Emerson designs, manufactures, and tests its products to meet many national and international
standards. Because these instruments are sophisticated technical products, you must properly install,
use, and maintain them to ensure they continue to operate within their normal specifications. You
must adhere to the following instructions and integrate them into your safety program when
installing, using, and maintaining Emerson's Rosemount™ products. Failure to follow the proper
instructions may cause any one of the following situations to occur: loss of life, personal injury,
property damage, damage to this instrument, and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Emerson representative for
clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the
product.
• Install equipment as specified in the Installation section of this Quick Start Guide. Follow
appropriate local and national codes. Only connect the product to electrical and pressure sources
specified in this Quick Start Guide.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and
maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts
specified by Emerson. Unauthorized parts and procedures can affect the product's performance,
place the safe operation of your process at risk, and may result in fire, electrical hazards, or
improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified people, to prevent electrical shock and personal
injury.
Note
The information contained in this document is subject to change without notice.
WARNING
Hazardous area installation
Installations near flammable liquids or in hazardous area locations must be carefully evaluated by
qualified on site safety personnel.
To secure and maintain intrinsically safe installation, use an appropriate transmitter/safety barrier/
sensor combination. The installation system must be in accordance with the governing approval
agency (FM, CSA, or BASEEFA/CENELEC) hazardous area classification requirements. Consult your
transmitter Reference Manual for details.
Proper installation, operation, and servicing of this sensor in a hazardous area installation are
entirely the operator's responsibility.
2Emerson.com/Rosemount
November 2020Quick Start Guide
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end
users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your
system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for
all systems used within the facility.
CAUTION
Sensor/process application compatibility
The wetted sensor materials may not be compatible with process composition and operating
conditions.
Application compatibility is entirely the operator's responsibility.
CAUTION
Sensor/process application compatibility
All pH/ORP sensors have a plastic enclosure which must only be cleaned with a damp cloth to
avoid the danger due to a build up of electrostatic charge.
All pH/ORP sensor models are intended to be in contact with the process fluid and may not meet
the 500 r.m.s.a.c. test to earth.
Take this into consideration at the time of installation.
1. Inspect the outside of the carton for any damage. If you detect
damage, contact the carrier immediately.
2. Inspect the hardware. Make sure all the items in the packing list are
present and in good condition. Notify the factory if any part is
missing.
Note
Save the original packing cartons and materials as most carriers
require proof of damage due to mishandling, etc. Also, if it is
necessary to return the sensor to the factory, you must pack the
sensor in the same manner as it was received. Refer to Return of
material for return instructions.
WARNING
Buffer solution in the vinyl boot may cause skin or eye irritation.
Wear personal protective equipment.
Avoid contact with skin and eyes.
CAUTION
Glass electrode must be wetted at all times (in storage and in line) to
maximize sensor life.
Postrequisites
If the sensor appears to be in satisfactory condition, proceed to Mount
sensor.
1.2
Quick Start Guide5
Mount sensor
Emerson has designed the sensor to be located in industrial process
environments.
Do not exceed temperature and pressure limitations at any time. A caution
label regarding this matter is attached to the sensor. Please do not remove
the label.
See Figure 1-1 for dimensional information to ensure proper installation into
your process.
Quick Start GuideNovember 2020
WARNING
Pressurized spray injury
US and foreign patents pending.
Do not exceed pressure and temperature specifications: 0 to 150 psig (0
to 1135 kPa), 32 to 212 °F (0 to 100 °C).
Remove pressure and allow to cool before removal.
Read and follow Quick Start Guide.
Figure 1-1: Dimensional Drawing
A. 1-in. (25.4 mm) male national pipe thread (MNPT), two places
B. 1-in. (25.4 mm) wrench opening
C. Reference junction
D. End tip option -41
E. Solution ground
F. Electrode
G. Sensor cable (or Variopol connector - not shown)
6Emerson.com/Rosemount
November 2020Quick Start Guide
WARNING
Internal electrolyte fill solution may cause skin or eye irritation.
Wear personal protection equipment (PPE).
Avoid contact with skin and eyes.
1. Shake the sensor in a downward motion to remove any air bubbles
that may be present inside the tip of the pH glass.
2. Do not install the sensor on the horizontal. The sensor must be 10
degrees off the horizontal to ensure accuracy.
3. Do not install the sensor upside down.
4. Air bubbles may become trapped in the sensor end between the
glass bulb and the sensor body. This problem is most commonly
encountered in areas of low flow or during calibration. Shake the
probe while immersed in solution to remove bubbles. To avoid this
problem, order the sensor with the slotted tip (option -41).
In most cases, you can simply install the pH sensor as shipped and obtain
readings with an accuracy of ±0.6 pH. To obtain greater accuracy or to verify
proper operation, calibrate the sensor as a loop with its compatible
transmitter.
1.2.1Flow through and insertion mounting
Rosemount 396P and 396PVP Sensors have a 1-in. (25.4 mm) male national
pipe thread (MNPT) process connection at the front of the sensor for
mounting into a 1½-in. (38.1 mm) tee or the process pipes.
See Figure 1-2 through Figure 1-8 for installation configurations.
Note
Do not use large pipe wrenches to tighten the sensor into a flange or other
type of mounting.
Quick Start Guide7
Quick Start GuideNovember 2020
Figure 1-2: Flow-Through Tee with Adapter (PN 915240-xx)
Pressure/temperature: 60 psig at 120 °F (514 kPa at 49 °C)
A. Adapter retrofit PN 33211-00
B. O-ring must be in place prior to use
C. Process connection threads - two places
D. 2-in. (50.8 mm) sched. 80 T
E. Sensor cable
Ordering option (xx)
03¾ in. (19.1 mm)
041 in. (25.4 mm)
051½ in. (38.1 mm)
8Emerson.com/Rosemount
Process connection threads
November 2020Quick Start Guide
Figure 1-3: Flow-Through and Insertion Installations
A. 1½-in. x 1-in. (38.1 mm x 25.4 mm) reducing bushing
B. 1½-in. (38.1 mm) pipe tee PN 2002011
C. 1½-in. (38.1 mm) pipe Y
D. Flow
Quick Start Guide9
Quick Start GuideNovember 2020
Figure 1-4: Rosemount 396P with Insertion Mounting Adapter (PN
23242-02)
Not for use with Rosemount 396PVP. Mounting adapter allows for sensor
removal without twisting or disconnecting interconnecting cable for ease of
maintenance.
