North America/Latin America+1.713.467.6000+1.713.827.4805
Flow Lifecycle Services for Daniel products+1.713.827.6314+1.713.827.6312
USA (toll free)+1.888.356.9001+1.713.827.3380
Asia Pacific (Republic of Singapore)+65.6777.8211+65.6777.0947.0743
Europe (Stirling Scotland, UK)+44 (0)1786.433400+44 (0)1786.433401
Middle East Africa (Dubai, UAE)+971 4 8118100+971 4 8865465
Daniel Measurement and Control, Inc. (Headquarters)
11100 Brittmoore Park Drive
Houston, TX 77041 USA
http://www.emerson.com
Email
•Customer Service: Daniel.SystemSales@Emerson.com
•Customer Support: Daniel.SystemSales@Emerson.com
•Asia-Pacific: danielap.support@emerson.com
•Europe: danielEMA.cst@emerson.com
Return Material Authorization (RMA)
A Return Material Authorization (RMA) number must be obtained prior to returning any equipment for any reason. Download the
RMA form from the Support Services web page by selecting the link below.
http://www.daniel.com/rma
Signal words and symbols
Pay special attention to the following signal words, safety alert symbols and statements:
Safety alert symbol
This is a safety alert symbol. It is used to alert you to potential physical injury hazards. Obey all safety messages that follow this symbol
to avoid possible injury or death.
DANGER!
Danger indicates a hazardous situation which, if not avoided, will result in death or serious injury.
WARNING!
Warning indicates a hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION!
Caution indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
NOTICE
Notice is used to address safety messages or practices not related to personal injury.
Important
Important is a statement the user needs to know and consider.
Tip
Tip provides information or suggestions for improved efficiency or best results.
Note
Note is “general by-the-way” content not essential to the main flow of information.
Important safety instructions
Daniel Measurement and Control, Inc. (Daniel) designs, manufactures and tests products to function within specific conditions.
Because these products are sophisticated technical instruments, it is important that the owner and operation personnel must
strictly adhere both to the information printed on the product and to all instructions provided in this manual prior to installation,
operation, and maintenance.
Daniel also urges you to integrate this manual into your training and safety program.
BE SURE ALL PERSONNEL READ AND FOLLOW THE INSTRUCTIONS IN THIS MANUAL AND ALL NOTICES AND PRODUCT WARNINGS.
WARNING!
Failure to follow the installation, operation or maintenance instructions for a Daniel product could lead to serious injury or death
from explosion or exposure to dangerous substances.
To reduce the risk:
•Comply with all information on the product, in this manual, and in any local and national codes that apply to this product.
•Do not allow untrained personnel to work with this product.
•Use Daniel parts and work procedures specified in this manual.
Product owners (Purchasers):
•Use the correct product for the environment and pressures present. See technical data or product specifications for
limitations. If you are unsure, discuss your needs with your Daniel representative.
•Inform and train all personnel in the proper installation, operation, and maintenance of this product.
•To ensure safe and proper performance, only informed and trained personnel should install, operate, repair and maintain
this product.
•Verify that this is the correct instruction manual for your Daniel product. If this is not the correct documentation, contact
Daniel at 1-713-827-6314. You may also download the correct manual from:
•Save this instruction manual for future reference.
•If you resell or transfer this product, it is your responsibility to forward this instruction manual along with the product to the
new owner or transferee.
•ALWAYS READ AND FOLLOW THE INSTALLATION, OPERATIONS, MAINTENANCE AND TROUBLESHOOTING MANUAL(S) AND
ALL PRODUCT WARNINGS AND INSTRUCTIONS.
•Do not use this equipment for any purpose other than its intended service. This may result in property damage and/or
serious personal injury or death.
Product operation (Personnel):
•To prevent personal injury, personnel must follow all instructions of this manual prior to and during operation of the
product.
•Follow all warnings, cautions, and notices marked on, and supplied with, this product.
•Verify that this is the correct instruction manual for your Daniel product. If this is not the correct documentation, contact
Daniel at 1-713-827-6314. You may also download the correct manual from: http://www.daniel.com/.
•Read and understand all instructions and operating procedures for this product.
•If you do not understand an instruction, or do not feel comfortable following the instructions, contact your Daniel
representative for clarification or assistance.
•Install this product as specified in the INSTALLATION section of this manual per applicable local and national codes.
•Follow all instructions during the installation, operation, and maintenance of this product.
•Ensure that all connections to pressure and electrical sources are secure prior to and during equipment operation.
•Use only replacement parts specified by Daniel. Unauthorized parts and procedures can affect this product's performance,
safety, and invalidate the warranty. “Look-a-like” substitutions may result in deadly fire, explosion, release of toxic
substances or improper operation.
•Save this instruction manual for future reference.
Notice
THE CONTENTS OF THIS PUBLICATION ARE PRESENTED FOR INFORMATIONAL PURPOSES ONLY, AND WHILE EVERY EFFORT HAS
BEEN MADE TO ENSURE THEIR ACCURACY, THEY ARE NOT TO BE CONSTRUED AS WARRANTIES OR GUARANTEES, EXPRESSED OR
IMPLIED, REGARDING THE PRODUCTS OR SERVICES DESCRIBED HEREIN OR THEIR USE OR APPLICABILITY. ALL SALES ARE GOVERNED
BY DANIEL'S TERMS AND CONDITIONS, WHICH ARE AVAILABLE UPON REQUEST. WE RESERVE THE RIGHT TO MODIFY OR IMPROVE
THE DESIGNS OR SPECIFICATIONS OF SUCH PRODUCTS AT ANY TIME.
DANIEL DOES NOT ASSUME RESPONSIBILITY FOR THE SELECTION, USE OR MAINTENANCE OF ANY PRODUCT. RESPONSIBILITY FOR
PROPER SELECTION, USE AND MAINTENANCE OF ANY DANIEL PRODUCT REMAINS SOLELY WITH THE PURCHASER AND END-USER.
TO THE BEST OF DANIEL'S KNOWLEDGE THE INFORMATION HEREIN IS COMPLETE AND ACCURATE. DANIEL MAKES NO
WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE WITH RESPECT TO THIS MANUAL AND, IN NO EVENT, SHALL DANIEL BE LIABLE FOR ANY INCIDENTAL,
PUNITIVE, SPECIAL OR CONSEQUENTIAL DAMAGES INCLUDING, BUT NOT LIMITED TO, LOSS OF PRODUCTION, LOSS OF PROFITS,
LOSS OF REVENUE OR USE AND COSTS INCURRED INCLUDING WITHOUT LIMITATION FOR CAPITAL, FUEL AND POWER, AND CLAIMS
OF THIRD PARTIES.
PRODUCT NAMES USED HEREIN ARE FOR MANUFACTURER OR SUPPLIER IDENTIFICATION ONLY AND MAY BE TRADEMARKS/
REGISTERED TRADEMARKS OF THESE COMPANIES.
Warranty and Limitations
1. LIMITED WARRANTY: Subject to the limitations contained in Section 2 herein, Daniel Measurement & Control, Inc. (“Daniel”)
warrants that the licensed firmware embodied in the Goods will execute the programming instructions provided by Daniel, and that
the Goods manufactured by Daniel will be free from defects in materials or workmanship under normal use and care and Services
will be performed by trained personnel using proper equipment and instrumentation for the particular Service provided. The
foregoing warranties will apply until the expiration of the applicable warranty period. Goods are warranted for twelve (12) months
from the date of initial installation or eighteen (18) months from the date of shipment by Daniel, whichever period expires first.
Consumables and Services are warranted for a period of 90 days from the date of shipment or completion of the Services. Products
purchased by Daniel from a third party for resale to Buyer (“Resale Products”) shall carry only the warranty extended by the original
manufacturer. Buyer agrees that Daniel has no liability for Resale Products beyond making a reasonable commercial effort to
arrange for procurement and shipping of the Resale Products. If Buyer discovers any warranty defects and notifies Daniel thereof in
writing during the applicable warranty period, Daniel shall, at its option, correct any errors that are found by Daniel in the firmware
or Services or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by Daniel to be defective, or
refund the purchase price of the defective portion of the Goods/Services. All replacements or repairs necessitated by inadequate
maintenance, normal wear and usage, unsuitable power sources or environmental conditions, accident, misuse, improper
installation, modification, repair, use of unauthorized replacement parts, storage or handling, or any other cause not the fault of
Daniel are not covered by this limited warranty, and shall be at Buyer's expense. Daniel shall not be obligated to pay any costs or
charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by Daniel. All costs of dismantling,
reinstallation and freight and the time and expenses of Daniel's personnel and representatives for site travel and diagnosis under
this warranty clause shall be borne by Buyer unless accepted in writing by Daniel. Goods repaired and parts replaced by Daniel
during the warranty period shall be in warranty for the remainder of the original warranty period or ninety (90) days, whichever is
longer. This limited warranty is the only warranty made by Daniel and can be amended only in a writing signed by Daniel. THE
WARRANTIES AND REMEDIES SET FORTH ABOVE ARE EXCLUSIVE. THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY
KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE OR ANY OTHER MATTER WITH RESPECT
TO ANY OF THE GOODS OR SERVICES. Buyer acknowledges and agrees that corrosion or erosion of materials is not covered by this
warranty.
2. LIMITATION OF REMEDY AND LIABILITY: Daniel shall not be liable for damages caused by delay in performance. The remedies of
Buyer set forth in this agreement are exclusive. In no event, regardless of the form of the claim or cause of action (whether based in
contract, infringement, negligence, strict liability, other tort or otherwise), shall Daniel's liability to Buyer and/or its customers
exceed the price to Buyer of the specific goods manufactured or services provided by Daniel giving rise to the claim or cause of
action. Buyer agrees that in no event shall Daniel's liability to Buyer and/or its customers extend to include incidental, consequential
or punitive damages. The term “consequential damages” shall include, but not be limited to, loss of anticipated profits, revenue or
use and costs incurred including without limitation for capital, fuel and power, and claims of Buyer's customers.
The meter body and piping may be extremely hot or cold.
Wear appropriate personal protective equipment when coming in contact with the meter.
Failure to comply may result in injury.
CAUTION!
TRANSPORTATION HAZARD
When moving the meter, do not insert the forks of a forklift into the bore.
Inserting the forks may cause the meter to become unstable, resulting in injury or damage to
the bore and sealing face.
CAUTION!
TRIPPING HAZARD
Clear all obstacles or obstructions from the work area when transporting, installing or removing the
meter.
Failure to comply may cause injury to personnel.
Maintenance and Troubleshooting manual1
Maintenance
WARNING!
CRUSHING HAZARD
Do not remove flange stabilizers.
Attempting to do so may allow the meter to roll, resulting in serious injury or equipment
damage.
1.2
A. Flange stabilizers
CAUTION!
ESCAPING GASES OR FLUIDS HAZARD
The purchaser of the meter is responsible for the selection of Daniel components/seals and materials
compatible with the chemical properties of gas flow measurement.
Failure to select the suitable meter component/seals may cause escaping gases or liquids,
resulting in injury or equipment damage.
Field hydrostatic pressure testing procedures
Daniel 3410 Series Ultrasonic Gas Flow Meters are supplied with T-Slot transducer
assemblies which are extractable while the line is pressurized.
CAUTION!
LEAKAGE OR PRESSURE CONTAINING PARTS FAILURE
Use precautions to eliminate hazards to personnel in the event of leakage or failure of the liquid
ultrasonic meter pressure containing parts or failure of the test equipment and to prevent overpressurization during the test procedure.
Failure to comply may result in injury to personnel or cause damage to the equipment.
2Gas Ultrasonic Flow Meters
1.2.1T-Slot Transducer assembly and mount
T-slot transducer mount and holderFigure 1-1:
A. Transducer mount
Maintenance
B. Captive screws
C. T-slot transducer holder
D. Cable nut
E. Chordset
Procedure
1.Slowly vent all line pressure on the 3410 Series Gas Ultrasonic Meter to atmosphere.
2.Disconnect transducer cable from the transducer holder.
3.If installed, remove the mount cover by loosening the two mount cover captive
screws.
4.Loosen the T-Slot transducer assembly with a 1 1/4 inch (32 mm) socket. Carefully
remove the T-Slot transducer assembly.
5.Place a label on the transducer assembly to marks its location (i.e., Model 3414-A1,
A2, B1, B2, C1, C2, D1, or D2; 3412-A1, A2, B1, B2; or 3411-A). Port locations are
marked on the transducer cable as well as on cast meter housings.
6.Inspect the transducer mount threads and ensure they are clean and free of debris.
7.Apply a small amount of Nickel antiseize compound (P/N 3-9960-134) to the
threads of the Hydrotest plug (P/N 1-360-01-212) from kit (P/N 1-360-01-220) and
install it into the mount. JuniorSonic kit part numbers are listed below.
• JuniorSonic™ SP Field Hydrotest Kit (P/N 1-360-01-221)
• JuniorSonic™ DP Field Hydrotest Kit (P/N 1-360-01-222)
Maintenance and Troubleshooting manual3
Maintenance
8.Repeat Step 1 through Step 6 for the other transducer(s) being careful to note the
location of each transducer in the meter assembly.
