E.M.E. M672B, M672P User manual

Mk II Infant FlowTM
Driver
Models M672B and M672P
Software Versions: V1.22 and all V2.x.x
Ref.:
Service
672-81
Issue
Service
Manual
2001
Manual
-01
-24
Section Content
Issue
Page Cover page Table of Contents
1 About this Manual
2
3
Product Specification Performance Verification Procedure
4 Routine Maintenance 5 Electrical and Electronic description
6
7
Electronic Calibration Procedure Mechanical description
AppendixA Circuit & Layout Diagrams Appendix B Mechanical Drawings
C
Appendix Appendix
Information on oxygen fuel cells
D
Spare Parts Lists
3 3 3 3
3
3 3 3
3
3 3 3 3
1
2
3
4
Appendix E Warranty
3
E.M.E. (Electro Medical Equipment) Ltd. 60 Gladstone Place, Brighton, Tel.:+44 (0)
1273
6451 00, Fax.:+44 (0)
BN2 3QD,
1273
UK
6451 01
Email: info@eme-med.co.uk, Internet: www.eme-med.co.uk
O
E.M.E. (Electro Medical ~qui~ment) Limited. Copyright Notice. No part of this document may be photocopied or otherwise reproduced
of
without the prior permission in writing
2
E.M.E. (Electro Medical Equipment) Limited except as permitted by copyright law.
The
Infant
FlowTM System
1
-
About
To assist the engineer and technician this manual provides detailed descriptions of the major components of
N1672B and R11672P versions of the lnfant FlowTM Driver from serial number 2000 up. For Drivers with
the serial numbers repair or re-calibration of these products fully familiarise themselves with the clinical aspects and usage of the lnfant FlowTM System as this will assist in front line troubleshooting and service. This manual must be
used in conjunction with the Operating Manual document numbers 672-830 (English, Arabic, Dutch, French
and German), 672-831 (English, Danish, Finnish, Norwegian and Swedish), 672-832 (English, Greek,
Portuguese, Spanish and Italian) or 672-833 (English, Chinese, Japanese and Turkish). Routine maintenance, service or repair of the EME Infant FlowTM Driver as defined in this manual should only
be undertaken by competent qualified individuals trained in the calibration, service and repair of devices of
this nature and who have access to the special tools and equipment that are required. Alternatively contact
EME or an authorised distributorto ensure full reliability and safety of the product.
below2000, refer to document M672SM. We suggest that all personnel involved in the service,
this
Manual
Defective Medical Equipment) Ltd..
,
.
.
.
.
,
.
The manual comprises several discreet sections, which deal with the major sub-assemblies in order. This
allows work to be undertaken in a logical and methodical manner on one section of the product without
disturbing other parts. However, we recommend that the complete performance and verification procedure
be carried out after any work has been carried out on any part of the lnfant FlowTM Driver. E.M.E. (Electro Medical Equipment) Ltd. has a policy of continual improvement and update of all products.
Technical bulletins are sent to all registered users of the company's products to keep them informed of any
product changes. We welcome suggestions from users for product improvement and enhancement - please
write to the either the Marketing or Research
orworn parts must be replaced only with parts manufactured, sold or approved by E.M.E. (Electro
&
Development departments with your comments.
.....,....
.
........,........
.....
.....
.
..........,....
.;
....,..
.
........,
.....
....,
.
,.....
;
,....
.,.
,...,.,..........
...
...........................
.......,.
.......
.....
................
Service
Manual
3
2
-
Product Specification
Gas Supply Range Power Supply Power Supply
page 28 for options), Internal charge with fifteen (1 5) minutes remaining after the low battery indicator illuminates.
Note: Weight
Dimensions AirlOxygen Mixer Flowmeter Auxiliary Outlet
M672B - DlSS underside mounted or Mini-Schrader Quick Connect underside mounted, both with integral check valve and
M672P - DlSS side mounted or Quick Connect front panel mounted, both with integral check valve and 15 Ipm capability.
-
Nominal 4 bar, clean, dry medical air and oxygen.
-
Minimum 2 bar, maximum 6 bar. Maximum differential pressure 2 bar.
-
M672B - 200 to 265VAC, 0.1 OA, or 95 to 135VAC, 0.20A, 50 to 60Hz.
-
M672P - 12VDC, 1A from approved AC adaptor (see approved accessories list on
12V
LeadJAcid Battery- The battery has a life of four (4) hours from a full
It can take up to sixteen (1 6) hours to recharge a fully discharged battery.
