The Elcometer Rotary Viscometer has been tested in accordance with EU regulations
governing Electro-magnetic compliance and it meets the required directives.
Note: Gauge readings may be affected if the unit is operated within a radio frequency
electromagnetic strength of greater than 3 V/m.
is a registered trademark of Elcometer Instruments Ltd.
All rights reserved. No part of this Document may be reproduced, transmitted, transcribed, stored
(in a retrieval system or otherwise) or translated into any language, in any form or by any means
(electronic, mechanical, magnetic, optical, manual or otherwise) without the prior written permission
of Elcometer Instruments Ltd.
A copy of this Instruction Manual is available for download on our Website via
www.elcometer.com/downloads.
Thank you for purchasing this Elcometer Rotary Viscometer. Welcome to Elcometer.
Elcometer are world leaders in the design, manufacture and supply of inspection equipment for
coatings and concrete. Our products cover all aspects of coating inspection, from development
through application to post application inspection.
The Elcometer Rotary Viscometer is a world beating product. With the purchase of this product you
now have access to the worldwide service and support network of Elcometer. For more information
visit our website at www.elcometer.com
1 ABOUT YOUR VISCOMETER
The Elcometer Rotary Viscometer measures the viscosity of liquids in accordance with ISO 2555
and many ASTM standards.
A spindle fitted with a disc or a cylinder is rotated in the sample to be tested. A spring is connected
between the spindle and the motor shaft which rotates at a fixed speed. The angle of deviation
between the spindle and the motor shaft is measured electronically and converted to torque. The
torque measured by the Viscometer is a function of:
•spindle rotation speed
•spindle geometry (disc/cylinder size)
•viscosity of the test sample
Taking into account the spindle rotation speed and the spindle geometry, viscosity is calculated and
displayed directly in SI or CGS units.
A range of spindles is supplied with the Viscometer; the combination of the appropriate spindle and
spindle rotation speed allows readings to be taken over a very wide range of viscosity.
Optional accessories enable measurement of:
•Low viscosity samples - at temperatures up to 100°C (212°F)
•Low viscosity samples - at temperatures up to 200°C (392°F)
•Small volumes of sample
•Difficult-to-measure samples such as creams, gels, waxes and pastes, etc.
See “Spare parts and accessories” on page 28 for details of all accessories.
R
1.1 These instructions
These instructions describe the operation of the following Elcometer Rotary Viscometers:
•Elcometer 2300 RV1-L (low to medium viscosity tests)
•Elcometer 2300 RV1-R (medium to high viscosity tests)
•Elcometer 2300 RV2-L (low to medium viscosity tests, PC controlled)
•Elcometer 2300 RV2-R (medium to high viscosity tests, PC controlled)
1
R
ABOUTYOUR VISCOMETER
1.2 What the box contains
•Viscometer head
•Column with head mounting bracket
•Spindles L1 to L4 or R2 to R7 (R2 to R7 shown in illustration)
•Base
•Box spanner tool
•Hexagonal wrench
•Spindle guard
•Power supply cable
•Storage rack
•PT100 Temperature probe
•RS 232 connection cable
•ViscosityMaster software
•Calibration certificate
•Operating instructions
•Carrying case
2
ABOUTYOUR VISCOMETER
1.3 Standards
Your Viscometer can be used in accordance with the following National and International
Standards:
•BS 3900 A7-2
•ISO 2884-2
•ASTM D2196
1.4 Features of your Viscometer
•19 speeds from 0.3 rpm to 200 rpm
•PT100 temperature probe
•RS 232 interface (bidirectional on RV2 models)
•Data download software
•Remote control via PC (RV2 models only)
•Audible warning if viscosity reading exceeds the optimum measuring range
•4 line display showing:
• Viscosity (cP or mPa•s)
• Spindle rotation speed
• % scale (Brookfield)
• Sample temperature
• Auto range
• Shear rate (RV2 models only)
• Shear stress (RV2 models only)
R
Your Elcometer Rotary Viscometer is packed in a cardboard and foam package. Please ensure that
this packaging is disposed of in an environmentally sensitive manner. Consult your local
Environmental Authority for further guidance.
To maximise the benefits of your new Elcometer Rotary Viscometer please take some time
to read these Operating Instructions. Do not hesitate to contact Elcometer or your Elcometer
supplier if you have any questions.
3
R
GETTINGSTARTED
2 GETTING STARTED
This section of the instructions is intended for first-time users of the Elcometer Rotary Viscometer.
It contains information on the parts and controls of the instrument and advice on safe use of the
equipment. When you have finished reading this section you will be ready to start using your
Elcometer Rotary Viscometer.
2.1 The parts of the instrument
Bubble level
Viscometer head with
display and control
panel
Temperature probe
4
Spindle
Base
Column
2.2 Assembling the Viscometer
To assemble your Viscometer:
1. Remove the nut from the bottom end of the column.
2. Fit the column onto the base with the slot facing
towards the front.
3. Fit the nut onto the bottom of the column and tighten
using the box spanner supplied with your Viscometer.
4. Place the base and column on a stable and flat surface.
5. Insert the rear rod of the Viscometer head through the
hole in the mounting bracket on the column.
6. Rotate the mounting bracket lever to lock the
Viscometer head in position.
R
GETTINGSTARTED
7. Rotate the two adjustable legs of the base until the
Viscometer is level; use the bubble level on the top of
the instrument as a guide.
8. Remove the plastic protection cap from the bottom of
the Viscometer; pull the cap vertically down until it is
well clear of the drive coupling and then remove to one
side.
Do not throw the protection cap away. Always
replace the protection cap when the instrument is
not in use.
9. Connect the instrument to the mains.
5
R
GETTINGSTARTED
2.3 Rear connections
PT 100
probe
socket
Power
supply
socket
RS 232
interface
Power
on/off
switch
2.4 Control panel
Your Viscometer is controlleda using the five buttons on the control panel on the front of the
Viscometer head.
Enter setup screen
ENTER
Start /
stop
motor
START
UP
DOWNSTOP
Increase /
decrease
speed and
modify setup
functions
a. Viscometer models RV2 can be connected to a PC and controlled remotely via the supplied ViscosityMaster
software - see the instructions supplied with the software.
6
GETTINGSTARTED
2.5 Configuring your Viscometer
Switch on the Viscometer by pressing the power on/off switch at the rear of the Viscometer head.
