Eliwell EWDR 981 User Manual

USER INTERFACE
The user interface has a 4 LED display to indicate status and alarms and four buttons for controlling instrument status and pro­gramming.
KEYS
UP Increase in value of
DOWN Decrease in value of
set Access to different levels of
menu Alarm display, set point and probes Access to programming of parameters
fnc Exit current level of
menu Confirmation of value of para­meter (Parameter programmable H33)
SIGNAL LED
The status of the external devices, functions and controllers is described by the device LEDs.
DISPLAY
This is used to display the inputs, the set point, the parameters and related values, alarms, functions and the status of the device.
DESCRIPTION OF MENU
Access to both menus is controlled by the ‘set’ button. If it is pressed and immediately released, the ‘machine status menu‘ is displayed. Hold the same button down for 5 seconds to access the ‘parameter programming menu‘. When one of the two menus has been accessed, you can navigate between the level 1 folders using the ‘UP’ and ‘DOWN’ buttons. The folders are opened by pressing the ‘set’ button once. You can now scroll through the contents of each folder, modify it or use its functions. You can exit each level of both menus in three ways: using the ‘fnc’ button, if a new value is confirmed by pressing the ‘set’ button or, when the time-out has elapsed (15 seconds inactivity on the device).
MACHINE STATUS MENU
The ‘machine status menu’ contains the folders and basic information on the device:
-AL: alarm folder
-SEt: Set point setting folder
-Pb1: ‘probe 1 value’ folder
-Pb3: ‘probe 3 value’ folder
If no alarms are present, the “SEt” label is displayed. From here, you can scroll down the other menu items using the UP’ and ‘DOWN’ buttons. Each folder can be accessed by pressing the ‘set’ button once. Values are modified using the ‘UP’ and ‘DOWN’ buttons and the ‘set’ button that confirms the selected value and takes you back to the higher level.
Setting the set point
Access the ‘machine status menu’. If no alarms are present, the “SEt” label is displayed. By pressing and immediately releasing the ‘set’ button, the set point value can be set using the UP’ and ‘DOWN’ buttons. Press and release the ‘set’ button again or press the ‘fnc’ button to go back to the main menu level. The set point setting folder is also closed when the time-out elapses.
Alarm on
If an alarm condition exists when the Machine Status menu is accessed the “ALfolder label appears.
PARAMETER PROGRAMMING MENU
Access the menu by pressing the ‘set’ button for at least 5 seconds. The menu structure enables all parameter folders to be divided into two levels. All the level 1 folders can be accessed by entering the password ‘PA1’. Scroll down the level 1 folders using the ‘UP’ and ‘DOWN’ buttons. Press and release the ‘set’ button next to the selected label to access the parameters. Scroll through the labels in the folder using the ‘UP’ and ‘DOWN’ buttons, press ‘set’ to display the current value of the selected parameter, use the ‘UP’ and ‘DOWN’ buttons and set the required value by pressing ‘set’. To access the level 2 folders in the ‘Cnf’ folder, select the ‘PA2’ label, enter the password ‘PA2’ and confirm with the ‘set’ button. All the parameters that cannot be changed at level 1 are in this level.
NOTE: Level 1 parameters will only be displayed if you quit the ‘parameter programming menu’ and repeat the steps for manipulation of level 1 folders.
The steps to follow for manipulation of level 2 parameters are the same as those
NOTE: It is strongly recommended that the instrument is switched off and on again each time parameter configuration is changed in order to prevent malfunctioning of the configuration and/or ongoing timings.
PASSWORD
The passwords “PA1and “PA2” are used to access level 1 and level 2 parameters. To change them and assign them the desired value, access the ‘parameter programming menu’ in the “diS” label folder.
The password is requested:
- PA1 when entering the ‘parameter programming menu’;
- PA2 in the “Cnf” folder containing the level 1 parameters.
USING COPY CARD
The Copy card function can be used to upload and download parameter maps of one or more of the same type of instrument. The accessory is connected to the device using the special TTL serial port on the instrument. The Copy card can be used for the following functions:
EWDR 981
electronic controllers for refrigeration units
cod. 9IS43091
rel. 3/05
GB
LED Description Indication
reduced set point Lights up when inserted LED set point is dis-
played, blinking when reduced set point is inserted
compressor LED ‘on’ with LED on.
