ElektroPhysik 80-174-0500, 80-174-0300, 80-175-0900, 80-175-0600 User manual

Page 1
Technical Manual /Operating Instructions
MiniTest 7200 FH / MiniTest 7400 FH Thickness Gauges
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
Contents
Contents
2. FIRST STEPS.................................................................................................................................6
NSERT BATTERIES AND CONNECT SENSOR.................................................................................6
2.1 I
WITCH ON AND TAKE READINGS..............................................................................................7
2.2 S
2.2.1 Setting language..................................................................................................................7
3. DESCRIPTION OF THE MEASURING SYSTEM...................................................................9
3.1 B
ASIC UNIT .................................................................................................................................9
3.1.1 General Remarks.................................................................................................................9
3.1.2 Operating keys ....................................................................................................................9
3.1.3 Sockets and interfaces.......................................................................................................10
3.1.4 Power supply.....................................................................................................................10
3.1.4.1 Batteries and storage batteries......................................................................................10
3.1.4.2 Power supply unit...........................................................................................................11
3.2 S
ENSORS FH 4 AND FH 10........................................................................................................11
3.2.1 General Remarks...............................................................................................................11
3.2.2 Sensor Models...................................................................................................................11
3.2.3 Sensor Stand......................................................................................................................12
HE USER MENU.........................................................................................................................13
4. T
4.1 G
ENERAL REMARKS
4.2 S
WITCH
-ON / S
EASURE SCREEN /STATISTICS SCREENS ..................................................................................13
4.3 M
....................................................................................................................13
TART SCREEN
......................................................................................................13
4.3.1 Measure Mode – Numerical Screen..................................................................................13
4.3.2 Measure Mode –Graphics Screen with MiniTest 7400 FH ..............................................14
4.3.3 Statistics Screens...............................................................................................................14
SER MENU..............................................................................................................................15
4.4 U
4.4.1 U
4.4.2 N
4.4.3 A
SER PREFERENCES...............................................................................................................15
UMERICAL ENTRIES .............................................................................................................16
LPHANUMERICAL ENTRIES (ON MINITEST 7400 FH)...........................................................16
5. MEASURE MODE......................................................................................................................17
MPORTANT NOTES ON THICKNESS MEASUREMENT..................................................................17
5.1 I
5.1.1 Limitations to magneto-static thickness measurement......................................................17
ATCH CONFIGURATION / NECESSARY SETTINGS.....................................................................17
5.2 B
5.2.1 Batch .................................................................................................................................17
5.2.2 Target Ball Size.................................................................................................................18
5.2.3 Data Logging Mode..........................................................................................................18
5.3 P
REPARING MEASUREMENT......................................................................................................18
5.3.1 Calibration........................................................................................................................18
5.3.2 Measurement without Sensor Stand..................................................................................19
5.3.3 Using the Sensor Stand.....................................................................................................19
AKE READINGS .......................................................................................................................19
5.4 T
5.4.1 Measurement without Sensor Stand..................................................................................19
5.4.2 Using the sensor stand......................................................................................................20
5.4.3 Delete Readings ................................................................................................................20
5.5 S
TORING READINGS INTO STATISTICS MEMORY .......................................................................20
5.5.1 Data Logging Mode..........................................................................................................20
5.5.2 Manual Data Storage........................................................................................................21
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2
Page 3
Contents
5.5.3 Automatic Data Storage (Auto).........................................................................................21
5.5.4 Storing Minimum Reading................................................................................................21
5.5.5 Storing Maximum Reading................................................................................................21
5.6.1 Correct Sensor Position....................................................................................................22
5.6.2. Incorrect Readings through tilting of the Sensor.............................................................22
5.6.3 Incorrect readings through Blocked Target Ball..............................................................23
5.6.4 Air gap...............................................................................................................................23
5.6.5 Measuring material with ferruginous particles................................................................23
6. CALIBRATION...........................................................................................................................24
ENERAL REMARKS ON CALIBRATION .....................................................................................24
6.1 G
ALIBRATION METHODS............................................................................................................25
6.2 C
OW TO CALIBRATE ..................................................................................................................26
6.3 H
6.3.1 Factory Calibration ..........................................................................................................26
6.3.2 Zero only...........................................................................................................................26
6.3.3 Multi-point calibration (Zero + 1 to 4 point(s) )..............................................................27
6.4. D
ELETE CALIBRATION PO INT ....................................................................................................28
ALIBRATION ERROR SOURCES ................................................................................................29
6.5 C
7. DATA STORAGE........................................................................................................................30
ATCHES...................................................................................................................................30
7.1 B
7.1.1 General..............................................................................................................................30
7.1.2 Batch Data ........................................................................................................................30
7.2 D
ATABASE OF THE MINITEST 7400 FH .....................................................................................31
7.2.1 General Remarks...............................................................................................................31
7.2.2 Tree structure....................................................................................................................31
7.2.3 Create a Directory............................................................................................................31
7.2.4 Create a New Batch ..........................................................................................................32
7.2.5 Select a Batch for Measurement .......................................................................................32
7.2.6 Delete a Batch / a Directory .............................................................................................33
7.2.7 Rename a Batch / Rename a Directory.............................................................................33
8. STATISTICS................................................................................................................................34
TATISTICAL PARAMETERS.......................................................................................................34
8.1 S
8.1.1 Upper and lower specifications limits (USL and LSL).....................................................34
8.1.2 Set Point (Differential Mode)............................................................................................35
8.2 R
EADINGS .................................................................................................................................35
8.2.1 View Readings...................................................................................................................35
8.2.2 Delete Single Readings .....................................................................................................35
8.2.3 Delete all Readings...........................................................................................................36
UMERICAL STATISTICS ...........................................................................................................36
8.3 N
ISTOGRAM ON MINITEST 7400 FH .........................................................................................37
8.4 H
8.5 T
REND DIAGRAM ON MINITEST 7400 FH.................................................................................37
9. DATA OUTPUT / DATA TRANSFER......................................................................................38
9.1 D
ATA PRINT-OUT
........................................................................................................................38
9.1.1 Screen Shot........................................................................................................................39
9.1.2 Print-out a Batch...............................................................................................................39
RANSFER DATA TO A PC.........................................................................................................40
9.2 T
10. MAIN MENU .............................................................................................................................41
10.1 D
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
ATABASE ..............................................................................................................................41
3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
Contents
TATISTICS..............................................................................................................................41
10.2 S
10.3 C
10.4 A
ALIBRATION..........................................................................................................................41
CTIVE BATCH .......................................................................................................................41
10.4.1 Batch Properties .............................................................................................................41
10.4.2 Readings..........................................................................................................................42
10.4.3 Configuration..................................................................................................................42
10.4.4 Calibration......................................................................................................................42
10.5 S
ETUP
.......................................................................................................................................43
10.5.1 General Remarks.............................................................................................................43
10.5.2 Display............................................................................................................................43
10.5.3 Date and Time.................................................................................................................43
10.5.4 Data Outputs...................................................................................................................44
10.5.5 User Preferences.............................................................................................................45
10.5.6 Gauge Specifications.......................................................................................................45
11. SPECIAL FUNCTIONS............................................................................................................46
11.1 I
NITIALIZING ...........................................................................................................................46
11.2 S
PECIAL FUNCTIONS
..................................................................................................................47
12. QUICK REFERENCE ..............................................................................................................48
13. CARE AND MAINTENANCE.................................................................................................49
13.1 C
ARE.......................................................................................................................................49
13.1.1 Using NiMH storage batteries........................................................................................49
13.2 M
AINTENANCE
.....................................................................................................................49
14. TECHNICAL SPECIFICATION.............................................................................................50
14.1 G
AUGE AND SENSORSPECIFICATION
14.2 D
ELIVERY SCHEDULE .............................................................................................................52
..........................................................................................50
14.2.1 Thickness gauge MiniTest 7200 FH / 7400 FH..............................................................52
14.2.2 Sensors............................................................................................................................52
14.3 A
CCESSORIES..........................................................................................................................53
15. APPENDIX.................................................................................................................................56
15.1 T
ROUBLESHOOTING
15.2 D
ETECTION OF ERROR SOURCES
TATISTICAL TERMS ...............................................................................................................60
15.3 S
AFETY NOTES........................................................................................................................62
15.4 S
15.5 D
ECLARATION OF CONFORMITY
15.6 A
FTER SALES SERVICE
...................................................................................................................56
(INF
FLAG
)
...............................................................................59
.................................................................................................64
..............................................................................................................65
INDEX...............................................................................................................................................66
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
4
Page 5
1. Introduction
Designed for non-destructive thickness measurement, the portable MiniTest 7200 / 7400 FH measuring systems are suitable for in the field or laborator y use. Working on the m agneto-static principle, the gauges enable quick and easy measurement on non-ferromagnetic mat er ials. The measuring system consists of a handy sensor, which is connected via a cable to the base and display unit for visualization and processing of readings. For m easurement , specially treated targ et balls made of a ferromag netic material are used as a r eference. For measurement , a target ball is placed on the material to be measured so that the material thickness between target ball and sensor will be measured. The measuring system comes with sensor st and (available as an option) to add additional comfort of handling during measurement. For documentation a portable data printer is available. The measuring system may also be connected to a PC for data communication. For taking reading s, the sensor is placed on the surf ace of t he measuring object, the target ball is placed on the opposite side of sample. The sensor is placed on the opposite side of sample. Equipped with a strong perm anent magnet, the sensor at tracts the target ball and holds it exactly over its sensor tip. W hen moving the sensor, the t arget follows accordingly. The presence of t he target ball will change the magnetic field near the sensor tip. The magnetic field changes according to the distance between target ball and sensor tip. T he chang e of magnetic f ield increases with the distance between target ball and sensor t ip getting smaller and can be taken as a ref er ence for the material thickness of the sample to be measured. The sensor incorporates a Hall element to capture the change of mag net ic field and to translate it into thickness.
For measuring hollow parts, a target ball is placed into the interior of sample and the sensor is placed on the opposite side in order to attract t he target ball exactly over the sensor tip. For taking readings, the sensor is moved over the surface of sample. For measuring sheets, the sensor is placed on the surface of sheet and the target ball is placed near the sensor on the opposite side of sheet. The target ball will automatically center on the correct measuring position and the measuring processes can be started.
