Emerson Rosemount 6888C Quick Start Manual

Page 1
Quick Start Guide
00825-0100-4891, Rev AB
October 2019
Rosemount™ 6888C In-Situ Combustion Oxygen Analyzer
For use in Hazardous Areas
Page 2
Quick Start Guide October 2019
This guide provides basic guidelines for Rosemount 6888 In-Situ Oxygen Analyzers. It does not provide instructions for configuration, diagnostics, maintenance, service, troubleshooting, or flame-proof installations. Refer to the Rosemount 6888C Manual for more instruction.
Essential instructions
Read this page before proceeding!
Emerson™ designs, manufactures, and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Emerson products. Failure to follow the proper instructions may cause any one of the following situations to occur: loss of life, personal injury, property damage, damage to this instrument, and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Emerson representative for
clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the
product.
Install equipment as specified in the installation instructions of the appropriate instruction manual
and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and
maintain the product.
Use only factory documented components for repair. Tampering and unauthorized substitution
and parts can affect product performance and cause unsafe operation of your process.
The Field Communicator must be upgraded to System Software 2.0 with graphic license for operation with the Rosemount™ 6888C O2 Transmitter. The AMS software must be upgraded to AMS 8.0 or above. Contact Emerson's Global Service Center (GSC) at +1-800-833-8314 to upgrade the Field Communicator software to System Software 2.0 with graphic license.
Symbols
Earth (ground) terminal
Protective conductor terminal
Risk of electrical shock
Refer to reference manual.
2 Emerson.com/Rosemount
Page 3
October 2019 Quick Start Guide
The following definitions apply to Warnings, Cautions, and Notices found throughout this publication.
WARNING
Highlights an operation or maintenance procedure, practice, condition, statement, etc., which if not strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION
Highlights an operation or maintenance procedure, practice, condition, statement, etc., which if not strictly observed, could result in damage to or destruction of equipment or loss of effectiveness.
Highlights an essential operating procedure, condition, or statement.
WARNING
EXPLOSION Do not open when an explosive atmosphere may be present.
WARNING
ELECTRIC SHOCK Do not open while energized.
CAUTION
EQUIPMENT DAMAGE For the standard housing probe and direct replacement probe, only use supply cables & certified cable glands rated > 105 °C (221 °F).
CAUTION
EQUIPMENT DAMAGE For the autocal housing, only use supply cables and certified cable glands rated > 85 °C (185 °F).
Contents
Install........................................................................................................................................... 5
Configuration, startup, and operation........................................................................................ 26
Rosemount™ 6888 product certifications................................................................................... 31
Declaration of Conformity..........................................................................................................35
Quick Start Guide 3
Page 4
Quick Start Guide October 2019
China RoHS Table....................................................................................................................... 37
4 Emerson.com/Rosemount
Page 5
October 2019 Quick Start Guide

1 Install

WARNING
Failure to follow safety instructions could result in serious injury or death.
Before installing this equipment, read the Essential Instructions at the front of this document.
WARNING
Electrical shock
Failure to install covers and ground leads could result in serious injury or death.
Install all protective covers and ground leads after installation.

1.1 Mechanical installation

Most combustion processes run only slightly negative or positive in pressure, so that the probe flange is for mechanical mounting only. The probe is not rated for high pressures. If this is a new installation, a weld plate for welding to the flue gas duct can be supplied.
WARNING
Electrical shock
Failure to install covers and ground leads could result in serious injury or death.
Install all protective covers and ground leads after installation.

1.1.1 Probe installation

Complete the following steps to install the probe in the duct.
Prerequisites
Ensure all components are available to install the probe.
Procedure
1. If using the optional ceramic diffusion element, the vee-deflector must be correctly oriented. Before inserting the Rosemount 6888C probe, check the direction of gas flow in the duct. Orient the vee­deflector so the apex points upstream towards the flow.
Quick Start Guide 5
Page 6
Quick Start Guide October 2019
2. If using the standard square weld plate or an optional flange mounting plate, weld or bolt the plate onto the duct.
The through hole diameter in the stack or duct wall and refractory material must be at least 2½ in. (63.5 mm).
3. Insert probe through the opening in the mounting flange and bolt the unit to the flange.
Figure 1-1: Rosemount™ 6888C Probe with Standard Terminations/Electronics Housing
Note
All dimensions are in inches with millimeters in parentheses.
Figure 1-2: 6888C Probe with Integral Autocal Housing
Table 1-1: Removal/Installation
Probe length DIM A insertion
depth
18 (457) probe 16.10 (409) 15.77 (401) 19.26 (490)
36 (910) probe 32.52 (826) 46.6 (1182) 50.1 (1271)
6 Emerson.com/Rosemount
DIM B removal envelope standard housing
DIM removal envelope accessory housing
Page 7
October 2019 Quick Start Guide
Table 1-1: Removal/Installation (continued)
Probe length DIM A insertion
depth
6 ft (1830) probe 68.52 (1740) 82.6 (2097) 86.1 (2186)
DIM B removal envelope standard housing
DIM removal envelope accessory housing
Figure 1-3: Probe Installation
Table 1-2: Mounting Flange
ANSI DIN
Flange diameter 8.25 (210)
Hold diameter .75 (20)
(4) holes eq sp on BC 6.00 (152) 6.70 (170)
Quick Start Guide 7
Page 8
Quick Start Guide October 2019
Table 1-3: Installation weld plate outline
ANSI DIN
A 7.75 (197) 8.5 (215)
B thread .625 (11) (M-16x2)
C dia 6.00 (152) 6.70 (170)
Figure 1-4: Drip Loop and Insulation Removal
A. Line voltage B. Logic I/O, 4-20 mA signal C. Stack or duct metal wall
D. Adapter plate
E. Insulation F. Note
Replace insulation after installing Rosemount 6888C.
Note
Standard housing probe shown. Accessory housing is similar. Probe may be vertical or horizontal.
8 Emerson.com/Rosemount
Page 9
October 2019 Quick Start Guide

