Elster EK260, 83 462 140 Operating Manual And Installation Instructions

EK260
Volume Conversion Device EK260
Operating Manual and Installation Instructions
Operating Manual: 73016960 SW Version: from V2.52
Issued: 13.03.08 (p) Edition:
All rights reserved. Copyright © 2008 Elster GmbH, D-55252 Mainz-Kastel
All the figures and descriptions in this operating and instruction manual have been compiled only after careful checking. Despite this however, the possibility of errors cannot be completely eliminated. Therefore, no guarantee can be given for completeness or for the content. Also, the manual cannot be taken as giving assurance with regard to product characteristics. Furthermore, characteristics are also described in it that are only available as options.
The right is reserved to make changes in the course of technical development. We would be very grateful for suggestions for improvement and notification of any errors, etc.
With regard to extended product liability the data and material characteristics given should only be taken as guide values and must always be individually checked and corrected where applicable. This particularly applies where safety aspects must be taken into account.
You can obtain further support from the branch or representative responsible for your area. You will find the address in the Internet or simply enquire at Elster GmbH.
Passing this manual to third parties and its duplication, in full or in part, are only allowed with written permission from Elster GmbH.
The guarantee becomes invalid if the product described here is not handled properly, repaired or modified by unauthorised persons or if replacement parts which are not genuine parts from Elster GmbH are used.
Mainz-Kastel, March 2008
2 Elster GmbH
Contents
I Safety instructions.......................................................................................................5
II Items supplied and accessories.................................................................................6
1 Brief description ..........................................................................................................7
2 Operation......................................................................................................................9
2.1 Front panel...............................................................................................................9
2.2 Display...................................................................................................................10
2.2.1 Line 1 = Labels...........................................................................................................10
2.2.2 Line 2 = Value with name and unit ............................................................................. 12
2.3 Keypad...................................................................................................................13
2.3.1 Changing values ........................................................................................................14
2.3.2 Entering sources........................................................................................................15
2.3.3 Entry errors................................................................................................................15
2.4 Access rights..........................................................................................................16
2.4.1 Calibration lock ..........................................................................................................16
2.4.2 Calibration logbook ....................................................................................................17
2.4.3 Supplier's lock and customer's lock............................................................................ 17
2.5 Formation of the list structure.................................................................................17
3 Functional description...............................................................................................22
3.1 Standard Volume (Volume at base conditions) list.................................................23
3.2 Actual volume (volume at measurement conditions) list..........................................25
3.3 Pressure list...........................................................................................................28
3.4 Temperature list.....................................................................................................31
3.5 Volume corrector list ..............................................................................................34
3.6 Archive list..............................................................................................................37
3.6.1 Device numbers and channel numbers for WinView and WinLIS............................... 40
3.6.2 Find function for checking the archive entries............................................................. 40
3.7 Status list................................................................................................................41
3.7.1 List of status messages .............................................................................................. 44
3.7.2 Status register addresses........................................................................................... 50
3.8 System list..............................................................................................................51
3.9 Service list..............................................................................................................53
3.10 Input list..................................................................................................................58
3.11 Output list...............................................................................................................64
3.11.1 Parameterising the HF output .................................................................................... 69
3.11.2 Brief summary of output parameterisation..................................................................70
3.12 Interface list............................................................................................................71
3.12.1 Printer log ..................................................................................................................78
3.12.2 Automatic setting of the clock by remote data transmission .......................................80
3.12.3 Modbus parameters...................................................................................................82
3.13 Energy list ..............................................................................................................85
3.14 User list..................................................................................................................87
4 Applications ...............................................................................................................88
4.1 Rated operating conditions for the various conversion methods............................88
4.2 Application as high flow display device ..................................................................90
4.3 Application as flow recording device......................................................................91
4.4 Connection of a counter with LF pulse transmitter .................................................91
4.5 Connecting a meter with encoder...........................................................................92
Elster GmbH 3
4.6 Application in areas subject to explosion hazards ................................................. 93
4.6.1 Applications in Zone 1................................................................................................ 93
4.6.2 Applications in Zone 2................................................................................................ 93
4.7 Applications for Interface 2.................................................................................... 94
4.7.1 Modem with control signals (standard modem) .......................................................... 94
4.7.2 Modem without control signals................................................................................... 94
4.7.3 FE260 Function Expansion with modem ....................................................................95
4.7.4 FE260 Function Expansion without modem ............................................................... 95
4.7.5 FE230 Function Expansion with modem ....................................................................95
4.7.6 Printer on the EK260 or on an FE260.........................................................................96
4.7.7 Other device with RS232 interface (e.g. PC).............................................................. 96
4.7.8 Modbus protocol.........................................................................................................96
4.7.9 Sending short messages by SMS...............................................................................97
4.7.10 Standard output data records for process data ("three-minute values")...................... 97
5 Installation and maintenance ................................................................................... 99
5.1 Installation procedure ............................................................................................ 99
5.2 Mounting.............................................................................................................. 100
5.2.1 Wall mounting ..........................................................................................................100
5.2.2 Meter superstructure................................................................................................ 101
5.3 Cable connection and earthing............................................................................ 102
5.4 Terminal layout.................................................................................................... 103
5.5 Connection of the serial interface ........................................................................ 105
5.5.1 Modem with control signals...................................................................................... 105
5.5.2 Modem without control signals.................................................................................106
5.5.3 Printer...................................................................................................................... 106
5.5.4 Other devices with RS-232 interface ........................................................................107
5.5.5 FE260 Function Expansion (optionally with modem)................................................ 107
5.5.6 FE230 Function Expansion ...................................................................................... 108
5.6 Connection of a low-frequency pulse transmitter (reed contacts) ........................ 108
5.7 Connection of an encoder.................................................................................... 109
5.8 Seals.................................................................................................................... 109
5.8.1 Seal layout of basic device....................................................................................... 110
5.8.2 Seal layout of temperature sensor............................................................................112
5.8.3 Sealing layout of pressure sensor ............................................................................113
5.9 Battery replacement............................................................................................. 114
A Approvals................................................................................................................. 116
A.1 EC Declaration of Conformance .......................................................................... 116
A.2 Approval for Ex Zone 1........................................................................................ 117
B Technical data ......................................................................................................... 122
B-1 General data (mechanical, terminals, ambient conditions).................................. 122
B-2 Batteries.............................................................................................................. 122
B-3 External power supply......................................................................................... 123
B-4 Pulse, status and encoder inputs......................................................................... 123
B-5 Signal and pulse outputs..................................................................................... 124
B-6 Optical serial interface......................................................................................... 124
B-7 Electrical serial interface (internal)....................................................................... 124
B-8 Pressure sensor .................................................................................................. 125
B-9 Temperature sensor ............................................................................................ 125
B-10 Measurement uncertainty .................................................................................... 125
C Index......................................................................................................................... 126
4 Elster GmbH
I Safety instructions
F The connections of the EK260 are freely accessible during setting up. Therefore, make
sure that no electrostatic discharge (ESD) can occur in order to avoid damage to the components. The person carrying out the installation can, for example, discharge himself/herself by touching the potential equalisation line.