For measuring large-sized parts, the sensor can be guided manually over the sample to be measured. For measuring small part s , it is recommended to use the sensor stand supplied with the sensor. Suitable for use in industrial production areas, the measuring systems MiniTest 7200 / 7400 FH measure hollow parts, containers and sheet material of all kinds such as bottles, cans, inject ions moldings, car bodies, glass, SMC components etc made of non-magnetic (non-ferr ite) materials.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5
Page 6
2. First Steps
This section refer s to persons to use the gauge for the first tim e. This section explains the main features of the g auge and how to take readings.
2.1 Insert Batteries and connect Sensor
a) Take gauge and batteries from the carrying case. b) Push the battery compartment lid at the gauge bottom in arrow direction (as shown below) c) Inser t bat teries into the battery compartment. Respect polar it ies ( as shown below). d) Close lid.
b) c)
e) Take requested sensor (FH4 or FH10) form t he carrying case. Plug sensor into the sock et on
top of the base unit (as shown below). Make sure the nib of plug is correctly inserted into the guide way of socket. The red mark ing at the probe plug must be positioned correctly to the red marking at the sensor plug .
f) Fully engage the sensor plug into the sensor sock et t o the stop.
Sensor Socket
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
6
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
2.2 Switch ON and Take Readings
2.2.1 Setting language
This setting can be made at initial use or as requested at any time (see section 11.1) Press the red ON/OFF button on the left side of gauge and ESC simultaneously to switch on and to go to initialization menu.
1. A banner appear s with a language option. User arrow keys ↑↓ to go to your requested lang uage option. Press OK to confirm or ESC to abor t.
2. On display appears “Total Reset”. A Total Reset restores factory settings. To abort Total Reset, press Function key “No” (section 11.1)
a) Start Screen appears showing company logo, gauge version
and sensor type that is currently connected (see illustr. on the left.
Flashing antenna symbol if sensor is
connected
b) After approx. 2 seconds, the gauge switches automatically into
measure mode and is ready to take readings. The Measure Screen will appear without any reading being shown.
c) At initial switch-on the gauge is in “Direct Mode“ (see section
7.1.1). For this mode, the factory calibration is preset (for more detailed information on “Calibr at ion” topics please refer to section
6). The Status Line provides inform ation on the active calibration method and the active batch (MiniTest 7400 FH) Please check whether the target ball size as adjusted mat ches your application (for more details on targ et ball size pleaser refer to sect ion 5.2.2 “Target ball size”).
Target ball diameter, hiere: 4.0 mm.
Selected calibration mode, here:
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
Zero calibration
www.GlobalTestSupply.com
Active batch, here: Pet bottle (only with Minitest 7400 FH)
Status line
7
Page 8
d) The factory calibration is for quick and easy measurement when larger measuring errors are
acceptable. For other calibration methods please refer to section 6.2.
e) For measurement, please use the sensor stand supplied with the gauge. Place the measur ing
object on the sensor. Place the targ et ball on the opposite side. For measuring hollow parts, place the ball inside the sample. Equipped with a strong permanent magnet, the sensor attracts the target ball and holds it exactly over its sensor tip. Hold t he measuring obj ect in rectangular position to the sensor and move it smoothly over the sensor. Press OK t o store the reading into memory. Press Function key “Statistics” to view readings and the numerical statistics.
For measuring large-sized sheeting material or large hollow parts, place the senor on the surface of sample. T he target ball is placed on the opposite side near the sensor. Keep the sensor in rectangular position to the surface and move it across the surface of sample.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
8
Page 9
3. Description of the measuring system
f
3.1 Basic unit
3.1.1 General Remarks
Graphics display 160 x 160 Dots
3.1.2 Operating keys
ON/OFF button
Command and navigation key block
Large backlit display for easy reading o measuring values, statistical data, histog ram and trend diagram.
Robust scratch resistant plastics housing.
Function keys
Use the ON/OFF button to switch the gauge ON or OFF. If you press ON/OFF button and ESC simultaneously, the initializing procedure will be perform ed (for more details please refer to section
11.1). The Function keys may assume different functions according to the menu being active. Their current properties are displayed in the line upper to the Function keys.
The command and navigation key block may also assume different functions:
-
Press OK to confirm setting s, st or e values or select m enu it ems.
-
Press ESC to abort actions or to quit subm enus.
-
Use ARROW k eys to navig at e t hrough the menu or change settings.
-
If the alphanumerical block has been activated, O K and ESC keys also assume navigation
functions.
The Function keys and the command and navigation k ey block may be backlit on request. This adds additional comfort in poor light conditions (for more details please refer to section 10.5.2)
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
Press the “Help” Function key for calling the context-or iented on-line “Help“.
3.1.3 Sockets and interfaces
Multi-purpose socket
The basic unit features a sensor sock et for connecting the FH 4 or the FH 10 sensor. The multi­purpose socket is used for connecting the mains unit, the foot switch, the alarm device or the RS232 interface cable.
3.1.4 Power supply
3.1.4.1 Batteries and storage batteries
Sensor socket
IR Interface
Both models, MiniTest 7400 FH and MiniTest 7200 FH, are powered by a set of four alkaline­manganese cells, 1.5V, AA LR6 size (Batteries included in the standar d supply schedule.) .
As an alternative, both models may be operated on rechargeable NiMH (type AA-HR6) storage batteries or via a power supply unit (available as an option). Please use only products as recommended by ElektroPhysik (See section 14.3, Accessories) .
If you wish to use the storage batteries, they must be recharge using the external charger unit (available as an option). Please also refer to sect ion 13.1 “Care and maintenance” for more details.
Note:
Remove batteries or storage batteries from the instrument if you are not going to use it for a longer period of time.
The battery symbol on display indicates the battery state.
For field use, replacement batter ies should be m ade available.
Erratic readings due to low battery voltage do not occur because the gauge switches off or
does not switch on if the battery is too low.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
10
Page 11
Used or defective batteries or storage bat t er ies m ay contain hazardous substances and must be disposed according to the legal provisions of your count r y.
3.1.4.2 Power supply unit
For mains operation, the power supply unit must be used. If operated via the power supply unit, batteries should be inserted to supply the internal time clock , otherwise, after approx. 1 minute, the settings of the real-t im e clock will go lost as soon as supply from the mains is cut.
The power supply unit comes with two different adapter plug versions ( Eur o and US plug). If the power supply unit comes with an adapter plug connected that does not match your socket, simply change the adapter plug accordingly.
Simply remove the adapter from the power supply unit and f ix the ot her one as r equired.
Note:
The adapter has not been designed for frequent change.
3.2 Sensors FH 4 and FH 10
3.2.1 General Remarks
In this new SIDSP (Sensor integrated digital signal processing) procedure, all necessary measuring signals are created and complet ely processed in the sensor itself. Only the completely processed digital readings are transferred to the base unit for display, statistical evaluation and data storage. Unlike the comm only used analog procedures, the new SIDSP procedure excludes any error influences on the measuring data during transfer over the pr obe cable. The result is a measuring accuracy and constancy of readings that has been unmatched so far. The complete sensor technique is integrated into a r obust st ainless- steel housing.
3.2.2 Sensor Models
Two sensor models are available:
-
FH 4 sensor to cover the 0 …4 mm range
-
FH 10 sensor to cover the 0…10 mm range.
Both sensors feature a wear-resistant sensor tip most suitable also for hard materials such as glass, for instance.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
11
Page 12
3.2.3 Sensor Stand
Both sensor types come with a spring-loaded V-grooved sensor stand. The use of the sensor stand ensures increased reproducibility of readings by adding stability of the sensor during measurement so that even small parts can be measured without problems.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
12
Page 13
4. The User Menu
r
A
t
E
4.1 General Remarks
4.2 Switch-ON / Start Screen
At switch on, the company logo, gauge version and type of sensor being connected appear on display.
After approx. 2 seconds the gauge switches to the m easur e screen of the last batch that was active before the gauge was switched off.
4.3 Measure screen /Statistics screens
Readings can be displayed numerically only or numerically combined with a diagram (only with MiniTest 7400 FH)
4.3.1 Measure Mode – Numerical Screen
Power supply indicato
Number of readings
djusted set poin
CALIBRATION MOD
The current function key proper-ties are indicated in the line above the function keys.
Time clock
Current reading Target ball size and measuring unit
Minimum reading Maximum reading
Active batch (MiniTest 7400FH only)
Status line
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
13
Page 14
4.3.2 Measure Mode –Graphics Screen w i t h M iniTest 7400 FH
r
H
T
t
r
Power supply indicato
Reading
Set poin
REAL-TIME TREND
CALIBRATION METHOD
The current function key proper-ties are indicated in the line above the function keys.
Time clock
Target ball size, measuring unit
DATA LOGGING MODE
UPPER LIMI
MAX = MAXIMUM READING N = NUMBER OF READINGS
MINIMUM READING
LOWER LIMIT
NAME OF ACTIVE BATC
STATUS LINE
For setting the Real-Time Trend display option, select “ Active Batch” form the Main Menu. Press OK to confirm. Select “Configuration” from the Batch Menu and press OK to confirm. Use the arrow keys to move to the “Graphics” option. O N or OFF is highlighted. Use arrow keys to make your selection and press OK to confirm or t he “ ESC” navigation button to restore previous setting and to go back to the previous menu level.
To take full advantage of the real-time trend it is necessary to toler ance lim its (see also section
8.11 „Upper specification limits (USL) and lower specification limits (LSL)“.
4.3.3 Statistics Screens
,
Power supply indicato
Process capability,
cp and cpk
Selected calibration mode
The current function key proper-ties are indicated in the line above the function keys.
Time clock
Active batch (MiniTest 7400 FH only)
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
14
Page 15
4.4 User Menu
The numerous MiniTest FH 7400 / 7200 features can be accessed via the dif ferent menus that are divided into submenus.
Press the Function key “Main menu“ in order to get access to the main menu.
Select an item from the Main Menu (e.g. Database) and press OK to confirm. You will go to a submenu. Use the arrow keys to select the request ed it em and press OK to confirm. To go back to the previous menu level press ESC.
