.50 ppm
Gas value: 1.00 Raw reading: 0.9 ppm Linearized value: 1.0 ppm
Gas value: 2.00 Raw reading: 1.9 ppm Linearized value: 2.0 ppm
Point to be measured: Point 1
nalyzer function: READY
HOME
ESCAPE
Gas concentrations
Point 1
Point 2
DAT
NEXT
INFO
Figure 3-21. Gas Concentrations Menu
In the “Gas concentrations” menu, enter the gas value for the desired point as ppm or percent of scan range in accordance with the previous choice made for the “Gas values shown as.”
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-15
Page 44
Instruction Manual
A
760000-C November 2002
Model NGA2000 CLD
Choose the point to be measured from the “Point to be measured” line.
At each point in succession, flow the gas of the correct value and, when the reading is stable, press the DATA (F3) softkey to record the gas value and raw reading for each point.
Move to the next two points with the NEXT (F4) softkey and move backwards with the ESCAPE (F2) softkey. After the desired number of points has been meas­ured and recorded, press ESCAPE from the Point 1 menu to return to the “Poly­nomial set up” menu.
In the “Polynomial set up” menu, press the “CALC” softkey.
The analyzer will calculate the best fit polynomial and store it as the coefficients in the current range’s linearizer function.
The order or the polynomial is optimized based on the number of data points pro­vided. At least 7 points are required for a fourth-order polynomial correction. The results can be modified with the “Midpoint correction” also provided. See Section 3­7e, page 3-15.
CAUTION.
The linearization curve must be mono­tonic. If it is not, the calibration routine will fail and the analyzer will not cali­brate. Test this by copying the values of the linearization coefficients into a spreadsheet program and plotting the results.
The analyzer does not test for monotonic­ity when it spans, but this test may not catch all possible errors.
Monotonic means that the curve does not change direction as the gas concentration increases.
Midpoint Correction Set Up
This function allows the adjustment of the primary variable output into a precise value using up to three midpoints. It does this with a piecewise-linear algorithm that occurs after and in addition to any poly­nomial linearization. .
The correction can be performed indi­vidually for each range.
Main Menu
Analyzer and I/O expert controls &
setup…
Analyzer module setup…
Gas measurement parameters…
Linearization functions…
Midpoint correction set up…
CLD 7.50 ppm
Correction: DISABLED Point being mesured: Point 1 Point 1 gas concentration: 2.50 ppm Point 2 gs concentration: 5.00 ppm Point 3 gas concentration: 7.50 ppm Point 1 reading: 2.52 ppm Point 2 reading: 5.12 ppm Point 3 reading: 7.56 ppm Span gas value: 10.00 ppm
nalyzer function: READY
HOME
Midpoint correction setup
Point 1
ESCAPE
SET
RANGE 2
INFO
Figure 3-22. Midpoint Correction
Setup Menu
First, disable the correction.
Set the point being measured to Point 1.
Then enter the first midpoint gas value, run the gas, and when stable, press “SET.” “Point 1 reading” will show the ac­tual reading, but the analyzer will adjust it to the correct value.
Repeat the above steps with the second and third points as desired.
When complete, se the “Correction:” line to “ENABLED” to activate the correction.
3-16 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 45
Model NGA2000 CLD
g
A
CAUTION.
Make sure that the corrections are not excessive. If the correction is too ex­cessive, the calibration routine will fail and calibration of the analyzer will not be possible.

f. Response Time

The response time menus allows the set­ting of the primary analyzer variable t90 times, the LON update rate and the output delay time.
Main Menu
Analyzer and I/O expert controls &
setup…
Analyzer module setup…
Gas measurement parameters…
Response time/delay parameters…
CLD 7.50 ppm
Range 1 t90 ti me: 3.0 s Range 2 t90 ti me: 3.0 s Range 3 t90 ti me: 3.0 s Range 4 t90 ti me: 3.0 s
LON update time: 10 per sec
Output delay time: 0.0 s
HOME
Response time/delay Parameters
ESCAPE
Figure 3-23. Response time/delay
The
t90 time adjusts the filtering or damp-
ing factor for the concentration outputs for each range. They are adjustable from 0.1 to 30 seconds with a default of 3.0 sec­onds.
The
LON Update Rate is the rate at
which the analyzer communicates over the system network. It is adjustable to “ASAP” (the fastest rate that the network communicates), “10 per sec,” and “1 per sec.”
Parameters
Output Delay Time establishes the
The delay for the DIO and analog outputs to respond to a change in concentration value. It is adjustable from 0.0 to 30.0 seconds.
To change a value, move the cursor to the desired line and press the softkey. Change the value using the and ↓ soft- keys and then press the softkey again to save the selection. Press the Back (F2) softkey before pressing the ↵ softkey to restore the previous value.

g. Automatic Range Change

This menu function allows the setting and enabling of the automatic range change for each of the four ranges. The automatic range switching for increasing values is effective at the upper range limit of each range, while for decreasing values it is the next lower range limit times the hysteresis percent times the upper range limit.
Main Menu
Analyzer and I/O expert controls &
setup…
Analyzer module setup…
Gas measurement parameters…
Automatic range change parameters…
CLD 7.50 ppm
Actual switch levels…
Switch level hysteresis: 20 %
Usage of range –1: Enabled
Usage of range –2: Enabled
Usage of range –3: Enabled
Usage of range –4: Enabled
Automatic range change control: Enabled
Absolute range upper limit: 10000 ppm
bsolute range lower limit: 10.0 ppm
Measure
-- Automatic Ran
Figure 3-24. Automatic Range Control
Set the desired hysteresis level in the range of 10% to 50% which is applied to each range transition. (20% is the default setting.) Enable the desired ranges and enable the automatic range change con­trol. To view the actual switching levels,
Instruction Manual
760000-C
November 2002
e Control --
Back…
Menu
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-17
Page 46
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
select the submenu “Actual switch lev­els…”.
CLD 7.50 ppm
-- Actual Switch Levels --
Range – 1 up: 10.0 ppm Range – 1 down: -1000000 ppm
Range – 2 up: 25.0 ppm Range – 2 down: 9.0 ppm
Range – 3 up: 100.0 ppm Range – 3 down: 22.5 ppm
Range – 4 up: 250.0 ppm Range – 4 down: 90.0 ppm
Measure
Figure 3-25. Actual Switch Levels

h. Display Units

This menu function is used to set the dis­played units for the various parameters. This only affects the displayed values. All outputs are in the basic SI units which, for example, for pressure is hPa. Gas con­centration values are for all ranges. Indi­vidual ranges cannot be set to different units.
Main Menu
Analyzer and I/O expert controls &
setup…
Analyzer module setup…
Gas measurement parameters…
]
CLD 7.50 ppm
Gas measurement units: ppm
Pressure measurement units: hPa
Temperature measurement units: C
Ppm to mg/Nm3 conversion factor: 1.000 Lower explosion limit (LEL): 0.00 % Upper explosion limit (UEL) 0.00 % Variable are still sent as the basic SI unit.
HOME
-- Units --
Figure 3-26. Display Units Menu
Menu
Units…
Back…
Select the desired parameter and change the units as follows:
Gas measurement units: ppm, %, ppb, mg/Nm3
Pressure measurement units: hPa, psig
Temperature measurement units: C, F
ppm to mg/Nm3 conversion factor: The ppm to mg/Nm
3
conversion factor in the range of 1 to 100000 according to the formula:
mg/Nm3 =
MW
Molecular weight
NO 30.0 NO
46.0
2
CO 28.0 SO O
64.1
2
32.0
2
Lower explosion limit (LEL) and upper explosion limit (UEL):
These parameters are used for the PMD, FID2 and NDIR modules and do not apply to the CLD. They should both be set at
0.00%.
i. Physical Measurements & Pressure
Limits
This menu is used to display the physical diagnostic parameters of the CLD ana­lyzer and to set the various pressure lim­its.
INFOESCAPE
In the Physical measurements display, the normal values for the displayed values are:
Sample capillary pressure: 7 psig Ozone supply pressure: 15 psig Ozonator status: On Ozonator power: Enabled Bypass flow: 800 cc/min
3-18 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 47
Model NGA2000 CLD
play
Instruction Manual
760000-C
November 2002
Converter temperature: 300 °C Ozonator temperature: 58 °C Sensor temperature: 0.5 °C Block temperature: 51.5 °C Capillary flow rate: 200 cc/min
CLD 7.50 ppm
Sample capillary pressure: 7.0 hPa Ozone supply pressure: 15.0 hPa
Converter temperature: 300.0 C Sensor temperature: 50.0 C Block temperature: 51.5 C
Physical Measurements Manufacturer’s settings
From the “Physical Measurements” menu (Figure 3-14), select “Pressure limits…” submenu to view and set the alarm pres­sure limits for the sample capillary and the ozonator.
CLD 7.50 ppm
-- Pressure Limits --
HOME
ESCAPE
INFO
Figure 3-29. Physical Measurements –
Manufacturer’s settings Display

j. Single Component Display Parameters

Sample capillary upper limit: 490.0 hPa Sample capillary lower limit: 50.0 hPa Ozone supply upper limit: 1050 hPa Ozone supply lower limit: 700.0 hPa
This function and menu is used to estab­lish the parameters to be displayed on the single component display.
Barometric pressure: 1013 hPa
HOME
ESCAPE
Figure 3-27. Pressure Limits Menu
The nominal values are:
INFO
Main Menu
Analyzer and I/O expert controls & setup…
Analyzer module setup…
Physical measurements parameters…
Capillary upper limit: 7.1 psig Capillary lower limit: 0.7 psig Ozone upper limit: 15.2 psig Ozone lower limit:: 10.2 psig
In the “Physical Measurements” menu, the “MORE” (F3) softkey displays the submenu selection for Temperature limits as shown on the next page. The “HIS­TORY” (F4) softkey displays the values for pressure and temperature as originally set at the factory.
CLD 7.50 ppm
Converter upper limit: 500.0 C Converter lower limit: 150.0 C Ozonator upper limit: 65.0 C Ozonator lower limit: 40.0 C Sensor upper limit: 55 C Sensor lower limit : 45 C Block upper limit: 55.0 C Block lower limit: 45.0 C
-- Temperature Limits --
HOME
ESCAPE
INFO
Figure 3-28. Temperature Limits Menu
CLD 7.50 ppm
-- Dis
ed Parameters --
First line’s parameter: Sample flow Second line’s parameter: Sample press Third line’s parameter: Ozonator Fourth line’s parameter Converter temp
Displayed concentration digits: 6 Digits after decimal point: 2
HOME
INFOESCAPE
Figure 3-30. Displayed Parameters
Menu
Here you can change the parameters that are displayed in the Single Component Display.
It is also possible to set the concentration value precision and number of digits. This does not change the inherent precision of the analyzer.
Select a line and press the softkey. To change the value use the and soft­keys. The values will scroll through the al­lowable selections as follows:
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-19
Page 48
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
Sample flow
Displays the sample flow rate in ml/min with a bargraph.
Sample press
Displays the sample pressure with a bar­graph.
Ozonator
Displays the status of the Ozonator: ON, Disabled, OFF-PRES.SW.
Converter temp
Displays the converter temperature with a bargraph.
Block temp
Displays the detector block temperature with a bargraph.
Detector temp
Displays the photodiode detector tem­perature with a bargraph.
NO/NOx
Displays the current mode, NO or NOx
Noise level
Displays the concentration noise level in ppm in the range 0.0 to 0.1 with a bar­graph.
t90 time
Displays the t90 response time for the displayed range with a bargraph.
Output delay time
Displays the output delay time setting with a bargraph.
Calibration status
Displays the calibration status: READY, Calibration
Linearizer
________________
Measurement mode
________________
Operational state
Standby, Calibration, Ready
Health
________________
Interference
Off, On
Validity
Validity of the concentration measure­ment: Valid, Invalid (Goes to Valid after a successful calibration)
Raw signal
Displays the raw concentration output of the detector before any linearization or other correction.
3-20 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 49
Model NGA2000 CLD
Instruction Manual
760000-C
November 2002