A welcome message will be displayed for a period of two seconds. The contents of the message
will depend upon the model, software version and language setting:
V2-R v4.3
English
While this message is displayed, press the [START] and then the [ENTER] buttons.
The language selection screen will appear:
Select Language
English
Using the [UP] or [DOWN] buttons, select the desired language (German, Catalan, Spanish,
French, English, Italian or Portuguese). Press [ENTER] to confirm your selection.
The viscosity units selection screen will appear:
R
Viscosity Units
SI (mPas)
Using the [UP] or [DOWN] buttons, select the viscosity units, SI (mPa•s) or CGS (cP). Press
[ENTER] to confirm your selection.
Model RV2 only: The shear stress units selection screen will appear:
Shear Stress Units
SI (N/m2)
Using the [UP] or [DOWN] buttons, select the shear stress (S.S.) units, SI (N/m
(dyne/cm
2
). Press [ENTER] to confirm your selection.
2
) or CGS
7
R
GETTINGSTARTED
The temperature units selection screen will appear:
Temperature Units
Celsius
Using the [UP] or [DOWN] buttons, select the temperature units, Celsius or Fahrenheit. Press
[ENTER] to confirm your selection.
The mode selection screen will appear:
Computer mode
Using the [UP] or [DOWN] buttons, select the required mode, Printer or Computer. Press [ENTER]
to confirm your selection.
See “Spindle selection tables” on page 19 for further details
10
TL5TL6TL7
2.8 Fitting the spindle
Note: This operation must be done very carefully to
ensure that the spindle is not bent and the motor drive
coupling or internal components are not damaged.
1. Consult “Spindle selection tables” on page 19 to identify
the correct spindle to use; the spindle type is stamped
into the spindle coupling.
2. Check that the threads and connecting surfaces of the
spindle and the motor drive coupling are clean.
3. If the spindle is of a disc type, it should be submerged
carefully in the sample to avoid bubbles forming under its
lower surface.
4. Grasp the motor drive coupling with one hand, raise it
slightly and then screw the spindle onto the coupling
gently. The screw has a left-hand thread.
5. Adjust the height of the Viscometer head until the spindle
is immersed in the sample up to the immersion point,
indicated by a groove on the spindle. Take care to avoid
knocking the sample container against the spindle during
this operation as this may damage the spindle and motor
drive coupling.
R
GETTINGSTARTED
2.9 Caution
The Elcometer Rotary Viscometer has been manufactured with your safety in mind.
However, improper use can result in damage to the machine.
Please observe the precautions discussed in these operating instructions.
•To reduce the risk of electric shock, do not open the head of the Viscometer. There are no
user-serviceable parts inside.
•To reduce the risk of fire or electric shock, do not expose the Viscometer to rain or excess
moisture.
•The mains plug on your Viscometer may be fitted with a fuse. When replacing this fuse, ensure
a fuse of the correct rating is used.
11
R
USINGYOUR VISCOMETER
3 USING YOUR VISCOMETER
This section describes how to take measurements with your Viscometer. The instructions apply to
RV1 Viscometers and to RV2 Viscometers under manual operation. For additional instructions on
using RV2 Viscometers interfaced to a PC, refer to “ViscosityMaster Software” on page 30.
3.1 Before you start
1. Attach the spindle guard (if necessary) - See “Fitting the spindle guard” on page 9.
2. Attach the spindle - See “Fitting the spindle” on page 11.
3. Lower the spindle into the test sample.
4. Insert the temperature probe connector into the socket on the rear of the Viscometer head and
place the probe into the sample.
3.2 Setting spindle type and rotation speed
1. Switch on the Viscometer by pressing the power on/off switch at the rear of the Viscometer
head. A welcome message will be displayed for a period of two seconds. The contents of the
message will depend upon the model, software version and language setting:
V2-R v4.3
English
After two seconds, the data screen will be displayed, showing the latest spindle and speed
parameters used (these were stored in memory the last time the instrument was used).
rpm60R5
mPas
temp20.5
o
C
range10000 mPas
2. The spindle type flashes. Use the [UP] or [DOWN] buttons to modify the spindle type. Press
[ENTER] to confirm your selection.
The rotation speed (rpm) flashes. Use the [UP] or [DOWN] buttons to modify the rotation
speed. Press [ENTER] to confirm your selection.
The parameter ‘range’ indicates the maximum viscosity value which can be measured using
the selected combination of spindle and rotation speed.
The message ‘press start’ will flash on the fourth line of the display, indicating that the
Viscometer is now ready to be started.
rpm60R5
mPas
temp20.5
o
C
Press start
12
USINGYOUR VISCOMETER
3.3 Starting a measurement
1. Press [START] to start the measurement.
Stable flow conditions are reached quickly and the viscosity readings on the display of the
Viscometer can be considered correct within a few seconds (depending on the selected speed
and the viscosity of the sample).
If the viscosity reading exceeds the optimum measuring range (<10% and >90% of selected
full scale), the instrument will give a warning beep.
The message ‘ERROR’ appearing on the screen indicates that the maximum viscosity value
has been exceeded. In this case, the speed should be reduced or a higher range spindle
should be used.
2. When you are satisfied that the reading is stable, press [STOP] to stop the motor. The motor
will slow down smoothly until 0 rpm is reached; this protects the delicate parts of the
instrument.
The last viscosity reading will be retained on the display.
rpm60R5
mPas 508050.8%
temp20.5
range10000 mPas
On pressing [START] again, the Viscometer will restart at the preset speed.
If you are using an RV2 Viscometer
The information shown on the display of an RV2 Viscometer is identical to that shown on an RV1
Viscometer unless the optional accessories are being used.
If the Adaptor for Small Sample Volumes or the Adaptor for Low Viscosity Materials is being used,
the bottom two lines of the display will show the Shear Rate (S.R.) and the Shear Stress (S.S.).
o
C
R
rpm60LCP
mPas 4.0540.9%
1/sec 73.3820.5
o
C
N/m2 3.12
Changing spindle rotation speed and spindle type
To modify the spindle rotation speed and spindle type, while the motor is running, press [ENTER]
to return to the data screen.
3.4 Viscosity measurement problems
Unstable viscosity readings may be caused by changes in the temperature of the sample or by the
behaviour of the product itself.