According selected submenu you can
-
View settings (e.g. Gauge specs = Gauge specifications in the Setup Menu)
-
Select preset parameters (e.g . 100% br ightness from the Display Menu)
-
Make numerical or alphanumerical entries (e.g. Ent er a directory name in the database / new
directory option).
4.4.1 User Preferences
Use the arrow keys to select your requested menu item, e.g. Language. Press OK to confirm. T he current selection will be highlighted. Use the arrows to select the requested parameter ( e. g. English) and press OK to confirm. To undo your selection, pr ess ESC and you will return t o t he previous menu level.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
15
Page 16
4.4.2 Numerical Entries
As soon as the alphanumeric block is activated, ESC and OK will assume navigation functions to the lef t or right. Choose your character by placing the cursor over the requested character. After a certain period of time, the entry will be accepted automatically without the need to confirm. This period of tim e can be specified under User Pref erences in the “Entry accepted” option. See also section 10.5.5 for further reference. Complete your entry by pressing Function key “Ready” and then confirm “Save changes?” by pressing OK.
4.4.3 Alphanumerical Entries (on MiniTest 7400 FH)
For alphanumerical entries you can use as many as 15 characters for entering a name. ESC and OK assume navigation f unctions to the left or right. Choose your character via the alphanumerical block. Your selection will be automatically accepted after the previously specified period of time. This period of time can be set under User Pref erences in the “Entry accepted” option. See also section 10.5.5 for further reference. Complete your entry by pressing Function key “Ready”. Complete your entry by pressing Function key “Ready” and then confirm “Save changes?” by pressing OK.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
16
Page 17
5. Measure Mode
5.1 Important Notes on Thickness Measurement
Make sure that the operator has been properly instructed r egarding the use of thickness gauges and has basic knowledge of the specific requirements for measurement of the application. The operator should have basic knowledge of the following:
Selection of a measuring device suitable f or his applicat ion
Fundamentals on the magneto-static m easuring principle
Influences on the magnetic f ields t hr ough the surrounding field
Influence of the surf ace pr oper ties of the material to be tested (roug hness, build- up on t he
surface)
Statistical evaluation of measuring ser ies
Gravity influencing the ref e r ence ball
5.1.1 Limitations to magneto-stati c t hi ckness measurement
The information obtained f rom thickness measurements according to the magneto-static principle only refer to t hose parts of the t est object that have been covered by the magnetic sensor. For that reason, be careful to draw conclusion on other parts of the measuring object that have not been covered by the sensor during measurement. I n general, such conclusions are only admissible if comprehensive experience and approved methods of statistical data acquisition are available.
5.2 Batch Configuration / Necessary Settings
Before taking reading s, it is necessary to make a few settings in the Configuration menu. The Configuration always refers to a certain measuring series (Batch).
5.2.1 Batch
-
You can continue to take readings in the active batch
-
You can create a new batch (only with MiniTest 7400 FH model)
-
You can choose an exiting batch from the database (please refer to section 7.2).
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
17
Page 18
5.2.2 Target Ball Size
According to sensor type and the shape of measuring object, a suitable target ball must be set. This can be m ade in the Configuration menu, menu item “Targ e t ball” (also refer to section 6 “Calibration”. Use arrow keys to move to “Target Ball” size and press OK. Use arrow keys to adjust the requested t arg et ball size. Press O K to confir m your setting or press ESC to undo your setting. If you undo, you will automatically quit target ball setting option and go back to the previous menu level.
5.2.3 Data Logging Mode
In the data logg ing mode you can specif y how you wish readings to be stored in memory. Several modes are available:
-
manually or via footswitch (Current reading, MIN, MAX)
-
Automatic (Auto)
Note: For logging data into memory please always proceed as follows: Put the sensor into the correct measuring position, press ESC. Then readings can be taken.
5.3 Preparing Measurement
Before measurement, the probe must be adapted to the ambient temperature. In case of high variations in temperature, the targ et ball must be lifted f rom the probe tip at regular intervals (e.g. every other minute). This is to compensate f or f r equency drif t due to variations in temperature. It is recommended to switch on the gauge for warming up around 2 to 3 minutes in order to ensure maximum measuring accuracy.
5.3.1 Calibration
According to your setting of task , you may use different calibration methods. Measuring accuracy depends on the selected calibration method. Please refer to section 6 for more details on this issue. There are three calibration methods available:
-
Factory
-
Zero only
-
Multi-point calibration (Zero + 1 to 4 Point(s)
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
18
Page 19
5.3.2 Measurement wit hout Sensor St and
For measuring large-sized parts, the sensor can be guided manually over the measuring sample. For measuring thick samples (6 m m and thicker) make sure to k eep the sensor in vertical position to the sample. When holding the sensor in horizontal position, due to the influence of gravity on the steel ball, the readings will be higher than the actual thickness.
5.3.3 Using the Sensor Stand
For measuring small part s, bottles or other cylindrical bodies the sensor stand (supplied with the sensor) and a footswitch (available as an option) should be used.
Put the sensor into the stand and secure with the red sensor stand retaining ring.
5.4 Take Readings
5.4.1 Measurement wit hout Sensor St and
For measurements on large-sized sheeting m aterial or large hollow parts, place the sensor onto the surface of the object to be measured. Place an appropriate tar get ball on the opposite side of object near the sensor. The t arget ball will be attr acted by the strong permanent m agnet of sensor and be positioned centrally just opposite of the sensor t ip. Keep the sensor in right-ang le position to the object to be measur ed and scan continuously over the surface. For storing readings, please refer to section 5.5.1.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
19
Page 20
5.4.2 Using the sensor stand
If you are using the sensor stand, position the sample on the sensor tip and keep it in right-angle position to the sensor. Place the appropriate t arget ball on the opposite side of obj ect. Make sure to keep the right -angle position while moving the object continuously over the sensor. For st oring readings, please refer to sect ion 5. 5. 1. When measuring small hollow parts such as bottles e.g., put the target ball inside the object and place the object onto the v-grooved device of sensor stand. Turn the object unt il the target ball is centered directly over the sensor tip. Then move the object continuously over the v-groove of sensor stand. For storing readings, please r efer to section 5.5.1.
5.4.3 Delete Readings
Please refer to section 8.2.2 “Delet e Single Readings“.
5.5 Storing Readings into Statistics Memory
5.5.1 Data Logging Mode
During measurement, readings can be stored into memory manually or automatically. Data in memory will be used for statistics.
The following storing options ar e available:
-
Storing the current reading manually or automatically
-
Storing the minimum reading (only manually)
-
Storing the maximum reading ( only manually).
The Data logging mode can be specif ied in t he “Conf ig uration” submenu. Use arrow keys to select the Data logging mode option and pr ess OK to conf irm. You can chose f rom t he following setting options:
-
Curr. reading (Current r eading)
-
Auto
-
Max
-
Min
Make your selection and press OK to confirm. If you wish to undo your setting (before OK has been pressed) press ESC and the gauge will return to t he pr evious menu level.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
20
Page 21
5.5.2 Manual Data Storage
During the measuring procedure, current readings can be stored manually into the active batch by pressing OK or by actuating a foot switch. For this option set Data logging mode to “ Cur r . r eading”.
5.5.3 Automatic Data Storage (Auto) Auto option
In the auto mode, readings will be stored automatically during the measuring procedure as soon as you press OK. To st op the automatic data storage pr ess OK again. If a footswitch is connected, readings are stored as long as the f ootswitch is actuated. Before starting the automatic data storage make sure the sensor is placed in the correct measuring position. Make sure to st op the automatic data storage BEFORE the sensor has left the measuring position. This is to avoid err atic readings to be stored into memory that m ight occur when putting the sensor to or lifting it from the sample.
After setting the Data logging option to the Auto option, a Logging rate option will appear. The logging rate defines t he time intervals at which data will be stored into memory (1, 2, 5, 10 or 20 readings per second). For slow measurement scans, a low logging rate should be set, for quick scanning, the logging rat e should be set to a higher rate.
5.5.4 Storing Minimum Reading
In the Minimum option, minimum readings will be stored to t he active batch by pressing OK during measurement or by activating a foot switch. To set Data logging mode t o the MIN option, go to the Configuration submenu and select Data logging mode. Use arrow keys to select MIN and press O K to confirm. Please note: The minimum reading can also be stored even after the sensor has been lifted from the sample.
5.5.5 Storing Maximum Reading
In the Maximum option, maximum readings will be stored to the active batch by pressing OK during measurement or by activating a footswitch. To set Data logging mode t o the MAX option, go to the Configuration submenu and select Data logging mode. Use arrow keys to select MAX and press OK t o confirm.
Important note:
-
Make sure to put the sensor into the correct measuring position and to press ESC before
storing the Maximum reading into memory.
-
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
OK must be pressed BEFORE the sensor is moved from its measuring position.
21
www.GlobalTestSupply.com
Page 22
5.6 Error Sources during Measurement
After calibrating the gauge according to your application and appropriate target ball size, the gauge is ready for measurement. The readings will be correct as long as you measure within the range as specif ied for this target ball. The target ball is placed on one side of the sample to be measured and will be attracted by the sensor tip on the other side of the sample. The tar get ball serves as a reference to the sensor. What is being m easur ed is t he distance between sensor and target ball. Make sure to place both, the sensor tip and the targ et ball correctly onto the sample. Both must be in good contact with the surface of sam ple. Otherwise erratic readings (higher readings) might occur. Make sure to keep the sensor in perpendicular position while moving it over the sample. The target ball on the other side must be freely movable.
5.6.1 Correct Sensor Position
The picture illustrates the correct posit ion of sensor tip and steel ball. Make sure to keep the sensor tip perpendicular t o t he surface of sample.
5.6.2. Incorrect Readings through tilt i ng of t he Sensor
Error source: The sensor tip is not positioned corr ectly in perpendicular position to the sample. The reading will therefore be higher t han the actual thickness.
Remedy: Make sure to keep the sensor in perpendicular position to t he sample. If necessary, use the sensor stand.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
22
Page 23
5.6.3 Incorrect readings through Blocked Target Ball
Error source: The target ball has got stuck through a bulge, e.g. inside the
sample. It is out off t he correct measuring position. Remedy: Move the sample in order to release the target ball.
5.6.4 Air gap
Error source: Due to the shape of sample, an air gap is between target ball and measuring object. The reading will be higher than the actual thickness.
