Eliwell EM300, EM300LX User Manual

USER INTERFACE
The user has a display and four buttons for controlling instrument status and pro­gramming.
BUTTONS AND MENUS ACCESSING
AND USING MENUS
The resources are arranged in a menu that can be accessed by pressing and quickly releasing the “set” button (Machine Status menu) or holding down the “set” button for more than 5 seconds (Programming menu). To access the contents of each folder indicated by the relevant label, just press the “set” button once. You can now scroll through the contents of each folder, modify it or use its functions. If you do not use the keyboard for over 15 seconds (time-out) or if you press the “fnc” button once, the last value shown on the display is confirmed and you return to the previ­ous screen mask.
MACHINE STATUS MENU (See Machine Status Menu Diagram)
To access the Machine Status menu, press the “set” button and quickly release it (The “SP1” label appears.
V/I
V/I
MODELS
MODELS
ONL
ONLY)Y)
(If alarms are active, with the exception of faulty probes/probe errors, the “AL” label appears). By using the “UP” and “DOWN” buttons you can scroll through the other folders in the menu: the folders are indi­cated below in the order they appear:
-AL: alarm folder
LXLXMODELS
MODELS
ONL
ONLYY
NOTE: The AL folder appears if high or low temperature alarms are present.
Alarm (IF PRESENT)
•ON for active alarm;
•blinking for silenced
alarm that is still present
LLEEDDSS AANNDD KKEEYYPPAADD
decimal point: associated with displaying decimal point
•ON if enabled-decimal point present (see ndt parameter);
•OFF if disabled-decimal
point NOT present (see
ndt parameter);
for NTC/PTC
or Pt100-TC
input
for V/I input
(current-
voltage std
signals)
bar for
V/I input
(pressure)
%RH for
V/I input
(humidity)
EM300(LX)
electronic digital indicator
code 9IS43049
rel. 6/06
GB
UP Button Scrolls through the menu items
Increases the values
DOWN button Scrolls through the menu items
Decreases the values
fnc button ESC function (quit)
Set point button 1-Accesses Machine Status Menu
(ACTIVE ALARMS, PROBE READ­ING) and related labels/values; 1-Accesses Programming Menu (PARAMETERS, COPY C ARD (LX models)) and related labels/val­ues; 3-Confirms commands
fnc
set
ALARM PRESENT?
If an alarm condition exists when the “Machine Status” menu is accessed, the “AL” folder label appears with the alarm codes. (see section on “Diagnostics”).
if alarm(s)
present
AL
set
alarms
set
scroll alarms
press and release
(single press)
MACHINE STATUS MENU DIAGRAM
HOW TO LOCK THE KEYBOARD
Keyboard operating can be locked by pro­gramming the “Loc” parameter (see folder with “diS” table). If the keyboard is locked you can only access the Programming MENU (see) Functions CANNOT be activat­ed with buttons (to silence alarms, for example).
TELEVIS SYSTEM (LX MODELS ONLY)
The Televis remote control systems can be connected using the TTL serial port. The TTL- RS 485 interface module must be used:
• 130 or 150 BUS ADAPTER.
• 350 BUS ADAPTER for EM300(LX) with V/I. Pt100/TC input.
To configure the instrument to do this, you need to access the folder with the “Add” label and use the “dEA” and “FAA” para­meters.
ONLY MODELS LX
AL
Add
dIS
CnF
set
level 1 par
level 1 par
level 1 par
level 1 par
set
set
set
set
PA1≠0
set
set PA1 value
Fpr
level 1 par
set
level 1
change
par value
scroll 
parameters
press for 5 sec
ONLY MODELS LX
ONLY MODELS LX
rE1
input VI
PROGRAMMING MENU
DIAGRAM
230V
PC with
Televis package
N
L
only use 350 Bus Adapter for model with V/I, Pt100/TC input
EM300(LX) 2/6
USING COPY CARD (LX MODELS ONLY)
The Copy Card is an accessory connected to the TTL serial port used for quick programming of the unit parameters (upload and download parameter map to one or more units of the same type). Operations are described below:
Fr-Format
This command is used to format the copy card. This is necessary if
•it is used for the first time,
•it is used with models that are incompatible,
•after use with a specific model, another model must be connected. Warning: when the copy card has been programmed using the “Fr” parameter all the data entered is can­celled. This operation cannot be undone.