F To avoid erroneous operation and problems, the operating manual must be read
before putting the EK260 into operation.
The EK260 Volume Conversion Device(Volume Conversion Device) is suitable for applications in Ex Zone 1 for gases in the temperature class T4 (ignition temperature > 135°C, e.g. natural gas) according to VDE 0170. (see Appendix A-2)
In this application it is essential to take note of the following information:
F Follow the regulations in the relevant standards, in particular DIN EN 60079-14 (VDE
0165 Part 1) and DIN EN 50014.
F Make sure that the limits quoted in the certificate of conformance (see Appendix A-2) for
the devices to be connected are not exceeded.
F The housing of the EK260 must be earthed directly to a potential equalisation strip. A
terminal screw is provided for this on the left housing wall.
Elster GmbH 5
II Items supplied and accessories
Items supplied:
The items supplied with the EK260 include:
a) EK260 Volume Conversion Device b) Dispatch list c) Configuration data sheet d) Operating Manual e) 3 blind insertion seals for gland-type cable entries f) Lead sleeves for sealing the pressure connection.
Ordering information and accessories
EK260 Volume Conversion Device, complete
EBL 50 Thermowell, complete with M10 x 1 weld-in sleeve
EBL 67 Thermowell, complete with M10 x 1 weld-in sleeve
EBL 160 Thermowell, complete with G 3/4" weld-in sleeve and sealing ring
EBL 250 Thermowell, complete with G 3/4" weld-in sleeve and sealing ring
Three-way test tap
Shut-off ball valve with Ermeto 6L test connection
Minimess test connection
Operating manual, German
Operating manual, English
Operating manual, French
Plug-in terminal, 2-pole black
Calibration covering cap
Order no. 83 462 140 73 012 634 73 014 456 73 012 100 73 012 100 73 008 403 73 016 166 73 016 167 73 016 960 73 017 115 73 017 218 04 130 407 73 016 879
Battery module, 13 Ah
6 Elster GmbH
73 015 774
1 Brief description
The EK260 Volume Conversion Device is used for the conversion of the gas volume measured at line conditions by a gas meter into the base conditions and into the appropriate energy.
The pressure and temperature are measured for the determination of the line conditions. The inverted compressibility factor ratio (K-value) can alternatively be calculated according to S-GERG 88 or AGA-NX19 or it can be entered as a constant. The volume is converted into energy using the adjustable calorific value.
The integral recording device contains meter readings and maxima from the last 15 months and the consumption profile of the last 9 months for a measurement period of 60 minutes.
Power supply:
Battery operation with a service life depending on operating mode ≥ 5 years.
Optional double battery life is possible by connection of an additional battery pack.
Battery replacement possible without loss of data and without violation of calibration seals.
Data retention without battery supply due to internal EEPROM.
Connection for external power supply unit.
Operator interface:
Alphanumeric display with two lines of 16 characters.
A display list freely assignable by the user.
Programming via keypad possible.
Calibration switch (separately sealed in the device).
Calibration logbook according to PTB-A 50.7 for changing values relevant to calibration
without the calibration lock.
Two user locks (supplier's and customer's locks) with numerical codes.
Access rights for each individual value can be set separately via interface (with appropriate
rights).
Counter / signal inputs:
3 inputs for reed contacts or transistor switches, programmable as pulse or signal inputs.
Connection provided for an Elster GmbH C1 Encoder Counter for digital transmission of
genuine counter readings (also in battery mode).
Maximum counting frequency 2 Hz (adjustable).
Pulse value for each input separately adjustable, also non-decade.
Various counters for Vb and Vm and for each input (main counter, original counter,
disturbance volumes, totaliser, adj. counter, measurement period counter, day counter).
Each input can be separately sealed and secured under official calibration.
Pulse / signal outputs:
4 programmable transistor outputs, each freely programmable as alarm / warning output, pulse output, signal output for limit monitoring.
Each output can be separately sealed and secured under official calibration.
Elster GmbH 7
Data interface:
Optical interface according to IEC 62056-21 (replaces IEC 61107 and EN61107).
Permanently wired serial interface (RS232 or RS485).
MODBUS protocol via the permanently wired serial interface.
Automatic setting of the clock by remote data transmission with a modem connected.
Sending short messages by SMS.
Programmable standard output data records for process data ("three-minute values").
Pressure and temperature sensors:
Pressure sensor type CT30 integrated in device,or external mounted.
Pt500 temperature sensor, variable length.
Mechanical details / housing:
Suitable for wall mounting and meter installation (with mounting bracket).
Mounting + device installation without breaking the calibration seals.
Ambient temperature range: -25°C...+55°C
Extended temperature range with restricted functions possible.
Approvals:
Approval by calibration authorities
- as Volume Conversion Device acc. MID-Directive 2004/22/EG,
- as flow recording device acc. PTB-A50.7,
- as high flow display device acc. PTB-A50.7.
Ex approval for use in Ex Zone 1 according to EEx ib IIC T4.
Monitoring functions
Monitoring of signalling inputs.
Monitoring of any values with respect to programmable limits.
All monitoring can trigger appropriate reactions such as for example, entries in the status
register, logbook or archives or signalling via outputs.
Archives
Counter readings and maxima from the last 15 months for Vb and Vm.
Mean values, maxima and minima from the last 15 months for pressure and temperature
as well as partially for the inverted compressibility factor ratio and conversion factor.
Measurement period values (consumption profile) from the last 9 months for Vb, Vm, p, T, K and C for a measurement period of 60 minutes. The measurement period can be set in the range from 1 to 60 minutes.
Automatic changeover to daylight saving time can be set.
Event logbook with 250 entries for events such as for example status changes, signalling
inputs, limit violations.
Changes logbook ("audit trail") with entry of the last 200 changes to settings (parameterising steps).