3-8 CALIBRATION PROCEDURE

The CLD analyzer module may require peri­odic calibration with known zero and span gases in order to maintain a desire level of analytical accuracy. It is recommended, after initial startup, that the CLD Analyzer Module is calibrated at least once every eight hours. This practice should continue until evidence indicates that some other interval is more ap­propriate depending on the analytical accu­racy required.
Calibration is the process of flowing known zero or span calibration gas into the analyzer for a specified period (averaging time), after which the analyzer will automatically set its zero or span factors so that the concentration measurement equals the calibration gas value. A limit can be set, beyond which any attempt by the analyzer to reset its concentra­tion measurement will cause a warning alarm. In this case, user intervention would be re­quired to reset the alarm and attempt another calibration.
There are three methods for performing a calibration with the CLD Analyzer Module and NGA Platform:

Basic Controls calibration:

(See Section 3-
8b, page 3-22)
This method allows the user to input a span calibration gas value and perform a zero or span calibration for each of the four ranges. This method uses calibration parameters es­tablished in the module setup menus. (See Section 3-8a, page 3-21)
Expert Controls calibration
: (See Section
3-8c, page 3-23)
This method allows the user to perform a zero or span calibration for each of the four ranges, enable or disable the calibration adjustment limits, view the results, and view or change the Factors that the analyzer uses to adjust the zero and span concentration reading. This method uses calibration parameters estab­lished in the module setup menus. (See Sec­tion 3-8a, page 3-21)

System calibration and setup:

This method allows the user to establish com­plex automated calibration sequences for modules bound to the Platform. This is fully described in the Platform manual.
NOTE
If zero calibration is done with other than a true zero gas followed by a span calibra­tion, it may be necessary to repeat the calibration. This is due to the slope/intercept effect where the subse­quent span calibration may change the zero crossing point.

a. Calibration Setup

Calibration Gas List
This menu is used to set the concentra­tion values of the calibration gases for each range. These values are used for all calibrations except in Basic where the span gas can be entered for a quick man­ual calibration.
Main Menu
Analyzer and I/O expert controls &
setup…
Analyzer module setup…
Calibration gas list…
CLD 7.50 ppm
Zero gas – range 1: 0.00 ppm NO span gas – range 1: 10.00 ppm NOx span gas – range 1: 10.00 ppm
Zero gas – range 2: 0.00 ppm NO span gas – range 2: 10.00 ppm NOx span gas – range 2: 10.00 ppm
Calibration…
HOME
-- Calibration Gas List --
MORE
INFOESCAPE
Figure 3-31. Calibration Gas List Menu
Use the MORE (F3) softkey to display and change the values for the remaining ranges.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-21
Page 50
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
The “Calibration…” submenu provides quick access to the Expert Controls Cali­bration menu, Section 3-8c, page 3-23.
Calibration Parameters
This menu provides various parameter settings for all calibration performed from Basic or Expert modes.
Analyzer and I/O expert controls &
Analyzer module setup…
Calibration parameters…
CLD 7.50 ppm
Calibration adjustment limits: Enabled Calibration averaging time: 5 s Calibration failure alarm: No Cal failure error allowed: 50 %
Calibration time out: 60 s Zero ranges: TOGETHER Span ranges: SEPARATELY
HOME
Figure 3-32. Calibration Parameters
Calibration adjustment limits:
Set to “Disable” to recover from a calibra­tion failure.
Calibration averaging time: Sets the time used by the analyzer to av-
erage its reading during calibration. A longer time will give a better calibration.
Main Menu
setup…
-- Calibration Parameters --
Display
Calibration time out:
Sets how long the analyzer will wait for the signal to stabilize before issuing a Warning.
Zero (Span) ranges:
Used to select whether to calibrate ranges TOGETHER or SEPARATELY. If to­gether, zeroing or spanning will go through each range one by one. If the change required is too great, it will fail and send an alarm if warning alarms are en­abled. In this case, Disable Calibration Adjustment Limits and try again. First check that the calibration gases are cor­rect. If non-zero gases are used, or the changes are great, zero and span may have to be repeated a few times.

b. Basic Controls Calibration

This method allows the user to input a span calibration gas value and perform a zero or span calibration for each of the four ranges.
INFOESCAPE
In the “Basic Controls” menu, move the cursor to the “Span gas concentration:” line and set the correct value for the cali­bration span gas. The zero gas is as­sumed to be 0.00 ppm.
Begin flowing the calibration gas (zero or span) and allow time for the analyzer to stabilize on the gas.
Press the appropriate ZERO (F3) or SPAN (F4) softkey to display the calibra­tion menu as shown below.
Calibration failure alarm: When turned on (Yes), issues a warning if
The “Calibration status” must be READY in order to initiate a calibration.
the analyzer has to change its calibration by more than the Cal Failure Error, if warning alarms are enabled.
Verify that the desired measurement range is active. If not, press ESCAPE (F2) to return to the previous menu and
Cal failure error allowed: The percentage by which the calibration
can change before an alarm is triggered if the Calibration Failure Alarm is enabled.
3-22 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
change the range.
The “Zero (Span) ranges” tag indicates if the ranges will be calibrated together or separately. See Section 3-8a, page 3-21 to change this.
Page 51
Model NGA2000 CLD
p
j
Instruction Manual
760000-C
November 2002
The “Calibration time” begins to count af­ter the calibration is started to show the elapsed time.
Press the ZERO (F3) [SPAN (F4)] softkey to begin the calibration. The “Calibration status” will display ZEROING-WAIT and the “Calibration time” clock will count the seconds. After the signal has stabilized the calibration will finish. If the signal does not stabilize within the timeout period (See Section 3-8a, page 3-21), the cali­bration will fail and another attempt will be started automatically. Make sure that the proper calibration gas is flowing.

c. Expert Controls Calibration

This method allows the user to perform a zero or span calibration for each of the four ranges, enable or disable the calibra­tion adjustment limits, view the results, and view or change the Factors that the analyzer uses to adjust the zero and span concentration reading.
In the “Expert Controls” menu (Figure 3-12), set the desired measurement range number to be calibrated. Change the measurement mode using the “NO/NOx toggle!” control.
Select “Zero/Span calibration…” menu line.
CLD 7.50 ppm
Measurement range number: 1 Zero gas concentration: 0.0 ppm Span gas concentration: 10.0 ppm Sample flow: 1300 ml/min Raw measurement signal: 5216063.5 Measurement gas: NO NO/NOx toggle! Status: READY Result… Calibration ad
HOME
-- Zero/s
ustment limits: Enabled
FACTORS INFOSPANZERO
Figure 3-33. Zero/Span Calibration
Menu
an calibration --
The zero and span gas concentrations (set in the calibration gas values menu) are displayed. See Section 3-8a, page 3­21 to change the values.
The sample flow rate is displayed along with the raw measurement signal. The sample flow rate should be in the range of 200 – 2200 ml/min.
The raw measurement signal can be used to set the FACTORS described in Section 3-8c, page 3-23.
The “Measurement gas:” tag shows the current measurement status as NO or NOx.
Press the ZERO (F3) [SPAN (F4)] softkey to begin the calibration.
CLD 7.50 ppm
Are you sure?
You must have zero gas flowing through the analyzer.
-- Analyzer Zero --
Calibration time: 0 s Measurement range number: 1 Zero ranges TOGETHER Calibration status: READY Error message for last zero: CAL OK
HOME
ESCAPE INFOZERO
Figure 3-34. Analyzer Zero Display
The “Calibration status” will display ZE­ROING-WAIT and the “Calibration time” clock will count the seconds. After the signal has stabilized the calibration will finish. If the signal does not stabilize within the timeout period (See Section 3­8a, page 3-21), the calibration will fail and another attempt will be started automati­cally. Make sure that the proper calibra­tion gas is flowing. See Section 3-8a, page 3-21 for “Calibration adjustment lim­its.”
Calibration Results The last calibration results can be viewed
by selecting the “Result…” submenu from the “Zero/span calibration” menu (Figure 3-33, page 3-23). The “Zero/span diag­nostic data” screen will be displayed.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-23
Page 52
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
CLD 7.50 ppm
Date of last zero: Apr 28, 2001 Error message for last zero: CAL OK Error percentage for last zero: -5 Raw signal at last zero: 0 Last zero gas would read: 0.000 ppm Date of last span: Arp 28, 2001 Error message for last span: CAL OK Error percentage for last span: 25 Calibration status: 789542.0 The last span would read: 10.000 ppm
HOME
-- Zero/span diagnostic data --
ESCAPE INFOFACTORS
Figure 3-35. Zero/Span Diagnostic
Data Menu
The errors are expressed as a percentage of the range.
The last zero and span readings are how the analyzer would read on those gases with the current calibration results or fac­tors.
Calibration Factors
Calibration Factors can be used to manu­ally set a calibration by flowing calibration gas and setting the Factor for zero or span until the reading conforms to the calibration gas value. In this way, the user can force the analyzer to any desired reading. Alternately, the Factors can be viewed and recorded after an automatic calibration.
NOTE
CLD 7.50 ppm
Only those factors appropriate for the current range Will affect the reading on the current range. Make sure you are using the right ones! Measurement range number: 1
Range 1 factors… Range 2 factors… Range 3 factors… Range 4 factors…
HOME
-- Calibration Factors --
ESCAPE INFO
Figure 3-36. Calibration Factors Menu
Move the cursor to the “Range _ fac­tors…” line corresponding to the selected range and press . The “Range _ Fac­tors” menu will display.
CLD 7.50 ppm
Zero offset: 524287.0 Span factor: 0.000021250 Full scale range at calibration: 250.0 PPM Measurement range number: 1
Raw measurement signal: 522819.7
-- Range 1 Factors --
HOME
STORE INFONEXT HISTORY
Figure 3-37. Range Factors Menu
Flow the appropriate gas (zero/span) while adjusting the corresponding factor (Zero offset/Span factor) until the concen­tration reading is stabilized on the desired calibration gas value.
If zero calibration is done with other than a true zero gas followed by a span calibration, it may be necessary to re­peat the calibration. This is due to the slope/intercept effect where the sub­sequent span calibration may change the zero crossing point.
In the “Zero/span Calibration” menu (Figure 3-33, page 3-23), verify the “Measurement range number:” is set to the desired range. If not, move the cursor to the “Measurement range number:” line
The factors take effect after pressing the softkey. With zero gas, the zero factor should be the same as the raw reading.
When the correct zero and span calibra­tion is achieved, press the STORE (F2) softkey to save the factors.
Use the “Measurement range number” line to change the range and the NEXT (F3) softkey to display the factors for the next range.
and change the setting. Press the FACTORS (F2) softkey. The “Calibration Factors” menu is displayed.
3-24 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 53
Model NGA2000 CLD
Instruction Manual
760000-C
November 2002
CLD 7.50 ppm
Zero offset: 0.000000 Span factor: 0.000000
Zero offset: 524287.0 Span factor: 0.00021250
HOME
NEXT INFORSTRMN RSTRST
Figure 3-38. Range Factors -
Manufacturer’s/Stored Settings Display
Press the HISTORY (F4) softkey to view the current stored factors versus the manufacturer’s (factory) settings. Use the RSTR MN (F3) softkey to restore the manufacturer’s settings and the RSTR ST (F4) softkey to restore the stored settings.