A. Cable or Variopol connector (not shown)
B. Peek adapter 1-in. (25.4 mm) female national pipe thread (FNPT) x ¾-in.
(19.1 mm) FNPT
C. 2-135 Viton™ O-ring. O-ring must be in place prior to use (PN 9550175).
D. Nut, hex union 2-in. (50.8 mm); 3-in. (76.2 mm) wrench opening (304
stainless steel)
E. 3.531.8 acme thread (typ)
F. Neck, union fitting (316 stainless steel) 2⅝-in. (66.7 mm) wrench
opening
G. 1½-in. (38.1 mm) male national pipe thread (MNPT). Insertion mounting
A. End cap PVC
B. Sensor cable
C. 1½-in. (38.1 mm) PVC pipe schedule 80
D. 1½-in. (38.1 mm) pipe clamp, three places
E. Sensor model Rosemount 396P. Regularly check to make sure
connections are water tight.
F. Sweep pipe with 1-in. (25.4 mm) FNPT adapter
G. Unistrut 1⅝ x 1⅝-in. (41.3 x 41.3 mm) aluminum
H. 1½-in. (38. mm) pipe clamp, two places
I. Coupling
J. Mounting channel aluminum, two places
K. Can be any convenient dimension
L. Customer handrail, two places
M. Locking pin with bead chain
Quick Start Guide13
Quick Start GuideNovember 2020
Figure 1-7: Junction Box
A. Junction box
B. 2-in. (50.8 mm) pipe mounting bracket (PN 2002565)
C. Flexible conduit if required
D. 1-in. (25.4 mm) pipe by others
E. 1-in. (25.4 mm) FNPT CPVC union (PN 9320057)
14Emerson.com/Rosemount
November 2020Quick Start Guide
Figure 1-8: Jet Spray Cleaner (PN 12707-00)
A. Cleaning solution by others
B. Solenoid valve or manual valve (supplied by others)
C. Corrosion resistant tubing (supplied by others)
D. Polypropylene ¼-in. (6.4 mm) compression fitting
E. ¼-in. (6.4 mm) 316 stainless steel
F. ¼-in. (6.4 mm) polypropylene
G. Stainless set screw for adjustable spray nozzle height
H. 2-in. (50.8 mm) NPT threads
I. Sensor
J. 1-in. (25.4 mm) PVC couping for submersible applications (supplied by
others)
K. 1-in. (25.4 mm) PVC or stainless conduit (supplied by others)
L. Cable
M. Timer supplied by others or use timer feature in Rosemount instrument
Quick Start Guide15
Quick Start GuideNovember 2020
1.2.2Submersion mounting
Rosemount 396P and 396PVP sensors also have a 1-in. (25.4 mm) male
national pipe thread (MNPT) process connection at the back of the sensor.
Using a standard 1-in. (25.4 mm) union, you can mount the sensor to a 1-in.
(25.4 mm) schedule 80 CPVC or PVDF standpipe.
Tapered pipe threads in plastic tend to loosen after installation. Therefore,
Emerson recommends using PTFE tape on the threads and checking the
tightness of the connection frequently to assure that no loosening has
occurred. To prevent rain water or condensation from running into the
sensor, Emerson recommends using a weatherproof junction box. Run the
sensor cable through a protective conduit for isolation from electrical
interference or physical abuse from the process. Install the sensor within 80
degrees of vertical, with the electrode facing down. Do not run the sensor's
cable with power or control wiring.
16Emerson.com/Rosemount
November 2020Quick Start Guide
2Wire
2.1General wiring guidelines
Figures in Wiring diagrams provide guidelines for wiring the Rosemount
396P sensor to various transmitters.
To determine which wiring diagram to use, locate the model number of the
sensor to be installed.
If you need to extend the cable, use a high quality eleven conductor doubleshielded instrument cable (part number 9200273) available from Emerson.
Note
If the cable is too long, loop up the excess cable. If you need to shorten the
cable, cut and terminate each conductor neatly and make sure that the
overall (outermost) drain wire is not shorted out with either of the two inner
drain wires (shields).
Run signal cable in a dedicated conduit (preferably an earth grounded
metallic conduit) and keep it away from AC power lines. For your
convenience, Emerson has furnished a wire nut kit (in a plastic bag wrapped
around the cable).
CAUTION
For maximum electromagnetic interference/radio frequency interference
(EMI/RFI) protection when wiring from the sensor to the junction box,
connect the outer braid of the sensor to the outer braided shield of the
extension cable. Terminate the outer braid of the extension cable to the
instrument at earth ground or use an appropriate metal cable gland fitting
to provide a secure connection to the instrument cable.
2.2Wiring diagrams
The Rosemount 396P has an optional built-in preamplifier and comes with a
shielded cable.
WARNING
Serious injury may result.
Do not connect sensor cable to power lines.
Quick Start Guide17
Quick Start GuideNovember 2020
CAUTION
Handle the cable carefully and keep it dry and free of corrosive chemicals
at all times.
Take extreme care to prevent it from being twisted, damaged, or
scraped by rough, sharp edges or surfaces.
Note
Remove electrical tape or shrink sleeve away from gray reference wire
before connecting wire to terminal.
Note
For additional information on this product, including sensor combinations
not shown here, please refer to our website Wiring Diagrams.
18Emerson.com/Rosemount
November 2020Quick Start Guide
Figure 2-1 and Table 2-1 are applicable to option codes -01, -03, -04, and
-05.
Figure 2-1: Wiring for Rosemount 396P (Gray Cable) and Rosemount
1056, 56, or 1057
If there is a shrink sleeve on the gray reference wire, remove it before
connecting the wire to the terminal. For Rosemount 1056, terminate inner
drain with a wire nut.
Table 2-1: Wiring for Rosemount 396P (Gray Cable) and Rosemount 1056, 56, or
1057
Terminal numberWire colorConnects to
N/AShieldN/A
1WhiteResistance temperature
2White/redRTD sense
3RedRTD in
4BlueSolution ground
5Green+5 Vdc
6Brown-5 Vdc
7White/blackpH shield
8BlackpH in
9White/grayReference shield
10GrayReference
Quick Start Guide19
device (RTD) return
Quick Start GuideNovember 2020
Figure 2-2 and Table 2-2 are applicable to option codes -01, -03, -04, and
-05.
Figure 2-2: Wiring for Rosemount 396P (Blue Cable) and Rosemount
1056, 56, or 1057
If there is a shrink sleeve on the gray reference wire, remove it before
connecting the wire to the terminal. For Rosemount 1056, terminate inner
drain with a wire nut.