9.Run the field hydrostatic test.
10.Reverse the steps above to reinstall the transducers into their appropriate ports.
Before reinstalling the transducer assemblies, ensure the transducer ports, mounts,
and transducer holders are clean and free of debris. Apply a small amount of Nickel
antiseize compound to the outer threads of the transducer holders before installing
them into the mounts.
1.3Routine maintenance
Routine maintenance operations requires adherence to all applicable regulations and laws
and safety training for personnel to perform the maintenance operations. Review your
organization’s best practices procedures before performing routine maintenance.
1.3.1Maintenance logs and reports
To monitor the performance health of the meter, and ensure it is operating within
acceptable specifications, routine diagnostics should be performed. Collecting a
maintenance log gives you a snapshot of the current health of the meter and you can
compare the inspection reports from previously saved logs. Use the Logs/Reports menu
and click Maintenance Logs and Reports. Daniel MeterLink displays the Maintenance Logs
and Reports dialog box. Choose the time duration, log format and collection rate for the
output file and click the Start button. You can open the file immediately after it is
generated or view it at a later time. It is recommended that a Maintenance log be collected
after an upset in the system.
In establishing a baseline to be used for the trending of the meter diagnostics, it is very
helpful if a set of log files are collected immediately after the meter has been installed in
the field. Preferably, collect the log files at several velocities within the operating range of
the meter. This helps establish that the flow profile is relatively constant throughout the
meters operating range (except velocities below 3 ft/sec where the profile may vary).
4Gas Ultrasonic Flow Meters
Maintenance log collection
Maintenance log collection parametersFigure 1-2:
Maintenance
Trend maintenance log collection
Merging the results of two or more Maintenance logs into a single file, allows you to build a
historical database of the meter’s performance. Trending the logs indicates changes from
the original installation of the meter, or over time. Looking at a single inspection report,
that is either collected monthly or quarterly, can give you an indication of the meter's
health.
Maintenance and Troubleshooting manual5
Maintenance
Trend log collectionFigure 1-3:
This is important since many diagnostics change slowly over time. Trending the
maintenance logs helps identify these changes and makes problems much more obvious
than merely viewing a single inspection report. The Trending feature is integral to Daniel
MeterLink which allows all important parameters to be trended. Daniel MeterLink supports
trending files in a Microsoft® Excel® workbook from multiple 3410 Series meter
maintenance logs. Some parameters like gain, signal level, and noise level level may show
a shift over time which can be useful in detecting changes in the meter and the
installation.
Maintenance logs or Trend files to be trended must all have matching column headings.
This means the logs must be in the same units (i.e. U.S. Customary or Metric), must have
the same pressure type (i.e. gage or absolute), and must have the same time base (1/
second, 1/minute, 1/hour, 1/day). If not, an error message will be displayed stating the
column headings do not match and the file will not be added to the Workbook to trend
list.
Archive log collection
Archive logs may be collected and the options include:
•Daily log - generated every 24 hours on the Contract Hour.
•Hourly log - generated every hour at the top of the hour.
•Event log - collects the alarm and event log records.
6Gas Ultrasonic Flow Meters
Maintenance
Archive log collection parametersFigure 1-4:
The logs may be collected in a single file or you can choose to collect one type of log. Each
of the Meter Archive logs include the Meter Configuration file.
Maintenance and Troubleshooting manual7
Maintenance
1.3.2Pipeline cleaning maintenance
WARNING!
BURST HAZARD
Before pipeline cleaning and maintenance ("pigging operations"), remove straightening vanes
or flow conditioners. Failure to do so may cause excessive pressure in the meter system,
resulting in serious injury/death or equipment damage.
Figure 1-5:
Figure 1-6:
3410 Series Gas Ultrasonic Flow Meter with flow conditioner for unidirectional flow
3410 Series Gas Ultrasonic Flow Meter with flow conditioner for bidirectional flow
Straightening vanes or flow profilers must be removed during pipeline cleaning
maintenance operations (“pigging operation”). If the meter run is pigged with a flow
conditioner in line, pressure may build up and cause the pipes and flanges to burst and
severely injure personnel. The excessive pressure may damage the meter or the transducer
ports may collect debris which may impede data acquisition and flow measurement.
1.3.3
8Gas Ultrasonic Flow Meters
Visual inspection
Periodically inspect meter and meter run for signs of components loosening or seals
leaking. This includes:
Maintenance
Procedure
1.Fluids leaking from seals. This could be visually noticed for liquids leaking. It may be
audible for gasses leaking. Ice may also form at a point of a gas leak.
2.Movement of components that should be rigid.
3.Excessive noise due to vibration could be sign of a loose component.
Inspection should be more frequent in systems with a large amount of vibration.
Maintenance and Troubleshooting manual9
Maintenance
10Gas Ultrasonic Flow Meters
2Troubleshooting
Topics covered in this chapter:
Meter status alarms
•
Troubleshooting the meter
•
Troubleshoot maintenance log files and trend files
•
2.1Meter status alarms
Run Daniel MeterLink and open the Meter Monitor (Summary) view to perform a
diagnostics health check.
Meter Monitor status alarmsFigure 2-1:
Troubleshooting
If the meter is measuring flow and operating within the calibration parameters the Meter
Status LED is green. If the Meter Status LED is red, an active alarm exists that requires you
to take corrective action. Click the Check Status button to display the Status Summary
screen. The alarms are shown with the primary causes listed first. Click the question mark
next to the alarm to display a help topic related to the alarm and recommended actions
to resolve the issue.
Maintenance and Troubleshooting manual11
Troubleshooting
Status summaryFigure 2-2:
2.1.1Check status
Click the Check Status button if any of the LEDs are yellow or red to see more specific
information causing the status alarm. Some alarms do not require an acknowledge and will
clear automatically when the alarm condition goes away. Alarms that require a user to
acknowledge them will have a button to the right titled ACK. Clicking the ACK button
changes the button text to Wait and sends a request to the meter to clear the alarm. The
alarm will disappear from the Check Status dialog once the alarm actually clears.
Click the Check Status button and Daniel MeterLink opens the Status Summary dialog box
that gives a short description of all alarms present.
Status SummaryFigure 2-3:
A. Active alarm conditions from Meter Monitor page
B. Status summary page with alarm examples
12Gas Ultrasonic Flow Meters
Following is a list and a brief description of the types of alarms:
•System
•Power Loss
•Field I/O
•Validity
•Comms
•Check Status
2.1.2System alarm
The System alarm indicates a failure in the hardware that should be addressed by a service
technician. This includes memory checksum errors and communication errors within the
hardware. A Red LED indicates a System alarm condition. Collect a Maintenance log and an
audit/alarm log and then, contact your Daniel service representative. This could be an
alarm condition that occurred and remains latched until the condition is resolved and the
alarm is cleared by clicking the ACK button on the Monitor|Check Status|Status
Summary page.
Troubleshooting
2.1.3Chord A, Chord B, Chord C and Chord D alarm
Chord A, Chord B, Chord C, and Chord D - These alarms indicate how a chord is
functioning.
LED Color
GreenNo alarms are present. Chord is operating properly.
YellowAt least one sample in the batch caused an alarm but it did not cause
RedThe chord has failed or is in acquisition. This chord is not used for this
GrayThe chord has manually been set to inactive.
Problem
the chord to fail. The sample will not be used in the batch. Discarding
occasional samples can occur during normal operation such as during
flow velocity changes.
batch. Chords that have failed or are shown to be in acquisition for repeated batches indicates that the meter should be inspected by a
service technician.
2.1.4Field I/O alarm
Reports various field I/O devices that are in alarm. Click the Check Status button for more
details on specific alarms. The field is grayed out if the Daniel 3410 Series Ultrasonic Gas
Flow Meter does not support this alarm.
Maintenance and Troubleshooting manual13
Troubleshooting
2.1.5Validity alarm
This alarm indicates that the meter may not be measuring accurately. Click Check Status
to see a description of which validity alarms are active. The validity alarms QMeter and
QFlow indicate an issue with the meter collecting enough information from the chords to
make an accurate measurement. The validity alarms for pressure and temperature indicate
that the value is above or below the alarm limits for these values. Red and green are the
only colors used for this alarm.
2.1.6Comms alarm
The Comms alarm indicates that communications between Daniel MeterLink and the
meter failed. This could be due to a poor communication link. Daniel MeterLink continues
to retry communications. Red and green are the only colors used for this alarm.
2.1.7Communications
The Communications Analyzer (via Daniel MeterLink Tools> Menu>CommunicationsAnalyzer menu path) displays communications between Daniel MeterLink and the
ultrasonic meter. This utility is useful for troubleshooting communications to the meter. It
displays many of the TCP/IP commands between Daniel MeterLink and the connected
meter.
2.2
For troubleshooting communications with the 475 Field Communicator for the HART
Protocol, refer to Section 5 of the Emerson 475 Field Communicator User’s Manual, Rev D.
This manual may be downloaded from the following location:
Table 2-1 and the following sections show errors that may occur with the meter hardware,
firmware or connections and recommended actions to resolve the problem(s).
14Gas Ultrasonic Flow Meters
Troubleshooting
TroubleshootingTable 2-1:
ErrorRecommended action(s)
Acquisition Module Error• Check interconnect cable between Acquisi-
tion Module and the CPU Module.
• Attempt the Program Download procedure
to install the firmware.
-Cycle power to the meter.
-Replace the Acquisition Module.
-If the Acquisition Module cannot be re-
programmed, collect a complete Archive log and contact your local area
Daniel service representative.
Acquisition Module is not compatible with firmware
Chord failure• Chord is hard failed (Model 3414 - Chord A,
• Upgrade the firmware in the meter to the
latest version using Daniel MeterLink. Contact your Daniel service representative to
obtain the latest firmware.
• Replace the Acquisition Module.
Chord B, Chord C or Chord D; Model 3412 Chord A or Chord B; or 3411 Chord A) and
meter is unable to obtain measurement data
from this pair of transducers.
-If Chord A is failed and no other trans-
ducers are failed or are reporting status
alerts, the issue is most likely isolated to
this pair of transducers or its cabling.
Check the transducer wiring for this pair
of transducers to make sure connections
are secure and wired correctly.
-Remove the transducer cable from the
transducer and measure the resistance
with an Ohm meter across the two pins
on the back of the transducer holder. If
the value is over 2 ohms, replace the
transducers.
-If transducer cabling allows, swap ca-
bling of failed transducer pair with a pair
with equal path lengths. If the alarm remains active for this chord, then the
transducers are working properly. If this
alarm clears but the chord that was
swapped now fails, the issue is with the
transducer.
-Collect a Maintenance Log, Configura-
tion file and Waveform stream file with
Daniel MeterLink and contact your Daniel service representative.
Maintenance and Troubleshooting manual15
Troubleshooting
Troubleshooting (continued)Table 2-1:
ErrorRecommended action(s)
CPU Module LINK LED• When connecting directly:
-Check the cross-over cable connection
(P/N 2-3-3400-079)
• When Using a Hub:
-Use straight-through patch cable be-
tween the meter and the hub and a
straight-through patch cable between
the hub and the PC.
-Do not connect either the Daniel 3410
Ultrasonic Flow Meter or PC to the hub
UPLINK port.
-Check the CPU Module LED 1 is on (ei-
ther solid red or flashing green). If the
LED is not on, check power to the meter.
-If the LED is on, check the Ethernet cable
connections
Communication line connected to the flow
computer but no signal is received
• Check for loose connections at the flow me-
ter and the flow computer.
• Check the CPU Module settings.
16Gas Ultrasonic Flow Meters
Troubleshooting (continued)Table 2-1:
ErrorRecommended action(s)
Communication issues due to blocked network
ports
• Blocked network ports on the computer run-
ning Daniel MeterLink or on a company LAN
can prevent connections to the meter or
prevent certain features from working.
These issues may occur over Ethernet, Modem and Direct serial connections. Reference the list of network ports used by Daniel
MeterLink in the Help file and the symptoms
of having blocked ports. Contact your IT department for assistance in resolving these issues.
• Error condition of a blocked network:
-Cannot connect to a meter
-Cannot collect Archive log files
-Cannot view or stream waveforms in
Waveform Viewer or Signal Analyzer
-Cannot upgrade firmware
-Communications lost over serial or mo-
dem connections while Daniel MeterLink
is idle on a screen.
• Symptoms of blocked network:
-If a PING is blocked on this network port,
serial or modem connections could be
lost after approximately 15 seconds of
inactivity. This issue can be confirmed by
checking the log_meter log file in the
Temp data folder. The path of the Temp
data folder is shown in the Daniel MeterLink About dialog.
-A blocked FTP port will generally not pre-
vent a connection to the meter, but will
prevent log collections and program
downloads. A blocked FTP port could
prevent a connection in the event the
meter is running a newer version of firmware for which Daniel MeterLink does
not currently have a database configuration file. If this is a case, a message stating “Error reading database config file
dbconfig<databaseversion>.xml from
the meter.” will be displayed.
-A blocked DB API port will report “Error
10001 opening database connection to
<IP address>”.