If the readings obtained are not stable, the temperature of the sample should be controlled and
stabilised. If the readings remain unstable once the temperature has been controlled, the instability
is due to the interaction between the product sample and the movement of the spindle.
In this instance, the product is called ‘Non-Newtonian’ and the behaviour of the product can be
described as thixotropic, rheopexy, dilatant, pseudoplastic, etc. The same settings (spindle, speed
13
R
PRINTINGTESTRESULTS
and temperature) should always be used to test this type of product and additional parameters
should be recorded such as the time elapsed before first measurement, sampling interval, etc.
3.5 After use
Clean the instrument and spindle.
Always remove the spindle from the instrument before cleaning. The spindle or motor
coupling will almost certainly be damaged if attempts are made to clean the spindle while it
is still attached to the instrument.
Do not use very aggressive solvents or wire brushes, metal scrapers, metal files or other
metallic tools. Do not use excessive force which might otherwise bend delicate spindles.
Clean the spindle and other parts using a suitable solvent only.
After cleaning and before storing the instrument ensure that all materials are removed and
that the instrument is dry.
4 PRINTING TEST RESULTS
Test results from your Viscometer can be printed using two methods:
•ViscosityMaster software
•Printing Mode
Use of ViscosityMaster software provides the greatest flexibility and allows for printing, storage
and graphical analysis of test results. Full instructions for ViscosityMaster can be found in
“ViscosityMaster Software” on page 30.
Printing Mode provides a quick and easy way to print
individual test results. Connect your Viscometer to a
suitable printer (see “Thermal printer” on page 29), press
the START button and the test results will be printed.
To use Printing Mode:
1. Switch on your Viscometer and set it to Printing
Mode - see “Configuring your Viscometer” on page 7.
2. Connect your Viscometer to the printer using a
null-modem
3. Whenever the START button is pressed the test
results will be sent to the printer.
Note: When Printing Mode is OFF, switch off your printer - otherwise a continuous stream of test
results will be printed.
b
cable.
b. Null-modem pin connections:
9 pin connector
(male)
Pin 7
Pin 2
Pin 3
(ground)
14
25 pin connector
(male)
Pin 5
Pin 3
Pin 2
USINGTHEOPTIONALACCESSORIES
5 USING THE OPTIONAL ACCESSORIES
A number of optional accessories are available which will extend the capabilities of your
Viscometer. For ordering information, See “Spare parts and accessories” on page 28
5.1 Adaptor for small sample volumes
The adaptor for small sample volumes allows precise
measurement of the viscosity of small sample volumes
(8 ml to 13 ml). The adaptor is suitable for use with all
models of the Elcometer Rotary Viscometer.
There are two types of this adaptor:
•without an integrated temperature sensor
•with an integrated temperature sensor
Although the following instructions describe the version
which does not have an integrated temperature sensor,
they may also be used for the version which does have a
temperature sensor as it is very similar in design.
The adaptor consists of a water jacket, sample chamber
and a mounting bracket. Special spindles types TL or TR
are used with this adaptor.
Water from an external heating bath is circulated through
the jacket surrounding the sample chamber; this allows the
sample to be maintained precisely at a set temperature of
-10°C to 100°C (14°F to (212°F).
See also “Small volume samples adaptor kit” on page 28.
Viscosity range
Version L: 3 mPa•s to 200 000 mPa•s
Version R: 40 mPa•s to 3 300 000 mPa•s
Using the adaptor
1. Attach the appropriate spindle to the Viscometer motor drive coupling using the screwed
coupling and mounting hook.
2. Attach one end of the mounting bracket to the water jacket, as shown; note the locating pin
which allows the bracket to be attached the correct way only.
3. Slide the water jacket slowly and carefully over the spindle and then attach the other end of the
bracket to the head of your Viscometer.
4. Fit the end cap to the bottom of the stainless steel sample chamber.
5. Fill the sample chamber with a suitable volume of the fluid to be tested; 8 ml to 13 ml (0.27 oz
to 0.44 oz) - depending upon spindle.
6. Slide the sample chamber up into the water jacket and rotate to lock into position.
7. Fit the cover around the spindle hook and lower it onto the top of the sample chamber.
8. Connect the external heating bath to the water jacket inlet and outlet and then leave it
operating for a sufficient amount of time for the temperature of the sample to stabilise.
The adaptor for small sample volumes is now ready to use.
R
15
R
USINGTHEOPTIONALACCESSORIES
5.2 Adaptor for low viscosity samples
The adaptor for low viscosity samples allows precise
measurement of the viscosity of low viscosity samples
(16 ml to 18 ml), both Newtonian and non Newtonian. The
adaptor is suitable for use with all models of the Elcometer
Rotary Viscometer.
There are two types of this adaptor:
•with water jacket - for indirect heating up to 100°C
(212°F)
•without water jacket - for sample temperatures up to
200°C (392°F)
Although the following instructions describe the version
which has a water jacket, they may also be used for the
unheated version which is very similar in design.
The adaptor consists of a water jacket, sample chamber, a
mounting bracket, a column extension piece and a special
cylindrical spindle (LCP).
Water from an external heating bath is circulated through
the jacket surrounding the sample chamber; this allows the
sample to be maintained precisely at a set temperature of
-10°C to 100°C (14°F to (212°F).
For ordering information see “Low viscosity adaptor kit with water jacket” on page 28 and “Low viscosity adaptor kit
- without water jacket” on page 28.
Viscosity range
Version L: 0.5 mPa•s to 2000 mPa•s
Version R: 4.8 mPa•s to 21 333 mPa•s
Using the adaptor
1. Fit the extension piece to the column. This raises the Viscometer head sufficiently to allow the
adaptor for low viscosity samples to be used correctly.
2. Attach the LCP spindle to the Viscometer motor drive coupling using the screwed coupling and
mounting hook.
3. Attach one end of the mounting bracket to the water jacket, as shown; note the locating pin
which allows the bracket to be attached the correct way only.
4. Slide the water jacket slowly and carefully over the spindle and then attach the other end of the
bracket to the head of your Viscometer.
5. Fit the end cap to the bottom of the stainless steel sample chamber.
6. Fill the sample chamber with a suitable volume of the fluid to be tested; 16 ml to 18 ml (0.54 oz
to 0.61 oz).