8 Elster GmbH
2 Operation
2.1 Front panel
The following are positioned on the front panel for operation:
Two-line alphanumeric display with 16 characters per line.
Six keys for the display and entry of values.
EK260
Arc. Status MenuPrefix
Instromet
Typenschild
Identification label
ESC
ENTER
Submenu
y
n
a m
r
e G n
i e
d
a M
Elster GmbH 9
2.2 Display
Basic display structure (with an example):
Prefix Archive
m a x á A W B V b
V b A 1 2 3 4 5 6 7 . 8
Both lines in the display are subdivided into fields which are described below.
Device status M e n u
à
m 3
Submenu
2.2.1 Line 1 = Labels
The first line is subdivided into five fields of which four are labelled on the front panel:
1. Prefix (Type of computation)
The type of computation identifies so-called "initial values" (also termed "capture values"). These are values which have been formed over a time period (e.g. the adjustable measurement period or one month). Labels:
- max Maximum – highest value within the time range
- min Minimum – lowest value within the time range
- Change – volume within the time range
- Mean – mean within the time range
2. Archive
If an arrow points upwards to the label "Archive", then the displayed value is an archived value. This was frozen at a defined point in time and cannot be changed.
10 Elster GmbH
3. Device status
Here a maximum of three of the most important items of status information are continually shown.
A flashing character signifies that the relevant state is still present and the relevant message is present in the momentary status. A non-flashing character signifies that the relevant state is past, but the message in the status register has not yet been cleared.
Meaning of the letters:
- A "Alarm"
At least one status message has occurred which has resulted in disturbance volumes being counted. Basically, all messages with numbers in the range "1" and "2" represent alarms (e.g. "Alarm limits for pressure or temperature violated" 3.7). Alarm messages are copied into the status register and are retained here, even after rectification of the cause of the error, until they are manually cleared.
- W "Warning"
At least one status message has occurred which is valid as a warning. Basically, all messages with numbers in the range "3" to "8" represent warnings (e.g. "Warning limits for pressure or temperature violated" or "Error on output" 3.7). Warning messages are copied into the status register and are retained here, even after rectification of the cause of the error, until they are manually cleared.
- B "Batteries flat"
The remaining battery service life is less than 3 months. This display corresponds to the status message "Batt. low" ( Page 48 ).
- L "Calibration logbook full"
The calibration logbook is full; some parameters can now only be changed with the calibration lock open. (PLogB, page 42) This display corresponds to the status message "PLogB full", ( page 0).
If the calibration lock is opened with the calibration logbook full, it can only be
F
closed again after clearing the calibration logbook.
- P "Programming mode"
The programming lock (calibration lock) is open. This display corresponds to the status message "Cal.lock o." ( Page 48 ).
- M "Measurement error"
The connected gas meter encoder does not supply a fault-free counter reading. A flashing "M" corresponds to the status message "Coder fault" ( Page 48 ).
- o "online"
A data transfer via the optical or permanently wired interface is running. In each case the other interface cannot then be used. This display corresponds to the status message "online" ( Page 48 ).
4. Menu
Here is displayed to which list according to Chapter 3 the currently displayed value belongs. In submenus (indicated by an arrow to the left, see below) its name is displayed which is identical to the abbreviated designation of the entry point.
Elster GmbH 11
∗ 1 2 3 4 5 6 7 . 8
5. Submenu
- Arrow to the right
indicates that the displayed value is the entry point of a submenu. This can be called with the key [ENTER].
- Arrow to the left
indicates that you are located in a submenu which can be quit with the key [ESC]. On pressing [ESC] you are returned to the entry point of the submenu.
2.2.2 Line 2 = Value with name and unit
In the second line the name, value and (when available) the unit of the data are always shown.
Uncalibrated values are identified for the user with an asterisk ("") after the abbreviated designation.
For use outside of applications subject to calibration, the device can also be obtained without the identification of uncalibrated values.
Example of uncalibrated values:
V b A
Example of calibrated values:
V b
1 2 3 4 5 6 7 . 8
m 3
m 3
12 Elster GmbH
2.3 Keypad
Key(s) Designation Effect
Down cursor key
Up cursor key
Pfeil rechts
Left cursor key
Downwards movement within the list: From the first value in the list movement is in the direction of the last value or from the last value directly to the first one.
Upwards movement within the list:
From the last value in the list movement is in the direction of the first value or from the first value directly to the last one.
Movement to the right to a different list:
From the first list movement is in the direction of the last list or from the last list directly to the first one. With similar lists (e.g.: Vb and Vm) skipping occurs to the appropriate value, otherwise to the first value.
Switchover to the second part of the value for values
displayed on two lines:
- Counter readings divided into pre- and post-decimal places.
- Date and time (together one value) divided.
Movement to the left to a different list:
From the last list movement is in the direction of the first list or from the first list directly to the last one. With similar lists (e.g.: Vb and Vm) skipping occurs to the appropriate value, otherwise to the first value in the adjacent list.
Depends on the value displayed (Data class, 2.3.1)
Activate the entry mode.
ENTER
Enter
Open the submenu.
Update measurement (by pressing twice).
Return from a submenu to the entry point in the higher
ESC
Escape
level main menu.
Cancel entry (the value remains unchanged).
Skips to first value in the list
+
Home / Clear
Updates a value in the entry mode
+
Help
Displays the address (reference code) of the value
In the entry mode the keys change their functions, see Chapter 2.3.1.
Elster GmbH 13
2.3.1 Changing values
The methods of entering and changing values differ depending on the value. These are therefore subdivided into so-called "data classes" (abbreviation: "DC"). Values in the same data class are treated identically during entry. A prerequisite for an entry is that the lock assigned to the value is open.
The following data classes (DC) are present in the EK260:
DC
1 2 3 4 5 6
7
Type Entry, change using <ENTER>
Display test No change possible. Function Triggers the function by entering "1". Constant No change possible. Measurement The value is updated by pressing <
ENTER
>. twice. Status Abbreviated text for the status messages can be called with < Initialisable
value Discrete
value
After <
ENTER
> value initialisation (standard setting) by pressing the
key combination <CLR> = + . After <
ENTER
> value change by selection from a list of possible values
with the keys and . Value initialisation is possible with + .
8
Permanent value
After <
ENTER
>, setting to any value within the valid range is possible. Selection of each individual character to be changed with and and changing with und .
Value initialisation is possible with + .
