Read this manual before working with the product. For personal and system safety,
and for optimum product performance, make sure you thoroughly understand the
contents before installing, using, or maintaining this product.
Within the United States, Emerson Process Management has two toll-free assistance
numbers:
Customer Central
Technical support, quoting, and order-related questions.
1-800-522-6277 (7:00 am to 7:00 pm CST)
North American Response Center
Equipment service needs.
1-800-654-7768 (24 hours—includes Canada)
Outside of the United States, contact your local Emerson Process Management
representative.
Title Page
March 2016
The products described in this document are NOT designed for nuclear-qualified
applications. Using non-nuclear qualified products in applications that require
nuclear-qualified hardware or products may cause inaccurate readings.
For information on Rosemount nuclear-qualified products, contact your local Emerson
Process Management Sales Representative.
Title Page
1
Title Page
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
March 2016
Reference Manual
00809-0100-4772, Rev FA
2
Title Page
Reference Manual
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
00809-0100-4772, Rev FA
Section 1Introduction
1.1Using this manual
The sections in this manual provide information on installing, configuring, troubleshooting,
and performing other procedures for the Rosemount
F
OUNDATION
The sections are organized as follows:
Section 2: Configuration and Operation describes operation, software functionality, and
configuration procedures.
Section 3: Installation provides assistance in hardware installation and wiring.
Section 5: Resource Block provides descriptions of resource block parameters, errors, and
diagnostics.
™
Fieldbus. Specifications and other important information are also included.
Introduction
March 2016
™
8800D Vortex Flowmeter with
Section 6: Analog Input Function Block provides descriptions of analog input function block
parameters, errors, and diagnostics.
Section 7: Troubleshooting provides troubleshooting information for the most common
problems that occur during operation. It describes the advanced troubleshooting features
provided by the flowmeter and provides the maintenance procedures required to
disassemble and re-assemble the hardware components.
Appendix A: Specifications and Reference Data provides reference and specification data for
the flowmeter and its applications.
Appendix B: Product Certifications provides the available product certfications and
hazardous area approvals.
Appendix C: Electronics Verification provides a procedure for verification of electronic
output to assist in meeting the quality standards for ISO 9001-certified manufacturing
processes.
1.2FOUNDATION Fieldbus technology
FOUNDATION Fieldbus is an all digital, serial, two-way communication system that
interconnects field equipment such as sensors, actuators, and controllers. Fieldbus is a Local
Area Network (LAN) with built-in capability to distribute control applications across the
network for instruments that are used in both process and manufacturing automation
Introduction
1
Introduction
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
March 2016
environments. The fieldbus environment is the base level group of digital networks in the
hierarchy of plant networks.
The fieldbus retains the desirable features of the 4–20 mA analog system, including a
standardized physical interface to the wire, bus-powered devices on a single pair of wires,
and intrinsic safety options. In addition, it enables the following capabilities:
Increased capabilities due to full digital communications
Reduced wiring and wire terminations due to multiple devices on one pair of wires
Increased selection of suppliers due to interoperability
Reduced loading on control room equipment with the distribution of some control
and input/output functions to field devices
Speed options for process control and manufacturing applications.
1.3System description
The Rosemount 8800D Vortex Flowmeter with FOUNDATION fieldbus consists of a meter
body and transmitter. It measures volumetric flow rate by detecting the vortices created by
a fluid passing by the shedder bar.
Reference Manual
00809-0100-4772, Rev FA
The meter body is installed in-line with process piping. A sensor is located at the end of the
shedder bar and creates an alternating sine wave signal due to the passing vortices. The
transmitter measures the frequency of the signal and converts it into a flowrate.
This manual is designed to assist in the installation and operation of the flowmeter.
This product is intended to be used as a flowmeter for liquid, gas, or steam
applications. Any use other than for which it was intended may result in serious injury or
death.