d. Unable to Calibrate

If the user is unable to calibrate the Ana­lyzer Module (i.e., when ZERO or SPAN is initiated, nothing happens), a possible solution relates to the use of an incorrect gas for zeroing or spanning (e.g., using a high concentration gas to zero or a zero gas to span the Analyzer Module). Simply recalibrating with the appropriate gas(es) will not correct the problem because the ZERO OFFSET or SPAN FACTOR has been set to an extreme value in the proc­ess.
To remedy the problem, do the following:
1. Select the following from the Main Menu: “Analyzer and I/O Expert
Range 1 Factors
Manufacturer’s settings
Stored settings
Controls & Setup,” “Analyzer Mod­ule Set Up,” and “Calibration Pa­rameters...”
2. Using the ↓ arrow, select “Zero
Ranges:”, press ENTER and, using the up/down arrows, toggle to SEPARATE. Do the same for the “Span Ranges:” selection. Do not press ESCAPE at any time unless retention of prior settings is desired.
3. Return to the “Main Menu” and make the following selections: “Ana­lyzer and I/O Expert Controls & Setup,” “Analyzer Module Controls,” “Zero/Span Calibration,” FACTORS (F3) softkey, and Range 1 (2, 3, 4) Factors (do Steps 4 and 5 for each range).
4. Select “Zero Offset,” press ENTER, adjust the value to 32700 with the and arrow softkeys, and press ENTER. Do not press ESCAPE at any time unless retention of prior settings is desired.
5. Select “Span Factor,” press ENTER, adjust the value to 0.00015 with the and arrow softkeys, and press ENTER. Do not press ESCAPE unless retention of prior settings is desired.
6. Attempt to recalibrate the Analyzer Module according to the procedure outlined in Section 3-8c, page 3-23. If recalibration fails, return to the Range Factors menu, readjust Zero Offset and Span Factor values, and try calibrating again.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-25
Page 54
Instruction Manual
y
y
y
760000-C November 2002
Model NGA2000 CLD
3-9 SYSTEM & NETWORK I/O MODULE CON-
TROLS (SETUP) – SYSTEM SIO
This menu provides access to several sub­menus for setting parameters of the SIO (Sig­nal Input/Output) and DIO (Digital Input/Output) of the analyzer.
Main Menu
Analyzer and I/O, expert controls &
setup…
System & network I/O module controls…
CLD 7.50 ppm
System SIO module… System DIO module…
stem & Network I/O Module Controls --
-- S
Measure
>>>Back…<<<
Figure 3-39. System & Network I/O Module
Controls Menu
Press the or softkeys to change to the desired submenu.
If there is no SIO module installed in the ana­lyzer, a corresponding message will be dis­played instead of the menu.
The “System SIO Module” menu provides submenus for setting up the output configura­tions of the SIO signals. The SIO board can contain 2 to 8 analog outputs, a serial inter­face (RS232 or RS485), and three relay out­puts. General configuration of the SIO board is contained in its own manual. If the SIO board is installed in the analyzer, the line “Module installed:” must be set to “Yes.”
To access this menu, in “System & network I/O module controls…”, select “System SIO module…”.
CLD 7.50 ppm
Analog output setup…
Serial interface setup…
Relay outputs setup…
Module installed: Yes
stem SIO Module --
-- S
Measure
Back…
Figure 3-40. System SIO Module Menu
Select a line with the or softkeys.
Select the variable or change to the submenu with the or softkeys.
Select the variable parameter with the or softkeys.
Confirm the new value with the Enter soft­key or cancel and return to the last value with the F2 softkey.

a. Analog Output Setup

In the System SIO Module menu (Figure 3-40, page 3-26), select Analog output setup…
CLD 7.50 ppm
-- Anal
zer Modules --
Measure
Figure 3-41. Analyzer Modules Menu
Output number: Choose the desired analog output (1-8) to
set the parameters. The number of out­puts depends on the analyzer configura­tion as 2, 4, 6, or 8.
CLD: 1.0 MLT/CH1 MLT/CH2 MLT/CH3
>>>Back…<<<
3-26 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 55
Model NGA2000 CLD
g
g
g
Instruction Manual
760000-C
November 2002
Choose signal source module…
Select the “Analyzer Modules” submenu by selecting the “Choose signal source module…” line and pressing the ↵ soft- key.
Select the tag of the desired reference channel with the or softkeys and then press the or softkey. The display will return to the previous menu automatically and the selected reference channel will be displayed in the “Source module:” line.
The available selections may be different depending on the installed modules. Choose signal…
CLD 7.50 ppm
Output number: 1 Choose signal source module… Choose signal… Signal value for 0 % output: 0.00 Signal value for 100 % output: 100.00 Output current: 0…20 mA Hold output during calibration: No Signal name: Sample flow Current signal value: 8.60 Source module: CLD
Measure
-- Analo
Output Setup --
More…Back…
CLD 7.50 ppm
-- Si
nals --
Sample flow:
Sample press:
Ozonator:
Converter temp:
Block temp:
Detector temp:
NO/NOx:
Noise level:
Measure
CLD 7.50 ppm
-- Si
nals --
>>>Back…<<<
t90 time:
Output delay time:
Measure
Calibration status:
Linearizer:
Measurement mode:
Operational status:
Health:
Interference:
>>>Back…<<<
Figure 3-43. Signals Menu
Signal value for 0% (100%) output:
It is possible to set the signal value for 0% output and for 100% output so as to out­put only a portion of the entire range.
Figure 3-42. Analog Output Setup Menu
Example:
Choose signal…
Select the “Signals” submenu by selecting the “Choose signal…” line and pressing the ↵ softkey. (The list of signals will de- pend on the module chosen.)
Range from 0 to 1000 ppm
0% value to be 400 ppm, 100%
value to be 700 ppm.
Analog output normally: 0V = 0
ppm, 10v = 1000 ppm
After changing the output scaling:
Press the F5 softkey to go to additional
0V = 400 ppm, 10V = 700 ppm
menus to choose the Primary Variable signal for the analog output. The Primary Variable is the actual NO or NOx concen­tration.
Move the cursor to the “Signal value for 0% output:” line and adjust the value to
400. Then change to the “Signal value for
100% output:” line and adjust the value to The signal chosen here will be applied to the analog output (1-8) chosen above.
See Section 3-7b, page 3-13 for a list of the signals and their values.
700.
NOTE
If the measurement range is changed,
the settings done in this menu will re-
vert back to the standard values of the
range. The output values can be
changed permanently in the menu
“Range Settings.” See Section 3-7a,
page 3-12.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-27
Page 56
Instruction Manual
p
760000-C November 2002
Model NGA2000 CLD
NOTE
The signal range of the analog output should not be less than the smallest range of the channel. Otherwise the analog output may exhibit excessive noise.
Output current:
Select the desired output range in the “Output current range:” line. The options are 0…20 mA or 4…20 mA.
Hold output during calibration:
Enable this option to hold the analog out­put to the last value during calibration.
Pressing the F5 (More…) softkey changes to the submenus “Output Signal if Assigned Module Fails” and “Fine Ad­justment.”
CLD 7.50 ppm
-- Out
Output(s) value on analyzer failure: BeginOfRange – 10 %
Output number: 1
Operation mode: Normal Fine adjustment for 0 % output: 4096 Fine adjustment for 100 % output: 819
ut Signal if Assigned Module Fails --
-- Fine Adjustment –
Operation mode:
Normal: The absolute measurement sig-
nal will be sent to the analog output.
Adjust 0V: Used to set the display equal
to the analog output for 0V and 0 mA. Life
zero signals (4-20 mA and 2-10V) are set
automatically and cannot be adjusted.
Adjust 10V: Used to set the display equal
to the analog output for 10V and 20 mA.
Select the “Fine adjustment for 0% out-
put” and/or “Fine adjustment for 100%
output” lines with the or softkey. Ad-
just to the desired value with the or
softkey and confirm with the softkey.
The range of values are:
3000 to 6000 for 0% (default 4096)
600 to 1000 for 100% (default 819)
The last three lines of the “Analog Output
Setup” menu are display only for configu-
ration values of the analog output.
Signal name: The name of the signal
chosen in the “Choose signal” menu.
Measure
More…Back…
Figure 3-44. Output Signal If Assigned
Module Fails Menu
Output(s) value on analyzer failure:
Choose the desired signal level to cause a failure condition. The choices are:
Actual BeginOfRange EndOfRange BeginOfRange-10% BeginOfRange+10%
Output number:
Choose the output number (1-8) for set­ting the fine adjustment.
Current signal value: The current value
of the variable.
Source module: The name of the module
chosen in the “Choose signal source
module” menu.
Pressing the F5 (More…) softkey
changes to the submenu “Special Scaling
for Concentration Signal.”
CLD 7.50 ppm
-- Special Scaling for Concentration Signal – (Scaling is the same as range limits)
Output #1: Yes Output #2: Yes Output #3: Yes Output #4: Yes Output #5: Yes Output #6: Yes Output #7: Yes Output #8: Yes
Measure
More…Back…
Figure 3-45. Special Scaling for
Concentration Signal Menu
3-28 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 57
Model NGA2000 CLD
g
p
Instruction Manual
760000-C
November 2002
This menu allows for the setting of each of the 8 outputs to be the same as the range limits “Yes” or as set on the previ­ous menus.
See Section 3-7a, page 3-12 for setting the range limits.
Pressing the F5 (More…) softkey changes to the submenu “Analog Output Updates per Second.”
CLD 7.50 ppm
Output #1: 10 Output #2: 0 Output #3: 0 Output #4: 0 Output #5: 0 Output #6: 0 Output #7: 0 Output #8: 0
Measure
-- Analo
Figure 3-46. Analog Output Updates per
This menu allows for the setting of the update rate for each of the 8 outputs.