UL-Upload
This operation unloads the programming para­meters from the instrument.
dL-Download
This operation downloads the programming parameters to the instrument.
NOTE:
• UPLOAD: instrument —> Copy Card
• DOWNLOAD: Copy Card —> instrument.
The operations are performed by accessing the folder identified by the “FPr” label and selecting the “UL”, “dL” or “Fr” commands. The operation is confirmed by pressing the “set” button. If the
operation is successful, a “y” is displayed where­as if it is unsuccessful an “n” will be displayed.
Download “from reset”
Connect the copy card with the instrument OFF. When the instrument is switched on the pro­gramming parameters will be downloaded into the copy card; when the lamp test has been completed, the following appear on the display for about 5 seconds:
• label dLY if copy operation is successful
• label DLn if operation fails PLEASE NOTE:
• after downloading the instrument will work with the parameter map settings that have just been downloaded.
PROGRAMMING MENU (See Programming Menu Diagram)
1) Displaying level 1 parameters
To access the Programming menu, hold the “set” button for more than 5 seconds.
Level 1 parameters can be protected by a PASSWORD* (defined by parameter DIS/PA1) If the PASSWORD is enabled, the label “PA1” will appear when you access the Programming Menu; press the “set” button and the value “0” will appear; enter the password using the “UP”/”DOWN” but­tons and press the “set” button again. This
allows you to access the level 1 parame­ters. Use the “UP” and “DOWN” buttons to scroll through the other folders; the
folders will only display level 1 parame­ters. PLEASE NOTE: It is strongly recommended
that the unit is switched off and on again each time parameter configuration is changed in order to prevent malfunction­ing of the configuration and/or ongoing timings (compulsory for selection of
probe type and count parameters).
*PASSWORD
The “PA1” password allows level 1 and level 2 parameters to be accessed. This password is not present in the standard configuration. To enable it (value 0) and assign it the desired value, access the Programming menu in the “diS” folder. If the password is enabled, it will be requested when entering the “Programming” menu.
The alarm condition is always signalled by a
buzzer (if present) and the alarm icon LED
(IF PRESENT)
The alarm signal from the
• control probe that measures values outside the nominal reading range
• control probe faulty/shorted/open probe for 4…20 mA* model appears directly on the instru­ment display as E1. Note: to prevent false alarms, the error condition must persist for more than 10 seconds. An error condition in probe 1 (regulation) leads to:
• E1 code appearing on display
PLEASE NOTE: In case of wrong connection of the 3rd wire (Pt100 sensor) in “AL” folder it will appear the label “Pt3”. For few seconds the display will shows a uncor­rect temperature.
MAXIMUM AND MINIMUM TEMPERATURE ALARM
If an alarm condition occurs and alarm exclu­sion times are not in progress (see alarm exclu­sion parameters), the fixed alarm icon comes on.
DIAGNOSTICS
DISPLAY
E1
*faulty/shorted/open probe for 4…20 mA model; the other probes, if shorted, will measure the 0V or 0mA value)
FAULT
Probe 1 (control) faulty*
Table of faulty probes
DISPLAY
AH1
AL1
Press any button to silence the alarm. The LED will start to blink.
ALARM
High temperature alarm (referring to thermostat control probe or probe 1) Low temperature alarm (referring to thermostat control probe or probe 1)
Alarm table
Alarms are considered as absolute values This alarm condition can be viewed in the “AL” folder with labels “AH1-AL1”.
The maximum temperature alarm occurs when the probe temperature is:
•higher than or equal to HAL/HA1
The minimum temperature alarm occurs when the probe temperature is lower than or equal to LAL/LA1
The maximum temperature alarm back swing occurs when the probe temperature is:
• higher than or equal to HAL/HA1 - AFd
Minimum temperature alarm back swing occurs when the probe temperature is:
•higher than or equal to LAL/LA1 + AFd
LAL/LA1
AFd
HAL/HA1
AFd
Max/Min. Alarm Diagram (minimum and maximum temperature)
UPLOAD
DOWNLOAD
EM300(LX) 3/6
GENERAL TECHNICAL DATA
Front protection: IP65. Casing: PC+ABS UL94 V-0 resin plastic body, polycarbonate front, thermoplastic resin buttons. Dimensions: front panel 74x32 mm,
•depth 30 mm (terminals excluded)
•’switching models’: depth 60 mm (termi­nals excluded) Mounting: on panel, with drilling template 71x29 mm (+0.2/-0.1 mm). Operating temperature: -5…55 °C. Storage temperature: -30…85 °C. Usage ambient humidity: 10…90 % RH (non-condensing). Storage ambient humidity: 10…90% RH (non-condensing).
