The present Operating Instructions are part of the delivered
equipment. They must be ready for use at any time and remain
with the unit in case of resale.
We reserve the right to carry out technical modifications on the
unit due to advanced development.
An operating manual cannot take account of every conceivable
use. An operating manual cannot take account of every
possible use. Contact your dealer or the manufacturer for
further information or in the event of problems which are not
covered or not sufficiently covered in this operating manual.
2 Important safety warnings
Please observe any additional national safety regulations that
may apply.
2.1 Instructions for the use of the present manual
Carefully read the Operating Instructions before you operate the
unit. Do not use the present electrical unit for any purpose other
than described in the Operating Instructions.
Warning symbols used in the present manual:
This symbol warns of the risk of injury caused by electricity.
This symbol warns of the risk of injury caused by explosion
and/or deflagration.
This symbol warns of the risk of injury caused by hot surfaces
and liquids.
This symbol warns of the risk of injury.
This symbol warns of the risk of damage to the equipment.
This symbol marks additional information.
Signal words used in the present manual
The signal word Danger warns of a potential risk of serious
injury and danger to life.
The signal word Warning warns of the risk of serious injury and
heavy damage to the equipment.
The signal word Caution warns of the risk of light injury or
damage to the equipment.
The signal word Attention warns of the risk of damage to the
equipment.
The present Elma ultrasonic cleaning unit has been designed
for the treatment of items and liquids only.
No cleaning of living beings or plants!
Operation of the unit by authorized and instructed staff only.
Observe the instructions given in the manual.
For safety reasons, the present unit must be connected to a
correctly grounded socket only. The technical details indicated
on the nameplate must correspond with the available mains
connection details, in particular those of the mains voltage and
current connected value.
For purposes of maintenance and care of the unit, in the case
of suspected humidity inside the unit or in the case of
malfunctions and after operation, unplug the mains plug.
The unit must only be opened by authorised specialised
personnel.
Fill the unit with a sufficient quantity of cleaning liquid before
switch-on. Flammable liquids must not be treated by ultrasound
directly in the cleaning tank: risk of fire and explosion!
Risk of burning and scalding! Depending on the operational
period of the unit, unit surfaces, cleaning liquid, basket and
cleaning items can heat up considerably.
Ultrasonic units can produce annoying sounds.
Wear personal ear protection devices when working close to an
ultrasonic unit which is operated without cover.
Do not reach inside the cleaning liquid or touch sound-carrying
parts (tank, basket, cleaning items, etc.) during operation.
The manufacturer cannot be held liable for damages on
persons, equipment or cleaning items caused by improper use.
The operator is responsible for the instruction of the operating
staff.
Temperature during storage: +5 C (+41 F) to +40 C (+104 F)
Temperature during transport: -15 C (+5 F) to +60 C (+140 C)
Humidity and air pressure during storage and transport:
10 % - 80 % relative humidity; non-condensing
Pressure range 500 hPa – 1060 hPa absolute
- rigid endoscopes and accessories
Only reusable medical devices that are approved for ultrasonic
cleaning and authorised for the reprocessing are permitted to
be cleaned (see information of the medical device manufacturer
according to, e.g. EN ISO 17664).
The pre-cleaning of medical devices with Elmasonic Easy-Line
products does not replace the subsequent cleaning, disinfection
or sterilisation in automated standard processes (e.g. cleaning
and disinfection unit (CDU) or autoclave).
The user is responsible for checking the cleaning result.
Ultrasonic cleaning machines are exclusively intended for
ultrasonic irradiation of objects and liquids. Thereby, no
flammable liquids are permitted to be used directly in the
cleaning tanks.
The machine is only permitted to be operated by trained
personnel and not by children. Operation and placement must
be performed in accordance with the conditions and media
defined in the operating manual.
The service intervals and regional regulations for checking the
equipment must be complied with.
2.4 Safety instructions on the machine
Observe operating instructions!
Observe warnings and safety instructions given in the operating
manual!
This symbol warns about the risk of injury from hot surfaces and
liquids.
The unit cannot be disposed with household waste! Observe
regional waste regulations!
Ultrasonic energy emitted through the air is not hazardous to
your health. However, high sonic emissions do arise during
ultrasound operation that may, under circumstances, cause
hearing damage to the foetus.
We recommend pregnant people not to spend long periods of
time near an ultrasonic cleaning device.
