Elitech TI-40 User Manual

Overview
The temperature indicator is mainly used in the fields of pharmaceuticals, vaccine, blood, food, flowers and plants, laboratories, etc. especially for the places that require using temperature indicator of high waterproof grade in cold chain storage and transportation.
About the indicator
Parameter conguration
Double-faced adhensive tape on the back
LCD screen
Start & stop botton
Bar coade & serial No.
1) Before delivery, the indicator’s parameters are configured. Just press START button to start it for use.
2) Default configuration:
Operation instruction
1. Press (START) botton for about five seconds, " " will be on, The indicator successfully starting.
2. If there is start delay, short press the button, " " will flash 30 seconds to off; if no start delay, then " " will be on for 30 seconds to off.
3. When the LCD temperature indicator is on, if the current status is start delay, you can not stop the indicator by stop (STOP) buttun. After start delay time, press the STOP button for about 5 seconds, the " " icon on the LCD screen will be off, " " lights up, it means that the LCD temperature indicator has stopped.when the LCD temperature indicator is stopped, it will not be possible to turn on the indicator by pressing the start (START) button.
4. short press the button to view the LCD temperature indicator's information, such as the maximum & minimum temperature, average teprature, alarm information, etc.
Working instruction
1. Not started state: the triangle in LCD screen is not displayed, see Figure 1
2. Start delay status: the triangle in LCD screen flashes for 30 seconds and goes off, see Fig. 2
Fig 1 LCD indicator not started
Fig 2 LCD indicator start delay
Start delay: 10min
Log interval: 5min Temp H2 (high): 45
Alarm type: cum (cumulative)
Alarm delay: 60min Temp H1 (high): 30
Alarm type: cum (cumulative)
Alarm delay: 10h Temp L1 (low): -0.5
Alarm type: sin (single)
Alarm delay: 60min Temp L2 (low): -5
Alarm type: sin (single)
Alarm delay: 60min
1
2
3
4
Technical parameters
Temperature measuring range : -30℃ ~ 70℃ Storage range : -25℃ ~ 60℃ Temperature accuracy : -20℃~+ 40℃±0.5℃, other±1℃ Sensor : Internal NTC Log interval : 5minutes (default) Start delay : 1min~12hr,default setting (one min as step unit) Power suply : internal CR2450 lithium battery Bettery life : 2 yr at 25℃ with log interval 5 min Protection grade : IP67 Dimensions : 77mm*46mm*12 mm(L*W*H) Weight : about 40g Start : press the botton for over 5 seconds Stop : press the botton for over 5 seconds
It is recommended to store at room temperature before the temperature label is not activated. In order to ensure the waterproof level of the indicator, please do not steep the indicator in alcohol or oleic acid and other corrosive liquids for long time.
User Manual
TI-40 LCD Temperature Indicator
3. Working status (no alarm): the triangle in LCD screen stays on for 30 seconds and goes off
4. Working status (alarm): LCD screen shows the wrong number and the corresponding temperature segment, the screen will be off after 30 seconds.
Fig 3 LCD indicator Working status (no alarm)
Fig 4 LCD indicator working status (alarm)
YY:MM:DD
HH:MM:SS
LCD display interfaceStatus interface
Max, Min, Average
System time
Temperature range configuration H2 Alarm threshold 45.0 degree, alarm type single (Sin), alarm delay 1 hour.
H1 Alarm threshold 30.0 degree, alarm type cumulative (cum), alarm delay 10 hours.
HH:MM:SS
YY:MM:DD
L2 Alarm threshold -0.5 degree, alarm type Single (Sin), alarm delay 1 hour.
L1 Alarm threshold -0.5 degree, alarm type Single (Sin), alarm delay 1 hour.
Start time: 10 hours 9 minutes.
Temperature range alarm cumulative time
H2 temperature range alarm cumulative time 1 hour and 8 minutes Initial alarm time
H2 initial alrm time 12 Jan., 2017, 19:35:08
Serial No. Stop status (no alarm)
The last five number of serial No. 00018
User Manual
TI-40 LCD Temperature Indicator
V2.0
Elitech Technology, lnc.
www.elitechlog.com 1551 McCarthy Blvd, Suite 112 Milpitas, CA 95035 USA