7. Slide the sample chamber up into the water jacket and rotate to lock into position.
8. Fit the cover around the spindle hook and lower it onto the top of the sample chamber.
9. Connect the external heating bath to the water jacket inlet and outlet and then leave it
operating for a sufficient amount of time for the temperature of the sample to stabilise.
The adaptor for low viscosity samples is now ready to use.
16
USINGTHEOPTIONALACCESSORIES
5.3 Adaptor for helical movement
The adaptor for helical movement allows precise
measurement of the viscosity of samples such as
cream, gel, wax, paste, putty, gelatine, etc. The
adaptor is suitable for use with all models of the
Elcometer Rotary Viscometer.
The adaptor moves the Viscometer head slowly up
and down. This movement causes the spindle to
trace a helical path through the sample, therefore
avoiding the creation of voids in the sample which
would otherwise make measurement impossible.
The adaptor consists of:
•drive mechanism with head mounting bracket
and mains lead
•spindle clamp/coupling and weight
•spacer tube
•ring clamp, x 2
•set of P series spindles
water jacket, sample chamber, a mounting
bracket, a column extension piece and a special
cylindrical spindle (LCP).
The adaptor is used with special T-type spindles PA, PB, PC, PD, PE, PF
See also “Helical movement adaptor kit” on
page 29.
Viscosity range
Version L: 156 mPa•s to 3 120 000 mPa•s
Version R: 1660 mPa•s to 33 300 000 mPa•s
Using the adaptor
1. Remove the standard head mounting bracket from the column.
2. Slide the spacer tube (A) onto the column until it rests on the base.
3. Slide one of the ring clamps (B) onto the column.
4. Press and hold lever (C) of the drive mechanism upwards and slide the drive mechanism (D)
onto the column. Release lever (C) to engage the drive mechanism with the column.
R
17
R
USINGTHEOPTIONALACCESSORIES
5. Slide the second ring clamp onto the
column.
6. Insert the rod of the Viscometer head
through the hole in the mounting
bracket on the drive mechanism and
rotate the mounting bracket lever to lock
the Viscometer head in position.
7. Screw the weight onto the spindle
clamp/coupling.
8. Select the appropriate spindle (See “P
series spindles” on page 24) and insert
it into the spindle clamp/coupling. Twist
the body of the clamp to lock the spindle
in position.
9. Attach the spindle clamp/coupling to the
Viscometer motor drive coupling - See
“Fitting the spindle” on page 11.
10. Adjust the position of the two ring
clamps to give the desired amount of
vertical movement and clamp into
position.
The adaptor for helical movement is now
ready to use. Connect to the mains supply
and press the power on/off switch to switch it
on.
D
C
B
A
18
SPINDLESELECTIONTABLES
6 SPINDLE SELECTION TABLES
6.1 L and TL Spindles (For use with RV1-L / RV2-L Viscometers).
a. Type L spindles are the standard spindles supplied with RV1-L/RV2-L Viscometers.
b. The minimum recommended reading is 15% of the full scale value.
c. The L1 spindle is used for low viscosity measurement. The spindle guard must be used with this spindle in
order to obtain correct viscosity readings. For higher accuracy when measuring low viscosity samples, use
the Adaptor for Low Viscosity Materials and the special cylindrical spindle (LCP); See “Spare parts and
accessories” on page 28.
a. Type TL spindles are the optional spindles for use with the Adaptor for Small Sample Volumes; See “Spare
parts and accessories” on page 28.
b. The minimum recommended reading is 15% of the full scale value.
c. TL5 shear rate: 1.32 x rpm with sample volume of 8.0 cm³
d. TL6 shear rate: 0.34 x rpm with sample volume of 10.0 cm³
e. TL7 shear rate: 0.28 x rpm with sample volume of 9.5 cm³
20
SPINDLESELECTIONTABLES
6.2 R and TR Spindles (For use with RV1-R / RV2-R Viscometers)
a. Type R spindles are standard spindles supplied with RV1-R/RV2-R Viscometers.
b. The minimum recommended reading is 15% of the full scale value.
c. Spindle R1 is used for measurement of samples having low viscosity. The spindle guard must be used with
this spindle in order to obtain correct viscosity readings. Because models RV1-R and RV2-R are normally
used for medium viscosity ranges, the spindle R1 is not very frequently used, therefore is not supplied as
standard equipment with your Viscometer. However, it is available on request; See “Spare parts and accessories” on page 28.
a. Type TR spindles are the optional spindles for use with the Adaptor for Small Sample Volumes; See “Spare
parts and accessories” on page 28.
b. The minimum recommended reading is 15% of the full scale value.
c. TR8 shear rate: 0.93 x rpm with sample volume of 8.0 cm³
d. TR9 shear rate: 0.34 x rpm with sample volume of 10.5 cm³
e. TR10 shear rate: 0.28 x rpm with sample volume of 11.5 cm³
f.TR11 shear rate: 0.25 x rpm with sample volume of 13.0 cm³
The Shear Rate (S.R.) has been calculated assuming the characteristics of newtonian products.
a. Type P series spindles are the optional spindles for use with the Adaptor for Helical Movement - See “Spare
parts and accessories” on page 28.
b. The minimum recommended reading is 15% of the full scale value.
24
CALIBRATION
7 CALIBRATION
Your Viscometer was calibrated at the factory before shipment and a calibration certificate was
included in the box. Additional calibration checks at regular intervals over the life of the Viscometer
are a requirement of quality management procedures, e.g. ISO 9000, and other similar standards.
Calibration checks may be carried out by the user of the Viscometer (see instructions below),
however if the calibration of the Viscometer is found to need adjustment, please contact Elcometer
or your local Elcometer supplier.
Checking Viscometer calibration
The following calibration checking procedure requires the use of the Adaptor for low viscosity
samples - See “Spare parts and accessories” on page 28.
Use of this adaptor helps to eliminate fluctuations in the temperature of the sample and variations
due to differences in the sample container.
It is recommended that calibrated viscosity oil is used for checking calibration. Oils between
100 mPa•s to 125 mPa•s for version R and 5 mPa•s for version L should be adequate for the
calibration procedure.
1. Fit the adaptor for low viscosity samples and fill with the recommended volume of calibration
oil - See “Adaptor for low viscosity samples” on page 16.
