Elster gas-lab Q1, gas-net Q1 Technical Documentation Manual

Gas quality analyser Q1
Technical documentation
commissioning and maintenance
FCQ1-BAShort-EN a 19.3.2013
2013 Elster GmbH Edition a 19.3.2013
Copyright
2013 Elster GmbH
GAS-WORKS, Z1, gas-lab Q1, and FLOW COMP are German registered trademarks of Elster.
Microsoft, Windows and Windows NT are registered trademarks of Microsoft Corporation.
HART is a registered trademark of HART Communication Foundation.
Elster GmbH
Schlossstraße 95a
D - 44357 Dortmund, Germany
Tel.: +49 231 937110-0 Fax: +49 231 937110-99
E-mail: info@elster-instromet.com
Page ii gas-lab Q1
Contents
Safety and warning notes ................................................................................. v
1 Introduction ............................................................................................... 1
1.1 The gas-net system idea ...................................................................... 1
1.2 The measuring principle ....................................................................... 1
2 Device view and design ............................................................................ 5
2.1 Sensor system ..................................................................................... 5
2.2 Evaluation computer ............................................................................ 7
3 Operating gas-net devices ..................................................................... 11
3.1 The keypad ........................................................................................ 11
3.2 Other operating elements: status LED, calibration switch .................. 12
3.3 Display ............................................................................................... 14
3.4 Displays / Menus / Dialogs ................................................................. 14
4 Primer for impatient operators: What do I have to do to …? ............. 25
4.1 ... view the gas quality error listing? ................................................... 25
4.2 … accept the gas quality measurement errors? ................................. 26
4.3 … check all parameter settings? ........................................................ 27
4.4 … check the input values? ................................................................. 28
4.5 … view and check the outputs? ......................................................... 28
4.6 … view the archives? ......................................................................... 28
5 Functional description ............................................................................ 29
5.1 Gas quality module ............................................................................ 29
5.2 Data logging module .......................................................................... 53
5.3 Monitoring module ............................................................................. 59
5.4 System module ................................................................ .................. 78
5.5 Integrated RDT module ...................................................................... 87
5.6 DSfG module, Data exchange module ............................................... 89
6 GAS-WORKS / GW-GNET+ ..................................................................... 97
6.1 Compiling and exporting a parameterisation: Brief description .......... 98
6.2 Importing and editing a parameterisation: Brief description ............. 101
6.3 Extras: GW-GNET+ service programs ............................................. 103
6.4 Optional: GW-Remote+ for downloading archives ........................... 105
gas-lab Q1 Page iii
7 Installation ............................................................................................. 107
7.1 Mounting the gas-net Q1 ................................................................. 107
7.2 Connecting the lines ........................................................................ 107
7.3 Mounting the sensor system ............................................................ 117
8 Commissioning ..................................................................................... 121
8.1 Parameter protection against unauthorised access ......................... 121
8.2 Parameterisation ............................................................................. 122
8.3 Sealing of the device (if applicable/when required) ......................... 122
8.4 Commissioning of Integrated RDT module ..................................... 122
8.5 Commissioning the sensor system .................................................. 123
9 Maintenance .......................................................................................... 125
9.1 Maintaining the gas-net Q1 evaluation computer ............................ 125
9.2 Maintaining the gas-lab Q1 sensor system ...................................... 128
10 Technical data: Q1 ................................................................................ 129
10.1 Device type ...................................................................................... 129
10.2 Sensor technology details (gas-lab Q1) ........................................... 129
10.3 Evaluation computer details (gas-net Q1)........................................ 131
11 Annex ..................................................................................................... 135
11.1 Error listing of the gas-lab Q1 .......................................................... 135
11.2 Menue structure of the Q1 ............................................................... 145
12 Index ...................................................................................................... 147
Page iv gas-lab Q1
Safety and warning notes
To ensure a safe and faultless operation of the device the instructions and notes contained in the documentation on hand must be observed. The device must be used as prescribed and connected according to the connection diagram. The national and local regulations for electrical installations must be observed. The mounting, electrical installation, commissioning and maintenance of
the entire measuring system must be performed by qualified staff who have been trained in the field of explosion protection and have read and understood all parts of the operating instructions necessary for the actions to be carried out. Please refer to the corresponding technical regulations (DIN, DIN EN, VDE, VDI, and DVGW) for general information on mounting, commissioning, taking out of service and maintaining. Below, we have listed the standards and guidelines that must be observed:
DIN EN 60079-14, Publication date: 1998-08, Electrical apparatus for
explosive gas atmospheres – Part 14: Electrical installations in hazardous areas (other than mines)
DIN EN 50110-1, Publication date: 1997-10, Operation of electrical
installations
DIN EN 60079-17, Publication date: 1999-08, Electrical apparatus for
explosive gas atmospheres – Part 17: Inspection and maintenance of electrical installations in hazardous areas (other than mines)
DIN VDE 0100-610, Publication date: 1994-04, Erection of power
installations with nominal voltages up to 1000 V; initial verification
gas-lab Q1 Page v
gas-lab Q1 sensor system (sensor technology):
The sensor system must neither be stored at temperatures below
-20°C nor above +55°C.
Connect the electrical cables only to the EEx-e approved junction box!
This junction box is located on the mounting plate of the gas-lab Q1.
The sensor system is supplied by 24 V DC and must be secured
externally by 2.5 A.
Include the housing and mounting plate in the local earthing system. A temperature between +5 deg. C and +40 deg. C must be guaranteed
during the operation of the sensor system (-x to +40 deg. C with additional heater, -x to +55 deg. C with cooling unit; the lower temperature limit depends on the heat output).
Only gases of the second gas family according to DIN DVGW 460 or
approved calibration gases are permissible. The oxygen content in the natural gas must not be higher than 2.0 percent by volume.
The ventilation must be connected to the exhaust gas manifold via a
stainless steel tube with a minimum inside diameter of 4 mm.
After a voltage failure the measuring system automatically purges
with process gas before it activates the sensors. After a new gas cylinder with flammable contents has been connected, the system must be purged manually before the normal measurement can be continued. This also applies if air may have got into a gas pipeline connected to the sensor system (see also Chapter 5.1.4).
The user must ensure by means of high-pressure reduction and safety
devices that all gases injected into the measuring system do not exceed an inlet pressure of 1,250 mbars absolute!
The opening of the housing as well as inspection and maintenance
must be carried out by specialists authorised by Elster!
Page vi gas-lab Q1
Only if the atmosphere is not explosive or in de-energised conditions
after a delay of 5 minutes after the electrical supply has been disconnected, the cover of the housing may be opened!
The cable entries for the electrical connections of the EEx d housing
of the measuring system must never be loosened if the atmosphere is explosive!
The loosening or unscrewing of the breather is prohibited! Only authorised personnel is allowed to exchange the breather, if
need be. In this case, the breather as a whole must be exchanged!
Only authorised personnel are allowed to exchange a cable entry
gland, if necessary.
The sensor system may only be switched on before or together with
the gas-net Q1 evaluation computer. It will result in an error (alarm) if you switch on the sensor system after having switched on the evaluation computer.
gas-lab Q1 Page vii
gas-net Q1 evaluation computer:
The gas-net Q1 evaluation computer must neither be stored at
temperatures below -20°C nor above +55°C.
A temperature between 0 deg. C and +40 deg. C must be guaranteed
during operation.
The gas-net Q1 must be installed outside ex-zone 2. The gas-net Q1 evaluation computer may contain subassemblies that
are approved as associated electrical apparatus of the ib category according to DIN EN 50020 with intrinsically safe circuits. This renders the electronic evaluation computer Q1 suitable for connection to sensors and pulse generators located in hazardous areas (e.g. zone 1). A mixed connection of intrinsically safe and not intrinsically safe circuits is not permitted in case of these subassemblies.
The evaluation computer is supplied by 24 V DC and must be secured
externally by 1 A.
The earthing is connected to PE of the power supply connector for
equipotential bonding.
Observe the regulations of the relevant standards, in particular of DIN
EN 50014, DIN EN 50020, and DIN EN 50029.
Observe the limit values stated in the respective certificates of
conformity of the boards to be connected.
Warning: The evaluation computer of the Q1 measuring system is a class A device that may cause interferences in living areas; in this case, the user may be asked for appropriate measures at his expense.
Page viii gas-lab Q1
Introduction 1
1 Introduction
1.1 The gas-net system idea
is the generic term for an Elster device family. The evaluation
computer of the gas-lab Q1, too, is based on the gas-net device family. All gas-net devices, including future device types, are characterised by uniformity in appearance, operation and parameterisation. Each Elster device always covers a multitude of measurement and control
functionalities. gas-net devices also provide this functional variety. To keep the operation and parameterisation of the devices well structured and user-friendly, the gas-net series is based on a modular concept. A module corresponds to a special functionality. Each module has its own main display within the device’s menu assistance, and each module has its own group of settings within the parameter data record. A particular module can be employed in different device types. This yields a modular system that is advantageous to the user as a particular module can always be operated and adjusted in the same way, no matter in which device type it has been installed.
1.2 The measuring principle
The gas-lab Q1 measuring system is a gas quality analyser that performs infrared absorption and thermal conductivity measurements. The primary target variables are the gross calorific value, standard density and CO2 content of the natural gas being measured. The entire measuring system consists of two components:
1. gas-lab Q1 sensor system
The actual sensor technology is located in an explosion-proof housing, which can be installed in a hazardous environment.
2. gas-net Q1 evaluation computer
The evaluation computer of the Q1 measuring system is located in a non­hazardous environment. The main tasks of the evaluation computer are controlling and monitoring the measuring process, evaluating the sensor measurements, calculating the target variables and supporting the user during calibration. Furthermore, the evaluation computer contains an
gas-lab Q1 Page 1
1 Introduction
gas-lab Q1 sensor system
Vent gas line
RS422,
4-wire
Hazardous
area
Non-hazardous
area
Gas supply
24 VDC
Sensor unit
Sensor
elektronics
Double-block&bleed
valve set
RS422 / optical fibre
converter
gas-net
Q1
evaluation computer
optical fibre
EEX-e
Communi-
cations
PTB
integrated data logging function, mainly for interval- and event-controlled data logging of analysis data as well as of error listing and logbook. The signal and message processing provides some digital and analogue output possibilities. The evaluation computer also controls the measuring process and calculates the target variables.
Page 2 gas-lab Q1
Introduction 1
The connection between sensor system and evaluation computer is implemented via interface cable, RS422/optical fibre-converter and fibre-optics cable, as illustrated above. The sensor system contains two infrared sensors for measuring the absorption of the hydrocarbons and carbon dioxide contained in the natural gas. Another sensor additionally determines the thermal conductivity of the natural gas and thus also measures gas components such as nitrogen, for instance, which cannot absorb infrared light. Afterwards the evaluation computer evaluates all three measurements. As a result, these measurements supply the gross calorific value, standard density and CO2 content of the natural gas. These variables are sufficient to establish with a flow computer the compressibility ratio k (real gas dependence) according to SGERG and the energy content of the natural gas. Besides, other variables such as the Wobbe and methane numbers are also determined. Other than that mentioned, the system provides also a non­calibratable sample analysis of the natural gas. The underlying algorithm is based on the systematic of the composition of natural gases. In case of non­typical gases a major deviation in the measured value of a single component can occur. The device measures continually and determines new measurements every second. It can therefore also be used for fast closed-loop control tasks. The manufacturer calibrates the gas-lab Q1 before delivery, i.e. he performs a zero point adjustment with nitrogen and afterwards a 3-point calibration with ultra-pure methane and two calibration gases. The correction values are saved in the sensor unit. This basic calibration is usually repeated during the commissioning and on the occasion of a recalibration. The evaluation computer automatically performs a 1-point calibration with ultra­pure methane after a mains failure and after each switching on of the measuring sensor technology.