-A blocked Streaming port will report an
error message “Unable to open a control
socket”. This will occur when opening
Troubleshooting
Maintenance and Troubleshooting manual17
Troubleshooting
Troubleshooting (continued)Table 2-1:
ErrorRecommended action(s)
the Signal Analyzer window or clicking
Read or Stream to File in the Waveform
Viewer.
Communicating with meter but all chords display failures
Cannot communicate with Daniel MeterLink
program
Cannot communicate with Field Communicator• Refer to the Emerson 475 Field Communica-
• Verify that the resistance of transducers is
within Specification (2 Ω).
• Check the Acquisition Module.
• Check the interconnect cables between the
Base assembly and the Transmitter Electronics Enclosure.
• Ensure that the meter is properly powered.
• Ensure that the computer cable is properly
connected and check your interface pins
(RS-485 or RS-232).
• Verify that the communication parameters
of the Daniel MeterLink program are correctly set.
• Check RS-485 or RS-232 communication
LEDs.
tion User’s Manual, Rev D. This manual may
be downloaded from the following location:
http://www.fieldcommunicator.com
Note
The 375 Field Communicator is no longer
available for purchase since the release of
the 475 Field Communicator. Customer supports remains available.
Cannot communicate with AMS Device Manag-erRefer to the AMS Books Online help documenta-
Connect to multiple meters via Ethernet when
they are on the same LAN
• Configure each meter with a unique user-
specified IP address (following the initial
communication quick start instructions (the
3414 Installation Manual, Section 3.5 in P/N
3-9000-759).
• Contact your IT department for valid IP ad-
dresses for your LAN and Gateway addresses.
• Disable the DHCP server.
18Gas Ultrasonic Flow Meters
Troubleshooting
Troubleshooting (continued)Table 2-1:
ErrorRecommended action(s)
Connect to multiple meters via Ethernet when
they are on the same hub but not connected to
an intranet LAN
Configuration changed• One or more parameters have been modi-
Configuration lost• The meter configuration has reset to default
• Configure each meter with a unique user-
specified IP address (following the initial
communication quick start instructions (the
3414 Installation Manual, Section 3.5 in P/N
3-9000-759).
• Assign each meter on the hub a unique IP
address within the range 192.168.135.150
through 192.168.135.254 (Gateway address
for each meter may be left unconfigured as
0.0.0.0).
• A PC may receive its IP address from an ex-
ternal DHCP server; in this case, one and only one meter must have its DHCP server enabled (the DHCP server will serve up to 10 IP
addresses to PCs attempting to talk to all
meters on the hub).
• Once a meter's IP address is configured, the
meter may be connected to the hub and accessed using that IP address.
fied in the meter's configuration
-Collect an Audit log using Daniel Meter-
Link in order to see what configuration
parameters changed and when they
changed.
-Run the Tools|Edit/Compare Con-
figuration utility and select Write All
or Write Checked values to write the
changes to the meter.
-Save the configuration file.
values and the meter is not configured correctly to measure flow and the meter has
performed a Cold Start.
-Unless the Cold Start occurred after up-
grading firmware, replace the CPU
board.
-If the cold start occurred after a firmware
upgrade, fully re-configure the meter
from a previously saved configuration
using the Tools|Edit/Compare
Configuration in Daniel MeterLink.
Maintenance and Troubleshooting manual19
Troubleshooting
Troubleshooting (continued)Table 2-1:
ErrorRecommended action(s)
Electronics Temperature is Out Of Nominal
Range
• Temperature of the electronics is out of
nominal operating range (below -40 °C or
above 100 °C) which could lead to a system
failure.
• Attempt to warm or cool the meter elec-
tronics housing.
• If the electronics is mounted to the meter
and the process fluid in the meter is over
60 °C, you must remote mount the electronics off of the meter body.
• Collect a Maintenance log using Daniel Me-
terLink while the meter is experiencing the
issue and contact your Daniel service representative.
20Gas Ultrasonic Flow Meters
Troubleshooting (continued)Table 2-1:
ErrorRecommended action(s)
Flow pressure is outside the alarm limits• Startup Issues:
-Verify that there is voltage to the pres-
sure sensor from either the meter's power supply board or from an external power supply.
-If using an analog pressure device, verify
that the pressure sensor is properly
wired to the connector.
-Verify the input is properly configured
for your pressure input.
-If using a flow computer to write pres-
sure to the meter, verify that it is properly writing to fixed flow pressure in the
proper units.
• Run Time Issues:
-Adjust pressure of process fluid to within
alarm limits.
-If using an analog pressure device and in-
put reading is 0, check if IsAI1Avail is
equal to 1 in the Meter Information dialog in Daniel MeterLink. If it is not 1, either the I/O Board has been removed or
is damaged. Reinstall or replace the
board if this value is 0.
-If using an analog pressure device, verify
that the pressure sensor is working properly.
-If using an analog pressure device, re-
check wiring and switch settings.
-If a flow computer is writing values to
the fixed flow pressure, verify that the
flow computer is still writing valid values
without Modbus write errors.
-Reverify the pressure input settings are
correct.
Troubleshooting
Maintenance and Troubleshooting manual21
Troubleshooting
Troubleshooting (continued)Table 2-1:
ErrorRecommended action(s)
Flow temperature is outside the alarm limits• Startup Issues:
-Verify that there is voltage to the tem-
perature sensor from either the meter's
power supply board or from an external
power supply.
-If using an analog pressure device, verify
that the temperature sensor is properly
wired to the connector.
-Verify the input is properly configured
for your temperature input.
-If using a flow computer to write tem-
perature to the meter, verify that it is
properly writing to fixed flow temperature in the proper units.
• Run Time Issues:
-Adjust temperature of process fluid to
within alarm limits.
-If using an analog temperature device
and input reading is 0, check if IsAI1Avail
is equal to 1 in the Meter Information dialog in Daniel MeterLink. If it is not 1, either the I/O Board has been removed or
is damaged. Reinstall or replace the
board if this value is 0.
-If using an analog temperature device,
verify that the temperature sensor is
working properly.
-If using an analog temperature device,
recheck wiring and switch settings.
-If a flow computer is writing values to
the fixed flow temperature, verify that
the flow computer is still writing valid
values without Modbus write errors.
-Reverify the temperature input settings
are correct.
22Gas Ultrasonic Flow Meters
Troubleshooting (continued)Table 2-1:
ErrorRecommended action(s)
Program download failed during firmware upgrade
• If the meter power fails during a firmware
upgrade process, perform a backup upgrade
in an attempt to connect to the meter and
download the program again.
-In Daniel MeterLink go to the File pull-
down menu and select Program Settings.
-Enable the Allow FTP-only connection.
-For Serial Port Connections: for Daniel
3410 Series Gas Ultrasonic meters connect to Port A. You may need to adjust
your Meter Directory settings for the
connection so that they match the port
default settings. Port A will default to
19200 baud with a Modbus address of
32.
-For Ethernet Connections: If you are con-
necting to the meter over an Ethernet
port, you should be able to connect with
the same IP address as normal. If this is
unsuccessful, the meter may have defaulted to 192.168.135.100 with a subnet of 255.255.255.0. Make sure your PC
has a compatible address and attempt a
connection using this IP address.
-Make sure your cabling and your Meter
Directory record are setup, then attempt
to connect to the meter.
-You will receive a message “Error 10001
opening database connection to …..”.
Click OK and you will be prompted to
“Attempt FTP-only connection ….”. Click
Yes and if successful, the Daniel MeterLink™ caption displays “…Connected to
<meter name>…”. Go to the Tools pull
down menu and select Program Download to attempt the firmware upgrade
again.
-If the firmware upgrade is successful, the
meter should start working as normally
as the meter’s configuration is not normally lost.
-If the configuration is lost, use Daniel
MeterLink Tools>Edit/Compare Configuration to write the saved configuration
back to the meter. The saved configuration files are typically stored in C:\Ultrasonic Data folder.
Troubleshooting
Maintenance and Troubleshooting manual23
Troubleshooting
Troubleshooting (continued)Table 2-1:
ErrorRecommended action(s)
-Restart the meter to install the firmware.
Daniel MeterLink prompts you with a
message that it must disconnect from
the meter. Once the firmware upgrade is
complete you will be able to reconnect
to the meter with Daniel MeterLinK.
-When the meter restarts, it takes about
two minutes before you will be able to
reconnect depending on the firmware
upgrade being performed. If the database does need to be reinitialized, it
could take up to five minutes.
-After an upgrade, it is recommended to
reconnect to the meter and repeat the
Program Download process.
-If all the program components are suc-
cessfully updated, they will show to be
the same date and version as the Currently Installed Versions and the Download button will be disabled.
-If one or more components are still not
updated, click Download to continue the
upgrade process.
No power to the unit
One or more of the chords is not indicating a
reading (reporting zeros)
• Check for correct voltage (10.4 - 36 VDC)
(refer to the System Wiring Diagram P/N
DMC-005324 in XREF Appendix).
• Check the main power source for blown fuse
or tripped circuit breaker. Reference your
“as built” installation drawings for your location.
• Check for loose connections at the cable
connectors.
• Check the resistance of the transducers
(should be approximately 2 Ω).
• Problem also may be caused by a bad Ac-
quisition Module or interconnect cable.
• Check system status in the Daniel MeterLink
program for any flagged errors.
• Check the CPU Module.
• If Chord A is not indicating, Change the
transducer cables from Chord B to chord A.
If Chord B then fails, the transducers are bad
on Chord A.
24Gas Ultrasonic Flow Meters
Troubleshooting
Troubleshooting (continued)Table 2-1:
ErrorRecommended action(s)
Power Failure• Meter has had power removed for a period
of time or the meter restarted itself such as
after a firmware upgrade. The Audit log in
the meter indicates the power fail time.
-If this was an unexpected restart of the
meter, verify the integrity of the power
to the meter and make sure that the
voltage level is the in the range of 11-36
VDC at the meter.
-If this was a known power fail or restart
of the meter, just acknowledge this
alarm.
Sound velocity is outside defined limits• The meter's measured average sound veloci-
ty is outside the defined limits.
-Verify that all chords are measuring the
same Speed of Sound within about
0.15%. Look for alarms that indicate
transducer problems and resolve any of
these issues. This could include failing
transducers, debris buildup on transducers, or incorrectly entered path lengths
in the configuration.
-If the chords agree, adjust the SSMin or
SSMax using the Edit/Compare Config
utility in Daniel MeterLinkTM so the meter's average speed of sound falls within
these limits.
-Collect a Maintenance log using Daniel
MeterLink and contact your Daniel service representative.
Maintenance and Troubleshooting manual25
Troubleshooting
Troubleshooting (continued)Table 2-1:
ErrorRecommended action(s)
Sound velocity is outside defined limits• The meter's measured average sound veloci-
ty is outside the defined limits.
-Verify that all chords are measuring the
same Speed of Sound within about
0.15%. Look for alarms that indicate
transducer problems and resolve any of
these issues. This could include failing
transducers, debris buildup on transducers, or incorrectly entered path lengths
in the configuration.
-If the chords agree, adjust the SSMin or
SSMax using the Edit/Compare Config
utility in Daniel MeterLinkTM so the meter's average speed of sound falls within
these limits.
-Collect a Maintenance log using Daniel
MeterLink and contact your Daniel service representative.
Waveform contains an excessive amount of
noise
Use the Daniel MeterLink Meter>Signal An-alyzer to increase the StackSize until noise
level decreases (settings can be 1 (none) 2, 4, 8,
or 16). If increasing the StackSize is not successful, try turning on the filter or consult with Daniel Customer Service if you are unsure of how
stacking a signal can affect the meter's operation.
2.2.1Meter maintenance
The Monitor (Summary) includes the direction of flow measurement, velocity rate, units of
measurement, uncorrected or corrected flow (if applicable for your meter) and a bar graph
for a visual comparison between the velocities for each chord. This is the default view
displayed when you select Meter|Monitor from the toolbar.
26Gas Ultrasonic Flow Meters
Troubleshooting
Meter Monitor (Summary) viewFigure 2-4:
Run Daniel MeterLink and open the Meter Monitor (Detailed) view to perform a
diagnostics health check and or adjust parameters for your site requirements. If you wish
to use the Monitor (Detailed) dialog as the default view, click the checkbox in the lower
portion of the dialog box.
Meter Monitor (Detailed) viewFigure 2-5:
The following details the information displayed in this dialog box.
•Flow Properties Table - the table at the top of the Meter Monitor dialog box shows
basic information about the condition of the flow in the meter.
•Flow Velocity/Flow Ratios Bar Graph - provides a visual comparison between the
velocities for each chord.
Maintenance and Troubleshooting manual27
Troubleshooting
•Chord Speeds of Sound Bar Graph - a visual comparison between the calculated
speeds of sound for each chord.
•Gain/Performance Bar Graph - provides either a visual comparison of the average of
the upstream and downstream gains for each chord or a visual comparison of the
average of the upstream and downstream performance for each chord.
•Signal to Noise Bar Graph - provides a visual comparison between the signal to noise
ratio for each chord direction.
•Meter Status Alarms - provides a visual indication of the meter’s status.
•Gas Comp - this dialog shows the gas composition that can be used by the AGA8 or
AGA10 gas calculations.