This section covers operation, software functionality, and configuration procedures for the
Rosemount
by block information. For detailed information about the function blocks used in the
flowmeter, refer to the Rosemount F
™
8800D Vortex Flowmeter with FOUNDATION™ Fieldbus. This section is organized
2.2Safety messages
Procedures and instructions in this section may require special precautions to ensure the
safety of the personnel performing the operations. Information that raises potential safety
issues is indicated by a warning symbol ( ). Refer to the following safety messages before
performing an operation preceded by this symbol.
Explosions can result in death or serious injury.
Do not remove the transmitter covers in explosive environments when the circuit
is live.
Transmitter covers must be fully engaged to meet explosion proof requirements.
Before connecting a configuration tool in an explosive atmosphere, make sure the
instruments in the loop are installed in accordance with intrinsically safe or
nonincendive field wiring practices.
OUNDATION Fieldbus Block Reference Manual.
Configuration and Operation
3
Configuration and Operation
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
March 2016
Electrical shock can result in death or serious injury.
Avoid contact with the leads and terminals. High voltage that may
be present on leads can cause electrical shock.
2.3User interfaces
The flowmeter has both DD and DTM™-based user interfaces. All device configuration and
maintenance tasks can be performed using either communication technology.
Device Descriptors (DDs) and DTM files for Rosemount products are available on the
Emerson Install Kits website
The DD capabilities supported will vary based on host supplier and host revision. Check with
the host supplier to determine and obtain the appropriate DD for your situation. The type of
DD your host supports may influence navigation between different functions.
Reference Manual
00809-0100-4772, Rev FA
.
2.4Commissioning
Commissioning the flowmeter consists of verifying the transmitter configuration and
operation. Typically this should be done before the flowmeter is installed in the application
environment. In most cases, the user-supplied variables are pre-configured in the
transmitter at the factory, so the meter is ready to use upon arrival at the end user
installation.
Configuration may be required if:
Any configuration variables have changed.
A replacement transmitter is being installed.
To commission, connect power to the device and connect a Field Communicator or
Foundation Fieldbus host system to the signal loop.
1.Verify the device powers up and communicates with the Field Communicator or
host system.
2.Verify transmitter configuration parameters. Following are typical parameters to
check:
Ta g
Tra nsmit ter Mode
Process Fluid
4
Reference K-factor
Flange Type
Mating Pipe ID
PV Units (configured in the AI block)
Configuration and Operation
Reference Manual
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
00809-0100-4772, Rev FA
Flow Damping
Fixed Process Temperature
Fixed Process Density and Density Units
Density Ratio (for Standard or Normal flow units only)
3.If transmitter functional verification is required, refer to Appendix C: Electronics
The flowchart in Figure 2-1 is a general guide for commissioning and meter installation. It
can serve as a checklist to be referenced before and during installation of the flowmeter.
Configuration and Operation
March 2016
Verif icati on.
Configuration and Operation
5
Configuration and Operation
Is
Configuration
OK?
Mount
Flowmeter
Wire
Flowmeter
Power
Flowmeter
DONE
Mount
Conduit
START HERE
FIELD
INSTALL
CONFIGURE
Ta g
Tra n sm it te r
Mode
No
Bench
Commissioning
?
Review
Configuration
Yes
No
Yes
Did you
Configure on
Bench?
No
Yes
Configure if
Necessary
Go to
Review
Configuration
A
A
A
Go to
B
Go to
B
Using
LCD Display?
Yes
Configure
Local
Display
No
Meter
Installed?
Yes
No
DONE
Process Fluid
Reference
K-Factor
Flange Type
Mating Pipe ID
PV Units
(configured in
the AI block)
Flow Damping
Process
Tem pe ra tu re
Dampin g
Fixed Process
Tem pe ra tu re
Auto Adjust Filter
Density Ratio
(for Standard or
Normal flow
units only)
Process Density
and Density
Units (for mass
flow units only)
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
March 2016
Figure 2-1. General Guide for Commissioning and Meter Installation
Reference Manual
00809-0100-4772, Rev FA
6
Configuration and Operation
Reference Manual
Resource Block
Tra ns du cer
Block
Analog Input
(AI Block)
Other
function
blocks
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
00809-0100-4772, Rev FA
2.5General block information
Reference information on the process control function blocks can be found in the Function
Block manual document number 00809-0100-4783.
