ELLEX itrack In-service Manual

STEP 1
Remove the iTrack
Canaloplasty
Microcatheter pouch from the box.
STEP 2
Transfer the product to a sterile area.
MICROCATHETER IN-SERVICE GUIDE
PULL TAB
3.1
3.2
STEP 3
Remove the tray lid and the ViscoInjector™ dispenser cover.
STEP 4
Remove the ViscoInjector™ from the tray.
STEP 5
Prepare the ViscoInjector™:
a. Remove the red cap.
b. Unscrew the injector body.
c. Insert a viscoelastic cartridge.
d. Reassemble the injector body.
e. Turn the injector knob to express viscoelastic.
REMOVE COVER
5b
5d
5c
5e
STEP 6
Remove the iTrack
Canaloplasty Microcatheter.
STEP 7
Remove the cover tube.
6
7
STEP 8
Prepare the iTrack Canaloplasty Microcatheter:
a. Remove the infusion line from the tray and attach the
ViscoInjector
.
b. Remove the fiberoptic line from the tray and insert the fiberoptic connector in the iLumin
.
c. Wet the microcatheter tip.
STEP 9
Prime the iTrack
Canaloplasty Microcatheter:
a. Turn the knob five full turns
8a
8b
8b
and wait 5-10 seconds.
b. Make a half turn and wait another 5-10 seconds. Repeat until viscoelastic flows from the tip.
NOTE:
Each click is a 1/8 turn.
Half and full turns are indicated by black and
red lines on the ViscoInjector
, respectively.
BLACK = HALF TURN
9a
HALF TURN
RED = FULL TURN
Headquarters
82 Gilbert Street Adelaide, SA, 5000 AUSTRALIA
+
61 8 8104 5200
Japan
3F, 3-2-22 Harumi Chuo-ku Tokyo 104-0053 JAPAN +81 3 5859 0470
USA
7138 Shady Oak Road Minneapolis, MN, 55344 USA 800 824 7444
Germany
ZPO floor 1, Carl-Scheele-Str.16 12489 Berlin GERMANY +49 30 6392896 00
® 2016. Ellex Medical Pty. Ltd. ABiC is a trademark of Ellex Medical Pty Ltd. Ellex is a registered trademark of Ellex Medical Pty Ltd. E&OE.
Australia
82 Gilbert Street Adelaide, SA, 5000 AUSTRALIA
+
61 8 8104 5264
France
La Chaufferie - 555 chemin du bois 69140 Rillieux la Pape FRANCE
+
33 4 8291 0460
FULL TURN
Loading...