2. Switch on and start the Viscometer; begin with viscosity readings at the highest speed
permitted by the combination of speed/spindle (taking into account the nominal viscosity and
temperature values of the calibration oil).
3. Once the desired sample temperature has been reached, check the viscosity reading on the
display. If the reading is identical to the nominal viscosity of the calibration oil, your Viscometer
is calibrated correctly.
R
8 MAINTENANCE
The Elcometer Rotary Viscometer is designed to give many years reliable service under normal
operating and storage conditions.
The Elcometer Rotary Viscometer does not contain any internal user-serviceable components. In
the unlikely event of a fault, the instrument should be returned to your local Elcometer supplier or
directly to Elcometer.
Details of Elcometer offices around the world are given on the outside cover of these Operating
Instructions. Alternatively visit the Elcometer website, www.elcometer.com
25
R
TROUBLESHOOTING
9 TROUBLESHOOTING
The instrument does not operate
Check:
•the connection to the mains electricity supply
•the power on/off switch position at the rear of the Viscometer head
The spindle does not rotate concentrically
Check:
•that the spindle is correctly adjusted to the shaft
•that the union between spindle and shaft is clean
The instrument does not read zero in vacuum
Check:
•that the instrument is levelled correctly
Viscosity reading is not stable or not very accurate
Check:
•that the instrument is levelled correctly
•that the rotation speed/spindle type selection is correct
•that the temperature of the sample is stable
Viscometer head slips down the column
•tighten the two nylon grub screws on the head mounting bracket using the hexagonal wrench
supplied
Viscosity range,Version L:3 mPa•s to 2 000 000 mPa•s in 76 ranges
19 speeds with 4 spindles.
(For low to medium viscosity test fluids)
Version R:20 mPa•s to 13 000 000 mPa•s in 114 ranges
19 speeds with 6 spindles.
(For medium to high viscosity test fluids)
Accuracy (measurement):± 1% of full scale
Repeatability:± 0.2%
Maximum altitude:2 000 m (6500 ft) above sea level
Surge class:Class II (domestic/light industry)
Contamination:Level 2
Thermometer,Range:-15°C to +180°C (5°F to 356°F)
Resolution:0.1°C (0.1°F)
Accuracy:± 0.1°C (0.18°F)
Operating voltage:UK, 240 V AC 50 Hz
EUR, 220 V AC 50 Hz
US, 110 V AC 60 Hz
Power consumption:23 W
Dimensions (of carry case):495 mm x 420 mm x 200 mm
(19.5" x 16.5" x 8")
Weight (including carry case):9 kg (20 lb)
27
R
SPAREPARTSANDACCESSORIES
11 SPARE PARTS AND ACCESSORIES
The following spare parts and accessories for your Elcometer Rotary Viscometer are available from
your local supplier or direct from Elcometer:
11.1 Standard spindles
Spindle set, standard, type L (L1 to L4)KT00230019698
Spindle set, standard, type R (R2 to R7)KT00230019699
Spindle, R1KT00230019700
11.2 Low viscosity adaptor kit - with water jacket
Adaptor kit for low viscosity samples, with water jacketKT00230019710
Adaptor kit for low viscosity samples, with water jacket,
plus integrated temperature sensorKT00230020174
(indirect heating by water up to 100°C)
(This adaptor kit is
supplied complete
with LCP cylindrical
spindle for low
viscosity samples)
11.3 Low viscosity adaptor kit - without water jacket
Adaptor kit for low viscosity samples, without water jacketKT00230019711
(for testing of samples up to 200°C)
(This adaptor kit is
supplied complete
with LCP cylindrical
spindle for low
viscosity samples)
(Comes complete with LCP cylindrical spindle for low viscosity samples)
11.4 Small volume samples adaptor kit
Adaptor kit for small volume samplesKT00230019702
Adaptor kit for small volume samples + integrated temperature sensor KT00230019784
(This adaptor kit requires a
spindle set for small volume
samples - see below)
Spindle set, small volume samples, (TL5, TL6 and TL7)KT00230019703
Spindle set, small volume samples, (TR8, TR9, TR10 and TR11)KT00230019704
28
RELATEDEQUIPMENT
11.5 Helical movement adaptor kit
Helical movement adaptor,240 V (UK)KT00230019705
220 V (EU)KT00230019706
110 V (USA)KT00230019707
(This adaptor kit is supplied
complete with six special T-type
spindles - PA, PB, PC, PD, PE,
PF)
11.6 Thermal printer
Thermal printer, null modem cable + adaptor 240 V (UK)KT00230020179
220 V (EU)KT00230020178
110 V (USA)KT00230020180
R
12 RELATED EQUIPMENT
In addition to the Elcometer Rotary Viscometer, Elcometer produces a wide range of other
equipment for determining the physical characteristics of surface coatings.
Users of the Elcometer 2300 may also benefit from the following Elcometer products:
•Elcometer 2000 Muller Laboratory Grinder
•Elcometer Fineness of Grind Gauges
•Elcometer Viscosity Cups
•Elcometer Film Applicators
For further information contact Elcometer, your local supplier or visit www.elcometer.com
29
R
ABOUT VISCOSITYMASTER
SECTION B: VISCOSITYMASTER SOFTWARE
13 ABOUT VISCOSITYMASTER
ViscosityMaster is powerful but easy-to-use software which simplifies and enhances the use of your
Viscometer:
•Design test processes on your PC and upload
Viscometer.
•Download readings data from your Viscometer to the ViscosityMaster data management
system.
•Keep your test notes, reports, photographs and all your other test information in one place ViscosityMaster.
•Create accurate and professional reports configured to allow your data to be understood easily
at all levels.
•Create PDF versions of your reports and e-mail direct from ViscosityMaster.
13.1 System Requirements
To use the software on this CD you will need an Elcometer 2300 Rotational Viscometer, a computer
and a data transfer cable.
The software will run on PC compatible computers under Windows
XP. The software will not run on Mac OS.
A data transfer cable to connect your Viscometer to the PC is required. This cable is supplied as
standard equipment with Elcometer 2300 RV2 Rotational Viscometer. Spare cables are available
as an accessory; contact Elcometer or your local supplier for ordering details.
a
them to your Elcometer 2300 RV2 Rotational
®
98, 98SE, 2000, ME, NT and
14 GETTING STARTED
This section of the instructions is intended for first-time users of ViscosityMaster. It describes the
steps you must take before you can start using the software, how to customise your working
environment and where to get help.