•Baseline - this dialog shows the meter’s flow characteristics in comparison to limits
defined for the Continuous Flow Analysis features. This dialog is only available for
four path meters that support a baseline and a valid Continuous Flow Analysis key.
•Run time - displays how long the monitor screen has been collecting data.
•Meter Time - the time displayed is the time from the Ultrasonic meter.
Note
If the time displayed has a yellow background, that is an indication that the meter’s time is
more than 10 minutes apart from the PC’s time.
•Meter Data List - displays read-only data selected from the drop-down list.
•Chart - the chart utility displays the data collected for the value selected from the
Chart drop-down list.
Refer to Table 2-1 for error resolutions and Table 2-2 for meter maintenance hardware
diagnostics.
MaintenanceTable 2-2:
Daniel MeterLink utilityDiagnosticsAction(s)
Meter Monitor (Summary) view • Check Status for active
alarms
• Meter Status LED is green if
there are no active alarms.
This indicates the meter is
measuring flow and operating within the calibrated
parameters.
• Meter Status LED is red.
This indicates an active
alarm. Resolve and acknowledge active alarms as
displayed on the Status
Summary page. Click the
Help button beside the
alarm description to display
information about the
alarm and recommended
actions to resolve the issue.
28Gas Ultrasonic Flow Meters
Troubleshooting
Maintenance (continued)Table 2-2:
Daniel MeterLink utilityDiagnosticsAction(s)
Meter Monitor (Detailed) view• Flow Profile• Flow profile velocity for me-
ters with four Chordal paths
- Chords A and D shown on
the Bar graph should be of
equal lengths and at 5 ft/s
should be 0.89 and Chords
B and C should be 1.0 indicating a good flow profile.
• If the velocity range is
greater than a 10% differential between chords, a degradation in the symmetry is
indicated by a change in
the flow velocity for the
chord and the measured
SOS. A chord that causes
the spread for the Speed of
Sound to vary more than
0.35% of the Average
Speed of Sound, the bar for
that chord turns yellow.
• Clicking one of the arrows
at the top left of the graph
will change the chart to
SOS Differences from the
average meter speed of
sound. This provides a
quick indication on how
much spread in speed of
sound is between the
chords. Negative values are
shown in blue.
• Compare Gains and Signal
to Noise (SNR) ratios decibel values with the values in
the Maintenance log Inspection report from the
meter flow calibration.
• Check chord average signal
amplitudes with the meter
baseline values in the Maintenance log Inspection report.
• The meter may not be in
measurement mode or
there are too few operating
chords.
Maintenance and Troubleshooting manual29
Troubleshooting
Maintenance (continued)Table 2-2:
Daniel MeterLink utilityDiagnosticsAction(s)
• If a chord is hard failed, the
Check Status LED will
change from green to red.
The issue may be the transducer pair for the failed
chord or the transducer cabling. Resolve the issue and
clear the alarm.
• If installed, check the flow
conditioner for blockage.
• If you have enabled the
Continuous Flow Analysis
feature, from the Meter
Monitor (Detailed) view,
click the Baseline button.
The Baseline Viewer displays the meter’s flow characteristics including: Flow
Velocity, Profile Factor,
Swirl Angle, Symmetry,
Crossflow, and Path Turbulences.
Meter Monitor (Summary) view
Meter Flow Properties Table
• Flow velocity• Check the flow direction. If
reverse flow is detected,
check for valve leaks.
• If the meter run typically
has reverse flow when flow
is stopped, reconfigure the
ReverseFlowVolLmt to allow a higher volume from
the Field Setup Wiz-
ard> General Page.
Meter Monitor (Detailed) view
Monitor Chart Selection list
• Speed of Sound• Compare Speed of Sound
deviation from measured
SOS relative to the average
SOS.
• Check the chord’s SOS.
• Check and correct geome-
try configuration (pipe diameter, distance between
the transducers (LA), and
delay time).
• If present, resolve transduc-
er issues (failed transducer,
cabling or debris buildup on
30Gas Ultrasonic Flow Meters
Maintenance (continued)Table 2-2:
Daniel MeterLink utilityDiagnosticsAction(s)
the transducer face, or path
length configured incorrectly).
• Adjust SSMin or SSMax only
if other checks pass (consult a Daniel Service representative before making
these adjustments).
Meter Monitor (Detailed) view
Meter Data List
• Electronics Temperature of
out range
• Temperature of the elec-
tronics is out of nominal
operating range below
-40 °C or above 100 °C
(-40 °F or above 212 °F).
-Heat or cool the meter
electronics housing. If
operating temperature
exceeds 60oC, remote
mount the Transmitter
Electronics Enclosure.
Troubleshooting
Meter Monitor (Detailed) view
Meter Data List
• Electronics voltage out of
range
• Systems voltages are valid
if 1.0V, 1.2V, 2.5V, 3.3V or
the Acquisition Module valid voltages are 1.2V, 2.5V
or 3.3V.
• Replace the CPU Module if
one or more of the System
Voltages is out of range.
• Replace the Acquisition
Module if one or more of
the voltages is out of range.
Maintenance and Troubleshooting manual31
Troubleshooting
Maintenance (continued)Table 2-2:
Daniel MeterLink utilityDiagnosticsAction(s)
MeterLink Tools Menu• Frequency output• Run the Frequency Outputs
test
• If the output reads zero,
you may require a pull up
resistor 1.2kOHM, 0.5W.
• Analog outputs• Run Analog Outputs test
• Verify outputs are within
4mA-20mA range
-0% = 4mA
-25% = 8mA
-50% = 12mA
-75% = 16mA
-100% = 20mA
• Digital outputs• Run the Digital Outputs
test.
• Digital Output Content is in
relation to frequency validity and flow direction configuration and polarity.
Meter Electronics• Acquisition Module com-
munications error
• Check firmware revision
and upgrade if necessary
using Daniel Meter-
Link Tools>Program
Download.
• If LED 4 and LED5 are not
Solid Green check interconnects cable between CPU
Acquisition Module. Check
IS Barrier Module or Acquisition Module.
32Gas Ultrasonic Flow Meters
Troubleshooting
Maintenance (continued)Table 2-2:
Daniel MeterLink utilityDiagnosticsAction(s)
MeterLink Logs/ Reports Menu• Meter performed a Warm
Start or a Warm start required
MeterLink Tools>Edit/
Compare Configuration
Menu
MeterLink Logs/ Reports Menu• Power failure• If this was a known power
• Meter performed a Cold
Start
• Meter performed a Warm
Start:
-Collect an Archive event
log (Audit log) using
Daniel MeterLink to
view configuration parameter changes and
when they changed.
• Warm start is required:
-When you make
changes to the transducer characteristics,
sample rates, the device number, or a Modbus map file.
• The meter configuration
has reset to default values
and the meter is not configured correctly to measure
flow.
• Unless the cold start occur-
red after upgrading firmware, you may need to replace the CPU Module.
• If the Cold Start occurred
after a firmware upgrade,
you must reconfigure the
meter from a previously
saved configuration file using the Edit>Compare
Configuration screen.
Then clear the latched
alarm on the Status Summary page.
fail or restart of the meter
just acknowledge this
alarm on the Status Summary page.
• If this was an unexpected
restart of the meter, verify
the integrity of the power
to the meter and make sure
that the voltage level is in
the range of 11-36 VDC at
the meter.
Maintenance and Troubleshooting manual33
Troubleshooting
Maintenance (continued)Table 2-2:
Daniel MeterLink utilityDiagnosticsAction(s)
• Collect a complete Archive
log and contact your local
area Daniel service representative
to one pair of transducers
or its cabling. Check the
transducer wiring for this
pair of transducers to make
sure connections are secure
and wired correctly.
• Verify the average gain of
this transducer pair is not
above 90dB. Read the value
from the Daniel MeterLink
Monitor Page or using AMS
under Service
Tools>Path performance.
• Remove the transducer and
clean the transducer face
( Transducer removal procedure).
• Only authorized personnel
may remove security seals.
Follow your standard operating procedure to report
seals that have been tampered with or removed and
replace the seals per instructions in
External power source connection and fuse in the In-
stallation Manual
(3-9000-765).
• Inspect ground lug wiring
and make sure the wiring is
tightly secured.
34Gas Ultrasonic Flow Meters
Troubleshooting
Maintenance (continued)Table 2-2:
Daniel MeterLink utilityDiagnosticsAction(s)
Conduit seals• Transmitter Electronics En-
closure
Flanges• Inspect for leaks
• Inspect flange stabilizers
• Inspect the conduit sealant
and follow your standard
operating procedure to report tampering with the
conduit sealant.
• Your operating procedures
may require a certified electrician and company witness to reseal the conduit.
• Perform leak tests on flang-
es
• Ensure flange stabilizers are
installed.
WARNING!
CRUSHING HAZARD
Do not remove flange stabilizers.
Attempting to do so may allow the meter to roll, resulting in serious injury or equipment
damage.
A. Flange stabilizers
2.2.2Unable to connect direct serial or external serial modem
If you are using Windows® XP, Windows® 7 or Windows® 8 make sure that you do not have
more than one modem driver installed to the same COM port. Typically this will only be
necessary if you use one COM port to talk direct (serial communications) and use the same
COM port to connect to an external modem. This is an apparent limitation in Microsoft’s
Dial-up Networking. If more than one modem driver is installed for a particular COM port,
Dialup Networking will always use the last driver installed regardless of what is selected.
Maintenance and Troubleshooting manual35
®
Troubleshooting
The only work around is to only install one modem driver per COM port on the PC at a
time. Refer to the Daniel MeterLink™ Quick Start Manual (P/N 3-9000-763) for phone and
modem details. The manual may also be downloaded from the Daniel website.
If you receive the error message “Unable to connect to meter” when trying to connect to a
Daniel 3410 Series Gas Ultrasonic Flow Meter, refer to the following:
•Ethernet Connections (Section 2.2.4)
•Direct Serial Connections (Section 2.2.5)
2.2.4Ethernet connections
If you received this message while trying to connect over Ethernet, verify you have the
correct IP address in the Meter Directory record. If the meter is to assign the IP address,
make sure the IP address is set to 192.168.135.100 and that the DHCP switch is in the ON
position on the CPU Module. If the meter has a fixed IP address, verify the IP address,
Subnet, and Gateway are correct in the meter. If going through a hub, verify that the
computer and meter are connected to the hub with straight-through patch cables.
2.2.5
2.3
Direct serial connections
Verify the switch settings on the CPU Module. Also verify your wiring between the meter
and the computer running Daniel MeterLink™ using the Field Wiring drawing DMC-005324
in Appendix B. Verify the Comms Address and Baud rate are correct in the Meter Directory
record.
For additional information on wiring and configuring the meter for the various
communication options refer to the Daniel 3410 Series Gas Ultrasonic Flow Meters
Installation Manual (P/N 3-9000-759, Section 3.5).
Troubleshoot maintenance log files and trend
files
2.3.1Files do not appear in workbook
Maintenance Log files and Trend files that exist on the PC do not appear in the Microsoft
Excel® workbooks tree under Trend Maintenance Logs. This is most likely caused by the
fact that the desired file or files are already open in Microsoft® Excel®. Open files can not be
verified as Maintenance Log files or Trend Files by Daniel MeterLink and are left out of the
list. Simply close the files in Microsoft® Excel® and then close and reopen the Trend
Maintenance Logs dialog box to include them in the list.
®
36Gas Ultrasonic Flow Meters
Troubleshooting
2.3.2Microsoft® Excel® Log/Export options are not available
In order for the Excel® log/export options to be available, Excel® 2000 or later must be
installed on the machine and at least one printer must be installed under Windows®. If
®
®
Excel
still unavailable, it may be because Excel
the Windows® default printer. If the Windows® default printer is a network printer and you
are not currently connected to the network, then Excel
2000 or later is installed and you have printers installed but the Excel
®
®
cannot access the printer driver information of
®
®
will most likely not be able to
access the printer driver information and Daniel MeterLink™ cannot use Excel
®
®
option is
®
®
to
generate reports or logs. One solution is to install a local printer on your machine tied to
LPT1. The local printer driver you installed can be for any printer and the printer does not
actually have to exist or be connected to the PC. If you install a local printer, you can
configure Daniel MeterLink to temporarily change your Windows® default printer over to
this local printer while running Daniel MeterLink. Do this by selecting this local printer for
the Override system default printer selection in the Program Settings dialog. Daniel
MeterLink will automatically change the Windows
override printer when it starts and will set the Windows
®
®
default printer to the selected
®
®
default printer back to its
original printer when it closes.
2.3.3Maintenance logs or trend files are not created
When using Excel® XP or later, some of the worksheets in the Maintenance Logs or Trend
files are not created. If the Inspection sheet of the Maintenance Log file or the Charts sheet
of a Trend files are not generated, it is probably because Excel® is not configured to allow
Daniel MeterLink to run the Visual Basic® script that generates the page. Excel® can be
configured to allow Daniel MeterLink to run the Visual Basic® script by following the
instructions below.
To enable Excel® 2000 to work with Daniel MeterLink,
Procedure
1.Select Tools>Macros>Security menu path.