Moreover, the gas-net Q1 offers a wealth of additional functions for monitoring tasks and data communications. The gas-net Q1 also always includes a data logging function that logs important measurements at defined intervals and when errors occur. See Chapter 5.2 for a detailed description of the archive structure.
gas-lab Q1 Page 3
View and design 2
breather
EEx-d junction
(voltage supply,
data connection)
3: Calibration gas connection (i.e. N2, H2-11K, L1-8K) 2: Calibration gas connection (methane) 1: Process gas connection
Vent gas line
2 Device view and design
2.1 Sensor system
The actual sensor technology of the gas-lab Q1, the sensor system, is mounted in an explosion-proof housing. The gas is supplied at approximately 80 mbars overpressure and via a double block & bleed solenoid valve set, which is also located inside the housing. There are a total of three gas inlets for different gases, one connection for the process gas and two connections for calibration gases. The out flowing gas is led via a vent gas line. The housing is mounted on a mounting plate in such a way that the gas and other process connections point downwards. In order to ensure that the housing withstands the maximum permissible inside pressure of 1,100 mbars, a breather is led out of the housing at the top. The following illustration shows the front of the sensor housing:
gas-lab Q1 Page 5
2 View and design
gas-lab Q1
exhaust gas
connection of
safety relief valve
power supply
connection
exhaust gas
connection of
gas-lab Q1
pressure controller
with integrated
safety relief and
safety shut-off valve
data line
connection
pressure gauge of
inlet pressure for gas-lab Q1
calibration gas 1
connection
process gas
connection
calibration gas 2
connection
release button of
safety shut-off valve
ex-junction box
The parameterisation of the evaluation computer defines which gas has to be connected to which inlet. The process gas is assigned to path 1 and the internal calibration gas (ultra-pure methane) to path 2. Nitrogen and further calibration gases are injected via path 3 for performing the basic calibration. The test gas can either be injected via path 1 or 3.
Not only the sensor housing is installed on the mounting plate, but also regulators for the inlet pressure of the different gases to be injected, safety shut­off and safety relief valves as well as the EEx-e junction box for the interconnecting cables (voltage supply, data link to the evaluation computer).
Page 6 gas-lab Q1
View and design 2
0987
,
6
5
4
_
3
2
1
Q1
gas-lab
Status
Cal.-
switch
close open
DSS
Gas Quality Meter
Data
interface
Key pad
Calibration
switch
Status LED
Display
The pressure regulators with integrated safety shut-off/pressure relief valves for the process and calibration gases are necessary, as the maximum absolute inlet pressure of the gases must never exceed 1,250 mbars. The pressure regulators have been set by the manufacturer to a control pressure of 80 mbars (gauge). The integrated pressure relief valve starts to vent at an overpressure of about 110 mbars and above. The safety shut-off valve has been set to a tripping pressure of 142 mbars (gauge) by the manufacturer.
2.2 Evaluation computer
The housing of the gas-net Q1 evaluation computer is designed as plug-in unit for a 19”-frame and is available in two housing sizes, i.e. with a mounting width of 1/3 for up to three process boards or a mounting width of 1/2 for up to six process boards. The device front includes one 8x32 characters-LCD, one keypad with 16 keys, one status LED and the calibration switch. The DSS data interface is also located on the device front. It serves the connection of a PC or laptop for servicing purposes. The following illustration shows, as an example, the front view of a gas-net Q1 in the narrow design with a mounting width of 1/3:
gas-lab Q1 Page 7
2 View and design
Most of the connection possibilities are located on the back of the device. The following interfaces are always available:
fibre-optics connection for the sensor system DSfG bus connection COM2 interface: serial interface according to RS232. In case of devices
with an integrated RDT, the modem is connected to COM2. A different software variant offers gateway functionality for interfacing a host protocol
instead. In this case, the COM2 interface can be used as protocol channel. DCF77 interface for connecting a radio clock 24 V DC power supply connection HSB bus connection (not used)
All process connections are implemented via process boards installed in the housing. The exact composition of the I/O boards depends on the tasks of each individual device (number of required output signals, etc.).
The name of the software variant, the version’s number and the checksum for
the identification of the software version can be invoked directly at the device via the main display of the System module.
The main functionality of the Q1 is measuring the gas quality. The sensor system is connected via an LMFA1-type process board, which provides 3 digital and 4 analogue outputs in addition to the fibre-optics connection for the sensor system.
Furthermore, the following boards can be used:
A multi-functional EXMFE4 input board for the connection of a pressure and
a temperature sensor (PT100) and two digital inputs (NAMUR). All channels
are intrinsically safe (EX-i). An MSER2 board with 2 serial channels (RS232, RS422 or RS485) for
interfacing communication protocols (e.g. MODBUS). An MFE11 input board with 8 digital and 3 analogue inputs.
An MFA6 output board with 4 digital and 2 analogue outputs.
Page 8 gas-lab Q1
View and design 2
0,63 ATT
LE
LA
D1 D2 D3
I1+ I3+
I­SH
DC
LMFA7
SH
GD
DT
VN
VP
PE
DCF7724V
HS-Bus
2
M
O
C
G
f
S
D
CD
RI
TX
RX
SD
DD
SH
DH
internal
Board 3Board 2Board 1
Output board LMFA7 (digital/analog) LA/LE:
D1-D3:Three digital outputs for digital signals / pulses. I1-I3: Three analog outputs for measurement output.
optical fibre connection for external I/O expansion (sensor system connection)
Any input or output board can be assigned to board locations 1 or 2.
gas-net
All in all, up to 6 process boards can be mounted in the broader housing and up to 3 boards in the narrow design. Please see Section Fehler! Verweisquelle konnte nicht gefunden werden. or the Technical Data Section in Chapter 10 for a description of the currently available boards.
The following illustration shows, as an example, a device in the narrow design with the always existing LMFA7 board.
gas-lab Q1 Page 9
gas-net Q1 rear view (example)
Operation 3
3 Operating gas-net devices
This chapter’s objective is to give you an understanding of the basic operating
and menu structures of gas-net devices. As already mentioned, all devices of the gas-net family have a uniform appearance and a comparable menu structure. This means for the user: If you have operated a gas-net device once, you will also be able to operate all other device types without any problems. According to our philosophy of how to parameterise gas-net devices, they are adjusted by means of a PC or laptop and not via the operator panel. The device operation via the operator panel mainly serves the indication of the most important information on the display. The content of the operator interface on the display depends on the individual gas-net device type.
3.1 The keypad
The keypad of gas-net devices consists of a numeric keypad for the entry of numbers, minus sign and decimal point keys and a group of four navigation keys. With these keys, you may move within the menu structure and invoke menus and displays. In some cases you may also trigger actions or change values via the navigation keys. The illustration below shows an overview of the keys’ meanings. The exact context-related meaning of each navigation key will be explained in connection with the menu structure in Section 3.4.
gas-lab Q1 Page 11
3 Operation
1 2 3
4 5 6
,
7 8 9 0
Menu key:
Opens/closes a menu. When starting from a display this means:
Pressing once opens the current display's submenu listing. Pressing twice opens the menu listing for branching to other modules.
Pressing three times closes the menu.
Arrow key left:
Previous entry. Input mode: delete previous character
Enter.
Menu selection. Enter input mode. Within input mode: Accept new value.
Numeric key pad
incl. minus sign and decimal point
Arrow key right:
Next entry. Input mode: Quit input mode without
changing values.
LED status
Meaning
red, blinking
An alarm is pending, i.e. an error has occurred that influences the gas quality measurement.
yellow, blinking
A warning is pending. That is, an event has occurred without affecting the gas quality measurement.
green, blinking
A green blinking light appears in the start-up phase after a mains failure.
red, steady light
An alarm has been pending but is no longer relevant. It can be removed from the error listing by accepting it.
yellow, steady light
A warning has been pending but is no longer relevant. It can be removed from the error listing by accepting it.
green, steady light
The device runs error-free.
3.2 Other operating elements: status LED, calibration switch
The status LED on the front of the device is a three-colour light emitting diode. The status of this LED indicates whether an error of the gas quality measurement is pending or has been pending. Please refer to the table below for the meaning of the individual colours:
Page 12 gas-lab Q1
Operation 3
The sequence of the LED status in the above table corresponds to the sequence the error management keeps to: The system always indicates the error with the highest priority. A pending error always takes precedence over an error that is no longer relevant. The exact meaning of the terms alarm, warning and hint is explained in Chapter
5.3.1. The calibration switch is on the lower right side of the front panel.
All gas-net devices are furnished with a two-level safety concept: All parameters being protected by the calibration switch can only be changed if the calibration switch is open. Such parameters are always modified with a PC or laptop and the associated parameterisation software GW-GNET+. Open the calibration switch by turning it anticlockwise as far as it will go. This first level of the safety concept is important for devices used for legal metrology and custody transfer. In this case, a seal may officially secure the calibration switch. The basic display of the device will automatically be invoked when you close the calibration switch.
Note: The User lock as the safety concept’s second level consists of one numerical lock for each of the two contract parties. The user lock is, in contrast to the calibration switch, implemented via the device software. This means that the locks are defined via the device parameterisation and opened or closed via the operator panel. Open locks allow the user to access certain parameters or actions. All parameters being subject to the user locks can be changed when both locks are or the calibration switch is open. .
gas-lab Q1 Page 13
3 Operation
3.3 Display
The display is an illuminated LCD consisting of 8 lines with 32 characters each.
After approximately 30 minutes without a keystroke, the display’s background
illumination switches off automatically.
3.4 Displays / Menus / Dialogs
One note at the outset: The following section describes the menu assistance and operation of all gas- net devices in general. Where appropriate, individual subjects have been illustrated with examples. These examples refer to currently available device types. Therefore, it may happen that a special menu illustrated in an example does not exist in your gas- net device type. In accordance with the gas-net concept, however, the operating mechanisms generally described here function in all devices in the same way.
Each module has a main display in which all important current values are indicated.
For example: Among others, the gas-net Q1 contains the gas-lab Q1 and Monitoring modules. The main display of the gas-lab Q1 module shows the current measurements, whereas the main display of the Monitoring module indicates the error listing.
The main display of the first module is also the basic display of the device, i.e. the display that is invoked automatically if there has been no keystroke for about 30 minutes.
For example: The basic display of the gas-net Q1 is the main display of the gas-lab Q1 module.
A display serves to present values. If there are more entries than can be made visible at once, little scroll arrows on the right side indicate whether or not you may scroll upwards or downwards.
Page 14 gas-lab Q1
Operation 3
1
Down scroll arrow pointing downwards: Scroll down­wards with the rightward arrow key.
„Up“ scroll arrow pointing
upwards: Scroll upwards with the leftward arrow key.
Each display which can be shown belongs to a module within the device software, therefore to a closed functionality. If you see the display of any module, there are two entirely different targets in the menu structure of the device. On the one hand a subordinated display / dialogue1 of the shown module and on the other hand the main display of any module.
To make the navigation within the menu structure as easy and fast as possible, the menu key is configured as follows: Unique pressing of the menu key opens up the list of submenues, wich is provided by the actual module display. Pressing the menu key again opens up the menu list to branch to any module. Repeated pressing closes the menu.
The list of submenus of a module is provided according to the current parameterisation: Menu items that relate to functionalities which are not parameterised, are not offered at all.
Submenus of a module are either calling other displays or dialogs, in which the user can manipulate values via control panel. The menu structure is aborescent:
A subordinated menu item of a module can offer subordinated menu items by himself.
A Dialog is a display window indicating values that can be changed by the operator.
gas-lab Q1 Page 15
3 Operation
Start: Basic display Gas quality
Menu list of the module Gas
quality.