-
M672B - 6.2Kg
M672P - 5.6Kg
-
21 2x1 70x1 40mm (excluding gas inlets, patient outlets and mounting brackets).
-
Range 21 to 100% oxygen, accuracy &3% of selected output.
-
Range 0 to 15 Ipm, accuracy
-
Mixed gas outlets, several types available dependant on model, all optional.
15
Ipm capability.
&I
0%
of selected output.
Pressure Relief
1 Patient safety
2 System/delivery circuit safety - mechanical internal relief valve preset at 205cm H20.
Manometer Oxygen Monitor Alarm System
alarms are set after 2 minutes operation without operator intervention although the operator can manually set or reset them if required.
1 Supply gases failure
limit of 2 bar or one gas fails completely an alarm will sound and the gas at the higher pressure only will be delivered to the patient.
2 High patient pressure
alarm automatically activates a pressure relieving solenoid which instantly reduces the pressure in the patient circuit to zero. The pressure is restored after should the cause of the alarm condition still exist. visual indication is set 3cm
3 Low patient pressure
measured CPAP pressure or at
4 Failure to deliver correct oxygen concentration -Audible and visual alarms are provided at
-
2 systems fitted
-
automatic electronic valve system preset to vent to ambient at 1 I cm H20.
-
Range, 0 to + 12cm H20, accuracy
-
Range 0 to 100% oxygen, accuracy i 2% of span.
-
Four separate alarm systems are provided all of which are automatic, the electronic
-
If the differential pressure between the two inlet gases falls outside of the
-
An audible and visual high pressure alarm is preset at Ilcm H20. This
H20 above the measured CPAP pressure.
-
An audible and visual low pressure alarm is set at 2cm H20 below the
0
cm H20 if this would otherwise be negative.
+I
cm H20.
3
seconds but will be reduced to zero
A
second high pressure alarm with audible and
&
5%
of
4
672-819
(3)
2001-01-24
The
Infant
FZolvTM
System
the measured Fi0,at the time of arming of the alarm with an upper maximum limit of 101% and a
18%
lower minimum limit of 20%. There is a low hazard warning at
EN
60601
-1
Classification M672B - Class M672P - Internally Powered Equipment, Type
This equipment is rated for Continuous Operation.
1,
Type
B,
IPX 0
B,
IPX
0
oxygen or below.
Storage Conditions
Safe for storage at Relative humidity 90% non-condensing.
Atmospheric pressure 0.6 to 1.4 Bar. Minimum Shipping Temperature - O°C Operating Environment - 10" to 40°C
Keep dry and do no expose to direct sunlight.
This product meets the EMC requirements of EN60601-1-2.
NOTE - Although the Infant FlowTM driver meets the requirements of current EMCIRFI legislation this
does not guarantee immunity from all sources of radiated energy. Some mobile telephones and other
products containing radio transmitting components may cause malfunction of the lnfant FlowTM driver
and should not be used in the vicinity of the device.
-All of the functional accessories supplied by EME for use with the lnfant FlowTM driver are for
NOTE
single patient use only. These accessories include the lnfant FlowTM generator, delivery breathing
circuits, humidification chambers, silencerlbacteria filters and fixation bonnets. Under no circumstances should sterilisation or re-use of these products be attempted. Please see the
disclaimer on page
-
O0
to 50°C.
22
of this manual.
.............................................................................................
............................................................................................
Service
Manual
3
-
Performance Verification Procedure
This performance verification procedure provides a means of determining whether the EME lnfant FlowTM Driver meets its design specifications. These tests are intended to be performed in the Hospital by qualified personnel. The procedure should be performed at least every four months or more frequently if desired or if
the device is suspect. The check list must be followed exactly and if the lnfant FlowTM Driver fails to meet any specifications it
should be removed from service until calibration
electrical safety tests be performed the
to tests with equipment such as a
NOTE: It is recommended that hospital personnel responsible for the Performance Verification Tests maintain records of their activities and identify equipment authorised for use.
WARNING: Oxygen vigorously accelerates combustion. To avoid explosion hazard do not use any instrument
or other equipment that may have been exposed to mineral oil or grease contamination.
N1672P is an internally powered device and should not be subjected
Rigel tester.
andlor service is accomplished. Please note that should
The performance verification procedure is divided into three tests, the first for the second for the pressure manometer and the third for the oxygen analyser.
