Emec KMS MF, KMSA MF, KMS AC MF, VMS MF, VMSA MF Operating Manual

SOLENOID DRIVEN METERING PUMPS
WITH DIAPHRAGM
EN
KMS MF - KMSA MF - KMS AC MF
OPERATING MANUAL
KMS MF
KMS AC MF
PRODUCT LABEL
This operating instructions contains safety information that if ignored can endanger life or result in serious injury.
Read these instructions carefully before use and keep them for future reference. Information and specifications on this manual could be uncorrect or could have printing errors. Specifications are subject to change without notice.
Version: R1-01-17
Danger!
Indicates a hazardous situation which, if not avoided, will result in death or serious injury.
Warning!
Indicates a hazardous situation which, if not avoided, could result in death or serious injury.
Important - A practice not related to personal injury or additional information.
Cross reference - An instance which refers to related information elsewhere in the same
document
GENERAL SAFETY GUIDELINES
Operating, installing, or maintaining the unit in any way that is not covered in this manual could cause death, serious personal injury, or damage to the equipment.
This manual use the following safety message icon:
ICON
KMS metering pump is tested and certified by WQA to NSF/ANSI 50 and 61 for materials safety.
NORME CE EC RULES (STANDARD EC) NORMAS DE LA CE
Direttiva Bassa Tensione Low Voltage Directive Directiva de baja tensión
Direttiva EMC Compatibilità Elettromagnetica EMC electromagnetic compatibility directive EMC directiva de compatibilidad electromagnética
Norme armonizzate europee nell’ambito della direttiva European harmonized standards underdirective Las normas europeas armonizadas conforme a la directiva
2014/35/UE
2014/30/UE
2006/42/CE
PURPOSE OF USE AND SAFETY
METERING PUMP IS INTENDED FOR CHEMICAL DOSING AND DRINKING WATER TREATMENT.
Do not use in explosive area (EX). Do not use with flammable chemicals. Do not use with radioactive chemicals.
Use after a proper installation.
Use the pump in accordance with the data and specifications printed on the label.
Do not modify or use in a manner inconsistent with the provisions of the operating manual.
Keep the pump protected from sun and water. Avoid water splashes.
In emergencies the pump should be switched off immediately. Disconnect the power cable from the power supply.
When using pump with aggressive chemicals observe the regulations concerning the transport and storage of aggressive fluids.
When installing always observe national regulations.
Manufacturer is not liable for any unauthorized use or misuse of this product that may cause injury, damage to persons or materials.
Pump must be accessible at all times for both operating and servicing. Access must not be obstructed in any way.
Feeder should be interlocked with a no-flow protection device.
Pump and accessories must be serviced and repaired by qualified and authorized personnel only.
Before any operation:
always read chemical Material Safety Data Sheet (MSDS);
always wear protective clothing;
always discharge the liquid end before servicing the pump.
empty and rinse the liquid end before work on a pump which has been used with hazardous or unknown chemicals.
This equipment requires regular maintenance to ensure potability requirements of the water and maintenance of improvements as declared by the manufaturer.
Work area
Always keep the pump area clean to avoid and/or discover emissions.
Recycling guidelines
EWC code: 16 02 14
Always recycle according to these guidelines:
1. If the unit or parts are accepted by an authorized recycling company, then follow local recycling
laws and regulations.
2. If the unit or parts are not accepted by an authorized recycling company, then return them to
the nearest representative.
Waste and emissions regulations
Observe these safety regulations regarding waste and emissions:
• Dispose appropriately of all waste.
• Handle and dispose of the dosed chemical in compliance with applicable environmental
regulations.
• Clean up all spills in accordance with safety and environmental procedures.
• Report all environmental emissions to the appropriate authorities.
ENVIRONMENTAL SAFETY
LABEL
Distributor
CODE: pump code MODEL: pump model
PUMP’S DATA
S/N: serial number
Data matrix
For spare parts orders or any other communication, refer to the pump’s label. Code (CODE) and serial number (S / N) uniquely identify the pump.
SPARE PARTS
Figura 1. Product label.
Figura 2. WQA label.
THIS METERING PUMP IS TESTED AND CERTIFIED BY WQA TO NSF/ANSI 50 AND 61 FOR MATERIALS SAFETY.
QR code
TRANSPORTATION AND STORAGE
A not suitable transportation or storage can cause damages.