Remedy: If the tar get ball is too large, use a smaller one. If the sensor is not in good contact, change measuring position and use a smaller target ball, if necessary.
5.6.5 Measuring material with ferruginous parti cl es
Ferruginous particles in the material to be measur ed m ay lead to er r at ic readings because the ferruginous particles may inf luence t he m easuring signal. This will lead to readings smaller than the actual thickness.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
23
Page 24
6. Calibration
Calibration is made in the batch being active and will always refer to this batch, i. e. each batch has its own calibration. Before calibration, an appr opr iat e target ball diameter must be set in the Calibration menu. Please refer to section 5.2.2 “Targ et ball size”.
There are two options to go to the Calibrat ion menu:
-
from the Main menu choose Calibration
-
from the Measure Mode menu press Calibration function key.
For selecting an appropriate targ et ball, the following aspects should be taken into considerat ion:
-
minimum radius of sample to be measured
-
maximum thickness to be measured
-
required measuring accuracy
Sensor Target ball to
be used
Min. Internal radius of sample
Measuring range
Recommended measuring range
FH 4 Ø 1.5 mm 0.75 mm 0 ... 1.5 mm 0... 1.5 mm FH 4 / FH 10 Ø 2.5 mm 1.25 mm 0 ... 2.5 mm 1... 2.5 mm FH 4 / FH 10 Ø 4.0 mm 2.0 mm 0 ... 4.0 mm 2... 4.0 mm FH 10 Ø 6.0 mm 3.0 mm 0 ... 10 mm 3... 6 mm
FH 10 Ø 9.0 mm 4.5 mm 0 ... 10 mm 6... 10 mm
6.1 General Remarks on Calibration
For achieving maximum accuracy, please observe the following calibration instruct ions:
Make sure to keep off str ong magnetic fields from t he sensor dur ing measurement.
Keep away from any ferromagnetic metal par ts in a safety distance of at least 30 cm.
It is recommended to switch on the gauge for warming up around 2 to 3 minutes before
starting calibration procedure.
Make sure the sensor tip, target balls and calibration standards are clean. Any foreign matter may lead to erratic readings.
For maximum accuracy of calibration and later measurements, choose the thickness of calibration standard within the same thickness range as the later measuring sample.
In some cases it may become necessary to recalibrate if a new target ball is used (also if it has the same diameter as the previously used one). To be sure, check calibration in
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
24
Page 25
measuring mode by putting the matc hing precision standard with the appropriate target ball on the sensor.
Please note that large targ et balls will compress soft material more than small tar get balls.
Please note:
If the gaug e switches off during the calibration procedure due to low battery voltage, calibration must be repeated after installing fresh batteries.
Please note:
Each time the target ball is removed f r om the sensor tip in a distance of more than 30 mm, t he gauge will automatically be adjusted to the infinite range. Any possible drift influences ( e . g. through variations in temperature) will be extensively compensated. The m or e often the target ball is removed form the sample, the more often compensation will take place so as to increase measuring accuracy.
For that reason it is recommended to lift the target ball fr om t he sensor tip at least every three minutes. A message will appear on display accordingly to remind you to repeat infinite setting (“Refresh infinite Value !!!).
To ensure measuring accuracy, check calibrat ion after one hour of use of the g auge and recalibrate if necessary.
6.2 Calibration methods
According to your setting of task, you may use different calibration methods. Measuring accur acy depends on the selected calibration method. The following calibration methods are available:
-
Factory calibration. This calibration method is suitable for quick and simple measurement and if larger measuring errors are admissible as compar ed t o zero calibr at ion.
-
Zero only (zero point calibration). This method is for quick calibration with a medium measuring accuracy.
-
Multi -point calibration (Zero + 1 to 4) Calibration standard(s). In most cases a Two-point calibration (Zero + one standard) will be sufficiently good as far as calibration is made in the rang e of thickness to be expected. Further calibration points are recommended if measurem ent will be over a larger thickness range and if accuracy requirements are high.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
25
Page 26
6.3 How to calibrate
Using the sensor stand will add additional comfort during the calibration procedure. For more details please refer to section 5.3.3.
For all calibration methods the following applies:
-
To call the Calibration menu from Measure mode m enu, pr ess t he Calibration function key.
-
Use the arrow keys to go to “Targ et Ball”. Press OK to confirm. Use arrow keys to adjust to
the requested target ball size. Pr ess O K t o confirm.
-
Use arrow key to go to “Calibration Mode”. Press OK to confirm and use arrow keys to
make your selection. Press OK to confirm.
-
6.3.1 Factory Calibration
Use arrow keys to select “Factory“. Press OK to confirm. The gauge is ready for measurement.
6.3.2 Zero only
Use arrow keys to select “Zero only“. Press OK to conf ir m .
For zero point calibration, the zero standards supplied with the gauge must be used together with the matching target balls. Make sure always first to place the target ball into the zero standard. Then put both together onto the sensor. Wait for the sound to bleep, remove the precision standard and keep it in a distance of at least 30 mm. For mean calculation it is recommended to repeat the procedure several times. (To g et reasonable values it is always recommended to take several readings and to have the mean calculated). Pr ess OK to confirm. A ticked box right to “Zero“ will confirm successful calibration. Now the gauge is ready for measurement.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
26
Page 27
Zero calibration standard wit h t arget bal l put on it
6.3.3 Multi-point calibrat ion (Zero + 1 to 4 point(s) )
For multi-point calibration you must f irst carry out a zero point calibrat ion. The precision standards must be used (1 to 4 as requested) to calibrate for the further calibration points. The precision standards have different thick ness values.
1. In most cases a Two-point calibration ( Zero + one standard) will be sufficiently good as far as calibration is made in the range of t hickness to be expected.
2. Further calibration points are only necessary if measurement will be made over a larger thickness range and if accuracy req uir em ents are high.
Now use arrow keys to select “Zero + 1 point” calibration. Pr ess OK to confirm. Carr y out zero calibration as described under 6. 3.2. Then use arrow keys to select “Calibr. Point 1”, press OK to confirm.
Please always use one of the precision standards supplied with the sensor. Make sure first to place the target ball on the precision standard. Then place t he precision standard tog ether with the target ball onto the sensor. W ait for the signal tone to sound and lift the precision standard more than 30 mm from the sensor. For calculation of average repeat several times.
In the “CAL = “ field, the calibration mean value will be shown which has been calculated form the set of single readings you have taken. If this value is the same as specified on the precision standard, press OK to complete the calibration procedure. If not, please use arrow keys to correct the value as requested.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
27
Page 28
In the alphanumerical input field, you can enter the t hickness of the precision standard along with the unit. ESC and OK assume navigation functions to the left or right. You can select the requested character via the numerical block. Your selection will be automatically accepted after the specif ied period of t ime. This period of time can be specified in the Setup menu. Chose User Pref er ences and select “Entry accepted” (see also section 10.5.5). Complete your entry by pressing Function key “Ready”. To store your setting, press OK. Then press OK again to quit .
For more calibration points repeat procedure as necessary. The current state of calibration is indicated in the calibration check box in the calibration menu. Ticked boxes will mark successful calibration points.
After successful calibration of the last calibration point, the gauge is r eady for measurement.
6.4. Delete calibration point
Use arrow keys to move to the calibration point to be deleted. Press “Delete” Function key and press OK to conf ir m or ESC to abort. Pressing “Delete” Function key a second time will remove the (X) from t he check box. If you press again “Delete” Function key, the calibration value will also be deleted.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
28
Page 29
6.5 Calibration Error Sources
The following situations may lead to errat ic calibr at ion:
Error: Remedy:
Wrong target ball: Select an appropriate target ball suitable for
your measuring sample and the thickness to be expected. Adjust to the correct target ball size via the menu.
Calibration range does not match to t he thickness to be expected:
During calibration, the precision standard has not been placed correctly on the sensor Tipp.
Calibration error due to changes in temper at ure Before calibr at ion, t he sensor m ust be adapted
Worn or dam aged precision standards: Please use new precision standards in
Build-up on the sensor tip. Clean sensor tip from metal parts, dirt particles
Calibrate within the thickness range to be expected.
Make sure to fully engage the precision standard on the sensor tip.
to the ambient temperature. Please wait 30 minutes after switch on before starting measurement.
impeccable condition.
etc. by using a soft cloth.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
29
Page 30
7. Data Storage
7.1 Batches
7.1.1 General
Readings and their statistics will be stored in batches. Batches can be named alphanumerically and stored in the database directly or under a certain direct or y. O ne bat ch is firmly installed under the name “Direct Mode”. This batch cannot be deleted nor can it be renamed. The model MiniTest 7200 FH does not include the database feat ure, but it offers one measuring batch.
7.1.2 Batch Data
7.1.2.1 Readings
One batch may include as many as 100,000 single readings plus statistics. The model MiniTest 7400 FH offers a maximum of 240,000 values in total.
7.1.2.2 Calibration Values
Calibration always refers to a certain measuring batch. As soon as a measuring batch is called, the corresponding calibration will become active. The calibration includes the following parameter settings: diameter of t ar get ball, type of calibration and sensor.
7.1.2.3 Other Parameters
predefined
The parameters as set in the Conf iguration submenu such as Data logging mode, Logging rate, Graphics (ON/OFF), Upper spec. lim it, Lower spec. limit, and Setpoint refer t o a cer tain batch and will become valid as soon as this batch is called.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
30
Page 31
7.2 Database of the MiniTest 7400 FH
7.2.1 General Remarks
The database includes numerous features for data management. Measuring batches can be stored in directories and subdirectories. In total, as many as 500 directories may be created in a maximum of 5 levels. Directories and batches can be named alphanumer ically.
7.2.2 Tree structure
Example for a tree-structured database divided in directories and batches. At first use of the MiniTest 7400 FH, the database only includes one batch named “Direct Mode”. You can create other directories or batches as required.
Note: For creating new directories or batches, make sure to place
7.2.3 Create a Directory
the cursor below
Press function key “ Main Menu”. Then select “Database” and press OK to confirm. Use arrow down key of the navigation block to place the cursor below “Direct mode” or, if other directories have already been created, to the request ed position where a new directory should be created. This may be below or between directories or batches. Now press function key “Actions” and press OK to confirm “New Directory” which is highlighted.
