Elcometer 456 Operating Instructions Manual

English
Elcometer 456
2
Coating Thickness Gauge

Basic Models

Operating Instructions
English
These instructions apply to the following Elcometer 456
• Ferrous (F)
• Non-Ferrous (NF)
• Dual Ferrous/Non-Ferrous (FNF)
Equipment described in these instructions is covered by the following Patents:
FNF UK Patent No: GB2306009B
FNF US Patent No: 5886522
F1 2 UK Patent No: 2367135B
F1 2 US Patent No. US 6,762,603
F1 2 German Patent Pending
2
models:
This product meets the emc directive 89/336/EEC, amended 92/31/EEC and 93/68/EEC.
trademarks of Elcometer Instruments Ltd.
and are registered
All other trademarks acknowledged.
© Copyright Elcometer Instruments Ltd. 2004-2005.
All rights reserved. No part of this Document may be reproduced, transmitted, transcribed, stored (in a retrieval system or otherwise) or translated into any language, in any form or by any means (electronic, mechanical, magnetic, optical, manual or otherwise) without the prior written permission of Elcometer Instruments Ltd.
Doc.No. TMA-0302 Issue 02
Text with Cover No: 18612

CONTENTS

Section Page
1 About your gauge................................................ 3
1.1 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 What this box contains . . . . . . . . . . . . . . . . . . . . . 5
1.4 Conventions in these instructions . . . . . . . . . . . . . 5
1.5 Quick-start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2 Getting started ..................................................... 6
2.1 Fitting the batteries . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2 Battery condition. . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 Fitting probes (separate versions only) . . . . . . . . . 7
2.4 The controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.5 Switching the gauge on . . . . . . . . . . . . . . . . . . . . 9
2.6 Switching the gauge off . . . . . . . . . . . . . . . . . . . . 9
2.7 The screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.8 Selecting a language . . . . . . . . . . . . . . . . . . . . . 11
2.9 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3 Taking a reading ................................................ 13
3.1 Before you start . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4 The reading screen and menus........................ 14
4.1 Reading screen . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.2 Main MENU . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.3 Main MENU - Extended menu off . . . . . . . . . . . . 16
4.4 Main MENU - Extended menu on . . . . . . . . . . . . 18
5 Calibration adjustment...................................... 23
5.1 Calibration method . . . . . . . . . . . . . . . . . . . . . . . 23
5.2 Calibration foils and standards . . . . . . . . . . . . . . 24
5.3 Calibration adjustment procedure . . . . . . . . . . . . 26
6 Statistics............................................................. 32
6.1 Enlarge stats . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
6.2 Stats on LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6.3 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6.4 Clear stats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6.5 Select stats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
1
7 Probes ................................................................ 35
7.1 Ferrous (F) probes . . . . . . . . . . . . . . . . . . . . . . . 35
7.2 Non-ferrous (N) probes . . . . . . . . . . . . . . . . . . . . 35
7.3 Dual ferrous/non-ferrous (FNF) probes . . . . . . . 35
7.4 Probe interchangeability . . . . . . . . . . . . . . . . . . . 36
7.5 F1 2 probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
7.6 F1 2 High temperature PINIP™ probes . . . . . . . 37
7.7 Coatings on galvanised or metallised steel . . . . 37
8 Personalised welcome screen ......................... 38
9 Storage and transit............................................ 39
10 Maintenance....................................................... 39
11 Statistics terminology ....................................... 40
12 Technical data.................................................... 41
13 Accessories ....................................................... 43
14 Related equipment ............................................ 45
15 Fitting the wrist harness ................................... 45
16 Probe measurement performance ................... 46
17 Probe capabilities.............................................. 47
18 Error messages ................................................. 54
19 Index ................................................................... 56
2
hank you for your purchase of this Elcometer 456 Coating Thickness Gauge. Welcome to Elcometer.
T
Elcometer are world leaders in the design, manufacture and supply of coatings inspection equipment. Our products cover all aspects of coating inspection, from development through application to post application inspection.
The Elcometer 456 beating product. With the purchase of this gauge you now have access to the worldwide service and support network of Elcometer. For more information visit our website at www.elcometer.com.
2
Coating Thickness Gauge is a world

1 ABOUT YOUR GAUGE

The Elcometer 4562 Coating Thickness Gauge is a handheld gauge for fast and accurate measurement of the thickness of coatings on metal substrates.
The gauge is available in three versions; Basic, Standard and Top. This manual describes the operation of the
Elcometer 456
All versions of the gauge feature an easy-to-use menu driven graphical interface which guides the user through tasks such as gauge configuration and calibration adjustment.
The gauge is available either with a built-in integral probe or as a separate probe version. A wide range of probes is available to suit requirements - see page 35. Separate probes may be standard, miniature or Plug in Integral Probes (PINIP™), and must be ordered separately.
2
Basic.
R
Figure 1. Elcometer 4562
Coating Thickness Gauge
2
3

1.1 Features

• A range of smooth and rough surface calibration adjustments.
• Menu driven backlit graphical user interface.
• Interchangeable probes (separate versions only).
• Simple statistics.
• Infrared interface.

1.2 Standards

The Elcometer 4562 can be used in accordance with the following National and International Standards:
Ferrous (F) Non-Ferrous (NF)
BS 5411 (11) BS 5411 (3)
BS 3900 (C5) BS 3900 (C5)
ISO 2178 BS 5599
ISO 2808 ISO 2360
BS EN ISO 1461 ISO 2808
DIN 50981 DIN 50984
ASTM B 499 ASTM D 1400
ASTM D1186 ASTM B 244
ISO 19840
SSPC-PA2 (2004)
4

1.3 What this box contains

• Elcometer 4562 Gauge with integral probe, or
Elcometer 456 (probe must be ordered separately)
2
Gauge and separate probe
• Calibration foils
• Gauge carrying pouch
• Wrist harness
• Batteries
• Operating instructions

1.4 Conventions in these instructions

The Elcometer 4562 is controlled using a simple menu structure which helps you get the most from your gauge
- see page 18.
As an example, the LANGUAGES option which is in SETUP from the MAIN MENU would be shown in these instructions as MENU/SETUP/LANGUAGES.
These instructions include images of Elcometer 456 screens with units set to microns (µm). Similar screens will be seen when the gauge is set to other units such as mils or inches.