Elma Schmidbauer products with the CE mark comply with the
European EMC and Low Voltage Directive and adhere to the
prescribed EMC limit values so that the electromagnetic
radiation emitted by the devices is harmless to healthy people.
A binding statement for people with implants, such as those
with cardiac pacemakers or implanted defibrillators, can only be
made at the specific occupational site and upon consulting the
manufacturer of the implants.
3 Functioning
Today, cleaning by ultrasound is the most modern fine cleaning
method.
The electric high-frequency energy created by an ultrasonic
generator is transformed into mechanical energy by piezoelectrical transducer systems and is then transmitted into the
bath.
This process creates millions of tiny vacuum bubbles which
implode due to the variations of pressure caused by the
ultrasonic activity. Highly energetic liquid jets are created.
These jets remove dirt particles from surfaces and even from
the smallest grooves and bores.
3.1 Ultrasonic cleaning factors
Basically, the cleaning result depends on four factors:
Ultrasonic energy is probably the most important mechanical
factor in the cleaning process. This energy must be transmitted
through a liquid medium to the surfaces which are to be
cleaned. The present Elmasonic unit is fitted with the innovative
sweep function device: electronic oscillation of the sound field
(sweep function) prevents the formation of zones of low
performance in the ultrasonic bath.
For saponification and removal of the dirt particles a suitable
cleaning agent is required. Elma has a large range of cleaning
media on offer.
Cleaning chemicals are also necessary to reduce the surface
tension. This increases considerably the efficiency of the
ultrasonic activity.
The effect of the cleaning medium is improved by the optimised
temperature of the cleaning liquid.
For Elma cleaning products please observe the instructions
given on the label or the product information leaflets.
The cleaning period depends on the degree and the kind of
contamination and on the correct selection of ultrasonic energy,
cleaning agent and temperature.
Transducer tank made of cavitation-resistant stainless
steel
Case made of hygienically easy to clean stainless steel
Sandwich power transducer systems
Sweep function for optimal sound field distribution in the
cleaning liquid
Pulse Mode for intensification of the ultrasonic cleaning
effect.
Liquid fast drain on the machine rear side
(starting from Elmasonic EASY 60 H)
Heating with dry-running protection *
Machine can be plugged in to mains power socket
Electronic selector switch
Drip-proof control unit
Plastic carrying handles
(starting from Elmasonic EASY 60 H)
Automatic safety shutdown after 8 hours operation
to prevent accidental continuous operation
Automatic safety shutdown at 90 °C bath temperature
Protection against unintended restart, e.g. after
automatic safety shutdown or power failure
*on units with heating
4.2 CE conformity
This Elma ultrasonic cleaner meets the requirements for the CE
marking based on the EC/EU Low Voltage, Electromagnetic
Compatibility (EMC) and RoHS Directives. Some models are
also registered as medical devices.
Refer to the EC/EU Declaration of Conformity that can be
obtained from the manufacturer for details.
4.3 RFI Statement (European Union)
This is a Class A product.
Please note:
This equipment has been approved for business purposes with
regard to electromagnetic interference.
In a domestic environment this product may cause radio
interference in which case the user may be required to take
adequate measures.
For this please contact your supplier or the manufacturer of the
unit.
Ultrasonic cleaning unit
Mains cable
Tube socket with tube clamp
(starting from Elmasonic EASY 60 H)
Operating Instructions
4.5 Unit front view / side view
Illustration 4.5 Front view / side view Elmasonic EASY 30 H
Filling line (not available on Elmasonic EASY 10 / S 10 H)
indicates the recommended maximum filling level. This level
should not be exceeded even with cleaning items inside.
Plastic carrying handles (from Elmasonic EASY 30 H) for the
safe transportation of the unit even with hot casing.
Turning knob for the draining of the tank (from Elmasonic
EASY 60 H) functional description see section 4.7.
Operating panel for the control of the operating functions.
Description see section 4.8 and 4.9.
Illustration 4.8 View operating elements (unit with heating)
On/Off switch
Cleaning time preselection * Setting options Short time
operation: 1; 2; 3; 4; 5; 10; 15; 20; 25; 30 min (with automatic
switch-off).
Continuous setting ON for continuous operation. The switch-off
must be performed manually here.
However, the machine is switched off automatically after 8
hours continuous operation for safety reasons.