The following special equipment is necessary to perform the tests fully, a manometer
accurate to f 0.2 cm H,O, adjustable pressure regulators, a calibrated oxygen analyser accurate to a1 % and a typical breathing circuit with an lnfant FlowTM Generator and prongs attached. Do not attempt to perform
the tests without suitable equipment.
First Test Observe the air inlet filter. If this is dirty or wet it should be replaced. Connect the lnfant FlowTM Driver to supply gases which are adjustable from
regulator sets to suit most pipeline systems. Please request ordering information from the customer services department.
Overall Accuracy Compare the oxygen analyser readings to mixer settings of 21, 40, 60, 80 and 100%. The mixer has a
specified accuracy of a 3%. If the oxygen analyser is accurate to within should agree within
Accuracy with varying inlet pressures
obtained with equal inlet pressures (4 Bar each). Vary the inlet pressures to the mixer to 3 Bar oxygen and 4
Bar air and then 4 Bar oxygen and
that the oxygen reading does not vary by more than 3% at each extreme. Alarm module test
reduce the air inlet supply to 2 Bar or just under. The alarm should sound and the oxygen analyser should read 100%. Increase the air supply pressure slowly, the alarm should cut out at around the Repeat the procedure varying the oxygen pressure. When the alarm sounds the oxygen analyser reading
should drop to 21%.
-
AirlOxygen Mixer
-
Set both supply gas pressures to 4 Bar. Select a flow of 8 Ipm on the Flowmeter.
&
4%
at the various settings.
-
Starting with the oxygen and air supplies at 4 Bar and 60% (the alarm should be silent)
0 to 4 Bar or more. EME supply
&
1 % the setting and the analyser
-
Select 60% and observe the reference oxygen analyser reading
3
Bar air. Note the oxygen analyser reading at each extreme and verify
airloxygen mixer, the
(e.g. Timeter RT200)
3
Bar level.
-
Second test
Set the mixer to 21 minute. If it is not adjust the lower potentiometer accessible from the left-hand side of the driver to obtain a reading of 21
Set the mixer to 100% and observe that the oxygen reading is stable at 100% after one minute. If it is not adjust the upper potentiometer accessible from the left-hand side of the driver to obtain a reading of 100%. If
adjustments are found to be necessary repeat this procedure two or three times as there may be some small
interaction between the controls.
Oxygen Analyser
%
and the flowrate to 8 Ipm. Observe that the oxygen reading is stable at 21% after one
%.
Once the end point calibration is satisfactory set the mixer to
57
and
between
63%.
60%
and check that the analyser reads
Third Test - Pressure Manometer Tee the test manometer with the hand pump bulb in line into the pressure input port on the front panel of
the lnfant FlowTM Driver. Check that both the bar graph manometer and the test manometer are at zero
10
then inflate the system to
10
indicates
cm H20. If it doesn't, referthe device for full calibration.
Reduce the pressure by
cm H20 and observe that the bar graph manometer display illuminates and
1
cm H,O at a time and ensure that the bar graph manometertracks the pressure.
Fourth Test -Alarm Functions
8
Attach the delivery circuit and lnfant Flow Generator to the lnfant Flow Driver. Set the flow to
5
occlude the prongs and check that the bar graph manometer shows
cm H20. Press the ArmIMute
Ipm,
button for 3 seconds to set the alarms and then occlude the exhaust tube of the lnfant Flow Generator to
11
create an over-pressure above
cm H20, at this point all of the LED'S in the bar graph illuminate, the
pressure vent valve operates immediately releasing the pressure to near zero and the alarm sounds. Remove the occlusion to the exhaust tube and press the
ArmIMute button for 3 seconds to reset the
alarms. Remove the occlusion to the prongs allowing the pressure to drop to zero. Observe that the low
pressure alarm activates immediately. Press the
the prongs again to restore the pressure to
Increase the flow to set the pressure to
ArmIMute button once, this will silence the alarm. Occlude
5
cm H20 and observe that the visual alarm cancels.
8
cm H20 and observe that the high alarm activates after
15
seconds. Reduce the flow to drop the pressure to 3 cm H20 and observe that the high alarm cancels and
15
that the low alarm activates after
seconds.
If any of the above alarm functions fail to perform refer the driver for full service and calibration.
Routine maintenance of the Infant FlowTM Driver is limited to regular checking of the oxygen analyser calibration
(4
and periodic status of the gas inlet filters, integrity of the alarm systems and cleaning of the exterior surfaces.
monthly) checking of the calibration of the airloxygen mixer and Electronic pressure manometer,
We recommend annual replacement of all filters, silencers, certain
kits containing all necessary parts are available for all models of the lnfant FlowTM Driver. See the spare parts
D
list Appendix
An lnfant FlowTM Driver in need of re-calibration, service or repair must not be used until the necessary
procedures are performed and the equipment has been tested to ascertain that it is functioning correctly.
