Emerson designs, manufactures, and tests its products to meet many national and international
standards. Because these instruments are sophisticated technical products, you must properly install,
use, and maintain them to ensure they continue to operate within their normal specifications. You
must adhere to the following instructions and integrate them into your safety program when
installing, using, and maintaining Emerson's Rosemount™ products. Failure to follow the proper
instructions may cause any one of the following situations to occur: loss of life, personal injury,
property damage, damage to this instrument, and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If this Quick Start Guide is not the correct one, call 1-800-854-8257 or 949-757-8500 to request
the correct Quick Start Guide. Save this Quick Start Guide for future reference.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the
product.
• Install equipment as specified in the installation instructions of the appropriate Quick Start Guide
and per applicable local and national codes. Connect all products to the proper electrical and
pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and
maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts
specified by Emerson. Unauthorized parts and procedures can affect the product's performance,
place the safe operation of your process at risk, and may result in fire, electrical hazards, or
improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified people, to prevent electrical shock and personal
injury.
Note
The information contained in this document is subject to change without notice.
WARNING
Hazardous area installation
Installations near flammable liquids or in hazardous areas must be carefully evaluated by on site
safety personnel.
To secure and maintain intrinsically safe installation, an appropriate transmitter/safety barrier/
sensor combination must be used. The installation system must be in accordance with the
governing approval agency (FM, CSA, or BASEEFA/CENELEC) hazardous area classification
requirements. Consult your transmitter Reference Manual for details.
Proper installation, operation, and servicing of this sensor in a hazardous area are entirely the
operator's responsibility.
2Emerson.com/Rosemount
November 2020Quick Start Guide
WARNING
Pressure and temperature
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig
and the process temperature is lowered to a safe level.
Do not insert or retract retractable sensors when process pressures are in excess of 64 psig (542
kPa) for option 21 or 35 psig (343 kPa) for option 25.
WARNING
Corrosive substance
The solution used during calibration is an acid; handle it with care.
Follow the directions of the acid manufacturer.
Wear the proper protective equipment.
Do not let the solution come into contact with skin or clothing.
If contact with skin is made, immediately rinse with clean water.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end
users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your
system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for
all systems used within the facility.
CAUTION
Special conditions for safe use
All pH/ORP sensors have a plastic enclosure which must be cleaned with a damp cloth to avoid the
danger due to a buildup of electrostatic discharge.
CAUTION
Sensor/process application compatibility
The wetted sensor materials may not be compatible with process composition and operating
conditions.
Application compatibility is entirely the operator's responsibility.
A. Temperature comp solution ground
B. Reference junction
C. pH electrode
D. ¾-in. male national pipe thread (MNPT)
E. 1-in. MNPT
F. Wrench flats 1.30 in. across
6Emerson.com/Rosemount
November 2020Quick Start Guide
Figure 1-3: Mount
A. 1½-in. x 1 in. reducing bushing
B. 1½-in. pipe tee PN 2002011
Straight flow shown
Note
Install sensor at least ten degrees above the horizon.
Table 1-1: Horizontal Pipe Tee (PN 2002011) Pressure/Temperature
Ratings
psig (kPa)°F (°C)
150 (1136)150 (65)
128 (984)160 (71)
102 (805)170 (77)
80 (653)180 (82)
57 (494)200 (93)
48 (432)210 (99)
Quick Start Guide7
Quick Start GuideNovember 2020
Figure 1-4: Low Flow Cell PN 24091-00/24091-02
Inlet and outlet connections are stainless steel and take ¼-in. OD tubing.
Flow cell is polycarbonate with ¼-in. female national pipe thread (FNPT)
fittings.
Wetted materials
Body and nut: polyester/polycarbonate
Fittings: 316 stainless steel
Seals: silicone
Flow cell ratings
Temperature: 32 to 158 °F (0 to 70 °C)
Max. pressure: 90 pisg (721 kPa)
A. Cleaning solution by others
B. Solenoid valve or manual valve (supplied by others)
C. Corrosion resistant tubing (supplied by others)
D. Polypropylene ¼-in. compression fitting
E. ¼-in. 316 stainless steel
F. ¼-in. polypropylene
G. Stainless set screw for adjustable spray nozzle height
H. 2-in. NPT threads
I. Sensor
J. 1-in. PVC coupling for submersible applications (supplied by others)
K. 1-in. PVC or stainless conduit (supplied by others)
L. Cable
M. Timer supplied by others or use timer feature in Rosemount instrument
Note
Sensor must be installed at least 10 degrees above the horizon.
10Emerson.com/Rosemount
November 2020Quick Start Guide
The jet spray cleaner eliminates routine, manual sensor maintenance by
cleaning the sensor with water or compressed air. Use a solenoid valve to
control flow through the cleaner.
Note
You can alos use the jet spray cleaner with the handrail mounting assembly
(PN 11275-01, not shown) or mount it through the conduit as shown in
Figure 1-6.
Figure 1-6: Jet Spray Cleaner with pH Sensor
Quick Start Guide11