Use origianal box to pack the pump.
Observe storage conditions also for transportation.
Although packed, always protect the unit against humidity and the action of chemicals.
Before return the dosing pump to the manufacturer Repair service, drain the chemical from pump head and rinse it. Refer to Shutdown procedure.
Fill the PRODUCT SERVICE REPAIR FORM and send it with the dosing pump. Repair service is not accepted if PRODUCT SERVICE REPAIR FORM is missing.
DO NOT TRASH PACKAGING. USE IT TO RETURN THE PUMP.
Transportation and storage temperature ..... 10–50°C (32–122°F)
Umidity ...................................................... 95% relative humidity (not condensed)
6
1. INTRODUCTION
1.1 KMS MF Series KMS MF is designed for low/middle dosing of chemicals.
The pump has different working modes: Constant, Divide, Multiply, ppm, perc, mlq, batch, volt, mA.
KMS MF has got:
- STAND-BY input
- SEFL (flow sensor) input
- LEVEL input
- ALARM contact output.
Flow rate is determined by the stroke length and by the stroke speed. The stroke length is adjustable from 0 to 100% using the stroke length adjustment knob. However dosing accuracy is guarantee within an adjustment range from 30% to 100%.
All control and setup parameters are available through a digital keyboard and they are displayed on a LCD backlit display.
Note: some functions described into this manual may need accessories not included into the
pump packaging.
1.2 KMSA MF Series KMSA MF is the KMS MF version with self-venting pump head.
Self-venting pump head must be used when using chemicals that produce gas (i.e. hydrogen peroxide, ammonium, sodium hypoclorite at particular conditions).
For connections Self-Venting pump head installation”.
1.3 KMS AC MF Series KMS AC MF is the KMS MF version with double supply: compressed air and power supply. Compressed air without lubrifiant and/or condensed water. Air supply pressure range must be from 6 to 10 bar.
For connections p. 5.
Pump can work in differents ways:
MODE WORKING MODES
CONSTANT
Pump doses at a constant rate set in “SPH” (strokes for hour), “SPM” (strokes for minute) or “LPH” (litres per hour) parameters set during program session.
DIVIDE
External pulses from a water meter are divided by a value set during program session. The pump doses with a rate determined by this parameter.
MULTIPLY
External pulses from a water meter are multiplied by a value set during program session. The pump doses with a rate determined by this parameter.
PPM
Dosing rate is determined by pulses from a water meter on the base of set PPM, chemical product concentration (%) and quantity for each single stroke set during program session.
PERC
Dosing rate is determined by pulses from a water meter on the base of set PERC (%), chemical product concentration (%) and quantity for each single stroke set during program session.
MLQ
Dosing rate is determined by pulses from a water meter on the base of set MLQ (milliliters per quintal), chemical product concentration (%) and quantity for each single stroke set during program session.
BATCH Signal from an external contact starts the pump to dose the set quantity.
VOLT
Voltage from an external device drives the pump that doses proportionally using a minimum and maximum of strokes for minute set during program session (0–10 VDC).
mA
Current from an external device drives the pump that doses proportionally using a minimum and maximum of strokes for minute set during program session.
In MULTIPLY, DIVIDE, PPM, PERC, MLQ working modes, the pump, connected to a pulse emitter water
meter, shows the instant flow.
1.4 Working modes
NSF/ANSI 61
NSF/ANSI 61
7
2. UNPACKING
Included into package:
n.4 Dibbles ø6 n.4 Self tapping screws 4,5 x 40 n.1 Delayed fuse 5 X 20 n.1 Foot filter with valve n.1 Injection valve n.1 Level probe m 2 Delivery pipe * (PVDF) m 2 Suction pipe * (transparent PVC) m 2 Venting pipe (transparent PVC) m 2,5 Signal cable for “Stand-by” and “Alarm” n.1 This installation manual
* If hose is 6x8 there is only a 4meters long hose. Cut to obtain suction and delivery hoses.
PLEASE DO NOT TRASH PACKAGING.
IT CAN BE USED TO RETURN THE PUMP.