1.5 Quick-start

To quickly configure the gauge and start taking readings:
1. Fit batteries: see page 6
2. Fit probe (separate versions only): see page 7
3. Switch on: see page 9
4. Select language: see page 11
5. Try taking a reading: see page 13
6. Adjust calibration: see page 23
The gauge is now configured and ready to use.
To maximise the benefits of your new Elcometer 456 Operating Instructions. Do not hesitate to contact Elcometer or your Elcometer supplier if you have any questions.
2
, please take some time to read these
2
5

2 GETTING STARTED

2.1 Fitting the batteries

1. Open battery compartment cover; press down in direction of arrow using thumb nail.
2. Insert 2 x LR03 (AAA), alkaline dry batteries taking care to ensure correct battery polarity (Figure 2).
3. Close battery compartment cover.
Rechargeable batteries can be used but they will only have 25% to 30% of the life of alkaline batteries.
Figure 2. Fitting batteries -
ensure correct battery polarity

2.2 Battery condition.

Symbol Battery condition/action required
100%
66% to 100%
33% to 66%, replacement recommended.
16% to 33%, replacement required.
<16%, gauge beeps every 10 seconds and symbol flashes - immediate replacement required.
5 loud beeps, gauge switches off automatically.
6

2.3 Fitting probes (separate versions only)

To ensure correct transfer of data from the probe and detection of the new probe, the gauge must be switched off when
A probe must be calibrated once it has been fitted ­see “Calibration adjustment” on page 23.
2.3.1 To fit the probe
Align connector keyway and push in direction shown. The connector locks automatically.
Note: The design of the probe connector allows some movement between the probe and the gauge. This is intentional and does not affect measurement performance.
2.3.2 To release the probe
Grasp knurled section and pull gently away from the gauge. The connection will unlock and the probe will release.
2.3.3 To fit the PINIP™
Twist the PINIP™ until the connector locates. Taking care not to cross the threads, lock the probe by turning the locking ring 1½ times clockwise, or until tight.
separate probes are fitted.
7

2.4 The controls

The gauge is operated by 5 keys (Figure 3).
• On/Off key : Switches the gauge on or off.
• Softkeys: The function of these keys varies and is described by symbols and writing on the bottom line of the screen.
• LED: Red/green flashes when the gauge is switched on, green flashes when a reading is taken.
LED
Softkeys
R
On/Off key
2
Figure 3. Elcometer 456
control keys
8

2.5 Switching the gauge on

Gauges with separate and PINIP™ probes:
Press key to switch on gauge.
R
Gauges with integral probes:
Press key to switch on gauge, or place the probe on a surface.
Note: Before switching the gauge on for the first time read “Selecting a language” on page 11.
09 : 30 09 / 1 0 / 20 00
456
CAL DATA STATS MEN U
F1
F
µm
R

2.6 Switching the gauge off

To switch off all gauge types, press and hold
key for 3 seconds. The gauge will beep, two single tones followed by a double tone.
The Elcometer 456 the last operation unless the Auto Switch Off time is changed (MENU/SETUP/AUTO SWITCH OFF). The Auto Switch Off feature can be set to a maximum of 10 minutes or can be deactivated - see “AUTO SWITCH OFF:” on page 21.
2
switches itself off 60 seconds after
3 seconds
R
9

2.7 The screen

Familiarise yourself with the Elcometer 4562 screen. The screen displays:
• Welcome information
• Measurement information
• Menus to configure the gauge and control functions
• Help and miscellaneous information
When the gauge is switched on a welcome information screen may be displayed briefly (Figure 4).
2
Figure 4. Typical Elcometer 456
The screen where readings are displayed is called the Reading Screen (Figure 5). The reading screen can also include statistics if this option is selected.
welcome screen
Figure 5. Typical Elcometer 456
10
2
reading screen

2.8 Selecting a language

The Elcometer 4562 has over 20 built-in languages. When the gauge is switched on for the first time after dispatch from the Elcometer factory the display will show the language selection screen (Figure 6).
Figure 6. Language selection screen
2.8.1 At first switch on
1. Press Up/Down softkeys to locate language
required.
2. Press SEL softkey to activate the selected
language.
The screen displays an Elcometer 456 screen (Figure 4) followed by the reading screen (Figure 5). The gauge operates in the new language until changed.
2
welcome
2.8.2 At any time
1. Switch gauge off.
2. Press and hold left hand softkey.
3. Press key to switch on gauge.
The display will show language selection screen with current language highlighted by cursor.
4. Release left hand softkey.
Follow the instructions given in 2.8.1.
Alternatively, select MENU/SETUP/LANGUAGES - see “Main MENU - Extended menu on” on page 18.
11

2.9 Interfaces

The gauge is fitted with an infrared interface (Figure 7) which allows information to be sent to a suitable printer.
RS232 5-pin connector
Figure 7. Infrared and RS232 interfaces
The RS232 5-pin connector next to the infrared interface is used with an optional PC connection cable to connect the gauge to a PC. This enables personalised screens to be downloaded to the gauge (see “Personalised welcome screen” on page 38).
Infrared interface
12

3 TAKING A READING

3.1 Before you start

Are you using the correct type of probe?
See “Probes” on page 35.
Is the probe calibrated?
See “Calibration adjustment” on page 23.
Do you require statistics?
See “Statistics” on page 32.
What units of measurement do you want to use?
See “UNITS:” on page 21.

3.2 Procedure

1. Press key to switch on gauge.
2. Place probe on surface to be measured. The
reading may be inaccurate if the probe is not held as shown in Figure 8.
09 : 30 09 / 1 0 / 20 00
F1
F
456
µm
R
CAL DATA STATS MEN U
R
Separate probe Integral probe
Figure 8. Taking a reading
3. Reading is displayed on screen (Figure 9).
Figure 9. Typical reading
13

4 THE READING SCREEN AND MENUS

4.1 Reading screen

The content of the reading screen (Figure 10, Figure 11) depends upon the type of measurement being made and how the gauge is set up.
Automatic-switching probe symbol (FNF probes only)
Battery symbol
Cal Method symbol
Figure 10. Example of reading screen with smooth
surface calibration method selected
CAL.
Operates selected calibration adjustment method.
MENU.
This opens the main MENU of the gauge and provides access to user-selectable features - see page 18.
Note: If CAL softkey symbol is flashing the gauge should be recalibrated. This is due to the calibration adjustment method having been changed or a probe change - see “Calibration adjustment” on page 23.
Figure 11. Reading screen in extended mode and
showing full set of statistical values.
Substrate
Units
Softkey function/ symbol
14

4.2 Main MENU

Gauge configuration and measurement functions are controlled using menus (Figure 12). The structure of the menus is shown on page 18.
Menu title
Menu contents
Softkey functions
2
Figure 12. Typical Elcometer 456
Some screens allow the status of a feature to be changed e.g. on to off or select or deselect, etc. A tick box indicates this type of feature. A tick against a menu item indicates the function is activated or selected.
SEL softkey selects the option displayed and in some cases toggles the status of a tick box off/on.
Up/Down softkeys K L move the cursor to the menu item required. The menus scroll up/down and a line across the screen indicates the start and end of the menu.
BACK softkey returns the gauge to a previous screen. Holding this softkey down will rapidly exit from any menu and return to the reading screen.
4.2.1 Simple and extended menus
The Elcometer 456 structures:
Extended menu off (simple menu mode): The gauge is shipped from the Elcometer factory with EXTENDED MENU turned off. In this simple menu mode the gauge can be calibrated and used to take measurements. This is the ideal setting for users who do not require access to advanced features of the gauge.
Extended menu on (extended menu mode): Additional items are automatically added to the MENU and the STATS softkey is activated. These give access to more advanced functions such as statistics, calibration method, print/output, setup, etc.
2
Basic Gauge has two menu
menu
15