Maintenance and Troubleshooting manual37
Troubleshooting
Excel® 2000 Tools MenuFigure 2-6:
2.The Security dialog appears. Select the Trusted Sources tab.
3.Click the Trust access to Visual Basic Project radio button and click OK to apply your
selections.
38Gas Ultrasonic Flow Meters
Troubleshooting
Excel® Trusted Access SettingFigure 2-7:
a. To enable Excel® 2007 to work with Daniel MeterLink, customize the Ribbon to
5.Select Macro Settings from the left panel, then click the Enable all macros (not
recommended; potentially dangerous code can run) radio button.
6.Place a check mark in “Trust access to the VBA project object model”.
7.Click OK to apply the changes and close the Trust Center dialog.
40Gas Ultrasonic Flow Meters
3Meter repairs
Topics covered in this chapter:
Prior Precautions
•
Field hydrostatic pressure testing procedures
•
T-Slot transducer removal and installation procedures
•
Transducer holder removal and installation
•
Transducer cable removal and installation
•
Replace the meter electronics
•
Replace the Adquisition Module
•
3.1Prior Precautions
This section includes discussion of the maintenance of Daniel 3410 Series Ultrasonic
Meters. For reference, you may download the Daniel MeterLink Quick Start Manual from:
The meter body and piping may be extremely hot or cold.
Wear appropriate personal protective equipment when coming in contact with the meter.
Failure to comply may result in injury.
CAUTION!
TRANSPORTATION HAZARD
When moving the meter, do not insert the forks of a forklift into the bore.
Inserting the forks may cause the meter to become unstable, resulting in injury or damage to
the bore and sealing face.
CAUTION!
TRIPPING HAZARD
Clear all obstacles or obstructions from the work area when transporting, installing or removing the
meter.
Failure to comply may cause injury to personnel.
Maintenance and Troubleshooting manual41
Meter repairs
NOTICE
Prior to lifting the unit, refer to the Daniel Gas Ultrasonic Flow Meter nameplate or outline
dimensional (general arrangement) drawing for the assembled weight.
WARNING!
CRUSHING HAZARD
Do not remove flange stabilizers.
Attempting to do so may allow the meter to roll, resulting in serious injury or equipment
damage.
A. Flange stabilizers
WARNING!
FLUID CONTENTS MAY BE UNDER PRESSURE
When the meter is under pressure, DO NOT attempt to remove or adjust the transducer holder.
Attempting to do so may release pressurized gas or fluid, resulting in serious injury or
equipment damage.
42Gas Ultrasonic Flow Meters
Meter repairs
WARNING!
CONTENTS MAY BE HAZARDOUS
The meter must be fully depressurized and drained before attempting to remove the
transducer holder. If gas or fluid begins to leak from the transducer holder, stop immediately
and reinstall the holder.
Failure to comply may cause serious injury or equipment damage.
3.2
A. Transducer holder
CAUTION!
ESCAPING GASES OR FLUIDS HAZARD
The purchaser of the meter is responsible for the selection of Daniel components/seals and materials
compatible with the chemical properties of gas flow measurement.
Failure to select the suitable meter component/seals may cause escaping gases or liquids,
resulting in injury or equipment damage.
WARNING!
CRUSHING HAZARD
During meter installation or removal, always place the unit on a stable platform or surface that
supports its assembled weight.
Failure to comply could allow the meter to roll, resulting in serious injury or equipment
damage.
Field hydrostatic pressure testing procedures
The Daniel 3410 Series Ultrasonic Gas Flow Meter are supplied with T-Slot transducer
assemblies which are extractable with or without line pressurize.
Maintenance and Troubleshooting manual43
Meter repairs
T-Slot transducer assemblies are available on 4-inch and larger meters for Daniel 3410
Series Ultrasonic Gas Flow Meters. The T-Slot assemblies are available on all meter sizes for
the Daniel JuniorSonic™ Meters (Models 3412 and 3411).
CAUTION!
LEAKAGE OR PRESSURE CONTAINING PARTS FAILURE
Use precautions to eliminate hazards to personnel in the event of leakage or failure of the liquid
ultrasonic meter pressure containing parts or failure of the test equipment and to prevent overpressurization during the test procedure.
Failure to comply may result in injury to personnel or cause damage to the equipment.
3.2.1T-Slot Transducer assembly and mount
T-slot transducer mount and holderFigure 3-1:
A. Transducer mount
B. Captive screws
C. T-slot transducer holder
D. Cable nut
E. Chordset
Procedure
1.Slowly vent all line pressure on the 3410 Series Gas Ultrasonic Meter to atmosphere.
2.Disconnect transducer cable from the transducer holder.
3.If installed, remove the mount cover by loosening the two mount cover captive
screws.
44Gas Ultrasonic Flow Meters
Meter repairs
4.Loosen the T-Slot transducer assembly with a 1 1/4 inch (32 mm) socket. Carefully
remove the T-Slot transducer assembly.
5.Place a label on the transducer assembly to marks its location (i.e., Model 3414-A1,
A2, B1, B2, C1, C2, D1, or D2; 3412-A1, A2, B1, B2; or 3411-A). Port locations are
marked on the transducer cable as well as on cast meter housings.
6.Inspect the transducer mount threads and ensure they are clean and free of debris.
7.Apply a small amount of Nickel antiseize compound (P/N 3-9960-134) to the
threads of the Hydrotest plug (P/N 1-360-01-212) from kit (P/N 1-360-01-220) and
install it into the mount. JuniorSonic kit part numbers are listed below.
• JuniorSonic™ SP Field Hydrotest Kit (P/N 1-360-01-221)
• JuniorSonic™ DP Field Hydrotest Kit (P/N 1-360-01-222)
8.Repeat Step 1 through Step 6 for the other transducer(s) being careful to note the
location of each transducer in the meter assembly.
9.Run the field hydrostatic test.
10.Reverse the steps above to reinstall the transducers into their appropriate ports.
Before reinstalling the transducer assemblies, ensure the transducer ports, mounts,
and transducer holders are clean and free of debris. Apply a small amount of Nickel
antiseize compound to the outer threads of the transducer holders before installing
them into the mounts.
3.3T-Slot transducer removal and installation
procedures
The T-Slot transducer assembly offers improved transducer alignment and superior
acoustic isolation between the transducer and the meter housing. The net result is
improved performance and stability. The assembly is used on Daniel 3414, 3410 and 3411
meters and is line pressure vented. The gas temperature ranges are as show in Table 3-1.
Temperature ranges for transducers, mounts and holdersTable 3-1:
Transducer typeTemperature rangeMount and holder type
To replace transducers under pressure, refer to Ultrasonic Split Clamp Extractor Tool
Operation Manual (Daniel P/N 3-9000-744).
NOTICE
To access the product datasheet from the Daniel products page (above link), select the Daniel
Gas Ultrasonic Flow Meter link, click the Documentation tab, expand the Data Sheets - Bulletins
- Catalogs tab, then select the Data Sheet.
The Daniel Split Clamp Extractor tool is used with Daniel Model 3414, 4-Path
(SeniorSonic™) meters only.
46Gas Ultrasonic Flow Meters
J-Mount transducer assemblyFigure 3-2:
A. Transducer cable (max. length 15 ft.)
B. Transducer cable nut and chordset
C. Mount
D. Transducer assembly, stalk and holder
Meter repairs
J-Mount transducer disassemblyFigure 3-3:
A. Transducer cable (max. length 15 ft.)
B. Transducer cable nut and chordset
C. Transducer holder
D. Transducer stalk (optional)
E. Transducer
F. Mount
Maintenance and Troubleshooting manual47
Meter repairs
3.3.2Transducer removal without extractor tool
This procedure requires venting the line pressure prior to removing the transducer(s).
Before removing and installing transducer(s), connect to the meter using Daniel MeterLink
and collect and save a Maintenance Log.
NOTICE
The following instructions are for transducer removal and installation without the use of an
extractor tool. It is recommended that one work on one transducer assembly at a time to
reduce the possibility of improper assembly with respect to transducer lengths and location.
•Transducers are always replaced in pairs.
•Update the calibration parameters for every chord when replacing a pair of transducers.
Refer to Section 3.3.5.
Tools and supplies for disassembly and assemblyTable 3-2:
Tool list, shop supplies and reference documents
3/8” drive ratchet
3/8” drive extension
3/8” drive 7/16” deep socket
3/8” drive 5/16” hex socket
3/8” drive 1/2” hex socket
3/8” to 1/2” driver adapter
1/2” drive ratchet
1/2” drive 1 1/4” Socket
1 1/4” combination wrench
1 1/8” combination wrench
1” combination wrench
3/16” flat Blade screw driver
1/16” Xcelite Hex driver
6 mm Allen wrench
N.A.S. Loctite Nickel Anti-Seize 16 oz. Brush Top P/N 77164
Molykote 111
Shop supplies
Shop rags
Adhesive stickers for the cable nuts Sharpie Fine Point Permanent Marker (Black) Series # 30000, to
label transducers:
• A1, A2, B1, B2, C1, C2, D1 and D2 for SeniorSonic meters
1.Blow the line down according to the site standard operating procedures.
2.Ensure that the line pressure is down to atmospheric pressure prior to disassembly.
3.Disconnect transducer cabling and chordset from the transducer assembly by
turning the cable nut counter-clock wise.
4.Disconnect the transducer cable nut. Pull the chordset from the transducer
assembly.
Note
Do not twist the chordset as you disconnect it from the transducer assembly.
5.If the meter is equipped with T-21 or T-22 transducers, remove the transducer
retainer using a 1 1/8” wrench and then disconnect and remove the transformer
(Figure 3-4).
Note
T-21 transducers use W-01 transformers and T-22 transducers use W-02 transformers.
Maintenance and Troubleshooting manual49
Meter repairs
T-21 and T-22 transducer assemblyFigure 3-4:
A. Transducer cable (max. length 15 ft.)
B. Transducer cable nut and chordset
C. Transformer retainer (Standard P/N 1-360-01-958 or High Temperature P/N 1-360-01-978)
D. Transformer assembly T-21 (W-01 P/N 1-360-03-090) or T-22 (W-02 P/N 1-360-03-110)
E. Transducer holder
F. Transducer holder O-rings
G. Set screw holes (end of transducer holder)
H. Transducer stalk (optional)
I. Transducer assembly
J. Mount and backup O-ring
6.Loosen the T-Slot transducer holder assembly with a 1 1/4” socket. Carefully remove
the T-Slot transducer assembly.
7.Loosen the three Allen setscrews with a 1/16” hex driver securing the transducer
assembly and stalk, if installed. Carefully remove the old transducer by pulling it
from the T-Slot transducer holder assembly without rotating.
Important
Record the “L” dimension of the removed transducers which is used to update the meter
configuration after all of the transducers are replaced. Make sure you have the report sheet
containing the “L” dimension, Delay Time, and Delta Delay Time for the replacement pair of
transducers to use during the Transducer Swap-out procedure in Daniel MeterLink.
8.Clean the transducer holder with a dry cloth.
50Gas Ultrasonic Flow Meters
3.3.3Transducers installation procedure
1.Ensure that the Daniel 3410 Series Ultrasonic Gas Flow Meter transducer port,
mount, and T-Slot transducer holder assembly are clean and free of debris.
2.Apply a small amount of Molykote 111 to the female contacts on the transducer.
3.Install the transducer assembly into the transducer holder or into the stalk (if
required). The parts are keyed and can only be assembled one way. As the
transducers are installed into the holder or stalk assembly, they must be labeled
with a marker for future reference (i.e., transducer #1 would be A-1 and transducer
#2 would be A-2).
4.Use a 1/16” hex driver to equally tighten the three Allen set screws on the
transducer holder to secure the transducer assembly and the stalks (if installed).
Transducer holder, stalk and transducer assemblyFigure 3-5:
Meter repairs
A. Transducer holder
B. Stalk
C. Transducer assembly
Note
Do not apply lubricant to the transducer or stalk O-rings.
NOTICE
Ensure that the transducers identified as belonging to end 1 are installed on end 1 of the
meter housing and those identified as belonging to end 2 are installed on end 2 of the
meter housing.
5.Replace the O-ring and Backup O-ring on the transducer holder. It is highly
recommended that the O-rings be replaced when the transducer is removed from
the holder or stalk. Make sure that the contoured side of the ring is facing away from
the mount. Lubricate with Molykote 111 Silicone Grease or equivalent.
Maintenance and Troubleshooting manual51
Meter repairs
Note
Replacing the O-rings at this point minimizes the chances of damaging the transducer by
dropping it.
6.Apply a small amount of nickel anti-seize (N.A.S.) compound (P/N 2-9-9960-134) to
the outer threads of the transducer holder (see Figure 3-5).
7.7. Carefully install the transducer holder assembly into the transducer mount. Make
sure the threads of the holder and mount are correctly aligned. Use a 1 1/4” socket
and screw the transducer assembly into the mount. Tighten to securely seat the
assembly in the mount. Do not over tighten (see Figure 3-6).