2.5.1Modes of operation
The Resource, Transducer, and all other function blocks in the device have modes of
operation. These modes govern the operation of the block. Every block supports both
automatic (AUTO) and out of service (OOS) modes. Other modes may also be supported.
For the procedures described in this manual, it will be helpful to understand the following
modes:
AUTO
The functions performed by the block will execute. If the block has any outputs, these will
continue to update. This is typically the normal operating mode.
Out of Service (OOS)
Configuration and Operation
March 2016
The functions performed by the block will not execute. If the block has any outputs, these
will typically not update and the status of any values passed to downstream blocks will be
“BAD”. To make changes to the configuration of the block, change the mode of the block to
OOS. When the changes are complete, change the mode back to AUTO.
MAN
In this mode, variables that are passed out of the block can be manually set for testing or
override purposes.
Other types of modes
Other types of modes are Cas, RCas, ROut, IMan and LO. Some of these may be supported
by different function blocks in the flowmeter. For more information, see the Function Block
manual, document 00809-0100-4783.
Note
When an upstream block is set to OOS, the output status of all downstream blocks will be
affected. The figure below depicts the hierarchy of blocks.
Configuration and Operation
Changing modes
To change the operating mode, set the MODE_BLK.TARGET to the desired mode. After a
short delay, the parameter MODE_BLOCK.ACTUAL should reflect the mode change if the
block is operating properly.
7
Configuration and Operation
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
March 2016
Permitted modes
It is possible to prevent unauthorized changes to the operating mode of a block. To do this,
configure MODE_BLOCK.PERMITTED to allow only the desired operating modes. It is
recommended to always select OOS as one of the permitted modes.
2.5.2Block instantiation
The Rosemount 8800D Device Revision 10 supports block instantiation. Previous device
revisions do not support block instantiation. When a device supports block instantiation,
the number of blocks and block types can be defined to match specific application needs.
The number of blocks that can be instantiated is only limited by the amount of memory
within the device and the block types that are supported by the device. Instantiation does
not apply to standard device blocks like the Resource, Sensor Transducer, Analog Input, and
PID Blocks.
By reading the parameter “FREE_SPACE” in the Resource Block you can determine how
many blocks you can instantiate. Each block that you instantiate takes up 4.5573% of the
“FREE_SPACE”.
Block instantiation is done by the host control system or configuration tool, but not all hosts
are required to implement this functionality. Refer to your specific host or configuration
tool manual for more information.
Reference Manual
00809-0100-4772, Rev FA
2.6Resource Block
The Resource Block contains diagnostic, hardware and electronics information. There are
no linkable inputs or outputs to the Resource Block.
2.6.1FEATURES and FEATURES_SEL
The FEATURES parameter is read only and defines which features are supported by the
flowmeter. Below is a list of the FEATURES the flowmeter supports.
FEATURES_SEL is used to turn on any of the supported features that are found in the
FEATURES parameter. The default setting of the flowmeter does not select any of these
features. Choose one or more of the supported features if any.
UNICODE
All configurable string variables in the flow meter, except tag names, are octet strings.
Either ASCII or Unicode may be used. If the configuration device is generating Unicode octet
strings, you must set the Unicode option bit.
REPORTS
The flow meter supports alert reports. The Reports option bit must be set in the features bit
string to use this feature. If it is not set, the host must poll for alerts. If this bit is set, the
transmitter will actively report alerts.
8
Configuration and Operation
Reference Manual
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
00809-0100-4772, Rev FA
SOFT W LOCK and HARD W LOCK
Inputs to the security and write lock functions include the hardware security switch, the
hardware and software write lock bits of the FEATURE_SEL parameter, the WRITE_LOCK
parameter, and the DEFINE_WRITE_LOCK parameter.