Blinking for delay, protection or acti­vation blocked
defrost LED ‘on’ during defrost-
ing. Blinking when activated manually or by digital input
alarm LED ‘on’ if alarm is pre-
sent. Blinking for silenced
alarm decimal point ‘on’ to indicate of
voltage when on stand-by and dis­play ‘off’
Upload (UL): The upload function reads the instrument parameters and writes them onto the Copy card Download(dL): The download function writes the parameter map that is on the copy card at that time onto the instrument memory.
NOTE: Upload = instrument—-> Copy card Download = Copy card —> instrument
The operations are performed by accessing the folder with the ‘FPr’ label and selecting the ‘UL’, ’dL’ or ‘Fr’ commands. The operation is confirmed by pressing the ‘set’ button. If the operation is successful, a “y” is displayed whereas if it is unsuccessful an “n” will be displayed.
Formatting the copy card
This function is necessary when using the copy card for uploading for the first time or when using the copy card on device models that are not compatible.
NOTE: formatting deletes all the data on the Copy card and cannot be undone. Download from reset
Once the copy card has been connected the instrument is switched on. When the lamp test is over, one of the following two labels will be displayed for 5 seconds:
- dLY if the operation is successful
- label DLn if operation fails The display will then go into default position (probe or set point).
NOTE: after downloading, the instrument will begin to work with the new parameter map that has just been downloaded.
ADVANCED FUNCTIONS
DOOR SWITCH INPUT
This is a clean contact digital input with programmable polarity. The door switch input functions are controlled by the values of the following parameters:
Parameter H11 is used to configure the door switch input with values between
-8 and +8. Positive and negative values are present in order to select the polarity assigned to the input and:
NOTE: the sign “-” indicates that the input is activated when the contact is
closed. The ‘+’ sign indicates that the input is activated when the contact is open
LIGHT CONTROLLER
Controls the light relay. The function can be started in two different ways. If you press the light button the light relay is switched on if it was off and is switched off if it was on. The light status is recorded by the device as soon as the button is pressed so that when power is restored after a black-out the device can continue to operate in the same way as before the power failure. The following parameters control how the light works:
The configuration of these parameters controls the status of the light relay by using the digital input as well as the normal button. The parameter dSd automatically switches on the light relay when the digital input is enabled and switches it off when the digital input is disabled following the delay set by parameter dLt. Parameter H06 activates the light button and enables the relay when the door is open even when the instrument is powered but is ‘off’. The dedicated button always deactivates the light relay even if the digital input is on or during the dLt delay only if the parameter OFL=Y.
DIAGNOSTICS
PROBE ALARMS
When one of the probes is outside the nominal operating range or the probe is open or has shorted, an alarm is generated if at least one of these conditions persists for at least 10 seconds. If at least one of these 3 alarms is signalled the alarm LED and relay are enabled. When activated code E1 appears on the display.
If several alarms are activated at the same time, they are displayed alternately for 2 seconds each. An error condition in the room probe leads to:
- E1 code appears on display
- activation of compressor as indicated by Ont and OFt
- deactivation of maximum and minimum alarm controller.
MINIMUM AND MAXIMUM TEMPERATURE ALARM
The alarm is regulated on the room probe. The temperature limits are defined by parameters HAL and LAL. The limits refer to the set point if parameter Att=1 and are absolute if Att=0.
NOTE: If the alarms are relative, the parameter HA1 is set to positive values and LA1 to negative values.
ALARM WITH THRESHOLD REFERRING TO PROBE 3
An alarm is associated with probe 3 that refers to a threshold that is reset at a specified differential. By setting parameter PbA=3 probe 3 will signal a high or low temperature alarm for exceeding the set value.
The alarm is handled in the same way as the other temperature alarms and standard signal delays are used.
EXTERNAL ALARM
It is set if the digital input is enabled with the delay specified by parameter dAd and remains enabled until the next time the digital input is deactivated. The alarm consists of an alarm LED that remains on, activation of a buzzer and deactivation of all the device loads (if specified by the EAL parameter). If an external alarm is present the EA label is displayed in the AL folder.
It is reset the next time the digital input is deactivated and the buzzer can be manually silenced.