The exterior air filter can be observed through the polycarbonate bowl. If it is discoloured or wet it should be
replaced.
the bowl.
CAUTION: The precision gas mixing device incorporated in this product may become non-functional or
damaged if used without the protective water trap and filters provided.
Oxygen Analyser Calibration The integral oxygen analyser is of the fuel cell type and as such requires a regular calibration check. To
perform this check set up the lnfant FlowTM Driver as for use and allow a minute forstabilisation of the device. Select an oxygen concentration of remove the small white plug adjacent to the
to give a reading of
remove the small white plug adjacent to the to give a reading of may be some small interaction between the set point controls if a gross adjustment is required, the process may need to be repeated two or three times. If this is the case it is indicative that the fuel cell is wearing out and should be replaced. Refer the lnfant FlowTM Driver to a competent service authority for replacement of
the fuel cell. Once calibration is completed please replace the white plugs.
on page
A
special tool (part number
35
for the appropriate kit to suit your driver.
21.
Set the mixerto
100.
Return the mixer to the
Mxxxxxx)
219'0,
loo%,
is supplied with the
wait one minute and verify that the display indicates
21
%
mark on the side if the device and adjust the potentiometer
wait one minute and verify that the display indicates
100%
mark on the side of the device and adjust the potentiometer
21
%
position and verify that the display reads
"0"
rings and the oxygen fuel cell. Service
M672P
driver to facilitate easy removal of
21.
100.
21
%.
If not,
If not,
There
Changing the Oxygen sensor The oxygen sensor must be changed annually
M672
lnfant FlowTM Driver the sensor is located in the mechanical section immediately behind the flowmeter.
Nlethods of access vary dependant on the model. In all cases ensure that the lnfant FlowTM Driver is disconnected from the power input supply.
M672B
6mm instrument head screws 8mm pan head screws on the rear cover and the remaining two
electronic module. Disconnect the grounding straps and slide the module forward to facilitate easy removal
of the oxygen sensor. Carefully remove the electrical connector from the old cell by using a small blade screwdriver to release the locking clip and unscrew the oxygen sensor from the flowmeter mounting block.
M672P
together with the forward to facilitate easy removal of the oxygen sensor. Carefully remove the electrical connector from the
old cell by using a
the flowmeter mounting block.
The new oxygen sensor must be carefully screwed into the block, ensure that the O-ring is fitted before screwing in. Fit the electrical connector into the new sensor observing the polarity. Re-assemble the lnfant
FlowTM Driver, apply stabilise before performing the only after calibration is satisfactory.
-To gain access remove the top cover of the enclosure. This is achieved by removing the four
(2
on each side), the top cover can then be lifted off. Remove the four
-
Remove the three
M3
srnall blade screwdriver to release the locking clip and unscrew the oxygen sensor from
t'he air and oxygen and switch on the power. Allow two hours for the oxygen sensor to
M4
screws on the bottom of the Driver, the
screw in the top left rear panel. The entire mechanical assembly may now be pulled
full two-point calibration. Return the Infant FlowTM Driver for use with patients
orwhen calibration is no longer possible. In all versions of the
M3
x
M3
x
M3
x 6mm instrurner~t head screws on the
M4
screw on the rear of the Driver
8
672-819
(3)
2001-01-24
The
Infant
FZOWTM
System
Dependant on storage and shipping conditions it is known that the output of galvanic oxygen fuel cells may
,
-
drift for the first few days of use and it is recommended that the calibration is checked daily for the first three days after a new sensor is fitted.
Cleaning
The exterior surfaces of the Infant FlowTM Driver can .be cleaned with a mild soap or liquid disinfectant solution. Do not use cleaning agents that contain abrasives.
Always disconnect the Infant FlowTM Driver from the power input supply before cleaning. CAUTION: Do not immerse any part of this device in water or gas or steam sterilise it. Damage will result.
Service
Manual
M672B
can be set to suit a range of input voltage conditions. Voltage selection is made by lifting a flap under the fuseholder and sliding the tray out. The voltage selector may then be withdrawn and refitted as required to suit the mains voltage available. Four settings are provided, these are 1
Power Supply -The M672B incorporates a multi-tapped transformer with a voltage selector which
OOV, 120V, 220V and 240V.
