Elcometer 214l Operating Instructions Manual

Elcometer 214L
Infrared Digital
Thermometer
English
Operating Instructions
English
The Elcometer 214L Infrared Digital Thermometer has been tested in accordance with EU regulations
governing electro-magnetic compliance. This product meets the requirements of:
The EMC directive 89/336/EEC with amendments 92/31/EEC and 93/68/EEC
The low voltage directive 72/23/EEC
is a registered trademark of
Elcometer Limited. All other trademarks acknowledged.
© Copyright Elcometer Limited. 2008. All rights reserved. No part of this Document may
be reproduced, transmitted, transcribed, stored (in a retrieval system or otherwise) or translated into any language, in any form or by any means (electronic, mechanical, magnetic, optical, manual or otherwise) without the prior written permission of Elcometer Limited.
Doc.No. TMA-0379 Issue 02
Text with Cover No: 19623
CONTENTS
T
Section Page
1 About this instrument ..............................2
2 Getting started .........................................3
3 Taking a reading ......................................6
4 Measurement area ...................................6
5 Emissivity .................................................7
6 Hints and tips ...........................................7
7 Troubleshooting .......................................8
8 Storage ......................................................8
9 Technical specification ...........................9
10 Maintenance ...........................................10
11 Related equipment .................................10
hank you for your purchase of this Elcometer 214L Infrared Digital Thermometer. Welcome
to Elcometer. Elcometer are world leaders in the design,
manufacture and supply of inspection equipment for coatings and concrete. Our products cover all aspects of coating inspection, from development, through application to post application inspection.
This Elcometer 214L Infrared Digital Thermometer is a world beating product. With the purchase of this product you now have access to the worldwide service and support network of Elcometer. For more information visit our website at www.elcometer.com
1
1 ABOUT THIS INSTRUMENT
The Elcometer 214L Infrared Digital Thermometer is an easy-to-use instrument which measures the surface temperature of objects. A laser pointer in the instrument helps the user to aim the thermometer at the object and pinpoint the exact area of measurement.
The Elcometer 214L measures the infrared energy emitted by a surface to calculate temperature. Because this is a ‘non-contact’ method of measurement, the instrument is ideal for electrical and mechanical maintenance, HVAC checks, automotive testing and other applications; anywhere that temperature is a factor.
The instrument is small and lightweight and may be carried easily in a pocket.
1.1 Instrument features
Precision glass optics for accurate non­contact measurement
Temperature range from -32°C up to 530°C (-20°F to 788°F)
Switchable between °C and °F
Laser alignment
Fast scanning of cold and hot spots
Exact measurement of objects as small as 13 mm (0.5”) at any distance less than 140 mm (5.5”)
Optical resolution 20:1
Angled shape allows simultaneous view of target object and display
Backlit display
Extremely lightweight
2
1.2 What the box contains
LASER RADIATION
DO NOT STARE INTO BEAM
CLASS 2 LASER PRODUCT
Elcometer 214L Infrared Digital Thermometer
Battery
Operating instructions
The Elcometer 214L Infrared Digital Thermometer is packed in a cardboard and foam package. Please ensure that this packaging is disposed of in an environmentally sensitive manner. Consult your local Environmental Authority for further guidance.
To maximise the benefits of your new Elcometer 214L, please take some time to read these Operating Instructions. Do not hesitate to contact Elcometer or your Elcometer supplier if you have any questions.
2 GETTING STARTED
Warning! This instrument is fitted
with a Class 2 Laser. Do not point the laser directly at
the eye or indirectly off reflective surfaces as this may cause serious damage to the eye.
2.1 Fitting the battery
Press and slide down the battery compartment cover at the rear of the instrument. Insert a single 9 V PP3 alkaline battery; take care to ensure correct battery polarity. Replace the battery compartment cover.
3
2.2 The controls
Selects °C
Toggles
laser on/off
UP
Toggles
between
MAX/MIN
MODE
Selects °F
Toggles
backlight
on/off
DOWN
Press to
take a
reading
TRIGGER
HOLD
MAX
MAX
0
0
Your Elcometer 214L is operated using 4 buttons.
2.3 Switching on/off
The instrument switches on when the TRIGGER button is pressed and switches off automatically 7 seconds after the TRIGGER button is released.
2.4 Switching the laser on/off
1. Press and hold the TRIGGER button.
2. Press the UP button to toggle the laser on/off.
2.5 Switching the backlight on/off
While the instrument is switched on, press the DOWN button to toggle the backlight on/off.
4
2.6 Symbols on the display
HOLD
MAX
MAX
0
0
Backlight, low battery and laser warnings
Max/Min reading
Temperature and units
Indicates ‘held’ temperature (7 seconds max.)
Emissivity
2.7 Selecting units (°C/°F)
To select °C:
1. While the instrument is switched off, press and hold the UP button.
2. Press the TRIGGER button.
The next time the instrument is switched on, the display will show temperatures in °C.
To select °F:
1. While the instrument is switched off, press and hold the DOWN button.
2. Press the TRIGGER button.
The next time the instrument is switched on, the display will show temperatures in °F.
2.8 Selecting maximum or minimum
Press the MODE button to toggle between
displaying MAX and MIN on the display.