The WRITE_LOCK parameter prevents modification of parameters within the device except
to clear the WRITE_LOCK parameter. During this time, the block will function normally
updating inputs and outputs and executing algorithms. When the WRITE_LOCK condition is
cleared, a WRITE_ALM alert is generated with a priority that corresponds to the WRITE_PRI
parameter.
The FEATURE_SEL parameter enables the user to select a hardware or software write lock or
no write lock capability. To enable the hardware security function, enable the HW_SEL bit in
the FEATURE_SEL parameter. When this bit has been enabled the WRITE_LOCK parameter
becomes read only and will reflect the state of the hardware switch. In order to enable the
software write lock, the SW_SEL bit must be set in the FEATURE_SEL parameter. Once this
bit is set, the WRITE_LOCK parameter may be set to “Locked” or “Not Locked.” Once the
WRITE_LOCK parameter is set to “Locked” by either the software or the hardware lock, all
user requested writes as determined by the DEFINE_WRITE_LOCK parameter shall be
rejected.
Configuration and Operation
March 2016
The DEFINE_WRITE_LOCK parameter allows the user to configure whether the write lock
functions (both software and hardware) will control writing to all blocks, or only to the
resource and transducer blocks. Internally updated data such as process variables and
diagnostics will not be restricted by the security switch.
Table 2 -1displays all possible configurations of the WRITE_LOCK parameter.
Table 2-1. Configurations of the WRITE_LOCK Parameter
1. The hardware and software write lock select bits are mutually exclusive and the hardware select has the highest priority. When the HW_SEL bit if set
to 1 (on), the SW_SEL bit is automatically set to 0 (off) and is read only.
FEATURE_SEL
SW_SEL bit
(1)
SECURITY
SWITCH
0
(unlocked)
WRITE_LOCK
WRITE_LOCK
1 (unlocked)Read onlyNAAll
Read/Write
DEFINE_WRITE_LOCK
access to
blocks
Blocks
Blocks
only
only
2.6.2MAX_NOTIFY
Configuration and Operation
The MAX_NOTIFY parameter value is the maximum number of alert reports that the
resource can have sent without getting a confirmation, corresponding to the amount of
buffer space available for alert messages. The number can be set lower, to control alert
9
Configuration and Operation
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
March 2016
flooding, by adjusting the LIM_NOTIFY parameter value. If LIM_NOTIFY is set to zero, then
no alerts are reported.
2.6.3PlantWeb™ alarms
The alarms and recommended actions should be used in conjunction with Tabl e 7-2 .
The Resource Block will act as a coordinator for PlantWeb alarms. There will be three alarm
parameters (FAILED_ALARM, MAINT_ALARM, and ADVISE_ALARM) which will contain
information regarding some of the device errors which are detected by the transmitter
software. There will be a RECOMMENDED_ACTION parameter which will be used to display
the recommended action text for the highest priority alarm. FAILED_ALARM will have the
highest priority followed by MAINT_ALARM and ADVISE_ALARM will be the lowest priority.
FAILED_A LARMS
A failure alarm indicates a failure within a device that will make the device or some part of
the device non-operational. This implies that the device is in need of repair and must be
fixed immediately. There are four parameters associated with FAILED_ALARMS specifically,
they are described below.
Reference Manual
00809-0100-4772, Rev FA
FAI LED_EN ABLED
This parameter contains a list of failures in the device which makes the device
non-operational that will cause an alarm to be sent. Below is a list of the failures with the
highest priority first.
1.NV memory failure
2.Sensor board electronics failure
3.Output board electronics failure
4.Thermocouple failure (MTA only)
FAILED_MASK
This parameter will mask any of the failed conditions listed in FAILED_ENABLED. A bit on
means that the condition is masked out from alarming and will not be reported.
FAI LED_PR I
Designates the alarming priority of the FAILED_ALM, see “Alarm priority” on page 17. The
default is 0 and the recommended value are between 8 and 15.
FAILED_ACTIVE
This parameter displays which of the alarms is active. Only the alarm with the highest
priority will be displayed. This priority is not the same as the FAILED_PRI parameter
described above. This priority is hard coded within the device and is not user configurable.