9
Archive
Branching to the appropriate archive.
heading
11
Combination Similar to "Permanent value" (see above) but masked entry, i.e. only
the character currently being edited is visible, all others are masked out by a minus sign. With a closed lock it is opened on entering the correct combination. With an open lock, the combination is changed by the entry.
ENTER
>.
12
Counters As "Permanent value" (see above.).
15
Computation
No change possible.
counter
16
Initial value No change possible,
sometimes branching to a submenu.
17
Archive value No change possible.
19
Status register
20
Flag No change possible.
21
Constant value with 0
Abbreviated text for the status messages can be called with <
ENTER
Initialisation is possible with the menu command Clr.
As "Permanent value" (DC = 8, see above), but the entry of "0" is always possible independent of the specified limits.
>.
If a value is accommodated in a submenu, it cannot be changed independent of its data class by the keypad, since the key <
ENTER
> is then used for branching into the submenu.
14 Elster GmbH
2.3.2 Entering sources
At a number of points the entry of a "source" is required for parameterisation (e.g. SC.Qb in the volume at base conditions list, SC.O1 in the output list). The address of the desired value is entered as the source. It can be found in the tables at the beginning of each list (Chapter 3.1 ff.). In comparison to the addresses shown there however, the following supplements must be given:
- Completion of leading zeroes so that a total of four numbers exist in front of the colon.
- If the address includes no underscore "_", then "_0" should be appended.
Example 1: Source: 2:300 (Address of the volume at base conditions Vb, see table in
3.1) Enter: 0002:300_0 (Supplements printed in bold)
Example 2: Source: 6:310_1 (Address of the temperature T, see table in 3.4) Enter: 0006:310_1 (Supplements printed in bold)
2.3.3 Entry errors
Entry error messages are displayed after invalid entries via the keypad. Display: ----x--- where x = Error code according to the following table
Code
1 2
Description
The archive is empty, no values are available yet. The archive value cannot be read.
The archive has possibly just been opened by the interface for reading out.
4 5
Parameter cannot be changed (constant). No authorisation for changing the value.
To change the value the appropriate lock must be opened.
6 7
Invalid value: The entered value is outside the permissible limits. Incorrect combination: The entered combination (numerical code) is incorrect and
the lock is not opened.
11 *
Entry not possible due to special setting or configuration
- The entry of Vm and VmD in the encoder mode (Md.I1 = 5) is not possible.
- Md.I1 cannot be set to "5" with devices without encoder capability.
12
The entry of this source (address) is not permitted. For the output source SC.O2, e.g. for output mode "8", the addresses of Qb, Qm, p and T are permissible, but no counter readings, etc.
13
The function can only by executed after the time (à3.8, Time) has been set (initialised) to its start value with the key combinations + .
14
Gas analysis parameters for AGA-NX-19 do not match. Example: For "H gas" (calorific value Ho.b over 11.055) the density ratio dr must not exceed the maximum value of 0.691. (à 3.5)
20
Value for the application-specific display is not defined. The value to be displayed can be defined by the user by entering the address. No value is displayed because this has not yet occurred.
* With an EK260 with a software version below 2.10, this error is displayed with code "8". Elster GmbH 15
2.4 Access rights
The EK260 differentiates between four access parties and the calibration logbook. Each access party has a lock and a corresponding code. The locks have the order of priority
Calibration lock – Manufacturer's lock1 – Supplier's lock – Customer lock.
The access rights apply both for keypad inputs as well as for accesses via the optical or electrical (permanently wired) interface. If the lock is locked, all attempts to set values are answered with an appropriate error message (see Chap. 2.3.3).
Also the reading of values via the interfaces is only possible, for reasons of data protection, when at least one of the locks is open.
Normally, in addition to the access rights assigned to each individual value, values can also be changed by the access parties with higher priority. A value, which for example has "S" ("Supplier") as access rights, can also be changed by calibration officials and a value subject to the customer's lock can also be changed by suppliers.
Each party with write access for a value can also change the access rights (write and read access for each party) for this value via the interface. The rights of parties with higher priority can be changed.
2.4.1 Calibration lock
The calibration lock is used for securing parameters subject to calibration regulations. This includes all values which affect the volume counting.
The calibration lock is implemented as a pushbutton located within the EK260 housing below the circuit board cover panel. It can be secured with an adhesive seal ( 5.8.1)
The parameters protected under calibration regulations are each identified with "C" in the lists in the functional description.
Depending on the applications, values, which are not included as inputs subject to calibration regulations, can be placed under the user lock via the WinPADS parameterising software, for example to be able to use them as signalling inputs.
The calibration lock is opened by pressing the pushbutton (the symbol "P" flashes in the display) and is closed again when it is pressed again (symbol "P" goes out). Closure is also possible by deleting the value "St.PL" ( 3.9) via the keypad or interface. With the aid of the "WinPADS" parameterisation software a time in minutes can also set after which the calibration lock automatically closes.
In particular for applications not subject to the German calibration regulations, the level of protection for all parameters can be changed on request. For example, parameters which are as standard subject to the calibration lock can also be protected with the supplier's lock.
1
The manufacturer's lock is reserved for Elster and is not described here.
16 Elster GmbH
2.4.2 Calibration logbook
With the aid of the "Calibration logbook" according to PTB-A 50.7 (PLogB, page 42 ) some parameters relevant to calibration regulations can be changed also with the calibration lock closed. Prerequisites for this are:
- The calibration lock (see below) must be open.
- At least three entries must be available in the calibration logbook.
The affected parameters (e.g. cp value, meas urement period) are identified in the lists in Chapter 3 with the access right "PL".
A data row for the value is entered before and after the change for each change to such a parameter with the calibration lock closed.
If the calibration logbook has been written full, it can be cleared with the calibration lock open using the command ClrPL ( page 43).
F
If the calibration lock is opened with the calibration logbook full, it canonly be closed again after clearing the calibration logbook.
2.4.3 Supplier's lock and customer's lock
The supplier's and customer's locks are used for securing all data which is not subject to calibration regulations, but which should also not be changed without authorisation.
The parameters which are write-protected under the supplier or customer locks are each identified with "S" or "K" in the lists in the functional description ( 3). All values which are
shown with a minus symbol "-" cannot be changed, because they represent, for example, measurements or constants. The locks can be opened by entering a code (the "combination") under Cod.S or Cod.C and closed by entering "0" for St.SL or St.CL (→ page 54). With the aid of the parameterising software "WinPADS" a time in minutes can also be set for each lock under the addresses 1:174 ... 4:174 after which it automatically closes.
