ElektroPhysik 80-124-1401 User manual

Page 1
Operating Instructions
for the Coating Thickness Gauges
MiniTest
650E F
650E FN
MiniTest
650B FN
Table of Contents
A. Start-up B. Operation C. Troubleshooting D. Technical Data
Field of Applications
Small and portable gauge for quick and easy on-site use, for example by surveyors, car dealers or on your construction site.
MiniTest 650E F and MiniTest 650B F work according to the principle of magnetic induction to measure non-magnetic coatings on iron and steel substrates.
MiniTest 650E FN and MiniTest 650B FN use a dual sensor working on two principles: magnetic induction and eddy currents to measure non-magnetic coatings on iron and steel as well as insulating coatings on non-ferrous metal substrates.
The dual FN sensor automatically identi fies the sub strate materia l and switches to the appropriate measuring principle accordin gly .
All models conform to: DIN EN ISO 1461, 2178, 2360, 2808, 3882; ASTM B244, B499, D7091, E376
Scope of delivery
- MiniTest 650E or MiniTest 650B
- 1 or 2 zero reference plate(s)
- 1 calibration foil
- 3 batteries AAA (Micro)
- Operating instructions German/ English/ French
- Soft pouch
A. Start-up
Inserting the batteries
MiniTest 650 is delivered including 3 batteries. Place the gauge front side down onto an ev en surface. Loosen the screw of the battery compartment, remove the lid and insert the three batteries (AAA). Make sure to respect correct polarities, following the markings in the battery compartment.
Replacing the batteries
Otherwise, the battery symbol will continue to flash after you switch OFF and ON again and the gauge will shut down after a second.
Caution:
Please respect polarities when changing batteries ! Make su re to inser t the fresh batteries within 30 seconds after removing the old ones. Otherwise, the calibration values will get lost.
B. Operation
Control keys
MiniTest 650E F and E FN ar e operat ed via one single con trol key, the models MiniTe s t 650B F and B FN are oper ated via three key s .
The following events require a battery change:
BATT flashing after switch-ON, E6 is shown and the gauge shuts down after one second: Change batteries immediately.
BATT flashing during measurement: You may continue measurement. Then
switch the gauge off and change batteries.
1. Switch ON
Briefly press ON. The last reading is shown along with the matching measuring principle (“Ferr“ or “Non-Ferr“).
2. Switch OFF
MiniTest 650 is programmed to switch off auto­matically approx.. 90 sec. after the last measurement or k ey action. Alternativ ely, use
3. Calibration
MiniTest 650E F and MiniTest 650E FN are factory-calibrated. No further calibration is required. After switch-ON you can directly proceed on measurement.
3.1 Zero-point calibration
MiniTest 650B F and B FN are also factory-calibrated and directly ready for measurement after switch-ON. Their factory calibration is sufficient for sim ple an d quic k meas ureme nt and if la rger t olera nces are acceptable. To increase the measuring accuracy, you should use the zero-point calibration. It is also recommended for substrates of a nature other than the one of the reference zero plates or for curved or rough samples.
During measurement, MiniTest 650B FN automatically identifies the substrate. In the calibration mode, however, the automatic switch to the appropriate measuring principle beco mes inacti ve. Therefore, prior to calibration, it is requested to ta ke a reading o n the uncoa ted sampl e (ferrous or non-ferrous), if calibration is to be done for a measuring principle other than the one used for the last measurement. Place the sensor on the sample. “Ferr” or “Non-Ferr” will be shown accordingly.
If, later on, you wish to measure on both substrates (ferrous and non­ferrous), make sure to calibrate for both substrates. On that purpose, please use an uncoated sample for each substrate.
the ON/ OFF button for switching off.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
For zero-point calibration, please proceed as follows:
© ElektroPhysik, Version 1.0 B4
Subject to change without notice.