The hyphen in front of the menu descriptions indicates that the list refers to subordinated menus.
In lower levels of the menu structure following contextual menu items are offered to return to the next upper level:
Menu item Back in a display Menu items OK / Cancel in a dialog
(OK means acceptance of the changed values too, Cancel means to reject the changes)
Independent of the menu level which is shown at the moment, with the arrow keys you can move for- and backward within every menu list and select a menu item. The selected target is presented in an inverted way, i.e. with green writing on a black background. Activate the menu item belonging to the selected entry by pressing the Enter key.
For example: We assume you want to change from the basic display of the Q1 to the main display of the Monitoring module. For this, press the menu key first to open the menu window. The first entry of the appearing listing is selected, i.e. it is backlit in black:
Page 16 gas-lab Q1
Operation 3
Module list: Without hyphens in front of the menu descriptions
Module list: Monitoring module is selected
Since we won’t activate a gas quality menu in our example, but branch to another module, please press the menu key again.
Now the menu shows a list of all modules the device software contains. Press the rightward arrow key several times until the module is selected, that you want to see. In our example it’s the Monitoring module.
Then press the Enter key and the display of the just selected module will be invoked.
gas-lab Q1 Page 17
3 Operation
Tips:
If you have pressed the rightward arrow key too often and went too far
down in the menu selection list, move upwards again by pressing the leftward arrow key.
If you want to quit an invoked menu selection window without having made
a selection, just press the Menu key as often, until the menu window is closed.
The selection of some menu items invokes a dialog. These dialogs are displays in which values can be modified. However, only a few values can be modified via the operator panel. In such input dialogs you move from one parameter to another by using the arrow keys. If a parameter must not be changed (for instance, because it is a parameter that is subject to the calibration lock, which is closed at that moment), it will be crossed out in the display. If you have selected a parameter that can be changed, you may switch into the edit mode via the Enter key. In order to render the operation more comfortable, there are different methods of defining a new value, depending on the type of the value to be changed:
Direct entry of a new numerical value
If you would like to replace individual characters only, delete the characters step by step from the right using the leftward arrow key. Then enter the new characters via the numerical keys including decimal point and minus sign. If it is easier to replace the entire value by a new one, just start with your entry right away: As soon as you press any numerical key the preset value will be deleted and overwritten by the new entry. Quit the edit mode via the Enter key. This initiates a test for consistency: If you have entered a value that does not make sense in the present context or is not permissible, you will not be able to quit the edit mode. This way, the user is forced to correct the value he has entered.
Page 18 gas-lab Q1
Operation 3
Start: Display Single message
Value to be modified is selected.
Edit mode: The insertion mark is now blinking behind the value.
In order to quit the edit mode without accepting the change, for instance after an erroneous entry, just press the rightward arrow key. In order to quit the entire dialog, invoke the menu and select either OK (the new values will be accepted) or Cancel (the values will be rejected).
For example: Change of the lower hint limit for the GCV in the Monitoring module of the gas-net Q1. After you have invoked the associated menu item, the following display is visible:
The currently set value is indicated. Switch to the edit mode via the Enter key.
gas-lab Q1 Page 19
3 Operation
1
,2,3,4,5,6,
7,8
,9,0,,,
Options on leaving Acceptance or rejection of the changed values.
Enter a new value now: Either directly via the numerical keypad or by deleting individual numbers from the right using the leftward arrow key and by entering new numbers.
Quit the edit mode by pressing the Enter key. Thereby the new value will be accepted. If you don’t want that, you should leave the edit mode via the rightward arrow key.
Invoke the menu now.
If you confirm OK by pressing the Enter key, the new value will be accepted. To reject the modification, go to Cancel by pressing the rightward arrow key and afterwards the Enter key. You will quit the dialog in both cases.
Page 20 gas-lab Q1
Operation 3
Start:
Monitoring - Switches
Selective list: All applicable values are offered. (Here: off and on)
New value by selection from a list
The device software offers a list of possible values in case of editable values the range of which is restricted to a fixed number of selectable values. Choose a suitable value from the list via the arrow keys and accept it by pressing the Enter key.
Example: Opening of the revision switch (Hint: Only possible with user locks opened!). The state of the revision switches can be changed in the submenu Switches of the Monitoring module. Therefore go to the related dialog via the menu items Monitoring - Switches.
In the above illustration the revision switch for both streams is shown as closed (Revision 1 = off, Revision 2 = off). Let’s assume that you want to activate the revision switch for the first stream. On entering the display this switch is already activated, so you can press the Enter key right now to switch to the edit mode.
gas-lab Q1 Page 21
3 Operation
Selective list: On is selected.
Revision switch 1 is opened now.
Options on leaving Acceptance or rejection of the changed values.
Select the desired value via the arrow keys, in our example select on. The display looks as follows:
Then press the Enter key to leave the edit mode.
Invoke the menu now:
If you confirm OK the new value will be accepted. To reject the modification, go to Cancel and press the Enter key afterwards.
Page 22 gas-lab Q1
Operation 3
Modifying several values at once
Most of the dialogs don’t offer individual values but whole sets of values for
being modified. In such a case, edit the first selected value first. Switch to the edit mode by pressing the Enter key. Move to the next value by pressing the Enter key.
Tip: If you don’t want to modify an offered value, skip it by pressing the
rightward arrow key.
Change the value either by directly entering the new value via the numerical keypad or by selecting a new value from a list. After having edited all values, press the Menu key. The invoked menu contains the menu items OK and Cancel. Selecting OK means accepting the modified values. Selecting Cancel means rejecting the changes. In both cases you will return to the display you invoked last.
gas-lab Q1 Page 23
Primer 4
Gas quality error listing
Monitoring error listing
Next listing
4 Primer for impatient operators:
What do I have to do to …?
Notice: The following instructions are based on the assumption that you are in
the basic device menu.
4.1 ... view the gas quality error listing?
If the status LED blinks red or yellow or is steadily illuminated, the error listing contains alarm or warning entries. Proceed as described below to view the error listing:
1) Invoke the menu (subordinated items) via the Menu key.
2) Invoke the next Menu (other modules) via the Menu key
3) Press the rightward arrow key until you have selected the Monitoring menu item.
4) Press the Enter key. The main display of the Monitoring module is invoked. The error with the highest priority level is displayed. Scroll through the error listing using the arrow keys.
5) Proceed to Chapter 5.3.2 should you need more detailed information. If you would like to accept an error, proceed to the very next chapter.
There may be up to two error listings. In addition to the Gas quality error listing, an error listing of the Monitoring module’s message processing is also available. Always switch to the display of the next error listing via the Next listing menu item. Please refer to Chapter 5.3 for further information on the error listing of the message processing.
gas-lab Q1 Page 25
4 Primer
4.2 … accept the gas quality measurement errors?
Errors can only be accepted and thus removed from the gas quality error listing if they are no longer pending. To accept an error, proceed as follows:
1) Open the error listing via the menu as described above in Chapter 4.1.
2) Select the error you want to accept via the rightward or leftward arrow key.
3) Press the Menu key. If the error indicated on the display cannot be accepted, the first menu item appears as being crossed out: Accept. In this case, press the Menu key again to quit the menu. If the error can be accepted, just select Accept and press the Enter key. The error disappears from the error listing, and the error with next lower priority level will be indicated on the display.
4) Repeat the steps explained above to accept further errors.
Note: Accepting errors as described above only refers to errors in connection with the device’s gas quality functions. How to accept messages of the general message processing (Monitoring error listing) is explained in Chapter 5.3.
Page 26 gas-lab Q1
Primer 4
4.3 … check all parameter settings?
The parameterisation of a gas-net device contains too many settings for them to be conveniently displayed via the operator panel. It is much easer to get a general idea of the device settings with the aid of the GW-GNET+ parameterisation program and a laptop. Proceed as described below:
1. Connect the COM interface at the PC to the DSS interface of the Q1 evaluation computer using a parameterisation cable.
2. Start GAS-WORKS on your computer. Activate the communication program by clicking the Import – Data interface tool in the GW-BASE toolbar.
3. After having successfully started the communication program, you are linked with the connected device data technology-wise. The window appearing on your display shows some important basic device information.
4. Select the Tools tab now. Double-click the Change parameters or the Edit parameterisation entry. The GW-GNET+ interface will appear on the screen. This is where you can invoke and check the parameter listings of the individual modules. Consult the GW-GNET+ online help for details on the operation of the user interface of GW-GNET+.
Please note: The Change parameters or Edit parameterisation service programs also offer the option of changing device settings. The current status of the protection mechanisms (calibration switch / user lock) is of course taken into consideration. Please refer to the GW-GNET+ comprehensive online help for further information.
gas-lab Q1 Page 27
4 Primer
4.4 … check the input values?
1) The input values are in the System module. Therefore, change to this module: Press the Menu key twice, then move to the System entry using the rightward error key and press the Enter key.
2) Then press the Menu key and select the Inputs entry via the rightward arrow key.
3) Press the Enter key afterwards. The Inputs display will be invoked.
5) Please refer to Chapter 5.4.2 for information on how to proceed further.
4.5 … view and check the outputs?
1) The output values are in the System module. Therefore, change to this module: Press the Menu key twice, then move to the System entry using the rightward arrow key and press the Enter key.
2) Then press the Menu key and select the Outputs entry by means of the rightward arrow key.
3) Press the Enter key afterwards. The Outputs display will be invoked.
4) Please refer to chapter 5.4.2 for information on how to proceed further.
4.6 … view the archives?
1) Change to the display of the Data logging module: Press the Menu key twice, move to the Data logging entry using the rightward error key and press the Enter key.
2) Select exactly the archive information you want to view in the appearing dialog. By the way it’s more comfortable to view the archives with the PC­Software. Please refer to Chapter 5.2.2 for a detailed description of the function.
Page 28 gas-lab Q1
Functional description 5
5 Functional description
5.1 Gas quality module
The gas-lab Q1 measuring system is a device for measuring the gas quality of natural gas. It measures the infrared absorption of hydrocarbons and carbon dioxide (CO2) and also the thermal conductivity. The following variables are directly determined based on these measurements:
Hs – Heating value superior, equal to GCV – gross calorific value standard density Rhos concentration xCO2 as mole fraction
Further variables are derived from these direct variables, such as:
Wobbe index superior Ws Methane number MN Composition of the natural gas consisting of 10 components from C1 to
C8+ as well as N2 and CO2. Hi – Heating value inferior, equal to NCV – net calorific value saturated Hs/GCV and Hi/NCV (gas with H2O saturation)
When the device is started, the evaluation computer reads the calibration data out of the non-volatile memory of the sensor system and starts to measure. Therefore, make sure to switch on the sensor system either before or together with the evaluation computer. The measuring operation may only start after the Q1 has been successfully calibrated (see 5.1.2) and if the process gas is injected to path 1 with sufficient pressure. The sensor system is furnished with a pressure switch that detects whether gas is flowing or not. The sensor system must have reached its operating temperature of about 55 to 60 °C before the device is ready to operate. It may take up to one hour to heat up a cold device. The evaluation computer waits until the operating temperature has been reached before indicating valid measurements. View the temperature of the sensor system in the Process values display (see 5.1.1.2).
gas-lab Q1 Page 29
5 Functional description
Indication of the error with the highest priority level
Time and date
Final values for GCV – gross calorific value, standard density, CO
2
concentration
Important: After the commissioning has been completed, operate the gas-net Q1 evaluation computer only with the calibration switch being closed. The calibration switch (the rotary switch on the front of the device) can be sealed for safety reasons. A closed calibration switch ensures that actions to be performed by trained and qualified staff, such as calibrating actions etc., cannot be started from the device.