4.3 Main MENU - Extended menu off

Figure 13. Main menu - extended menu off
4.3.1 BACKLIGHT
Switches backlight on and off. Toggle tick box to activate/deactivate. With BACKLIGHT activated the display is illuminated for approximately 5 seconds when a reading is taken or a key pressed.
Note: The battery life is reduced by about one third when the backlight is activated.
4.3.2 CALIBATION LOCKED
Protects against inadvertent calibration adjustment. Toggle tick box to activate/deactivate. If CAL softkey is pressed while CALIBRATION LOCKED is activated the gauge displays CALIBRATION LOCKED USE MENU
TO UNLOCK. The message disappears after 3 seconds.
4.3.3 Extended MENU
Provides access to additional features. Toggle tick box to activate/deactivate. See “Main MENU - Extended menu on” on page 18.
4.3.4 ABOUT
Provides information on Gauge, Probe, Contact information and Help (Figure 14):
GAUGE INFORMATION:
software versions, etc.
PROBE INFORMATION:
CONTACT:
if programmed, the contact details for the Supplier or Local Distributor.
Details of Elcometer offices worldwide and,
Elcometer 4562 model,
Probe type, range, etc.
16
HELP:
Explains symbols used on Elcometer 456
display screens.
Figure 14. About menu
4.3.5 RESET
Selects Factory Calibration or Gauge resets. The RESET menu option (Figure 15) allows one of three gauge resets to be selected:
FACT ORY CA L:
settings created at time of manufacture of the probe.
Factory calibration will not necessarily restore precise calibration values. The calibration of the gauge should be adjusted before use, or at least checked to ensure that it has been previously adjusted correctly for the conditions of use.
INTL GAUGEa:
settings e.g. DD/MM/YY date format and metric units.
US GAUGEb:
e.g. MM/DD/YY date format and imperial units.
Returns gauge to calibration
Resets gauge to International default
Resets gauge to USA default settings
2
A confirmation screen will be displayed. Press YES to reset, NO to cancel.
a. International settings can also be activated at
switch on. Press and hold softkey 3 and switch on gauge.
b. US settings can also be activated at switch on.
Press and hold softkey 4 and switch on gauge.
Figure 15. Reset menu
17

4.4 Main MENU - Extended menu on To toggle EXTENDED MENU on/off select

MENU/EXTENDED MENU/SEL
PRINT /OUTPUT
Current statistics
DELETE
MENU
Backlig ht Calibra tion locked
Extended menu Print/ Output Delete Cal method Setup About Reset
STATS
Enlarge stats Stats on LCD Display F1
Clear stats Select s tats
CAL
Probe calibration
 z z z z z z
Last reading
CAL METHOD
Smooth 2 Point
Rough Special su b Zero offset Set offse t
SETUP
Statist ics
Probe Units Output Beep volume 3 Langu ages Auto sw itch off 1 Opening screen
ABOUT
Gauge information Probe information
Contact Help
RESET
Factory cal Intl g auge US gauge
    
z
z
  
18
STATISTICS
Softkey enabled Select stats Display F1
OUTPUT
Infra red printer
 z
SELECT STATS
No. of readings Mean
Std devia tion Coef of variat’n Highest reading Lowest reading
     
19
The following features are added to the MENU when EXTENDED MENU is active:
4.4.1 PRINT/OUTPUT
Outputs the statistical summary to the infrared interface for printing on a suitable printer.
The Elcometer 456 Infrared printer interface and is programmed to work with the Elcometer Infrared Portable Printer.
Other Infrared printers with different print protocols may not be supported. Consult your local Elcometer supplier.
2
Basic version is fitted with an
To use this function first setup to print to infrared printer:
MENU/SETUP/OUTPUT/INFRA RED PRINTER - see “SETUP” on page 21.
Elcometer 456 summary to be printed to the infrared printer.
2
Basic allows only the current statistical
4.4.2 DELETE
Deletes last reading only. Gauge displays ARE YOU SURE? (Figure 16).
Figure 16. Delete last reading confirmation screen
Press NO softkey to include reading in statistical summary or YES softkey to delete reading.
If there is no reading to delete, the gauge displays:
Figure 17. Delete last reading not available screen
Press OK to return to Delete menu.
20
4.4.3 CAL METHOD
Allows selection of calibration method from list of options
- see “Calibration method” on page 23 for more details.
4.4.4 SETUP
Used to select, change or activate gauge features:
STATISTICS:
Stats softkey on/off.
Select stats - see “Select stats” on page 34.
Display - see “Display” on page 33.
PROBE:
Changes probe mode.
FNF probes - select from Automatic, F or N.
F1 2 probes - select from F1 or F2.
UNITS:
however the user can manually override the automatic setting. Select from µm, mm, mil, thou or inch.
OUTPUT:
interface. Toggle tick box to activate/deactivate. When activated, readings are sent to the infrared portable printer as they are taken. See “Infrared portable printer” on page 44 for sales part number of this optional accessory.
BEEP VOLUME:
•0 = off
• 5 = loudest
• Default = 3
LANGUAGES:
AUTO SWITCH OFF:
switches off when displaying Reading Screen.
• Minimum = 1 minute
• Maximum = 10 minutes
• Default = 1 minute
Auto switch off may be disabled by selecting ‘off’ (In this case, switch off using On/Off key .)
Activates simple statistics feature.
Only for dual function probes (FNF and F1 2).
Units are automatically set by the probe type,
Activates data output via the infrared
Changes volume.
Allows selection of language.
Changes delay before gauge
21
OPENING SCREEN:
screens so that the gauge switches on to display the reading screen.
If a personalised welcome screen has been downloaded into the gauge (see page 38), Opening Screen must be activated to display this screen.
Disables the opening (welcome)
22

5 CALIBRATION ADJUSTMENT

Calibration adjustment is the process of setting the gauge to known values of thickness to ensure accuracy on different substrate types, shapes and surface finishes.
Note: When using an FNF probe it must be calibrated in both the ferrous mode and in the non-ferrous mode to ensure accuracy of reading.

5.1 Calibration method

The calibration of the gauge can be adjusted (MENU/CAL METHOD) using several different methods described in National and International Standards.
The calibration adjustment method chosen is dependant on the condition of the substrate to be measured and is indicated on the screen by a symbol (Figure 18):
Figure 18. CAL METHOD screen
SMOOTH.
is set to zero on the uncoated surface and a known thickness above the expected thickness of the coating.
2POINT.
either side of the expected thickness. This enhances the accuracy of the gauge over the thickness range defined by the two values.
ROUGH.
enhances the accuracy of the gauge over the thickness range defined by the two values.
SPECIAL SUB.
calibration for unusual substrate materials such as cast iron, certain types of stainless steel, high carbon steel, special aluminium alloys, etc.
ZERO OFFSET.
ISO 19840 for coatings on steel surfaces roughened by
Smooth surface calibration where the gauge
Calibration on a thin value and a thick value
A calibration method similar to 2-Point. This
This method uses the 2-Point
This is the method described in
23
blast cleaning. The calibration uses the smooth surface technique, and a correction value (zero offset) is applied to each reading to account for the effect of the roughened surface; the value depends on the surface profile - see Table 1.
SET OFFSET.
for different surface roughness This value is used only with the Zero Offset calibration method.
Table 1: Correction values as detailed in ISO 19840
Note: When the calibration method is changed, e.g. from
Smooth to Rough, the gauge will display a message (Figure 19).
If the NO softkey is pressed the CAL softkey symbol on the Readings Screen will flash to warn that calibration adjustment is still required.
If the YES softkey is pressed the calibration adjustment procedure is activated - see “Calibration adjustment procedure” on page 26.
This screen sets and changes the offset
Profile according to ISO 8503-1
Fine 10
Medium 25
Coarse 40
Figure 19. Recalibration required screen
Correction Value (µm) (Zero Offset)