14.1 Installing ViscosityMaster
To install ViscosityMaster software on your PC:
1. Place the CD in the CD drive of your PC. The software installation wizard should run
automatically. If it does not run automatically:
• Choose Start > Run
• Type D:ViscosityMasterSetup (where ‘D’ is the letter of your CD drive)
• Click OK. The software installation wizard will run.
2. Follow the on-screen installation instructions.
a. Uploading of test processes to Elcometer 2300 RV1 models is not possible.
30
GETTINGSTARTED
14.2 Uninstalling ViscosityMaster
To uninstall the software follow the usual Windows procedure for removing programs. For Windows
XP:
•Choose Start > Control Panel > Add or Remove Programs > ViscosityMaster 1.xx > Remove
14.3 Starting ViscosityMaster
Start ViscosityMaster just as you would any software application:
You must connect the Viscometer to your computer if you want to transfer information to and from
your Viscometer
To connect your Viscometer:
1. Plug the 9-pin female D-type connector on one end of the data transfer cable into the COM
port
2. Plug the 9-pin connector on the other end of the data transfer cable into the RS232 connector
at the rear of the Viscometer.
3. Switch on the Viscometer.
b
c
on the PC.
.
14.5 Preferences
Use this menu to customise your ViscosityMaster working environment. Choose between metric
and imperial units of measurement, quality of report page thumbnails and whether e-mail report
R
b. If your Viscometer is not connected to your computer, you can still use ViscosityMaster to create batch
templates, view data downloaded previously and produce reports.
c. An RS232 to USB transfer cable is available which allows you to connect your Viscometer to a USB port on
your computer; part number T99916716. Contact Elcometer or your local supplier for ordering details.
31
R
GETTINGSTARTED
format is PDF or JPEG image. Also set default values for Viscometer model, spindle type, company
and operator and which COM port your Elcometer 2300 Rotational Viscometer is connected to.
14.6 Help
Help for your ViscosityMaster software is provided by this help file. A PDF version of this help file is
included on the ViscosityMaster installation CD. If you need to print any sections of this help file,
print from the PDF version as this will give a better quality document.
If you need help with your Elcometer 2300 Rotational Viscometer, refer to the operating instructions
which were included with your Viscometer.
If you cannot find the answer to your question, please do not hesitate to contact Elcometer
or your Elcometer supplier.
32
THEUSERINTERFACE
15 THE USER INTERFACE
The ViscosityMaster user interface is very similar to a standard Windows Explorer interface:
2. Explorer Window: Lists the files and folders on your PC and on your Viscometer. All data is
stored in \My Documents\My ViscoMeter Data.
3. List Window: Lists the files in a folder or the files on your Viscometer.
4. Report Window: Displays the contents of a file (click on to maximise/minimise the
window).
Many menu commands and actions can be accessed by clicking using the right mouse button in
one of the user interface windows; experiment as you use ViscosityMaster to find which method of
working (buttons or right click) you find most comfortable.
R
33
R
EXAMPLEDATA
16 EXAMPLE DATA
When you install Elcometer ViscosityMaster software, a set of example data is loaded into your local
folder; click on ‘Viscometer Data’ in the Explorer Window and then look for ‘Example Batch’ in the
List Window.
Double click on Example Batch to load the data into the Report Window and spend some time
looking at the readings and charts - you will soon learn the basics of ViscosityMaster and what it
can do.
Deleting the example data (when you have finished learning about ViscosityMaster, for example)
will not affect the operation of your ViscosityMaster software.
34
OPERATINGMODES
17 OPERATING MODES
Your ViscosityMaster software has two operating modes; manual and automatic.
•In manual mode your ViscosityMaster software records measurements as they are taken by
the instrument but it does not control the operation of the Viscometer.
•In automatic mode your ViscosityMaster software records measurements and it controls the
operation of the Viscometer.
Note: RV1 models can only be used in manual mode. RV2 models can be used in Manual mode
and in Automatic mode.
18 INSTRUCTIONS FOR RV1 VISCOMETERS
Note: RV1 Viscometers can only be used in manual mode - they cannot be used in automatic mode.
18.1 Setting up a batch
1. Click on ‘New’ and select ‘New Batch’.
2. Type a name for the batch and then press the Return key.
R
35
R
INSTRUCTIONSFOR RV1 VISCOMETERS
3. The Model, Spindle type, Company and Operator will already be set from ‘Preferences’ when
ViscosityMaster was run for the first time. If necessary, amend these details as required.
4. Enter details in the Identifier and Notes fields. This information is used later, typically within
test reports.
36
INSTRUCTIONSFOR RV2 VISCOMETERS
18.2 Running a batch
1. Lower the Viscometer spindle into the test sample.
2. Click on ‘Manual’ and then press ‘Start’ on the Viscometer.
3. The measurements and charts will be shown in real time.
R
4. If necessary, adjust the content of the chart X and Y axes to show the information you require.
5. Click on ‘Stop’ when you have obtained enough measurements.
6. Review the measurements/charts. If you are not happy with the results, click on ‘Clear’ to clear
the measurements and start the test again.
After you have stopped the test you can:
•review readings
•review charts
•create reports
•export readings to Excel or csv files
•print, create pdf or email charts and reports.
See “Working with Viscometer data” on page 41 for further details.
19 INSTRUCTIONS FOR RV2 VISCOMETERS
Note: The instructions in this section apply to RV2 models only - RV1 models can only be used in
manual mode.
19.1 Setting up a batch
The first section of the procedure is identical to the instructions given for operation in manual mode:
1. Click on ‘New’ and select ‘New Batch’.
37
R
INSTRUCTIONSFOR RV2 VISCOMETERS
2. Type a name for the batch and then press the Return key.
3. The Model, Spindle type, Company and Operator will already be set from ‘Preferences’ when
ViscosityMaster was run for the first time. If necessary, amend these details as required.
4. Enter details in the Identifier and Notes fields. This information is used later, typically within
test reports.
The procedure for automatic mode now differs from the procedure for manual mode because of
‘Processes’ - see Setting up processes .