Start/Stop button for ultrasound operation
Activation of the Pulse mode by pressing and holding for more
than 3 seconds.
Ultrasound operation LED indicator
Fault indicator (flashes - see chapter 11)
Temperature preselection * for machines with heater.
Temperature selection range in 5 °C steps from 30°C – 80°C
Heater operation LED indicator for machines with heater,
lights when the heater is active.
Fault indicator (flashes - see chapter 11)
* setting the preselection: turn clockwise
reset of the preselection: turn anticlockwise
Risk of electrocution due to humidity inside the unit!
Protect the unit from entering humidity.
The unit inside is splash-water-proof. Keep workplace and
casing dry in order to prevent electrical accidents and damages
on the unit.
Packing
Check for transport
damages
Placement
DANGER
Ambient conditions
Proceed as follows
5 Initial operation
Please keep the original packing or dispose of it according to
the relevant waste disposal regulations. You can also return the
packing to the manufacturer destination (to your account).The
machine must only be sent in the original packaging for
transport (e.g. in the case of service).
Check the Elmasonic EASY for possible transport damages
before initial operation. In case of visible damage do not
connect the unit to the mains. Contact your supplier and
forwarding agent.
For operation, place the unit on a dry and solid surface. Ensure
that the workplace is sufficiently ventilated! Do not use a soft
surface (e.g. a carpet) as this may impede the ventilation of the
unit.
Allowed ambient temperature during operation:
+5 °C (+41 °F) to +40 °C (+104 °F)
Allowed relative humidity of air during operation: max. 80 %
In-door operation only
5.1 Set up of the liquid drain (EASY 60 H – 300 H)
On the delivered unit, the drain duct for the cleaning liquid is
closed off with a plastic screw cap. For setting up the liquid
drain fix the delivered tube socket to the drain duct.
1. Unscrew (anti-clockwise) the plastic screw cap
(see illustration 5.1.1)
2. Screw the tube socket (included in delivery) onto the inside
thread of the drain duct (clockwise).
3. Turn the tube socket into the required drain position
(see illustration 5.1.2).
The plastic thread is self-sealing when the socket has been
screwed in by hand as far as possible.
Note: Unscrewing the tube socket (anti-clockwise) can
cause a leak of the thread.
4. The drain duct is now ready for connection to a customerprovided discharge system. Use a standard tube (dia 1/2").
Push the tube onto the socket and fix it with the clamp
included in the delivery.
Illustration 5.1.1 Illustration 5.1.2
Drain with plastic screw capDrain fitted with standard tube
The power supply must be protected by an earth leakage circuit
breaker.
Use the plug-in mains cable delivered with the unit. Connect the
unit to a grounded shockproof socket only. Ensure that the
values indicated on the nameplate of the unit correspond with
the available connecting conditions.
The mains plug must be connected to an easily accessible
socket only, as it serves as interrupted device!
Risk of fire and explosion!
Never use flammable liquids or solvents directly in an ultrasonic
cleaning bath.
Use the cleaning chemicals listed in section 8.3.
Risk of damage to the transducer tank!
Do not use any acid cleaning agents (pH value < 7) directly in
the stainless steel tank if the cleaning items or the
contamination of the cleaning items contain halogenides
(fluorides, chlorides or bromides). The same applies to NaCl
solutions.
Use the cleaning chemicals listed in section 8.3.
Shut the drain
Observe filling level
Suitable cleaning
agents
Prohibited cleaning
agents
DANGER
CAUTION
CAUTION
6 Putting unit into operation
6.1 Filling of the unit
Shut the drain duct before filling the tank. (Turning knob for
draining of the tank into horizontal position (see section 4.7).
Fill the cleaning tank with a sufficient quantity of a suitable
cleaning liquid before switch-on.
The optimum filling level is approx. 2/3 of the tank volume.
The marked maximum filling level of the tank (not available on
EASY 10 / EASY 10 H) indicates the recommended filling level
with cleaning items in the bath (see also section 4 Illustration 4.5).
Ensure that the chosen cleaning agent is suitable for treatment
in an ultrasonic bath and observe the instructions on dosage
and the compatibility of the material.
We recommend the use of the cleaning agents listed in
section 8.3.
Flammable products are generally not allowed for use in an
ultrasonic bath. Observe the safety warnings given in
section 8.1.