5
3 TAKING A READING
13 20 37
50
140
300
700
1000
spot size (mm)
distance (mm)
1. Aim the lens of the instrument at the object and press the TRIGGER button.
2. If you are using the laser, adjust your aim until the red spot is on the target.
The temperature will be shown on the display.
3. Release the TRIGGER button when you have finished taking a reading.
The display will show HOLD, indicating that the temperature shown is the last temperature recorded.
4. The display will also show the maximum or minimum temperature recorded while the TRIGGER button was pressed. To toggle between maximum and minimum, press the MODE button.
Seven seconds after the last button press the instrument will switch off automatically and the reading will be lost.
4 MEASUREMENT AREA
Your Elcometer 214L measures a circular spot. The diameter of this spot varies with distance from the lens.
Distance from lens Spot size (diameter)
< 140 mm 13 mm > 140 mm < 1000 mm see diagram below > 1000 mm distance divided by 20
The target area must be at least as large as the spot size.
6
5 EMISSIVITY
Temperature Infrared energy
1
Emissivity
--------------------------
×=
The intensity of infrared radiation which is emitted by a body depends not only upon the temperature of the body, but also upon the ability of the surface of the body to emit infrared energy.
This ability of the surface of the body to emit infrared energy is characterised by a factor known as the ‘emissivity’ of the material.
Your Elcometer 214L Infrared Digital Thermometer has a pre-set fixed emissivity value of 0.95; this value is very common for most organic materials and painted or oxidised surfaces.
The fixed emissivity value of 0.95 is shown in the bottom right hand corner of the display.
6 HINTS AND TIPS
Shiny or metallic surfaces may result in inaccurate readings. To prevent this, cover the measuring surface with either matt black paint or with a suitable tape. Remember to leave extra time for the surface covering to reach the temperature of the body being measured before taking a reading.
Infrared thermometers cannot measure the temperature of objects through transparent materials such as glass or plastic.
Always keep the lens clear of steam, dust, smoke or other particles which would otherwise cause inaccurate measurement.
7
7 TROUBLESHOOTING
Code/fault Problem Action
HHH Above
Maximum measuring limit
LLL Below
Minimum measuring limit
Blank display
Battery may be exhausted
Battery may
Laser does not work
be exhausted Laser is
switched off
8 STORAGE
Choose target within measuring specifications
Choose target within measuring specifications
Check/replace battery
Check/replace battery
Switch on laser (see page 4)
This gauge incorporates a Liquid Crystal Display (LCD). If the display is
heated above 60°C (140°F) it may be damaged. This can happen if the gauge is left in a car parked in strong sunlight.
Remove the batteries from the gauge and store them separately if the gauge is to remain unused for a long period of time. This will prevent damage to the gauge in the event of malfunction of the batteries.
8
9 TECHNICAL SPECIFICATION
Performance
Measurement range: -32°C to 420°C
(-20°F to 788°F)
Accuracy
Repeatability 0°C to 420°C: ±0.5% or ±0.7°C
a
0°C to 420°C: ±1% or ±1°C
-32°C to 0°C: ±1°C ±0.07°C/°C
-32°C to 0°C: ±0.7°C ±0.05°C/°C
Resolution (optical): 20:1
13 mm 140 mm
(distance:spot)
Resolution (display): 0.2°C (0.5°F) Response time (95%): 300 ms Emissivity: 0.95 (fixed) Laser: <1 mW class 2
9 mm offset
Spectral range: 8 µm to 14 µm
Environmental
Ambient temperature: 0°C to 50°C
(-32°F to 122°F)
Storage temperature: -20°C to 60°C (without battery) (-4°F to 140°F)
Relative humidity: 10% to 95%
(at <30°C)
Physical
Weight: 150 g (5.3 oz) Dimensions: 190 x 38 x 45 mm
Battery: 9 V, PP3/6LR61 Battery life: 40c hours
a. Strong RF fields may adversely affect
measurement accuracy.
(7.5" x 1.5" x 1.8")
b
9
10 MAINTENANCE
The Elcometer 214L Infrared Digital Thermometer is designed to give many years reliable service under normal operating and storage conditions.
Regular calibration checks over the life of the Elcometer 214L Infrared Digital Thermometer are a requirement of quality management procedures e.g. ISO 9000 and other standards. For checks and certification contact Elcometer or your local Elcometer supplier.
The instrument does not contain any user­serviceable components. In the unlikely event of a fault, the instrument should be returned to your local Elcometer supplier or direct to Elcometer.
Details of Elcometer offices around the world are given on the outside cover of these operating instructions. Alternatively visit the Elcometer website, www.elcometer.com
11 RELATED EQUIPMENT
In addition to the Elcometer 214L, Elcometer produces a wide range of other test equipment. Users of the Elcometer 214L may also benefit from the following Elcometer products:
Elcometer Pocket Digital Thermometer
Elcometer Infrared Thermometers
Elcometer Bi-metallic Thermometers For further information contact Elcometer, your
local supplier or visit www.elcometer.com
b. Alkaline batteries must be disposed of carefully to
avoid environmental contamination. Please consult your local environmental authority for information on disposal in your region.
Do not dispose of any batteries in fire.
c. 20 hours with laser and backlight on 50%
10
Loading...