Par Description
dOd Digital input switches off loads dAd D.I. activation delay OAO Alarm signal delay after dis-
disabling the digital input (door closed)
tdO Time out door open. Time out
signal after activation of D.I.
(door open) H11 Digital output configurability/polarity 1 H21...H25 Digital output configurability 1...5
Par Description
H06 Button/aux input/door switch light active
when instrument off dSd Enabling light relay by door switch dLt Delay Light Relay
deactivation delay OFL Light switch always disables light relay OAO Alarm signal delay after
disabling the digital input
(door closed) tdO Time out door open.
Time out signal after activation of D.I.
(door open) H11 Digital output configurability/polarity 1 H21...H25 Digital output configurability 1...5
Signal Description
E1 Faulty room probe E3 Faulty display probe
EWDR 981 2/8
Att=0 absolute
Maximum temperature alarm room probe temperature ≥ HAL
Minimum temperature alarm room probe temperature ≤ LAL
Att=1 relative
Maximum temperature alarm room probe temperature Set + HAL
Minimum temperature alarm room probe temperature Set + LAL
Signal Description AH3 High temperature alarm AL3 Low temperature alarm
Signal Description EA External alarm
OPEN DOOR ALARM
This alarm is generated when the door is open for longer than the value specified in parameter tdO. The door open alarm is signalled in the alarm folder (AL) and the LED and alarm relay are switched on. The OPd label is also displayed.
The relay can be deactivated by pressing the silencing button. The LED continues to flash intermittently and the OPd label remains in the alarm menu until the door is closed.
MECHANICAL ASSEMBLY
The EWDR 981 is designed to be wall­mounted (removable brackets) or mounted on DIN rail (Omega 3). The unit operates correctly with an ambient temperature range of between -5 and 55 °C. Units must not be installed in excessively humid and/or dirty locations.
TECHNICAL DATA
Casing: plastic 4 Din modules Dimensions: front 70x85 mm, depth 61mm.
Mounting: on DIN rail (Omega 3) or wall-mounted. Connections: on screw terminal block for 2.5mm
2
conductors (only one conductor per terminal block for power connections) Operating temperature: -5...55°C.
Storage temperature: -30...85°C. Operating and storage ambient humidity:
10...90 % RH (non-condensing). Display: 3 digits + sign
Analogue inputs: 2 NTC/PTC inputs (parameter configurable)
Digital inputs: 2 voltage-free parameter­configurable digital inputs.
Serial: TTL for Copy Card
Digital outputs: 1 on relay: 16A 1hp 250V SPDT;
Configurable 12Vc/24mA output Resolution: 1 or 0.1°C.
Accuracy: better than 0.5% of bottom scale +1 digit
Consumption: 5VA
Power supply: 230Va
ELECTRICAL CONNECTIONS
Warning! Always switch off machine before working on electrical connections.
The instrument has screw terminal blocks for connecting cables with a maximum diameter of 2.5 mm
2
(only one conductor per terminal block for power connections): for terminal capacity, see the label on the instrument. The relay outputs are voltage free. Do not exceed the maximum current allowed. For higher loads, use a suitable contactor. Make sure that the power voltage complies with the device voltage. Probes have no connection polarity and can be extended using an ordinary bipolar cable (note that if probes are extended this affects the electromagnetic compatibility (EMC) of the instrument: special care must be used when wiring).
The TTL serial cable must be kept separate from the power cables.
EWDR 981 3/8
CONDITIONS OF USE
PERMITTED USE
For safety reasons the instrument must be installed and used in accordance with the instructions supplied. Users must not be able to access parts with dangerous voltage levels under nor­mal operating conditions. The device must be suitably protected from water and dust depending on the specific application and only be accessible using special tools (except for the front key­pad). The device is ideally suited for house­hold use and/or similar use in the refrig­eration sector and has been tested with regard to safety in accordance with the European harmonized reference stan­dards. It is classified as follows:
• as an automatic electronic control device to be independently mounted as regards its construction;
• as a 1 B type operated control device as regards its automatic operating fea­tures;
• as a Class A device as regards the cate­gory and structure of the software.