5.1.1 Display and operation
5.1.1.1 Main display Gas quality
The gas-lab system measures continuously during normal operation and constantly calculates actual values for the target variables gross calorific value GCV, standard density Rhon and CO2 content of the measured gas. The basic display of the gas-net Q1 evaluation computer supplies a survey of the measurement status and currently calculated values of the gas analysis.
The first two lines show the pending gas quality measurement error with the
highest priority level on the left as well as time and date on the right.
The last three lines of the basic display indicate the current gas analysis
values GCV, Rhon, and CO2, written in large numbers. The values shown in the basic display are only current and valid if the system operates normally in undisturbed conditions. This means that the measurement is not in the alarm status and a calibration is not carried out.
Page 30 gas-lab Q1
Functional description 5
„Down“ scroll arrow pointing
downwards: Scroll down­wards with the rightward arrow key.
The following illustration shows an example of the basic display in case of an undisturbed operation:
If you scroll downwards in the main display, further derived values will be indicated as illustrated by the following figures:
A calculated gas analysis with 10 components and the molar percentage of the C2+ components are indicated in addition to the Wobbe (Ws) and methane numbers (MN). Hi is the NCV – net calorific value. The variables Hs’, Rhon and CO2 are corrected values used in the automatic adjustment by means of a process gas chromatograph (PGC). The values satHon and sat Hun are the GCV and NCV in case of saturated concentration of H2O for the defined reference condition. At last the density ratio is displayed.
gas-lab Q1 Page 31
5 Functional description
5.1.1.2 Process values display
The gas-net Q1 evaluation computer provides a special display to indicate the source data measured by the sensor technology, the intermediate values that have been calculated based on these source values and all determined final values. This display is called Process values and is invoked as described below:
1) Invoke the menu in the basic display.
2) Select the Process values entry by pressing the rightward arrow key. Press the Enter key afterwards.
The Process values display contains too many values to be indicated all at once. Scroll up or down the display with the arrow keys.
The upper part of the display shows the last valve positioning command sent by the computer to the valve set as bit string (10010001). Also the number of the current measurement is displayed, which is consecutively numbered. Further­more the current state of measurement is indicated.
All other values are intended for service and maintenance purposes. If you have questions for them please contact Elster.
Page 32 gas-lab Q1
Functional description 5
5.1.2 Calibration
Each gas-lab Q1 measuring system has been calibrated when being delivered, i.e. the device has been calibrated by the manufacturer. During operation an automatic calibration with methane (1-point calibration) is carried out periodically. During the routine testing of the device after one year, we recommend to perform a manual basic calibration with nitrogen, methane and two other calibration gases (3-point calibration).
5.1.2.1 1-point calibration / automatic calibration
Precondition for a successful completion of a 1-point calibration is that the methane is properly connected on gas path 2 (see Cylinder change 5.1.4). For this purpose the cylinder regulator must set be to approx 2 bars and the precision pressure regulator (M2R) of the second gas path must be set to about 80 mbars. If the methane cylinder is not already opened, open it now and do a purging right after it (see 5.1.5). Also the corresponding shut-off valve must be opened.
It is not necessary to open the calibration switch at the gas-lab Q1 for this calibration type.
An automatic 1-point calibration with methane can be carried out during normal measuring in the following conditions:
In a fixed, configurable cycle every 1, 2, 3, 4, 5 or 6 days or 1, 2, 4, 8 or 12
weeks on a specified day at xx o’clock, typically every 7 days.
After a configurable time in hours after each switching on of the supply
voltage for the evaluation computer or sensor system, typically 12 hours later.
After a configurable time in hours after each manual purge of a gas path,
typically 12 hours later.
After a configurable time in hours after errors that may affect the
measurement have been cleared, for instance Pressure disturbed
evaluation computer (A607), typically 12 hours later.
gas-lab Q1 Page 33
5 Functional description
Manually by performing the 1P-cal. start command in the Calibration menu
what is described below.
By setting a parameterised digital input. By means of a DSfG command.
The following reactions of the evaluation computer are indicating the status of the 1-point calibration:
Via measurement outputs and on the controller’s basic display the last valid
gas quality values are maintained provided any values are available.
The basic display shows 1P-cal.
during calibration.
A Revision message will not be
generated.
If you want to abort the function prematurely, do so by keeping to the following instruction sequences:
Invoke the menu from the basic display and select Calibration.
Confirm the selection and invoke the menu again. Then select Cal. Cancel and confirm the selection
The gas-lab Q1 terminates the calibration process and returns to normal operation.
Page 34 gas-lab Q1
Functional description 5
The process will be aborted automatically if an alarm occurs during calibration. Afterwards, the gas-lab Q1 uses the previous calibration data, provided such data exists.
To manually start the 1-point calibration, do so by keeping to the following instruction sequences:
Invoke the menu from the basic display select Calibration and confirm.
Measurement values of different sensors are displayed now.
By pressing the menu key again, the following submenu appears. Select the menu item 1P-cal. start and confirm.
gas-lab Q1 Page 35
5 Functional description
Now the actual status of the 1-point calibration is indicated on the display. Here e.g. the required operation – switch on gas (methane gas path 2)
In case the methane is properly connected to gas path 2 (see beginning of chapter), please invoke the menu and select the menu item gas switched on.
After confirming the calibration with methane proceeds automatically, whereas the remaining time is displayed.
This process starts after approx. 2 minutes for purging and lasts for 10 minutes.
Page 36 gas-lab Q1
Functional description 5
The end of calibration is indicated by the following display:
Then the display changes to the following condition (same as before starting):
After the 1-point calibration is finished, the sensor system automatically injects the process gas and starts measuring.
The new correction values determined during the calibration are written to the calibration archive.
We also recommend observing the deviations of the automatic 1-point calibration in the quality archive.
If these values are getting too high, a new basic calibration should be performed.
gas-lab Q1 Page 37
5 Functional description
5.1.2.2 Basic calibration / 3-point calibration
Such a basic calibration is generally performed on the occasion of the yearly routine testing. Trained and qualified staff must perform the basic calibration. Although it runs automatically, the process steps are supported via the operator panel of the evaluation computer.
The calibration switch at the gas-lab Q1 must be open to facilitate a basic calibration.
The following reactions of the evaluation computer are indicating the status of the 3-point calibration:
During the parameterisation of the gas-lab Q1 you may determine whether
or not the last valid gas quality values shall be maintained via measurement outputs and on the basic display of the computer during this time.
The display shows 3P-cal. during
calibration.
A Revision message will be
generated.
If you want to abort the function prematurely, do so by keeping to the following instruction sequences:
In case you are not in the Calibration module: Invoke the menu from the basic display, select Calibration and confirm the selection.
Invoke the menu and then select Cal. Cancel.
Page 38 gas-lab Q1
Functional description 5
By confirming the selection the gas-lab Q1 terminates the calibration process. The process will be aborted automatically if an alarm occurs during calibration.
Afterwards, the gas-lab Q1 uses the same calibration data as before, provided such data exists.
The basic calibration consists of the following steps:
1. Preparation
2. Zero point adjustment with nitrogen via path 3
3. Calibration with methane (1. calibration gas) via path 2
4. Calibration with H2-11K (2. calibration gas) via path 3
5. Calibration with L1-8K (3. calibration gas), via path 3 optional with a binary gas mixture (5% CO2 in methane)
1. Preparation
For successful completion of a 3-point calibration, the methane cylinder, also used for the automatic calibration, has to be connected correctly to the second gas path (see 5.1.4). The cylinder regulator must be set to approximately 2 bars and the precision pressure regulator (M2R) of the second gas path to about 80 mbars.
If the methane cylinder has not been opened yet, open it and perform a purge first (see 5.1.5).
Furthermore nitrogen and two other calibration gases (H2-11K, L1-8K or binary mixture) will be connected one after another to the third gas path during the calibration process. Since they are not needed for normal operation, they have to be connected first.
The nitrogen is needed first, so his connection to the third gas path is described next and stands also for the other two gases.
gas-lab Q1 Page 39
5 Functional description
As no precision pressure regulator (M2R) has been installed on the mounting plate for the third path normally, mount an additional precision pressure regulator behind the cylinder regulator of the nitrogen cylinder for you to adjust the operating pressure of the sensor system to approximately 80 mbars. This is done in the following way:
Make sure that the nitrogen cylinder has been turned off and the cylinder
regulator is set to zero.
Connect a precision pressure regulator to the cylinder regulator of the
nitrogen cylinder and attach a high-pressure tube to the precision pressure regulator.
Open the precision pressure regulator, so that some gas can flow off in the
next step, whereby the high-pressure tube is purged.
Open the nitrogen cylinder. The cylinder regulator must be set to about 2
bars. Set the precision pressure regulator to approx. 80 mbars Connect the high-pressure tube to the third gas path of the sensor system
finally
For performing a basic calibration the gas-lab Q1 needs to know the exact com­position of the used calibration gases also.
For that purpose the gas analysis of the used calibration gases has to be entered in the actual parameterization.
Support in changing an existing parameterization is given in chapter 6.2 and regarding the affected parameters in chapter Fehler! Verweisquelle konnte nicht gefunden werden. under periphery.
Page 40 gas-lab Q1
Functional description 5
2. Zero point adjustment with nitrogen
While being in the gas quality main display press the menu key and activate the Calibration command in the invoked menu.
A display with measurements of the different sensors appears.
By pressing the menu key again, the following submenu appears. Select the menu item 3P-cal. start and confirm.
Now the actual status of the 3-point calibration is indicated on the display. Here e. g.: The required operation – switch on gas (nitrogen channel 3)
gas-lab Q1 Page 41
5 Functional description
In case the nitrogen is properly connected to gas path 3 (see beginning of chapter), please invoke the menu and select the menu item gas switched on.
After confirmation the zero point adjustment with nitrogen proceders automatically.The process takes about 25 minutes. Meanwhile the display shows the remaining time. It will not count down beforet 2 minutes of flushing time.
3. Calibration with methane (1. calibration gas)
Now the prepared methane has to be switched on, which the following display indicates:
Please invoke the menu and select the menu item gas switched on.
Page 42 gas-lab Q1
Functional description 5
After confirming the calibration with methane (1. calibration gas) runs automatically. The adjustment starts after approx. 2 minutes for purging and lasts for 15 minutes, whereas the remaining time is displayed.
Meanwhile turn off the nitrogen cylinder and remove the tube on both sides.
(Caution: Unpressurise the line first!).
Furthermore you can prepare the second calibration gas (H2-11K) on the third gas path (see step 1. Preparation) When the calibration process with methane (1. calibration gas) is finished, the following display appears:
4. Calibration with H2-11K (2. calibration gas)
Now the prepared H2-11K has to be switched on. Please invoke the menu and select the menu item gas switched on.
gas-lab Q1 Page 43
5 Functional description
After confirming the calibration with H2-11K (2. calibration gas) runs auto­matically, whereas the remaining time is displayed.
The adjustment starts after approx. 2 minutes for purging and lasts for 15 minutes. When the calibration process with H2-11K (2. calibration gas) is finished, the following display appears:
Now turn off the H2-11K cylinder and remove the tube on both sides.
(Caution: Unpressurise the line first!).
5. Calibration with L1-8K or binary gas mixture (3. calibration gas)
Now the prepared L1-8K or binary gas mixture has to be connected. (see step 1. Preparation)Then the prepared 3rd calibration gas is started by opening the menu and selecting the menu item gas switched on.
Page 44 gas-lab Q1
Functional description 5
After confirmation the calibration with the 3rd calibration gas (L1-8K or a binary gas mixture) proceeds automatically. The remaining time is indicated on the displayed. In any case “L1-8K” is shown on the display, even if you use a binary gas mixture instead.