Enter a name in the alphanumerical input field. For more details, please refer to section 4.4.3 “Alphanum er ical Entries”.
this Direct Mode directory.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
31
Page 32
7.2.4 Create a New Batch
From Measure Menu pr ess function key “Main Menu”. Then select “Database” and press OK. Use arrow down key of the navigation block to place the cursor below “Direct mode” or, if other directories have already been created, to the req uested position where a new directory should be created. This may be below or between directories or batches. Now press function key “Actions” and move cursor to “New Batch” which is highlighted. Press OK to confirm. Enter a name in the alphanumerical input field. For more details, please refer to section 4.4.3 “Alphanum er ical Entries”.
After you have entered the bat ch name, the alphanumerical screen appears. Now you can enter a user name below the batch name according to the procedure as described above.
After completion of user name input, the confirmation as shown on the left appears. If you wish to accept the parameters (calibration and configuration) as set f or t he active batch for t he new batch to be created press OK. If you press ESC, preset parameters of the factory calibration will become valid. Parameter settings can be made in the “Configuration” submenu. Please refer to section 10.4.3 for more details.
7.2.5 Select a Batch for Measurement
From Main Menu select “Database”. Use arrow keys to select the requested batch. Press OK to confirm. The data stored under this batch appear on display. Press OK to switch to Measure Menu.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
32
Page 33
7.2.6 Delete a Batch / a Directory
From Measure Menu press function key “Main Menu“ and select „Database“. Press OK to confirm . Use arrow keys to select the directory or batch to be deleted. Press function key “Actions”. Select “Delete” from the “Actions” submenu. A confirmation appears. Press OK t o confirm or ESC to abort.
If you have chosen a directory, the delete action will only delete empty directories. If the directory includes subdirectories and batches, these must be deleted previously.
Please note: The batch “Direct Mode” is a permanent part of the database. It can neither be deleted nor r enam ed. Nor you can assign a username to it. However, you can delete readings or statistical values.
Once you have deleted data they cannot be restored.
7.2.7 Rename a Batch / Rename a Directory
From Measure Menu press function key “Main Menu“ and select “Database“. Press OK to confirm . Use arrow keys to select the directory or batch to be renamed. Your selection will be highlighted. Press function key “Actions”.
Use arrow keys to go to “Rename”. Press OK t o confirm.
On the alphanumerical input field the name t o be changed appears. This name can be changed (see section 4.4.3 “Alphanumerical Entries”.) This action will not change the username.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
33
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 34
8. Statistics
From Measure Menu press function k ey “Statist ics“ . From Main Menu use arrow keys to select “Statistics” and press OK to confirm .
8.1 Statistical Parameters
8.1.1 Upper and lower specifi cat i ons l imits (USL and LSL)
Tolerance limits can be set as requested and offer you the following features:
-
Readings beyond the specificat ion limits will be indicated by a signal tone and marked in the statistics.
-
The process capability indices Cp and Cpk will be calculated according to the set specification limits.
-
In the trend diagram and histog ram, readings lying within the set specification limits will be highlighted. This range will determ ine the real- time tr end (only available with MiniTest 7400 FH).
The specification limits can be set in the “ Configuration” submenu. . Use arrow keys to go to “Upper spec. limit ” and/or “Lower spec. limit”
Specify limits and measur ing system (metric/imperial) as requested via the alphanumerical input field. See also section 4.4.2 “Alphanumerical Entries”.
Complete your setting by pressing Function k ey “Ready”. The gaug e will check automatically whether your spec. limits are reasonable. This is to prevent input er rors such as entering a wrong measuring unit. Press OK to confirm your chang es. Spec. limits can be deleted via the function key “Delete” .
Spec. limits can be set before or after readings have been taken.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
34
Page 35
8.1.2 Set Point (Differential Mode)
In quality control, it may be interesting to measure the difference between a set point and the actual thickness. Once you have specified a set point, the dif ference between this set point and the actual thickness will be displayed. A set point must be entered BEFO RE stor ing t he f irst reading . I f you change a preset set point, the stored reading s must be deleted previously. The set point can be set in the “Conf iguration” submenu. Use arrow keys to g o to “Set point” and enter a value as requested along with the correct measuring system (metric/imperial) via the alphanumerical input field. See also section 4.4.2 “ Alphanum er ical Entries”. Complete your setting by pressing Function key “Ready”. Set point set tings can be deleted via the function key “Delete”.
8.2 Readings
8.2.1 View Readings
From Main Menu go to “Statistics”. Press OK and select “Readings” subm enu. From Measure Menu press function key “Statistics” . Then press function key “Batch” followed by function key “Readings”.
Use arrow keys to scroll through the batch as requested. Readings will be indicated along with their current reference number. If spec. limits have been set previously, information is included whether the readings have been above (>>) or below (<<) the spec. limit.
8.2.2 Delete Single Readings
Erratic readings may be deleted from t he list . Move the cursor to the reading to be deleted.
Press function key “Help/Extra”. Move cursor to “Delete Readings” and press OK to confirm or press ESC to abort.
Readings that have been deleted are marked by (X) .
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
35
Page 36
8.2.3 Delete all Readings
From Main Menu chose “Statistics“. Use arrow keys to move to “Delete all readings“. Press OK to confirm.
From Measure Menu press function key “Statistics” followed by function key “Help/Extras”. Use arrow keys to move to “Delete all readings“. Press OK to confirm.
To abort, press ESC.
The delete action will delete all readings of the active batch. Batch name, sensor config uration and calibration f or this batch will remain in memory.
8.3 Numerical Statistics
From Main Menu chose “Statistics“. Use arrow keys to move to “Numerical Statistics”. Press OK to confirm. From Measure Menu press function k ey “Statist ics” .
Number of readings Average ( Standard deviation (s) Minimum Maximum Variation coefficient Process capability indices Cp and Cpk (if spec. limits have been set).
For more detailed information on statistics please refer to section
15.3.
x
)
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
36
Page 37
8.4 Histogram on MiniTest 7400 FH
From Main Menu chose “Statistics“. Use arrow keys to move to “Histogram“. Press OK to confirm. From Measure Menu press function k ey “Statistics” followed by function key “Histogram”.
A histogram is a kind of plot that summarizes how data are distributed. It gives you information on how the production process might be changed in order to be optim ized.
The limit rang e is the gray-highlighted area. See also 8.1.1 “Upper and lower spec. limits”.
Please note: For a histogram, a minimum number of 5 readings are required.
8.5 Trend Diagram on MiniTest 7400 FH
From Main Menu chose “Statistics“. Use arrow keys to move to “Trend Diagram“. Press OK to confirm. From Measure Menu press function key “Statistics” followed by function key “Histogram”. Then press function key “Trend”.
A trend illustrates the tendency of reading s so that deviations can be quickly identified.
The limit range is the gray-highlighted-area.
For a trend-diagram, at least 5 readings are required.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
37
Page 38
9. Data Output / Data Transfer
Both models, MiniTest 7200 FH and MiniTest 7400 FH are equipped with a bi-directional RS232 interface and an infrared interface ( I rDA 1.9). Readings and statistics of a batch can be tr ansferred to a PC or to the MiniPrint 7000 data printer (available as an option).
9.1 Data Print-Out
Data transfer of readings and statistics to the dat a printer MiniPrint 7000 is made via the infrared interface (IrDA 1.0). Before printing out, go to the Main Menu and select “Setup”. Use arrow keys to move to “Data Output” and press OK to conf irm. Use arrow keys to select “Infrared (IrDA)”. For more details, please refer to sect ion 10.5.4.2.
Readings and statistics may be printed out from all statistical views such as statistics, batch, histogram or trend (MiniTest 7400 FH). Press “Help/Extra” Function key from your select ed statistical view. The Help/Extra window opens (Fig. 1)
Fig. 1
Fig. 2
Select “Print” form t he list and confirm by “OK”.
A “Printing sequence” menu opens (Fig. 2) Move the cursor to the first item on the list and press OK. Use arrow keys to make your selection (Fig. 3). The following options are available:
-
Logo (Company logo or other customized logo in graphics format (160x80 Pixel). For inserting such logo, MSoft 7000 software is required.
-
Name (of batch) / or name ( of batch) + date (MiniTest 7400 FH).
-
Operator (with MiniTest 7200 FH Notes-Field only)
-
Statistics (number of readings, Average (x) Standard deviation (s), Min., Max. Variation Coefficient, Cp and Cpk ).
-
Measuring Values or Measuring Values + Deleted Values (see also section 8.2.2).
-
Trend (MiniTest 7400 FH only)
-
Histogram (MiniTest 7400 FH only)
-
Fig. 3
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Notes-Field (space for remarks with 5 lines)
-
----- (remains empty)
38
Page 39
Press OK to confirm your selection and go to the next position on the printing list and proceed as described above.
The print-out list can also be limited to fewer items (e.g. statistics and measuring values only, Fig. 4)
Fig. 4
Now align the infrared windows of the MiniTest and MiniPrint 7000. Press function key “Print” to start the print-out. To abort, press “ESC”.
9.1.1 Screen Shot
Please refer to section 11.2 “Special functions“.
9.1.2 Print-out a Batch
Please refer to section 9.1 “Data Pr int - out”
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
39
Page 40
9.2 Transfer Data to a PC
Before transferring data, go to the Main Menu and select “Setup”. Use arrow keys to move to “Data Output” and press OK to confirm. Use arrow keys to select “RS232 cable” or “Infrared (IrDA). Press OK to conf irm. See also section 10.5.4. I f you adjust to RS232, please mak e sure to make the correct settings for baud r at e and t r ansm ission protocol.
Readings and statistics may be transfer r ed to a PC from all statistical views such as statistics, batch, histogram or trend (MiniTest 7400 FH). Press “Help/Extra” Function key from your selected statistical view. The Help/Extra window opens (Fig. 1) Select “Print” form t he list and confirm by “OK”.
A “Printing sequence” menu opens. Move the cursor to the first item on the list and press OK. The following opt ions ar e available:
-
Name (of batch) / or name ( of batch) + date (MiniTest 7400 FH).
-
Statistics (number of reading s, Average (x) Standard deviation (s), Min., Max. Variation Coefficient, Cp and Cpk ).