LEGEND:
a. Alternating Current;
b. DC,
c. Protective Earth;
d. Standby;
e. Warning -
8
3. PUMP’S DESCRIPTION
KMS MF (standard PVDF pump head)
Stroke length knob
Level probe input
External signal input
Power supply
Flow sensor input (SEFL)
Delivery connection
Venting knob
Venting connection
Suction connection
Compressed air input (mod. KMS AC MF)
STAND-BY / ALARM input
Venting valve (mod. KMS AC MF)
Manual stroke length adjustment
Max CC/stroke ( Construction Materials and Technical info) are referred to Stroke length
knob on 100%. If Stroke length knob is on 50% cc/stroke will be halved. To regulate pump’s capacity: turn on the pump then press and rotate the knob.
Dosing accuracy is guarantee within an adjustment range from 30% to 100%.
Note: if knob isn’t on 100% position then the pump will dose at a pressure greater than the one declared on label.
9
KMSA MF (standard PVDF pump head)
Stroke length knob
Level probe input
External signal input
Power supply
Flow sensor input (SEFL)
Delivery connection
Venting connection
Suction connection
Compressed air input (mod. KMSA AC MF)
STAND-BY / ALARM input
Venting valve (mod. KMSA AC MF)
10
4. BEFORE TO INSTALL WARNINGS
Pump’s installation and operativity is made in 4 main steps:
Pump’s installation
Hydraulic Installation (hoses, level probe, injection valve)
Electrical Installation (main power connection, SEFL installation, priming)
Programming the pump.
Before to start, please read carefully the following safety information.
Protective clothes
Wear always protective clothes as masks, gloves, safety glasses, ear plugs or ear muffs and further security devices during ALL installation procedure and while handling chemicals.
Installation location
Pump must be installed in a safety place and fixed to the table / wall to avoid vibration problems!
Pump must be installed in a easy accessible place!
Pump must be installed in horizontal position!
Avoid water splashes and direct sun!
Hoses and Valves
Suction and delivery hoses must be installed in vertical position! All hoses connections must be performed using only hands’ force! No tongs required!
Delivery hose must be firmly fixed to avoid suddenly movements
that could damage near objects!
Suction hose must be shorter as possible and installed in vertical position to avoid air bubbles suction!
Use only hoses compatibles with product to dose! See chemical compatibility table. If dosing product is not listed please consult full compatibility table or contact chemical’s manufacturer!
Feeder should be interlocked with a no-flow protection device to automatically shut-off the pumps when there is no flow!
Adequate measures shall be taken to prevent cross connection of chemicals!
Chemical feeding must be stopped during backwash cycles and periods of noflow as these conditions may introduce the potential for chemical overdosing. Not doing so may result in elevated chemical concentrations and hazerdous gas introduction into the pool or spa.
11
5. INSTALLATION DRAW
Pump must be installed in a stable support (for example a table) at a maximum height (from tank’s bottom) of 1,5 meters.
1 - Dosing Pump 2 - Suction Hose 3 - Delivery Hose 4 - Injection Valve 5 - Air venting 6 - Level Probe 7 - Foot Filter 8 - Power Cable 9 - Standby / Alarm
1
2
3
4
5
6
7
8
9
12
6. HYDRAULIC INSTALLATION
Suction Hose
Tightening Nut
Holding Ring
Pipe Holder
O-ring
Valve
Hydraulic connections are:
Suction Hose with level probe and foot filter
Delivery Hose with injection valve
Venting Hose
Suction Hose.
Completely unscrew tightening nut from pump’s head and remove assembling components: tightening nut, holding ring and pipe holder.
Assembly as shown in fig. (A). Insert hose into pipe holder until it reaches the bottom.
Lock hose on pump’s head by screwing down the tightening nut. Use only hands to do it!
Connect other side of the hose to the foot filter using the same procedure.
fig. (A)
13
Assembling foot filter with level probe.
Level probe must be assembled with foot filter using the provided kit. Foot valve is made to be installed into tank’s bottom without sediments priming problem.
Connect BNC from level probe into pump’s level input (front side of the pump). Put level probe assembled with foot filter into tank’s bottom.
Warning: If there is a mixer installed into tank, install a suction lance
instead of level probe / foot filter.
Delivery Hose.
Completely unscrew tightening nut from pump’s head and remove assembling components: tightening nut, holding ring and pipe holder.
Assembly as shown in fig. (A). Insert hose into pipe holder until it reaches the bottom.
Lock hose on pump’s head by screwing down the tightening nut. Use only hands to do it!
Connect other side of the hose to the injection valve using the same procedure.