The adjustment starts after approx. 2 minutes for purging and lasts for 15 minutes, whereas the remaining time is displayed
The end of calibration is indicated by the following display:
Then the display changes to the following condition (same as before starting):
After the basic calibration is finished, the sensor system automatically switches back to the process gas and starts measuring.
Now turn close the L1-8K or binary gas mixture cylinder and remove the tube on both sides.
(Caution: Unpressurise the line first!)
gas-lab Q1 Page 45
5 Functional description
5.1.3 Test gas injection
The injection of test gases serves to check the gas quality meter with a known gas or to measure unknown gases on a non-routine basis.
The following reactions of the evaluation computer are indicating the status of the test gas injection:
Via measurement outputs and on the controller’s basic display the last
valid gas quality values are maintained provided any values are available.
A Revision message will be generated.
Please proceed as described below:
1. Decide whether you like to inject the test gas via gas path 1 or 3.
2. Make sure the test gas cylinder is closed and the cylinder regulator is set to zero. Set the associated precision pressure regulator to zero. Connect the outlet of the precision pressure regulator to the first or third gas path of the sensor system using a high-pressure tube. When using gas path 1, make sure the process gas, which is also connected to this path, has been shut­off tightly and thus is not able to affect the measurement by contaminating the test gas.
3. Open the test gas cylinder. The cylinder regulator must be set to about 2 bars.
4. Set the precision pressure regulator to approx. 80 mbars.
5. Now purge the first or third gas path (see 5.1.5).
Page 46 gas-lab Q1
Functional description 5
6. Press the Menu key while being in the gas quality main display and activate the TG path 1 start or TG path 3 start command in the invoked menu.
7. Now, the sensor system will be purged with test gas for about 2 minutes. Afterwards, the measurement begins. The measurement results are shown in the main display.
8. To finish the test gas injection, press the Menu key and activate the TG path 1 end or the TG path 3 end command.
9. Now please turn off the test gas cylinder and remove the tube on both sides.
(Caution: Unpressurise the line first!).
A running test gas injection will be aborted automatically after 1 hour (which can be parameterised), and process gas will be injected instead.
When gas path 1 was used, make sure that after the test gas injection the process gas is reconnected and opened to this path. Otherwise the test gas is handeled as process gas and the measurement results are mapped incorrectly.
gas-lab Q1 Page 47
5 Functional description
5.1.4 Cylinder change
Different gases must be available in the plant for calibration and measurement tasks. Therefore it might be necessary from time to time to connect another gas cylinder to a gas supply connection of the gas-lab sensor housing; for instance, if the pressure of the methane cylinder is insufficient for the 1-point calibration to be carried out.
To remove an existing cylinder, proceed as follows:
1. Close the old cylinder at the main valve on top of the cylinder and unpressurize the connected gas line. You possibly have to wait until the remaining gas has been used up or use an available venting possibility (preferably low pressure side).
2. Remove the cylinder connection from the cylinder and close the cylinder with the corresponding nut.
3. Place the protective cap on the cylinder and fasten it..
4. Not until now detach the protection against falling down (safety chain or clamp) and remove the cylinder.
Page 48 gas-lab Q1
Functional description 5
To connect a new cylinder, proceed as follows:
5. Secure the new cylinder with a chain or clamp against falling down at first.
6. Remove the protective cap from the cylinder by unscrewing it.
7. Make sure that the main valve on top of the cylinder is closed.
8. Remove the nut from the valve connection now.
9. Make sure that the pressure reducing valve and the shut-off valve of the high-pressure regulator or cylinder pressure regulator to be connected are closed. Close the pressure reducing valve by unscrewing its setting screw.
10. Now connect the cylinder connection of the high-pressure regulator or cylinder pressure regulator to the valve connection of the gas cylinder.
11. Open the main valve of the gas cylinder and adjust the outlet pressure of the high-pressure regulator or cylinder pressure regulator to the operating pressure needed by the sensor system for the gas stream you are preparing just now. (i.e.: 1 – 5 bars if in front of an additional low pressure regulator (M2R) or only 80 mbars if directly connected to gas path 3 of sensor system)
To purge the high-pressure or cylinder pressure regulator, proceed as follows:
12. Close the main valve of the gas cylinder and wait until the high-pressure regulatoror cylinder pressure regulator is almost drained, i.e. until the outlet pressure has almost dropped down to 0 bar. Now open the gas cylinder again.
13. Repeat this draining and filling of the high-pressure regulator or cylinder
pressure regulator twice to ensure that the regulator’s dead volume does
not contain gas from the previous use or air.
14. Pay attention to the fact that the gas cylinder must be open finally to make a calibration or measurement possible.
gas-lab Q1 Page 49
5 Functional description
5.1.5 Purging
After having exchanged a gas cylinder with a flammable content, always make sure that no air gets into the sensor system. This is why a purge must be initiated manually after each cylinder change. This also applies if air may have entered in a gas path connected to the sensor system.
The following reactions of the evaluation computer are indicating the status of the test gas injection:
The last valid gas quality values are maintained via measurement outputs,
provided any values are available at all.
The Revision status will not be indicated.
Proceed as described below:
1. Connect the gas cylinder to the relevant gas path. Set the pressure of the cylinder regulator output to 1 - 5 bars and the pressure of the downstream precision pressure regulator to 80 mbars.
2. Press the key while menue being in the main display, select the purging command and confirm it by pressing the key enter.
3. Gas-channel 1 appears on the display.
Page 50 gas-lab Q1
Functional description 5
4. Press the Enter key again. The following selection list appears: Select the gas path you would like to purge via the arrow keys.
Confirm your selection by pressing the enter key again.
5. Press the menu key now and start the purge with OK or quit the dialog via Cancel.
6. The purge lasts for about 7 minutes. The display shows purging.
After the purge has been finished automatically, the gas-lab Q1 starts measuring the process gas again.
gas-lab Q1 Page 51
5 Functional description
5.1.6 Revision
The gas-lab Q1 sets the revision status in case of a zero point adjustment, a 3­point calibration or a test gas injection.
A set revision status means that the gas quality is not measured under normal and proper operating conditions.
All entries in the interval archive are marked with the Revision status note.
Furthermore, the Revision switch open hint will be generated and entered in the error listing and logbook.
The termination of the revision operation results in the ending of the Revision
switch open hint with an entry in the logbook and induces a last entry marked Revision in the interval archive.
The gas quality measurement runs normally again.
Page 52 gas-lab Q1
Functional description 5
5.2 Data logging module
5.2.1 Function
The gas-net Q1 evaluation computer is always equipped with an integrated data logging function. The responsible Data logging module, however, provides only the data logging service. The actual data to be logged is generated by other modules of the module group. When parameterising the Data logging module, you only have to define which of the available archive groups shall actually be logged and which storage depth shall be applicable. All archives defined in that way are designed as ring storage. The data logging depth determines how many entries an archive is able to write at most. If an archive is full, the respective oldest entry will be overwritten by each new entry.
The following sections list the types of archive groups each gas-net Q1 module provides:
Gas quality module: The gas quality module provides 3 archive groups:
The PTB archive logs the billing-relevant data, which consists of the gross
calorific value GCV, CO2 content and status of the measurement system. The data is recorded every 15 minutes. The PTB archive has a size of 180 days.
The Interval archive logs the gross calorific value GCV, the standard
density Rhon, the CO2 content and the status of the measuring system. The data is logged at the full hour and when an error occurs. The Interval archive has a size of about 60 days.
The Quality archive logs 6 quality factors during each 1-point calibration.
These factors provide information on the corrections that have been made during the individual calibrations with methane. The Quality archive has a depth of 365 entries.
gas-lab Q1 Page 53
5 Functional description
Monitoring module:
The Monitoring module facilitates the compilation of process value archives (archives with any measurements or count values). Moreover, it runs the gas
quality measurement’s error logbook. The beginning and ending of all error
types (alarm, warning, hint – see Chapter 5.3.1) are entered in the listing in clear text and together with a time mark. There is an additional, separate error listing for internal message processing, if this function is used. This error listing can also be logged.
System module: The System module runs a parameter change archive (changed settings archive) ), in which changes of the parameterisation are logged. If individual parameters are modified, the old and new values will be logged in addition to the time mark. The module to which the changed parameter belongs is also displayed. A completely new parameterisation will be entered in the changed settings archive as New parameterisation via the data interface when the calibration switch is open. A change of the parameterisation’s operational part is marked as New operational parameterisation.
Note: The archive depth, exact composition and order of archives can be configured by the user via the parameterisation. However, when changing the archive structure, you have to delete the old archives already existing in the device. The devices have already been provided with a pre-defined archive structure corresponding to the common requirements before delivery.
Page 54 gas-lab Q1
Functional description 5
Current calendar time
Input mask for selection of the archive information to be indicated in the next step.
}
5.2.2 Display and operation
Main display (Data logging module) All existing archive entries can be made visible at the operator panel. The main display of the Data logging module consists of a mask in which you may choose the data you want to view more closely. The following illustration shows an example:
After the main display of the Data logging module has been invoked, the most recent entry of the first channel of the first archive group is always indicated first.
Choose the archive information you would like to view :
The archive type is selected at the beginning, i.e. it is backlit in black.
Press the Enter key to get into the edit mode.
A list opens up from which you may choose the desired archive type via the arrow keys.
gas-lab Q1 Page 55
5 Functional description
2
Afterwards, confirm your choice by pressing the Enter key. The next field is selected.
Specify the archive channel you want to view in detail in the next selection field. Several possibilities are offered, depending on the archive group type you have selected in the first step.
For instance, for the PTB archive type you may choose between GCV, CO2 and Status2.
You may specify the point of time to be filtered by changing day, month, year and time one after the other.
You get into the edit mode (as indicated by the black underscore) for each respective field by pressing the Enter key and are thus able to specify date and time step by step.
Use the arrow- or number keys to set the date and time. Confirm your entry via the Enter key
Status stands for a bit string that provides an overview of the status of the gas quality measurement system according to the DSfG Regulations. The meaning of the individual bits is defined in the DVGW documentation Technische Spezifikation für DSfG- Realisierungen (see Bibliography).
Page 56 gas-lab Q1
Functional description 5
3
The next field is selected. Select the entry you want to view via the ordinal number3 in this field.
As there is an unambiguous relation between date and ordinal number, the associated ordinal number is indicated automatically after the date has been entered (see above).
If you enter a different ordinal number, the date in the second line of the selection mask will be adjusted automatically. (See diverent times in the example)
Invoke the selected archive via the menu View (see next section).
Each individual measured variable of an archive group, which is filed by a device, has an ordinal number according to the DSfG Regulations. The sequence of the ordinal numbers is arranged in such a way that the very first archive entry receives the number 1. The ordinal number for each further entry is increased by one. For instance, ordinal numbers are needed for the polling of archive data via DSfG.
gas-lab Q1 Page 57
5 Functional description
The "up" and "down" scroll arrows indicate that you may view even more archive entries.(Scroll with leftward or rightward arrow key.)
Archive entry with ordinal no. 110
Subordinate menu items (Data logging module)
View
By activating the View menu item, you invoke the display of the archive channel you have selected in the main display.
The archive type and current time are shown in the first two lines. Each entry is provided with a time mark. Name, value and unit of the logged data are shown. Exactly two successive archive entries are displayed at once. For example (interval archive gas quality, GCV channel):
Return to the main display, i.e. to the selection mask for viewing archives, via the subordinate menu item Back. To be able to switch to the entries of other channels with the same time mark, exactly the same mask you have specified before will be invoked.
Tip: The archive presentation on the display is limited due to a lack of
space. To analyse larger archive areas it is much more practical to read the archives out of the device and into a PC or laptop before, for example via the data interface by means of the GAS-WORKS module GW-REMOTE+. View the data afterwards in a table or as diagram by means of the GW-XL+ program. The configuration of the data logging function is more comfortable via the PC-Software and therefore not described here. See the Online-Help for further questions.