10
FAILED_ALM
Alarm indicating a failure within a device which makes the device non-operational.
Configuration and Operation
Reference Manual
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
00809-0100-4772, Rev FA
MAINT_ALARMS
A maintenance alarm indicates the device or some part of the device needs maintenance
soon. If the condition is ignored, the device will eventually fail. There are five parameters
associated with MAINT_ALARMS, they are described below.
MAINT_ENABLED
The MAINT_ENABLED parameter contains a list of conditions indicating the device or some
part of the device needs maintenance soon. If the condition is ignored, the device will
eventually fail.
Below is a list of the conditions with the highest priority first.
1.Low pass filter over range
2.Low-flow cutoff over range
3.Density calc using fixed temp (MTA only)
4.Trigger over range
Configuration and Operation
March 2016
5.Electronics temp beyond limit (MTA only)
MAINT_MASK
The MAINT_MASK parameter will mask any of the failed conditions listed in
MAINT_ENABLED. A bit on means that the condition is masked out from alarming and will
not be reported.
MAINT_PRI
MAINT_PRI designates the alarming priority of the MAINT_ALM, “Process alarms” on
page 17. The default is 0 and the recommended values is 3 to 7.
MAINT_ACTIVE
The MAINT_ACTIVE parameter displays which of the alarms is active. Only the condition
with the highest priority will be displayed. This priority is not the same as the MAINT_PRI
parameter described above. This priority is hard coded within the device and is not user
configurable.
MAINT_ALM
An alarm indicating the device needs maintenance soon. If the condition is ignored, the
device will eventually fail.
Advisory alarms
Configuration and Operation
An advisory alarm indicates informative conditions that do not have a direct impact on the
device's primary functions There are five parameters associated with ADVISE_ALARMS.
ADVISE_ENABLED
The ADVISE_ENABLED parameter contains a list of informative conditions that do not have a
direct impact on the device's primary functions. Below is a list of the advisories with the
highest priority first.
11
Configuration and Operation
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
March 2016
1.Flow simulation mode
2.PWA simulation active
3.Low flow cutoff active
4.Flow signal injection
5.PV range exceeded
6.Sensor range exceeded
7.Process temp above USL (427 °C) (MTA only)
8.Process temp below LSL ( –50 °C) (MTA only)
9.Process temp above sat steam limit (MTA only)
10.Process temp below sat steam limit (MTA only)
ADVISE_MASK
Reference Manual
00809-0100-4772, Rev FA
The ADVISE_MASK parameter will mask any of the failed conditions listed in
ADVISE_ENABLED. A bit on means the condition is masked out from alarming and will not
be reported.
ADVISE_PRI
ADVISE_PRI designates the alarming priority of the ADVISE_ALM, see “Process alarms” on
page 17. The default is 0 and the recommended values are 1 or 2.
ADVISE_ACTIVE
The ADVISE_ACTIVE parameter displays which of the advisories is active. Only the advisory
with the highest priority will be displayed. This priority is not the same as the ADVISE_PRI
parameter described above. This priority is hard coded within the device and is not user
configurable.
ADVISE_ALM
ADVISE_ALM is an alarm indicating advisory alarms. These conditions do not have a direct
impact on the process or device integrity.
Recommended actions for PlantWeb alarms
The RECOMMENDED_ACTION parameter displays a text string that will give a
recommended course of action to take based on which type and which specific event of the
PlantWeb alarms is active Ta b le 5 - 1. Reference Index Number 78.
12
Configuration and Operation
Reference Manual
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
00809-0100-4772, Rev FA
2.7Transducer Block
The Transducer Block contains sensor and process fluid information used by the transmitter
for accurate measurements.
2.7.1Parameters
To make parameter changes, the Block Mode must be in the Out of Service (OOS) mode of
operation.
The Transducer Block parameters must be correct for an accurate measurement. These
parameters are factory configured specifically for the measurement when the transmitter is
sized using Rosemount Instrument Toolkit
2.7.2XMTR_MODE
Transmitter Mode
Without Temperature Sensor—No process temperature sensor available (Non MTA
electronics)
™
program.