-
Measuring Values or Measuring Values + Deleted Values (see also section 8.2.2).
-
Operator (with MiniTest 7200 FH Notes-Field only)
-
----- (remains empty)
Press OK to confirm you selection. Connect the MiniTest FH via the RS232 cable to your computer or
align the infrared windows of your computer or of your IrDA adapter respectively. Then press Function key “Send->PC”.
To abort, press “ESC”.
Please note
: The “MSoft 7000 Pro” software recalculates the histogram and trend diagram (MiniTest 7400 FH oly) and optimizes their display in graphical format .
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
40
Page 41
10. Main Menu
10.1 Database
Please refer to section 7.2.
10.2 Statistics
Please refer to section 8.
10.3 Calibration
Please refer to section 6.
10.4 Active Batch
10.4.1 Batch Properties
From Main Menu chose “Active Batch” and press OK to confirm. Now you can select from the options as shown on the left.
Select “Batch properties” and press OK to confirm. Now you can view: Batch name (with MiniTest 7400 FH) Create date, Modified date User name. Press function key “Statistics” to view the statistics of t his batch..
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
41
Page 42
10.4.2 Readings
10.4.3 Configuration
This menu is for viewing readings. To view the corresponding statistics, press function key “Batch” followed by function key “Statistics”. See also section 8.
In this menu you can set the following parameters to define your batch as requested:
Target ball size (see section 5.2.2) Data logging mode (see section 5. 5) Logging rate (see 5.5) Graphics on/off (see sect ion 4. 3. 2)
10.4.4 Calibration
Upper spec. limit (see section 8.1.1) Lower spec. limit (see section 8.1.1) Set point (see section 8.1.2)
Please refer to section 6.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
42
Page 43
10.5 Setup
10.5.1 General Remarks
In this menu both models, MiniTest 7200 FH and 7400 FH can be adjusted to customer requirements.
From Measure Menu press function key “Main Menu”. Use arrow key to select “Setup”. Press OK to confirm.
Use arrow keys to go to the requested item and pr ess OK to confirm.
10.5.2 Display
In this menu you can adjust the backlight of display and keys (on/off); brightness and contrast). Please refer to section 4.4.1 for more details.
10.5.3 Date and Time
In this menu you can adjust date and time, day, month and year. Please refer to section 4.4.1 for more details. The gauge f eatures a quartz-controlled time clock . The current time is displayed on the top right of display. If connected to the data printer or for PC evaluation, t he create date and time and date and time of last modificat ion m ay be included on the bat ch r epor t.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
43
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 44
10.5.4 Data Outputs
In the “Data Output” menu you can choose RS232 cable or Infrared (IrDa) . Please refer to section
4.4.1 for more details.
10.5.4.1 Data Output RS232 Cable
If “RS232 Cable” has been selected, you can adjust the data f ormat to the requested baud rate:
57600 (default), 38400, 19200, 9600
The settings 8 data bits, 1 Stop- Bit , No Parity do not change.
The Transmission Protocol option must be “off” when using the Msoft 7000 Basic. (Description of transmission protocol available on request).
If you have adjusted to “RS232 Cable“, you cannot connect a footswitch, charger unit or alar m device to t he m ulti-purpose socket.
10.5.4.2 Data Output Infrared (IrDA)
If you have adjusted to “Infrared (IrDA)”, data transmission to the MiniPrint 7000 data printer or to a PC will be wireless.
In this setting, you can connect a footswitch, power supply unit or alarm device to the multipurpose socket. If a footswitch has been connected, a power unit and an alarm device may be connected to the footswitch. Via the footswitch you can “Save” readings to the statistics or “delete” Minimum and Maximum values during measurement.
The duration of the alarm signal can be adj usted individually so that you are able to distinguish the alarm signals for exceeding upper specification limit and lower specification limit . The duration can be adjusted to 50, 100, 150, 200, 250, 300, 350, 400 ms or to
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
44
Page 45
10.5.5 User Preferences
permanent signal.
In the User Preferences menu the following settings can be made:
-
Language
-
Measuring system (metric / imperial)
-
Loudspeaker (ON/OFF).
-
Signal Volume (low, high, medium)
-
Auto Switch OFF (after 1, 3, 10, 30 minutes, disabled)
-
Entry accepted (numeric or alphanumeric entries accepted
(after 0,5, 1, 1,5 or 2 seconds) .
-
Modification lock**
** You can assign a password for the Modification lock (pay attention to upper and lower case so that the modification lock will be password protected. If you have forgotten the password you can
10.5.6 Gauge Specifications
make a total r eset . Please note: a total reset deletes all reading s and settings stored in memory. Please also refer to section 4.4.1.
From Setup Menu use arrow keys to go to “Gauge Specs“. Pr ess OK to confirm. Now you can view gauge specifications.
For viewing sensor data press function key “Sensor”.
In case of service requests please make available these data for further reference.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
45
Page 46
11. Special functions
11.1 Initializing
At first use or on demand the init ializing sequence may be called. Press ON/OFF button and ESC simultaneously to g o to the initializing sequence. A language flag symbol appears. Use arrow keys to make your selection. Press OK to confirm or ESC to abort. Now a Total Reset query appears. Press “No” if you do not wish to perfor m a T otal Reset to delete all data.
Attention!
Pressing “Yes“ will irrevocably delete all data and settings st ored in memory. However, your language setting remains and the gauge will be reset to factory setting. All directories and batches will be deleted along with their readings, statistics and calibration values. Press ESC to abort the Total Reset process. Only the language settings remain in memory.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
46
Page 47
11.2 Special functions Press ON/OFF button for more than one second while the gaug e is switched on. Now a Special
functions menu appears with the following setting options:
-
Print Screen (print-out of screenshot on MiniPrint 7000 data printer)
-
Backlight (of display and keyboard) ON
-
Backlight (of display and keyboard) OFF
-
Metric system
-
Imperial system
All these setting options except the Print Screen item may also be accessed via the menu system. The Special functions menu is meant for quick access to these functions at any time during operation. To quit the Special f unctions menu and to g o back to your current operat ion, press OK to confirm your setting or ESC.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
47
Page 48
12. Quick Reference
g
pp
p
A
y
guag
y
y
g
Measure Mode
Calibration
Target Ball Calibration Mode
Backlight
Brightness
Contrast
Da Month Year Hour Minute
Data output
Main Menu
Database
(MiniTest 7400 FH only)
Statistics Calibration Active Batch
Setup
Display Date / Time Data Output User Preferences Gauge Specs
Batch properties Readings Configuration Calibration
Statistics
Numerical Statistics Histogram Trend diagram
(MiniTest 7400 FH only)
Readings Delete all Readin
(MiniTest 7400 FH only)
s
Lan Meas. S Loudspeaker Signal Volume
Entr Modification lock
e
stem
uto Switch OFF
accepted
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Meas. Mode
et Ball Size
Tar Data logging mode Logging rate Graphics
ers spec. limit
U Lower s Set point
ec. limit
48
Page 49
13. Care and Maintenance
13.1 Care
Use a soft damp cloth with water or a mild detergent t o clean t he gauge and accessories.
Caution:
Do not use solvents because they might damage t he plastic parts. Do not use metal brushes or other tools for cleaning the sensor t ip.
13.1.1 Using NiMH storage batteries
To achieve optimal service life of the NiMH storage batteries, please respect the following instructions.
Make sure to use the full battery capacity befor e your recharg e it. Freq uent short discharg e and recharge cycles prematurely reduce their service life.
The storage batteries should always be fully recharged. Avoid prematur e interr upt of reload process, otherwise batteries may not be recharged to their full capacity any more.
Try to save battery lif e e. g. by operating the MiniTest 7200 FH / 7400 FH in Auto Switch off mode, instead of continuous service.
13.2 Maintenance
Generally, no maintenance work is required for MiniTest 7200 FH / 7400 FH.