Page 58 gas-lab Q1
Functional description 5
5.3 Monitoring module
5.3.1 Function
The Monitoring module basically performs the signalling and monitoring tasks of the device:
Monitoring the gas quality measurement for errors.
The error listing serves the user to analyse and remove gas quality measurement errors and to be able to understand the history of the course of error.
Monitoring of any logged measurements. Signalling of operating states. Log-on of process value archives (archives for measurements and count
values which can be used to replace a recording instrument, for example)
Switching functionality
5.3.1.1 Monitoring the gas quality measurement and error listing
There are three different kinds of gas quality measurement errors, which are also treated differently by the error management:
An Alarm is generated if a gas quality-relevant error has occurred.
Beginning and end of an alarm are entered in the associated error listing together with a time mark (<Alarm> begins/ends). If an alarm is no longer pending, i.e. it is no longer relevant, it can be accepted at the device and thus removed from the error listing.
A Warning is generated if a gas quality-relevant variable is affected but
does not influence the result of the gas quality measurement. If a warning is no longer pending, i.e. it is no longer relevant, it can be accepted at the operator panel of the device and thus removed from the error listing.
gas-lab Q1 Page 59
5 Functional description
A Hint is generated if a measurement violates a hint limit specified by the
operator (upper/lower hint limits for gas quality data). In addition, a hysteresis can be defined to avoid fluttering messages when a measurement fluctuates around a limit value. The message <Hint limit begins> will then be generated or removed only if the measurement is not within the hysteresis range. A hint will also be generated if an event occurs which has no influence on the gas quality measurement (e.g. Calibration switch open). The beginning of a hint is entered in the error listing together with a time mark. Hints cannot be accepted and disappear from the error listing automatically after they have ended.
A complete list of all errors together with their classification is included in the annex to this documentation.
The status LED of the device indicates the message with the highest priority level (pending alarm, pending warning, unaccepted alarm, unaccepted warning) according to the table in Chapter 3.2.
5.3.1.2 Measurement indication
In addition to monitoring measurements for hint limits and gradients, the device may also file up to 32 measurements. For each defined measurement the minimum and maximum values measured since the last reset are indicated. The operator can view these values together with the associated time marks on the display.
Page 60 gas-lab Q1
Functional description 5
5.3.1.3 Message processing; messages and groups
Note: The message processing function of the monitoring module runs
completely independent of the device’s gas quality measurement monitoring. Therefore, this functionality of the monitoring module is provided with a separate error listing and an own logbook. Please take into consideration that the rules of this message processing differ considerably from those of the monitoring of the gas quality measurement (see above); for instance, the acceptance mechanism is totally different. If you know the DIN 19 235 standard (Signalling of operating conditions), you may be familiar with many terms mentioned in the description below. The message processing of the Q1 is oriented towards this standard wherever it has proved to be useful and possible with regard to the monitoring function.
The message processing of the Q1 may manage up to 64 messages defined by parameterisation. In principle, there are different types of conditions to be signalled which can be mapped to such a message. The operator himself defines the events that shall result in messages within the message processing by parameterising the device. First, the status of a digital message input (set/not set) can be evaluated as message. Messages generated by the device itself during operation may also be integrated in the message processing, such as a gas quality alarm. Second, measurements can be monitored for limit violation. This applies to measurements supplied via analogue inputs or measured values generated internally. In addition to monitoring upper and lower hint limits, it is also possible to monitor gradients. The gradient monitoring observes measurement changes within defined time ranges: If the difference between the highest and lowest values of a measured quantity exceeds the defined maximum value within a parameterised time range, the condition for generating the message Hint limit gradient has been met.
Note: All single messages referring to measurements are also pending if the associated input value is regarded as being disturbed.
gas-lab Q1 Page 61
5 Functional description
measured
value
Lower hint limit
occurs
Lower hint limit
ends
Lower hint limit
Hysteresis
range
Upper hint limit
Hysteresis
range
Upper hint limit
ends
Upper hint limit
occurs
There are two different methods of avoiding fluttering messages:
1. Hysteresis monitoring (only in case of measurement monitoring) A hysteresis can be defined to avoid fluttering messages if a measurement fluctuates around a limit value. The message <Hint limit begins> will then be generated or removed only if the measurement leaves the hysteresis range.
2. Considering a minimum pending time When a minimum pending time is evaluated, a message is only considered pending if the condition to be signalled (set digital input, hint limit violation, etc.) is pending longer than the minimum pending time lasts.
The target of the message-processing concept is to combine messages in groups. Owing to group monitoring and evaluation the monitoring itself remains clearly understandable even if the amount of signals is rather large.
The user himself may define the scope of the associated error listing. Via the device parameterisation he determines for each message whether a single message shall be entered in the error listing and logbook of the monitoring module. Only if this is the case, the message will be entered in the error listing as soon as it begins. There is no acceptance required for single messages; they are automatically removed from the error listing when they end.
Page 62 gas-lab Q1
Functional description 5
4
If a message is entered in the error listing, beginning and end of this message are also entered in the logbook. The archive module is able to log the logbook of the monitoring module. Individual messages can be disabled during operation. This is rather helpful during a plant inspection, for instance. The message processing completely disregards a disabled message, which is always considered as being not pending. The state of a message (disabled / enabled) can be changed via the operator panel of the device, the parameterisation and a DSfG adjusting telegram. Single message data (characteristics, current values) can be enquired via DSfG. Certain properties are also adjustable via DSfG, such as hint limits and gradients4.
In order to simplify the monitoring of up to 64 messages, single messages can be combined by parameterisation in up to 8 groups in any composition whatever. Each group generates 3 different messages that are created by linking the single messages contained in that group. This way, the operator is able to monitor the status of content-related messages without getting lost in the observation of single messages. The different message types of a group are called: Group message, Held group message and Centralized message.
Group message
A group’s group message results from the OR operation of the single messages belonging to the group. Thus, a group message is pending if at least one of the messages of the group is pending. A group message is not linked with an acceptance.
The DSfG data elements for message processing are included in the DSfG data element tree of the Control entity. Please ask FLOW COMP for a precise list of the supported or used DSfG data elements, when required.
gas-lab Q1 Page 63
5 Functional description
Group
Single message M1
Single message M2
Group message
Held group message
= holding time
Held group message
The held group message only differs from the normal group message by always keeping to a hold time; for instance, to take the telecontrol response times into consideration. A held group message is pending at least as long as the hold time lasts, even if the normal group message ends during the hold time.
Centralized message
A special feature of the centralized message is that it can link the states of the single messages with an acceptance signal. The property Acceptance required or No acceptance required is defined as group characteristic in the parameterisation.
All centralized messages are entered in the logbook and error listing. When a group is marked with No acceptance required, the result of the centralized message is always the same as the result of the group message. In contrast to the group message, however, the no acceptance requiring centralized message is entered in error listing and logbook.
Page 64 gas-lab Q1
Functional description 5
Single message M1
Single message M2
Centralized message group marked no acceptance required (= group message)
Centralized message group marked acceptance required
Group
Acceptance Acceptance Acceptance
Acceptance
The centralized message of a group characterised Acceptance required links the states of the single messages contained in that group with an acceptance signal. The acceptance can either be implemented via an assigned digital input or via a corresponding menu item, i.e. via the operator panel. An acceptance accepts all groups at once, no matter via which source it has been generated.
The centralized message of a group marked Acceptance required is pending if at least one single message of the group is pending. However, it will only end with the ending of the last single message if all pending messages have been accepted. A currently pending single message is also considered accepted if it had been accepted before it ended.
The fact that a centralized message is still pending after the last single message has ended indicates that at least one of the single messages has begun again after the last message has been accepted. In such a case the single message is or the single messages are to be accepted.
The following illustration shows how the centralized message of a group marked No acceptance required differs from the one of a group marked Acceptance required.
These groups with their centralized and group messages are intended to provide an overview of the plant conditions.
gas-lab Q1 Page 65
5 Functional description
This is why the output of these message types is supported in different ways:
In contrast to single messages, centralized and group messages can be
output via digital outputs.
The events Group message begins / Group message ends / Any message
of the group begins / Any message of the group ends can be routed to the DSfG bus with an attention telegram. It can be determined for each group upon which events such attention telegrams shall be generated and of which type they shall be (Alarm, Hint or Warning).
The status of centralized, group and held group messages can be enquired
via DSfG.
Only the centralized message of the messages linked within a group is
entered in the message processing error listing and logbook to keep them well structured.
A Maintenance switch (M-switch) is used to suppress group and/or centralized messages if maintenance work has to be carried out. The setting of this M-switch affects each group differently:
- No influence
A set M-switch does not have any effect on the signalling reactions of this group.
- Suppression “towards telecontrol”
This means that, if the M-switch is set, the group does no longer trigger any DSfG attention telegrams and the Held group message is considered not pending.
- Suppression of all signalling reactions without exception
In this case all messages generated by a group are suppressed if the M-switch is set. This means that the group neither generates held group nor group nor centralized messages. Nevertheless, the centralized message is still entered in the logbook and error listing. If the M-switch is set, the main signaller no longer takes the group into account.
Page 66 gas-lab Q1
Functional description 5
Group
Single message M1
Single message M2
Acceptance Acceptance
Main signaller message
The M-switch is activated via an assigned digital input. It cannot be set via the operator panel for safety reasons.
Status and signalling reactions of single messages remain unaffected by the M­switch. Single messages are suppressed via the disabling mechanism explained above.
In addition to the above-mentioned messages a message named Main signaller is also available. It usually triggers a hooter if being routed to an output. An additional acceptance input is assigned to the main signaller.
The main signaller may include any number of groups.
It starts operating when a message arrives that belongs to one of the involved groups, and it ends upon acceptance, no matter whether or not messages are still pending. The acceptance can be implemented via the digital input assigned to the main signaller. Besides, if all groups are accepted via the operator panel, the main signaller is also accepted. It starts operating again when the next involved message arrives.
The following illustration shows the behaviour of the main signaller in connection with the acceptance (in case of one involved group):
gas-lab Q1 Page 67
5 Functional description
5.3.1.4 Logging on measurement or counter archives
The Q1 is able to run archives for process values, i.e. archives for measurements and count values. Up to four archive groups with up to 8 channels each are available. It can be determined for each archive group in which conditions the contained archive channels shall be logged. The standard procedure used is cyclic data logging (with configurable data logging cycles of one second up to one hour). Additionally or alternatively, the data logging can be made conditional upon certain events occurring during the monitoring of the process values:
It is possible to link the data logging with one or more group(s) of single
messages. In this case the logging takes place upon the beginning and end of each message contained in one of the involved groups.
If the development of a particular process value shall be logged in detail, a
maximum change rate can be defined for an archive channel. Then the entire archive group is logged each time the difference between the current measurement and the value measured during the last data logging exceeds the maximum change rate. This way, the data is logged more often in case of quickly changing values. This option is also available for counter archives, which means you can specify that the data logging shall always start when a count value has increased by a defined difference.
Page 68 gas-lab Q1
Functional description 5
In addition, measurements and count values are always logged if the input value has been identified as being disturbed.
It is possible to “freeze” process value archives ". If an assigned message begins, the corresponding archive continues to log data and stops after a certain number of entries have been reached. The point of time of the last entry will be marked with hint PVA <no.> freeze begins (PVA stands for ProcessValueArchive) in the error listing and logbook of the monitoring module. The data logging will only be continued if the assigned message ends. Simultaneously, hint PVA <no.> frozen also ends.