8.If replacing T-21 or T-22 transducers, install the keyed transformer assembly into the
transducer holder (see Figure 3-6).
a. Apply a small amount of Molykote 111 to the transformer assembly O-ring.
b. Insert the keyed transformer into the back end of the transducer holder.
E. Transducer retainer (Standard P/N 1-360-01-958) (High temperature P/N 1-360-01-978)
F. Cable nut and chordset:
• 100° C -5’ length (P/N 3-3400-190)
• 100° C -15’ length (P/N 3-3400-194)
52Gas Ultrasonic Flow Meters
Meter repairs
9.For T-21 and T-22 transducers, Place the retainer over the transformer assembly,
ensure the retainer threads are aligned correctly and hand-tighten. Use a 1 1/8”
wrench and turn clockwise until the transformer assembly is fully seated in the
transducer holder.
Note
Do not over tighten the retainer.
10.Reconnect the transducer chordset to the transducer holder or the retainer if
installing T-21 or T-22 transducers. The internal connector of the transducer
chordset is keyed and will only go on one way. Secure the transducer cabling nut by
turning clock-wise. Ensure the cable nut threads are correctly aligned.
11.Repeat Step 1 through Step 10 for the remaining transducer assemblies which were
replaced.
12.Check that the Daniel 3410 Series Ultrasonic Gas Flow Meter is pressure tight.
Pressurize the meter to line pressure. Check for leaks around all mounts and
transducer holders, which were removed, using soapy water or other recognized
leak detector. If leaks are found, the meter must be vented to atmosphere and the
problem corrected. Check for leaks again. Continue the process until there are no
leaks.
13.Continue with Section 3.3.5 to use the Daniel MeterLink Transducer Swap-out
Wizard.
3.3.4
Replace transformers procedure
If your meter is equipped with T-21 or T-22 transducers and you need to replace the
transformers, use the following procedure. Refer to Figure 3-4 and Figure 3-6.
Procedure
1.Disconnect the transducer chordset from the transducer retainer by turning the
cable nut counter-clock wise.
Note
Do not twist the chordset as you disconnect it from the transformer.
2.Unscrew the retainer from the holder using a 1 1/8” wrench.
3.Pull the transformer assembly from the transducer holder assembly.
4.Apply a small amount of Molykote 111 to the O-rings on the replacement
transformer assembly.
5.Plug the replacement transformer assembly into the transducer holder assembly.
The transformer is keyed and can only be installed one way.
Note
T-21 transducers use W-01 transformers and T-22 transducers use W-02 transformers.
Maintenance and Troubleshooting manual53
Meter repairs
6.Place the retainer over the transformer assembly. Ensure the retainer threads are
aligned correctly and hand-tighten. Use an 1 1/8” wrench and turn clockwise until
fully seated in the transducer holder.
Note
Do not over tighten the retainer.
7.Plug the keyed chordset into the transformer assembly.
8.Thread the cable nut onto the transducer retainer. Ensure the cable nut threads are
aligned correctly and hand-tighten.
3.3.5Modifying the calibration parameters
When transducer pairs, mounts, stalks or transducer holders are replaced, the
corresponding meter calibration parameters must be updated for accurate operation. This
means modifying the affected chord "L" dimension (LA... LD) (see Determining the “L”
Value below), average delay time (AvgDlyA... AvgDlyD) and delta delay time (DltDlyA...
DltDlyD) using the Daniel MeterLink Transducer Swap-out Wizard (see Figure 3-7).
Average delay time and delta delay time modifications
The transducer pair average delay time and delta delay time are located on the transducer
pair calibration sheet. These values must be downloaded to the appropriate meter data
points (AvgDlyA... AvgDlyD, DltDlyA... DltDlyD). The lengths of the transducers are also
included on the calibration sheet and are etched on the transducers. Likewise the lengths
of the stalk assemblies, transducer holders, and mounts are etched on the individual
components. The length of the meter body is found on the original calibration sheet
supplied with the meter.
Determining the “L” value
The value "L" is determined by adding the length of the meter body to the lengths of the
two mounts and subtracting the lengths of the transducer holders, stalk assemblies, and
transducers. This value should be written to the appropriate meter data points for each
chord that received new transducers (LA... LD). See Equation 3-1 for the “L” dimension
calculation.
54Gas Ultrasonic Flow Meters
Daniel MeterLink Transducer Swap-out WizardFigure 3-7:
Chord “L” Dimension Calculation
Meter repairs
The chord “L” dimension is calculated from the meter housing length as well as the
transducer pair lengths, mount lengths, holder lengths, and stalk lengths as shown in
Equation 3-1. The transducer lengths are etched on the transducers. Likewise, the lengths
of the mounts, stalk assemblies, and transducer holders are also etched on the individual
components. The length of the meter body is found on the original calibration sheet
supplied with the meter.
Chord “L” Dimension Equation 3-1:
ℎ
=
sin
1
2
+
1
2
1
1
2
+
2
where
L
chord
L
MeterHousing
L
Mount1
L
Mount2
L
Xdcr1
L
Xdcr2
L
Stalk1
L
Stalk2
L
Hldr1
chord “L” dimension (in) (LA ... LD)
meter housing length (in)
transducer 1 mount length(in)
transducer 2 mount length (in)
transducer 1 length (in)
transducer 2 length (in)
transducer 1 stalk length (in)
transducer 2 stalk length (in)
transducer 1 holder length (in)
Maintenance and Troubleshooting manual55
Meter repairs
L
Hldr2
Tip
The transducer “L” dimension is re-calculated when you run the Transducer Swap-out utility.
transducer 2 holder length (in)
3.4Transducer holder removal and installation
WARNING!
CONTENTS MAY BE UNDER PRESSURE
When the meter is under pressure, DO NOT attempt to remove or adjust the transducer holder.
Attempting to do so may release pressurized gases, resulting in serious injury or equipment
damage.
WARNING!
CONTENTS MAY BE HAZARDOUS
The meter must be fully depressurized and drained before attempting to remove the
transducer holder. If gas or fluid begins to leak from the transducer holder, stop immediately
and reinstall the holder.
Failure to comply may cause serious injury or equipment damage.
A. Transducer holder
56Gas Ultrasonic Flow Meters
Meter repairs
WARNING!
CUTTING HAZARD
Sharp edges may be present on the meter.
Wear appropriate personal protective equipment when working on the meter. Failure to
comply may cause serious injury.
3.4.1Remove the transducer holder with the extractor tool
To replace transducer holder while under pressure, refer to Ultrasonic Split Clamp
Extractor Tool Operation Manual (Daniel P/N 3-9000-744).
Note
The extractor tool is used on Daniel Model 3414 Ultrasonic Flow Meters and is not applicable for use
with Daniel Model 3412 and Model 3411 Ultrasonic Flow Meters.
3.4.2Remove the transducer holder without the extractor
tool
Daniel 3410 Series Ultrasonic Gas Flow Meters utilize transducer holders that contain the
transducer assemblies and act as the pressure barrier between the transducers and the
fluid. Under normal maintenance such as transducer replacement, the transducer holders
do not need to be removed. If it is necessary to remove the transducer holders, the
following steps detail how to safely remove and reinstall them. Before removing and
installing the transducer holder, connect to the meter using Daniel MeterLink and collect
and save a Maintenance Log.
Procedure
1.Blow the line down according to the site standard operating procedures.
2.Ensure that the line pressure is down to atmospheric pressure prior to disassembly.
3.Disconnect the transducer cable nut. Pull the chordset from the transducer
assembly.
Note
Do not twist the chordset as you disconnect it from the transducer assembly (see Figure 3-6).
4.If the meter is equipped with T-21 or T-22 transducers, remove the transducer
retainer using a 1 1/8” wrench and then disconnect and remove the transformer
(Figure 3-4).
T-21 transducers use W-01 transformers and T-22 transducers use W-02
transformers.
Maintenance and Troubleshooting manual57
Meter repairs
5.Use a 1 1/4” (32 mm) wrench on the hex of the transducer holder and slowly
unscrew in a counterclockwise direction from the meter. If you hear gas leaking
from the threads, immediately stop and reinstall the holder as the meter has not
been fully drained and/ or pressure has not been relieved from the meter. Correct
the issue before attempting to remove the holder.
6.The transducer holder has now been removed from the meter with the transducer
still installed inside the transducer holder.
7.Make a note of the removed transducer holder length which is used to update the
meter configuration during the Transducer Swap-out procedure in Daniel MeterLink,
after all of the transducer holders are replaced.
Transducer holder length and set screw identificationFigure 3-8:
A. Transducer holder set screws
B. Transducer holder length identification
8.Loosen the three Allen setscrews with a 1/16” hex driver securing the transducer
assembly and stalk, if installed. Carefully remove the transducer by pulling it from
the T-Slot transducer holder (or stalk if installed) without rotating.
9.Clean the holder with a dry cloth.
58Gas Ultrasonic Flow Meters
Meter repairs
3.4.3Install the transducer holder with the extractor tool
To replace transducer holder while under pressure, refer to Ultrasonic Split Clamp
Extractor Tool Operation Manual (Daniel P/N 3-9000-744).
Note
The extractor tool is used on Daniel Model 3414 Ultrasonic Flow Meters and is not applicable for use
with Daniel Model 3412 and Model 3411 Ultrasonic Flow Meters.
3.4.4Install the transducer holder without the extractor tool
1.Ensure that the Daniel 3410 Series Ultrasonic Gas Flow Meter transducer port,
mount, and T-Slot transducer holder assembly are clean and free of debris.
2.Insert the transducer (parts are keyed and can only be assembled one way) into the
stalk or into the new transducer holder if no stalk is required. Do not use any
lubricant on the O-rings or contacts of the transducers
NOTICE
Ensure that the transducers identified as belonging to end 1 are installed on end 1 (A1,
B1, C1 or D1) of the meter holder and those identified as belonging to end 2 (A2, B2, C2
or D2) are installed on end 2 of the meter housing.
Transducer holder, stalk and transducer assemblyFigure 3-9:
A. Transducer holder
B. Stalk
C. Transducer assembly
3.Replace the O-rings and backup rings on the transducer holder. Make sure the
contoured side of the backup ring faces away from the transducer holder. It is highly
recommended that the O-rings be replaced when the transducer is removed from
the holder/stalk.
Maintenance and Troubleshooting manual59
Meter repairs
4.Use a 1/16” hex driver to equally tighten the three Allen set screws on the
transducer holder to secure the transducer assembly and the stalks (if installed).
5.Apply a light coat of Molykote 111
(1)
Silicone grease or equivalent to the transducer
holder O-rings.
6.Ensure that the transducer port, mount, and T-Slot transducer assembly are clean
and free of debris.
7.Apply a small amount of nickel anti-seize compound (P/N 2-9-9960-134) to the
outer threads of the transducer holder.
8.Insert the T-Slot transducer assembly into the meter transducer port. Tighten with
crescent wrench to securely seat the assembly in the mount. Do not over tighten.
9.Align the keyed chordset and securely seat into the transducer holder.
Note
The transducer chordset is keyed and will only go on one way.
Hand-tighten the transducer cable nut to secure the chordset onto the transducer
holder.
(1) Molykote 111 is a trademark of Dow Corning Corporation, U.S.A.
60Gas Ultrasonic Flow Meters
Meter repairs
T-Slot transducer holder removalFigure 3-10:
A. Base enclosure transducer cable glands
B. Cable ties (two locations)
C. 3410 Series Ultrasonic Meter transducer port
D. Transducer mount and transducer chordset
10.Repeat Step 1 through Step 9 for all transducer holders to be replaced.
11.Slowly repressurize the meter to line pressure. Check for leaks as the meter is
pressurized. If you hear gas leaking from the threads, recheck all connections and
resolve the problem. Then, slowly repressurize the meter to line pressure.
12.Connect to the meter with Daniel MeterLink and update the transducer parameters.
From the Tools|Transducer Swap-out menu, run the Transducer Swap-out
Wizard. This utility allows you to update the parameters for the components that
are replaced.
Note
Running the Transducer Swap-out utility is required when transducers, mounts, holders, or
stalks are replaced for a chord.
Maintenance and Troubleshooting manual61
Meter repairs
a. After writing all of the changes to the meter’s configuration, open the Monitor
(Detailed View) and verity the meter is acquiring data, the transducers have
good signals and flow profiles for the all of the chords displayed.
b. Collect and save a Maintenance Log and verify the meter is optimally performing.
Save the meter configuration file. If communicating with the meter via Modbus
or HART, manually update the parameters (see Section 3.3.5).
3.5Transducer cable removal and installation
Daniel 3410 Series Ultrasonic Gas Flow Meters have blue transducer cables and a
transducer cable nut that threads directly into the back of the transducer holder. The
cables use plastic glands that are installed on the meter’s Base Enclosure.
Note
Make a note of the exiting cabling path layout to allow proper tie-wrap configuration later in this
procedure.
62Gas Ultrasonic Flow Meters
Meter repairs
Figure 3-11:
Daniel 3410 Series Ultrasonic Gas Flow Meter transducer cables and
ports
3.5.1
A. Base enclosure transducer cable glands
B. Cable ties (two locations)
C. 3410 Series Ultrasonic Meter transducer port
D. Transducer mount and transducer cable nut and chordset (P/N 1-360-01-310)
WARNING!