1.2 Electrical installation

All wiring must conform to local and national codes. Multiple wiring diagrams are shown in this section. Always refer to the diagrams that apply to your transmitter configuration and disregard all other wiring diagrams.
WARNING
Electrical shock
Failure to install covers and ground leads could result in serious injury or death.
Disconnect and lock out power before connecting the power supply. Install all protective covers and ground leads after installation. To meet the safety requirements of IEC 61010-1 (EC requirement) and
ensure safe operation of this equipment, connect the main electrical power supply through a circuit breaker (min 10 A) which will disconnect all current-carrying conductors during a fault situation.
This circuit breaker should also include a mechanically operated isolating switch. If it does not, keep another external means of disconnecting the power supply from the equipment located close by. Circuit breaks or switches must complay with a recognized standard such as IEC 947.
To maintain proper earth grounding, ensure a positive connection exists between the transmitter housing and earth. The connecting ground wire must be 14 AWG minimum.
Line voltage, signal, and relay wiring should be rated for at least 221 °F (105 °C).

1.2.1 Wire Rosemount™ 6888 Transmitter Probe only (no Rosemount 6888 Xi Electronics)

The Rosemount 6888 Transmitter Probe has built-in electronics that control the heater temperature and amplify the raw O2 millivolt signal to a linear 4-20 mA.
You can run the 4-20 mA signal lines can be run directly to the control room and also power the transmitter electronics. There is no O2 display or keypad on the probe, so you must set up the transmitter through HART communications via a Field Communicator or Asset Management Solutions (AMS).
Procedure
1. Remove the cover from the probe.
2. Connect the line (L1 wire) to the L1 terminal, the neutral (L2 wire) to the L2/N terminal, and the ground wire to the ground lug.
Quick Start Guide 9
®
Page 10
Quick Start Guide October 2019
The Rosemount 6888C accepts line voltage at 120/240 Vac ±10%, 50/60 Hz. No setup is required.
3. Connect the 4-20 mA signal wires at the transmitter. Use a shielded twisted wire pair. Do not allow bare shield wires to contact the circuit boards. Insulate the shield wires prior to termination.
The transmitter electronics are loop-powered (i.e., the 4-20 mA signal wires supply 24 Vdc from the DCS or an external power supply).
Figure 1-5: Rosemount 6888C Standard Probe Housing
4. Terminate the shield only at the transmitter electronics housing unless using a Rosemount 6888 Xi. When using the Rosemount 6888 Xi Advanced Electronics, terminate the shield at both ends.
5. Reinstall cover on transmitter.