Configuration and Operation
March 2016
With Temperature Sensor—Process temperature sensor installed (MTA electronics only)
2.7.3PROCESS_FLUID
Liquid—Process fluid is liquid
Gas/Steam—Process fluid is a gas
T Comp Sat Steam—Temperature Compensated Saturated Steam (MTA electronics only)
2.7.4Reference K_FACTOR
The factory calibrated K-Factor stamped on the meter body
2.7.5FIXED_PROC_TEMPERATURE
Fixed Process Temperature. The operating temperature of the process
2.7.6FIXED_PROC_DENSITY
Fixed Process Density. The density of the process fluid at flowing pressure and temperature
2.7.7FLANGE_TYPE
The flange rating and type to match the vortex meter body
2.7.8PIPE_INSIDE_DIAMETER
2.7.9DAMPING
Configuration and Operation
The inside diameter (I.D.) of the mating pipe where the meter body is installed.
The flow damping value. Default is 2 seconds. Valid range is 0.2 to 255 seconds.
13
Configuration and Operation
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
March 2016
2.7.10FILTER_AUTO_ADJUST
Optimize Signal Processing based on the density of the process fluid. Select the incremental
value equal to or less than the process fluid density.
2.7.11PRIMARY_VALUE_RANGE
Range of the flow measurement. The units match the XD_SCALE units from the AI Block
assigned to Flow
2.7.12SENSOR_RANGE
Range the meter is capable of measuring. The maximum measurable flow rate the vortex
meter will measure
The units match the XD_SCALE units from the AI Block assigned to Flow
2.7.13SECONDARY_VALUE
The shedding frequency measured by the vortex sensor in units of Hz
Reference Manual
00809-0100-4772, Rev FA
2.7.14INSTALLATION_EFFECTS
Installation Effect correction factor. Use when the meter is installed in less than ideal piping
configuration. See Installation Effects document 00816-0100-3250 for details.
Valid range -1.5 to 1.5
2.7.15PROCESS_DENSITY_RATIO
Density Ratio = Density at flowing P and T/Density at Standard P and T
The transmitter only uses density ratio when the flow output is in Standard or Normal units.
Examples of Standard unit: Standard Cubic Feet per Minute, SCFM
Example of Normal unit: Normal Cubic Meters per hour, NCMH
2.7.16METER_DISPLAY
Data displayed locally on LCD display.
The following measurements can be displayed on the flowmeter LCD display:
Flow
Transducer Percent Range—Percent range of maximum accurate flow
Process Temperature—MTA electronics only
Process Density—MTA electronics only
Shedding Frequency
14
Electronics Temperature—MTA electronics only
Integrator Block Out—Totalized flow from Vortex meter Integrator Block
Configuration and Operation
Reference Manual
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
00809-0100-4772, Rev FA
2.7.17PROC_TEMP_DAMP
Process Temperature Damping. Only vortex meters with the MTA option measure process
temperature.
The units are seconds, default is 2 seconds. Valid range is 0.4 to 32 seconds.
2.7.18VEL_MEAS_BASE
Velocity Measurement Base. The velocity calculation can be based on the Mating Pipe inside
diameter or the Meter body inside diameter. Default is the Mating pipe ID.
2.7.19TC_FAILURE_MODE
Thermocouple Failure Mode. The MTA optioned meters measure process temperature with
a thermocouple. If the thermocouple fails, the transmitter will indicate the failed sensor
input. There are two options available for the transmitter operation after the sensor fails.
Sensor Failure; Status is BAD—Mass Flow measurement status is BAD.
Use Fixed Temp; Status is UNCERTAIN—Mass Flow calculation uses fixed temperature values
when calculating density and status of measurement is UNCERTAIN .
Configuration and Operation
March 2016
2.7.20LFC_RESPONSE
Low Flow Cutoff Response. The flow measurement will be 0 when the measured flow is less
than the Low Flow Cutoff.
The transition from the flow measurement down to 0 flow can be Damped or Stepped.