CRUSHING HAZARD
During meter installation or removal, always place the unit on a stable platform or surface that
supports its assembled weight.
Failure to comply could allow the meter to roll, resulting in serious injury or equipment
damage.
Remove transducer cables
The meter body ports are identified by stamped or cast lettering adjacent to the
transducer port (i.e. Model 3414 - A1, A2, B1, B2, C1, C2, D1, and D2; Model 3412 - A1, A2,
B1, and B2; Model 3411 - A1 and A2).
Maintenance and Troubleshooting manual63
Meter repairs
Procedure
1.Remove power to the meter.
2.Unscrew the cable nut from the transducer holder and then, pull the cable chordset
from the transducer holder (see Figure 3-11).
3.Cut the tie wraps for the transducer cable you are replacing.
4.Use a wrench and loosen the cable gland (where the transducer cable enters the
Base Electronics Enclosure) until the transducer cable freely slides inside the gland.
5.Remove the two bolts holding the Base Enclosure cover to the Transmitter
Electronics Enclosure using a 6mm Allen wrench.
6.Lift the Transmitter Electronics Enclosure from the Base Enclosure. It may be
necessary to remove the ground lug wire and loosen the conduit connections prior
to removal.
7.Carefully prop the Transmitter Electronics Enclosure to the side.
8.Use a flat-blade screw driver and disconnect the transducer wiring terminal block
from the Acquisition Module (see Figure 3-12).
9.Place the Transmitter Electronics Enclosure in a clean dry area. Be careful not to
dislodge or damage the enclosure gasket.
10.Remove the transducer cable wires from the terminal block and slide the existing
cable through the Base Enclosure gland.
3.5.2
Install transducer cables
1.Use the existing cable and cut the new cable (Daniel P/N 1-360-01-310) for each
chordset to same length.
2.Insert the keyed chordset into the transducer holder. Make sure the keyed parts are
correctly aligned.
3.Screw the cable nut onto the transducer holder turing clock-wise until hand-tight.
Ensure the threads are correctly aligned and do not overtighten the cable nut.
4.Route the cable through the gland on the Base Enclosure and pull the cable up
through the Base Enclosure to allow enough slack to strip the cable wire.
5.Strip the outer insulation, outer shield, and inner insulation to just inside the cable
gland using a wire stripper. Verify that insulation of individual wires were not cut
while removing outer layers. Strip each wire 1/4 inch and wire them to the
Acquisition Module terminal block. Check the label number (i.e. A1) on the
Acquisition Module and match it with the label on the cable. Securely tighten the
mounting screws of terminal block J1 and J2 as shown in Figure 3-12.
64Gas Ultrasonic Flow Meters
Meter repairs
Model 3414 Ultrasonic Meter Acquisition Module wiringFigure 3-12:
6.Only connect wires for Chords A1, A2, B1 and B2 for 3412 meter applications or A1
and A2 for 3411 meter applications. The relative position of the contacts is shown
on the Acquisition Module label adjacent to the terminal block. When terminating
the connector wires, ensure that the contacts clamp on the bare wires and not on
the wire insulation. Leave the connector plugged into the Acquisition Module while
terminating the individual wires.
7.Once the transducers are wired correctly, tighten the cable gland so that the
transducer cable is held securely in place. Pull the cable back through the gland to
remove the slack and configure the cable to follow the same path of the existing
cable (seeFigure 3-11) and note in Section 3.5.
8.Repeat Step 1 through Step 7 if you are replacing other cables.
9.Once all of the cables are replaced, dress with tie wraps (P/N 2-4-9158-001) in
groups of two; A1 and C1, D1 and B1, A2 and C2, D2 and B2. Install one cable tie
three inches from the Base Enclosure and another near the point the cables start to
bend and separate out into their respective port (see Figure 3-11).
10.Inspect the Transmitter Electronics Enclosure gasket for wear and replace it if
necessary.
11.If replacing the gasket, lubricate it with Molykote 111
(2)
(P/N 2-9-9960-135).
12.Prop the Transmitter Electronics Enclosure at an angle on top of the Base Enclosure.
(2) Molykote 111 is a trademark of Dow Corning Corporation, U.S.A.
Maintenance and Troubleshooting manual65
Meter repairs
13.Plug the Acquisition cable terminal block to J3 on the Acquisition Module. Use a flat
blade screw driver and securely tighten the terminal block mounting screws to the
Acquisition Module.
14.Wrap the excess cable around the Acquisition Module below the lip of the Base
Enclosure (this prevents pinching the cable when the Transmitter Electronics
Enclosure is installed).
15.Attach one desiccant pack to the underside of the Base Enclosure cover.
16.Place the Transmitter Electronics Enclosure onto the Base Enclosure. Rotate the
Transmitter Electronics Enclosure until the mounting holes are correctly aligned
with the holes in the Base Enclosure.
17.Install the two hex head bolts with a 6 mm Allen wrench to secure the Transmitter
Electronics Enclosure to the Base Enclosure.
18.Reattach the external ground wire to the ground lug, reconnect the conduit, and
power the meter.
19.If required, install the security seal wire into and through one of the two holes in the
end cap. Choose holes that minimize counterclockwise rotation of the end cap when
the security wire is taut (maximum wire diameter 0.078 inch; 2.0 mm).
20.Adjust the security wire, removing all slack and thread into the lead seal.
21.Cut wire ends to remove excess wire.
22.If required, attach the security wire seals on the Base Enclosure.
66Gas Ultrasonic Flow Meters
Meter repairs
Base Enclosure security sealsFigure 3-14:
A. Transmitter Electronics Enclosure
B. Security wire seals
C. Transmitter Electronics endcap security latch
D. Base Enclosure
23.If required, install a security seal wire (maximum wire diameter 0.09 inch; 2.3mm)
on the Transducer Cable Connector to the Transducer Cable Nut by rotating the
Transducer Cable Nut clockwise; compressing the seal on the Transducer Cable
Connector (see Figure 3-15).
Maintenance and Troubleshooting manual67
Meter repairs
Transducer cable nut, chordset and security sealFigure 3-15:
3.6
A. Transducer cable nut
B. Transducer cable chordset
C. Security wire seal
24.Twist and adjust wire removing all slack and seal. Remove excess wire (see
Figure 3-15).
25.If required by the site operations manager, have an electrician fully test the
connections. After the Acceptance Test is witnessed and approved, seal the conduit.
26.Power down the system and apply the sealing compound to the conduit and allow
to set in accordance with manufacturer specifications.
Replace the meter electronics
The following procedure should be performed by a qualified service technician or trained
personnel. Observe all warning labels on the meter before starting this procedure.
The Daniel 3410 Series Gas Ultrasonic Flow Meter Transmitter Electronics Enclosure
consists of the following:
•CPU Module assembly (P/N 1-360-03-010)
•Optional I/O Module (RS-232 or RS-485)
•I.S. Barrier Board (P/N 360-03-004)
•Power Supply (P/N 360-03-003)
•Backplane Board (P/N 360-03-007)
68Gas Ultrasonic Flow Meters
Meter repairs
The Daniel 3410 Series Gas Ultrasonic Flow Meter Base Enclosure consists of the following:
•Acquisition Module (P/N 1-360-03-008)
•Acquisition Cable (P/N 360-01-595)
Should the Daniel 3410 Series Ultrasonic Gas Flow Meter require disassembly in the field
(i.e., check boards, change switch settings, or replace boards), to prevent electrostatic
damage to the electronic boards, always use a ground strap while handling the circuit
boards. If one is not available, make sure you are electrically discharged before touching
the boards by first touching a metal surface such as a ground lug on the meter or a table.
3410 Series electronicsFigure 3-16:
A. Terminal end of Transmitter Electronics Enclosure
B. Backplane board location
C. End cap security latch
D. Base enclosure with Acquisition Module
3.6.1
Maintenance and Troubleshooting manual69
Replace the CPU Module or Optional I/O Module
1.Remove power to the meter.
Meter repairs
2.Disconnect security seals on the Transmitter Electronics Enclosure (see Figure 3-11),
loosen the end cap security latches using a 3 mm Allen wrench (see Figure 3-12) and
remove both end caps from the Transmitter Electronics Enclosure.
3.If replacing the CPU Module (terminal end of the enclosure) or the Optional I/O
Module, use a 6 mm flat blade screw driver and disconnect the CPU Module terminal
blocks (or the optional I/O Module terminal blocks).
CPU or I/O Module replacementFigure 3-17:
A. Terminal end of Transmitter Electronics Enclosure
B. CPU Module
C. Optional I/O Module
D. Power Supply board
E. Fuse
F. Internal chassis ground
70Gas Ultrasonic Flow Meters
CPU or I/O Module replacementFigure 3-18:
A.CPU Module Type 2 - side label
Meter repairs
B. CPU Module Type 4 - side label
4.Grasp the outer ends of the module you want to replace and pull it out of the
enclosure.
5.Insert the new CPU Module or I/O Module into the enclosure and firmly push until
the board is fully seated into the Backplane Board connectors and the lock is
engaged.
6.Replace the terminal blocks for the CPU Module and/or the Optional I/O Module and
tighten the flat head screws with a 3 mm flat blade screw driver.
7.If you are not replacing other electronics, replace the end caps and security latches
(requires a 3 mm Allen wrench). If required, install the security seal wire into and
through one of the two holes in the end cap. Choose holes that minimize
counterclockwise rotation of the end cap when the security wire is taut (maximum
wire diameter .078 inch; 2.0 mm).
8.Adjust the security wire, removing all slack and thread into the lead seal.
9.Cut wire ends to remove excess wire.
10.If replacing other electronics or the fuse, continue with Section 3.6.2, Section 3.6.3
and Section 3.7 before replacing the end caps and sealing the enclosure.
11.If you encounter problems replacing the electronics, see the Lifecycle Customer
Service contact information on the back cover of this manual.
This completes the CPU Module or I/O Module replacement procedure.
Replace the Fuse
1.Remove power to the meter.
2.Disconnect the Transmitter Electronics Enclosure security seals (see Figure 3-13),
loosen the end cap security latch (requires a 3 mm Allen wrench) on the terminal
end of the enclosure (see Figure 3-16) and remove the end cap.
3.Use a 1/4 inch standard flat head screw driver and remove the Fuse Holder cap (see
Figure 3-17).
4.Insert the replacement fuse (Littlefuse #218002.HXP) into the Fuse Holder.
5.Install the fuse cap into the holder and push until it is flush with the holder.
6.Turn the fuse cap clockwise 1/8 turn using a 1/4 inch standard flat head screw driver.
7.If replacing other electronics, continue with Section 3.6.3, and Section 3.7 before
replacing the end caps and sealing the enclosure.
72Gas Ultrasonic Flow Meters
Meter repairs
8.If you are not replacing other electronics, replace the end caps and security latches
(requires a 3 mm Allen wrench). If required, install the security seal wire into and
through one of the two holes in the end cap. Choose holes that minimize
counterclockwise rotation of the end cap when the security wire is taut (maximum
wire diameter .078 inch; 2.0mm) (see Figure 3-19).
9.Adjust the security wire, removing all slack and thread into the lead seal.
10.Cut wire ends to remove excess wire.
11.Apply power to the meter.
This completes the fuse replacement procedure.
3.6.3Replace Backplane, I.S. Barrier or Power Supply board
The following sections detail removal of the Backplane board, the I.S. Barrier Board and the
Power Supply Board.
Replace the Backplane
1.If replacing the Backplane board, remove power to the meter.
2.Disconnect the Transmitter Electronics Enclosure security seals, loosen the end cap
security latches (3mm Allen wrench required) and remove both end caps (see
Figure 3-19).
3.Remove the CPU Module and the Optional I/O Module (if installed). See Figure 3-17
for board locations and associated terminal blocks.
4.Use a Phillips head screw driver and remove the four Backplane board screws and
captive star washers from the enclosure standoffs.
5.Pull the Backplane board out of the enclosure. This disconnects the I.S. Barrier
Board. Lay the Backplane board down with the Acquisition Cable still attached (the
Power Supply board may remain attached to the Backplane when you remove it
from the enclosure).
Maintenance and Troubleshooting manual73
Meter repairs
Figure 3-20:
A. Non-terminal end of Transmitter Electronics Enclosure
B. Power Supply board
C. I.S. Barrier board (inside the Guide Plate)
D. Acquisition cable
E. Backplane board
6.Use a 3 mm flat head screw driver and disconnect the Acquisition Cable terminal
block from the Backplane. Unplug the Acquisition Cable from the Backplane.
7.Remove the Power Supply (if it was not removed with the Backplane board) and I.S.
Barrier boards from the enclosure. The I.S. Barrier Board has a notched tab that
secures the board to the Guide Plate.
8.Attach the Acquisition Cable terminal block to the new Backplane Board and plug
the Power Supply Board and I.S. Barrier board into the Backplane board.
9.Insert the Backplane (with the Power Supply and I.S. Barrier Boards attached to the
Backplane) into the enclosure.
74Gas Ultrasonic Flow Meters
Meter repairs
10.Fully seat the CPU Module and Optional I/O Module onto the Backplane board.