STEP 5
STEP 2
INSERT FLOATER
STEP 1
INSERT RING AS SHOWN
STEP 4
INSERT RING AS SHOWN
STEP 3
INSERT PROBE
WITH N.O. CONTACT
UNTIL TO HEAR
A CLICK
fig. (B)
14
Venting hose.
Insert one side of venting hose into venting connector as shown in fig (C).
Insert other side of venting hose into product’s tank. During priming procedure product exceeding will flow into tank. For priming procedure see the paragraph “Priming”.
Flow direction is indicated by the arrow on the valves.
fig (C): pump head connections.
fig (C1): PMMA pump head connections.
Injection Valve.
Injection valve must be installed on plant from water’s input. Injection valve will open at pressure greater than 0,3bar.
Venting knob
Venting knob
to suction hose
to suction hose
1
2
3
4
5
federico.renzi
10/04/2008
Progettato da Controllato da Approvato da
Data
Data
to delivery hose
to delivery hose
to manual venting hose
to manual venting hose
15
Self-venting pump head.
Self-venting pump head must be used when using chemicals that produce gas (i.e. hydrogen peroxide, ammonium, sodium hypoclorite at particular conditions).
Hoses assembling procedure (including purge hose) is described in fig. (A).
Notes:
- suction, delivery and purge valves are DIFFERENT! Do not exchange them!
- delivery and purge hoses are made of same material!
- it’s allowed to lightly bend venting hose!
- during calibration procedure (“TEST”) insert venting hose into BECKER test-tube!
1
2
3
4
5
to venting hose
to venting hose
to suction hose
to suction hose
to delivery hose
to delivery hose
fig (D): pump head connections. fig (D1): PMMA pump head connections.
Flow direction is indicated by the arrow on the valves.
16
7. ELECTRICAL INSTALLATION
All electrical connections must be performed by AUTHORIZED AND QUALIFIED personnel only. Before to proceed, please, verify the following steps:
- verify that pump’s label values are compatible with main power supply.
- pump must be connected to a plant with a differential switch (0,03A
sensitivity) if there isn’t a good ground.
- to avoid damages to the pump do not install it in parallel with heavy
inductance load (for example: engines). A relay switch must be used. See below picture.
P - Dosing Pump
R - Relay
I - Switch or safety device
E - Electrovalve or inductance load
A - Main Power
WARNING
IF EQUIPMENT IS NOT SUPPLIED WITH A PLUG:
a) a switch or circuit-breaker shall be included in the building installation
b) it shall be in close proximity to the equipment and within easy reach of the operator
c) it shall be marked as the disconnetting device for the equipment
WARNING
IF EQUIPMENT IS SUPPLIED WITH A PLUG:
If an appliance coupler or separable plug is used as the disconnecting device, it shall be readily identifiable
and easily reached by the operator. For single-phase portable equipment, a plug on a cord of length not
greater than 3m is considered to be easily reached.
17
Once verified previous steps proceed as follows:
- check that “BNC” of level probe has been connected as described in “Hydraulic
Installation” chapter.
- connect “BNC” and external signal to pump’s “INPUT” connectors.
-braided shield cable; +center conductor
This input may be used as follows:
- as pulse sender water meter or
- as startup contact for “BATCH” mode or
- as voltage input for “VOLT” mode or
- as current input for “mA” mode
- connect alarm and/or stand-by signal as described below fig (E):
- proceed to “SEFL” connection (Flow sensor is optional) as described in page 56.
fig (E)
1
4
1: (Green) “Stand-by” Input 4: (Yellow) “Stand-by” Ground
2: (Brown) Alarm (Relay) Common 3: (White) N.O. Alarm(Relay) Contact
Max load relay output: free contact 1A 24VAC
2
3
Notes: - “Alarm” signal isn’t fuse protected
- “Standby” signal has main priority on pump’s enabling / disabling.
External Signal Input
Level Probe Input
IF NOT USED, PROTECT THE MINI DIN PLUG WITH THE BLACK RUBBER CAP LOOSE IN THE ACCESSORIES BAG.
white - signal
yellow - ground
brown - + 12V
Connection to water meter with HALL effect (option)
Metering pumps for connection to an Hall effect water meter have got a three wires signal cable. If the water meter is equipped with the pump, there is a MPM plug to connect the pump.
Refer to Main board connection.