38
INSTRUCTIONSFOR RV2 VISCOMETERS
19.2 Setting up processes
In automatic mode the Viscometer is controlled using processes. Processes are groups of data
which specify settings such as spindle speed, rotation duration, measurement frequency, etc.
A single batch can incorporate any number of processes.
1. Click on the ‘Processes’ tab.
2. Click on ‘New Process’.
A Process Wizard will start. Follow the instructions given in the wizard to create a single
process. You will be asked to define the following process variables:
IdentifierA name or number for the process
Start RPMSpindle speed at the start of the process
End RPMSpindle speed at the end of the process
TimeDuration of the process (seconds)
Start delayTime delay before measurements start (seconds)
Start temperature
Measurements
Delay start until when above this temperature, or
Delay start until when below this temperature
Number of measurements to take during the
process
R
3. Click on ‘Finish’ at the end of the process.
The process will appear as a line item under the ‘Processes’ tab.
4. Go on defining as many processes as you need to define the test procedure.
39
R
INSTRUCTIONSFOR RV2 VISCOMETERS
19.3 Running a batch
1. Lower the Viscometer spindle into the test sample.
2. Click on ‘Automatic’.
The Viscometer will start under control of ViscosityMaster and the measurements and charts
will be shown in real time.
3. If necessary, adjust the content of the chart X and Y axes to show the information you require.
4. The Viscometer will stop automatically at the end of the test procedure. Alternatively, click on
‘Stop’ to stop the procedure at any time.
5. Review the measurements/charts. If you are not happy with the results, click on ‘Clear’ to clear
the measurements and start the test again.
After you have stopped the test you can:
•review readings
•review charts
•create reports
•export readings to Excel or csv files
•print, create pdf or email charts and reports.
See “Working with Viscometer data” on page 41 for further details.
40
WORKINGWITH VISCOMETERDATA
20 WORKING WITH VISCOMETER DATA
This section describes what you can do with your Viscometer batch data once it has been
downloaded into ViscosityMaster:
•Viewing Measurement Data and Standard Reports
•Customising report window tabs
•Creating Custom Reports
•Viewing a Custom Report
•Exporting Measurement Data
•Printing, Creating PDF Files and E-mailing from ViscosityMaster
20.1 Viewing Measurement Data and Standard Reports
Viewing measurement data and producing standard reports is quick and easy using
ViscosityMaster software.
To view measurement data for a batch:
1. In the Explorer Window, navigate to the folder where your data is stored.
2. In the List Window, click on the batch or batches you want to view. To select more than one
batch, hold down the ‘Ctrl’ key and click on the batches required.
3. In the Report Window, click on the Measurements tab.
R
41
R
WORKINGWITH VISCOMETERDATA
4. Adjust the contents of the chart using the X axis and Y axis drop down lists.
5. To zoom in to an area on the chart, use your mouse to left click and drag. To reset the zoom
click on the Reset Zoom button.
To scroll the chart, use your mouse to right click and drag
To view standard reports for a batch:
1. In the Explorer Window, navigate to the folder where your data is stored.
2. In the List Window, click on the batch or batches you want to view. To select more than one
batch, hold down the ‘Ctrl’ key and click on the batches required.
3. In the Report Window, click on the tab required: Paged Run Chart, Overview Run Chart.
4. Click on a thumbnail of a report on the left side of the Report Window to view an individual
page of the report. Use the Zoom pull-down list to adjust the size of the report on the screen.
42
WORKINGWITH VISCOMETERDATA
20.2 Customising report window tabs
The Header, Measurements and My Reports tabs are always visible in the Report Window. All other
tabs (and user defined tabs) can be added or deleted:
1. In the Explorer Window, click on Reports.
2. In the List Window, all reports in the ‘Tab Reports’ column will appear as tabs in the Report
Window. Use the arrows between the two columns to add or remove reports from this column.
•All reports with a blue icon are standard reports which cannot be deleted.
•All reports with a brown icon are user defined reports which may be deleted if needed.
R
20.3 Creating Custom Reports
ViscosityMaster software includes a built-in report designer which can be used to create reports
customised to your requirements.
To create a custom report:
1. In the Explorer Window, click on ‘Reports’.
2. In the Toolbar, click ‘New Report’.
3. Choose a name for your report and press ‘Return’.
43
R
WORKINGWITH VISCOMETERDATA
4. In the Report Window, click ‘New Page’ and then choose between ‘Opening’, ‘Inner’ and
‘Closing’ page types.
• Opening pages: Typically used for report title page; can contain text, shapes and graphics;
cannot be used to display reading data; prints first when printing a report.
• Inner pages: Typically used for showing reading data (run charts, measurements); prints in
the middle when printing a report.
• Closing pages: Typically used for report notes/summary page; can contain text, shapes
and graphics; cannot be used to display reading data; prints last when printing a report.
The new page will then appear in the Report Window.
5. Click on the page in the Report Window and then click on the Edit button. The Page Designer
will start.
6. Follow the instructions in “The page designer” on page 47 to define your report.
7. Click ‘Save’ to save your report layout when you are finished.
20.4 Viewing a Custom Report
To view a custom report:
1. In the Explorer Window, navigate to the folder where your data is stored.
2. In the List Window, click on the batch or batches you want to view. To select more than one
batch, hold down the ‘Ctrl’ key and click on the batches required.
44
WORKINGWITH VISCOMETERDATA
3. In the Report Window, click on My Reports tab and then select the report from the Reports
drop-down list.
4. Click on a thumbnail of a report on the left side of the Report Window to view an individual
page of the report. Use the Zoom drop-down list to adjust the size of the report on the screen.
R
20.5 Exporting Measurement Data
Measurement data can be exported in Microsoft Excel format and in CSV (comma separated
values) format.
Highlight the batch required and then click on ‘Export’. Choose between Excel and CSV.
20.6 Printing, Creating PDF Files and E-mailing from ViscosityMaster
Any report currently displayed in the Report Window can be:
•Printed
•Saved as an Adobe PDF (Portable Document Format) file
45
R
USINGDOCUMENTSANDIMAGES
•Sent via e-mail as an attachment
Click the appropriate button in the Report Window and follow the standard Windows instructions
which follow.
d
21 USING DOCUMENTS AND IMAGES
Use ViscosityMaster to store your test notes, reports, spreadsheets and all your other test
information. Group these files by product, customer, project, production shift, etc.