UNPERMITTED USE
The use of the unit for applications other than those described above is for­bidden. It should be noted that the relay contacts supplied with the device are functional and therefore exposed to potential faults. Any protection devices required to comply with product requirements or dictated by common sense due to obvious safety reasons should be installed externally.
RESPONSIBILITY AND RESIDUAL RISKS
Eliwell shall not be liable for any damages deriving from:
- installation/use other than that prescribed and, in particular, which does not comply with the safety standards specified in the regulations and/or those given herein;
- use on boards which do not guarantee proper protection against electric shock, water or dust when assembled;
- use on boards which allow dangerous parts to be accessed without the use of tools;
- tampering with and/or alteration of the product;
- installation/use on boards that do not comply with the standards and regulations in force.
Signal Description OPd Open door alarm
PAR. DESCRIPTION
RANGE DEFAULT VALUE LEVEL U.M.
SEt Set point with range falling between the minimum
LSE set point and the maximum HSE set point. The set point value is in the machine status menu
menu
LSE...HSE 0.0 °C/°F
diF When the compressor stops when it reaches the
set point value, it restarts at a value corresponding to the set point plus the value of the differential It
must not be 0
0.1...30.0 2.0 1-2 °C/°F
HSE Maximum set point value
LSE...302 50.0
1-2 °C/°F
LSE Minimum set point value
-58.0...HSE -50.0 1-2 °C/°F
OSP Offset point. Value to be added to set point if
reduced set point is activated Economy function).
-30.0...30.0 0 2 °C/°F
Cit Minimum compressor ON time. Minimum time for
activating a compressor before deacti vation Not active if=0
0...250 0 2 min
CAt Maximum compressor ON time. Maximum time for
activating a compressor before deacti vation Not active if=0
0...250 0 2 min
Ont (1) Compressor activation time if probe is faulty. If set
to 1 with OFt=0 the compressor always remains on whereas if OFt>0 it operates in duty cycle mode (see Duty Cyclediagram)
0...250 0 1-2 min
OFt (1) Compressor shut-down time if probe is faulty. If set
to 1 with Ont=0 the compressor always remains off whereas if OFt>0 it operates in duty cycle mode (see Duty Cyclediagram)
0...250 1 1-2 min
dOn Delay in activating compressor relay after start-up
0...250 0 1-2 sec
dOF Delay after shut-down; between compressor relay
shut-down and subsequentstart-up the specified time must elapse.
0...250 0 1-2 min
dbi Delay between switch-ons; the specified time must
elapse between two subsequent switch-ons
0...250 0 1-2 min
OdO Delay in activating outputs after start-up of instru-
ment or after a power failure. Not active if=0
0...250 0 1-2 min
dty Type of defrost.
0 = electrical defrosting; 1 = cycle reversing defrosting (hot gas); 2 = Free mode defrosting (compressor disabled).
0/1/2 0 1-2 num
dit Period of time elapsing between the start of two
defrosts 0=function disabled
0...250 6h 1-2 hours/min/sec
dt1 Unit of measurement for defrost times (par. dit)
0=”dit” expressed in hours 1=”dit” expressed in minutes 2=”dit” expressed in seconds
0/1/2 0 2 num
NOTE: At level 1 the folders will only display all the level 1 parameters. At level 2 the folders will only display all the level
2 parameters. The symbol 1-2 indicates the parameters that are displayed at both menu levels
Defrosting controller-dEF label Compressor controller-CP label
dt2 Unit of measurement for duration of defrosting
(dEt parameter) 0= “dEt” parameter expressed in hours 1= “dEt” parameter expressed in minutes 2= “dEt” parameter expressed in seconds
0/1/2 1 2 num
dCt Selection of defrosting time count mode.
0=compressor operating hours DIGIFROST® method). Defrosting active only if compressor is on. 1 = device operating hours; defrost counting is always active when the machine is on 2=compressor stop Each time the compressor stops a defrosting cycle defrost cycle is performed according to par. dtY
3=disabled
0/1/2/3 1 1-2 flag
dOH Delay between start of first defrosting operation
and start-up of instrument.
0...59 0 1-2 min
dEt Defrosting time-out; determines maximum dura-
tion of defrosting.
