Eli Ezer ESL-7800 Operation Manual

Preface
Thank you for purchasing our ESL-7800 slit lamp. The slit lamp will have another capability of photography, if you purchase the appliance with the photography accessories. At the same time, in order to meet the needs of the customs, we provide sorts of accessories, which can be used widely. Please read this manual carefully for the sake of your best use.
General Requirements for Safety
Please read carefully about following precautions to avoid unexpected personal injury as well as the product being damaged and other possible dangers.
Precautions
1.
Do not use this instrument in the environment prone to fire and blast or where there is much dust and with high temperature. Use it in the room and simultaneously be careful to keep it clean and dry.
2.
Check that all the wires are correctly and firmly connected before using. Ensure that the instrument is well grounded.
3.
Please pay attention to all the ratings of the electrical connecting terminal.
4.
Only use fuse according to the specifications and rated values stipulated by our product.
5.
Use the power cable supplied with this instrument.
6.
Don’t touch the surface of the lens and prism with hand or hard objects.
7.
Turn off the main power first before replacing the main bulb, flash lamp and fuse.
8.
To prevent the instrument from falling down to floor, it should be placed on the floor where the inclination angle is less than 10°.
9.
Turn off the power and cover the instrument with dust-prove hood when it is not in use.
10.
In case there is any trouble, please first refer to the trouble-shooting guide. If it still can’t work, please contact with the authorized distributor or our Repair Department.
THE SAFETY MARKS USED IN THIS INSTRUMENT
TYPE B ATTENTION PLEASE TERMINAL OF REFER TO THIS MANUAL THE PROTECTIVE GROUNDING
Fig.1
Contents
1 Nomenclature...........................................................................................................1
2 Assembly ..................................................................................................................3
2.1 Components.....................................................................................................3
2.2 Assembly procedure........................................................................................5
2.3 Checking procedure after assembling............................................................9
3 Operation procedures...........................................................................................10
3.1 Diopter compensation and Pupil Distance adjustment...................................10
3.2 Patient position and fixation target ..............................................................11
3.3 Base operation..............................................................................................11
3.4 Operation illumination unit..........................................................................12
3.5 Fundus observation with Hruby lens (optional)...........................................14
4 Maintenance.........................................................................................................15
4.1 Replacing the illumination bulb...................................................................15
4.2 Replacing the reflecting mirror....................................................................15
4.3 Replacing the fuse........................................................................................16
4.4 Replacing the chin-rest paper
.......................................................................16
4.5 Adjusting the tightness of the slit width knob..............................................16
4.6 Adjusting the inclination of the illumination part........................................16
4.7 Cleaning ........................................................................................................17
4.8 Protecting .....................................................................................................17
4.9 Consumables................................................................................................19
5 Trouble shooting guide .........................................................................................20
6 Responsibility.........................................................................................................21
7 Transportation and storage..................................................................................21
8 Optional accessories (purchase in addition) .......................................................22
8.1 Applanation tonometer...................................................................................19
9 Specifications .........................................................................................................22
1 Nomenclature
1 Joystick
Incline joystick to move the instrument slightly on the horizontal surface and rotate it to adjust the elevation of the microscope.
2 Base Locking Screw
The base will be locked when fastening this screw.
3 Rail Cover
Protect the rail surface.
4 Base
Support the microscope and the illumination arms with the joystick controlling its movement.
5 Work Table 6 Accessory Drawer
Store the focusing test rod and other accessories.
7 Brightness Control Switch
The brightness can be continuous adjusted. Avoid working continuously at high setting, as the service life of the bulb will be shortened.
8 Main Power Switch 9 Pilot Lamp 10 Microscope Arm Locking Knob
Lock the rotational movement of the microscope arm.
11 Angle Mark Ring
Marks on the angle mark ring of the illumination arm, which relates to the long mark of the microscope arm, represent the two arms’ ring relates to the short mark at one side of the operator, the right eyepiece may be
blocked,
and the side of the patient the left eyepiece.
12 Chin-rest Elevation Adjustment Knob
Rotate the knob to adjust the elevation of the chin-rest
13 Location Roller
When it is in the
middle,
it stands for included angle of 0o between the microscope
arm and the illumination arm. And the right or left side the included angle of 10o.
14 Microscope and Illumination Arm Couple Bolt
Fasten this bolt and the illumination arm and the microscope arm could rotate together
in couple state
. Loosen it and the illumination arm then can rotate
separately.
15 Hruby Lens Guide Plate
Also used as an assembly plate for the applanation tonometer.
16 Breath Shield 17 Chin-rest 18 Magnification Select Dial
Five different magnifications are provided.
19 Prism Box
Separate the prism box to adjust the interpupillar distance.
20 12.5X Eyepiece
Before using the slit lamp, adjust the proper diopter for each eyepiece to obtain a
angle.
When the ‘0’ on the
1
definite image.
21 Microscope Fixation Screw 22 Accessories Mount
Accept the model R-900 Goldmann applanation tonometer as well as other accessories
23 Horizontal Mark
When the horizontal center of the patient’s eye is in line with this mark, the elevation of the microscope controlled by joystick is also in its center position.
24 Forehead Belt 25 Diffusion Lens
Used for observing and photographing at a low magnification, and for enlarging
the illumination field.
26 Lamp Cap 27 Aperture Slit Height and Display Window 28 Filter Selection Lever
There are four filters for selection
29 Aperture and Slit Height Control Knob
Rotate this knob to adjust the spot and the slit height. Swing the knob horizontally to revolve the slit.
30 Fixation target
Two kinds of fixation targets are available. One is for diopter adjustment to help the patient to see the target clearly while the other is an illuminated fixed spot.
31 Reflecting Mirror
Both long and short reflecting mirrors are provided. The long mirror is routinely used for most examination procedures. The short mirror is used when the long mirror interferes with the observation pathway, such as during funduscopy.
32 Hruby Lens
Used for observation of fundus and the posterior segment of vitreous body.
33 Hruby Lens Holder 34 Centering Knob
Loosening the knob allows the illumination light to be moved from the center of the vision field for indirect retro-illumination. Fastening the knob brings the illumination light back to the center.
35 Slit Width Control Knob
The slit width is continuously adjustable within the range from 0 to 9mm. The marks on the left knob stand for the approximate value of the width.
36 Illumination Inclination Lever
Four 5o inclination stops are availableup to 20o.
37 Brightness Control Connector
2
2 Assembly
This section of the manual describes how to assemble parts should be taken out with great care from the packing case before assembling.
2.1 Components
ESL-7800
slit lamp. All
Fig2-1
3
Fig.2-2
Name Quantity
A Illumination Part 1 B Converging Binocular Tubes and base part 1 C Head-rest Part 1 D Breath Shield 1 E Worktable with Power Box 1 F Rail Cover 2 G Input Power Cable 1 H Hruby Lens 1* I Hruby Lens Guide Plate 1* J Spare Illumination Bulb 1 K Chin-rest Paper 1 L Focusing Test Rod 1 M Protection Cap 2 N Dust-proof Cover 1 O Fixation Target 1 P Spare Long Reflecting Mirror 1 Q Spare Short Reflecting Mirror 1 R Brush 1 S Spare Fuse 2 T Cross Screw Driver with Wood Handle 1 U Watch Screw Driver 1 V Spanner 1 W
Hexagonal wrench 1
(*Optionally available in some region.)
4