Table 2-2: Wiring for Rosemount 396P (Blue Cable) and Rosemount 1056, 56, or
1057
Terminal numberWire colorConnects to
1WhiteRTD return
2White/redRTD sense
3RedRTD in
4BlueSolution ground
5Inner drain+5 Vdc
6White/gray-5 Vdc
7ShieldpH shield
8OrangepH in
9N/AReference shield
10GrayReference
N/AGreenN/A
20Emerson.com/Rosemount
November 2020Quick Start Guide
Figure 2-3 and Table 2-3 are applicable to option codes -02, -07, and -0.8
Figure 2-3: Wiring for Rosemount 396P and Rosemount 1056, 56, or
1057
If there is a shrink sleeve on the gray reference wire, remove it before
connecting the wire to the terminal. For Rosemount 1056, terminate inner
drain with a wire nut.
Table 2-3: Wiring for Rosemount 396P and Rosemount 1056, 56, or 1057
Terminal numberWire colorConnects to
N/AShieldN/A
1WhiteRTD return
2White/redRTD sense
3RedRTD in
4BlueSolution ground
5N/A+5 Vdc
6N/A-5 Vdc
7BraidpH shield
8OrangepH in
9BraidReference shield
10GrayReference
Quick Start Guide21
Quick Start GuideNovember 2020
Figure 2-4 and Table 2-4 are applicable to option codes -02, -07, and -08.
Figure 2-4: Wiring for Rosemount 396P (Blue Cable) and Rosemount
1056, 56, or 1057
If there is a shrink sleeve on the gray reference wire, remove it before
connecting the wire to the terminal. For Rosemount 1056, terminate inner
drain with a wire nut.
Table 2-4: Wiring for Rosemount 396P (Blue Cable) and Rosemount 1056, 56, or
1057
Terminal numberWire colorConnects to
N/AShieldN/A
1WhiteRTD return
2White/redRTD sense
3RedRTD in
4BlueSolution ground
5N/A+5 Vdc
6N/A-5 Vdc
7BraidpH shield
8OrangepH in
9White/grayReference shield
10GrayReference
N/AGreenGround
22Emerson.com/Rosemount
November 2020Quick Start Guide
Figure 2-5: Wiring for Rosemount 396PVP without Integral Preamplifier
and Rosemount 1056, 56, or 1057
If there is a shrink sleeve on the gray reference wire, remove it before
connecting the wire to the terminal. For Rosemount 1056, terminate inner
drain with a wire nut.
Table 2-5: Wiring for Rosemount 396PVP without Integral Preamplifier and
Rosemount 1056, 56, or 1057
Terminal numberWire colorConnects to
N/AClearN/A
1WhiteRTD return
2White/redRTD sense
3RedRTD in
4BlueSolution ground
5N/A+5 Vdc
6N/A-5 Vdc
7ClearpH shield
8OrangepH in
9ClearReference shield
10GrayReference
Quick Start Guide23
Quick Start GuideNovember 2020
Figure 2-6: Wiring for Rosemount 396PVP-70 and Rosemount 1056, 56,
or 1057
Table 2-6: Wiring for Rosemount 396PVP-70 and Rosemount 1056, 56, or 1057
Terminal numberWire colorConnects to
1WhiteRTD return
2White/redSense
3RedRTD in
4BlueSolution ground
5Inner drain+5 Vdc
6White/gray-5 Vdc
7ClearpH shield
8OrangepH in
9N/AReference shield
10GrayReference
N/AGreenNo connection or earth
ground
24Emerson.com/Rosemount
November 2020Quick Start Guide
Figure 2-7 and Table 2-7 are applicable to option codes -01, -03, -04, -05,
and -06.
Figure 2-7: Wiring for Rosemount 396P (Gray Cable) and Rosemount
1066
Note
• If ground lead is present, terminate it to green ground screw on inner
enclosure.
• TB5, TB7, and TB8 not used for pH/ORP sensor wiring.
Table 2-7: Wiring for Rosemount 396P (Gray Cable) and Rosemount 1066
Terminal numberWire colorConnects to
TB3WhiteRTD return
TB3White/redRTD sense
TB3RedRTD in
TB4GreenPreamplifier: +volts
TB4BrownPreamplifier: -volts
TB1Black/whitepH shield
TB1BlackpH in
TB2BlueSolution ground
TB2White/grayReference shield
TB2GrayReference in
Quick Start Guide25
Quick Start GuideNovember 2020
Figure 2-8 and Table 2-8 are applicable to option codes -01, -03, -04, -05,
and -06.
Figure 2-8: Wiring for Rosemount 396P (Blue Cable) and Rosemount
1066
Note
• If ground lead is present, terminate it to green ground screw on inner
enclosure.
• TB5, TB7, and TB8 not used for pH/ORP sensor wiring.
Table 2-8: Wiring for Rosemount 396P (Blue Cable) and Rosemount 1066
Terminal numberWire colorConnects to
TB3WhiteRTD return
TB3White/redRTD sense
TB3RedRTD return in
TB4Inner drainPreamplifier: +volts
TB4White/grayPreamplifier: -volts
TB1ShieldpH shield
TB1OrangepH in
TB2BlueSolution ground
TB2GrayReference in
26Emerson.com/Rosemount
November 2020Quick Start Guide
Figure 2-9 and Table 2-9 are applicable to option codes -02, -07, and -08.
Figure 2-9: Wiring for Rosemount 396P (Gray Cable) and Rosemount
1066
Note
• If ground lead is present, terminate it to green ground screw on inner
enclosure.
• TB5, TB7, and TB8 not used for pH/ORP sensor wiring.
Table 2-9: Wiring for Rosemount 396P (Gray Cable) and Rosemount 1066
Terminal numberWire colorConnects to
TB3WhiteRTD return
TB3White/redRTD sense
TB3RedRTD return in
TB1Braid or clearpH shield
TB1OrangepH in
TB2BlueSolution ground
TB2Braid or clearReference shield
TB2GrayReference in
Quick Start Guide27
Quick Start GuideNovember 2020
Figure 2-10 and Table 2-10 are applicable to option codes -02, -07, and -08.
Figure 2-10: Wiring for Rosemount 396P (Blue Cable) and Rosemount
1066
Note
• If ground lead is present, terminate it to green ground screw on inner
enclosure.
• TB5, TB7, and TB8 not used for pH/ORP sensor wiring.