1...250 30min 1-2 min/sec
EWDR 981 4/8
PAR. DESCRIPTION
RANGE DEFAULT VALUE* LEVEL** U.M.
dPO Determines when instrument starts up if
the defrosting cycle must be activated (if the temperature on the evaporator allows this) y=defrosting activated at start-up n=defrosting not activated at start-up
n/y n 1-2 flag
tcd minimum time for each compressor state
before defrosting “Ontime if >0; “Offtime if >0
-31...31 0 2 min
Cod Compressor “Off” time before defrost cycle.
The compressor is not turned on if a defrost cycle is expected in the time indicated by the parameter. 0=Function excluded
0...60 0 2 min
Defrosting controller-dEF label
NOTE: At level 1 the folders will only display all the level 1 parameters. At level 2 the folders will only display all the
level 2 parameters. The symbol 1-2 indicates the parameters that are displayed at both menu levels
Alarms-AL label
Att Determines if “LAL” and “HAL” are expressed as
absolute values or as a differential related to the set point 0=absolute value 1=value related to set point
0/1 0 2 flag
AFd Alarm differential
1.0...50.0 2.0 1-2 °C/°F
HAL (2) Maximum alarm. Temperature limit (whose
absolute or relative value status is regulated by “Att”) above which the alarm is activated.
LAL...150.0 50.0 1-2 °C/°F
LAL (2) Minimum alarm. Temperature limit (whose
absolute or relative value status is regulated by “Att”) below which the alarm is activated.
-50.0...HAL -50.0 1-2 °C/°F
PAO (3) Alarm exclusion time after start-up of instrument
following a power failure
0...10 0 1-2 hours
dAO Alarm exclusion time after defrosting
0...999 0 1-2 min
EWDR 981 5/8
OAO High and low temperature alarm delay after
disabling digital input (door closed)
0...10 0 2 hours
tdO Time out after alarm signal when digital input is
disabled (door open)
0...250 0 2 min
tAO (3) Temperature alarm delay time
0...250 0 1-2 min
EAL External alarm to lock controllers
n=does not lock, y=locks
n/y n 2 flag
AOP Polarity of alarm output:
0 = alarm active and output disabled; 1 = alarm active and output enabled
0/1 1 2 flag
PbA Configuration of temperature alarm on
probe 1 and/or 3: 0=on probe 1 (thermostat control) 1=on probe 3 (display) 2=on probe 1 and 3 (thermostat control and display) 3=on probe 1 and 3 (thermostat control and display)
on external threshold
0/1/2/3 0 2 num
SA3 Probe 3 alarm set point
-50.0...150.0 50 2 °C/°F
dA3 Probe 3 alarm differential
-30.0...30.0 2.0 2 °C/°F
dt Dripping time
0...250 0 1-2 min
Light & digital inputs
Lit Label
dLt Light relay disabling delay after closing
door if “dSd”=y
0...31 0 2 min
OFL Light relay disabled even if disabling delay
“dLt” is active
n/y n 2 flag
dAd Delay in activating digital input
0...255 0 2 min
dOd Digital input switches off loads
n/y n 2 flag
dSd Light relay enable from door switch.
n = door open, light does not turn on; y = door open, light turns on (if it was off)
n/y y 2
flag
PAR. DESCRIPTION
RANGE DEFAULT VALUE* LEVEL** U.M.
NOTE: At level 1 the folders will only display all the level 1 parameters. At level 2 the folders will only display all the
level 2 parameters. The symbol 1-2 indicates the parameters that are displayed at both menu levels
EWDR 981 6/8
LOC Keyboard locked. Parameters can still be
programmed. n= keyboard not locked y= keyboard locked
n/y n 1-2 flag
PA1 Contains the password for level 1 parameters.
Enabled if not 0
0...250 0 1 num
PA2 Contains the password for level 2 parameters.
Enabled if not 0
0...250 0 2 num
ndt Display with decimal point.
n= without decimal point (only whole numbers) y= with decimal point
n/y n 1-2 flag
CA1 Temperature value to be added to that read
by probe 1 as specified by parameter CAI
-12.0...12.0 0 1-2 °C/°F
CA2 Temperature value to be added to that read
by probe 2 as specified by parameter CAI
-12.0...12.0 0 1-2 °C/°F
CA3 Temperature value to be added to that read
by probe 3 as specified by parameter CAI
-12.0...12.0
0
1 °C/°F
Display - diS label
CAI Intervention of offset on display, thermostat
control or both: 0 = only modifies the temperature displayed 1 = adds to the temperature used by controllers
not the temperature displayed that remains unchanged.