Damped will transition to 0 flow based on the Damping setting.
Stepped will transition to 0 flow as soon as the Low Flow Cutoff activates.
2.7.21CALC_PROC_DENSITY
Calculated Process Density. The MTA optioned transmitters calculate the density of
Saturated Steam based on the process temperature measurement.
The Calculated Process Density is used to determine the Mass Flow rate of saturated steam.
2.7.22PROC_TEMP_RANGE
MTA option only. Process Temperature Range.
The process temperature range is the minimum and maximum temperature of the vortex
sensor.
The Units will match the XD_SCALE units of the AI Block assigned to Process Temperature
2.7.23ELEC_TEMP_RANGE
Configuration and Operation
MTA option only. Electronics Temperature Range.
The electronics temperature range is the minimum and maximum temperature rating for
the electronic components.
15
Configuration and Operation
PV_FTIME
63% of Change
OUT (mode in man)
OUT (mode in auto)
PV
Time (seconds)
FIELD_VAL
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
March 2016
The Units will match the XD_SCALE units of the AI Block assigned to Electronics
Tem p er at ur e.
2.8Analog Input (AI) Function Block
For information on the Analog Input (AI) Function Block, refer to Section 6.
2.8.1Filtering
The filtering feature changes the response time of the device to smooth variations in output
readings caused by rapid changes in input. Adjust the filter time constant (in seconds) using
the PV_FTIME parameter. Set the filter time constant to zero to disable the filter feature.
Figure 2-2. Analog Input PV_FTIME Filtering
Reference Manual
00809-0100-4772, Rev FA
2.8.2Low cutoff
16
When the converted input value is below the limit specified by the LOW_CUT parameter,
and the Low Cutoff I/O option (IO_OPTS) is enabled (True), a value of zero is used for the
converted value (PV). This option is useful to eliminate false readings when the flow
measurement is close to zero.
Note
Low Cutoff is the only I/O option supported by the AI block. Set the I/O option in Manual or
Out of Service mode only.
Configuration and Operation
Reference Manual
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
00809-0100-4772, Rev FA
2.8.3Process alarms
Process Alarm detection is based on the OUT value. Configure the alarm limits of the
following standard alarms:
High (HI_LIM)
High high (HI_HI_LIM)
Low (LO_LIM)
Low low (LO_LO_LIM)
In order to avoid alarm chattering when the variable is oscillating around the alarm limit, an
alarm hysteresis in percent of the PV span can be set using the ALARM_HYS parameter. The
priority of each alarm is set in the following parameters:
HI_PRI
HI_HI_PRI
LO_PRI
LO_LO_PRI
Configuration and Operation
March 2016
2.8.4Alarm priority
Alarms are grouped into five levels of priority:
Priority
number
0The alarm condition is not used.
1An alarm condition with a priority of 1 is recognized by the system, but is
2An alarm condition with a priority of 2 is reported to the operator.
3-7Alarm conditions of priority 3 to 7 are advisory alarms of increasing
8-15Alarm conditions of priority 8 to 15 are critical alarms of increasing
Priority description
not reported to the operator.
priority.
priority.
2.8.5Status options
Status options (STATUS_OPTS) supported by the AI block are shown below:
Propagate fault forward
If the status from the sensor is Bad, Device failure or Bad, Sensor failure, propagate it to OUT
without generating an alarm. The use of these sub-status in OUT is determined by this
option. Through this option, the user may determine whether alarming (sending of an alert)
will be done by the block or propagated downstream for alarming.
Configuration and Operation
Uncertain if limited
Set the output status of the Analog Input block to uncertain if the measured or calculated
value is limited.
BAD if limited
Set the output status to Bad if the sensor is violating a high or low limit.
17
Configuration and Operation
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
March 2016
Uncertain if man mode
Set the output status of the Analog Input block to uncertain if the actual mode of the block
is Man.
Note
The instrument must be in Out of Service mode to set the status option.