Buzzer output: only in certain models.
(LX MODELS ONLY)
Serial: TTL for connection to
•Copy Card;
•TelevisSystem.
Consumption:
• 230V model: 1,8 W max.;
• 12V model: 0.5 W max.
• ‘switching’ models: 3 W max.
Power supply:
• 230Va ±10% 50/60 Hz or
• 12Va/c ±10% 50/60 Hz or
• 12-24Va ±10%, 12-36Vc ±10% SELV
(only models NTC/PTC & Pt100-TC)
Warning: check the power supply specified on the instrument label: contact the Sales Office for further information
EM300(LX) TECHNICAL DATA
WITH PTC/NTC INPUT
Display range:
• NTC probe: –50.0…110.0°C (–58…230°F);
• PTC probe: –55.0…140.0°C (–67…284°F) on display 3 1/2 digits + sign. Analogue input: one PTC or NTC input (parameter-selectable). Measurement range: from -55 a 140 °C. Accuracy: better than 0.5% of bottom scale + 1 digit. Resolution: 0.1°C (0.1°F up to +199.9°F; 1°F above).
WITH V INPUT - I INPUT
Display range:
-99…100 (ndt = n), -99.9…100.0 (ndt = y),
-999…1000 (ndt = int) on display 3 1/2 digits + sign. Analogue input (see table):
• V input (0-1V, 0-5V, 0-10V) or
• I input (0…20mA, 4…20mA);
Measurement range: from –999 to 1000. Accuracy: better than 0.5% of bottom scale + 1 digit. Resolution: 1 or 0.1 digits depending on parameter settings.
Current input impedance: 100 ohm Voltage input impedance: 20K ohm NOTE: for 3-wire sensor connections the maximum current supplied by the device is 25mA (measurement + sensor power supply)
WITH PT100-TC INPUT
Viewing range:
• Pt100 model : -200…800°C (-328...1472°F),
• TcJ model -40…760°C (-40...1400°F),
•TcK model -40…1350°C (-40...2462°F)** ** (-40..1999°F) above which SuP is displayed, with decimal point, selectable through parameter ndt on 3 digit &
1
/2+ mark display.
Analogue input: one PT100 input or TcJ or TcK depending on model. Measuring range: from -200 to 1999. Accuracy:
• Pt100 model : 0,5% for all scale + 1 digit; 0,2% from -150 to 300°C.
• TcJ model: 0,4% for all scale + 1 digit;
• TcK model 0,5% for all scale + 1 digit; 0,3% from -40 to 800°C. Resolution:
• Pt100 model: 0,1°C (0,1°F) within 199,9 °C, 1°C (1°F) over
• TcJ/TcK model 1°C (1°F). Power supply: 12/24 Va/c ±10% or 230Va
±10% 50/60 Hz.
NOTE: in model Pt100 the thermal drift in the -5...55°C range can reach 3°C.
ELECTRICAL WIRING
Warning! Always switch off machine before working on electrical connec­tions.
The instrument has screw terminals for connecting electrical cables with a diame­ter of 2,5 mm
2
max. (only one conductor per terminal for power connections). for terminal capacity, see the label on the instrument. Make sure that the power voltage complies with the device voltage. Probe cables, power supply cables and the TTL serial cable should be kept separate from power cables.
The sensor can be extended using an ordi­nary bipolar cable (note that extending the probe may affect the electromagnetic compatibility (EMC) of the instrument: special care must be used when wiring).
• version with PTC/NTC input:
The sensor has no connection polarity.
• version with V/I/Pt100-TC input: NOTE: Check the probe connection polarity.
NOTE: The technical characteristics in this document concerning measure­ments (range, accuracy, resolution, etc.) refer to the instrument in the strictest sense and not to any accessories provid­ed such as probes, for example. This means that an error introduced by the probe is added to any error that is in the instrument.