Wires cable connections
18
8. BASIC SETTINGS
LCD backlit display
Storke Lenght Knob
Scroll and increase digit
Turn the pump on or off and exit from setup menu (without saving parameters)
Enter / exit from setup menu (saving parameters)
The “KMS MF” pump is equipped with a keyboard.To avoid any misunderstanding during next chapters all keys will be described as shown on this legend:
“UP” key “ESC” key
“RIGHT” key “E” key
Menu navigation:
To enter into programming mode press and keep pressed “E” key from main screen (fig.3):
Main screen (fig.3) may appear different if “PPM” or “BATCH” mode is enabled. After about 4 seconds the pump will show the password screen (fig.5):
Default password is “0000”. Just press “E” key. Otherwise insert password using “UP” and “RIGHT” keys.
E
ESC
fig.3
password
0000
fig.5
ON/OFF
STROKES 100 SPM
batch
waiting
upkeep
000m 00s
19
OFF
------------------
Saving / Discarding changes / Activating working mode
Once edited data into setup menu it’s possible to save them by pressing “E” key or to discard them by pressing “ESC” key.
To activate a working mode (Constant, Divide, Multiply, PPM, PERC, MLQ, Batch, Volt, mA) select the required mode and confirm it using “E” key.
Turning on and off the pump
“ESC” key has a double function. It can be used to discard all changes made into setup mode or to turn on/off the pump. To
turn on/off the pump press and keep pressed this key while in main screen (fig.3). The pumps will show:
To return into operating mode press “ESC” key.
Alarm output logical working.
The pump has an “Alarm” output that changes its status (from N.O. to N.C. or viceversa) when a signal is received from “LEVEL” and/or “SEFL” and/or “STAND-BY”. To set this alarm refer to related chapter.
Full menu / Short Menu mode
When entering into SETUP display shows access mode menu:
If this is the first time into SETUP menu then the pump will automatically set itself into “FULL” menu mode as shown in fig. A. Just press “E” key to confirm. This mode will show all pump functions and working modes.
Next time the SETUP menu will be reached it will possible to operate with “SHORT” menu configuration to change only selected mode parameters as shown in fig. B. Press “E” to confirm.
Note: “SHORT” menu option is not available during first time into SETUP menu or after a reset.
fig.6
FULL
MENU
SHORT
MENU
fig. A fig. B
20
9. PRIMING
the pump will show the “Delay” (pump’s activation delay) as shown fig.2:
Press any key to skip the “Delay”. Pump will show “Srokes” (actual strokes) as shown in fig.3:
In any working mode, if a SEFL is installed and enabled (see SEFL Setup procedure), the display will show the icon (asterisk as in fig. 4):
- if SEFL works correctly, the asterisk blinks to any pulses given by the solenoid;
- if the asterisk does not appear, there is an anomaly (i.e.: hoses and/or valves are obstructed, SEFL in not connected, etc.).
Press and keep pressed the “RIGHT” key to enter into priming mode. Pump will go for 30 seconds into priming mode as shown in fig.5.
When the chemical begins to flow out from the outgassing hose then completely close the outgassing knob (except for self­venting pump heads). This ends the priming procedure. If countdown for priming is not yet ended press “ESC” key.
Now the pump is operative. Proceed to setup and programming.
fig.1
fig.2
fig.3
fig.4
PRIMING
To proceed follow these steps: connect all hoses to the pump; open dischage valve by completely turning the discharging knob (counter clock-wise). Power up the pump and turn stroke lenght knob to 100%. After pump’s intro (fig.1):
*
KMS MF
R: 1.xx
WAITING
00:59
STROKES 100 SPM
STROKES 100 SPM
PRIMING
30 Sec.
fig.5
21
10. PUMP’S FUNCTIONS SUMMARY
Pump’s functions summary
During pump’s working mode is it possible to see furthers working information. Press more times the “UP” key to cycle through following information:
key
key
key
Working Mode
Power Supply Voltage
key
key
divide
000.10
key
cc/st
0.20
External signal value (in Volt and mA working mode) or actual system flow (Multiply, Divide and PPM working mode). This item is not present in Constant working mode.
key
STROKES
100 SPM
MODE
divide
SUPPLY
220 VAC
DOSING
0.000 lph
input
196 lph
Liters/Hour or Gallons/ Hour Dosage
Loading...
+ 47 hidden pages