Please note: Repairs may only be carried out by authorized ElektroPhysik staff.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
49
Page 50
14. Technical specification
14.1 Gauge and Sensor Specification
Measuring principle:
Measuring rate: 20 readings per second (factory set)
Number of readings to be logged for statistics:
Sensor types: FH4 FH10
Low range resolution: 0.1 μm 0.2μm
Display resolution Ranges Resolution:
magneto-static principle, suitable for nonmagnetic materials
1, 2, 5, 10 or 20 readings per second (user adjustable)
0.0 μm – 59.9 μm
60.0 μm – 149.9 μm
150.0 μm – 299.5 μm
300.0 μm – 599.0 μm
600.0 μm – 998.0 μm
1.000 mm – 1.498 mm
1.500 mm – 2.995 mm
3.00 mm – 5.99 mm
6.00 mm – 10.12 mm
0.1 μm
0.2 μm
0.5 μm 1 μm 2 μm 2 μm 5 μm 10 μm 20 μm
FH4 Sensor
Target ball Ø
1.5 mm 0 ... 1.5 mm ± (1.5 μm + 0.5%)* ± (10 μm + 3%)* ± (5μm + 1.5%)* ± (3μm + 1%)*
2.5 mm 0 ... 2.5 mm ± (2.5 μm + 0.5%)* ± (15 μm + 3%)* ± (8μm + 1.5%)* ± (5μm + 1%)*
4.0 mm 0 ... 4.0 mm ± (5 μm + 0.5%)* ± (30 μm + 3%)* ± (15μm + 1.5%)* ± (10μm + 1%)*
FH10 Sensor
Target ball Ø
2.5 mm 0 ... 2.5 mm ± (2.5 μm + 0.5%)* ± (15μm + 3%)* ± (8μm + 1.5%)* ± (5μm + 1%)*
4.0 mm 0 ... 4.0 mm ± (5 μm + 0.5%)* ± (30μm + 3%)* ± (15μm + 1.5%)* ± (10μm + 1%)*
6.0 mm
9.0 mm 0 ... 10 mm ± (10 μm + 0.5%) * ± (50μm + 3%)* ± (30μm + 1.5%)* ± (20μm + 1%)*
Measuring range
Measuring range
0 ... 6 mm ± (10 μm + 0.5%)* ± (50μm + 3%)* ± (30μm + 1.5%)* ± (20μm + 1%)*
6 ... 10 mm ± 0.5%* ± 4%* ± 2.5%* ± 1.5%*
Reproducibility
Reproducibility
Measurement on horizontal surfaces, mean value calculated from a minimum of 10 readings
Factory
calibration:
Factory
calibration:
All percentages are related to the measuring value
Tolerance: Zero-point
calibration
Tolerance: Zero-point calibration
Multi-point calibration
Multi-point
calibration
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
50
Page 51
Calibration modes:
Display: LCD 160 x 160 pixel, backlit (user adjustable)
Measuring system: metric (μm, mm) and imperial (mils)
Statistics functions: Single readings, number of readings, min, max, average, standard deviation,
Factory calibration, Zero point calibration and multi-point calibration ((Zero plus 1 to 4 calibration points)
variation coefficient and process capability indices Cp and Cpk (only available with MiniTest 7400 FH)
Data memory:
Maximum number of directories:
Maximum number of batches:
Maximum number of readings:
Differential mode:
Statistics:
Languages: English, German, French, Italian, Spanish, Portuguese
Interfaces: IrDA, 1.0, Multi-purpose socket for RS232 TTL, footswitch, power supply unit and
Operating temperature: -10°C to + 60°C
Storing temperature: -20°C to + 80°C
Protection class of gauge housing:
Dimensions and weight of gauge: 153 mm x 89 mm x 32 mm / 310 g including batteries
7200 FH 7400 FH 0 500 subdivided into 5 levels 1 200
100.000 240.000 Display of the difference between current reading and preset set point 7200 FH:
numerical
alarm device
IP 50
7400 FH: numerical, histogram, trend diagram
Dimensions and weight of FH 4 sensor
Dimensions and weight of FH 10 sensor:
Dimensions of plastics carrying case:
Power supply: 4 x 1,5 V LR 6 – AA baby cells, Power supply unit (90 to 240V AC /
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
Ø 17 mm x 96 mm / 90 g
Ø 30 mm x 125 mm / 300 g
365 mm x 450 mm x 140 mm
48 - 62 Hz)
51
www.GlobalTestSupply.com
Page 52
14.2 Delivery Schedule
14.2.1 Thickness gauge MiniTest 7200 FH / 7400 FH
Description
Thickness gauge MiniTest 7400 FH for non-magnetic materials 80-176-0000 or Thickness gauge MiniTest 7200 FH for non-magnetic materials 80-177-0000 including:
Instruction manual in German/English/French on CD-Rom Short instructions 4 x batteries AA (LR06), Plastics carrying case, rubber protection case with positioning device and shoulder belt, 1 magnetic screwdriver
Order number
14.2.2 Sensors
Description
Sensor FH4 for thickness measurement measuring range: 0 ... 4.0 mm
including: shielding tube 3 precision standards 0.25 mm, 1 mm, 3 mm 1 zero calibration standard for the target balls with 1.5 mm, 2.5 mm, 4 mm dia. (1 zero calibration standard per size) 1 spring loaded sensor stand 1 set of target balls with 1.5 mm and 2,5 mm dia.(100 balls each size) 1 set of target balls with 4 mm dia (50 balls)
Order number
80-174-0000
Sensor FH 10 for thickness measurement measuring range: 1…10.0 mm
including shielding tube 3 precision standards 1 mm, 3 mm, 8 mm set of zero calibration standards for the target balls with 2.5mm, 4mm and 6mm dia. (1 pc each ball size) 1 spring-loaded sensor stand 1 set of target balls with 2.5 mm dia. (100 balls) 1 set of target balls with 4 mm dia (50 balls) 1 set of target balls with 6 mm dia (25 balls)
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
80-175-0000
52
Page 53
14.3 Accessories Description
Order number
Data printer MiniPrint 7000 incl. charger unit Quick charger for NiMH storage batteries NiMH storage baby cells, AA HR6 1,2V (4 pcs required f or MiniTest FH) Baby cells AA LR6 1,5V (4 pcs required for MiniTest) Power supply unit Footswitch for storing readings into the data memory, incl. adapter unit for mains operation Rubber protection case with positioning device (included in standard supply schedule) Belt for rubber protection case Case with shoulder belt for MiniTest 7200 FH / 7400 FH Plastics carrying case (incl. in standard supply schedule) Dust protection pouch IR / USB converter IR/RS232 converter Connecting box for footswitch, USB, power unit, alarm deice, headphones
80-202-0000 02-070-0001 02-064-0001 02-064-0008 80-900-0004 80-901-1800
82-010-0003
82-010-0016 82-010-0007 82-020-0033 85-199-0001 85-139-0014 85-139-0013
MSoft 7000 basic data transmission software MSoft 7000 pro data management software
80-901-1600 80-901-xxxx
Manufacturer’s Test Certif icat e accor ding to DIN 55350 M for gauges and sensors
Accessories for sensor FH4
Sensor stand, spring-loaded, for s Sensor FH4 (incl. in standard delivery) 70-174-0015
Zero calibration standard for FH4 sensor 1. 5 mm dia. target ball (stand. delivery) 82-174-0024 Zero calibration standard for FH4 sensor 2. 5 mm dia. target ball (stand. delivery) 82-174-0025 Zero calibration standard for FH4 sensor 4 mm dia. target ball (stand. delivery) 82-174-0026
Precision standard for sensor FH4 0.3 mm (standard delivery) 82-174-0020 Precision standard for sensor FH4 1 mm (standard delivery) 82-174-0021 Precision standard for sensor FH4 3 mm (standard delivery) 82-174-0023
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
53
Page 54
1 set precision standards for sensor FH 4 for multi-point calibration for 1.5 mm dia. target ball size, comprising: 1 precision standard set for approx. 0.15 mm, 0.25 mm, 0.43 mm, 0.75 mm and
1.3 mm 1 set precision standards for sensor FH4 for multi-point calibration for 2.5 mm target ball size, comprising: 1 precision standard set for approx. 0.25 mm, 0.43 mm, 0.75 mm, 1.3 mm and
2.2mm (1 pc each) 1 set precision standards for sensor FH4 for multi-point calibration for 4,0 mm target ball size, comprising: 1 precision standard set for approx. 0.43 mm, 0.75 mm, 1.3 mm, 2.2 mm and
3.6mm ( 1 pc each)
1 set of target balls, 1.5 mm dia. (100 pcs) 82-174-0004 1 set of target balls, 2.5 mm dia. (100 pcs) 82-175-0004 1 set of target balls, 4 mm dia. (50 pcs) 82-175-0005
Manufacturer’s Test Certificate according to DIN 55350 M for precision standards Manufacturer’s Test Certificate according to DIN 55350 M for 1 set precision standards for multi-point calibration
82-174-0010
82-174-0011
82-174-0012
Accessories for sensor FH10
Sensor stand, spring-loaded, for s Sensor FH 10 (incl. in standard delivery) 70-175-0014 Zero calibration standard for FH 10 sensor 2. 5 mm dia. target ball (standard
delivery) Zero calibration standard for FH 10 sensor 4 mm dia. target ball (standard delivery) Zero calibration standard for FH 10 sensor 6 mm dia. target ball (standard delivery) Zero calibration standard for FH 10 sensor 9 mm dia. target ball (standard delivery)
Precision standard for sensor FH10 1 mm (standard delivery) 82-175-0030 Precision standard for sensor FH10 3 mm (standard delivery) 82-175-0031 Precision standard for sensor FH10 8 mm (standard delivery) 82-175-0032
1 basic calibration set for sensor FH10 for target ball dia. 9.0 mm, comprising: Zero calibration standard, and 10 pcs target balls 9 mm dia., comes in a plastics box
1 set precision standards for sensor FH10 for multi-point calibration for target ball dia. 2.5 mm, comprising: 1 set precision standards approx. 0.25 mm, 0.43 mm, 0.75 mm, 1.3 mm and
2.2mm
82-175-0033 82-175-0034 82-175-0035 82-175--0036
82-175-0038
82-175-0020
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
54
Page 55
1 set precision standards for sensor FH10 for multi-point calibration for target ball dia. 4.0 mm, comprising: 1 set precision standards approx 0.43 mm, 0.75 mm, 1.3 mm, 2.2 mm and 3.6 mm 1 set precision standards for sensor FH10 for multi-point calibration for target ball dia. 6.0 mm, comprising: 1 set precision standards approx 1.0 mm, 1.7 mm, 3.0 mm, 5.2 mm and 9.0 mm 1 set precision standards for sensor FH10 for multi-point calibration for target ball dia. 9.0 mm, comprising: 1 set precision standards approx 1.0 mm, 1.7 mm, 3.0 mm, 5.2 mm and 9.0 mm
82-175-
82-175-
82-175-
0017
0010
0037
1 set of target balls, 2.5 mm target ball size (100 pcs) 1 set of target balls, 4 mm target ball size (50 pcs) 82-175-0005 1 set of target balls, 6 mm target ball size (25 pcs) 82-175-0006 1 set of target balls, 9 mm target ball size (10 pcs) 82-175-0014
Manufacturer’s Test Certificate according to DIN 55350 M for precision standards Manufacturer’s Test Certificate according to DIN 55350 M for 1 set precision standards for multi-point calibration
82-174-0004
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
55
Page 56
15. Appendix
15.1 Troubleshooting
Error message Problem Remedy
“Please check clock settings“ If you insert batteries at initial use
or if the gauge has been cut from power supply for more than 5 minutes
Reset clock (s. section 10.5.3)
“Low Bat !!!” Though you can continue to
measure, it is recommended to change batteries. If battery voltage is below minimum, the gauge will switch off
automatically. “Please hold probe in air(infinite range) !! “
(Briefly appears on display at switch on)
“Confirm sensor in infinite range” a) a target ball is placed on the
Please check
a) whether you have removed the
shielding tube or protection cap
b) whether a target ball is
positioned on the sensor
c) whether the sensor tip is in
contact with a steel part
See also section 6.1
sensor tip or the sensor is near an
ferritic object or magnetic field.
b) Infinite adjustment required.
See also section 6.1
tip
Change batteries. Dispose batteries according to your local regulations.
a) Remove shielding tube or
protection c
Remove target ball from sensor
b)
tip
c) Enlarge distance between
sensor head and steel part
Now the message on display should disappear and the sensor is ready for measurement. The message does not disappear, and the following message appears after about 10 sec:
“Make sure there is no target ball near the sensor. Please keep the sensor away from metallic objects or magnetic fields.”
ap
If this message persists, press
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
56
Page 57
Function key “Infinite”. Now the message should disappear and the gauge is ready for measurement.