The unit has been designed to be panel-mounted: Drill a 29x71 mm hole, insert the keyboard and fix it in place with the special brackets provided. Do not assemble the key­board in excessively dirty and/or dirty locations because it is designed to be used in locations with normal pollution levels. Always make sure that the area next to the unit cooling slits is adequately ven­tilated.
switching models
only TTL models
MECHANICAL ASSEMBLY AND CUT-OUT
Model
EM300(LX) EM300(LX) EM300(LX) EM300(LX) EM300(LX) EM300(LX)
NOTE: For TECHNICAL DATA and connections using EWPA and EWHS probes refer to the relevant technical data sheets and/or diagram on last page. * external power supply. For external power supply characteristics, see characteristics of connected sensor.
Analogue input
NTC-PTC probe Pt100-TcK-TcJ probe EWPA 007/030 EWHS 280/300/310* 0…1/5/10 V voltage input 0/4…20mA current input
Analogue input table
PARAMETER
AFd HAL (!)
LAL (!)
PAO (!)
tAO
tP
dEA (!)
FAA (!)
LOC
PA1
ndt
CA1
LdL HdL dro(*)
H00 (!) rEL tAb
UL
dL
Fr
* DEFAULT column: for these parameters the default depends on the model.
** VALUE column: to be compiled manually with any custom settings (if different from default value).
EM300(LX) 4/6
PARAMETER TABLE FOR EM300(LX) with NTC/PTC probe
DESCRIPTION
ALARMS (folder with “AL” label)LXLX
MODELS
MODELS
ONL
ONLYY
Alarm Fan differential. Alarm differential. Higher ALarm. Maximum temperature alarm. Temperature value (considered as absolute value) which if gone above triggers the alarm signal. Lower ALarm. Minimum temperature alarm. Temperature value (considered as absolute value) which if gone below triggers the alarm signal. Power-on Alarm Override. Alarm exclusion time after instrument start-up, after a power failure. Temperature Alarm Override. Temperature alarm signal delay time. silences alarm with button
COMMUNICATION (folder with “Add” label) LX MODELS ONLY
DEA= device number within the family (valid values: from 0 to 14) FAA= device family (valid values: from 0 to 14) The value couple FAA and dEA represents the network address of the device and it is indicated in the following way: "FF.DD" (where FF=FAA and DD=dEA).
DISPLAY (folder with “diS” label)
Keyboard locked (set point and buttons). However, you can still access the parameter programming menu and modify parameters including the status of this parameter to allow keyboard unlock­ing. y = yes; n = no. Password 1. When enabled (value is not 0) it represents the access key to level 1 parameters. number display type. Display with decimal point. y = yes; n = no, int= whole numbers. Calibration 1. Positive or negative temperature value that is added to the value read by control probe (analogue input) before being displayed or used for regulation. Minimum value that can be displayed. Maximum value that can be displayed. Select °C or °F to display temperature read by probe. N. B.: when changing from °C to °F
or vice versa the temperature values are NOT converted (e.g. 10°C becomes 10°F)
CONFIGURATION (folder with “CnF” label)
Selection of input type, PTC/NTC Device version. Read only parameter. Reserved. Read only parameter.
COPY CARD (folder with “Fpr” label) LXLX
MODELS
MODELS
ONL
ONLYY
UpLoad: transfer of programming parameters from instrument to Copy Card. downLoad: transfer of programming parameters from Copy Card to instrument. Format. Cancels all data entered in the copy card.
N.B.: if “Fr” parameter (formatting of copy card ) is used the data entered in the card will be permanently lost. This oper­ation cannot be undone. After the operation with the Copy Card the device must be switched off and then switched back on
DEFAULT*
2.0
50.0
-50.0
0
0
y
0 0
n
0
n
0
-50.0 (-200 (§))
140.0 (1500 (§)) 0
*
/ /
/
/
/
RANGE
1.0...50.0
LAL...150.0
-50.0...HAL
0...10
0...250
n/y
0…14 0…14
n/y
0...250
n/y
-12.0...12.0
(-30,0...30,0) (§)
-55.0…HdL (-328,0...HdL) (§) LdL…302 (LdL...1999) (§)
0/1
0/1
/ /
/
/
/
VALUE**
U.M.