1.2.2 Wire standard housing transmitter probe with Rosemount™ 6888 Xi Electronics

The Rosemount 6888 Xi Electronics serves as an operater interface unit with a back-lit display and keypad. It is capable of two channels, serving up to two Rosemount 6888 probes.
Procedure
1. Remove cover screws from the front cover of the Rosemount 6888 Xi. Swing down the front cover of the interface box.
10 Emerson.com/Rosemount
Page 11
October 2019 Quick Start Guide
2. Pull out the I/O board on the right side of the card rack inside the Rosemount 6888 Xi.
If your system is configured to operate two transmitter probes, there are two I/O interface boards.
3. Connect the 4-20 mA signal wires at J4 of the I/O board. Attach the supplied ferrite clamp over the 4-20 mA OUT wires that extend past the shield.
Note
Installation of the ferrite clamp over the 4-20 mA OUT wires is required for compliance with the European EMC directive.
4. Terminate the shield of the 4-20 mA signal wires at the designated ground terminal of the Rosemount 6888 Xi. Do not allow bare shield wires to contact the circuit boards. Insulate the shield wires prior to termination.
5. Connect the signal wires from the SPS or IMPS (if used) to the applicable terminals of J3.
Refer to the SPS or IMPS reference manual for wiring details.
6. Reinstall the I/O board in the card rack of the Rosemount 6888 Xi.
7. If your system is configured for two channel operation, repeat Step 2 through Step 6 to connect the other probe's signal wires.
8. Remove the probe's connector from the power supply board located on the left side of the card rack inside the Rosemount 6888 Xi.
9. Connect the line, or L1, wire to the L1 terminal and the neutral, or L2, wire to the N terminal.
10. Reinstall the power supply connector to the power supply board.
Quick Start Guide 11
Page 12
Quick Start Guide October 2019
Figure 1-6: Wiring Diagrams - Single/Dual Channel Wiring Diagram
Note
A. Except for JP5, JP7, and JP8 on IO board, jumper and switch settings are
factory set and are shown for reference only.
B. IO board 4-20 mA loop power settings.
JP5: Pins 1-2 internal power Rosemount 6888 Xi to Rosemount 6888
Transmitter. Pins 2-3 external power Rosemount 6888 Xi to Rosemount 6888
Transmitter (requires 250 Ω resistor across J4, PR+ to PR-).
JP7/JP8: Pins 1-2 internal power Rosemount 6888 Xi to DCS.
Pins 2-3 external power Rosemount 6888 Xi to DCS.
12 Emerson.com/Rosemount
Page 13
October 2019 Quick Start Guide
Figure 1-7: Wiring Diagrams - Single/Dual Channel Wiring Diagram
A. Power supply board B. Channel #2 IO board C. Shield ground
D. Channel #1 IO board
E. AC input to P/S
F. Plug G. Channel #2 alarm relay SPS/IMPS H. Channel #2 4-20 mA / HART output
I. Channel #1 alarm relay SPS/IMPS J. Channel #1 4-20 mA/HART output
1.2.3 Connect the transmitter probe with integral autocal to HART
®
communications
This probe contains gas-switching solenoids so that the Rosemount™ 6888 Xi Electronics can control the introduction of calibration gases. Calibrations can be initiated via a calibration recommended diagnostic, time since last
Quick Start Guide 13
Page 14
Quick Start Guide October 2019
calibration, manually via external dry contact, HART communications, or from the Rosemount 6888 Xi local operator interface keypad. The integral autocal feature can only be implemented when the probe is used with a Rosemount 6888 Xi.
Procedure
1. Remove the two covers from the transmitter.
2. Connect the line (L1 wire) to the L1 terminal, the neutral (L2 wire) to the L2/N terminal, and the ground wire to the ground lug.
The Rosemount 6888C accepts line voltage at 120/240 Vac ±10%, 50/60 Hz. No setup is required.
3. Connect the 4-20 mA signal wires from the Rosemount 6888 Xi to the connections in the side chamber of the transmitter.
Do not connect the signal wires to the terminals in the main chamber where the AC input wires are connected. Use a shielded twisted wire pair. Do not allow bare shield wires to contact the circuit boards. Insulate the shield wires prior to termination. The 24 Vdc loop power is sourced from the Rosemount 6888 Xi.
4. Terminate the shield at both the probe and the Rosemount 6888 Xi Advanced Electronics.