Gauge type
MiniTest 650E F
MiniTest 650E FN
Measuring range
F: 0...3 mm / 120 mils
F: 0...2 mm / 80 mils N: 0...2 mm / 80 mils
Measuring principle
magnetic induction / eddy
current
Measuring uncertainty*
± (3 μm + 5 % of reading) /
± (0.12 mils + 5 % of reading)
± (3 μm + 5 % of reading) /
± (0.12 mils + 5 % of reading)
Low range resolution
2 μm / 0,08 mils
2 μm / 0,08 mils
Min. curvature radius, convex
50 mm / 2
50 mm / 2
Min. curvature radius, concave
Min. area for measurement*
2
ø 50 mm / ø 2”
ø 50 mm / ø 2”
Min. substrate thickness*
F: 0.70 mm / 28 mils
N: 0.1 mm / 4 mils
Measuring units:
according to model: μm/mm or mils/inch
Calibration
factory calibration (no
calibration required)
factory calibration (no
calibration required)
Operating temperature
gauge: 0…50 °C / 32 °…122 °F
sensor: –10 °…70 °C / 14 °…158 °F
Power supply
3 Micro-AAA batteries for more than 10,000 readings
Standards
DIN EN ISO 1461, 2178, 2360,2808, 3882
ASTM B 244, B 499, D7091, E 376
Dimensions
housing: 70 mm x 122 mm x 32 mm / 2.7” x 4.8” x 1.26” sensor: Ø 15 mm x 62 mm / Ø 0.60” x 2.44”;
Weight including batteries
approx. 225 grams / 7.93 ozs
Gauge type
MiniTest 650B F
MiniTest 650B FN
Measuring range
F:
F: 0...2 mm / 80 mils N: 0...2 mm / 80 mils
Measuring principle
magnetic induction / eddy
current
Measuring uncertainty*
± (2 μm + 3 % of reading) /
± (0.08 mils + 3 % of reading)
± (2 μm + 3 % of reading) /
± (0.08 mils + 3 % of read ing)
Low range resolution
1 μm / 0.04 mils
1 μm / 0.04 mils
Min. curvature radius, convex
10 mm / 0.4“
10 mm / 0.4“
Min. curvature radius, concave
Min. area for measurement
ø 50 mm / ø 2”
ø 50 mm / ø 2”
Min. substrate thickness
F: 0.70 mm / 28 mils
N: 0.1 mm / 4 mils
Measuring units:
according to model: μm/mm or mils/inch
Calibration
factory calibration , zero-
point calibration
factory calibration , zero-
point calibration
Operating temperature
gauge: 0…50 °C / 32 °…122 °F
sensor: –10 °…70 °C / 14 °…158 °F
Power supply
3 Micro-AAA batteries for more than 10,000 readings
Standards
DIN EN ISO 1461, 2178, 2360,2808, 3882
ASTM B 244, B 499, D7091, E 376
Dimensions
housing: 70 mm x 122 mm x 32 mm / 2.7” x 4.8” x 1.26” sensor: Ø 15 mm x 62 mm / Ø 0.60” x 2.44”;
Weight including batteries
approx. 225 grams / 7.93 ozs
MiniTest 650B F or B FNare switched ON.
1. Press ZERO-key to intialize calibration. ‘ZERO’ flashing and ‘MEAN’ is shown to indicate that a mean value willl be shown.
2. Take your uncoated sample (coating thickness = zero) and place the sensor onto it. Wait for the signal tone to sound and lift the sensor. Repeat this procedure several times. The mean value calculated from your set of measurements is shown. If you wish to abort the zero-point calibration, press CLEAR-key.
3. Press ZERO-key to complete zero-point calibration. “ZERO“ (non-flashing) will be shown on display.
Now you can take readings!
The calibration needs to be effected f or each principle o f measuremen t (”Ferr“ or “Non-Ferr“) individually.
3.2 Delete calibration
Press ZERO followed by CLEAR. The ZERO-calibration will be deleted. Note: The factory calibration for use on flat and even surfaces will be enabled automatically.
C. Trouble Shooting
The following list of error codes explains how to identify and eliminate
faults. All error codes start with an “E” for “Error”.
Errors causing the gauge to shut-down: E 3: Sensor failure. This error code appears at switch ON. E 4: Sensor providing unstable readings (due to strong magnetic
fields in close proximity or if readings are taken on soft coatings).
E 5: At switch-ON, the sensor was held in close proximity to metal.
E 6: Low battery. E11: Remove batteries for 2 minutes
Failures without any error code shown:
x The gauge will not switch off automatically x The gauge is unable to take any further readings x Keys are without function x Illogical readings
Remedy with MiniTest 650B: Total Reset Note: A Total Reset will delete all calibration values!
Switch the gauge off. If the gauge will not switch off via the ON/OFF­key, please remove batteries prior to the Total Reset.
Press and hold CLEAR and ZERO key, then press the ON-key simultaneously.
A long bleep sounds to confirm the Total Reset. The last reading and the calibration values have been deleted.
6-N 07/2014
D. Technical Data
1
magnetic induction
100 mm / 4” 100 mm / 4”
2
F: 0.70 mm/ 28 mils
*1
acc. to DIN 55350 part 13
0...3 mm / 120 mils
magnetic induction
1
50 mm / 2” 50 mm / 2
F: 0.70 mm/ 28 mils