Ultrasonic activity increases the vaporisation of liquids and
creates a very fine mist which can catch fire on any ignition
source.
Observe the instructions on limitations of use given in
section 8.1.
The stainless steel tank can be destroyed by crevice corrosion
in a very short time. Substances that cause crevice corrosion
can be contained in household cleaners.
Observe the instructions on limitations of use given in
section 8.2. For queries please contact the manufacturer or
your supplier.
Danger of damage to the transducer system!
Fill no liquid > 60 °C and < 10 °C in the ultrasonic tank.
Do not reach inside the tank during ultrasonic operation!
Cell walls can be damaged by prolonged exposure to ultrasonic
activity.
For placing and taking out the cleaning items always switch off
the unit.
CAUTION
No cleaning items on
the bottom of the tank
Use cleaning basket
Acid tank
Ultrasound button
6.2 Placement of cleaning items
Caution! The ultrasonic bath has been designed for the
ultrasonic treatment of items and liquids only.
Do not clean living beings or plants!
Do not place the cleaning items directly onto the bottom of the
cleaning tank, as this might lead to damages to the unit.
Place the cleaning items into the stainless steel cleaning basket
(accessory equipment).
For the use of cleaning chemicals which might destroy or
damage the stainless steel tank use a separate container. For
the special plastic cleaner tank for acid chemicals please
contact your supplier.
6.3 Degassing of liquid
Freshly mixed cleaning liquids are saturated with air which
lessens the cleaning effect of the ultrasonic activity. By
sonification of the liquid over a period of several minutes before
the cleaning process the tiny air bubbles in the liquid are
eliminated.
Degas the fesh cleaning liquid for approx. 5 - 10 minutes.
Risk of scalding by hot surfaces and cleaning liquid!
Ultrasonic energy is physically transformed into heat.
The unit and the cleaning liquid in the tank heat up during
ultrasonic operation even with the heating switched off.
During permanent operation with cover temperatures exceeding
60°C can be reached.
During permanent operation with cover and heating
temperatures exceeding 80°C can be reached.
Do not reach inside the bath.
If necessary touch unit and basket with protecting gloves!
Ultrasonic units can produce annoying sounds.
Wear personal ear protection devices when working close to an
ultrasonic unit which is operated without cover.
Sensitive surfaces can be damaged when exposed to
ultrasound over prolonged periods, particularly at low cleaning
frequencies.
Ensure that sensitive surfaces are exposed to ultrasonic
acitivity for a suitable period only.
If in doubt check the cleaning progress regularly and observe
the state of the surface material.
Ultrasonic energy is physically transformed into heat.
The unit and the cleaning liquid in the tank heat up during
ultrasonic operation even with the heating switched off.
During permanent operation with cover temperatures exceeding
60 °C can be reached.
For the cleaning of temperature-sensitive items please take into
consideration the heating-up of the cleaning liquid.
Please observe that the temperature of the cleaning media
remains below 42 °C when cleaning parts contaminated with
fresh protein or blood.
CAUTION
CAUTION
ATTENTION
ATTENTION
7 Ultrasonic cleaning process
Please observe the following instructions before starting the
ultrasonic cleaning process. It is the user’s responsibility to
High temperatures! Risk of burning and scalding!
Cleaning liquid, cleaning tank, casing, cover, basket and
cleaning items can heat up considerably.
Do not reach inside the bath.
If necessary wear protective gloves when touching unit and
basket!
Cleaning temperature recommendations in the medical sector:
Please observe that the temperature of the cleaning media
remains below 42 °C when cleaning parts contaminated with
fresh protein or blood.
Please observe the temperature even when using low or no
heating.
CAUTION
How to proceed
Heating control by
turning temperature
knob
7.1 Heating up of the cleaning liquid
(units with heating)
Depending of the degree and kind of contamination and on the
cleaning medium used it might be required to heat up the
cleaning liquid. For a quick heating-up process and in order to
prevent unnecessary energy losses we recommend to use a
cover (optional accessory equipment).
The ultrasonic energy is transformed physically into heat. Low
set temperatures can be exceeded during ultrasonic operation.
The cleaning effect through ultrasonic cavitation is reduced
when cleaning with high temperatures. We recommend not to
exceed a temperature of 80°C inside the tank.
For the recommended cleaning temperature please observe the
product information of the used elma clean cleaner.