2.8.6Advanced features
The AI Function Block provides added capability through the addition of the following
parameters:
ALARM_TYPE
ALARM_TYPE allows one or more of the process alarm conditions detected by the AI
function block to be used in setting its OUT_D parameter.
OUT_D
Reference Manual
00809-0100-4772, Rev FA
OUT_D is the discrete output of the AI function block based on the detection of process
alarm condition(s). This parameter may be linked to other function blocks that require a
discrete input based on the detected alarm condition.
2.9Flow simulation
The electronics is capable of internally generating a flow signal that may be used to simulate
a sensor signal. The simulated signal amplitude is based on the transmitter required
minimum process density. The simulated signal can be a constant frequency or the
simulated signal can be a varying frequency to represent a ramping flow rate.
Simulating the flow signal requires the simulate ENABLE jumper on the transmitter
electronics board to be in the ‘ON’ position. The transmitters are shipped with the jumper
in the ‘OFF’ position.
2.9.1Using transducer block parameters
SIMULATION_CONTROL
Sim Disabled—Normal operation, no simulated flow signal.
Sim—Internal Generator—The internal frequency generator will produce the frequency
signal.
18
Sim—Ex ternal Generator—An Ex ternal Frequency generator can be connected to Freq In and
Ground connections on the electronics board.
SIMULATION_UNITS
PV Engineering Units—Simulated flow will be in engineering units.
PV Percent of Range—Simulated flow will be in percentage of Primary Value flow range.
Configuration and Operation
Reference Manual
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
00809-0100-4772, Rev FA
SIMULATION_HIGH_POINT
High simulation value in engineering units or percentage.
SIMULATION _LOW_POINT
Low simulation value in engineering units or percentage.
SIMULATION_RAMP_PERIOD
The time, in seconds, between low and high simulation points.
2.9.2Configuration software
Configuration Software, such as AMS™ Wireless Configurator, simplifies the simulation
process.
AMS Revision 12 with Rosemount 8800D Device Revision 9
Set Target Mode to Out of Service by unchecking Auto and checking Out of Service boxes.
Configuration and Operation
March 2016
Set the simulation values as desired and return the mode to Auto.
The flow value will be simulated until the simulation is disabled. Simulation is also disabled
with a power cycle.
Device Revision 10
The Device Rev 10 interface steps through the simulation configuration after clicking the
‘Simulate Flow’ button. Follow the steps for the desired flow simulation.
2.10Device capabilities
2.10.1Link active scheduler
The flowmeter can be designated to act as the Backup Link Active Scheduler (BLAS) in the
event that the LAS is disconnected from the segm ent. As the backup LAS, the flowmeter will
take over the management of communications until the host is restored.
The host system may provide a configuration tool specifically designed to designate a
particular device as a backup LAS. Otherwise, this can be configured manually as follows:
1.Access the Management Information Bose (MIB) for the flowmeter.
2.To activate the LAS capability, write 0x02 to the
BOOT_OPERAT_FUNCTIONAL_CLASS object (Index 605). To deactivate, write
0x01.
Configuration and Operation
3.Restart the processor.
19
Configuration and Operation
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182
March 2016
2.10.2Capabilities
Virtual Communication Relationship (VCRs)
There are a total of 20 VCRs. One is permanent and 19 are fully configurable by the host
system. Twenty-five link objects are available.
Table 2-2. Network Parameters and Values
Network parameterValue
Slot Time6
Maximum Response Delay4
Maximum Inactivity to Claim LAS Delay47
Minimum Inter DLPDU Delay7
Time Sync class4 (1ms)
Maximum Scheduling Overhead21
Per CLPDU PhL Overhead4
Maximum Inter-channel Signal Skew0
Required Number of Post-transmission-gab-ext Units0
Required Number of Preamble-extension Units1
Reference Manual
00809-0100-4772, Rev FA
Host timer recommendations
T1 = 96000
T2 = 1920000
T3 = 480000
Block execution times
Analog Input = 15 ms
PID = 20 ms
Arithmetic = 20 ms
Integrator = 25 ms
20
Configuration and Operation
Loading...
+ 162 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.