5.2 Calibration foils and standards

Calibration adjustment should be carried out with the appropriate probe on the same type of metal, the same curvature and similar finish to the item to be measured.
24
It is best to use an uncoated sample of the items to be tested.
Calibration can be carried out using measured foils or coated standards.
Foils (shims) .
which have been measured using techniques independent of the gauge. They are ideal for calibration because they provide a known value of thickness on the actual substrate to be measured. Calibration certificates for foils are available upon request.
When using foils care must be taken to keep the foils clean and free from dust and to avoid damage by creasing particularly the thinner foils. Always remove a foil from its storage wallet before use.
When calibrating a High Temperature PINIP™ Probe use the special thickness standards supplied with the probe - see “Calibrating High Temperature PINIP™ Probes” on page 31.
Note: To calibrate 5 mm (200 mils) and 13 mm (500 mils) range gauges it will be necessary to stack the foils (Figure 20). Care must be taken to avoid errors due to placing the foil labels between the foils.
These are coating thickness standards
Serial no.
TK1009
Inspected by
micron mils/thou
502 19.75
TK1010
micron mils/thou
Serial no.
Inspected by
176.1 6.93
Figure 20. Stacking foils to increase thickness
Coated standards.
substrate materials coated with hardwearing materials and measured using techniques independent of the gauge.
Thickness standards using typical
25
Coated standards are most often used to confirm that the gauge meets its specifications if it is not possible to use foils (shims).

5.3 Calibration adjustment procedure

Calibration adjustment can be carried out at any time by pressing CAL softkey from the reading screen. To prevent inadvertent calibration adjustment the CAL softkey can be locked (MENU/CALIBRATION LOCKED).
The user is guided through the operation of the chosen calibration procedure by means of instructions and illustrations on the graphics screen. Audible warnings are also provided when action is required, e.g. when the probe must be placed down to get a reading.
If the routine is interrupted in any way the previous settings will be restored until after the full calibration routine has been completed or the reset has been completed.
The screen detail depends on the calibration method chosen, but the calibration is in two steps.
The following example is for a Smooth Calibration adjustment.
Step 1
1. Hold probe in air and press CAL softkey (Figure 21).
Figure 21. Step 1 - on thickness standard
ESC softkey returns the gauge to the Reading
Screen from the Calibration Procedure without making any changes. TEST softkey allows the user to take readings to verify the accuracy of the current calibration. These readings do not affect statistical calculations.
26
2. Place probe on calibration standard. The gauge will
display a reading.
3. Lift probe and then replace on calibration standard.
Gauge displays the average ( ) of these readings and the last reading. Repeat this action until a stable reading is obtained.
Average
First reading Second reading
Figure 22. Step 1 - Calibration adjustment on
To reject the displayed reading and start the calibration procedure again, press both the Up and Down softkeys at the same time. To adjust the displayed reading until it is correct relative to the thickness standard use the Up/Down softkeys.
4. Press SET softkey to accept the value.
Note: - - - indicates over-range (Figure 23). Taking a reading within range clears this screen.
Step 2
1. Place probe on uncoated standard or zero plate.
The gauge will take and display a reading.
2. Lift probe and then replace on uncoated standard or
zero plate. Gauge displays the average ( ) of these
thickness standard
Figure 23. Over-range reading
Last
27
readings and the last reading. Repeat this action until a stable reading is obtained.
First reading Second reading
Figure 24. Step 2 - Calibration adjustment on
To reject the displayed reading and start Step 2 of the calibration procedure again, press the Reset softkey .
3. Press ZERO softkey to zero the display (Figure 25).
4. Press SET softkey to accept this value.
The gauge will display the option to test the calibration of the gauge.
5. Either press NO softkey to complete the calibration
adjustment procedure and return the gauge to the reading screen, or proceed to take test readings ­see “Taking test readings” on page 29.
uncoated sample
Figure 25. Zero the display
Figure 26. TEST READINGS screen
28

Taking test readings

Press YES softkey (see previous section) to take test readings. This allows the calibration of the gauge to be tested without contributing to the statistical calculations.
Figure 27. TEST READINGS screen
CAL softkey returns the gauge to Step 1 of the
calibration adjustment procedure.
OK softkey completes the calibration adjustment procedure and returns the gauge to the reading screen.

Other calibration methods

For the 2-POINT, ROUGH and SPECIAL SUBSTRATE calibration methods Step 2 requires readings to be taken on a thin standard value instead of on an uncoated base.
Figure 28. Step 2 - On thin standard
Repeating the reading will display the average. This is particularly useful for rough surfaces as it allows variations in the surface to be accounted for in the
29
calibration adjustment, therefore improving the accuracy of the gauge.
First reading Second reading
Figure 29. Step 2 - Calibration adjustment on thin
standard
30

Calibrating High Temperature PINIP™ Probes

Special thickness standards are supplied with F1 2 High Temperature PINIP™ Probes - see page 37. These thickness standards should be used in place of the calibration standard in Step 1 of “Calibration adjustment procedure” on page 26.
1. Place appropriate thickness standard over end of
PINIP™ probe.
2. Press CAL softkey.
3. Place probe on hot
4. Lift probe and then replace on hot surface to take
second reading.
5. Repeat as necessary until reading is stable.
6. Press SET to accept value.
7. Remove thickness standard from end of PINIP™
probe.
8. Proceed with Step 2 - see page 27.
c
surface and take a reading.
c. The temperature of the surface used for
calibration should be equal to the temperature of the substrate being measured.
31

6 STATISTICS

The Elcometer 4562 Basic has a Simple Statistics feature (STATS) which calculates and displays a statistical analysis of readings as they are taken.
When EXTENDED MENU is active (MENU/EXTENDED
MENU/SEL), press STATS softkey to access STATS MENU (Figure 30).
Figure 30. Stats menu
The statistical values available are:
• Number of Readings
• Mean
• Standard Deviation
• Coefficient of Variation
• Highest Reading
• Lowest Reading
See “Select stats” on page 34 and “Statistics terminology” on page 40.
If you require statistics in batches, or memory for readings, Elcometer 456 provide this facility. Contact Elcometer or your local Elcometer supplier for more details.
2
models Standard and Top

6.1 ENLARGE STATS

Displays the chosen statistical values as double-height characters. The example screens (Figure 31) appear when all the statistical values are selected. The
32
Up/Down softkeys can be used to move through the list. OK softkey returns to the Reading Screen.
Figure 31. Enlarged statistics

6.2 STATS ON LCD

Activates the presentation of the chosen statistical values on the reading screen.
Combined statistics symbol
Figure 32. Reading screen with all statistics

6.3 DISPLAY

Only applies when using dual function probes. Allows selection of the types of readings used in the statistical calculation when a dual function probe is connected.
Probe Options
FNF F, N or F and N combined
F1 2 F1, F2 or F1 and F2 combined
d
d

6.4 CLEAR STATS

Resets to zero all statistical values selected in STATS MENU/DISPLAY.
d.When readings are combined a symbol will be
displayed on the Reading Screen (Figure 32).
33

6.5 SELECT STATS

Allows the user to chose which statistical values are displayed. The default condition is all values (Figure 33).
Figure 33. Select stats menu
Use Up/Down softkeys to move cursor and SEL softkey to select or deselect the statistical values.
34

7 PROBES

An extensive range of probes is available for the Elcometer 456 ferrous (F), non-ferrous (N) and dual ferrous/non-ferrous (FNF) operation are available as either integral (built-in) or separate options.
Separate probes are fully interchangeable and are available in standard, PINIP™ and miniature formats.
PINIP™ format (Plug-In Integral Probe) is an integral style probe which plugs in to a separate gauge. This provides all the benefits of an integral gauge and the flexibility of a separate gauge in a single unit.
Miniature separate probes permit measurements where space is restricted.
2
Coating Thickness Gauge. Probes for

7.1 Ferrous (F) probes

F probes measure the thickness of non-magnetic
coatings on magnetic substrates. They can be used on paint, plastic, galvanising, enamel, powder paint, hard chrome and other coatings such as electro-less nickel applied to steel or iron.