M672P
lead acid battery and integral battery charger. The power switch on the rear panel only switches the power to the electronics and
driver to be switched off whilst also allowing the battery to remain on charge. A fully charged battery will
enable the
NOTE:
within 20 minutes
external power not be applied the battery will continue to discharge even after the
ceased to operate.
The charger is capable of recovering a totally discharged battery this could take up to 16 hours depending
on the level of discharge. It should however be noted that once an external power source is connected the
unit will perform perfectly even though the full battery backup facility may not be available for an extended
time period.
Display display board are routed through the 40 way connector "Alarms Armed"
D5 are driven by driven by
button,
luminous intensity for the displays is controlled by a signal derived from LDRI. The two display drivers
require,different signals to change the brightness of the displays. To accomplish this two op-amp circuits are
provided on the processor board.
Power Supplv -The M672P is a battery operated instrument which incorporates a 12 volt sealed
not
the external power to the battery charger circuitry. This enables the lnfant FlowTM
lnfant FlowTM Driver to operate independently for at least 4 hours.
It is essential that the M672P be connected to an external power source whenever possible and
ofthe red low battery indicator illuminating. Should the power switch not be turned off 3 the
lnfant FlowTM Driver has
-
The display and processor board are mounted as a pair on the front panel. The signals to the
CONl. The pressure bar graph, Dl3 to D26 and the
LED'S Dl 8, are driven by IC3, a MM5450N LED driver. The alarm LED's, Dl 1, D27, D2 and
IC2, a UCN5812A. The seven segment oxygen concentration displays, LDI to LD3, are
ICI, an ICM7212AMIPL display driver. Also fitted on the display board is the "ARMIMUTE" push
SWI, and a NORP12 light dependant resistor, LDRI, for the auto-dimming of the displays. The
The
M672P Display board has three additional LED's fitted which are the battery status indicators. These are
connected to the power supply board with a 5 way lead.
-
Processor
CON1
programmable and, should it become defective, can only be replaced with a pre-programmed device obtained
from EME or an authorised distributor. The micro-controller contains the following elements:
The Microprocessor qlock signal is generated by XTLI and C2, a 4 MHz external crystal and capacitor.
There is no provision for adjusting the frequency of the oscillation.
The processor board fits behind the display board and is connected to it by a 40 way connector,
.
The instrument is based around a Motorola 68705 type micro-controller ICI. This device is one time
6805 Micro Computer
bi-directional I10 Lines
24
8
Input Only Lines 4 Channel Analogue to Digital Converter 3776 Byte Programmable Read Only Memory 112 Byte Random Access Memory 8 Bit Timer
Pre-Scaler
7 Bit Memory Mapped I10
The pressure sensor PTI, is a Sensym solid state device having a range of 0 to 1 PSI which is fitted directly
to the processor board. It can withstand up to 20 PSI without sustaining damage. This item is not a standard
component as it is specially selected to our specification for low noise and zero temperature drift. It must only be replaced with a device to the same specification obtained from EME or an authorised distributor.
The reference section of the
produce 12 V to energise the pressure sensor. The pressure sensor with potentiometer VR3 being used to adjust the stage gain. Stage 1 of used in conjunction with VR7 to swing the output of the amplified output from the pressure sensor circuit is applied to the analogue input
In parallel with this signal is
ensures that the inputs to the micro-controller cannot go outside the limits of the ADC reference. On the
mixer monitor PCB is mounted on the front panel of the mixer unit. The potentiometer is used to monitor the oxygen control shaft position. It is connected between OV and depending upon where the mixer shaft is set. Stage 1 of and
100%
the oxygen alarm limits, IC4 is again used to clamp the signal to the micro-controller. On the
alarms are set on the measured oxygen value. Input AN2 on provides an offset to trim out any zero error from the oxygen sensor, whilst stage 3 provides the necessary
amplification. VR8 adjusts the offset and VR4 adjusts the gain. IC4 is used to clamp the signal. Input AN4 on
IC1 is spare and is connected to ground to prevent noise pick-up from getting into the measurement circuitry.
On the The display brightness levels are generated from the
resistive circuit, VR2 and
this. The brightness of the 7 segment displays is set by the difference between this level and
brightness of the bar graph
the voltage at its wiper.
M672B only input AN1 on IC1 is used for the signal conditioned mixer monitor potentiometer. The
IC8 stage 2, provides an offset to ensure that the output achieves a level of 0.05V at 21
0,.