2= adds to temperature displayed that is
also used by controllers.
0/1/2
2
2 num
LdL Minimum value that can be displayed
-55.0...140
-50.0
2 °C/°F
HdL Maximum value that can be displayed
-50.0...302
140.0
2 °C/°F
ddL display during defrosting:
0= displays temperature read by thermostat control probe 1= displays temperature read entering defrost cycle until set point is reached 2= displays “dEF” label during defrosting until
set point is reached (or when Ldd expires)
0/1/2 1 1-2 flag
Ldd Time out for unlocking display (with ddL=2) if
defrosting lasts too long
0...255
0
1-2 min
dro (4) Select °C or °F to display temperature:
0= °C 1= °F
0/1
0
1-2 flag
ddd Value to be displayed:
0 = Set point 1 = probe 1 (thermostat control) 2 = disabled 3 = probe 3 (display)
0/1/2/3
1
2 num
H00 Selects probe, PTC or NTC 0= PTC 1= NTC
0/1
1
1-2 flag
H02 Quick activation time for functions with configured
buttons. Not possible for aux (time expected = 1 second)
0...15
5
2 sec
H11 (5) Configuration of digital inputs/polarity:
0= disabled 1 = defrost 2 = reduced set point 3 = auxiliary 4 = door switch 5= external alarm 6= disabled 7= stand-by (On/Off) 8= maintenance request
-8...8
4
2 num
H08 Stand-by operating mode
0= only display switched off; 1= display on and controls locked; 2= display off and controls locked
0/1/2
2
2 num
H06 Button/input aux/door switch light active when
instrument is off
n/y
y
2 flag
H12 (5) Configuration of digital inputs/polarity
Same as H11
-8...8
0
2 num
Configuration- CnF label
PAR. DESCRIPTION
RANGE DEFAULT VALUE* LEVEL** U.M.
Configuration- CnF label
EWDR 981 7/8
H21 Digital output configurability (C):
0= disabled 1= compressor 2= defrosting 3= fans 4= alarm 5= auxiliary 6= stand-by 7= light 8= buzzer 9= defrosting on 2nd
evaporator
0...9
1
2 num
H31 UP button configurability
0=disabled 1=defrosting 2=auxiliary 3=reduced set point 4=disabled 5=disabled 6=light 7=stand-by 8= maintenance request
H32 DOWN button configurability
Same as H31(0=disabled default)
0...8
0
2 num
H33 ESC button configurability
Same as H31(0=disabled default)
0...8
0
2 num
0...8
1
2 num
H41 Presence of control probe:
n= not present y= present
n/y
y
2 num
H43 Presence of display probe:
n= not present y= present (display probe)
n/y
y
2 flag
PA2*** in the CnF folder when you enter the correct password you can access level 2 parameters from label PA2 by pressing the “set” button
rEL Device version. Read only parameter
/
/
1 /
tAb Parameter table; Reserved; Read only parameter
/
/
1 /
Copy Card
Fpr label
UL Transfer of parameter map from instrument to
Copy Card
dL Transfer of parameter map from Copy Card to
instrument
Fr (6) Formatting. Cancels all data in the Copy Card
/
/
1 /
/
/
1 /
/
/
1 /
NOTES:
(1) See Duty Cycle diagram (2) If relative values are present (par. Att=1) parameter HAL is set to positive values and the parameter LAL is set to negative values (-LAL) (3) Refers exclusively to high and low temperature alarms (4) when changing from °C to °F or vice versa the set points, differentials, etc. are NOT converted (for example, “set=10 °C becomes set=10°F”) (5) CAUTION: positive or negative values change polarity, Positive values: active input when contact is closed; Negative values: active input when contact is open. (6) If the Fpr parameter is used, the data previously stored on the Copy Card will be permanently lost. This operation cannot be undone
* Value: to be compiled manually by user with any custom settings (if different from default settings) ** Level: indicates the visibility level of parameters accessed using a password (see relevant paragraph) *** PA2 is visible at level 1 in the CnF folder
whereas it can be modified at level 2 in the diS folder
H25 Digital output configuration. (E) Same as H11.