7.2 Non-ferrous (N) probes

N probes measure the thickness of non-magnetic
coatings on non-magnetic metals. They can be used on anodising, paint, plastic coatings, powder paint, etc. applied to aluminium, brass, non-magnetic stainless steel, etc.
e

7.3 Dual ferrous/non-ferrous (FNF) probes

FNF probes are dual function, F and N in one probe. FNF
gauges will automatically detect the type of substrate and set the mode accordingly. Alternatively the mode can be set manually - see “PROBE:” on page 21 and “Coatings on galvanised or metallised steel” on page 37.
e. Using an N probe (or an FNF probe manually set
to N1) on a ferrous substrate will give a reading, but the reading will be incorrect.
35

7.4 Probe interchangeability

The following table shows which probes can be used in the three types of Elcometer 456
2
Gauge.
Probe type Gauge Type
FNFNF
F1 * 989 F2 * 989 F1 2 * 989 F3 * 989 F1 right angle 989 F2 right angle 989 F1 2 right angle 989 F1 telescopic 989 F2 telescopic 989 F1 PINIP™ 989
FERROUS
F2 PINIP™ 989 F1 2 PINIP™ High Temp 989 F3 PINIP™ 989 F3 miniature straight 989 F3 miniature 45° 989 F3 miniature 90° 989 N1 * 899 N2 899 N1 right angle 899 N1A anodiser’s 899 N1 PINIP™ 899 N3 miniature straight 899
NON-FERROUS
N3 miniature 45° 899 N3 miniature 90° 899 FNF1 * 889 FNF1 right angle 889
DUAL
FNF PINIP™ 889
* indicates probes available for integral type gauges.
36

7.5 F1 2 probes

The F1 2 scale combines the F1 scale with the F2 scale in a single probe. The user selects the appropriate range for the work in hand. The resolution of the gauge is dependent on the scale selected on the gauge.

7.6 F1 2 High temperature PINIP™ probes

These probes are capable of measuring on surfaces up to 250°C (480°F). Wear appropriate protective clothing and take care to avoid bodily contact with the hot
Do not exceed the maximum measurement speed when using these probes - see page 41.
Note: Refer to special calibration procedure - see “Calibrating High Temperature PINIP™ Probes” on page 31.
surface during measurement.

7.7 Coatings on galvanised or metallised steel

The FNF probe in fixed N1 mode may be used to measure paint coatings on galvanised, aluminium (Al) or zinc (Zn) sprayed steel substrates.
1. Set the gauge to the N1 mode
(MENU/SETUP/PROBE).
2. Zero and calibrate the gauge on a sample of the
coated steel - see “Calibration adjustment” on page 23.
Care must be taken to ensure that the calibration conditions are not affected by changes in the zinc or aluminium coating thickness. This can be determined by checking the zero over an area of the galvanised or metal-coated steel. Metal coatings on steel above 50 µm (2 mil/thou) should be consistent enough to obtain a stable zero on the layer of metal.
3. Take readings.
37

8 PERSONALISED WELCOME SCREEN

A personalised welcome screen can be created and downloaded into the gauge.
Screen dimensions are 128 pixels x 64 pixels. The welcome screen is typically used to personalise the gauge with a logo, serial number, user name, etc. This is the first screen displayed when the gauge is switched on.

8.1 Creating the screen

1. Download Elcometer ‘Welcome Screen Wizard’
software. This software is available free of charge from the downloads section of the Elcometer website, www.elcometer.com.
2. Connect gauge to PC using optional 456 to PC
connection cable - see “PC Connection Cable” on page 44.
3. Press key to switch on gauge.
4. Ensure Reading Screen is displayed.
5. Run ‘Welcome Screen Wizard’ software and follow
the on-screen instructions.

8.2 Deleting the screen

1. Run the ‘Welcome Screen Wizard’ software.
2. Click ‘Next’.
3. Select ‘Create a new screen setup’.
4. Click ‘Next’.
5. Select ‘Disabled’.
6. Click ‘Next’.
Follow the remaining on-screen instructions to delete the welcome screen.
38

9 STORAGE AND TRANSIT

This gauge incorporates a Liquid Crystal Display (LCD). If the display is heated above 50°C (120°F) it may be damaged. This can happen if the gauge is left in a car
Always store the gauge in its carrying pouch when it is not being used.
Remove the batteries from the gauge and store them separately if the gauge is to remain unused for a long period of time. This will prevent damage to the gauge in the event of malfunction of the batteries.
parked in strong sunlight.

10 MAINTENANCE

You own one of the finest hand-held coating thickness gauges in the world. If looked after, it will last a lifetime.
The gauge does not contain any user-serviceable components. In the unlikely event of a fault, the gauge should be returned to your local Elcometer supplier or directly to Elcometer. Contact details are stored in the gauge - MENU/ABOUT/CONTACT.
Worldwide: sales@elcometer.com
Or USA/Canada: inc@elcometer.com
Note: Probes will eventually wear. Probe life will depend on the number of measurements taken and how abrasive the coating is. Probe life can be prolonged by careful positioning of the probe on the surface.
Replacement separate and PINIP™ probes can be fitted by the user without the need to return the gauge for service.
Gauges with an integral probe have to be returned for re-programming or replacement if the probe becomes worn or damaged.
39

11 STATISTICS TERMINOLOGY

Term Meaning
COEF OF VARIAT’N Coefficient of Variation. The
HIGHEST READING The value of the maximum
LOWEST READING The value of the minimum
MEAN The average of a group of
NO. OF READINGS Number of Readings. The
STD DEVIATION Standard Deviation. A
standard deviation divided by the mean for a group of readings, expressed as a percentage.
thickness in a group of readings.
thickness in a group of readings.
readings; the sum of the individual readings divided by the number of readings.
running value for the number of readings taken in a group. In the case of the averaging or counted average, the Number of Readings is the number of values recorded, not the total number of readings taken.
statistical measure of the spread of values in a group of readings.
40

12 TECHNICAL DATA

12.1 Measurement speed

>60 readings per minute.
This rate is reduced to >40 readings per minute when Infrared printer output is active - see “OUTPUT:” on page 21.
When measuring high temperature materials measurement speed must be reduced to prevent overheating of the probe. The maximum measurement speed of the High Temperature PINIP™ probe at 250°C (480°F) is 4 readings per minute.

12.2 Minimum substrate thickness

Ferrous: 300 µm (12 mils)
Non-ferrous: 100 µm (4 mils)
Measurements can be taken on thinner substrates if 2-point calibration is carried out either side of the required substrate thickness, however gauges will have reduced range when adjusted for thin substrates.