5. Reinstall both covers on transmitter.
14 Emerson.com/Rosemount
Page 15
October 2019 Quick Start Guide
Figure 1-8: Integral autocal and HART communications
A. Ferrite clamp B. Signal C. Test points
D. #8 pan htd scr (internal ground)
E. Power F. Test point group
Note
A. Except for JP5, JP7, and JP8 on IO board, jumper and switch settings are factory set and are
shown for reference only.
B. IO board: 4-20 mA/HART loop power settings
JP5
Quick Start Guide 15
Page 16
Quick Start Guide October 2019
Pins 1-2: internal power Rosemount 6888 Xi to Rosemount 6888 transmitter
Pins 2-5: external power Rosemount 6888 Xi to Rosemount 6888 transmitter
(requires 2,500 resistor across J4, PR+ to PR-)
JP7/JP8Pins 1-2: internal power to DCS
Pins 2-3: external power Rosemount 6888 Xi to DCS
Note
I/O board - Channel 2 is a duplicate of Channel 1.
1.2.4 Connect the transmitter probe with integral autocal to FOUNDATION Fieldbus communications
This probe contains gas-switching solenoids so that the Rosemount™ 6888 Xi Electronics, Field Communicator, or AMS can control the introduction of calibration gases. Calibrations can be initiated via a calibration recommended diagnostic, time since last calibration, manually via external dry contact, or from the Rosemount 6888 Xi local operator interface keypad. The integral autocal feature can only be implemented when the probe is used with a Rosemount 6888 Xi.
Procedure
1. Remove the two covers from the transmitter.
2. Connect the line (L1 wire) to the L1 terminal, the neutral (L2) wire to the L2/N terminal, and the ground wire to the ground lug.
The Rosemount 6888C accepts line voltage at 120/240 Vac ±10%, 50/60 Hz. No setup is required.
3. Connect the FOUNDATION Fieldbus wires from the Rosemount 6888 side housing to the FF segment.
Note that the Rosemount 6888 probe is not rated as intrinsically safe and will render any IS or FISCO segment it is wired to as non-IS. Use a shielded twisted wire pair. Do not allow bare shield wires to contact the circuit boards.
4. Terminate the shield at both the probe and the Rosemount 6888 Xi Advanced Electronics.
Note
The FOUNDATION Fieldbus signal represents the O2 value and also powers the probe-mounted electronics.
5. Reinstall both covers on transmitter.
16 Emerson.com/Rosemount
Page 17
October 2019 Quick Start Guide
Figure 1-9: Integral Autocal and FOUNDATION Fieldbus Communications without Optional Rosemount 6888 Xi
A. Signal B. Not used C. #8 pan htr scr (internal ground)
D. Power
E. Probe test point group
Quick Start Guide 17
Page 18
Quick Start Guide October 2019
Figure 1-10: Integral autocal and FOUNDATION Fieldbus Communications and Optional Rosemount 6888Xi
A. Ribbon cable to display board J2 Sensor 1 B. Signal C. HART® connection (Used as a communication bus from probe
transmitter electronics to optional Rosemount 6888 Xi. Not accessible to Field Communicator or AMS)
D. #8 pan htr scr (internal ground)
E. Power F. Probe test point group
IO board switch/jumpers
Jumper settings
JP1: Pins 2-3
18 Emerson.com/Rosemount
Page 19
October 2019 Quick Start Guide
JP2: Pins 2-3
JP5Pins 1-2 internal power
Pins 2-3 external power
JP7Pins 1-2 internal power
Pins 2-3 external power
JP8Pins 1-2 internal power
Pins 2-3 external power
SW4 switch settings
Position 1: Off
Postion 2: Off
Position 3: Off
Position 4: Off
A. Except for JP5, JP7, and JP8 on IO board, jumper and switch settings are
factory set and are shown for reference only.
B. IO board 4-20 mA/HART loop power settings
JP5
Pins 1-2 internal power Rosemount 6888 Xi to Rosemount 6888
Transmitter.
Pins 2-3 external power Rosemount 6888 Xi to Rosemount 6888
Transmitter (requires 2,500 resistor across J4, PR+ to PR-).
JP7/JP8
Pins 1-2 internal power Rosemount 6888Xi to DCS.
Pins 2-3 external power Rosemount 6888Xi to DCS.
Quick Start Guide 19
Page 20
Quick Start Guide October 2019
Figure 1-11: Wiring Diagrams - Integral Autocal and FOUNDATION Fieldbus, Communications with Rosemount 6888Xi
A. Power supply board B. Channel #2 IO board C. Shield ground
D. Channel #1 IO board
E. AC input to P/S
F. Plug G. Channel #2 alarm relay, SPS/IMPS H. Channel #2 4-20 mA/HART output
I. Channel #1 alarm relay, SPS/IMPS J. Channel #1 4-20 mA/HART output