VR6 adjusts the offset and VR5 sets the span. This section of the electronics is used for determining
M672P input AN1 is also spare and is tied to ground in a similar fashion.
LM10, IC7, produces 200 mV which is amplified by the op-amp section of IC7 to
AMP02, IC6, provides the signal conditioning for the
IC5, a LM324, is
APM02 so that any offset can be trimmed out. The
AN0 of the micro-controller.
IC4, a Thomson L9700 or Texas Instruments TL7726 precision clamp. This
5V and feeds an output of a proportion of this
IC8, an LM324 quad op-amp, amplifies the signal
%
0,
and 5V at
M672P the
IC1 is used forthe oxygen monitor. IC8 stage 4
NORP12 signal fed back from the Display Board. A
R11, is used to set the overall brightness level and the fed-back signal modulates
15V. The
LED'S is adjusted by using trimmerVR1 to set a constant current dependant upon
IC5 stage 3 and transistor TR2 form a voltage controlled constant current source.
The audible alarm is fitted on the Processor Board and is driven by the micro-controller via MOSFET The audible alarm is powered from the +I 5V rail and is a high efficiency type, selected for its high output at
low voltages.
An over-pressure relief valve is fitted in the Infant FlowTM Driver to vent any excessive pressures to atmosphere.
This slave over-pressure valve is driven by the micro-controller using a solenoid valve as a 'master. The
a
power source for the solenoid valve is a three terminal regulator, REGI, connected as source. The solenoid valve is operated by a MOSFET, TR3, connected in series with opto-coupler, rnonitor the function of the solenoid valve circuit.
Also fitted on the Processor Board is a two-stage watchdog circuit. The first stage,
SG3548N which has four sense inputs, an inverter and a threshold input. The inverter is connected so as to allow monitoring of the minus 15V power rail by Sense 4 input. The 5V and 15V power rails are monitored by Sense 2 and Sense
the set levels. The tolerance on the over and under voltages is set by using the threshold input. The resistors,
R6 and R8, set the window width to plus and minus 5%. The line input, which is normally used to monitor an
AC input, is used here to gate the watchdog reset pulses from the micro-controller with the power supply
monitorsignal to produce outputs at pins 7, 9 and 10. Whilst all is well, this pulsating signal is applied to IC3 pin 6 to prevent timing out of the watchdog. IC3 is a Maxim AD694AN microprocessor supervisory circuit, of which only the 1.6 second timer is used. If any power rail goes above its permitted value, the value, the
the micro-controller will reset after 1.6 seconds delay. C12 provides a short delay on operation of the sense
inputs to avoid operation on transients.
UN
3
inputs respectively. The Sense inputs are checked for voltages both over and under
MAX694CPA, Supertex MP694P or an AMD
ON
output at pin 9 will go low; if any power rail goes below its permitted
output at pin 10 will go low. In either case, the watchdog reset pulses will be interrupted and
IC2, is a Silicon General
constant current
TRI.
IC9, to
Service Manual
11
M672B
and
M672P
Electronics Trouble Shooting
Guide
Problem or fault
M672B - The electronic module does not function at all
M672B - Electronic module continually resetting
M672P - The electronic module does not function
M672P - Electronic module continually resetting
att all
Possible cause
No mains power
Power supply fault
+
5 VDC supply faulty
+
1 5VDC supply faulty
-
15 VDC supply faulty
Battery discharged
Power supply fault
+
5 VDC supply faulty
+
1 5VDC supply faulty
Corrective action
Power cord connected?
Mains switch turned on? Voltage selector set correctly3 Check primary fuses Check plug fuse (if fitted)
Power connector plugged in?
PSU
Check BRllBR2 bridge rectifiers and voltage regulators
External power source connected?
External power source switch turned on?
Check primary fuses Check PSU fusible link and
DC to DC converters
secondary fuses,
1
-
15 VDC supply faulty
All displays dim
Audio alarm works but visual does not
Visual alarm works but buzzer does not
Manometer display off Parts of manometer display Faulty display driver Check
missing
Auto dimming circuits faulty or not set correctly
Alarm
LED'S defective
Buzzer faulty
Buzzer drive circuit fauty Faulty display driver
Faulty
LED'S Check LED'S
Check
IC 5
Check alarm
Check buzzer
Check TR3
VR1 , VR2, LDRI and
LED'S
(ZVN4306A)
IC2 (MM5450N)
12
The
Infant
FlowTM
System
Loading...
+ 26 hidden pages