2.5 Formation of the list structure
The data display in the EK260 is structured in a tabular form. The individual columns in the table each contain associated values.
Values identified with U and Arc are submenus or archives which you can view by entering <ENTER> and leaving again with <ESC>. They each have, subordinate to the main menu,
a dedicated list structure, which is written in the corresponding list ( 3). The archives are subdivided into a number of data rows (also termed "data records" ). All
values in the same data row are saved ("archived") at the same point in time. The maximum number of data rows and the number of values in a data row depends on
the relevant archive. Within an archive the number of values and their meaning are the same for each data row. Switching to another archive data row occurs with the keys (for "younger" data rows) and (for "older" data rows). After the last data row, the first follows again and vice versa. Switching to another value within a data row occurs with the keys and . After the last value, the first follows again and vice versa.
A summary of the standard main menu (list structure) is shown on the following pages. By changing the value Menu (see page 87 ), a minimum main menu can be selected.
Elster GmbH 17
Volume at base
conditions
To
"User"
Flow at base cond. Actual flow Lower warn. limit Lower warn. limit
Disturbance quant. Disturbance quant. Upper warn. limit Upper warn. limit
Total quantity Total quantity Lower alarm limit Lower alarm limit
Adjustable counter Adjustable counter Upper alarm limit Upper alarm limit
Source monitoring Source monitoring Meas. range bottom Meas. range bottom
Upper warn. limit Upper warn. limit Meas. range top Meas. range top
Lower warn. limit Lower warn. limit Substitute value Substitute value
Meas. per. counter Meas. per. counter
Month's maximum Month's maximum Pressure mode Temperature mode
Daily counter Daily counter Press. sensor type Temp. sensor type
Month's maximum Month's maximum Sensor serial no. Sensor serial no.
Std.V.
Vb Vm p T
Vol. at base cond. Actual volume Pressure Temperature
Qb Qm VbD VmD p.UW VbT VmT pMin VbA VmA pMax SC.Qb SC.Q MRL.p QbUW QmUW MRU.p QbLW QmLW p.F VbMP
VbMP max S VbDy VbDy max S
Equ. coefficient 1 Equ. coefficient 1
Equ. coefficient 2 Equ. coefficient 2
Equ. coefficient 3 Equ. coefficient 3
Actual volume
Act.V.
VmMP
VmMP max S
VmDy VmDy max S
Eq1p Eq2p Eq3p Eq3T
p1Adj
Adjustment val. 1 Adjustment val. 1
p2Adj
Adjustment val. 2 Adjustment val. 2
Prog
Accept adjust. Accept adjust.
p.Amb
Air press. fixed val. Temp. meas.
p.Mes
Pressure meas. Meas. period mean
p.Abs
Abs.pressure meas. Month's maximum
Meas. period mean Month's minimum
Month's maximum
Month's minimum
Pressure
Press.
p.LW
Pb
Pressure at base conditions
Md.p Typ.p SNp
p.MP
p.Mon max S
p.Mon min S
Temperature
Temp.
T.LW T.UW TMin TMax MRL.T MRU.T T.F Tb
Temp. at base conditions
Md.T Typ.T SNT Eq1T
Eq2T
T1Adj T2Adj Prog T.Mes T.MP
T.Mon max S T.Mon min S
To "Conv."
18 Elster GmbH
To
"Temp."
Conversion
Conv.
C
Conversion factor Month archive 1 Status register Date and time
K
Gas law dev. factor Month archive 2 Momentary status Daylight saving: y / n
Ho.b
Calorific value Meas. per. archive Clear S. Reg Meas. cycle time
CO2 MPer
(depends on Md.K) Rem. time mea. p. Audit Trail Disp. switch-off time
(depends on Md.K) ArMP frozen
K substitute value
K mode Ambient temp.
Software version
Software checksum
Carbon dioxide content
H2 / N2 MP.Re
Rhob / dr FrMP
K.F
Md.K
Meas. period Logbook Operating cycle time
Archive Status + Logbook
Archiv
ArMo1 Arc ArMo2 Arc ArMP Arc
Clr MCyc
Status
SReg S Stat S
Logb. Arc
AudTr Arc
PlogB Arc
Calibration Logb.
ClrPL SNo
clear Cal. Logb.
Ta.Rg Vers Chk
System
System
Time
MdTim
OCyc
Disp
Aut.V
Disp. changeover
time
Serial number
To "Serv."
Elster GmbH 19
Service Inputs Outputs
To
"System"
Bat.R
Remaining bat. life Orig. counter Input 1 Mode for Output 1
Bat.C cp.I1 SC.O1
Battery capacity Cp value Input 1 Source Output 1
Serv.
Vo
Inputs
Md.O1
Outp.
"Ser.IO."
St.SL Md.I1 cp.O1
Supplier's lock Mode for Input 1 cp value Output 1
Cod.S V1 SpO1
Supplier's code Adj. counter Input 1 Status pointer O1
St.CL q.max Md.O2
Customer's lock
(only for encoder mode)
Mode for Output 2
Cod.C cp.I2 SC.O2
Customer's code Cp value Input 2 Source Output 2
St.PL Md.I2 cp.O2
Calibration lock Mode for Input 2 cp value Output 2
Contr V2 SpO2
Display contrast Adj. counter Input 2 Status pointer O2
Adj.T St.I2 Aj1O2
Clock adjust. factor Status on Input 2 HF adjust factor 1
Sel.p MdMI2 Aj2O2
Select. press. sensor Mode monitorino I2 HF adjust factor 2
Save SC.I2 f1.O2
Save all data Source monitoring I2 Freq. for Aj1.O2
Clr.A L1.I2 f2.O2
Clear archives
Limit 1 for I2 Freq. for Aj2.O2
Clr.V L2.I2 Md.O3
Clear counters Limit 2 for I2 Mode for Output 3
Clr.X SpI2 SC.O3
Initialise device Stat. pointer mon. I2 Source Output 3
Bin.T St.I3 cp.O3
Temp. raw value Status on Input 3 cp value Output 3
Bin.p MdMI3 SpO3
Pressure raw value Mode monitoring I3 Status pointer O3
Addr SC.I3 Md.O4
User display address Source monitoring I3 Mode for Output 4
... L1.I3 SC.O4
User display Limit 1 für I3 Source Output 4
WRp SpI3 cp.O4
Repair counter W Stat. pointer mon. I3 cp value Output 4
VbRp SNM SpO4
Repair counter Vb Serial no. gas meter Status pointer O4
VmRp
Repair counter Vm
Rep.