11.Install the four Phillips head screws to secure the Backplane to the enclosure
standoffs.
12.Reinstall the terminal blocks on the CPU Module, Optional I/O Module (if installed),
and the Power Supply board using a 3 mm flat head screw driver.
NOTICE
Ensure the terminal blocks are aligned with the Guide Plate openings.
13.Recheck the connections, wiring and switch settings before replacing the end caps.
14.If replacing other electronics, continue with the following sections before replacing
the end caps and sealing the enclosure.
15.If you are not replacing other electronics, replace the end caps, security latches,
reseal the meter and apply power. If required, install the security seal wire into and
through one of the two holes in the end cap. Choose holes that minimize
counterclockwise rotation of the end cap when the security wire is taut (maximum
wire diameter.078 inch; 2.0 mm) (see Figure 3-13).
16.Adjust the security wire, removing all slack and thread into the lead seal.
17.Cut wire ends to remove excess wire.
18.Apply power to the meter.
19.This completes the Backplane Board replacement procedure.
20.If you encounter problems with this procedure, see the Lifecycle Customer Service
contact information on the back cover of this manual.
Replace the I.S. Barrier Board
1.If replacing the I.S. Barrier board, remove power to the meter.
2.Disconnect the Transmitter Electronics Enclosure security seals, loosen the end cap
security latches with a 3mm Allen wrench and remove both end caps (see
Figure 3-13).
3.Use a 3 mm flat head screw driver and remove the terminal blocks from the Power
Supply board, the CPU Module and the Optional I/O Module (if installed). See
Figure 3-17 for board locations and associated terminal blocks.
4.Use a Phillips head screw driver and remove the four Backplane board screws from
the enclosure standoffs. If the Local Display Module is installed on the Backplane,
use a flat blade screw driver and remove the four flat-head screws from the
standoffs.
5.Pull the Backplane board out of the enclosure. This disconnects the I.S. Barrier
Board. Lay the Backplane board down with the Acquisition Cable still attached (the
Power Supply board may remain attached to the Backplane when you remove it
from the enclosure).
Maintenance and Troubleshooting manual75
Meter repairs
I.S. Barrier board replacementFigure 3-21:
A. Non-terminal end of Transmitter Electronics Enclosure
B. Power Supply board
C. I.S. Barrier board (inside the Guide Plate)
D. Acquisition cable
E. Backplane board
6.Remove the I.S. Barrier Board from the Guide Plate on the right side of the
enclosure.
7.Install the new I.S. Barrier board onto the Backplane Board and seat the Power
Supply board onto the Backplane board.
8.Insert the Backplane, I.S. Barrier board and the Power Supply Board into the
enclosure.
9.Fully seat the CPU Module and Optional I/O Module onto the Backplane Board.
76Gas Ultrasonic Flow Meters
Meter repairs
10.Attach the Backplane to the enclosure standoffs with the four Phillips head screws. If
the Local Display Module is installed on the Backplane, use a flat blade screw driver
and install the four flat-head screws into the enclosure standoffs.
11.Reinstall the J7 terminal block, if removed, on the Backplane board. Re-install the
CPU Module, Optional I/O Module (if installed) and the Power Supply.
12.Recheck the connections, wiring and switch settings before replacing the end caps.
13.If replacing other electronics, continue with the following procedures before
replacing the end caps and sealing the enclosure.
14.If you are not replacing other electronics, replace the end caps and security latches
(3 mm Allen wrench required). If required, install the security seal wire into and
through one of the two holes in the end cap. Choose holes that minimize
counterclockwise rotation of the end cap when the security wire is taut (maximum
wire diameter .078 inch; 2.0 mm).
15.Adjust the security wire, removing all slack and thread into the lead seal.
16.Cut wire ends to remove excess wire.
17.Apply power to the meter.
18.This completes the I.S. Barrier Board replacement procedure.
19.If you encounter problems with this procedure, see the Lifecycle Customer Service
contact information on the back cover of this manual.
Replace the Power Supply Board
1.If replacing the Power Supply board remove power to the meter.
Maintenance and Troubleshooting manual77
Meter repairs
2.Disconnect the Transmitter Electronics Enclosure security seals, loosen the end cap
security latches with a 3mm Allen wrench and remove both end caps (see
Figure 3-19).
3.Use a flat head screw driver and remove the terminal blocks from the Power Supply
board, the CPU Module and the Optional I/O Module (if installed). See Figure 3-17 for
board locations and associated terminal blocks.
4.Use a Phillips head screw driver and remove the four Backplane board screws from
the enclosure standoffs. If the Local Display Module is installed on the Backplane,
use a flat blade screw driver and install the four flat-head screws into the enclosure
standoffs.
78Gas Ultrasonic Flow Meters
Meter repairs
Power Supply board replacementFigure 3-23:
A. Non-terminal end of Transmitter Electronics Enclosure
B. Power Supply board
C. I.S. Barrier board (inside the Guide Plate)
D. Acquisition cable
E. Backplane board
5.Pull the Backplane board out of the enclosure. This disconnects the I.S. Barrier
Board. Lay the Backplane board down with the Acquisition Cable still attached (the
Power Supply board may remain attached to the Backplane when you remove it
from the enclosure).
6.Plug the new Power Supply board and the I.S. Barrier Board onto the Backplane
Board.
7.Insert the Backplane, I.S. Barrier board and the Power Supply Board into the
enclosure and fully seat the CPU Module and Optional I/O Module.
Maintenance and Troubleshooting manual79
Meter repairs
8.Attach the Backplane to the enclosure standoffs with the four Phillips head screws. If
the Local Display Module is installed on the Backplane, use a flat blade screw driver
and install the four flat-head screws into the enclosure standoffs.
9.Use a flat blade screw driver and install the terminal blocks on the CPU Module,
Optional I/O Module, I.S. Barrier Board and the Power Supply.
10.Recheck the connections, wiring and switch settings before replacing the end caps.
11.If replacing other electronics, continue with Section 3.7 before replacing the end
caps and sealing the enclosure.
12.If you are not replacing other electronics, replace the Transmitter Electronics
Enclosure end caps, install the end cap security latches (3 mm Allen wrench
required). If required, install the security seal wire into and through one of the two
holes in the end cap. Choose holes that minimize counterclockwise rotation of the
end cap when the security wire is taut (maximum wire diameter .078 inch; 2.0 mm)
(see Figure 3-21).
13.Apply power to the meter.
14.This completes the Power Supply Board replacement procedure.
15.If you encounter problems with this procedure, see the Lifecycle Customer Service
contact information on the back cover of this manual.
3.7Replace the Adquisition Module
WARNING!
CRUSHING HAZARD
During meter installation or removal, always place the unit on a stable platform or surface that
supports its assembled weight.
Failure to comply could allow the meter to roll, resulting in serious injury or equipment
damage.
Procedure
1.Remove power to the meter.
2.If the installation has rigid conduit, use a medium size crescent wrench and loosen
the hex nuts on the Transmitter Electronics Enclosure. This should allow enough
slack to remove the Transmitter Electronics Enclosure from the Base Enclosure. If
the installation uses flexible conduit, you may not need to disconnect it from the
Transmitter Electronics Enclosure.
80Gas Ultrasonic Flow Meters
Conduit removalFigure 3-24:
A. Transmitter electronics enclosure
Meter repairs
B. Conduit nuts
3.If the meter is equipped with security seals, remove the seals from the bolts on the
Base Enclosure.
Maintenance and Troubleshooting manual81
Meter repairs
Figure 3-25:
Transmitter Electronics Enclosure and Base Enclosure security seal
removal
A. Transmitter electronics enclosure
B. Base enclosure bolts and security seals
4.Use a 6 mm Allen wrench and remove the two hex head bolts securing the
Transmitter Electronics Enclosure to the Base Enclosure.
5.Use a 3 mm flat head screw driver and disconnect the Acquisition cable terminal
block and the transducer wire terminal blocks from the Acquisition Module inside of
the Base enclosure.
Maintenance and Troubleshooting manual83
Meter repairs
Acquisition Module cable and transducer wiringFigure 3-27:
A. Base Enclosure gasket
B. Acquisition cable
C. Acquisition wiring terminal blocks
D. Acquisition Module
NOTICE
Ensure that the transducers identified as belonging to end 1 are installed on end 1 (A1,
B1, C1 or D1) of the meter holder and those identified as belonging to end 2 (A2, B2, C2
or D2) are installed on end 2 of the meter housing.
6.Remove the three Acquisition Module flat head screws and split lock washers, then
remove the Acquisition Module from the Base Enclosure.
7.Insert the new Acquisition Module into the Base Enclosure and secure with the three
split lock washers and flat head screws.
8.Reattach the terminal blocks onto the Acquisition Module (3 mm flat head screw
driver required) for the corresponding chord (3414 four path meters A1, A2, B1, B2
C1, C2, D1, and D2; 3412 two path meters A1, A2, B1, B2; or 3411 single path
meters A1 and A2). Make sure the transducer wires have good contact with the
terminal block and the terminal block screws are tight.
9.When you have completed attaching the Transducer wire terminal blocks and the
Acquisition cable terminal block to the Acquisition Module, check the Base
Enclosure o-ring and reinstall if necessary.
84Gas Ultrasonic Flow Meters
Meter repairs
10.Reattach the Transmitter Electronics Enclosure to the Base Enclosure with the two
hex head bolts and lock washers. Tighten bolts with a 6 mm Allen wrench.
11.Retighten or reattach the conduit to the Transmitter Electronics Enclosure using a
crescent wrench or channel lock pliers.
12.If required, install security wire seal into and through the hole in the socket head
screw on the Base Enclosure cover (maximum wire diameter 0.078 inch; 2.0mm).
Base Enclosure wire seal installationFigure 3-28:
A. Base Enclosure cover
B. Security wire seals (optional)
13.Position the wire to prevent counterclockwise rotation of the screws when the seal
wire is taut.
14.Feed the security wire beneath the Transmitter Electronics Enclosure and through
the adjacent socket head screw. Twist the wire, removing all slack and seal.
Maintenance and Troubleshooting manual85
Meter repairs
Base Enclosure security sealsFigure 3-29:
A. Transmitter Electronics Enclosure
B. Security wire seals (optional)
C. Transmitter Electronics endcap security latch
D. Base Enclosure
15.Cut wire ends to remove excess wire.
16.Apply conduit sealing compound according to manufacturer’s recommendations.
17.Apply power to the meter.
This completes the Acquisition Module replacement procedure.
86Gas Ultrasonic Flow Meters
Conversion factors
Appendix A
Conversion factors
Topics covered in this appendix:
Conversion factors per units of measurement
•
K-Factor and inverse K-Factor
•
A.1Conversion factors per units of measurement
The following table includes conversion factors for many of the Metric and U.S. Customary
units of measure used with Daniel 3410 Series Ultrasonic Gas Flow Meters and Daniel
MeterLink.
Conversion factors per units of measurementTable A-1:
Conversion factorsUnit of measurement
(°F-32)x(5/9)—>°C
(°C+273.15)—>K
1K/°C
5/9°C/°F
-6
10
0.006894757MPa/psi
0.1MPa/bar
0.101325MPa/atm
0.000133322MPa/mmHg
0.3048m/ft
0.0254m/in
3
10
-6
10
(0.3048)
(0.0254)
3600s/h
86400s/day
10
0.45359237kg/lbm
231in3/gal
42gal/bbl (barrel)
0.0037854gal/m
3
3
3
MPa/Pa
3
dm3/m
m3/cc (=m3/cm3)
3
m3/ft
3
m3/in
g/kg
3
Maintenance and Troubleshooting manual87
Conversion factors
Conversion factors per units of measurement (continued)Table A-1:
Conversion factorsUnit of measurement
6.289811bbl/m
-3
10
1.488Pa•s/(lb/(ft•s))
3
Pa•s/cPoise
A.2K-Factor and inverse K-Factor
Frequency volumetric flow rate K-Factor Equation A-1:
=
and
3600ℎ
*
=
where
KFactor
InvKFactor
FreqQ
FullScale
MaxFreq
* TimeUnit
** Volume
Frequency volumetric flow rate inverse K-Factor Equation A-2:
3600ℎ
frequency “K-Factor” (pulses/volume**) (Freq1KFactor and Freq2KFactor)
frequency “Inverse K-Factor” (volume**/pulse) (Freq1InvKFactor and
Freq2InvKFactor)
frequency full-scale volumetric flow rate (volume**/time unit*)
(Freq1FullScaleVolFlowRate and Freq2FullScaleVolFlowRate)
maximum frequency (Hz = pulses/time unit*) (Freq1MaxFrequency and
Freq2MaxFrequency)
time conversion factor depends on the VolFlowRate Time Unit data point:
•volume/second = 1 s/s
•volume/minute = 60 s/m
•volume/hour = 3600 s/h
•volume/day = 86400 s/d
where the volume is selected via data points:
•Units System
•VolUnitUS
-gallons
*
88Gas Ultrasonic Flow Meters
-barrels
•VolUnitMetric
-cubic meters
-liters
Conversion factors
Maintenance and Troubleshooting manual89
Conversion factors
90Gas Ultrasonic Flow Meters
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.