1.2.5 Connect traditional architecture system to the direct replacement probe

Use a traditional architecture configuration to provide for remote location of the transmitter electronics. All electronics are housed inside the
20 Emerson.com/Rosemount
Page 21
October 2019 Quick Start Guide
Rosemount™ 6888 Xi. A multi-conductor power/signal cable connects the probe to the Rosemount 6888 Xi. Use the following procedure to connect the traditional architecture probe to the Rosemount 6888 Xi.
Note
The traditional architecture cable is provided at the specified length and is ready for installation. The cable glands must be properly terminated to maintain EMC/EMI noise protection.
Procedure
1. Run the 7-conductor cable between the traditional architecture probe and the installation site for the Rosemount 6888 Xi.
Use new cable conduit or trough as needed.
2. Install the cable and lead wires to the probe per manufacturer's instructions.
3. Install the cable at the probe housing and at the Rosemount 6888 Xi enclosure.
a) Unscrew locking nut from gland assembly and slide locking
nut back along cable.
b) Pull the gland body away from the plastic insert.
Use care not to damage the cable shield braid.
c) Insert the cable wires into the proper entry port in either the
probe housing or the Rosemount 6888 Xi enclosure.
d) At the probe housing, apply Teflon tape or similar sealing
compound to the tapered pipe threads. Thread the gland body into the probe housing until properly seated.
e) At the Rosemount 6888 Xi enclosure, insert the gland body
into the left front cable port from the inside of the enclosure. Use the rubber O-ring provided to seal the cable port.
f) Ensure the cable shield braid is evenly formed over the gray
insert. When properly formed, the braid should be evenly spaced
around the circumference of the insert and not extend beyond the narrow diameter portion.
g) Carefully press the gray insert into the gland body.
The grooves on the insert should align with similar grooves inside the gland body. Press the insert in until it bottoms out in the gland body.
h) Slide the locking nut up and thread it onto the gland body.
Tighten the locking nut so the rubber grommet inside the
Quick Start Guide 21
Page 22
Quick Start Guide October 2019
plastic insert compresses against the cable wall to provide an environmental seal.
4. At the Rosemount 6888 Xi, connect the cable leads to the connectors on the transmitter I/O board.
Figure 1-12: Wiring Diagrams - Traditional Architecture with Direct Replacement Probe (no Electronics Inside)
Note
A. See the Rosemount 6888 Xi Reference Manual for additional installation
and operating instructions.
B. All wiring marked with an asterisk (*) is factory wiring inside the
Rosemount 6888 Xi.
C. Except for JP7 and JP8 on IO board, jumper and switch settings are
factory set and are shown for reference only.
22 Emerson.com/Rosemount
Page 23
October 2019 Quick Start Guide
Figure 1-13: Wiring Diagrams - Traditional Architecture with Direct Replacement Probe (no Electronics Inside)
A. Power supply board B. DR board C. Shield ground
D. IO board
E. Plug
F. Probe cable G. AC input H. Alarm relay, SPS/IMPS
I. 4-20 mA/HART® output
Quick Start Guide 23
Page 24
Quick Start Guide October 2019

1.3 Pneumatic installation

1.3.1 Calibration gas

The transmitter uses two calibration gas concentrations, low gas - 0.4% O2, balance N2, and high gas - 8% O2, balance N2.
CAUTION
Calibration reading errors
Failure to use proper gases will result in erroneous readings.
Do not use 100% nitrogen as a low gas (zero gas). Emerson suggests that gas for the low (zero) be between 0.4% and 2.0% O2.
Do not use gases with hydrocarbon concentrations of more than 40 parts per million.
CAUTION
Equipment damage
If the ducts will be washed down during outage, verify that 6888C units have been powered down and removed from the wash areas.
CAUTION
Equipment damage
Damage can result from having a cold unit exposed to the process gases.
Upon completing installation, ensure that the transmitter is turned on and operating prior to firing up the combustion process.
During outages, if possible, leave all units running to prevent condensation and premature aging from thermal cycling.
See Figure 1-14 for the calibration gas connections.
24 Emerson.com/Rosemount
Page 25
October 2019 Quick Start Guide
Figure 1-14: Rosemount 6888C Calibration Gas Connections
A. Calibration gas in B. Reference air vent C. Reference air in
Quick Start Guide 25
Page 26
Quick Start Guide October 2019

2 Configuration, startup, and operation

WARNING
Electrical shock
Failure to install covers and ground leads could result in serious injury or death.
Install all protective covers and ground leads after installation.
CAUTION
Equipment damage
If external loop power is used, the power supply must be a safety extra low voltage (SELV) type.

2.1 Power up transmitter without Rosemount™ 6888 Xi

Procedure
1. Apply AC line power to the transmitter.
2. Apply 24 Vdc loop power to the transmitter.
3. Using either the DCS control or a Field Communicator, verify communications to the transmitter. The transmitter probe takes approximately 45 minutes to warm up to the 1357 °F (736 °C) heater setpoint. The 4-20 mA signal remains at a default value of 3.5 mA, and the O2 reading remains at 0% through this warm-up period. After warm-up, the probe begins reading oxygen, and the 4-20 mA output is based on the default range of 0-10% O2.
If there is an error condition at startup, an alarm message is displayed.
2.2

Power up transmitter with single/dual channel or single channel

Complete the following steps to power up the transmitter using the Rosemount™ 6888 Xi Electronics.
Procedure
1. Apply AC line power to the transmitter.
2. Apply AC line power to the Rosemount 6888 Xi. Run the Quick Start Wizard as described below. At the Auto Cal Device screen, select the
26 Emerson.com/Rosemount
Page 27
October 2019 Quick Start Guide
calibration method based on the Rosemount 6888 Transmitter as follows:
Standard Probe Housing Configuration: Select None, SPS, or IMPS
as appropriate. Do not select Integral or calibration will not be possible.
Integral Autocal Probe housing: Select Integral only. If you don't
select Integral, calibration will not be possible.
3. Verify communications between the transmitter and the Rosemount 6888Xi.
The Rosemount 6888 Xi display is preconfigured to display O2 and cell temperature for single channel configurations and both O readings for dual channel configurations.
The transmitter probe takes approximately 45 minutes to warm up to the 1357 °F (736 °C) heater setpoint. The 4-20 mA signal remains at a default value of 3.5 mA, and the O2 reading remains at 0% through this warm-up period. After warm-up, the probe begins reading oxygen, and the 4-20 mA output is based on the default range of 0-10% O2.
If there is an error condition at startup, an alarm message is displayed.
2