Press the on/off key to start the unit.
Set the desired cleaning temperature at the temperature
selector switch for the temperature preselection.
The heater is active until the preselected temperature is
reached.
The LED indicator lights while the heater is active.
The LED extinguishes as soon as the preselected temperature
Switch on the machine using the on/off button.
Turn the selector switch clockwise to the desired cleaning time
for short time operation.
Press the ► button to start the ultrasound operation.
The ultrasound is automatically switched off after expiry of the
setpoint time.
Turn the selector switch clockwise to the ON position for
continuous operation. No automatic switch-off is performed in
the "Continuous operation" position. The ultrasound function
must be switched off by the user after the cleaning by pressing
the ► button. Or put the selector switch back into the OFF
position.
Attention: Only turn the selector switch anticlockwise back to
the 0 position!
Elmasonic EASY machines are equipped with safety shutdown
to prevent inadvertent continuous operation. The machine is
completely shut down automatically after 8 hours of continuous
operation. If you want to continue operating the machine
immediately, you only need to restart it.
The selector switch of the time setting must first be turned to
the "0" position to prevent any unintended operation after
external interruption of the mains power supply. For further
operation, select the desired cleaning time again to activate the
ultrasound and heating functions.
7.3 Sweep function
Elmasonic EASY machines are also equipped with an
automatic Sweep function.
Homogeneous irradiation in the cleaning tank is reached by
constant shifting of the maximum zones of the sound pressure.
7.4 Pulse Mode
Special function for intensification of the ultrasonic cleaning
effect. Particularly beneficial for stubborn soiling.
The ultrasound effect is increased by 20 % by increasing the
amplitude of the ultrasound signal.
Keep the Start/Stop button for the ultrasound operation ►
pressed for at least 3 seconds.
This function can also be activated at any time during operation.
This function activates an optical signal when a pre-set
temperature limit is reached.
This allows the user to take action in good time to prevent
temperature-sensitive items from becoming damaged or from
coagulating if blood or egg white is being cleaned.
The LED indicator for heating mode will flash rapidly when the
pre-set temperature limit is reached (4 sec.).
The unit emits an optical signal as a warning only.
The user is responsible for any action required, such as
switching off the unit or removing the cleaned item.
1. Turn the "Cleaning time rotary switch" to "OFF".
2. Turn the "Temperature rotary switch" to the required
temperature limit.
3. Press the "Start/Stop button" for at least 2 seconds.
The heater LED indicator will light up for around 3 seconds.
The pre-set temperature limit is now programmed and will also
be stored in the memory when the unit is switched off.
Proceed as follows to delete the programmed temperature limit:
1. Turn the "Cleaning time rotary switch" to "OFF".
2. Turn the "Temperature rotary switch" to "0".
3. Press the "Start/Stop button" for at least 2 seconds.
The heater LED indicator will light up for around 3 seconds.
The pre-set temperature limit is now deleted.
7.6 After the cleaning
When the cleaning process is finished rinse the cleaning items,
e.g. under the tap.
Drain the liquid as soon as it is dirty or when the unit is not
operated over a prolonged period of time. Certain residues and
types of contamination may destroy or damage the stainless
steel tank.
Use the quick-drain duct to drain the cleaning tank
(see section 4.7).
The cleaning chemical to be used must be suitable for the use
in an ultrasonic bath to prevent damage to the tank or injuries to
the user. Use the recommended cleaners mentioned in
section 8.3. Observe the restrictions to cleaners containing
solvents and aqueous cleaners mentioned in sections 8.1
and 8.2.
For queries please contact the manufacturer or your supplier.
Damages caused by non-compliance with the instructions given
in sections 8.1 and 8.2 will not be covered by the
manufacturer's warranty!
8.1 Limitations of use of cleaners containing solvents
Never use flammable liquids or solvents directly in an ultrasonic
cleaning tank. Risk of fire and explosion!
Ultrasound increases the volume of vaporisation of liquids and
creates a very fine mist that can catch fire on any ignition
source at any time.
Do not fill potentially explosive substances and flammable
solvents
marked in compliance with the EEC directives by symbols
and safety warnings R 1 to R 9
or E, F+, F, O or R 10, R 11 or R 12 for flammable
substances
into the stainless steel tank for ultrasonic treatment.
In compliance with the general regulations on the protection of
labour, certain limited volumes of flammable liquids (max.