b. Serial Interface Setup

The submenu “Serial Interface Setup” is used to set the parameters for data trans­fer between the analyzer and external de­vices. The choices in this menu depend on the configuration of the analyzer. The full specification of the serial interface is described in its own manual.
In the System SIO Module menu (Figure 3-40, page 3-26), select “Serial interface setup…”
CLD 7.50 ppm
Baud rate: 19200 Data bits: 8 Stop bits: 1 Parity: None Echo mode: Disabled Handshake: Xon/Xoff Transmission delay: 0 Type of installed serial interface: RS232 Communication protocol: AK Special protocol definitions…
Measure
Figure 3-47. Serial Interface Setup
Output Updates per Second --
Back…
Second Menu
-- Serial Interface Setu
Menu
--
Back…
Options:
Baud rate: 300, 1200, 2400, 4800, 9600, 19200
Data bits: 7, 8 Stop bits: 1, 2 Parity: None, Even, Odd Echo mode: Enabled, Disabled Handshake: None, Xon/Xoff Transmission delay: 0…100.
Type of installed serial interface: RS232, RS485/2w, RS485/4w, RS485/4w bus, None
Communication protocol: AK, MODBUS RTU, None (not applicable to CLD)
The “special protocol definitions…” line accesses a submenu for setting the parameters of the AK and MODBUS TRU communication protocols.
CLD 7.50 ppm
Device address (RS-485 only): 1
Measure
Figure 3-48. AK Protocol Definitions
The value can range from 1 to 50.

c. Relay Outputs Setup

There are three relays on the SIO board. The contact logic can be set with a jumper on the SIO board to select NO (normally open) or NC (normally closed). Full details of the SIO board are contained in its own manual.
In the System SIO Module menu (Figure 3-40, page 3-26), select “Relay outputs setup…”
NOTE
-- AK Protocol Definitions --
Back…
Menu
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-29
Page 58
Instruction Manual
y
g
t
760000-C November 2002
Model NGA2000 CLD
CLD 7.50 ppm
Output number: 1 Invert signal: Disabled Choose source module… Choose signal…
Signal comes from: Control module Signal name: Failure Actual status: Off
Measure
-- Rela
Outputs Setup --
Back…
Figure 3-49. Relay Outputs Setup Menu
Output number:
Corresponds to the relay number 1-3.
The list of modules will depend on the in­stalled modules.
Choose signal
CLD 7.50 ppm
Measure
-- Choose Si
nal --
Function control
Maintenance reques
Zero in progress
Span in progress
Back…
Failure
Cal. In progress
Zero failed
Span failed
Figure 3-51. Choose Signal Menu
Choose desired signal for the relay output
Invert signal:
number (1-3) being configured. “Disabled” signal is normal, “Enabled” signal is inverted.
The list of signals will depend on the cho-
sen module. If available, press the >>>
Choose source module..
CLD 7.50 ppm
-- Choose Source Module --
Control module: 0.0
CLD: 1.0
(F5) softkey for additional signals.
The three lines displayed at the bottom of
the “Relay Outputs Setup” menu show the
current status of the selected relay output.
Signal comes from: The module chosen
from the “Choose Source Module” menu.
Measure
Back…<<< >>>
Figure 3-50. Choose Source Module
Menu
Signal name: The signal chosen from the
“Choose Signal” menu.
Actual status: The current status of the
signal; Off or On.
Choose desired source module for the re­lay output number (1-3) being configured.
3-30 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 59
Model NGA2000 CLD
y
Instruction Manual
760000-C
November 2002
3-10 SYSTEM & NETWORK I/O MODULE CON-
TROLS (SETUP) – SYSTEM DIO
Selecting “System DIO module…” from the “System & Network I/O Module Controls” menu (Figure 3-39, page 3-26) provides sub­menus for setting up the output configurations of the DIO signals. The DIO board is com­prised of 8 digital inputs and 24 digital out­puts. Functions of supported analyzer modules can be attached to each input and a signal to each output. Further detailed infor­mation about the DIO board is contained in its own manual.
If there is no DIO module installed in the ana­lyzer, a corresponding message will be dis­played instead of the menu.
Main Menu
Analyzer and I/O, expert controls & setup…
System & Network I/O Module Controls…
System DIO module…
CLD 7.50 ppm
Input number: 1 Output number: 1 Choose module… Choose signal… Invert signal: No Module status: ??? Slot ID: ??? Signal name: ??? Signal level: 000.0 Signal comes from: ???
Measure
stem DIO Module --
-- S
Back…
Figure 3-52. System DIO Module Menu
For detailed information on the installation and setup of the DIO module, see the NGA 2000 Platform manual P/N 760006.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-31
Page 60
Instruction Manual
y
y
760000-C November 2002
Model NGA2000 CLD
3-11 SYSTEM CONFIGURATION AND DIAGNOS-
TICS
This menu and its submenus provides for setup of the system parameters for the plat­form.
Main Menu
System configuration and diagnostics…
CLD 7.50 ppm
stem Configuration and Diagnostics --
-- S
System calibration… Diagnostic menus… Load/Save configuration (CM/MCA) Date and time… Security codes… Network module management… System reset… Pump 1: Off Pump 2: Off System tag:
Measure
Channel
Back…
Figure 3-53. System Configuration and
Diagnostics Menu
The following is a short overview of the con­tents of the menus:
1
System calibration…
Diagnostic menus…
Control module diagnostics…
1
Analyzer module diagnostics… Software error messages
Loading/saving configuration parame­ters
Sending or loading of analyzer configuration data by the serial interface
Date and time…
Date and time setup of the analyzer

a. Diagnostic Menus

From the “System Configuration and Di-
agnostics” menu (Figure 3-53, page 3-
32), select “Diagnostic menus…”.
CLD 7.50 ppm
Control module diagnostics… Analyzer module diagnostics…
Measure
Figure 3-54. Diagnostic Menu
This menu has two submenus for viewing
and resetting any software errors.
The “Control module diagnostics” menu is
not applicable to this analyzer model.
Analyzer Module Diagnostics
From “Diagnostic Menus” (Figure 3-54,
page 3-32), select “Analyzer module di-
agnostics…”
CLD 7.50 ppm
Power supply voltages… Primary variable parameters… Physical measurement… Temperature control parameters… NO/NOx flow balance… Miscellaneous control parameters… Trend display control… Software diagnostic s… Alarm message valid for: FAILURE Start up analyzer… NOx converter efficienc
HOME
ESCAPE
-- Diagnostics Menus --
<<<
-- Analyzer Diagnostics --
Back…
>>>
INFO
Security codes…
Setup of security codes for the different oper­ating levels
Network module management…
1
System reset…
System reset and re-initializing of the analyzer
Figure 3-55. Analyzer Diagnostics Menu
This menu provides access to several
submenus for viewing and adjusting the
various CLD module parameters.
Choose one of the submenus to view or
adjust the parameters.
The “Alarm messages valid for:” functions
1
Not used in this module. See NGA 2000 Platform man-
ual PN 760006.
3-32 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
allows a choice of annunciation for these
choices:
Page 61
Model NGA2000 CLD
Instruction Manual
760000-C
November 2002
WARNING FAILURE SAFETY FAILURE
“Calibration factors…” see Section 3-8c. ANY
Physical measurement… Power supply voltages…
CLD 7.50 ppm
+15V analog is: 15.06 V +15V analog was: 0.0 V
-15V analog is: 15.26 V
-15V analog was: 0.0 V +5V digital is: 4.980 V +5V digital was: 0.0 V +25V power is: 23.73 V +25V power was: 0.0 V +12V analog is: 11.98 V +12V analog was: 0.0 V
HOME
-- Analyzer Diagnostics -­Power supply voltages
ESCAPE
MORE
INFO
Figure 3-56. Power Supply Voltages Menu
The “is” lines show the current value of the power supplies. The “was” values show the values when the analyzer was manufactured. Changes of more than a few percent should be tracked. The 24V power supply may differ substantially if the Rosemount Analytical power supply is not used. Use the MORE (3) softkey to see additional values.
Refer to Figure 3-14, page 3-13. These are measurements made by the analyzer module to verify proper functioning and appropriate flows of sample and support gases, if any. See Section 3-7i, page 3­18 for a complete description.
Temperature control parameters…
CLD 7.50 ppm
Converter set point: 300 C Converter P gain: 0.015 Converter I gain: 0.000125 Converter bias: 0.6 C Converter temperature: 28.6 C Detector set point: 51.5 C Detector P gain: 0.100 Detector I gain: 0.000170 Detector bias: 0.4 C Detector temperature: 29.2 C
HOME
Temperature control
ESCAPE
MORE
INFO
Figure 3-58. Temperature Control Menu
(Screen 1)
Displays the parameters used by the temperature control PID algorithms. Al­lows setting of the warning alarms for the temperatures.
Primary variable parameters…
CLD 7.50 ppm
Raw measurement signal: 521651.7 Signal gain setting: 64 Current range: 1 Barometric pressure compensation: Enabled Pk-pk noise: 39.9 ppm Peltier device voltage: Failure Reference duty cycle: 0.2785000 Duty cycle coefficient: -0.0681700 (Temperature compensation coefficients…) Calibration factors…
HOME
Primary variables parameters
ESCAPE
INFO
Figure 3-57. Primary Variable Parameters
Menu
Shows the value of internal parameters used in the primary variable calculation.
Barometric pressure compensation may be enabled if another analyzer has a
Press the MORE (F3) softkey for the Ozonator settings.
CLD 7.50 ppm
Ozonator set point: 58.0 C Ozonator P gain: 0.100 Ozonator I gain: 0.000170 Ozonator bias: 0.4 C Ozonator temperature: 29.2 C
Minimum full-on block FET current: 1.000 A
Temperature control
HOME
ESCAPE
INFO
Figure 3-59. Temperature Control Menu
(Screen 2)
Displays the parameters used by the temperature control PID algorithms.
Allows setting of the warning alarms for the temperatures.
pressure measurement device and is set to report its reading.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-33
Page 62
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
NO/NOx flow balance

b. Load/Save Module Configuration

CLD 7.50 ppm
NO/NOx flow balance
From the “System Configurations and Di­agnostic” menu (Figure 3-53, page 3-32), select “Load/save configuration
NOx correction factor: 1.0000
Calculate factor using pressure ratio…
Calculate factor using span gas response ration…
HOME
ESCAPE
INFO
Figure 3-60. NO/NOx Flow Balance Menu
Since the sample flows through separate paths in the NO and NOx mode of meas-
(CM/MCA)…
This menu provides several functions to send or load configuration data of the analyzer through the serial interface. These functions are only available if an SIO with serial interface is installed.
NOTE
When loading configuration data all of the current configuration in the mem­ory will be overwritten.
urement, there are differences in the flows entering the reaction chamber of the two modes. The CLD analyzer uses a correc­tion factor in the NOx mode to compen­sate for the reduced flow rate. This correction factor can be entered manually or it can be calculated by the analyzer us­ing one of two methods. One is based on adjusting to respond equally to a span gas and the other is based on adjusting for equal capillary pressure.
To start the automatic measurement and flow balance calculation cycle, enter the appropriate menu and follow the direc­tions.
CLD 7.50 ppm
This procedure will calcul ate the NO/NOx flow balance correction factor by measuring the capillary head both the NO pressure used in both the NO and NOx modes and using the ratio between the two pressures as the correction factor.
Press the CALC softkey to start the measurement and calculation.
Status: Factor entered manually
HOME
Calculate factor using pressure ratio
ESCAPE
CALC
INFO
Figure 3-61. Calculate Factor Using
Pressure Ratio Menu
It is recommended to flow the gas at the same pressure and flow rate as the sam-
CLD 7.50 ppm
Send configuration to s erial interf ace! Load configuration from serial interface!
-- Load/Save Configuration (CM/MCA)
- BE CAREFUL with this functi on -
Replace current configuration with fac tory settings!
Measure
Back…
Figure 3-62. Load/Save Configuration
(CM/MCA) Menu
If asked, confirm with the F2 (Yes) softkey or cancel and go back to the menu page with the F4 (Back…) or softkey.
Send configuration to serial interface !
The configuration data in memory will be sent through the serial interface of the analyzer to an external computer or other device.
Load configuration data from serial in­terface !
Configuration data will be loaded into memory from an external computer or other device through the serial interface of the analyzer. The current configuration in memory will be overwritten.
ple gas.
3-34 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 63
Model NGA2000 CLD
A
Y
Instruction Manual
760000-C
November 2002
Replace current configuration with fac­tory settings !
Deletes the configuration in memory and re-establishes the factory default setting from the Flash-EPROM.

c. Date and Time

From the “System Configurations and Di­agnostic” menu (Figure 3-53, page 3-32), select “Date and time…”
CLD 7.50 ppm
Minutes: 0 Hours: 12
ear: 2000 Day: 1 Month: 2
Network updating: Enabled Current time: 08:45:35 February 01, 2000
Measure
Figure 3-63. Date and Time Menu
This menu is used to set the date, time and format for the analyzer.
Select a line with the or softkeys.
Press the or softkeys to select the parameter.
Select any digit with the or softkey and set a new value with the or ↓ soft- key.
Network updating:
Not used with this model.
-- Date and Time --
Set!
Back…
This menu is used to set the security codes for the three levels of security.
CAUTION
If a security code is lost or forgotten, there is no possibility of entering the locked security level.
CLD 7.50 ppm
Basic level security: Disabled
Expert level security: Disabled
System level security: Disabled
Define basic level security PIN… Define expert level security PIN… Define system level security PIN…
Measure
-- Security Codes --
Back…
Figure 3-64. Security Codes Menu
Use the function softkeys F1 to F5 to en­ter the numerical security code in the de­sired sequence. The numbers will appear in the “Actual PIN” line as they are en­tered. The characters displayed on the function softkeys cannot be entered as code numbers.
Setting the code numbers
Enter the submenu for the desired secu­rity level to set the PIN. The default val­ues are:
Basic level: 12345 Expert level: 54321 System level: 12345
Set up a new date or time:
Set the “Minutes,” “Hours,” “Year,” “Day,” or “Month” lines and make any desired adjustments. Press the F3 softkey to set the new time and date. The “Current time” line will change to reflect the new time and date set.

d. Security Codes

From the “System Configurations and Di­agnostic” menu (Figure 3-53, page 3-32), select “Security codes…”
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-35
Example:
CLD 7.50 ppm
The actual PIN is prepresented by the order in which they
Actual PIN: 12345
BCDE1
-- Define Basic Leve l Security PIN --
Press five softkeys in any order to define the PIN
are pressed, and shown numberically below.
Press the left arrow key when you are done.
FGHIJ2
KLMNO3
PQRST4
UVWXYZ5
Figure 3-65. Define Basic Level Security
PIN Menu
Page 64
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
Enable Security Code
Select the desired security level line to enable. Change the parameter to “En­abled.”
Once a locked security level has been entered, it will remain unlocked even after exiting to a different security level. To protect the level, press the F4
NOTE
(Lock..) softkey in the Main Menu after
CAUTION
If System level is enabled, it will not be possible to re-enter the Security Setup
returning from the locked level.
Press the softkey to return to the “Se­curity setup” menu.
and change back to Disabled without the code.
Entering a level locked by security code
For example, return to the Main Menu by pressing the F4 softkey twice.
Attempt to enter an enabled level by choosing the menu line. A new menu will appear requesting entry of the security code. Enter the correct code using the correct sequence of function softkeys. The asterisk (*) symbol will appear for each entry.

e. System Reset

From the “System Configurations and Di­agnostic” menu (Figure 3-53, page 3-32), select “System reset…”
CLD 7.50 ppm
-- System Reset --
System reset!
Measure
Are you sure?
Back…
If the code is incorrect, the message “Ready” will appear in the line and access to the locked level is prevented. If the code is correct, the display will change to the locked level after the last digit of the correct code is entered.
Figure 3-66. System Reset Menu
Resets the analyzer to the initializing mode which is the same as switching the power off and then on.
3-36 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 65
Model NGA2000 CLD
Instruction Manual
760000-C
November 2002
3-12 CONVERTER TEMPERATURE ADJUST-
MENT
The vitreous carbon converter used in this analyzer module must be checked periodically to assure that it is working at peak efficiency. The efficiency of the converter is typically 95% to 98%, that is, 95% to 98% of the nitrogen dioxide introduced to the module is reduced to nitric oxide. That is well above the 90% mini­mum required by the Environmental Protec­tion Agency. (Refer to 40 CFR 60, App. A, Method 20)
Two conditions reduce the efficiency of the converter:
The converter is operating at too low a tem-
perature and the efficiency drops or
The converter is operating at too high a
temperature and the nitrogen dioxide is re­duced to nitrogen, which is not detectable by the chemiluminescence reaction.
Initially, an interval of one week between con­verter efficiency checks is recommended be­cause high temperature operation changes conditions inside the converter. The active surface area of the vitreous carbon increases through use. Initially, when the surface area is low, the temperature at which converter effi­ciency peaks is relatively high. This peak temperature moves downscale as surface area increases, and less external energy is required to cause adequate conversion.
The nominal range of converter operational temperatures is 300°C to 400°C (572°F to 752°F). The current converter temperature can be viewed in the “Physical measurement” menu. To access the converter temperature adjustment:
Main Menu
Analyzer and I/O expert controls &
setup…
Analyzer module setup…
Physical measurements parameters…
Follow this procedure to optimize the operat­ing temperature of the converter:
1. Power up the module and allow it to stabi­lize at operating temperature (about one hour).
2. Check the Converter Temperature in the Physical Measurements menu (See Sec­tion 3-7b, page 3-13). Note the value for future reference.
3. Introduce a calibration gas of known (NO
) concentration into the analyzer and
2
note the concentration value determined when the full response has been achieved.
4. Change Converter Setpoint in the “Tem­perature Control” menu to 300°C (See Section 3-11a, page 3-32). Allow module 15 minutes to stabilize, recheck the con­centration value and note the value for later use.
5. Increase the Converter Setpoint value by 20°C, wait 15 minutes, and note the con­centration value. Repeat this step until ei­ther a converter efficiency of between 95% and 98% is obtained or the final 20°C increment produces an efficiency in­crease of less than one percent.
6. Decrease the Converter Setpoint value by 5°C, which places the converter at a tem­perature suitable for low ammonia inter­ference and efficient NO
7. Recheck the Converter Temperature value in the “Physical Measurements” menu, and compare it to the initially re­corded value.
conversion.
2
NOTE
Converter temperature is not a direct measure of converter efficiency. Tempera­ture measurement is for reference pur­poses only.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-37
Page 66
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
3-13 MEASUREMENT OF CONVERTER EFFI-
CIENCY
It is the responsibility of the user to measure efficiency of the NO2-to-NO converter during initial startup and thereafter at intervals ap­propriate to the application (normally once a month).
The reactant material used in the converter provides the optimum combination of high conversion efficiency and low ammonia inter­ference. Unlike most competitive analyzers, the NGA 2000 CLD Analyzer Module utilizes a reactant material that gradually becomes more efficient at a given temperature. Thus, after a period of use, operation at a lower temperature setpoint than initially required is possible.
a. Test Setup for Measurement of Con-
version Efficiency
A typical setup for measurement of con­version efficiency is shown in Figure 3-67, page 3-41. The test setup includes:
A cylinder of nitric oxide standard gas consisting of NO in N
The concentration of NO in the standard gas should be about the full-scale value of the range under test. The test sample supplied to the analyzer should contain a concentration of NO comparable to that in the samples that are to be analyzed. Al­ternatively, a higher concentration NO standard may be used if the test setup in­cludes provision for diluting it appropri­ately with zero air. Suitable standard gases are available from various suppli­ers. Stainless steel cylinders are com­monly used, but specially treated aluminum is preferred for low parts-per-million NO samples.
An ozone generator utilizing an ultraviolet lamp, not a corona discharge
A corona discharge ozone generator is undesirable because it may produce oxy­gen atoms, which can then combine with atmospheric nitrogen to form NO. The re-
2
sult can be an erroneously high value for the measured conversion efficiency.