°C/°F °C/°F
°C/°F
hours
min
flag
num num
flag
num
flag
C/°F
C/°F C/°F
flag
flag
/ /
/
/ /
LX
LX
MODELS
MODELS
ONL
ONL
Y
Y
LX
LX
MODELS
MODELS
ONL
ONL
Y
Y
LX
LX
MODELS
MODELS
ONL
ONL
Y
Y
RESPONSIBILITY AND RESIDUAL RISKS Eliwell & Controlli s.r.l. shall not be liable for any dam-
ages deriving from:
- installation/use other than that prescribed and, in par­ticular, which does not comply with the safety standards specified in the regulations and/or those given herein;
- use on equipment that does not guarantee adequate protection against electric shock, water or dust when assembled.
- use on equipment that allows dangerous parts to be accessed without the use of tools;
- tampering with and/or alteration of the product;
- use on equipment that does not comply with the stan­dards and regulations in force.
DISCLAIMER This document is exclusive property of Eliwell & Controlli s.r.l. and cannot be reproduced and circulat­ed unless expressly authorized by Eliwell & Controlli s.r.l.. Although Eliwell & Controlli s.r.l. has taken all possible measures to guarantee the accuracy of this document, it declines any responsibility for any damage arising out of its use. The same applies to any person or company involved in preparing and writing this manual. Eliwell & Controlli s.r.l. reserves the right to make any changes or improvements without prior notice and at any time.
CONDITIONS OF USE
PERMITTED USE
For safety reasons the instrument must be installed and used in accordance with the instructions supplied. Users must not be able to access parts with dangerous voltage levels under normal operating conditions. The device must be protected from water and dust depending on the specific appli­cation and only be accessible by using special tools. (except for the front panel). The device is ideally suited for use on household appliances and/or similar refrigera­tion equipment and has been tested with regard to safety in accordance with the European harmonized reference standards.
It is classified as follows: Installation: Class II where applicable (front keypad) Protection class: 2 Material group: IIIa Device status: Permanently connected fixed device. Measurement category: III
UNPERMITTED USE
The use of the unit for applications other than those described above is forbidden.
(°) The mathematical conversion for temperature is °F=(9/5)* °C+32. For example: 32°F=0°C;
50°F=10°C. dro parameter: when changing from °C to °F or vice versa the mathematical conversion is NOT performed and the values are NOT modified.
All the temperature values set will therefore need reviewing. e.g. with a set point set at 10°C, when changing the value to °F the set point will become 10°F and not 50°F (according to conversion table)
(§) Pt100/TC model
EM300(LX) 5/6
PARAMETER
HA1 (!)
LA1 (!)
AFd PAO (!)
tAO
tP
H00 (!) H03
H04 rEL tAb
* DEFAULT column: for these parameters the default depends
on the model.
(!) CAUTION!
• If one or more parameters marked with (!) are modified, the device must be switched off after the modification and then switched back on
PLEASE NOTE: We strongly recommend that you switch the instrument off and on again each time parameter configuration is changed in order to prevent malfunction­ing of the configuration and/or ongoing timings.
DESCRIPTION
ALARMS (folder with “rE1” label)LXLX
MODELS
MODELS
ONL
ONLYY
Higher ALarm. Maximum temperature alarm. Temperature value (considered as absolute value) which if gone above triggers the alarm signal. Lower ALarm. Minimum temperature alarm. Temperature value (considered as absolute value) which if gone below triggers the alarm signal.
ALARMS (folder with “AL” label) LXLX
MODELS
MODELS
ONL
ONLYY
Alarm Fan differential. Alarm differential. Power-on Alarm Override. Alarm exclusion time after instrument start-up, after a power failure. Temperature Alarm Override. Temperature alarm signal delay time. silences alarm using button
COMMUNICATION (folder with “Add” label) LX MODELS ONLY
SEE
SEE
EM300
EM300 TT
ABLE above DISPLA
ABLE above DISPLA
Y (folder with “
Y (folder with “
diS” label)
diS” label)
SEE
SEE
EM300
EM300 TT
ABLE above
ABLE above
V/I
V/I
MODELS
MODELS
ONL
ONLY:Y:
NOTE 1: ndt parameter has range n/y/int int=whole numbers.