1 litre) can be used in an ultrasonic cleaning unit under the
following conditions: these liquids must be filled into a suitable
separate vessel (e.g. beaker) with sufficient ventilation; this
vessel (beaker) can then be put into the stainless steel tank
which is filled with non-flammable liquid (water with a few drops
of surfactant).
Do not use aqueous cleaning media with pH values in the acid
range (pH < 7) directly in the ultrasonic tank if fluoride (Fˉ),
chloride (Clˉ) or bromide (Brˉ) ions can be taken in by the
removed dirt or through the cleaning chemical. These can
destroy the stainless-steel tank by crevice corrosion within a
very short period of ultrasonic operation.
Other media which can destroy the stainless-steel tanks when
used in high concentrations or with high temperatures during
ultrasonic operation are: hydrochloric acid, nitric acid, sulphuric
acid, formic acid, hydrofluoric acid (even diluted).
(Completeness of list not guaranteed.)
Risk of damage to the unit: do not use cleaning solutions
containing more than 0.5 mass % alkali (KOH and/or NaOH) in
an ultrasonic cleaning tank.
The above limitations for the use of chemicals in an ultrasonic
bath also apply for the aforementioned chemicals when these
are brought into an aqueous (particularly distilled water) bath
through entrainment or from the removed dirt.
For the ultrasonic treatment with the above mentioned media
use an acid-resistant tank (available as accessory equipment).
The limitations of use also apply to the standard cleaners and
disinfectants if these contain the above mentioned compounds.
Observe the safety warnings indicated by the manufacturer of
the chemicals (e.g. goggles, gloves, R and S phrases).
For queries please contact the manufacturer or your supplier.
8.3 List of recommended cleaning media
Elma has a large range of suitable cleaning products on offer
developed by chemical engineers in the Elma laboratory.
Please contact your supplier to find the most suitable cleaning
chemical for your application.
Product informations and safety data sheets are available from
the manufacturer (www.elmaultrasonic.com/produkte/reinigungsmittel).
Pull the mains plug before carrying out any maintenance
works!
The present Elmasonic EASY unit is maintenance-free.
Check the casing and the mains cable for damage regularly in
order to prevent electrical accidents.
Lime deposits on the stainless-steel tank can be cleaned gently
e.g. with elma clean 40 or elma clean 115C (operate the unit
with concentrate + water).
Check regularly the grid of the air fan at the bottom of the unit
(not existent in all units).
Remove dirt if necessary to allow sufficient ventilation inside the
unit.
Residues of cleaning media can be wiped away with a
household cleaner or decalcifier depending on the kind of
contamination. Do not put the unit in or under water!
If the unit is used for medical and sanitary purposes it is
necessary to disinfect the transducer tank and the surfaces
regularly (standard surface disinfectants).
9.2 Service life of the transducer tank
The transducer tank and particularly the ultrasound transmitting
surfaces are wear parts. The changes on the surfaces that
occur after a certain operating period are visible first as grey
areas and later on as material abrasions, the so-called
cavitation erosion.
To prolong the service life of your ultrasonic unit even more we
recommend to observe the following instructions:
Regularly remove any cleaning residues, in particular metal
particles and rust films.
Use suitable cleaning chemicals, with particular caution
concerning the kind of removed contamination (see
instructions section 8.2).
Abrasive particles from removed contaminations (e.g.
polishing pastes) must be drained and removed from the
cleaning tank as frequently as possible (exchange the
cleaning bath).
Exchange the cleaning medium before it is too heavily
contaminated.
Do not operate the ultrasound unnecessarily; switch off after
* EASY 10 – EASY 20 H: Pulsed ultrasound; EASY 30 H – EASY 300 H: Double half-wave sound.
The selection of the signal form has been adapted to the bath geometry. Due to the signal form, the
quadruple or octuple value is produced for the maximum peak value of the ultrasound power.
Subject to technical and visual modifications. Elma_ch_042016
Technical data
Technical data
Technische und optische Änderungen vorbehalten Elma_ch_06/02
12 Putting out of action and waste disposal
The machine components can be disposed of through
electronics and metal recycling facilities. The manufacturer also
accepts old components for disposal.
Do you have any queries or suggestions concerning the present
unit, its operation or the Operating Instructions?
Please contact us, we will be glad to assist: