Eliwell ID 961 LX, ID 961 A User Manual

USER INTERFACE
The user has a display and four keys for controlling status and programming of the instrument.
KEYS AND MENUS
UP key Scrolls through the menu
items Increases the values Activates manual def. function
DOWN key Scrolls through the menu items
Decreases the values Programmable by parameter
fnc key ESC function (exit)
Programmable by parameter
set key Accesses the setpoint
Accesses the menus Confirms the commands Displaying the alarms (if present)
At start-up the instrument performs a Lamp Test; for few seconds the display and the leds blink, in order to verify their integrity and correct operation. The instru­ment has two main menus: the “Machine Status” and “Programming” menu.
ACCESSING AND USING MENUS
Resources are arranged in a menu, which can be accessed by pressing and quickly releasing the “set” key (“Machine Status” menu) or by holding down the “set” key for more than 5 seconds (“Programming” menu). To access the contents of each folder, indi­cated by the relevant label, just press the “set” key once. You can now scroll through the contents of each folder, modify it or use its func­tions. If you do not use the keyboard for over 15 seconds (time-out) or if you press the “fnc” key once, the last value shown on the display is confirmed and you return to the previous screen mask.
MACHINE STATUS MENU
To access the “Machine Status” menu Press and quickly release the “set” key. If alarms are not present, the label “SEt” appears. By using the “UP” and “DOWN” keys you can scroll through the other fold­ers in the menu:
-AL: alarm folder (if present);
-Pb1: probe 1 value folder;
-SEt: Setpoint setting folder.
Set Setting
Access the “Machine Status” menu by pressing and quickly releasing the “set” key. The label of the “SEt” folder appears. To display the Setpoint value press the “set” key again. The value appears on the display. To change the Setpoint value, use the “UP” and “DOWN” keys within 15 seconds. If the parameter is LOC = y the Setpoint cannot be changed.
Alarm on
If alarm condition exists, when accessing The “Machine Status” menu the “AL” fold­er label appears (see the “Diagnostics” section).
Displaying Probes
By pressing the “set” key when the appro­priate label appears, the value of the probe associated to it is displayed.
PROGRAMMING MENU
1) Level 1 P
arameters
To access the “Programming” menu, press the “set” key for more than 5 seconds. If specified, the level 1 access PASSWORD will be requested (see parameter “PA1”) and (if the password is correct) the label of the first folder will follow. If the pass­word is wrong, the display will show the PA1 label again.
To scroll other folders, use the “UP” and “DOWN” keys; the folders contain only
the level 1 parameters. NOTE: At this point level 2 parameters are NOT visible, even if they aren’t protected by password.
2) Level 2 P
arameters
In the Programming Menu go into the “CnF” folder, scroll all the parameter until you reach the PA2 label. By pressing and releasing the “set” button you will enter to level 2 parameters and the label of the first folder in the programming menu will follow. The level 2 parameters may be protected by a second password (see “PA2” parame­ter inside “diS” folder, not to be confused with PA2 label inside “CnF” folder. If specified, level 2 parameters are hidden to user; accessing the “CnF” folder the level 2 access PASSWORD will be request­ed and (if the correct password is entered) the label of the first folder in the pro­gramming menu will follow.
NOTE: At this point you will see only level 2 parameters.
Level 1 parameters will NOT be visible; to reach them you shall exit the Programming Menu and re-entry the Programming Menu section (see step 1). To enter the folder, press “set”. The label of the first visible parameter appears. To scroll through the other parameters, use the “UP” and “DOWN” keys; to change the parameter, press and release “set”, then set the desired value using the “UP” and “DOWN” keys, and confirm with the “set” key. Move to the next parameter. PLEASE NOTE: It is suggested to switch­off and switch-on again the instrument everytime it is changed the configuration of the parameters: this prevents malfunc­tioning on regulation and delay time occuring.
PASSWORD
The passwords “PA1” and “PA2” allow access respectively to level 1 and level 2 parameters. In the standard configuration passwords are not present. To enable them (value 0) and assign them the desired value, access the “Programming” menu, within the folder with the “diS” label. If passwords are enabled, they will be requested:
- PA1 at the entrance of the “Programming” menu (see the “Programming Menu“ section);
- PA2 within the folder with the “Cnf” label containing level 1 parameters.
ID 961 LX/A
electronic controllers for refrigerating units
fnc
set
LED
Position Related Function Status
Compressor or relay 1 ON when the compressor is started up; blinking in case of delay,
protection or blocked enabling
Defrost ON when defrosting; blinking in case of manual
enabling
Alarm ON when the alarm is enabled; blinking when the alarm is silenced
MANUAL ACTIVATION OF THE DEFROSTING CYCLE
To manually activate the defrosting cycle, press the “UP” key for 5 seconds. If defrosting conditions are not present, (for example the evaporator probe temperature is higher than defrost stop temperature), the display will blink three (3) times, in order to indicate that the operation will not be performed.
COPY CARD
The Copy Card is an accessory connected to the TTL serial port which allows pro­gramming quickly the instrument parame­ters. The operation is performed as fol­lows:
FrFormat
This command allows copy card formatting, an operation recommended in case of first use. Warning: if the copy card has been pro­grammed, using the “Fr” the data entered are erased. This operation cannot be can­celled.
UL-Upload
This operation loads the programming parameters from the instrument.
dL-Download
This operation downloads to the instru­ment the programming parameters.
NOTE:
• UPLOAD: instrument --> Copy Card
• DOWNLOAD: Copy Card --> instru­ment.
The operations are performed accessing the folder identified by the “FPr” label and selecting, according to the case, “UL”, “dL” or “Fr” commands; the operation is con­firmed by pressing the “set” key. If the operation is successful an “y” is displayed, on the contrary, if it fails a “n” will be dis­played.
TELEVIS SYSTEM
The link with the Televis system can go through the serial port TTL (the BUS ADAPTER 100 TTL RS-485 interface module must be used). In order to configure the instrument for this purpose you must access the folder identified by the “Add” label and use the parameters “dEA” and “FAA”.
KEYBOARD LOCKING
The instrument includes a facility for dis­abling the keyboard, by programming the “Loc” parameter (see folder with “diS” label). If the keyboard is locked, you can still access the programming menu by pressing the “set” key. The Setpoint can also be viewed.
DIAGNOSTICS
The alarm condition is always signalled by the buzzer (if present) and by the led of
the alarm icon The alarm signal produced by a faulty
thermostat probe (referred to probe 1) is shown as E1 on the instrument display
When the sensor detects an error condi­tion:
• the code E1 is displayed
• the compressor is activated as indicated by the "On" and "Off" parameters if pro­grammed for the duty cycle or:
Other signalling alarm are not shown on the instrument display, but from the “Machine Status” menu within the “AL” folder.
MINIMUM AND MAXIMUM TEMPERA­TURE ALARMS
Regulation of the minimum and maximum temperature alarms refers to the thermo­stat probe. The temperature limits defined by the “HAL” (maximum temperature alarm) and “LAL” parameters (minimum temperature alarm) are in absolute temperature value. When an alarm status occurs, if no alarm exclusion phases are underway (see alarm exclusion parameters), the alarm set icon is lit up and the buzzer, and/or the relay configured as an alarm, is activated. The occurrence of this alarm does not in any­way effect the control activities in progress. This alarm status can be viewed in the “AL” folder with AH1-AL1 labels.
INSTALLATION
The instrument is designed for panel mounting. Make a hole of 29x71 mm, insert the instrument and fix it using the brackets provided. Do not mount the instrument in humid and/or dirty places; it is suitable for use in ordinary polluted places. Ventilate the place in proximity to the instrument colling slits.
ELECTRICAL WIRING
Attention! Never work on electrical connections when the machine is switched on.
The instrument is equipped with screw ter­minal boards for connection of electrical cables with a diameter of 2.5 mm
2
(one conductor only per terminal for power connections). For the capacity of the terminals, see the label on the instrument. The relay contacts are voltage free. Do not exceed the maximum current allowed; in case of higher loads, use an appropriate contactor. Make sure the power supply voltage complies with the one required by the instrument. In 12V versions the power supply must be provided by a security transformer with the protection of a delayed 250 mA fuse. Probes have no connection polarity and can be extended using a regular bipolar cable (note that the extension of the probes affects the EMC electromagnetic compatibility of the instrument: pay extreme attention to wiring). Probe cables, power supply cables and the TTL serial cables should be distant from power cables.
ID 961 LX/A 2/6
DISPLAY
E1
If simultaneous they will appear on display with 2 seconds alternation
ERROR
Thermostat probe fault
Probe Error table
Ont
0 0 >0 >0
Oft
0 >0 0 >0
Compressor output
OFF OFF ON dc
DISPLAY
AH1
AL1
EA oPd To silence alarms press any key. If silenced the LEDs will blink Alarms are in absolute value or related to Setpoint and (considered as the distance from the Setpoint itself) depending on Att parameter.
ALARM
High temperature alarm (referred to the thermostat probe or probe 1) Low temperature alarm (referred to the thermostat probe or probe 1) Esternal alarm Open door alarm
Loading...
+ 4 hidden pages