Table 2-10: Wiring for Rosemount 396P (Blue Cable) and Rosemount 1066
Terminal numberWire colorConnects to
TB3WhiteRTD return
TB3White/redRTD sense
TB3RedRTD return in
TB4Inner drainPreamplifier: +volts
TB4White/grayPreamplifier: -volts
TB1ShieldpH shield
TB1OrangepH in
TB2BlueSolution ground
TB2GrayReference in
28Emerson.com/Rosemount
November 2020Quick Start Guide
Figure 2-11: Wiring for Rosemount 396PVP and Rosemount 1066
Note
• If ground lead is present, terminate it to green ground screw on inner
enclosure.
• TB5, TB7, and TB8 not used for pH/ORP sensor wiring.
Table 2-11: Wiring for Rosemount 396PVP and Rosemount 1066
Terminal blockWire colorConnects to
TB3WhiteRTD return
TB3White/redRTD sense
TB3RedRTD return in
TB1Braid or clearpH input - pH shield
TB1OrangepH input - pH in
TB2BlueSolution ground
TB2White/grayReference shield
TB2GrayReference in
Quick Start Guide29
Quick Start GuideNovember 2020
Figure 2-12: Wiring for Rosemount 396PVP-70 and Rosemount 1066
Note
• If ground lead is present, terminate it to green ground screw on inner
enclosure.
• TB5, TB7, and TB8 not used for pH/ORP sensor wiring.
Table 2-12: Wiring for Rosemount 396PVP-70 and Rosemount 1066
Terminal blockWire colorConnects to
TB3WhiteRTD return
TB3White/redRTD sense
TB3RedRTD return in
TB4Inner drainPreamplifier: +volts
TB4White/grayPreamplifier: -volts
TB1ShieldpH shield
TB1OrangepH in
TB2BlueSolution ground
TB2GrayReference in
30Emerson.com/Rosemount
November 2020Quick Start Guide
Figure 2-13 and Table 2-13 are applicable to option codes -01, -03, -04, -05,
and -06.
Figure 2-13: Wiring for Rosemount 396P (Gray Cable) and Rosemount
5081-P-HT
If there is a shrink sleeve on the gray reference wire, remove it before
connecting the wire to the terminal.
Table 2-13: Wiring for Rosemount 396P (Gray Cable) and Rosemount 5081-P-HT
Terminal
number
1N/AN/A9White/blackDrain
2ShieldN/A10BlackmV in
3WhiteRTD return11Brown-5 Vdc
4White/redRTD sense12Green+5 Vdc
5RedRTD in13N/AAnode
6White/grayDrain14N/ACathode
7GrayReference15N/A-24 Vdc
8BlueSolution
Quick Start Guide31
Wire colorConnects toTerminal
number
16N/A+24 Vdc
ground
Wire colorConnects to
Quick Start GuideNovember 2020
Figure 2-14 and Table 2-14 are applicable to option codes -01, -03, -04, -05,
and -06.
Figure 2-14: Wiring for Rosemount 396P (Blue Cable) and Rosemount
5081-P-HT
If there is a shrink sleeve on the gray reference wire, remove it before
connecting the wire to the terminal.
Table 2-14: Wiring for Rosemount 396P (Blue Cable) and Rosemount 5081-P-HT
Terminal
number
N/AGreenN/A9ShieldDrain
1N/AN/A10OrangemV in
2N/AN/A11White/gray-5 Vdc
3WhiteRTD return12Inner drain+5 Vdc
4White/redRTD sense13N/AAnode
5RedRTD in14N/ACathode
6N/ADrain15N/A-24 Vdc
7GrayReference16N/A+24 Vdc
8BlueSolution
32Emerson.com/Rosemount
Wire colorConnects toTerminal
number
ground
Wire colorConnects to
November 2020Quick Start Guide
Figure 2-15 and Table 2-15 are applicable to models -02, -07, and -08.
Figure 2-15: Wiring for Rosemount 396P (Gray Cable) and Rosemount
5081-P-HT
If there is a shrink sleeve on the gray reference wire, remove it before
connecting the wire to the terminal.
Table 2-15: Wiring for Rosemount 396P (Gray Cable) and Rosemount 5081-P-HT
Terminal
number
1N/AN/A9BraidDrain
2ShieldN/A10OrangemV in
3WhiteRTD return11N/A-5 Vdc
4White/redRTD sense12N/A+5 Vdc
5RedRTD in13N/AAnode
6BraidDrain14N/ACathode
7GrayReference15N/A-24 Vdc
8BlueSolution
Quick Start Guide33
Wire colorConnects toTerminal
number
16N/A+24 Vdc
ground
Wire colorConnects to
Quick Start GuideNovember 2020
Figure 2-16 and Table 2-16 are applicable to models -02, -07, and -08.
Figure 2-16: Wiring for Rosemount 396P (Blue Cable) and Rosemount
5081-P-HT
If there is a shrink sleeve on the gray reference wire, remove it before
connecting the wire to the terminal.
Table 2-16: Wiring for Rosemount 396P (Blue Cable) and Rosemount 5081-P-HT
Terminal
number
N/AGreenN/A9BraidDrain
1N/AN/A10OrangemV in
2ShieldN/A11N/A-5 Vdc
3WhiteRTD return12N/A+5 Vdc
4White/redRTD sense13N/AAnode
5RedRTD in14N/ACathode
6White/grayDrain15N/A-24 Vdc
7GrayReference16N/A+24 Vdc
8BlueSolution
34Emerson.com/Rosemount
Wire colorConnects toTerminal
number
ground
Wire colorConnects to
November 2020Quick Start Guide
Figure 2-17: Wiring for Rosemount 396PVP and Rosemount 5081-P-HT
If there is a shrink sleeve on the gray reference wire, remove it before
connecting the wire to the terminal.
Table 2-17: Wiring for Rosemount 396PVP and Rosemount 5081-P-HT
Terminal
number
N/AGreenN/A9ClearDrain
1N/AN/A10OrangemV in
2ClearN/A11N/A-5 Vdc
3WhiteRTD return12N/A+5 Vdc
4White/redRTD sense13N/AAnode
5RedRTD in14N/ACathode
6White/grayDrain15N/A-24 Vdc
7GrayReference16N/A+24 Vdc
8BlueSolution
Quick Start Guide35
Wire colorConnectstoTerminal
number
ground
Wire colorConnects
to
Quick Start GuideNovember 2020
Figure 2-18: Wiring for Rosemount 396PVP-70 and Rosemount 5081
Table 2-18: Wiring for Rosemount 396PVP-70 and Rosemount 5081
Terminal
number
N/AGreenN/A9ClearDrain
1N/AN/A10OrangemV in
2N/AN/A11White/gray-5 Vdc
3WhiteRTD return12Inner drain+5 Vdc
4White/redRTD sense13N/AAnode
5RedRTD in14N/ACathode
6N/ADrain15N/A-24 Vdc
7GrayReference16N/A+24 Vdc
8BlueSolution
Wire colorConnects toTerminal
number
ground
Wire colorConnects to
36Emerson.com/Rosemount
November 2020Quick Start Guide
3Start up and calibrate
3.1Calibrate Rosemount 396P and 396PVP pH sensors
3.1.1Prepare sensor
Procedure
1. Shake down the sensor to remove any air bubbles that may be
present at the tip of the pH glass bulb.
2. To obtain greater accuracy or to verify proper operation, calibrate
the sensor as a loop with its compatible transmitter.
3.1.2Calibrate pH
1. Establish a temporary connection between the sensor and the
transmitter.
2. Perform a buffer calibration.
3. Consult the appropriate pH/ORP transmitter Reference Manual for
specific calibration and standardization procedures, or see Calibrate
with two buffer solutions for the recommended two-point buffer
calibration procedure.
Calibrate with two buffer solutions
Prerequisites
Select two stable buffer solutions, preferably pH 4.0 and 7.0. (You can use
pH buffers other than pH 4.0 and pH 7.0 as long as the pH values are at least
two pH units apart).
Note
A pH 7 buffer solution reads an mV value of approximately zero, and pH
buffers read approximately ±59.1 mV for each pH unit above or below pH 7.
Check the pH buffer manufacturer specifications for millivolt values at
various temperatures, as it may affect the actual value of the buffer solution
mV/pH value.
Procedure
1. Immerse the sensor in the first buffer solution. Allow the sensor to
adjust to the buffer temperature (to avoid errors due to temperature
differences between the buffer solution and sensor temperature) and
wait for readings to stabilize.
The transmitter can now acknowledge the value of the buffer.
2. Once the first buffer has been acknowledged by the transmitter,
rinse the buffer solution off the sensor with distilled or deionized
water.
Quick Start Guide37
Quick Start GuideNovember 2020
3. Repeat Step 1 and Step 2 using the second buffer solution.
Once the transmitter has acknowledged both buffer solutions, a sensor
slope (mV/pH) is established (the slope value can be found within the
transmitter). The slope value should read about 59.1 mV/pH for a new
sensor and will decrease over time to approximately 47-49 mV/pH.
Postrequisites
Once the slope reads below the 47-49 mV/pH range, install a new sensor to
maintain accurate readings.
Standardize pH sensor
For maximum accuracy, you can standardize the sensor in-line or with a
process grab sample after performing a buffer calibration and conditioning
the sensor to the process. Standardization accounts for the sensor junction
potential and other interferences. Standardization does not change the
sensor's slope, but simply adjusts the transmitter's reading to match that of
a known process pH.
Procedure
1. While obtaining a process solution sample, record the pH value that
is shown on the transmitter display.
Emerson recommends taking the sample close to the sensor.
2. Measure and record the pH of the process solution sample with
another temperature compensated, calibrated pH instrument.
For best results, perform standardization at the process temperature.
3. Adjust the transmitter to the standardized value.
3.2
Calibrate Rosemount 396P and 396PVP oxidation
reduction potential (ORP) sensors
Most industrial applications have a number of ORP reactions occuring in
sequence or simultaneously. Reagents can oxidize or reduce several
components. Theoretically, the ORP potential is absolute, because it is the
result of the oxidation/reduction equilibrium. However, the actual measured
potential is dependent on many factors, including the condition of the
surface of the ORP platinum electrode. Therefore, allow the sensor one to
two hours to become conditioned to the stream when it is first set up or
after cleaning it.
Procedure
1. Make a temporary electrical connection between the sensor and the
instrument.
2. Obtain an ORP standard solution (PN R508-8OZ) or make one by
adding a few crystals of quinhydrone to either pH 4 or pH 7 buffer.
38Emerson.com/Rosemount
November 2020Quick Start Guide
Quinhydrone is only slightly soluble; therefore, use only a few
crystals.
3. Immerse the sensor in the standard solution. Allow one to two
minutes for the ORP sensor to stabilize.
4. Adjust the standardized control of the transmitter to the solution
value shown in Table 3-1.
The resulting potentials, measured with a clean platinum electrode
and saturated KCl/AgCL reference electrode, should be within ±20
millivolts of the value shown in Table 3-1. Note solution temperature
to ensure accurate interpretation of results. The ORP value of
saturated quinhydrone solution is not stable over long periods of
time. Therefore, make these standards fresh each time they are used.
Table 3-1: ORP of Saturated Quinhydrone Solution
pH 4pH 7
Temp ºF
(°C)
mV
potential
68 (20)77 (25)86 (30)68 (20)77 (25)86 (30)
268264260948780
5. Remove the sensor from the buffer, rinse, and install in the process.
Quick Start Guide39
Quick Start GuideNovember 2020
4Maintenance
4.1General maintainence information
The sensors require mimimal maintenance.
Keep the sensor clean and free of debris or sediment at all times. The nature
of the solution measured determines the frequency of cleaning. To clean,
wipe with a soft cloth or brush with a brush. Remove the sensor from the
process periodically and check it in buffer solutions.
WARNING
Before removing the sensor, be absolutely certain that the process pressure
is reduced to 0 psig and the process temperature is lowered to a safe level.
4.2Automatic temperature compensator
The temperature compensator element is a temperature sensitive resistor
and can be checked with an ohmeter. Resistance increases with
temperature.
PT-100 reads 110 ohms. Resistance varies with temperature and can be
determined according to Table 4-2 or the following formula:
RT = RO [1 + R1(T-20)]
Where RT = Resistance and T = Temperature in °C
Refer to Table 4-1 for RO and R1 values.
Table 4-1: R
Temperature elementR
PT-100107.70.00385
and R1 Values for Temperature Compensation Elements
O
O
R
1
Table 4-2: Temperature vs. Resistance of Automatic Temperature
Compensation Elements
Temperature °F (°C)Resistance (Ohms) ±1% PT-100
32 (0)100.0
50 (10)103.8
68 (20)107.7
77 (25)109.6
86 (30)111.5
40Emerson.com/Rosemount
November 2020Quick Start Guide
Table 4-2: Temperature vs. Resistance of Automatic Temperature
Compensation Elements (continued)
Temperature °F (°C)Resistance (Ohms) ±1% PT-100
104 (40)115.4
122 (50)119.2
140 (60)123.1
158 (70)126.9
176 (80)130.8
194 (90)134.6
212 (100)138.5
4.3Clean electrode
If the electrode is coated or dirty, clean as follows:
Procedure
1. Remove the sensor from process.
2. Wipe the glass bulb with a soft, clean, lint free cloth or tissue. If this
does not remove the dirt or coating, go to Step 3.
Detergents clean oil and grease; acids remove scale.
3. Wash the glass bulb in a mild detergent solution. If this does not
clean the glass bulb, go to Step 4.
4.
WARNING
Corrosive substance
The solution used during the following step is an acid.
Handle with care.
Follow the directions of the acid manufacturer.
Wear the proper protective equipment.
Do not let the solution come in contact with skin or clothing.
If contact with skin is made, immediately rinse with clean water.
Wash the bulb in a dilute five percent hydrochloric acid solution and
rinse with clean water.
Soaking the sensor overnight in the acid solution can improve
cleaning action.
Quick Start Guide41
Quick Start GuideNovember 2020
Note
You may get erroneous pH results immediately after acid soak due to
reference junction potential buildup. Replace the sensor if cleaning
does not restore sensor operation.
4.4Check platinum electrode
Check the platinum electrode as follows. There are two types of standard
solutions which may be used to check the oxidation reduction potential
(ORP) electrode/transmitter system.
Type 1: One type of commonly used ORP standard solution is the saturated
quinhydrone solution (PN R508-8OZ). Refer to Calibrate Rosemount 396P
and 396PVP oxidation reduction potential (ORP) sensors.
WARNING
The solution used during the following check is an acid.
Handle the solution with care.
Follow the manufacturer's directions.
Wear the proper protective equipment.
If contact with skin or clothing is made, immediately rinse with plenty of
clean water.
Type 2: A second ORP standard solution can be prepared from the following
recipe:
Procedure
1. Dissolve 39.2 grams of reagent grade ferrous ammonium sulfate,
Fe(NH4)2(SO4)2 ● 6H2O and 48.2 grams of reagent grade ferric
ammonium sulfate, FeNH4(SO4)2 ● 12 H2O in approximately 23.7 oz.
(700 ml) of water.
Distilled water is preferred, but tap water is acceptable.
2. Slowly and carefully add 1.9 oz. (56.2 ml) of concentrated sulfuric
acid.
3. Add sufficient water to bring the total solution volume up to 33.8 oz.
(1000 ml).
This standard ORP solution, although not as simple to prepare as the
quinhydrone recipe, is much more stable and will maintain its millivolt value
for approximately one year when stored in glass containers. This solution
(ferric/ferrous ammonium sulfate) produces a nominal ORP of 476 +20 mV
at 77 °F (25 ºC) when used with a saturated KCL/AgCl reference electrode
and platinum measuring electrode. Expect some tolerance in mV values due
to the rather large liquid reference junction potentials that can arise when
42Emerson.com/Rosemount
November 2020Quick Start Guide
measuring this strongly acidic and concentrated solution. However, if you
keep measuring electrodes clean and in good operating condition, you can
carry out consistently repeatable calibrations using this standard solution.
4.5Clean platinum electrode
To restore the electrode to normal operation, clean the platinum electrode
with baking soda. Polish it by rubbing it with a damp paper towel and baking
soda until it appears bright and shiny.
Quick Start Guide43
Quick Start GuideNovember 2020
5Diagnostics and troubleshooting
5.1Transmitter troubleshooting
Many Rosemount instruments and transmitters automatically search for
fault conditions that would cause an error in the measured pH value. Refer to
the applicable Reference Manual for a complete description of the
transmitter's fault conditions.
The sections below list some of the diagnostic messages that indicate a
possible sensor problem as well as a description of the problem and a
suggested remedy.
5.1.1Calibration warning
CALibrAte
Potential cause
Aged glass.
Recommended action
Perform buffer calibration.
Potential cause
Sensor not immersed.
Recommended action
Make sure electrode measuring tip is fully submerged in the process
liquid.
5.1.2Cracked glass failure
GLASS fAIL
Potential cause
Broken or cracked glass.
Recommended action
Replace sensor.
5.1.3High reference impede
rEF fAIL or rEF WArn
Potential cause
Liquid junction coated.
44Emerson.com/Rosemount
November 2020Quick Start Guide
Recommended action
Clean sensor; replace if necessary.
Potential cause
Reference cell gel depleted.
Recommended action
Replace sensor.
Potential cause
Sensor not fully immersed.
Recommended action
Make sure electrode tip is fully immersed in the process solution.
5.1.4Input voltage high or input voltage low
Potential cause
pH input shorted or sensor miswired.
Recommended action
Check wiring. Replace sensor if necessary.
5.1.5Old glass warning
GLaSSWArn
Potential cause
Glass electrode worn out.
Recommended action
Replace sensor.
Potential cause
Sensor not fully immersed.
Recommended action
Make sure electrode tip is fully immersed in the process solution.
5.1.6Reference offset err
Std Err
Offline only.
Quick Start Guide45
Quick Start GuideNovember 2020
Potential cause
Reference electrode poisoned.
Recommended action
Replace the sensor.
5.1.7Ref voltage high or ref voltage low
Potential cause
Reference shorted or miswired.
Recommended actions
1. Check wiring and installation.
2. Replace sensor if necessary.
Potential cause
Sensor not fully immersed.
Recommended action
Make sure electrode tip is fully immersed in the process solution.
5.1.8Sensor miswired
Potential cause
Open wire between sensor and transmitter.
Recommended action
Check wiring.
Potential cause
Bad preamplifier.
Recommended action
Replace preamplifier (code -01 only).
5.1.9Temp error high or temp error low
tEMP HI or tEMP LO
Potential cause
Open or shorted resistance temperature device (RTD).
Recommended action
Replace the sensor.
46Emerson.com/Rosemount
November 2020Quick Start Guide
Potential cause
Temperature out of range.
Recommended action
Check process temperature.
5.2Troubleshooting without advanced diagnostics
The sections below list common problems, causes, and remedies typically
encountered in process measurement.
5.2.1Reading is off scale
Display reads overrange.
Potential cause
Defective preamplifier.
Recommended action
For code -02 sensors, replace preamplifier. For code -01 sensors, replace
sensor.
Potential cause
Temperature element shorted.
Recommended action
Check temperature element and replace sensor if defective.
Potential cause
Sensor not in process. Sample stream is low or air bubbles are present.
Recommended action
Make sure sensor is in process with sufficient sample stream.
Refer to Install for installation details.
Potential cause
Open glass electrode.
Recommended action
Replace the sensor.