The freezing of measurement archives is useful if you are interested in how a measurement develops in case a certain event occurs. It may be possible, for instance, to link the freezing with a centralized message. Then, the freezing will be activated upon the first beginning of any message of the group. Normal data logging will not be continued after the end of the last message until the group has been accepted explicitly, for the centralized message only ends in this case.
gas-lab Q1 Page 69
5 Functional description
5
Current time
Error type (Alarm, Warning or Hint)
DSfG error number
Error type (clear text)
Date and time of Beginning and End (the latter is only indicated if the error is no longer pending).
„Up“ and „down“ scroll arrows: There are at least two more errors in the listing.
5.3.2 Display and operation.
Main display (Monitoring module) The display indicates the error listing of the gas quality measurement when you
invoke the Monitoring module. The error message with the highest priority level5 is displayed, provided the error listing is not empty. If the error listing contains further errors, you may scroll through the error listing via the arrow keys. The little black scroll arrows on the right side of the display indicate whether or not there is more than one entry in the listing. The following information is provided for each error:
An alarm has the highest priority level, i.e. it is of greatest importance. A warning has a higher priority level than a hint. Cf. Chapter 5.3.1 for an explanation of the terms Alarm, Warning and Hint.
Page 70 gas-lab Q1
Functional description 5
6
6
Subordinate menu items (Monitoring module)
Accept The Accept menu item only exists for the gas quality measurement error listing.
If this menu item is crossed out, the indicated message cannot be accepted because it is still pending. If Accept is not crossed out, however, the currently indicated error message can be accepted and thus removed from the error listing. After the message has been accepted, the menu is cleared and the next error will be displayed. Just repeat the procedure if you would like to accept this error, too: Invoke the menu and confirm Accept.
Accept all (only available for the Monitoring error listing)
The Accept all menu item is only available if you are in the error listing display of the monitoring module; via this menu item you accept all groups requiring an acceptance together with the main signaller (hooter). Accept all is crossed out if there is no need to accept any messages.
Next listing Via Next listing you may switch to the respective following error listing. There are two error listings: the Gas quality measurement error listing and the Monitoring error listing.
The following messages are entered in the error listing of the monitoring module:
Messages named System messages
have top priority. These messages are usually not visible in the error listing as they are pending for only one second.
The next position shows an M-switch message if the maintenance switch
has been activated.
Then, all pending centralized messages of groups 1 to 32 are indicated.
System messages are generated for the events Re-start performed, Supply voltage
failure, New parameterisation, and Parameter changed.
gas-lab Q1 Page 71
5 Functional description
Name of the shown measurement. Pressing the Enter key opens up a selection list containing all measurement names. You may switch to the indication of another measurement via this list.
Current measured quantity of the associated measurement, incl. unit. Blinks if the value is disturbed.
Line before last: Minimum value since last resetting, with time mark. Last line: Maximum value since last resetting, with time mark.
Pending single messages follow according to their priority level.
In order to keep the error listing clearly understandable only such single messages are entered in the error listing that have been parameterised accordingly.
In principle, all single messages of the message processing and the M-switch message do not require an acceptance. Only centralized messages of groups requiring an acceptance must be accepted. This means that all single, group and centralized messages of groups that do not require an acceptance automatically disappear from the error listing when they stop pending. Centralized messages of groups requiring an acceptance will only end after they have been accepted and if none of the single messages of the group is pending any longer. The messages can be accepted even before the last single message has ended.
Measurements
The gas-net Q1 evaluation computer may have up to 32 parameterised measurements for maximum/minimum indication on file. If you invoke the corresponding display, the first measurement will be presented. You can switch to the display of another measurement via the name.The display contains the following information in detail:
If the associated measurement is indicated as being disturbed, the display keeps showing the last valid value. In this case, however, the value is blinking.
Page 72 gas-lab Q1
Functional description 5
Name of the indicated group
Initial message, i.e. the message of the group that has begun first. The text is blinking if the message has not been accepted via the group yet.
New message, i.e. the message of the group that was the last one to begin. The text is blinking if the message has not been accepted via the group yet.
The measurement display contains the following subordinate menu items:
Resetting
The Resetting menu item sets the minimum and maximum values of the measurement to the current measurement.
Reset all
The Reset all menu item resets all filed measurements.
Groups
The Groups menu refers to the message processing of the monitoring module. You may invoke the associated display only if message groups have been created. The display facilitates an overview of the event status of the messages linked in a group. If you invoke the associated display, you will see information on the first group. For switching to the display of another group you have to get into the edit mode (by pressing the Enter key) and select the name of the group you want to view from the appearing list. Confirm your choice by pressing the Enter key again.
If the centralized message of the invoked group is currently pending, the single message of the group that has begun first is indicated below Initial
message.
gas-lab Q1 Page 73
5 Functional description
Name of the selected group
Currently indicated single message of the group. A list with all messages existing in this group opens up if you press the Enter key. You may also switch to the indication of another single message of the group via this selection list.
As this is always the message that has initially caused the beginning of the centralized message, this information helps to analyse the error status. The message text is blinking if no message has been accepted yet after the beginning of the centralized message. Below the New message menu item you can read off the most recent single message of the group that has begun. As long as the centralized message itself is pending the New message indication is updated upon each beginning of a new message of the group. A New message, too, does not stop blinking on the display before a message has been accepted after the beginning of the new message. Only if neither initial nor new message are indicated, a centralized message is not pending. Activate the View submenu to check the status of all single messages contained in the group:
View This display provides information on the single messages contained in the group invoked before. Besides, it provides information on the status of each contained single message. Due to the mass of information there is always only one single message indicated on the display at a time. As the Name selection list contains all single messages of the corresponding group you can switch to the indication of any other single message of the currently selected group.
Page 74 gas-lab Q1
Functional description 5
Seconds counter for minimum pending time
The following information is indicated for each individual message:
Name
Each message is identified via its parameterised name. You may change to the indication of any other single message of the group via the Name selection list.
Type
A single message can be derived from a Message, the monitoring of an Upper or Lower limit value or Gradient monitoring.
Status
The status of a single message is characterised by several specifications: On or Off indicates whether the message is pending. Here, a possibly parameterised minimum pending time is taken into consideration: Even though the status to be signalled is currently pending, the associated single message will only be set if the status is pending longer than the minimum pending time. To be able to make out whether a single message would be pending if no minimum pending time would have been assigned to it, the display includes a seconds counter.
gas-lab Q1 Page 75
If the seconds counter starts counting backwards, the status to be signalled has already been pending. A minimum pending time is assigned to each single message. This is why the message itself is only generated if its status is still pending after the minimum pending time has elapsed. Therefore, the counter constantly shows how many seconds of the minimum pending time are remaining. Only after the counter has reached 0, the status indication of the message will change from Off to On.
5 Functional description
If the status to be signalled ends before the minimum pending time has elapsed, the counter will be set to 0 but the message will not begin. The counter always shows 0 if a minimum pending time has not been assigned to a single message. In this case the message begins immediately after the status to be signalled has been detected. However, it depends on its state (disabled / enabled) whether a single message is being routed at all: If a single message has been disabled, the entire subsequent message processing considers it as not pending.
Tip: The main advantage of the GroupsView menu described here is the possibility to analyse the status of a group and the single messages contained therein. This menu only provides information but does not allow to change the properties of a single message, for instance to set a disabling mark. For this purpose the Single messages menu must be invoked. The display of this menu is structured similar to the one described above.
Single messages
The Single messages menu refers to the message processing of the monitoring module. It does not only serve to indicate basic information relating to single messages but also provides the possibility to disable such messages via the operator panel of the device and to change limit values for single messages of the measurement monitoring. In order to switch to the display of another single message you have to get into the edit mode (by pressing the Enter key) and select the name of the message you want to view from the appearing list. Confirm your choice by pressing the Enter key again. The Single messages display is basically identical with the display of the Groups – View menu described above. The parameterised limit value is indicated as additional information in case of a message derived from measurement monitoring.
Page 76 gas-lab Q1
Functional description 5
Pressing the Enter key opens the selection list containing all currently blocked messages.
Limit, here: upper limit value. Note: An additional hysteresis may have been adjusted via PC and with the parameterisation program to avoid fluttering messages
Limit
The Limit menu item indicates the currently set limit value of the limit monitoring or the maximum permissible change of the measurement within a parameterised monitoring period when monitoring gradients.
This dialog allows disabling single messages via the operator panel. It is also possible the change the limit value to be monitored in case of measurement monitoring messages. However, the numerical locks must have been opened for such actions to be carried out.
Disabled messages
The Disabled messages menu refers to the message processing of the monitoring module. The corresponding display lists all currently disabled single messages. Additional information indicated on the display comprises the type of the message, the real message status before the disabling, and possibly the counter for a monitored minimum pending time.
The Disabled messages menu described here merely provides information. Please switch to the Single messages menu (see above) if you want to change the status of a single message (disabled / enabled).
gas-lab Q1 Page 77
5 Functional description
This is the last subordinate menu of the Monitoring module: Switches (Change of switch outputs)
The task of the Switch menu is to display the status of the Switches. A switch is a message with a preset status (on or off), which can be changed manually, i.e. via the operator panel of the device. Switch-type messages are routed to digital outputs (cf. Chapter 5.4.1) to trigger switching operations of any kind from the gas-net device. Three switches are available (Switch 1 to Switch 3). To change the status of a switch you need to invoke the Switches menu, which indicates the current status of each switch. Select the switch the status of which you would like to change and change into the edit mode. Afterwards, you may select the other switching status and confirm via OK. Changing a switch status is subject to the user lock.
Note: Message outputs may be parameterised in such a way that they invert the output signal during operation. Please consider that this setting also applies to the output of switches.
Note: The Switches menu offers with Revision 1 and Revision 2 two additional switching options. However, the gas-net Q1 does not use these options.
5.4 System module
5.4.1 Functions
The System module comprises all basic functionalities of the gas-net device.
Inputs
One such basic functionality of the System module is the processing of the input information and the transmission of this information to other modules. The availability of input channels depends on the process board equipment of each individual device. A Q1 always contains at least one LMFA7 board for the connection of the sensor system. Besides that, there are further process boards providing input channels (see Chapter 7.2.2).
The System module’s display offers a special menu for viewing the source, source value and final value of the input signal on the device display (see Chapter 5.4.2).
Page 78 gas-lab Q1
Functional description 5
Outputs
The evaluation computer always contains one LMFA7 board. One LMFA7 output board offers 3 transistor outputs for the output of messages. Furthermore, it contains 4 current outputs for the output of measurements.
In addition to that, there are further gas-net process boards providing output channels (see Chapter 7.2.2). The gas-net evaluation computer provides a special menu for viewing and checking outputs (see Chapter 5.4.2). Checking outputs is only possible if the locks are open.
DSfG (optional feature for the European market) DSfG is a digital interface especially developed for data communications
between gas meters (such as gas-net devices).
If you need further information, please contact your local agent.
Please see www.elster.com
gas-lab Q1 Page 79
5 Functional description
7
7
Locks
The parameter list of the Q1 contains various parameters that are not subject to the calibration lock but protected against wilful changes by an operational safety mechanism. This protection mechanism is called user or numerical lock. Each of the two contract parties may define a lock of up to 6 digits. These locks are opened and closed via the keypad (see Chapter 5.4.2). The device parameterisation defines the locks that shall be valid.
Setting the clock
A free setting of the device internal clock is only possible if the calibration switch is open. If the latter is closed, you may only reset the clock within a range of ± 20 seconds (and only once within 24 hours). This restriction applies to all time sources, especially to the manual clock setting via the device’s operator panel. If you were successful in resetting the time outside the ± 20 seconds-range, two warnings will be generated. Unsuccessful adjustments result in Clock-synch failed warnings.