12.3 Probe operating temperature

Separate ferrous probes: 150°C (300°F)
High temperature PINIP™ probes: 250°C (480°F)
Miniature probes without outer sleeve: 150°C (300°F)
All other probes: 80°C (176°F)

12.4 Physical

Weight (including batteries):
Dimensions: 130 mm x 70 mm x 35 mm
Gauge operating temperature:
Case: High impact ABS
Separate Probe (FNF1), 190g (6.7oz) Separate Probe (PINIP™), 155g (5.5oz) Integral Probe, 130g (4.6oz)
(5.12" x 2.76" x 1.38")
0°C to 50°C (32°F to 120°F) Operation outside these limits depends upon climatic conditions.
41

12.5 Power supply

Internal batteries, 2 x LR03 (AAA), alkalinef dry batteries or rechargeable
Battery life
h
hours to 40 hours continuous use with alkaline dry
30 batteries. (15 000 to 20 000 readings at an average of 8 readings per minute.) Battery life is reduced by one third when using the backlight.
g
equivalents.

12.6 Packaging

The gauge is packed in cardboard and plastic packaging. Please ensure that this packaging is disposed of in an environmentally sensitive manner. Consult your Local Environmental Authority for further guidance.
f. Alkaline batteries must be disposed of carefully to
avoid environmental contamination. Please consult your local environmental authority for information on disposal in your region.
Do not dispose of any batteries in fire.
g.Rechargeable batteries can be used if they are
charged outside the gauge.
h.Battery life is reduced to approximately 25% of dry
battery life when using rechargeable batteries. Follow the instructions provided by the battery manufacturer when charging and disposing of rechargeable batteries.
42

13 ACCESSORIES

The Elcometer 4562 is complete with all the items required to get started and take measurements.
Many of the following accessories are optional. However, some are consumable items that may need to be replaced over the lifetime of the gauge.
All these accessories are available from Elcometer, or your local Elcometer supplier. At time of ordering please quote the sales part number which follows the description of each accessory.

13.1 Consumable items

Alkaline LR03 (AAA) Battery (2 required):
T9999329-

13.2 Probes

Full details of the extensive range of 456 probes can be obtained from Elcometer, your local Elcometer supplier or the Elcometer website, www.elcometer.com.

13.3 Foil sets

2.2 mm (85 mils) 8 pieces: T9904199F
1.3 mm (51 mils) 3 pieces: T9904199G
5.5 mm (220 mils) 4 pieces: T9904199J 15 mm (595 mils) 4 pieces: T9904199K
Individual foils in the range 12.5 µm to 20 mm (0.5 mil to 790 mils) and customised sets chosen from this range are also available. Consult your local Elcometer supplier.

13.4 Calibration certificates for foils

Certificates traceable to National Standards including UKAS and NIST are available on request.

13.5 Test certificates

A certificate with results of a standard test on known foil values over the full range of the probe. Order using sales part number TEST-456.
43

13.6 Coated thickness standards including certificate

Ferrous Standard (4 Values): T995111261 Non-Ferrous Standard (4 Values): T995111271 Ferrous Standard (2 Values): T995166001 Non-Ferrous Standard (2 Values): T995166011

13.7 Probe placement jig

To aid probe positioning on small components a probe placement jig is available and an adapter suitable for use with the full range of miniature probes is also available as an accessory.
Probe placement jig: T95012880

13.8 Probe adapters

Jumbo Hand Grip (F and N probes): T9997766­Jumbo Hand Grip (FNF probes): T99913225 V Adapter for pipes (F & N probes): T9997381­V Adapter for pipes (FNF probes): T99913133

13.9 Infrared portable printer

Infrared Portable Printer: X45613877 Thermal Printer Paper: T99913878

13.10 Bench Stands

Integral Probe Version: T45616161 Integral/Separate Probe Version: T45616162

13.11 PC Connection Cable

456 to PC Connection Cable (9-pin): T45616217
Note: A 9-pin to 25-pin adapter may be required for certain PC RS232 ports.
44

14 RELATED EQUIPMENT

Elcometer produces a wide range of coating thickness gauges and associated paint inspection equipment. Users of the Elcometer 456 following Elcometer products:
2
may also benefit from the
• Uncured powder thickness gauges
• Coatings analyser
• Inspection management software
• Mechanical coatings thickness gauges
• Appearance testers
• Adhesion testers
For further information contact Elcometer, your local Elcometer supplier or visit www.elcometer.com

15 FITTING THE WRIST HARNESS

1. Pass harness
round pin
2. Pass harness
through loop
3. Pull tight
45
a
Resolution in range
0 µm to 99.9 µm
100 µm to 1500 µm
0 mil to 4.99 mils
5 mils to 60 mils
0 mm to 0.99 mm
1.0 mm to 5.0 mm
0 mil to 49.9 mils
50 mils to 200 mils
0 mm to 1.99 mm
2 mm to 13 mm
0 mil to 99.9 mils
100 mils to 500 mils
0 µm to 99.9 µm
100 µm to 500 µm
0 mil to 3.99 mils
4 mils to 10 mils
1.0 µm
0.1 mil
10 µm
1mil
10 µm
1mil
1.0 µm
0.1 mil
0 µm to 1500 µm ±1% to ±3% or ±2.5 µm 0.1 µm
0 mil to 60 mils ±1% to ±3% or ±0.1 mil 0.01 mil
0 mm to 5.0 mm ±1% to ±3% or ±0.02 mm 1.0 µm
0 mil to 200 mils ±1% to ±3% or ±1 mil 0.1 mil
0 mil to 500 mils ±1% to ±3% or ±2.0 mils 0.1 mil
0 µm to 500 µm ±1% to ±3% or ±2.5 µm 0.1 µm
0 mil to 10 mils ±1% to ±3% or ±1.0 mil 0.01 mil

16 PROBE MEASUREMENT PERFORMANCE

46
Scale Total range Accuracy
F1
F1 2 (F1 mode)
FNF1
N1, N1A
F2
F1 2 (F2 mode)
N2
F3 0 mm to 13 mm ±1% to ±3% or ±0.05 mm 1.0 µm
FM3
a.Whichever is the greater. Lower value achieved when calibrated close to the thickness to be measured.
NM3
a
Cal foil value
Minimum
sample
Headroom
Minimum
concave
Minimum
convex surface
diameter
surface radius
4 mm (0.16”) 25 mm (0.98”) 130 mm (5.1”) 4 mm (0.16”) 250 µm (10 mil)
4 mm (0.16”) 25 mm (0.98”) 135 mm (5.3”) 8 mm (0.32”) 1 mm (40 mil)
diameter
(or F1 2 set for F1
(or F1 2 set for F2
F2
F3 15 mm (0.59”) 40 mm (1.57”) 150 mm (5.9”) 14 mm (0.55”) 2.5 mm (100mil)
N1 (N) 35 mm (1.38”) 25 mm (0.98”) 130 mm (5.1”) 6 mm (0.24”) 250 µm (10 mil)
F1
Probe type
operation)