b. Test Procedure

1. Measure converter temperature in
the sub-menu structure as de­scribed in Section 3-7i, page 3-18. Note present reading as a reference for comparison with subsequent readings.
2. Lower converter temperature to
300°C as described in Section 3­11a, page 3-32, Temperature Con­trol Parameters, and wait 15 min­utes for temperature equilibration.
3. (See Figure 3-67, page 3-41) Con-
nect the Model 958 Converter Effi­ciency Tester to the CLD Module, and follow Steps 4 through 17 below (as adapted from 40 CFR 60):
4. Attach the NO/N
air supply to C1, and the CLD Mod­ule inlet fitting to C3.
5. With the variable transformer off,
switch the CLD Module to NO mode, and close valve MV1.
6. Open valve MV2 until the CLD
Module SAMPLE Pressure Gauge reaches operating pressure and the BYPASS flowmeter indicates some bypass flow. Wait until stable read­ings are obtained by the CLD Mod­ule.
7. Zero and span the Analyzer output
to indicate the value of the NO con­centration being used. This value should be about 80% of full-scale. Record this concentration.
8. Open valve MV1 (air supply meter-
ing valve) and adjust to blend enough air to lower the NO concen­tration (as noted above) about 10%. Record this concentration.
9. Power up the ozonator, and in-
crease its supply voltage until the
supply to C2, the
2
3-38 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 67
Model NGA2000 CLD
Instruction Manual
760000-C
November 2002
NO concentration noted in Step 8 is reduced to about 20 percent of the concentration noted in Step 7. Wait for stabilization. NO formed from the NO + O
is now being
2
reaction.
3
There must always be at least 10 percent un-reacted NO at this point. Record this concentration.
10. Switch the CLD Module to NOx mode. Total NOx concentration is now output to the network for dis­play. Record this concentration.
11. Turn off the ozonator, and allow the Analyzer reading to stabilize. Total NOx concentration of the dilute NO span gas initially used is displayed. Record this concentration.
12. Close valve MV1. The NO concen­tration should be equal to or greater than the reading in Step 7. This in­dicates whether the NO contains any NO
.
2
In the initial measurement, after lowering the temperature setpoint in Step 2, the efficiency will nor­mally be less than 92%.
14. Reset converter temperature setpoint
20°C higher, wait 15 minutes for temperature equilibration, and measure conversion efficiency by re­peating Steps 3 through 13. Conver­sion efficiency should be improved.
15. Repeat Step 14 until: a) 95% to 98% efficiency is attained or b) the final 20°C converter temperature adjust­ment yields an increase in efficiency of less than 1%.
16. Reset converter temperature setpoint 5°C lower. Converter temperature is now set to the front edge of the pla­teau on the effi­ciency-vs.-temperature curve. (See
NOTE
Figure 3-68, page 3-42.) This setting
13. Calculate the efficiency of the NO converter by substituting the con­centrations obtained during the test
should provide the optimum combi­nation of high conversion efficiency and low ammonia interference.
in the equation below:
17. Wait 15 minutes for temperature
%efficiency = 1 + x 100
a - b
-
Where:
a = recorded concentration in Step 10
equilibration, and check converter temperature. Compare present tem­perature with original value. Nor­mally, converter temperature should be in the range of 300°C to 400°C (572°F to 752°F).
b = recorded concentration in Step 11 c = recorded concentration in Step 8 d = recorded concentration in Step 9
In the example in Figure 3-67B, page 3-41, the following calculations would apply:

c. Subnormal Conversion Efficiency

If a measured conversion efficiency of be­tween 95% and 98% is unobtainable within the normal temperature range, the most probable cause is depletion of the catalytic material within the converter.
%efficiency = 1 + x 100 = 92%
Efficiency checks should be made on each analyzer range, using an NO span gas concentration appropriate to the instrument range.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-39
80 - 85
-
However, before concluding that the con­verter is defective, ensure that the con­version efficiency measurement is accurate. Though the measured efficiency is less than the 95% to 98% range, the actual efficiency may be somewhat higher.
Page 68
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
An apparent subnormal efficiency can be due to a problem external to the Analyzer Module, perhaps located either within the test setup or between it and the Analyzer Module. Check the following:
1. Leakage.
2. Loss of NO Analyzer Module. Such loss can oc­cur by reaction with a rubber dia­phragm in a pressure regulator or flow controller. Stainless steel dia­phragms are preferred. Loss can also occur during passage through filter media.

d. Replacement of Converter

If the subnormal conversion efficiency is real, and not due to measurement error introduced by the test setup, the converter must be replaced. See Section 4-4, page 4-1.
between test setup and
2
The usual cause of converter failure is destruction of a large part of the catalytic material by excessive heat. This is due ei­ther to an excessively high temperature setpoint or failure of the converter tem­perature control circuitry.

e. Capillaries

Replacement vent and pump capillaries should be installed finger-tight. Use of a wrench can constrict capillaries, thus changing flow rate. Sample capillary is metal; use a wrench for tightening.

f. TEA Scrubber

The presence of NO2 in the NO cylinders can cause inaccurate converter efficiency values. The TEA Scrubber accessory (P/N NL635741) can be used to remove residual NO of this accessory allows an NO2-free NO calibration gas.
from the NO cylinders. Use
2
3-40 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 69
Instruction Manual
760000-C November 2002
A. TYPICAL TEST SETUP
FLOWMETER FM2
METERING VALVE MV2
Model NGA2000 CLD
CONNECTOR C3
TO SAMPLE INLET OF
OFF
115
ON
VAC
VARIABLE TRANSFORMER
T1
OZONATOR UTILIZING ULTRAVIOLET LAMP
FLOWMETER FM1
CLD ANALYZER MODULE
FLOW APPROXIMATELY 3 LITERS PER MINUTE
CONNECTOR C2
STANDARD GAS: NO IN N
BACKGROUND
2
ALL LINES AND FITTINGS STAINLESS STEEL OR TEFLON
B. TYPICAL TEST RESULTS
NO in
2
N
85
90
80
AIR ADDED
OZONATOR OFF
OZONATOR ON
CONNECTOR C1
ZERO AIR
METERING VALVE MV1
CAUTION: Externally limit sample flow rate to less than 2200 cc/min.
a-b
c-d
20
0
c
NO Mode
a
d
TO NOX MODE
NOX Mode
b
% Efficiency = 100
a - b
1+
( )
c - d
Figure 3-67. Measuring Efficiency of NO2 to NO Converter
3-42 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Page 70
Model NGA2000 CLD
A
bout
Instruction Manual
760000-C
November 2002
97%
Final 2 degree higher
adjustment
Final 1 degree lower
Converter Efficiency
adjustment
Converter Temperature
Figure 3-68. Conversion Efficiency as a Function of Converter Temperature
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-43
Page 71
Model NGA2000 CLD

MAINTENANCE AND SERVICE

Instruction Manual
760000-C
November 2002
SECTION 4
WARNING.
ELECTRICAL SHOCK HAZARD
Disconnect power to the module(s) prior to replac­ing components.
This equipment should not be adjusted or repaired by anyone except properly qualified service per­sonnel.

4-1 OVERVIEW

The CLD Analyzer Module requires very little maintenance during normal operation.
Occasionally, the intake fan screen may require cleaning, refer to Section 4-3, page 4-1.
Also, the detector's reaction chamber and sap­phire window may require cleaning, refer to Sec­tion 4-7, page 4-3.
White crystal deposits on the windows of the reac­tion chamber and plugging of capillaries and vent are usually due to sample contaminates such as ammonia reacting with the high ozone levels and NO components. To eliminate the contaminates, the sampling system should be reworked or a pre­ventive maintenance program developed (if drop­out is not excessive). Another source of crystalline formation is contaminated air.
Several components may require replacement. These are discussed in the following sections.

4-2 FUSES

The main power fuse may require replacement.
NOTE
Before replacing the fuse, remove power to the Analyzer Module.
See Figure 1-3, page 1-3 for the location of the Power Supply Board main power fuse, which protects 24 VDC input to the module.

4-3 FANS

Refer to Figure 4-1, page 4-2. To replace ei­ther rear panel fan, remove the cover of the Analyzer Module and then the rear panel. Disconnect connectors and remove screws. Assemble in reverse order.

4-4 CONVERTER

Refer to Figure 4-1, page 4-2 and Figure 4-2, page 4-3. To replace the converter or sensor, disconnect the two pneumatic tubes and two electrical connections. Unlace the heater blanket, and remove the converter. Reassem­ble in reverse order, ensuring that the con­verter is oriented with the glass cloth at the bottom and the sensor is oriented correctly in­side the heater jacket.

4-5 OZONATOR

Refer to Figure 4-1, page 4-2.To replace the ozonator, remove the two large straps and all tie-wraps, and disconnect the one electrical connection. Reassemble in reverse order.

4-6 PRINTED CIRCUIT BOARDS

All four printed circuit boards can be replaced, if necessary. Refer to Figure 4-1, page 4-2 for location of the Driver, Power Supply, Signal and Computer Boards.
To remove any PCB (except the Computer Board), disassemble the enclosure side first. Ribbon and other cables are long enough to allow the entire side to be folded out from the remainder of the components. This makes PCB removal much simpler.
Tag each connector and its location before disconnecting any wiring. This helps in reas­sembly.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-1
Page 72
Instruction Manual
760000-C November 2002
Optional Bypass Flow Sensor.
4
Ground wires from ozonator shown.
3
Flow direction of intake fan (shown) is into case. Flow direction of exhaust fan is out of case.
2
1
Fan guard between fan and EMI filter is on the intake fan only (shown).
PCB Insulators
Flow Balance Metering Valve 903207
Ozonator 659494
Signal Board 655580
Insulator
Computer Analysis Board 655520
Transistor 655264
Driver Board 655620
Sample Pressure Sensor 655253
Sample Regulator 655269
NO/NOx Solenoid 655263
Thermostat 657298
4
Flow Sensor 902931
Glass Tube Connectors
Pressure Switch 662298
Ozonator Power Supply 657716
Ozone Pressure Sensor 655254
Fuse, Power 903347
3
LON/Power Module
Fan Guard
Figure 4-1. CLD Module Assembly
Model NGA2000 CLD
Converter Assembly 655250 See Figure 4-2
1
Fan Guard
EMI Shield
2
Fan 655245
Detector Assembly 658160 See Figure 4-3
Detector Case Insulator
Power Supply Board 657520
4-2 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Page 73
Model NGA2000 CLD
r
r
r
r
Instruction Manual
760000-C
November 2002
ASSEMBLED SIDE VIEW
Senso
Glass
Cloth
4-7 DETECTOR DISASSEMBLY
Refer to Figure 4-3, page 4-5.

a. Reaction Chamber Removal

Disconnect the stainless steel tubing lines at the fittings. Remove the (4) nuts holding the Detector Assembly to the chassis. Discon­nect the plug from connector J1 on the Sig­nal Board and remove the assembly from the chassis.
NOTE
Care should be taken to avoid getting heatsink compound on optical surfaces. If this substance is removed during the dis­assembly process, a zinc-oxide-filled, silicone grease (e.g., Dow Corning 340 or EG&G Wakefield Engineering's Series 120 Thermal Joint Compound) be reapplied in the reassembly of this component.
Heate
Jacket
655228
Converte
Tube 655227
Wrap with aluminum foil
Figure 4-2. Converter Assembly
Remove the (2) screws holding the tube assembly in place. Hold the tubing with one hand while inverting the Detector Housing with the other, allowing the Re­action Chamber O-ring and window to be removed from below.