To store a document, image or any other type of file, use the standard Windows techniques of ‘drag
and drop’ or ‘copy and paste’ to place your file in the List Window.
d. Select the format of your e-mail attachment (PDF file or JPEG image) in Preferences .
46
THEPAGEDESIGNER
22 THE PAGE DESIGNER
This section describes how to use the Page Designer - a utility built in to ViscosityMaster which
allows you to create your reports:
•Designing a Page
•The Property Editor
•The Text Editor
R
1. Object Picker Window
2. Toolbar
3. Object Selector Window
4. Property Editor
5. Page Layout Window
47
R
THEPAGEDESIGNER
22.1 Designing a Page
The contents of a page are ‘Objects’ such as shapes, memos, readings, etc. To add an object to
the page, choose the object in the Object Picker Window (1), move the cursor to the Page Layout
Window (5) and then click and drag the object to the required size. Once the object has been
created it can be moved by dragging, and resized by dragging on the handles around the edge of
the object.
Precise control over the layout, attributes and content of an object is obtained by modifying the
properties of the object in the The Property Editor (4).
When an object is added to the page, the name of the object appears in the Object Selector Window
(3).
The buttons on the Toolbar (2) Zoom, Save, Delete, Cut, Copy and Paste using standard Windows
techniques.
48
THEPAGEDESIGNER
22.2 The Property Editor
Precise control over the layout, attributes and content of an object is obtained by modifying the
properties of the object using the Property Editor.
Object properties which are adjustable using the Property Editor are listed below. The list of
properties available depends upon the type of object selected (shape, memo, graphic, overview
chart, etc.):
PropertyDescription
3D:
Alignment:
Anchor:
AspectRatio:Defines how a graphic fits within its bounding box.
Bold:Make text in a memo block bold.
ColumnWidthDate:In measurements object - Width of date column (mm)
ColumnWidthPercentFS:
ColumnWidthRPM:In measurements object - Width of RPM column
ColumnWidthSheerRate:In measurements object - Width of Sheer rate column
ColumnWidthSheerStress:
ColumnWidthTemperature:
ColumnWidthTime:In measurements object - Width of Time column
ColumnWidthViscosity:In measurements object - Width of Viscosity column
FillColor:Fill colour of shapes.
FillStyle:
FontColour:Colour of text in a memo block.
FontName:Defines font type (Arial, Times, etc.)
FontSize:Size (in points) of text in a memo block.
Graphic:Selects graphic file.
Grid:Displays grid lines on histogram and run charts.
Height (mm):Height of an object in mm.
Italic:
Left (mm):
LineColor:Defines colour of bounding line of shapes.
LineStyle:
LineWidth:Defines thickness of bounding line of shapes.
Select between 3 dimensional or 2 dimensional run
charts.
Alignment of text in a memo block (top, middle, left,
right, etc.)
Defines how a block is anchored to the page. Anchor
position remains constant if page size is changed.
In measurements object - Width of percentage of full
scale column (mm)
In measurements object - Width of Sheer stress
column
In measurements object - Width of Temperature
column
Fill style of shapes (solid, diagonals, cross hatched,
etc.)
Defines whether text in a memo block is italic or
upright.
Defines gap between left edge of page and left edge
of block (see also Top).
Defines style of bounding line of shapes (solid, dash,
dot, etc.)
R
49
R
THEPAGEDESIGNER
PropertyDescription
Name:
PageWidth (mm):Defines width of the page in mm.
PageHeight (mm):Defines height of the page in mm.
PageType:Defines page type; Opening, Inner or Closing.
PageOrientation:Portrait (default) or Landscape.
PaperSize:Paper dimensions; A3, A4, Letter, etc.
ReadingsPerPage:
RowsPerPage:
Shape:
ShowColumnHeader:
ShowLegend:Defines whether colour key bar and legend is visible.
Strikeout:
Text:Opens The Text Editor .
Top (mm):
Underline:
Width (mm):Width of an object in mm.
WordWrap:
Defines name of object. Object names are displayed
in Object Selector Window.
Number of readings of a batch displayed on one
page of a report.
Number of rows of a batch displayed on one page of
a report.
Shape object type; Rectangle, Horizontal Line or
Vertical Line.
Defines whether titles at the top of readings columns
are visible.
Defines text style in a memo block as strikeout
(strikeout
).
Defines gap between top edge of page and top edge
of block (see also Left).
Defines text style in a memo block as underline
(underline
).
Defines whether text in a memo block will wrap onto
the line below.
22.3 The Text Editor
Use the Text Editor to enter text and data fields onto the Page Layout Window of The page designer .
•To enter text, click in the text editor window and type your text. Use the carriage return key to
enter a new line.
50
SUPPORT
•To enter data fields, click in the text editor window and then click on a data field button.
Explanations of the data field buttons are given in “Text Editor Data Field Buttons” on page 51.
When you have finished, click on the OK button to enter your text onto the Page Layout Window.
22.4 Text Editor Data Field Buttons
Descriptions of the data field buttons are given below:
Data Field Button Description
Batch ID:The batch name
File Path:The location of the data file on your computer
Todays Date:Today’s date
Page No.Page number in the report
Model:Viscometer type (RV1/RV2 etc.)
Spindle:Spindle type (R1, L1, etc.)
Company:Information added in header when creating a New Batch
Operator:Information added in header when creating a New Batch
Identifier:Information added in header when creating a New Batch
Note:Information added in header when creating a New Batch
R
23 SUPPORT
For support with this software, information on the Elcometer product range and details of your local
supplier, contact any of the Elcometer offices worldwide. Contact details can be found under ‘Help’
in ViscosityMaster, on the outside cover of these instructions and at www.elcometer.com
51
INDEX
24 INDEX
R
A
Accessories 28
Overview 1
Adaptor for helical movement 17
Adaptor for low viscosity samples 16
Adaptor for small sample volume 15
Assembly (of Viscometer) 5
B
Batch
Creating
RV1 35
RV2 37
Running
RV1 37
RV2 40
Box spanner 5
Buttons 6
C
Calibration 25
Caring for your Viscometer 25
Cleaning 14
Clock, setting 8
Closing page (ViscosityMaster) 44
Computer mode 8
Configuring (Viscometer) 7
Connections at rear 6
Contents of the box 2
Control panel 6
CSV files, exporting to 45