Attention: Make sure not to press the Infinite function key unless you have removed the error sources (target ball, magnetic field or metallic parts).
“Refresh infinite value !!!”
(The infinite value should be actualized)
“Sensor type not compatible with batch ! “ “Service failure” (MiniTest 7400 FH only) “Batch readings do not match connected sensor !”
“Delete readings?” “Please check for correct target ball size!”
“Please check sensor and target ball position!”
“Carry out Zero measurement and the gauge will be ready for measurement again.
To ensure the guaranteed
measuring accuracy, the infinite
value must be refreshed.
See also section 6.1
Another sensor has been
specified for the active batch.
Another sensor has been
specified for the active batch.
You have probably selected a
wrong target ball.
Make sure there is not target ball
positioned laterally to the sensor
tip. Make sure to keep the sensor
out off the influence of ferrite
objects or magnetic fields.
The magnetic field of the sensor
has been influenced e.g. by a
target ball laterally to the sensor
tip or by any magnetic field.
Remove target ball from sensor. If the message persists, please press Refresh function key.
During a measurement you can press ESC in order to cancel the Refresh procedure. Select or create another batch.
Select or create another batch or add sensor data of the sensor currently being used.
Make sure the adjusted target ball size matches the target ball used for measurement. Remove target ball from sensor. Keep away from ferrite objects or magnetic fields.
Place target ball matching to the adjusted target ball size on the sensor tip. This step is required to make the gauge ready for
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
57
Page 58
Target ball: (adjusted ball size) measurement. “Full memory capacity !” Probable cause:
You have reached maximum
number of storable readings.
You have reached maximum
number of entries in the data
base (MiniTest 7400 only) “Memory optimization in progress” “Check sensor connection !” Possible cause:
“No sensor connected ! “
“Changes not possible !”
“Sensor failure “” Please contact after-sales
Message appears briefly on
display.
the sensor has been
removed
loose probe connection damaged cable defective sensor
Sensor connecting error.
Possible cause:
no sensor connected loose probe connection damaged cable
defective sensor
Delete readings, batches or directories in the database which you do not need any more.
Automatic procedure. No further action required. Please check the options. If the sensor is defective, please change it or contact after sales service.
Please check the options. If the sensor is defective, please change it or contact after sales service.
service.
The following errors may be remedied by a total r eset (see section 11.1)
-
keyboard failure
-
no measuring admitted
-
illogical readings
If the gauge cannot be switched of f via the ON/OFF button, remove and reinsert batter ies.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
58
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 59
15.2 Detection of Error sources (INF flag)
Measurement may be influenced by static or alternating m agnetic fields and by metallic objects near the sensor.
If no target ball is placed on the sensor , the “INF” flag appears on display. That means that the probe is in the infinity range.
INF flag
INFdis­appeared
Make sure that INF is shown so that correct readings can be taken.
If the sensor is subject to some int er ference (e.g. through electro-motors, transformers, potential distribution or if a steel plate is near t he sensor) the INF flag disappears.
Try to move the sensor a little bit in order to locate the source of interference. If you have detected the error source, move the sensor away from it. As soon as the “INF” flag appears on display, the sensor is out of the error source range and the gaug e is r eady again to take correct measurements.
To be sure, take double the distance form the error source.
In the default measure mode using a target ball, it is not of importance if t he “INF”-flag disappears.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
59
Page 60
=
15.3 Statistical Terms
The statistical evaluation will help you to assess the quality of your product.
Average (Mean)
The sum of single readings divided by the tot al num ber of readings.
x
=
x
n
Standard Deviation (STD. DEV.) s (s = σ = sigma) The sample standard deviation is a statistic that measur es how “dispersed“ t he sam ple is ar ound
the sample mean. The sample standard deviation increases with increasing spr ead out. The standard deviation of a set of numbers is t he r oot mean square of the variance s².
2
=
s
ss
2
xx
()
n
1
2
Variance
The variance of a list is the square of the standard deviation of the list, that is, the average of the squares of the deviations of the numbers in the list from their mean divided by the (number of readings minus 1).
Variation coefficient (Var.-coeff.)
The variation coefficient is the st andar d deviation divided by the arithm etic mean. The variation coefficient is indicated in percent.
K
s
var %100
x
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
60
Page 61
Process capability index (Cp)
s
s
The cp index is a measure of the spread of the readings as related to the specification lim it s. Only the spread is of importance here. The cp index is calculated as follows:
LSGLUSL
Cp
w
=
s
6
here USL = Upper specification limit
where LSL = Lower specification limit
Process capability index (Cpk)
In addition to the spread, the cpk index also takes into account the location of midpoint as related to the specification limits.
LSLx
Cpku
Cpko
=
3
xUSL
=
3
Cpk = Min { Cpkl, Cpku }
where USL = Upper specification limit where LSL = Lower specification limit
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
61
Page 62
15.4 Safety Notes
Safe operation will be ensured as far as the instr uctions and notes in this manual and/or gauge will be observed. For installation work, please cut power supply. Use only original spare parts and/or accessories. If not in use, the sensor must always
Risk through exposure to magnetic fields
Within a distance of 15 cm, the sensor creates static magnetic f ields of up to 1 mTesla.
Pacemakers
Patients with heart pacemakers should consult their doctor to clarify whether using the magnetic sensor may imply any health risk for them.
Magnetic data carriers
The magnetic field created y the sensor may impair the operability of magnetic data carriers, electronic devices and all kinds of encoded magnetic media. Do not place the sensor alongside computer disks, credit cards, travel cards and other magnetic media. The information contained on disks or cards m ay be affected by the sensor. Make sure to keep a safety distance of at least 30 cm.
be stored with the protection tube put on it.
Keep away from monitors and electronic devices
The magnetic field created by the sensor may impair the operability of computer or video monitors, elect ronic devices and measuring gauges or even destroy them. Make sure to keep a safety distance of at least 20 cm from such devices.
Keep away from watches and metallic parts
The static magnet ic field creat ed by the sensor may magnetize watches or other metallic par ts. Make sure to keep a saf ety distance of at least 20 cm from such items.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
62
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 63
Risk of injury
Within a distance of 15 cm, the sensor creates static magnetic f ields of up to 1 mTesla Keep a safety distance of at least 15 cm in order to avoid injury through metallic parts being attracted by the sensor
Risk of injury through metallic splinters
The magnetic field may attract metal parts damaging the hard metal sensor tip. Throug h the impact, the hard metal housing of sensor might go into splinters.
Storage batteries and accessories
Make sure to use only original accessories and batteries supplied/recommended by the manufacturer of gauge. Connect only to compatible peripheral devices.
Connecting other devices
If you connect the gauge to another device as recommended by the manufacturer, please ref er to the corresponding instruction manual for detailed information on safety issues. Do only connect original accessories recommended by the manufacturer of the MiniTest 7200 FH / 7400 FH.
Keep away from water
The measuring unit is not waterproof. Keep in a dry place.
Keep away from explosion-hazardous area
Approved after-sales service
The gauge may only be repaired by approved and qualified af ter-sales service personnel.
Medical facilities
Please ask for permission bef or e using the gauge in medical facilities.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
63
Page 64
15.5 Declaration of Conformity
We, Elek troPhysik, Pasteurstr. 15, D-50735 Colog ne, Germany, declare in sole responsibility that the products MiniTest 7400 FH, MiniTest 7200 FH and MiniPrint 7000 to which this declaration relates is in conformity with the provisions of direct ive 89 / 336 / EEC ( Elect r om agnetic compatibility), in Germany: EMVG (Gesetz über die elektromagnetische Verträglichkeit) of November, 9
th
, 1992.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
64
Page 65
15.6 After Sales Service
MiniTest 7200 FH / MiniTest 7400 FH models are manufactured according t o st at e-of-the-art production methods using high-class components. Car eful production controls along with a Certified Quality Management according to DIN EN ISO 9001 ensure optimum product q ualit y.
In case of errors please contact Elek t roPhysik. If repairs should become necessary, please send the gauge to ElektroPhysik or contact your local ElektroPhysik representative for r eturn and repair instructions.
Please note that the gauge should only be repaired by authorized, sk illed and trained personnel. Service attempts by untrained personnel could cause extensive damage to the gauge and possibly void any and all warranties.
For more detailed inf ormation on the use, applications, service or technical data, please contact your local ElektroPhysik representative.
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
65
Page 66
Index
A
average ...................................................51, 60
Average
...................................................38, 40
B
backlight.......................................................43
batteries
inserting......6, 10, 11, 25, 49, 51, 56, 58, 63
battery symbol..............................................10
C
Calibration....18, 24, 25, 26, 29, 30, 41, 42, 51
charger unit
external...............................................10, 44
Configuration....14, 17, 18, 20, 21, 30, 32, 34,
35
D
Directory ................................................31, 33
display........5, 7, 10, 11, 13, 14, 32, 43, 47, 56
E
entries
alphanumerical.............................15, 16, 45
numerical......................................15, 16, 45
F
factory calibration ................................7, 8, 32
foot switch....................................................10
G
gauge specifications.....................................45
H
histogram................................................34, 37
I
Infrared.........................................................38
L
language...................................................7, 46
lower specification limit ..............................14
LSL..................................................14, 34, 61
M
Main Menu..14, 15, 31, 32, 33, 34, 35, 36, 37,
40, 41, 43
Measuring Values
...........................................38
modification lock.........................................45
multi-point calibration...........................27, 51
multi-purpose socket..............................10, 44
P
power supply unit ............................10, 11, 51
precision standard................25, 26, 27, 28, 29
Process capability
Cp
.............................................................38
Cp
.............................................................40
Cpk
...........................................................40
Process capability Cpk
....................................38
S
sensor stand............................5, 12, 19, 20, 26
set point..................................................35, 51
single readings.................................27, 30, 60
Standard deviation
....................................38, 40
switch off
automatic..................................................49
Switch ON .....................................................7
T
Total Reset...............................................7, 46
tree-structure................................................31
trend diagram...............................................34
U
user name...............................................32, 41
USL..................................................14, 34, 61
© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
66
Loading...