Potential cause
Reference element open: no contact.
Quick Start Guide47
Quick Start GuideNovember 2020
Recommended action
Replace the sensor.
5.2.2Display reads between 3 and 6 pH regardless of actual pH of solution
or sample
Potential cause
Electrode cracked.
Recommended action
Replace the sensor.
5.2.3Meter or display swings or jumps widely in AUTO T.C. mode
Potential cause
Temperature element open.
Recommended action
Ohm out temperature element and replace sensor if defective.
5.2.4Span between buffers extremely short in AUTO T.C. mode
Potential cause
Temperature element open.
Recommended action
Ohm out temperature element and replace sensor if defective.
5.2.5Sluggish or slow meter indication for real changes in pH level
Potential cause
Electrode coated.
Recommended actions
1. Clean sensor as recommended in Clean electrode or Clean
platinum electrode.
2. Replace sensor if cracked.
Potential cause
Electrode defective.
Recommended action
Replace the sensor.
48Emerson.com/Rosemount
November 2020Quick Start Guide
5.2.6Transmitter cannot be standardized
Potential cause
Electrode coated.
Recommended actions
1. Clean sensor as recommended in Clean electrode or Clean
platinum electrode.
2. Replace sensor if cracked.
Potential cause
Defective preamplifier.
Recommended action
Replace preamplifier.
5.2.7Transmitter short spans between two different buffer values
Potential cause
Aged glass electrode or high temperature exposure.
Recommended action
Replace the sensor.
Potential cause
Electrode coated.
Recommended actions
1. Clean sensor as recommended in Clean electrode or Clean
platinum electrode.
2. Replace sensor if cracked.
Potential cause
Air bubbles trapped in sensor end between glass bulb and sensor body.
Recommended action
Shake the sensor in solution.
See Install for mounting guidelines.
Quick Start Guide49
Quick Start GuideNovember 2020
6Return of material
For all repair or warranty inquiries, please contact our Customer Care
department at 800-999-9307.
50Emerson.com/Rosemount
November 2020Quick Start Guide
7Rosemount pH/ORP sensor(s) product
certifications
Rev 0.5
7.1European directive information
A copy of the EU Declaration of Conformity can be found at the end of the
Quick Start Guide. The most recent revision of the EU Declaration of
Conformity can be found at Emerson.com/Rosemount.
7.2Ordinary location certification
As standard, the transmitter has been examined and tested to determine
that the design meets the basic electrical, mechanical, and fire protection
requirements by a nationally recognized test laboratory (NRTL) as accredited
by the Federal Occupational Safety and Health Administration (OSHA).
7.3Installing equipment in North America
The US National Electrical Code® (NEC) and the Canadian Electrical Code
(CEC) permit the use of Division marked equipment in Zones and Zone
marked equipment in Divisions. The markings must be suitable for the area
classification, gas, and temperature class. This information is clearly defined
in the respective codes.
7.4
USA
7.4.1FM Intrinsic Safety
Certificate
Standards
Markings
Specific Conditions for Safe Use (X):
1. Sensors with Model 1700702 preamplifier:
Quick Start Guide51
FM17US0198X
FM Class 3600:1998, FM Class 3610:2010, FM Class 3611:
2004, FM Class 3810: 2005
IS/I,II,III/1/ABCDEFG/T6 Ta = –20 °C to 60 °C
I/0/AEx ia IIC/T6 Ta = –20 °C to 60 °C
NI/I/2/ABCD/T6 Ta = –20 °C to 60 °C
S/II,III/2/EFG/T6 Ta = –20 °C to 60 °C
a. Model 385+-a-b-c. Triple junction pH/ORP sensor
b. Model 389-a-b-c-d-e. pH/ORP sensor
c. Model 389VP-a-b-c-d. pH/ORP sensor
d. Model 396VP-a-b-c-d. Submersion/insertion pH/ORP sensor
Quick Start GuideNovember 2020
e. Model 396P-a-b-c-d-e. Submersion/insertion pH/ORP sensor
f. Model 396PVP-a-b-c-d-e. Submersion/insertion pH/ORP
sensor
g. Model 396RVP-a-b-c-d-e. Retraction/submersion/insertion
pH/ORP sensor
h. Model 398RVP-a-b-c-d-e-f. pH/ORP sensor
i. Model 3200HP-00. High purity water pH sensor
j. Model 3300HTVP-a-b-c-d. High performance pH and ORP
sensor
k. Model 3400HTVP-a-b-c-d-e. High performance pH and ORP
sensor
l. 3500P-a-b-c-d-e-f. High performance pH and ORP sensor
m. 3500VP-a-b-c-d-e-f. High performance pH and ORP sensor
n. Model 3900-a-b-c. General purpose pH/ORP sensor
o. Model 3900VP-a-b. General purpose pH/ORP sensor
The polymeric surface of all the apparatus listed above may store
electrostatic charge and become a source of ignition. Clean surface
should only be done with a damp cloth.
2. Sensors without Model 1700702 preamplifier (simple apparatus):
a. Model 385-a-b-c-d-e. Retractable pH/ORP sensor
b. Model 385+-a-b-c Triple junction pH/ORP sensor
c. Model 389-a-b-c-d-e. pH/ORP sensor
d. Model 389VP-a-b-c. pH/ORP sensor
e. Model 396-a-b-c. Submersion/insertion pH sensor
f. Model 396VP-a-b. Submersion/insertion pH sensor
g. Model 396P-a-b-c-d-e. Submersion/insertion pH/ORP sensor
h. Model 396PVP-a-b-c-d. Submersion/insertion pH/ORP sensor
i. Model 396R-a-b-c-d-e. Retraction/submersion/insertion
pH/ORP sensor
j. Model 396RVP-a-b-c-d. Retraction/submersion/insertion
pH/ORP sensor
k. Model 397-a-b-c-d-e. pH sensor
l. Model 398-a-b-c-d-e. pH/ORP sensor
52Emerson.com/Rosemount
November 2020Quick Start Guide
m. Model 398VP-a-b-c. pH/ORP sensor
n. Model 398R-a-b-c-d-e-f. pH/ORP sensor
o. Model 398RVP-a-b-c-d-e-f. pH/ORP sensor
p. Model 3200HP-00. High purity water pH sensor
q. Model 3300HT-a-b-c-d. High performance pH and ORP sensor
r. Model 3300HTVP-a-b-c-d. High performance pH and ORP
sensor
s. Model 3400HT-a-b-c-d-e-f. High performance pH and ORP
sensor
t. Model 3400HTVP-a-b-c-d-e-f. High performance pH and ORP
sensor
u. Model 3500P-a-b-c-d-e-f. High performance pH and ORP
sensor
v. Model 3500VP-a-b-c-d-e-f. High performance pH and ORP
sensor
w. Model 3800-a. Autoclaveable and steam sterilizable pH
sensors
x. Model 3800VP-a. Autoclaveable and steam sterilizable pH
sensors
y. Model 3900-a-b-c. General purpose pH/ORP sensor
z. Model 3900VP-a-b. General purpose pH/ORP sensor
The polymeric surface of all the apparatus listed above may store
electrostatic charge and become a source of ignition. Clean surface
should only be done with a damp cloth.