Revision mode
ArCal Arc
Frozen values
Frz
Freeze
-
Display test
To
20 Elster GmbH
Interfaces Energy User list
To
"Outp."
Md.S2
Mode, Interface 2 Energy Vb total
DF.S2 P VmT
Data format, Interface 2 Power Vm total
Ser.IO
W
Energ.
VbT
User
Bd.S2 WD p
Baud rate, Interface 2 Disturbance quantity Pressure
TypS2 WT T
RS-232 / RS-485 Total quantity Temperature
Num.T WA K
No. of rings bef. answrg. Adjustable counter Gas law dev. factor
M.INI Ho,b C
Initialise modem Calorific value Conversion factor
PrLog S
Printer log Source monitoring Status register
CSync S
Clock set by rem. trans. Upper warning limit Month's maximum Vb
GSM.N
GSM network Lower warning limit Of month's max. Vb
GSM.L
Reception field strength Meas. per. counter Of month's max. Vb
StM
Modem status (GSM) Month's maximum Flow at base cond.
P.Sta
Response to PIN code Daily counter Actual flow
PIN
PIN code Month's maximum Display menu
Resp1
Response to Message 1
Response to Message 2
Send message
Baud rate, Interface 1
Call window 1 start
Call window 1 end
(depends on setting)
(depends on setting)
ANT2 SEND Bd.S1 CW1.S CW1.E CW2.S / M.CW1 CW2.E / M.onl
SC.W P.UW P.LW
WMP WMP max S
WDy WDy max S
SReg VbMP max Date Time Qb
Qm Menu
To
"Std.V."
Elster GmbH 21
3 Functional description
The data display is structured in tabular form (list structure) (→ 2.5). The individual columns in the table each contain associated values. The following functional description is orientated to this list structure.
Here, the following abbreviations are used:
- AD Abbreviated designation
Designation of the value in the display
- Access Write access
Indicates which lock must be opened to change the value ( 2.4.1, 2.4.3):
- C = Calibration lock
- PL = Calibration logbook (PTB logbook, page 42)
- M = Manufacturer's lock
- S = Supplier's lock
- K = Customer's lock If the letter is located in brackets, the value can only be changed via the
interface and not via the keypad.
- Address Address of the value.
This is required especially for data transmission via the serial interface. The address can be displayed by pressing the keys + simultaneously.
- DC Data class
The data class shows, amongst other properties, whether and how the value can be changed. ( 2.3.1)
22 Elster GmbH
3.1 Standard Volume (Volume at base conditions) list
AD Designation / value Unit Access Address DC
Vb Volume at base conditions m3 PL 2:300 12 Qb Flow at base conditions m3/h - 2:310 4 VbD Vb disturbed m3 S 2:301 12 VbT Vb total m3 - 2:302 15 VbA Vb adjustable m3 S 2:303 12 SC.Qb Source for warning Qb - S 7:154 8 QbUW Upper warning limit Qb m3/h K 7:158 8 QbLW Lower warning limit Qb m3/h K 7:150 8 VbMP
Measurement period counter Vb m3 - 1:160 16 VbMP max Max. meas. per. count. Vb current month m3 - 3:160 16 VbDy
Daily counter Vb m3 - 2:160 16 VbDy max Max. daily count. Vb current month m3 - 4:160 16
Vb Volume at base conditions
The Volume at base conditions computed from the measured "actual volume" is summed here provided no alarm is present.
"Alarm" means "any message with the number "1" or "2" ( 3.7 ). Vb = Vm ž C where Vm = Actual volume ( 3.2) C = Conversion factor ( 3.5)
Qb Flow at base conditions
Momentary flow at base conditions (standard flow rate). Qb = Qm ž C with Qm = Actual flow ( page 25) C = Conversion factor ( page 35) The measurement inaccuracy of Qb is maximum 2% to 11% depending on the boundary conditions quoted for Qm ( page 25). In the alarm state Qb is computed with the substitute values of the disturbed
measurements.
VbD Vb disturbed
Here the volume at base conditions is summed while ever an alarm is present, i.e. a message with the number "1" or "2" is located in any momentary status ( 3.7). In the alarm state the volume at base conditions is computed with the substitute values of the disturbed quantities. ( 3.3: p.F, 3.4: T.F)
VbT Vb total
Here the sum of Vb + VbD is always displayed. Entries for Vb or VbD therefore also have an effect here. No entry for VbT itself can be carried out.
VbA Vb adjustable
Here, as with VbT, the total quantity, i.e. disturbed and undisturbed volumes, are counted. In contrast to VbT, VbA can however be changed manually. The counter is typically used for tests.
Elster GmbH 23
SC.Qb Source for Warning Qb QbUW Upper warning limit Qb QbLW Lower warning limit Qb
Using these three values, the flow at base conditions can be monitored in various ways. When the value to be monitored exceeds the upper limit QbUW or falls
below the lower limit QbLW, the message "Vb Warn Lim." is entered in St.2 ( page 47 ).
Various actions can in turn be programmed for this message, such as for example, entry of the change of status in the logbook (→ 3.7) or activation of a signalling output ( 3.11). With SC.Qb you can set which value is to be monitored.
SC.Qb Value to be monitored 0002:310_0 Qb Flow at base conditions 0001:160_0 0002:160_0
For further details on entering a source for SC.Qb: See Chap. 2.3.2.
VbMP Measurement period counter Vb
VbMP ∆ is restarted at "0" at the beginning of each measurement period (→ 3.6) and indicates the progress of VbT (see above). The measurement period MPer can be set in the archive list ( 3.6).
At the end of each measurement period VbMP is saved in the measurement period archive ( 3.6). VbMP ∆ can be monitored by appropriate programming of SC.Qb and QbUW (see above) in order, for example, to issue a warning signal to a special-contract customer when a limit is exceeded.
VbMP max Max. measurement period counter Vb in current month
By entering <ENTER> you can branch to the submenu where the time stamp of the maximum is displayed.
The maxima of the last 15 months can be interrogated in the month archive 1 ( 3.6).
VbDy Daily counter Vb
VbDy is restarted with "0" at the start of each day and indicates the progress of VbT (see above). As standard, the start of day is set to 06:00 hrs. and can be
changed with the calibration lock open via the serial interfaces under the address "2:141".
VbDy max Max. daily counter Vb in current month
By entering <ENTER> you can branch to the submenu where the time stamp of the maximum is displayed. The maxima of the last 15 months can be interrogated in the month archive 1 ( 3.6).