b. Reaction Chamber Installation

To reinstall, hold the housing in the in­verted position while sliding the Reaction Chamber O-ring and window into position and the tubing into the slot in the housing. Hold the Reaction Chamber in place while rotating the housing upright. Replace the hold-down screws.
The procedure described above is for the purpose of maintaining the relative positions of windows and O-ring to the Reaction Chamber during installation.
Senso
655282
NOTE
Although the heater and thermostat can be removed to facilitate handling, contact with the white heatsink compound can be mini­mized by leaving these items in place. Re-
Replace the top cap and screws. Reverse the removal procedure to reinstall the De­tector Assembly into the Analyzer Mod-
ule. move the (2) screws holding the top plate of the Detector , and move the plate along the wires and away from the Detector .
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-3
Page 74
Instruction Manual
760000-C November 2002
Model NGA2000 CLD

c. Photodiode Removal

Remove the Detector Assembly as described above. Invert the housing to access the mounting bracket. Remove the (3) screws and shoulder washers from the bracket. Re­move the bracket, insulating disk and bottom plate as a unit to minimize the spread of the heatsink compound.
Remove the (2) screws holding the lower section of the Detector Housing, then slide the section along the cable and remove.
Remove the (2) screws holding the socket, thermistor and photodiode in place, being careful not to lose the washers that are used as shims.
Grasp the socket and photodiode base while slowly rotating to separate the photodiode from the housing. Some friction will be felt as an O-ring is used around the photodiode as a seal.

d. Photodiode Installation

To replace the photodiode, carefully remove the diode from the green socket, and replace with a new one. Before mounting the new di-
ode, the top cap of the enclosure should
be temporarily removed and the (2)
screws holding the Reaction Chamber
loosened about two turns. This allows air
which is trapped between the O-ring seals
to escape when the diode is inserted. It
also maintains the position of the O-ring
and window in the upper compartment.
The new photodiode should be slowly in-
serted into the housing while gradually ro-
tating the body. This allows the O-ring to
properly seat. Continue replacing screws,
washers, thermistors, etc., with the thicker
shim (washer) on the opposite side of the
socket from the thermistor.
Replace the lower section of the housing,
then the bottom cover, insulator and
bracket with the shoulder washers and
screws.
Re-tighten the screws in the Reaction
Chamber (upper section). Replace the top
cap and its screws.
To reinstall in the Analyzer Module, re-
verse the procedure for removal as indi-
cated above.
4-4 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Page 75
Model NGA2000 CLD
A
Instruction Manual
760000-C
November 2002
Sapphire Window
Photodiode
Reaction Chamber
Thermistor
Assembly
Exhaust
Heater
*
M3X0.5 x 16 mm Screw (2) 3mm Spring Washer (2)
Photodiode Socket
ssembly
Detector Mounting Bracket
Heater
*
Thermostat
Tubing Cover
M3X0.5 x 25 mm Screw (2) 3mm Spring Washer (2)
Detector Header
Retainer Gasket
*
Reaction Chamber
O-Ring 854540
Sapphire Window
Cushioning Gasket
Ozone
Sample
Photodiode Cable
Lower Cover
Insulator (between Lower Cover and Mounting Bracket)
Photodiode Case Ground
M3X0.5 x 16mm Screw (3)
Nylon Shoulder Washers (3)
O-Ring 876478
Photodiode Assembly (see detail below)
M3X0.5 x 20 mm Screw (2)
(see detail below)
3mm Spring Washer (2)
Detector Cover
M3X0.5 x 16 mm Screw (2) 3mm Spring Washer (2)
Heater/Thermostat Assembly 655235.
*
Photodiode 655258
Thermistor 655216
Thermistor Shim
Thermistor Spacer
No. 6 Flat Washer (2)
Photodiode Socket Assembly
Assembly of Photodiode
Figure 4-3. Detector Assembly
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-5
Page 76
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
4-6 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Page 77
Model NGA2000 CLD

TROUBLESHOOTING

5-1 LEAKS

Liberally cover all fittings, seals, and other possible sources of leakage with a suitable leak test liquid such as SNOOP (part
837801). Bubbling or foaming indicates leak­age. Checking for bubbles will locate most leaks but could miss some, as some areas are inaccessible to the application of SNOOP. For positive assurance that system is leak free, perform one of the tests above.
Instruction Manual
760000-C
November 2002
SECTION 5
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 5-1
Page 78
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
5-2 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
Page 79
Instruction Manual
760000-C
Model NGA2000 CLD
November 2002
SECTION 6

REPLACEMENT PARTS

WARNING
PARTS INTEGRITY
Tampering with or unauthorized substitution of components may adversely affect safety of this product. Use only factory-approved components for repair.

6-1 MATRIX

Each analyzer is configured per the customer sales order. Below is the CLD sales matrix which lists the various configurations avail-
To identify the configuration of an analyzer, locate the analyzer name-rating plate. The sales matrix identifier number appears on the analyzer name-rating plate.
able.
CLD CHEMILUMINESCENCE NO/NOx ANALYZER - CLD (ANALYZER MODULE)
Code Software Version
01 Current Version Software 02 2.2.1 Version Software 03 3X Version Software - specify version
Code Configuration Identifier(1)
A Calibrated Low Ranges: 0-10, 0-25, 0-100, 0-250 ppm B Calibrated Low Ranges: 0-10, 0-30, 0-100, 0-300 ppm C Calibrated Low Ranges: 0-10, 0-50, 0-100, 0-500 ppm D Calibrated High Ranges: 0-250, 0-1000, 0-2500, 0-10000 E Calibrated High Ranges: 0-300, 0-1000, 0-3000, 0-10000 ppm F Calibrated High Ranges: 0-500, 0-1000, 0-5000, 0-10000 ppm G Calibrated Low Range 0-5 ppm 9 Special
Code Materials
1 Brass and Neoprene Back Pressure Regulator 5 psi 2 Brass and Neoprene Back Pressure Regulator 2 psi 3 Stainless Steel and Viton Back Pressure Regulator 5 psi 4 Stainless Steel and Viton Back Pressure Regulator 2 psi 9 Special
Code Flow Sensor
01 Standard (No Flow Sensor) 02 400 - 2000 cc/min 99 Special
Code Special Requirements
00 None G1 Customer Option 99 Special
CLD 01 A 3 01 00 Example
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-1
Page 80
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
6-2 REPLACEMENT PARTS
658157 Air Restrictor 430 cc/min 12 psig 657473 Capillary, Pump 5 psig 660405 Capillary, Pump 2 psig 660404 Capillary, Sample 200 cc/min 2 psig 659658 Capillary, Sample 200 cc/min. 5 psig 659657 Capillary, Sample 70 cc/min. 5 psig 655520 Computer Board 655250 Converter 659754 Detector 200 cc/min. 659753 Detector 70 cc/min. 655620 Driver Board 655245 Fan (Exhaust and Intake) 903347 Fuse, Main Power 6A 250V 659494 Lamp, Ozone Generator 657719 Ozone Generator 657520 Power Supply Board 657716 Power Supply, Ozone Generator 655253 Pressure Sensor, Sample 662298 Pressure Switch 659895 PROM 660400 Regulator, Brass/Neoprene 2 psig 655269 Regulator, Brass/Neoprene 7 psig 660401 Regulator, SS/Viton 2 psig 659063 Regulator, SS/Viton 5 psig 902931 Sensor, Flow 655254 Sensor, Ozone Pressure 655580 Signal Board 655263 Solenoid Valve, NO/NOx 657298 655264 Transistor Assembly 903207 Valve, Flow Balance
Thermostat 80°C
655250 Converter Replacement Parts
655228 Heater Jacket 655227 Tube, Aged 655282 Sensor, Temperature
659753, 659754 Detector Replacement Parts
655235 Heater/Thermostat Assembly 655216 Thermistor 655258 Photodiode 854540 O-Ring, Viton .739ID .875OD 876478 O-Ring, Viton .737ID .943OD
6-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management
Page 81
Model NGA 2000 TO2

RETURN OF MATERIAL

7-1 RETURN OF MATERIAL

If factory repair of defective equipment is re­quired, proceed as follows:
1. Secure a return authorization from a Rosemount Analytical Inc. Sales Office or Representative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted.
Rosemount CSC will provide the shipping address for your instrument.
In no event will Rosemount be responsi­ble for equipment returned without proper authorization and identification.
2. Carefully pack the defective unit in a sturdy box with sufficient shock absorbing material to ensure no additional damage occurs during shipping.
3. In a cover letter, describe completely:
The symptoms that determined the
equipment is faulty.
The environment in which the equip-
ment was operating (housing, weather, vibration, dust, etc.).
Site from where the equipment was
removed.
Whether warranty or non-warranty
service is expected.
Complete shipping instructions for the
return of the equipment.
4. Enclose a cover letter and purchase order and ship the defective equipment accord­ing to instructions provided in the Rose­mount Return Authorization, prepaid, to the address provided by Rosemount CSC.
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
SECTION 7

7-2 CUSTOMER SERVICE

7-3 TRAINING

Instruction Manual
760000-C
November 2002
If warranty service is expected, the defective unit will be carefully inspected and tested at the factory. If the failure was due to the condi­tions listed in the standard Rosemount war­ranty, the defective unit will be repaired or replaced at Rosemount’s option, and an oper­ating unit will be returned to the customer in accordance with the shipping instructions fur­nished in the cover letter.
For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions.
For order administration, replacement Parts, application assistance, on-site or factory re­pair, service or maintenance contract informa­tion, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
A comprehensive Factory Training Program of operator and service classes is available. For a copy of the Current Operator and Service
Training Schedule
vices Department at:
Rosemount Analytical Inc.
Customer Service Center
contact the Technical Ser-
1-800-433-6076
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 7-1
Page 82
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
7-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management
Page 83
Model NGA 2000 TO2
SECTION 8
A
ACKN, 3-14 Actual status, 3-30 Alarm delay, 3-14 Alarm generation, 3-14 Alarm messages valid for, 3-32 analyzer module diagnostics, 3-32 analyzer will calculate the best fit polynomial, 3-16 Applying power, 3-7 automatic measurement and flow balance
calculation, 3-34
Automatic range change control, 3-10
B
Barometric pressure compensation, 3-33 binding, 3-7 Block temp, 3-20 Brightness and Contrast, 3-11 bypass flow, 3-13
Instruction Manual
760000-C
November 2002