DCF-77 (optional feature for the European market)
Optional the gas-net Q1 can have a DCF77 interface for connecting a radio ¬ clock that receives the PTB time normal about DCF-77 radio signal.
If you need further information, please contact your local agent.
Please see www.elster.com
According to the DSfG Specification: W811: Clock set new W820: Clock set old W812: Clock-synch failed
Page 80 gas-lab Q1
Functional description 5
5.4.2 Display and operation.
Main display (System module) The main display of the System module shows basic device information. The following details are indicated:
device name with software variant product version of the device software serial number of the relevant device check sum for software authentication (will be calculated after the display
has been invoked)
number of operating hours of the device
Subordinate menu items (System module)
Inputs The Inputs display enables you to view detailed information of individual
analogue input channels (name, board/channel, input value, etc.). After you have activated the Inputs menu item, the first channel of the input
board will always be indicated at board location 1. For instance, the display of the first channel of an EXMFE4 board looks as follows:
To switch to the display of another channel change to the edit mode (by pressing the Enter key) and select the name of the desired input value from the appearing list. After you have confirmed your choice by pressing the Enter key, the display of the selected board will be invoked.
gas-lab Q1 Page 81
5 Functional description
Name
Name of the output channel. Serves to identify the channel and is defined by the device parameterisation.
Board
Board location, board type, version number of the board software.
Channel
Channel number with operating mode.
Phys. value
The final value of the respective physical variable with physical unit.
Output value
In case of analogue values: value for the current output (in mA) calculated on the basis of the final value. In case of messages: output contact closed / open. In case of pulse outputs: frequency in Hz, number of pulses pending in the pulse buffer.
Test value
Only for output testing (see below).
Outputs The system menu Outputs enables you, similar to the input channels’ display, to
view the assignment and current status of the output channels. After you have activated the corresponding menu item, details of the first defined output channel will be presented:
To change to the display of another channel, activate the edit mode (by pressing the Enter key). A list with the names of all parameterised output channels appears. After you have chosen the desired output channel and have pressed the Enter key, the display of the selected channel will be invoked.
Page 82 gas-lab Q1
Functional description 5
Output testing In addition to just viewing the current process values you may also check
the output channels. This means that you can output any current value between 0 and 20 mA via an analogue output for testing purposes. A message output can be set to Contact closed or open, and you may also output pulses for testing reasons. However, only one output can be switched into the test mode at the same time. Proceed as described below:
1) Select the desired output channel, and press the Menu key afterwards.
2) Select the Test on submenu.
The Test value line is selected automatically. Press the Enter key to switch to the edit mode. If you enter a new value now and confirm it by pressing the Enter key, this value will be output via the selected output. In the activated test mode the Test value indicated in the last line is no longer crossed out. The Output value menu item also indicates the output test value.
3) To quit the test mode, press the Menu key again and confirm the
Test off menu item.
Note: You may also quit the test mode by quitting the current display, for instance via Back or by selecting a different channel.
gas-lab Q1 Page 83
5 Functional description
DSfG (optional feature for the European market)
DSfG is a digital interface especially developed for data communications between gas meters (such as gas-net devices).
Sub menu DSfG
Operaitng mode menü
Among the „own participants” bus addresses (EADRs) lists the Q1 internal instances..
Menu known participants
Menu line: Bus activityät:
The Q1 can get an attention-Telegram .
If you need further information, please contact your local agent.
Please see www.elster.com
Page 84 gas-lab Q1
Functional description 5
DCF77 (optional feature for the European market) Optional the gas-net Q1 can have a DCF77 interface for connecting a radio ¬
clock that receives the PTB time normal about DCF-77 radio signal.
If you need further information, please contact your local agent.
Please see www.elster.com
Locks. After you have activated the Locks menu item, a display is invoked indicating the current status of the user locks. The display is empty if no lock has been defined. New lock digits must always be defined by parameterisation.
The submenu contains the following items:
Open This menu item is crossed out if all locks are open. If at least one lock
is closed, you may activate the Open menu item by pressing the Enter key. A dialog will be invoked in which you may enter the numerical lock(s):
If only one lock has been defined, only one line will be shown here. Select the lock you would like to open in the first step. Change to the
edit mode via the Enter key as usual and enter the lock’s string of
digits. Confirm the code by pressing the Enter key. If the entered string of digits is incorrect, it is not possible to quit the edit mode via the Enter key. You have to correct the code first (or quit the menu via the rightward arrow key). It is also possible to change the status of just one lock. Press the Menu key after you have successfully entered the numerical lock(s). Open the locks with OK or leave them closed by confirming Cancel.
gas-lab Q1 Page 85
5 Functional description
8
Close All defined locks are closed.
This menu item appears being crossed out and is thus not selectable if all locks are closed. If the calibration switch is open, all locks are automatically open, too, and cannot be closed.
Set clock Activate this menu item to adjust the internal device time.
You may re-enter time and date step by step via the numerical keypad in the edit mode, i.e. after you have pressed the Enter key. The time zone menu item offers a selection list (summer/winter time) for you to choose the seasonal time zone as soon as you are in the edit mode.
However, it is also possible to operate the device throughout the year without a seasonal time change. The seasonal time change can be prevented by parameterisation. In such a case, the phrase not used appears next to Time zone, and it is not possible to change to the edit mode here.
After having set the time correctly, invoke the menu. Return to the main display without accepting the new time adjustment via Cancel. You may only confirm your setting via OK if the acceptance of the new time is permissible8.
Display test You may check whether or not the display and status LED function correctly by
performing the display test. After you have activated the Display test menu item, the pixels of the display are alternately switched on and off. At the same time the status LED shows all three colours after each other. Finish the display test via Back.
If the calibration switch is closed, resetting the clock is only permissible within a range of
± 20 seconds and only once within 24 hours.
Page 86 gas-lab Q1
Functional description 5
5.5 Integrated RDT module
(optional feature for the European market and, not available with all software variants)
5.5.1 Function
The optional function block has the task of realizing a data connection between the gas-net device and a remote control center. The center is a couple of him during an active data connection to be logically equal participant in the data traffic.
If you need further information, please contact your local agent.
Please see www.elster.com
5.5.2 Display and operation
Main display (Modul Integrated RDT)
The main display of the module Integrated RDT indicates in what state the RDT is.
Optional feature for the European market and, not available with all software variants.
If you need further information, please contact your local agent.
Please see www.elster.com
gas-lab Q1 Page 87
5 Functional description
Subordinate menu items (modul intregrated RDT)
Optional feature for the European market and, not available with all software variants.
Historie
Menu RDT-statistics
GSM (only when connected to a wireless modem)
Optional feature for the European market
If you need further information, please contact your local agent.
Please see www.elster.com
Page 88 gas-lab Q1
Functional description 5
9
5.6 DSfG module, Data exchange module
(Gateway functionality, not available with all software variants)
The DSfG protocol is a digital data protocol, especially developed for the communication between gas meters in gas measuring and regulating stations. The gas-net Q1 is always equipped with a DSfG interface. All important measurements and characteristic data are provided as DSfG data elements via the DSfG protocol according to the specification. These data elements are addressed via special addresses. However, gas measuring and regulating stations often also contain a PLC (programmable controller) the tasks of which are the open- and closed-loop control of the station and the telecontrol interfacing to a remote control centre. Most of the PLCs are not DSfG-capable but use other digital communication protocols9. Consequently, they cannot be simply included in the DSfG station group. In such cases a protocol interfacing between DSfG and host protocol is necessary (gateway functionality). This functionality is only available in Q1 software variants containing the DSfG and Data exchange modules.
5.6.1 DSfG module (Optional feature for the European market)
The optional module controls the DSfG side of data exchange.
If you need further information, please contact your local agent.
Please see www.elster.com
Typical protocols are, for instance, 3964R/RK512 (e.g. in case of Siemens S5/S7) or
Modbus-RTU (e.g. in case of Cegelec Modicon).
gas-lab Q1 Page 89
5 Functional description
5.6.2 Data exchange module
5.6.2.1 Function.
The Data exchange module controls the host side of the gateway functionality. All data points that are available during operation and shall be forwarded to a host must be positioned on registers or data blocks/data words for the host. It is also possible to scale the value in a suitable way before forwarding it, depending on the data type. Usually, the data to be forwarded is the content of DSfG data elements that have been polled by the DSfG module via a DSfG query telegram. The forwarding of the received DSfG data to the host is defined in the Data exchange module. The data transport in the direction to the host is processed cyclically. If a new telegram arrives from the DSfG side, the associated export values will be updated. In addition to the specification of the destination in the host, the description of the export value also defines the way of positioning the data point in the host (e.g. size, scaling, limits, and bit position). A particular import value describes for the reverse data flow direction at which position in the host a data point shall be collected and in which way it shall be converted. It can be defined in the parameterisation whether the transmission to the DSfG side shall be cyclic or event-driven. Any number of import values from the host can be combined to a job group. All data points of a job group are always put into the transfer memory together and simultaneously. However, only after all of them have been collected again once more. Moreover, a job group may be linked with a trigger. A trigger is a defined position in the host and initiates a new processing of a job when its content has changed.
Page 90 gas-lab Q1
Functional description 5
5.6.2.2 Protocol-specific facts: RK512 / 3964R
The gateway functionality of the gas-net Q1 supports the protocol consisting of 3964R (data link layer) and RK512 (data presentation). Look up the settings of the protocol parameters in exactly this form in the manuals of the programmable controller and adjust them in the communications processor procedure. The implementation of the RK512 protocol in the gas-net device is restricted to the data transmission between data blocks. The smallest information unit is a data word. Such a data word has a length of 16 bits and may contain, for instance, several messages, a status overview in the form of a bit string, or a binary measurement. The structure of the management of these data words is rather simple; there are up to 256 larger units, called data blocks, each of which may contain up to 256 data words. Data blocks and data words are numbered consecutively from 0 to 255, with the numbers serving as addresses. A single data word is clearly defined by exactly three details:
the address of the data block (0..255) the address of the data word (0..255) the content of the data word
The addresses to be assigned to the host data in the gas-net parameterisation therefore consist of a data block/data word (DB/DW) pair. Data block 0 is usually used for intrasystem host variables. This is why you should use data block 0 neither for exporting values towards the host nor for importing host values.
In special cases it may happen that the host only masters the subordinate 3964R protocol layer. This means that the following restriction applies: As no DB/DW information is transmitted on the 3964R protocol level, all data to be exported or imported must be arranged as a string. The number of dates is thus limited to 64 data words per transmission. This is why in the GW-GNET+ interface (Data exchange module) the positions for import and export values are not indicated in the DB/DW form but as serial register numbers.
gas-lab Q1 Page 91
5 Functional description
Communication protocol:
RK512/3964R – point-to-point
Interface:
V24 (RS232) with PC/AT pin assignment at the gas-net device (COM2-interface). RS232 or RS422 via serial interface of an MSER2-board in the gas-net device.
Transmission rate:
adjustable to 2400, 9600, 19200 or 38400 baud
Transmission parameters:
8 data bits, even parity, 1 stop bit
3964R telegrams:
Max. 128 bytes (64 DW) per telegram
RK512 response telegrams:
A maximum of 64 data words per telegram, no 3964R response telegrams. In the RK512 slave operating mode (i.e. host = master), the gas-net device generates the response telegrams listed in the table below:
The necessary programming of the host can be minimised by selecting the Master operating mode for the gas-net Q1. In this case it is not necessary to compile a job list for the host. The entire data block assignment is carried out via the gas-net device parameterisation. However, at least the agreed data blocks and the basic protocol functionality must have been created in the host of course. The communication between the host computer and the gas-net Q1 is subject to several basic requirements. The following list provides a summary:
Page 92 gas-lab Q1
Loading...