24 Elster GmbH
3.2 Actual volume (volume at measurement conditions) list
AD Designation / value Unit Access Address DC
Vm Actual volume m3 PL 4:300 12 Qm Actual flow rate m3/h - 4:310 4 VmD Vm disturbed m3 S 4:301 12 VmT Vm total m3 - 4:302 15 VmA Vm adjustable m3 S 4:303 12 SC.Q Source for Warning Q - S 8:154 8 QmUW Upper warning limit Qm m3/h K 8:158 8 QmLW Lower warning limit Qm m3/h K 8:150 8 VmMP
Measurement period counter Vm m3 - 8:160 16 VmMP max Max. meas. per. count. Vm current month m3 - 10:160 16 VmDy
Daily counter Vm m3 - 9:160 16 VmDy max Max. daily count. Vm current month m3 - 11:160 16
(Legends: see page 22)
Vm Actual volume
The volume V1 (→ 3.10) measured on the input is summed here provided no alarm is present.
"Alarm" means "any message with the number "1" or "2" ( 3.7 ). The entry of Vm is not possible in the encoder mode (Md.I1 = 5). An attempt to make an entry leads to entry error "11" ( 2.3.3 ) In the encoder mode, to bring Vm to the gas meter reading and to clear VmD, Md.I1 ( 3.10 ) can first be set to "0" and then set again to "5".
Qm Actual flow rate
Momentary actual flow (actual flow rate). The measurement inaccuracy of the displayed actual flow rate depends on
whether a pulse transmitter or an encoder is connected: With a connected pulse transmitter: With a pulse interval of a maximum of 15 minutes (at least four pulses per hour) and cp.I1 1 (→ page 58) the measurement uncertainty of Qm is maximum 1%.
With a pulse interval of more than 15 minutes Qm = "0" is displayed. After a change of the gas flow the accurate value can only be shown once the gas meter has sent at least two pulses. With a connected encoder: If the counter reading changes every 2 seconds or quicker, the measurement uncertainty of Qm is maximum 1%. With counter reading changes every 200 seconds or quicker the measurement uncertainty is maximum 10%. It can be reduced down to 2% at MCyc = 4 seconds by reducing the measuring cycle MCyc ( page 51). If the encoder counter reading does not change for longer than 200 seconds, Qm = "0" is displayed.
Elster GmbH 25
VmD Vm disturbed
Here the actual volume is summed while ever an alarm is present, i.e. a message with the number "1" or "2" is located in any momentary status ( 3.7). The entry of VmD is not possible in the encoder mode (Md.I1 = 5). An attempt to make an entry leads to entry error "11" ( 2.3.3 ) In the encoder mode, to clear VmD and to bring Vm to the gas meter reading, Md.I1 ( 3.10 ) can first be set to "0" and then set again to "5".
VmT Vm total
Here the sum of Vm + VmD is always displayed. Entries for Vm or VmD therefore also have an effect here. No entry for VmT itself can be carried out.
With the connection of an encoder (Md.I1 = 5, 3.10), VmT corresponds with two exceptions to the original counter Vo (→ 3.10 ) and so also to the gas meter
reading:
When the gas meter runs backwards, VmT stops and only continues synchronously with the gas meter again when the meter has a higher reading than before it started to run backwards.
VmT always has 9 pre-decimal and 4 post-decimal places, whereas the encoder always has 8 significant places and has between 6 and 9 pre-decimal places depending on the cp value of the gas meter. With a cp value of "1" VmT has one pre-decimal place more which is incremented with each overflow of the gas meter.
VmA Vm adjustable
Here, as with VmT, the total quantity, i.e. disturbed and undisturbed volumes are counted. In contrast to VmT, VmA can however be changed manually.
This counter is typically set to the same reading as the gas meter in order to be able to easily detect deviations by comparison of the two counter readings.
SC.Q Source for Warning Q QmUW Upper warning limit Qm QmLW Lower warning limit Qm
Using these three values, the actual flow can be monitored in various ways. When the value to be monitored exceeds the upper limit Qm.UW or falls below the lower limit Qm.LW, the message "Vm Warn Lim." is entered in St.4 ( page 47 ). Various actions can in turn be programmed for this message, such as for example, entry of the change of status in the logbook (→ 3.6) or activation of a signalling output ( 3.11). With SC.Q you can set which value is to be monitored.
SC.Q Value to be monitored
0004:310_0 Qm Actual flow rate 0008:160_0 0009:160_0
VmMP Measurement period counter Vm VmDy Daily counter Vm
For further details on entering a source for SC.Q: See Chap.2.3.2.
26 Elster GmbH
VmMP Measurement period counter Vm
VmMP ∆ is restarted with "0" at the beginning of each measurement period ( 3.6) and indicates the progress of VmT (see above). The measurement period MPer can be set in the archive list ( 3.6).
At the end of each measurement period VmMP is saved in the measurement period archive ( 3.6). VmMP ∆ can be monitored by appropriate programming of SC.Qb and QbUW (see above) in order, for example, to issue a warning signal to a special-contract customer when a limit is exceeded.
VmMP max Max. measurement period counter Vm in current month
By entering <ENTER> you can branch to the submenu where the time stamp of the maximum is displayed.
The maxima of the last 15 months can be interrogated in the monthly archive 1 ( 3.6).
VmDy Daily counter Vm
VmDy ∆ is restarted with "0" at the beginning of each day and indicates the progress of VmT (see above). As standard, the start of day is set to 06:00 hrs. and can be changed if required via the serial interfaces under the address "2:141".
VmDy max Max. daily counter Vm in current month
By entering <ENTER> you can branch to the submenu where the time stamp of the maximum is displayed. The maxima of the last 15 months can be interrogated in the month archive 1 ( 3.6).