2.3 Power up the Rosemount™ 6888 direct replacement probe (no electronics inside) with traditional architecture Rosemount 6888 Xi

Procedure
1. Apply AC line power to the Rosemount 6888 Xi.
2. Run the Quick Start Wizard as described in Run Rosemount™ 6888 Xi
Quick Start Wizard.
3. At the Auto Cal Device screen, select None, SPS, or IMPS as appropriate.
Do not select Integral or calibration will not be possible.
The transmitter probe takes approximately 45 minutes to warm up to the 736 °C (1357 °F) heater setpoint. The 4-20 mA signal remains at a default value of 3.5 mA, and the O2 reading remains at 0% through this warm-up period. After warm up, the probe begins reading oxygen, and the 4-20 mA output is based on the default range of 0-10% O2.
If there is an error condition at startup, an alarm message is displayed on the Rosemount 6888Xi.
Quick Start Guide 27
Page 28
Quick Start Guide October 2019

2.4 Run Rosemount™ 6888 Xi Quick Start Wizard

When you first power up the Rosemount 6888 Xi, a short wizard program guides you through the basic setup. Once configured, the Rosemount 6888Xi retains the setup, and the wizard will not repeat.
Procedure
1. Apply power to the Rosemount 6888 Xi. Once boot-up is complete, the Quick Start Wizard screen appears. With a dual channel Rosemount 6888 Xi, the wizard runs for both channels in succession.
2. Press Enter to continue.
3. At the Sensor Type screen, use the Up and Down keys to select O2.
Do not select CO as this option is reserved for future use.
4. Press Enter to continue.
5. At the Device Type screen use the Up and Down keys to select HART® or FF (FOUNDATION™ Fieldbus), whichever applies.
6. At the Auto Cal Device screen, use the Up and Down keys to select the calibration method to be used. The methods are defined as follows:
None: Manual calibration with the standard probe housing
configuration.
SPS: Automatic calibration with the standard probe housing
configuration using the Rosemount SPS 4001B.
IMPS: Automatic calibration with the standard probe housing
configuration using the Rosemount IMPS.
Integral: Automatic calibration with the integral autocal probe
housing configuration.
7. Press Enter to continue.
Note
If you select SPS, IMPS, or Integral, you must still configure automatic calibration as On. Verify other parameters, such as test gas values and gas times, as well. Refer to the Rosemount 6888 Xi Reference
Manual for calibration setup details.
8. When prompted by Setup Correct?, use the Up and Down keys to select Yes.
If you select No, the wizard restarts.
9. Press Enter to continue.
28 Emerson.com/Rosemount
Page 29
October 2019 Quick Start Guide
The Rosemount 6888 Xi displays several screens while saving the configuration, resets itself, and then returns to the main screen.

2.5 Calibration

2.5.1 Manual/semi-automatic calibration

A technician can calibrate the Rosemount 6888 probe with standard housing by following prompts via the display of the Rosemount 6888 Xi Electronics or via HART® communications to a Field Communicator or AMS console. The technician needs to manually switch the gases based upon these prompts. Emerson recommends using 0.4% O2 and 8% O2, balance nitrogen as calibration gases. Always use a two-stage pressure regulator set to 20 psi. Set the calibration gas flow meter for a maximum of 5 scfh with the cal gas fitting removed from the probe. A diffuser/filter that is plugged over time may cause the flow meter to deliver less flow to the sensing cell, but never readjust the flow rate until a new diffuser is installed. Readjusting the flow meter back up to the 5 scfh level could pressurize the cell during calibration and cause the O2 reading to shift downwards.
The electronics determine if the calibration was successful and calculate new calibration values. New calibration values are not automatically loaded into the electronics after a successful calibration, however. The technician has the opportunity to accept or reject the new values.
CAUTION
Reading errors
A loose or missing cap can permit fresh air to bias the O2 readings high in processes that run at negative pressure.
Make sure the calibration gas port is capped tightly between calibrations.