INDEX

installation location temperature restrictions, 2-2
power connection, 2-4 communication protocol, 3-29 Computer Board, 4-1 Contaminants, 2-4 converter operational temperatures, 3-37 Converter temp, 3-20 correction factor, 3-34 Current signal value, 3-28 Cursor Keys, 3-1 cylinder gases, 2-2
D
Date and time, 3-32 Detector Assembly, 4-3 Detector Assembly, 4-4 Detector temp, 3-20 Diagnostic menus, 3-32 DIO module, 3-31 Dow Corning 340, 4-3 Driver Board, 4-1
C
Cal failure error allowed, 3-22 calibration, 3-21
basic controls level, 3-21 expert controls level, 3-21
system, 3-21 Calibration adjustment limits, 3-22 Calibration averaging time, 3-22 calibration factors, 3-24 Calibration failure alarm, 3-22 Calibration Gas List, 3-21 Calibration Gases, 2-3 Calibration Parameters, 3-22 Calibration Results, 3-23 Calibration status, 3-20 Calibration time out, 3-22 capillaries, 3-40 cause of converter failure, 3-40 change the alarm limits, 3-13 chemiluminescence method of detection, 1-1 choose signal, 3-27 Choose signal, 3-30 choose source module, 3-30 choose the analyzer module, 3-27 CLD Analyzer Module
Installation into Platform, 2-1
E
efficiency of the NO2-to-NO converter, 3-38 EG&G Wakefield Engineering's Series 120 Thermal
Joint Compound, 4-3 Enable Security Code, 3-36 Enter Key, 3-1 Entering a level locked by security code, 3-36
F
fan screen, 4-1 Flash-EPROM, 3-35 Function Keys, 3-1
Back, 3-2
BasicCal, 3-2
Channel, 3-2
Display, 3-2
ESCAPE, 3-2
FACTORS, 3-24
HISTORY, 3-25
HOME, 3-2
INFO, 3-2
Lock, 3-2
Main, 3-2
Measure, 3-2
MFG Data, 3-2
Rosemount Analytical Inc. A Division of Emerson Process Management Index 8-1
Page 84
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
More, 3-2 NEXT, 3-24 RSTR MN, 3-25 RSTR ST, 3-25 Status, 3-2 STORE, 3-24
Function Line, 3-2
G
Gas measurement, 3-18
H
Health, 3-20 heater, 4-1, 4-3 heater jacket, 4-1 hold output during calibration, 3-28 hysteresis level, 3-17
I
I/O Module
installation, 2-1 Inlet/Outlet fittings, 2-2 Interference, 3-20 invert signal, 3-30
L
LCD brightness and contrast, 3-7 Leak Test, 2-4 linear polynomials, 3-14 linearization curve must be monotonic, 3-16 Linearizer, 3-20 Load configuration data from serial interface, 3-34 Loading/saving configuration parameters, 3-32 LON update rate, 3-17
M
main power fuse, 4-1 Maximum permissible interval between calibrations,
3-7 measured conversion efficiency, 3-39 Measurement mode, 3-20 Measurement range number, 3-10, 3-12 Menu Line, 3-1 Midpoint Correction Set Up, 3-16 Monotonic, 3-16
N
NO/NOx, 3-20 NO/NOx flow balance, 3-34 Noise level, 3-20
O
operation mode, 3-28
adjust 0V, 3-28 adjust 10V, 3-28
normal, 3-28 Operational state, 3-20 output current, 3-28 output delay time, 3-17, 3-20 output number, 3-30 Output number, 3-28 output(s) value on analyzer failure, 3-28 Ozonator, 3-20 Ozonator Power, 3-10 Ozonator Source Gas, 2-3
P
photodiode, 1-1, 3-20, 4-4 PIN, 3-35 point to be measured, 3-16 power failure, 3-1 Power Supply Board, 4-1 Power Supply Board main power fuse, 4-1 Power supply voltages, 3-33 ppm to mg/Nm3 conversion factor:, 3-18 Pressure, 2-3 Pressure measurement, 3-18 Primary variable parameters, 3-33
R
Range and functional control, 3-12 Raw signal, 3-20 Reaction Chamber, 4-3, 4-4 Replace current configuration with factory settings,
3-35 response time, 3-17 Return Authorization, 7-2
S
Sample, 2-3
maximum flow, 2-3 sample contaminates, 4-1 Sample flow, 3-20 Sample line tubing type, 2-2 Sample press, 3-20 security codes, 3-32 Security codes
enabled, 3-36
lost or forgotten, 3-35 select analog output number, 3-26 Send configuration to serial interface, 3-34 serial interface options, 3-29 Set up a new date or time, 3-35
8-2 Index Rosemount Analytical Inc. A Division of Emerson Process Management
Page 85
Model NGA 2000 TO2
Instruction Manual
760000-C
November 2002
Setting the code numbers, 3-35 shipping damage, 2-1 Signal Board, 4-1, 4-3 Signal comes from, 3-30 Signal name, 3-28, 3-30 signal value for 0% (100%) output, 3-27 SIO module, 3-26 SNOOP, 5-1 Source module, 3-28 Span gas concentration, 3-10 special scaling for concentration signal, 3-28 subnormal efficiency, 3-40 Switch automatically to, 3-11 Switch off backlight, 3-11 System calibration, 3-32 System reset, 3-32
T
t90 time, 3-17, 3-20 Tag Line, 3-2
TEA Scrubber, 3-40 temperature control PID algorithms, 3-33 Temperature measurement, 3-18 Text Line, 3-2 thermistor, 4-4
V
Validity, 3-20 Variable Line, 3-2 vitreous carbon converter, 3-37
reduced efficiency, 3-37
W
warm-up period, 3-7 warning alarms for the temperatures, 3-33
Z
Zero (Span) ranges, 3-22
Rosemount Analytical Inc. A Division of Emerson Process Management Index 8-3
Page 86
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
8-4 Index Rosemount Analytical Inc. A Division of Emerson Process Management
Page 87
Model NGA2000 CLD
MATRIX POSITION 12, VALUE “G”
1.0 OVERVIEW
The following special modifications to your standard CLD Analyzer (PN 194005) are made as specified by the “G” designator cho­sen for the 12 and supercedes the information supplied in this manual.
2.0 SPECIFICATIONS
Range: 10, 50, 300 PPM Noise: 1% Range T90 Response: 1-2 sec. Sample Flow: 3 L/min. Sample Pressure: 2 psig Conformity: Non-Linear
3.0 PHYSICAL MODIFICATIONS Flow Sensor Removal
The flow sensor (PN 902931) has been re­moved. This does not affect the operation, accuracy or safety of the analyzer. It means that the bypass flow is no longer measured. Therefore, the platform display has been modified to show:
th
position of the product matrix
Addendum
748371-E
February 2000
ADDENDUM
Sample pressure Block temperature Ozonator Converter temperature
The CLD physical parameter will now indicate zero sample flow, and generate a validity and warning alarm which will not be displayed and should be disregarded.
Component Changes
The following parts have replaced the stan­dard components:
Pump Capillary PN 660405
Detector 200 cc/min 2 psig PN 660404
Back Pressure Regulator 2 psig PN
660400
4.0 FLOW DIAGRAM
CLD module with value “G” in matrix position 12, Figure 1-1 Flow Diagram is replaced with Figure 1-1 on this addendum.
http://www.processanalytic.com
Page 88
Addendum
748371-E February 2000
Model NGA2000 CLD
TUBING, 1/8 O.D. FLOUROCARBON
OZONE GENERATOR
CONN, GLASS
TUBE (2)
BLACK
1/8 O.D.
1/8T
BRASS
NOx/NO CONVERTER
NC NO
NOx/NO SOLENOID VALVE
PRESSURE SWITCH
4-15 psig (NO)
SENSOR, OZONE
PRESSURE 30 psig
CONN, MALE
1/8T – 3/8NPT
BRASS
RESTRICTOR, AIR
130 cc/min. @ 12 psig
CONN
1/8T – 1/8NPT
BRASS
TUBING, VITON
.116 I.D.
CONN, MALE
1/8T – 3/8NPT
BRASS
CROSS
REDUCER, BULKHEAD
1/4T – 1/8T BRASS
E AIR
Unless otherwise specified all tubing is 1/8 O.D. teflon.
OZONE
CHAMBER
REACTION
DETECTOR
UNION
1/8T CRES
TEE
1/8T
KYNAR
VENT CAPILLARY
70 cc/min. 4 psig
PN 634398
TEE 1/8T KYNAR
C
METERING VALVE
.125 SWL CRES
EXHAUST
UNION
1/4T - 1/8T CRES
PUMP CAPILLARY
1/8 O.D. TEFLON
PN 657473
REDUCER
TUBING, 1/4O.D. TEFLON
ELBOW, MALE
1/4T – 1/4NPT
CRES
BACK
PRESSURE
TEE ASSEMBLY
1/8T – 1/4NPT
SAMPLE
SAMPLE CAPILLARY
70 cc/min. 5 psig PN 659657
TUBING 1 / 4 O.D. TEFLON
REDUCER
1/4T - 1/8T
CRES
1/8T - 1/16T CRES
SENSOR, SAMPLE
PRESSURE 15 PSIG
CROSS 1/8T
CRES
TEE
1/4T
REGULATOR
REDUCER, BULKHEAD
1/4T – 1/8T CRES
MPLE
200 cc/min. 2 psig PN 659658
CONN, MALE
CRES
BULKHEAD
1/4T CRES
1/8T-3/8NPT CRES
UST
TEE 1/8T
KYNAR
TUBING, 1/4O.D. TEFLON TUBING, 1/4O.D. TEFLON
Figure 1-1. CLD Flow Diagram – Matrix Position 12, Value “G”
Teflon® is a registered trademark of E.I. duPont de Nemours and Co., Inc. Kynar® is a registered trademark of Pennwalt, Inc.
2 of 2 Addendum - Matrix Position 12, Value "G"
Rosemount Analytical Inc. A Division of Emerson Process Management
Page 89

WARRANTY

Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and ma­terial under normal use and service for a period of ninety (90) days from date of shipment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the com­mercial unit shall not operate to condemn such commercial unit when such goods, part(s) and consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WAR­RANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DE-
LAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WAR­RANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF AC­TION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDEN­TAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVE­NUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT. IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond
Seller's direct control.
Page 90
Instruction Manual
760000-C November 2002
Model NGA2000 CLD
EMERSON PROCESS MANAGEMENT
ROSEMOUNT ANALYTICAL INC. PROCESS ANALYTIC DIVISION
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com
ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd.
1 Pandan Crescent Singapore 128461 Republic of Singapore T 65-777-8211 F 65-777-0947
http://www.processanalytic.com
© Rosemount Analytical Inc. 2002
EUROPEAN TECHNOLOGY CENTER Fisher-Rosemount GmbH & Co.
Industriestrasse 1 63594 Hasselroth Germany T 49-6055-884 0 F 49-6055-884209
EUROPE, MIDDLE EAST, AFRICA Fisher-Rosemount Ltd.
Heath Place Bognor Regis West Sussex PO22 9SH England T 44-1243-863121 F 44-1243-845354
LATIN AMERICA Fisher - Rosemount
Av. das Americas 3333 sala 1004 Rio de Janeiro, RJ Brazil 22631-003 T 55-21-2431-1882
Loading...