17 PROBE CAPABILITIES

17.1 Integral probes

operation)
FNF1 (N) 38 mm (1.50”) 25 mm (0.98”) 135 mm (5.3”) 8 mm (0.32”) 250 µm (10 mil)
measurements conditions
a.This is the recommended maximum calibration foil value to achieve the specified accuracy under these
FNF1 (F) 4 mm (0.16”) 25 mm (0.98”) 135 mm (5.3”) 4 mm (0.16”) 250 µm (10 mil)
47
a
Cal foil value
Minimum
sample
Headroom
Minimum
concave
Minimum
convex surface
diameter
surface radius
diameter
4 mm (0.16”) 25 mm (0.98”) 28 mm (1.10”) 4 mm (0.16”) 250 µm (10 mil)
4 mm (0.16”) 25 mm (0.98”) 32 mm (1.26”) 8 mm (0.32”) 1 mm (40 mil)
48
(or F1 2 set to F1) 4 mm (0.16”) 25 mm (0.98”) 85 mm (3.35”) 4 mm (0.16”) 250 µm (10 mil)
(or F1 2 set to F2) 4 mm (0.16”) 25 mm (0.98”) 89 mm (3.50”) 8 mm (0.32”) 1 mm (40 mil)
F2
F1 Right Angle
F2 Right Angle
F1 Telescopic 4 mm (0.16”) 25 mm (0.98”) 32 mm (1.26”) 4 mm (0.16”) 250 µm (10 mil)
F1
Probe type

17.2 Separate ferrous probes

(or F1 2 set to F1)
F2 Telescopic 4 mm (0.16”) 25 mm (0.98”) 36 mm (1.42”) 8 mm (0.32”) 1 mm (40 mil)
(or F1 2 set to F2)
F3 15 mm (0.59”) 40 mm (1.57”) 102 mm (4.02”) 14 mm (0.55”) 2.5 mm (100 mil)
measurements conditions
a.This is the recommended maximum calibration foil value to achieve the specified accuracy under these
a
Cal foil value
Minimum
sample
diameter
Headroom
Minimum
concave
surface radius
Minimum
convex surface
diameter
Probe type

17.3 Separate non-ferrous probes

35 mm (1.38”) 25 mm (0.98”) 85 mm (3.35”) 6 mm (0.24”) 250 µm (10 mil)
measurements conditions
N1 35 mm (1.38”) 25 mm (0.98”) 85 mm (3.35”) 6 mm (0.24”) 250 µm (10 mil)
N1 Right Angle 35 mm (1.38”) 25 mm (0.98”) 28 mm (1.10”) 6 mm (0.24”) 250 µm (10 mil)
N1A Anodiser’s
Probe
N2 100 mm (3.97”) 150 mm (5.90”) 85 mm (3.35”) 14 mm (0.55”) 1 mm (40 mil)
a.This is the recommended maximum calibration foil value to achieve the specified accuracy under these
49
a
Cal foil value
Minimum
sample
Headroom
Minimum
concave
Minimum
convex surface
diameter
surface radius
38 mm (1.50”) 25 mm (0.98”) 34 mm (1.34”) 8 mm (0.32”) 250 µm (10 mil)
diameter
4 mm (0.16”) 25 mm (0.98”) 34 mm (1.34”) 4 mm (0.16”) 250 µm (10 mil)
50

17.4 Separate dual FNF

measurements conditions
FNF1 (N) 38 mm (1.50”) 25 mm (0.98”) 88 mm (3.46”) 8 mm (0.32”) 250 µm (10 mil)
FNF1 (F) 4 mm (0.16”) 25 mm (0.98”) 88 mm (3.46”) 4 mm (0.16”) 250 µm (10 mil)
Probe type
FNF1 Right Angle
a.This is the recommended maximum calibration foil value to achieve the specified accuracy under these
(N)
FNF1 Right Angle
(F)
a
Cal foil value
Minimum
sample
diameter
Headroom
Minimum
concave
surface radius
Minimum
convex surface
diameter
(or F1 2 set to F1) 4 mm (0.16”) 60 mm (2.36”) 155 mm (6.10”) 4 mm (0.16”) 250 µm (10 mil)
(or F1 2 set to F2) 4 mm (0.16”) 60 mm (2.36”) 159 mm (6.25”) 8 mm (0.32”) 1 mm (40 mil)
F2
F3 15 mm (0.59”) 45 mm (1.77”) 169 mm (6.65”) 14 mm (0.55”) 2.5 mm (100mil)
N1 35 mm (1.38”) 50 mm (1.97”) 155 mm (6.09”) 6 mm (0.24”) 250 µm (10 mil)
F1
Probe type

17.5 PINIP™ probes

FNF1 (N) 38 mm (1.50”) 55 mm (2.17”) 156 mm (6.15”) 8 mm (0.32”) 250 µm (10 mil)
measurements conditions
a.This is the recommended maximum calibration foil value to achieve the specified accuracy under these
FNF1 (F) 4 mm (0.16”) 55 mm (2.17”) 156 mm (6.14”) 4 mm (0.16”) 250 µm (10 mil)
51

17.6 Separate miniature ferrous probes

52
Overall length
(headroom)
Minimum
access
width
Minimum
access
height
Minimum
sample
diameter
Minimum
concave
surface
radius
Minimum
convex
surface
diameter
Probe Type
1.5 mm (0.06”) 6.5 mm (0.26”) 3 mm (0.12”) 6 mm (0.24”) 150 mm (5.91”)
1.5 mm (0.06”) 6.5 mm (0.26”) 3 mm (0.12”) 6 mm (0.24”) 260 mm (10.24”)
1.5 mm (0.06”) 6.5 mm (0.26”) 3 mm (0.12”) 18 mm (0.71”) 7 mm (0.28”) 145 mm (5.71”)
1.5 mm (0.06”) 6.5 mm (0.26”) 3 mm (0.12”) 18 mm (0.71”) 7 mm (0.28”) 250 mm (9.84”)
1.5 mm (0.06”) 6.5 mm (0.26”) 3 mm (0.12”) 16 mm (0.63”) 7 mm (0.28”) 140 mm (5.51”)
1.5 mm (0.06”) 6.5 mm (0.26”) 3 mm (0.12”) 16 mm (0.63”) 7 mm (0.28”) 245 mm (9.64”)
F, St r a i g h t ,
45 mm (1.77”)
F, St r a i g h t ,
150 mm (5.9”)
F, 45°,
45 mm (1.77”)
F, 45°,
150 mm (5.9”)
F, 90°,
45 mm (1.77”)
F, 90°,
150 mm (5.9”)

17.7 Separate miniature non-ferrous probes

Overall length
(headroom)
Minimum
access
width
Minimum
access
height
Minimum
sample
diameter
Minimum
concave
radius
Minimum
convex
diameter
Probe Type
3 mm (0.12”) 25 mm (0.98”) 4 mm (0.16”) 6 mm (0.24”) 150 mm (5.91”)
3 mm (0.12”) 25 mm (0.98”) 4 mm (0.16”) 6mm (0.24”) 260 mm (10.24”)
3 mm (0.12”) 25 mm (0.98”) 4 mm (0.16”) 18 mm (0.71”) 7 mm (0.28”) 145 mm (5.71”)
3 mm (0.12”) 25 mm (0.98”) 4 mm (0.16”) 18 mm (0.71”) 7 mm (0.28”) 250 mm (9.84”)
3 mm (0.12”) 25 mm (0.98”) 4 mm (0.16”) 16 mm (0.63”) 7 mm (0.28”) 140 mm (5.51”)
3 mm (0.12”) 25 mm (0.98”) 4 mm (0.16”) 16 mm (0.63”) 7 mm (0.28”) 245 mm (9.64”)
NF, Straight,
45 mm (1.77”)
NF, Straight,
150 mm (5.9”)
NF, 45°,
45 mm (1.77”)
NF, 45°,
150 mm (5.9”)
NF, 90°,
45 mm (1.77”)
NF, 90°,
150 mm (5.9”)
53