2.5.2 Fully automatic calibration

For fully automatic calibration, the Rosemount™ 6888 Xi Electronics must manage the actuation of solenoids to introduce gases into the probe.
Probe with standard electronics housing
In addition to the Rosemount™ 6888 Xi, this arrangement requires a separate single probe sequencer (SPS), which is a solenoid box for switching calibration gases or a larger intelligent multiprobe sequencer (IMPS) which can handle the autocal for up to four probes in one box.
The automatic calibrations can be initiated in several ways:
Calibration recommended diagnostic that is periodically checking cell
impedance.
Push button on the Rosemount 6888 Xi Electronics.
Quick Start Guide 29
Page 30
Quick Start Guide October 2019
HART® communications from a Field Communicator or AMS.
External contact closure.
Time since the last successful calibration.
If using the O2 measurement for automatic control, always place the O
2
control loop into manual prior to calibrating. Always inform the operator prior to calibrating. The Rosemount 6888 Xi Electronics provides an in cal contact closure for this purpose as well as an initiate cal contact.
The Rosemount 6888 Xi Electronics sequences the calibration gases in turn into the sensing cell. A 300 second flow time is the factory default for both gases and also for the purge cycle, which lets the probe signal come back to the normal flue gas readings. The 4-20 mA signal representing O2 can be held during the calibration cycle or permitted to vary with the bottled gases, in which case a record of the calibration can be trended at the DCS.
Calibration setup is found under the detailed setup menu.
Probe with integral autocal housing
This probe contains the autocal solenoids within the blue electronics housing, eliminating the need and cost for an SPS or IMPS solenoid enclosure. Both calibration gases are permanently piped into two ports on the probe. It's important to confirm that there are no piping leaks or the calibration bottles will leak down permanently.
Note
The calibration sequence from the Rosemount™ 6888 Xi Electronics is identical to that for the manual/semiautomatic calibration, but note that with the integral autocal version of this probe it is not possible to conduct a manual calibration. The factory offers a probe rebuild capability if solenoid or other failures occur.
CAUTION
Leaks
Calibration gas bottles ARE piped and under pressure at all times, so be sure to leak-check all fittings, tubing, and connections.
Always use dual-stage pressure regulators.
30 Emerson.com/Rosemount
Page 31
October 2019 Quick Start Guide

A Rosemount™ 6888 product certifications

Rev 1.9

A.1 European Directive information

The most recent revision of the EU Declaration of Conformity can be found at Emerson.com/Rosemount.

A.2 Ordinary location certification

As standard, the transmitter has been examined and tested to determine that the design meets the basic electrical, mechanical, and fire protection requirements by a nationally recognized test laboratory (NRTL) as accredited by the Federal Occupational Safety and Health Administration (OSHA).

A.3 Installing equipment in North America

The US National Electrical Code® (NEC) and the Canadian Electrical Code (CEC) permit the use of Division marked equipment in Zones and Zone marked equipment in Divisions. The marking must be suitable for the area classification, gas, and temperature class. This information is clearly defined in the respective codes.