7.4.2CSA Intrinsic Safety
Certificate
Standards
Markings
Quick Start Guide53
70164066
C22.2 No 0-10, C22.2 No 0.4-M2004, C22.2 No 94-M1991,
C22.2 No 142 – M1987, C22.2 No 157-M1992, CAN/CSA
E60079-0:07, CAN/CSA E60079-11:02, UL 50-11th Ed, UL
508-17th Ed, UL 913-7th Ed, UL 60079-0: 2005, UL 60079-11:
2002
Preamplifier assembly:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups
EFG; Class III; Class I, Division 2, Groups ABCD; ambient
temperature rating –20 °C to +60 °C; Ex ia IIC; T6: Class I, Zone
0, AEx ia IIC ; T6
Quick Start GuideNovember 2020
Sensor apparatus with preamplifier:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups
EFG; Class III; Class I, Division 2, Groups ABCD; ambient
temperature rating –20 °C to +60 °C; Ex ia IIC; T6: Class I, Zone
0, AEx ia IIC ; T6
Sensor apparatus:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups
EFG; Class III; Class I, Division 2, Groups ABCD; Ex ia IIC; T6;
ambient temperature rating –20 °C to +60 °C: (simple
apparatus)
7.5Canada
7.5.1CSA Intrinsic Safety
Certificate
Standards
Markings
70164066
C22.2 No 0-10, C22.2 No 0.4-M2004, C22.2 No 94-M1991,
C22.2 No 142 – M1987, C22.2 No 157-M1992, CAN/CSA
E60079-0:07, CAN/CSA E60079-11:02, UL 50-11th Ed, UL
508-17th Ed, UL 913-7th Ed, UL 60079-0: 2005, UL 60079-11:
2002
Preamplifier assembly:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups
EFG; Class III; Class I, Division 2, Groups ABCD; ambient
temperature rating –20 °C to +60 °C; Ex ia IIC; T6: Class I, Zone
0, AEx ia IIC ; T6
Sensor apparatus with preamplifier:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups
EFG; Class III; Class I, Division 2, Groups ABCD; ambient
temperature rating –20 °C to +60 °C; Ex ia IIC; T6: Class I, Zone
0, AEx ia IIC ; T6
Sensor apparatus:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups
EFG; Class III; Class I, Division 2, Groups ABCD; Ex ia IIC; T6;
ambient temperature rating –20 °C to +60 °C: (simple
apparatus)
7.6Europe
7.6.1ATEX Intrinsic Safety
Certificate
Standards
Markings
54Emerson.com/Rosemount
Baseefa10ATEX0156
EN 60079-0: 2012+A11: 2013, EN 60079-11: 2012
pH/ORP sensors with no preamplifier fitted
II 1 G Ex ia IIC T4 Ga (–20 °C to +60 °C)
November 2020Quick Start Guide
pH sensors with integral smart preamplifier fitted
II 1 G Ex ia IIC T4 Ga (–20 °C to +60 °C)
ORP sensors with integral standard preamplifier fitted
II 1 G Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C)
pH sensors with integral standard preamplifier fitted
II 1 G Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C)
Specific Conditions for Safe Use (X):
1. All pH/ORP sensor models with a plastic enclosure or exposed plastic
parts may provide an electrostatic ignition hazard and must only be
cleaned with a damp cloth to avoid the danger of ignition due to
build-up of electrostatic charge.
2. All pH/ORP sensor models with a metallic enclosure may provide a
risk of ignition by impact or friction. Care should be taken during
installation to protect the sensor from the risk.
3. External connections to the sensor must be suitably terminated and
provide a degree of protection of at least IP20.
4. All pH/ORP sensor models are intended to be in contact with the
process fluid and may not meet the 500V r.m.s. test to earth. This
must be taken into consideration at installation.
7.7
International
7.7.1IECEx Intrinsic Safety
Certificate
Standards
Markings
Quick Start Guide55
IECEx BAS 10.0083X
IEC 60079-0: 2011, IEC 60079-11: 2011
pH/ORP sensors with no preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +60 °C)
pH sensors with integral smart preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +60 °C)
ORP sensors with integral standard preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C)
pH sensors with integral standard preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C)
Quick Start GuideNovember 2020
Specific Conditions for Safe Use (X):
1. All pH/ORP sensor models with a plastic enclosure or exposed plastic
parts may provide an electrostatic ignition hazard and must only be
cleaned with a damp cloth to avoid the danger of ignition due to
build-up of electrostatic charge.
2. All pH/ORP sensor models with a metallic enclosure may provide a
risk of ignition by impact or friction. Care should be taken during
installation to protect the sensor from the risk.
3. External connections to the sensor must be suitably terminated and
provide a degree of protection of at least IP20.
4. All pH/ORP sensor models are intended to be in contact with the
process fluid and may not meet the 500V r.m.s. test to earth. This
must be taken into consideration at installation.
7.8China
7.8.1Nepsi Intrinsic Safety
7.9
Certificate
Standards
Markings
Specific Conditions for Safe Use (X):
1. It is strictly forbidden to rub the plastic shell parts of the product to
2. When the product shell contains light metals, it should be prevented
Technical Regulations Customs Union (EAC)
GYB19.1035X
GB 3836.1-2010, GB 3836.4-2010, GB 3836.20-2010
Ex ia II C T4 Ga (–20 °C to +60 °C)
prevent the risk of static ignition.
in a zone 0 environment.
7.9.1EAC Intrinsic Safety
Certificate
Markings
TC RU C-US .MIO62. B.06011
pH/ORP sensors with no preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +60 °C)
pH sensors with integral smart preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +60 °C)
ORP sensors with integral standard preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C)
pH sensors with integral standard preamplifier fitted
The Emerson logo is a trademark and service
mark of Emerson Electric Co. Rosemount is a
mark of one of the Emerson family of companies.
All other marks are the property of their
respective owners.