Elster GmbH 27
3.3 Pressure list
AD Designation / value Unit Access Address DC
p Pressure bar - 7:310_1 4 p.LW Lower warning limit pressure bar S 10:150 8 p.UW Upper warning limit pressure bar S 10:158 8 pMin Lower alarm limit pressure bar C 7:3A8_1 8 pMax Upper alarm limit pressure bar C 7:3A0_1 8 MRL.p Pressure meas. range lower limit bar C 6:224_1 8 MRU.p Pressure meas. range upper limit bar C 6:225_1 8 p.F Pressure substitute value bar S 7:311_1 8 pb Pressure at base conditions bar C 7:312_1 8 Md.p Pressure mode - C 7:317 7 Typ.p Pressure sensor type - C 6:223 8 SNp Serial no. of pressure sensor - C 6:222 8 Eq1p Coefficient 1 of pressure equation - C 6:280 8 Eq2p Coefficient 2 of pressure equation - C 6:281 8 Eq3p Coefficient 3 of pressure equation - C 6:282 8 p1Adj Adjustment value 1 for pressure bar C 6:260_1 8 p2Adj Adjustment value 2 for pressure bar C 6:261_1 8 Prog Accept pressure adjustment - C 6:259 2 p.Amb Air pressure fixed value bar C 6:212_1 8 p.Mes Pressure meas. bar - 6:211_1 4 p.Abs Absolute pressure measurement bar - 6:210_1 4 p.MP Ø Measurement period mean pressure bar - 19:160 16 p.Mon max Month's maximum pressure bar - 21:160 16 p.Mon min Month's minimum pressure bar - 22:160 16
(Legends: see page 22)
p Press.
p is the pressure which is used for computing the conversion factor (→ 3.5) and hence the volume at base conditions ( 3.1). In disturbance-free operation the measurement p.Abs (see below) is used: p = p.Abs. With a relevant disturbance (alarm), the substitute value p.F (see below) is used: p = p.F. In addition, disturbance quantities are then counted ( → 3.1, 3.2) and the message "p Alarm Lim." displayed ( page 45 ). Relevant disturbances are:
p.Abs is located outside of the set limits pMin and pMax (see below).
With a connected encoder (Md.I1 = 5, 3.10) no actual volume can be
measured for more than 20 seconds. Since the gas flow variation is not known for the time period and therefore no gas quantities can be assigned to the measured values of pressure and temperature, the correction occurs as disturbance quantities with substitute values for pressure and temperature.
( 4.5)
28 Elster GmbH
p.LW Lower warning limit pressure p.UW Upper warning limit pressure
These values are used for monitoring the gas pressure p: When p exceeds the upper limit p.UW or falls below the lower limit p.LW, the message "p Warn Lim." is entered in St.7. (page 47) Various actions can in turn be programmed for this message, such as for example, entry of the change of status in the logbook (→ 3.6) or activation of a signalling output ( 3.11).
pMin Lower alarm limit pressure pMax Upper alarm limit pressure
The validity of the measured pressure p.Abs (see below) is checked based on these alarm limits. This monitoring does not occur when pMin = pMax. If p.Abs is located within the alarm limits, it is used as p (see below) for correction: p = p.Abs. If p.Abs is located outside the alarm limits, the substitute value p.F (see below) is used: p = p.F. In addition, disturbance quantities are counted in this case (→ 3.1,
3.2) and the message "p Alarm Lim." is displayed. ( page 45)
MRL.p Pressure meas. range lower limit MRU.p Pressure meas. range upper limit
These details of the measurement range are used to identify the pressure sensor. They have no effect on the measurements.
p.F Pressure substitute value
If the measured pressure p.Abs is outside the alarm limits pMin and pMax (see below), p.F is used as pressure p for correction. p = p.F.
pb Pressure at base conditions
The pressure at base conditions is used for computing the conversion factor ( 3.5) and hence the volume at base conditions.
Md.p Pressure mode
With Md.p = "1" the measured pressure p.Abs (see below) is used for correction, provided it does not violate the alarm limits.
With Md.p = "0" the fixed value (substitute value) p.F is always used for correction. No disturbance quantities are counted.
Typ.p Pressure sensor type
Here the designation of the pressure sensor is displayed which was selected in the service list with Sel.p. ( 3.9).
SNp Serial no. of pressure sensor
Identification of the pressure sensor associated with the EK260.
Eq1p Coefficient 1 of pressure equation Eq2p Coefficient 2 of pressure equation Eq3p Coefficient 3 of pressure equation
The coefficients of the quadratic equation for calculating the pressure p.Mes from the raw pressure value Bin.p ( 3.9):
p.Mes = Eq1p + Eq2p ž Bin.p + Eq3p ž Bin.p2 To adjust the pressure measurement circuit, the three coefficients of the quadratic
equation can either be found by the EK260 itself after entry of Prog (see below) or calculated and entered by the user.
The latter occurs based on three values for Bin.p and the corresp. set values.
Elster GmbH 29
The EK260 itself assumes a linear characteristic and after the entry of Prog (see below) only determines the coefficients Eq1p and Eq2p. The third coefficient Eq3p remains unchanged. The standard value for Eq3p is "0".
p1Adj Adjustment value 1 for pressure p2Adj Adjustment value 2 for pressure Prog Accept pressure adjustment
These values are used for the adjustment of the pressure measurement circuit, i.e. for the internal computation of the equation coefficients for the pressure (see above). The adjustment takes place in three steps:
1. Apply measurement pressure 1 (= reference value 1) to the pressure sensor and enter as p1Adj.
2. Apply measurement pressure 2 (= reference value 2) to the pressure sensor and enter as p2Adj.
3. Enter Prog = "1" so that the EK260 calculates the equation coefficients.
After applying the measurement pressure, you should either wait about one minute each time before entering the adjustment or press <ENTER> number of times during the display of the pressure measurement p.Mes (see below) until the displayed value is stable.
Approx. 0.4 ž pMax and approx. 0.9 ž pMax should be selected as adjustment values.
p.Amb Air pressure fixed value p.Mes Pressure measurement p.Abs Absolute pressure measurement
p.Abs is the sum of p.Amb and p.Mes: p.Abs = p.Amb + p.Mes. "0" should be entered when using an absolute pressure sentors and the ambient
pressure when using a gauge pressure sensor. p.Mes is the absolute or gauge pressure depending on the pressure sensor. If the measured pressure p.Abs is within the alarm limits pMin and pMax (see
above), then it is used as the pressure p for correction: p = p.Abs.
p.MP Ø Measurement period mean pressure
p.MP Ø is the averaged value of all pressure measurements within the current measurement period.
At the end of each measurement period p.MP Ø is saved in the measurement period archive ( 3.6).
p.Mon max Month's maximum pressure p.Mon min Month's minimum pressure
p.Mon max is the highest, p.Mon min the lowest of all pressure measurements within the current month.
By entering <ENTER> you can branch to the relevant submenu where the corresponding time stamp is displayed.
The maxima and minima of the last 15 months can be interrogated in the monthly archive 2 ( 3.6).
30 Elster GmbH
Loading...
+ 96 hidden pages