A.4 USA/Canada

Model String Option Code
CSA Certificate
Standards
Markings
Quick Start Guide 31
C
70100758
CAN/CSA Standard C22.2 No.0-10:2015, CAN/CSA C22.2 No.61010-1-12 Harmonized UL 61010-1:2012 (3rd Edition), CAN/CSA Standard C22.2 No. 94.1-15 Harmonized ANSI/UL Standard 50 (2nd Edition), CAN/CSA Standard C22.2 No. 94.2-15 Harmonized ANSI/UL Standard 50 (2nd Edition), CAN/CSA C22.2 No. 60529:16 and ANSI/ISA 60529:04, CAN/CSA Standard C22.2 No. 30­M1986: 2016, CAN/CSA-C22.2 No. 60079-0: 2015, CAN/ CSA-C22.2 No. 60079-1:2016, FM 3600:2011, FM 3615:2006, ANSI/UL-60079-0:2013 (6th Edition), ANSI/ UL-60079-1:2015 (7th Edition)
Type 4X, IP66, Class 1, Division 1, Groups B, C and D; T3, Class 1, Zone 1, AEx db IIB+H2 T3 Gb, Ex db IIB+H2 T3 Gb: -40 °C ≤ Ta ≤ +70 °C ( Autocal Housing and Probe assembly); -40 °C ≤ Ta ≤ +90 °C (Standard Housing and Probe assembly eq. “DR Probe”)
Page 32
Quick Start Guide October 2019
Conditions of acceptability
1. The unit is intended to be connected to supply mains by qualified personnel in accordance with national (e.g. CEC, NEC, etc) and local codes.
2. Suitable APPROVED switch and fuse or a circuit breaker shall be provided to facilitate the disconnection of mains power.
3. The maximum operating ambient is considered as follows: 90 °C for 6888C DR Probe, 70 °C for 6888C Transmitter.
4. Mounting Flange temperature shall not exceed 190 °C during combustion process.
5. Calibration air lines and reference air lines shall not contain pure oxygen or combustible gas other than inert/oxygen gas mixture of which oxygen represents no more than that normally present in air.
6. The pressure within the enclosure and gas lines shall not be higher than 1.1 times the atmospheric pressure during the normal operations of the equipment.
7. The 6888C O2 Analyzers are used with the 6888 Xi Advanced Electronics, which must be installed in a Non-Hazardous Location, per wiring diagram 6R00131.
8. Meets Enclosure Type 4X & IP66 ratings when the reference air vent is routed to a dry area.
9. Units installed with conduit runs must have suitably certified conduit seals installed at the enclosure.
10. Units installed with other than conduit runs and conduit seals, must be fitted with certified or listed cable glands for use in “Class I, Zone 1, Ex/AEx d IIB+H2” and “Class I, Division 1, Group B, C and D” or better, suitable for the ambient temperature range.
11. Flameproof joints are not intended to be repaired.
A.5
32 Emerson.com/Rosemount
Europe
Model String Option Code
ATEX Certificate
Standards
Markings
A
Sira 14ATEX1031X
EN 60079-0:2012/A11:2013, EN 60079-1:2014
II 2 G Ex db IIB+H2 T3 Gb; IP66; -40 °C ≤ Ta ≤ +70°C ( Autocal Housing and Probe assembly); -40 °C ≤ Ta ≤ +90°C (Standard Housing and Probe assembly eq. “DR Probe”)
Page 33
October 2019 Quick Start Guide
Special conditions for safe use (X):
1. Mounting flange temperatures shall not exceed 190 °C during combustion process.
2. The 6888C O2 Analyzers are used with the 6888 Xi Advanced Electronics (associated equipment not part of this certification) which must be installed in a Safe Area.
3. Calibration air lines and reference air lines shall not contain pure oxygen or combustible gas other than inert/oxygen gas mixture of which oxygen represents no more than that normally present in air.
4. The pressure within the enclosure and gas lines shall not be higher than 1.1 times the atmospheric pressure during the normal operations of the equipment.
5. Fasteners property class must be A2-70 Stainless Steel.
6. Flameproof joints are not intended to be repaired.
Model String
A
Option Code
IECEx Certificate
Standards
Markings
IECEx CSA 14.0044X
IEC 60079-0:2012/A11:2013, IEC 60079-1:2014
Ex db IIB+H2 T3 Gb; IP66; -40°C ≤ Ta ≤ +70°C ( Autocal Housing and Probe assembly ); -40 °C ≤ Ta ≤ +90°C (Standard Housing and Probe assembly eq. “DR Probe”)
Special conditions for safe use (X):
1. Mounting Flange temperature shall not exceed 190 °C during combustion process.
2. The 6888C O2 Analyzers are used with the 6888 XI Advanced Electronics (associated equipment not part of this certification) which must be installed in a safe area.
3. Calibration air lines and reference air lines shall not contain pure oxygen or combustible gases other than inert/oxygen gas mixture of which oxygen represents no more than that normally present in air.
4. The pressure within the enclosure and gas lines shall not be higher than 1.1 times the atmospheric pressure during the normal operations of the equipment.
5. Fasteners property class must be A2-70 Stainless Steel.
6. Flameproof joints are not intended to be repaired.
Quick Start Guide 33
Page 34
Quick Start Guide October 2019
34 Emerson.com/Rosemount
Page 35
October 2019 Quick Start Guide

B Declaration of Conformity

Quick Start Guide 35
Page 36
Quick Start Guide October 2019
36 Emerson.com/Rosemount
Page 37
October 2019 Quick Start Guide

C China RoHS Table

Quick Start Guide 37
Page 38
Quick Start Guide October 2019
38 Emerson.com/Rosemount
Page 39
October 2019 Quick Start Guide
Quick Start Guide 39
Page 40
*00825-0100-4891*
00825-0100-4891, Rev. AB
Quick Start Guide
October 2019
GLOBAL HEADQUARTERS
Emerson Automation Solutions 6021 Innovation Blvd Shakopee, MN 55379, USA
+1 800 999 9307 or +1 952 906 8888 +1 952 949 7001 gas.csc@emerson.com
EUROPE
Emerson Automation Solutions Neuhofstrasse 19a P.O. Box 1046 CH-6340 Baar Switzerland
+ 41 (0) 41 768 6111 + 41 (0) 41 768 6300 gas.csc@emerson.com
Linkedin.com/company/Emerson-
Automation-Solutions
twitter.com/rosemount_news
Facebook.com/Rosemount
youtube.com/RosemountMeasurement
NORTH AMERICA
Rosemount 8200 Market Boulevard Chanhassen, MN 55317
+1 800 999 9307 +1 952 949 7001 gas.csc@emerson.com
MIDDLE EAST AND AFRICA
Emerson Automation Solutions Emerson FZE Jebel Ali Free Zone Dubai, United Arab Emirates, P.O. Box 17033
+971 4 811 8100 +971 4 886 5465 gas.csc@emerson.com
©
2019 Emerson. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.
Loading...