0...9
4
2 num
if alarm(s)
present
AL
Pb1
SEt
set
alarms
Pb1 value
SEt value
set
change SEt
value
show alarms
press and release
(quickly/single press)
Pb3
Pb3 value
is present
MACHINE STATUS MENU
Ont, OFt parameters programmed for Duty Cycle
Ont OFt Compressor output
00OFF 0>0OFF >0 0 ON >0 >0 dc
OUT
On
Off
OFtOnt
Ont
The error condition for probe 1
(compressor) causes the following:
- E1 code appears on display
- the controller is activated as indicated by the “Ont” and “OFt” parameters if programmed for Duty Cycle
DUTY CYCLE DIAGRAM
CP
dEF
AL
level 1 par
level 1 par
level 1 par
PA1›0
set PA1 value
diS
CnF
Fpr
level 1 par
level 1 par
level 1 par
level 2 par
level 2 par
level 2 par
level 2 par
level 2 par
CP
dEF
diS
CnF
PA2›0
set PA2 value
level 2
level 1
change
par value
scroll
parameters
press for 5 sec
level 2 par
Lit
AL
PA2
PROGRAMMING MENU
EWDR 981 8/8
The technical characteristics in this document concerning measurements (range, accuracy, resolution, etc.) refer to the instrument in the strictest sense and not to any accessories provided such as probes, for example. This means that an error introduced by the probe is added to any error that is in the instrument.
Minimum temperature alarm
Maximum temperature alarm
LAL
AFd
HAL
AFd
1
set+ LAL
AFd
Off
set+HAL
AFd
set
2
Temperature lower than or equal to LAL (LAL with sign)
Temperature lower than or equal to set point+LAL (set-|LAL|) (LAL only negative)
Temperature higher than or equal to HAL (HAL with sign)
Maximum temperature alarm back swing
Minimum temperature alarm back swing
Temperature higher than or equal to LAL+AFd
Temperature lower than or equal to HAL-AFd
Temperature expressed as an absolute value (par “Att0) Abs(olute)
Temperature expressed in relation to set point (par “Att”=0) reL(ative)
if Att=reL(ative) LAL must be negative: therefore set point+LAL<set point because set point+(-|LAL|)=set point-|LAL|
MIN/MAX ALARM MENU
WIRING DIAGRAM
TERMINALS
1-2 Probe input 1 (room probe) 1-4 Probe input 3 (display or 2nd evaporator probe) 5-6 Digital input 1 5-7 Digital input 2 8-9 Auxiliary output 12Vc/20mA (E) see par. H25
(alarm default) 13-14 Power supply 230Va A TTL input for Copy Card 20-21-22* N.O. relay output (C) see par. H21 (compressor default) 19-22* N.C. relay output (C) see par. H21 (compressor default)
EWDR 981
1
245678910 11 12
13
14 19 20 21
22
TTL
A
(C)
AUX OUT
12VC/20mA
P
o
w
e
r
S
u
p
p
ly
Pb1
Pb3
D.I.1
D.I.2
(E)
Eliwell & Controlli s.r.l.
Via dell'Industria, 15 Zona Industriale Paludi 32010 Pieve d'Alpago (BL) ITALY Telephone +39 0437 986111 Facsimile +39 0437 989066 Internet http://www.eliwell.it
Technical Customer Support:
Email: techsuppeliwell@invensys.com Telephone +39 0437 986300
Climate Controls Europe An Invensys Company
3/05 eng cod. 9IS43091
* Relay characteristics
Relay output (C) 16A 1hP 250Va
DISCLAIMER
This document is the exclusive property of Eliwell and cannot be reproduced and circulated unless expressly authorized by Eliwell. Although Eliwell has taken all possible measures to guarantee the accuracy of this document, it declines any responsibility for any damage arising out of its use. The same applies to any person or company involved in preparing and writing this document. Eliwell reserves the right to make any changes or improvements without prior notice and at any time.
Temperature higher than or equal to set point+HAL (HAL only positive)
Temperature higher than or equal to set point + LAL + AFd
set point -|LAL|+AFd
Temperature lower than or equal to set point+HAL-AFd
Loading...