NOTE 2: LdL has range -99...HdL or -99.9...HdL or -999...HdL according to ndt parameter setting HdL has range -LdL...100 or
-LdL...100.0 or -LdL...1000 according to ndt parameter setting
CONFIGURA
CONFIGURA
TION (folder with “CnF” label)
TION (folder with “CnF” label)
Selection of probe type Minimum value of voltage / current input
Maximum value of voltage / current input Device version. Read only parameter. Reserved. Read only parameter.
COPY CARD (folder with “Fpr” label) LXLX
MODELS
MODELS
ONL
ONLYY
SEE EM300 TABLE above
DEFAULT*
*
(§) 1200
*
(§) -200
2.0
0
0
y
(§) 2
*
* / /
RRAANNGGEE ((VV//II))
LA1...150.0
-50.0...HA1
1.0...50.0
0...10
0...250
n/y
NOTE 3:
NOTE 3:
the dro parameter is
the dro parameter is
not present
not present
§§§
-99.0…100.0 (ndt= y)
-999…1000 (ndt= int)
as above
/ /
VALUE**
U.M.
°C/°F
°C/°F
°C/°F
hours
min
flag
num C/°F
C/°F
/ /
PARAMETER TABLE FOR EM300(LX) with V/I & Pt100-TC input
LX
LX
MODELS
MODELS
ONL
ONL
Y
Y
LXLXMODELS
MODELS
ONL
ONLYY
LXLXMODELS
MODELS
ONL
ONLYY
§§§
010/05/01 for VOLTAGE INPUT
420/020 for CURRENT INPUT
(§) Pt100/TC model
RRAANNGGEE ((PPtt110000))
LA1...1999
-328...HA1
1.0...50.0
0...10
0...250
n/y
NOTE 3:
NOTE 3:
the dro
the dro
parameter is
parameter is
present
present
(§) Jtc/HtC/Pt1
not present
not present
/ /
EM300(LX) 6/6
Supply
1,8 W max
3 4 5
+-
+12V
V
3 4
Pb1
NTC/PTC
V
1 2
3 4 5
A
Supply
0.5 W max
EM300(LX) - 12V & 230V
LX models ONLY
Pt100-TC
3 4 5
+
-
(230V model)
(12V model)
3 4 5
+-
+12V
I
I
WIRING DIAGRAMS
6/2006 - GB ­cod. 9IS43049
TERMINALS
1 - 2 Power supply
• 230Va ±10% 50/60 Hz or
• 12Va/c ±10% 50/60 Hz or
• 12-24Va ±10%, 12-36Vc ±10% SELV (only models
NTC/PTC & Pt100-TC)
NTC/PTC Models
3 - 4 Pb1 probe input
V Models
3 - 4 - 5 Voltage input (3=GND; 4=signal; 5=+12V)*
I Models
3 - 4 - 5 Current input (3=GND; 4=signal; 5=+12V)*
Pt100-TC Models
4 - 5 TcJ-TcK Input 3 - 4 - 5 Pt100 3 wires Input
LX MODELS ONLY
A TTL input for Copy Card and connection to
TelevisSystem
* according to model
* Check probe connection polarity.
EWHS 300/310 3 wires Power Supply from EM300(LX)
Probe
EWHS 280 2 wires Power Supply from EM300(LX)
Probe
EWPA 007/030 2 wires External Power Supply for Transducer
Transducer
brown
blue
brown
white
GND (only EWHS310)
V+
Power Supply for Transducer
RH/T
3 4 5
3 4 5
3 4 5
EWPA 007/030 2 wires Power Supply from EM300(LX)
Transducer
brown
white
3 4 5
EWHS 310 HUMID. + TEMP. Power Supply from external Power Supply
GND
V+
RH
3 4 5
3 4 5
T
230V
GND
GND IN V+
GND IN V+
EWHS 310 HUMID. + TEMP. Power Supply from EM300(LX)
GND
V+
RH
3 4 5
3 4 5
T
GND
GND IN V+
GND IN V+
EWHS 310
EWHS 310
3 4
Pb1
NTC/PTC
1 2
3 4 5
A
Supply
3 W max
EM300(LX) - 12-24Va 12-36Vc
LX models ONLY
Pt100-TC
+
-
1 2
Supply
3 W max
EM300 current
EM300 current
Loading...