18 ERROR MESSAGES

54
Under certain conditions the gauge will display error messages (Figure 34). These messages are normally
cleared by pressing any one of the softkeys. The cause of the error will be indicated by the message and
should be corrected before proceeding.
Error message Causes Action to take
Separate gauge - remove probe and refit.
If error persists, return to Elcometer*.
Separate gauge - try new probe.
If error persists, return to Elcometer*.
a) Gauge model F and gauge model N
cannot be used with a dual FNF probe.
b) Return to Elcometer* for software
upgrade.
a) Probe is dual FNF, but gauge is ferrous
only, or non-ferrous only. See page 35.
b) Old gauge software does not support
new probe.
Probe has been changed. Calibration adjustment required.
#1 - PROBE Probe-to-gauge communication failure. Integral gauge - return to Elcometer*.
#2 - PROBE Corrupt data output from probe. Integral gauge - return to Elcometer*.
#3 - PROBE Internal error. Return to Elcometer*.
PROBE
UNSUPPORTED
PROBE
CHANGED
Error message Causes Action to take
a) Isolate from electro-magnetic
interference.
a) External electro-magnetic interference.
b) Gauge faulty.
UNSTABLE
READING
b) Return to Elcometer*.
Numerical error. Switch gauge off then on again.
VALUE TOO
If error persists contact Elcometer.
LARGE
CALIBRATION Incorrect calibration calculation. Re-calibrate.
If error persists, return to Elcometer*.
Software error. Return to Elcometer*.
Figure 34. Example error message - no probe is connected to gauge
LANGUAGE
MEMORY
* Contact Elcometer or your local Elcometer Supplier to arrange return.
55

19 INDEX

Numerics
2 Point calibration method 23
A
Accessories 43 ASTM B 244 4 ASTM B 499 4 ASTM D 1400 4 ASTM D1186 4
B
Batteries
Fitting 6 Life of 42 Ordering 43 Precautions 39 Rechargeable
Charging 42 Life of 6
Specification 42
Beep
Changing volume 21
Switching off 21 BS 3900 (C5) 4 BS 5411 (11) 4 BS 5411 (3) 4 BS 5599 4 BS EN ISO 1461 4
C
Cable
PC to gauge 44 CAL softkey
Flashing 24 Calibration
Adjustment 24
Coated standards 25
Coated thickness
standards 25
FNF probes 23
Foils 25
Locking 16
Methods 23
56
Shims 25
Calibration Adjustment 26
Procedure, 2Point 29 Procedure, Rough 29 Procedure, Smooth 26 Procedure, Special
substrate 29 Calibration certificates 43 Calibration methods
2 POINT 23
ROUGH 23
SET OFFSET 24
SPECIAL SUB 23
ZERO OFFSET 23 Coated thickness standards
25
Spares 44 Coatings
On galvanised steel 37
On metallised steel 37 Coefficient of Variation 40 Computer
Connection cable 44
D
Delete last reading 20 Dimensions 41 DIN 50981 4 DIN 50984 4
E
Elcometer 456
Features 4
Overview 3 Error messages 54 ESC 26 Extended menu
Contents 20
F
Factory calibration 17 FNF 35 Foils 25
Ordering 43
G
Galvanised steel 37 Gauge information 16
H
High temperature 37 Highest reading 40
I
Infrared interface 12, 20
Effect on measurement speed 41
Interface 12
Infrared 12 RS232 12
International gauge settings
17 ISO 19840 4, 24 ISO 2178 4 ISO 2360 4 ISO 2808 4
K
Keypad 8
L
Language, selecting 11 Lowest reading 40
M
Maintenance 39 Mean 40 Measurement speed 41 Menu items
2-POINT 23 ABOUT 16 AUTO SWITCH OFF 21 BACKLIGHT 16
BEEP VOLUME 21 CAL METHOD 21 CALIBRATION LOCKED
16 CLEAR STATS 33 CONTACT 16 DELETE 20 DISPLAY 33 ENLARGE STATS 32 EXTENDED MENU 18 FACTORY CAL 17 GAUGE INFORMATION
16 HELP 17 INTL GAUGE 17 LANGUAGES 21 OPENING SCREEN 22 OUTPUT 21 PRINT/OUTPUT 20 PROBE 21 PROBE INFORMATION
16 RESET 17 ROUGH 23 SELECT STATS 34 SET OFFSET 24 SETUP 21 SMOOTH 23 SPECIAL SUB 23 STATISTICS 21 STATS MENU 32 STATS ON LCD 33 UNITS 21 US GAUGE 17 ZERO OFFSET 23
Menus
Extended off (simple) 16 Extended on 18 Overview 15 Structure 18
Metallised steel 37 Minimum substrate thickness
57
41
N
Number of readings 40
O
On/off 9 Opening screen
Creating 38 Disabling 22
P
Packaging 42 PINIP probes
Fitting of 7
High tempe rature 37 Power supply 42 Printer
Infrared 44
Output to 20 Printing
Statistical summary 20 Printing readings 21 Probe positioning jig 44 Probes 35
Adapters 44
Capabilities 47-53
Extending life of 39
F (Ferrous)
What used for 35
F1 2
High Temperature PINIP 37 Substrate selection
modes 21 F1 2 overview 37 Fitting of 7 FNF (Ferrous/Non-ferrous)
Substrate selection
modes 21, 35
What used for 35 Maximum temperature 41 N (Non-ferrous)
What used for 35 Performance 46
58
PINIP
What used for 35 Replacement of 39 Wear of 39 Which one to use? 36
Q
Quick-start 5
R
Reading screen 10
Typical 14
Reading, taking 13 Resetting the gauge 17 Rough calibration method 23 RS232 interface 12
S
Scale selection, F1 F2 21 Screen
Welcome information 10
Screen symbols
Battery condition 6 Definition of 17
Screen symbols, explanation of 17 Set offset 24 Setting up the gauge 21 Shims 25
Ordering 43
Simple menu
Turning on/off 16
Smooth calibration method 23 Softkey
BACK 15 CAL 14, 29
Flashing 14
Procedure 26 ESC 26 MENU 14 OK 29 SEL 15 SET 27, 28 TEST 26 UP/DOWN 15
ZERO 28
Special sub calibration method
23 SSPC-PA2 4 Standard deviation 40 Standards 4
Coated thickness 44 ISO 19840 24
Statistics
Activating 21 Choice of 34 Clearing 33 Displaying on screen 33 Dual function probes 33 Increasing size on display
32 Overview 32 Terminology 40
Substrate selection
Automatic 21, 35 Manual 21, 35
Substrate thickness
Minimum 41
Switching on/off 9
Automatically 9, 21
T
Temperature
Effect on measurement speed 41 High, meas uring of 37 Operating 41 Probes, maximum 41
Test certificates 43 Test readings 28 Testing calibration of gauge
29
Tick box 15
U
Units
Automatic setting of 21 Manual setting of 21
Up/Down 15 US gauge settings 17
W
Weight 41 Welcome screen
Creating 38 Disabling 22
Wrist harness, fitting of 45
Z
Zero offset calibration method
23
59
Loading...