EL-CELL ECC-PRESS, ECC-PRESS-DL User Manual

© 2010 - 2017
EL-CELL GmbH
ECC-PRESS (-DL)
User Manual
Release: 3.6
2017-06-12
Page 2 of 18 Manual_ECC_PRESS_Release_3.6.docx / 12.06.2017
The information in this manual has been carefully checked and believed to be accurate; however, no responsibility is assumed for inaccuracies.
EL-CELL GmbH maintains the right to make changes without further notice to products described in this manual to improve reliability, function, or design. EL-CELL GmbH does not assume any liability arising from the use or application of this product.
Contents
1 Product Description .......................................................................................... 3
2 Features and Specification .............................................................................. 3
3 Safety Precautions ........................................................................................... 7
4 Unpacking ........................................................................................................ 7
5 Assembly and Connection ............................................................................... 8
6 Sample Valve with Septum Port .................................................................... 10
7 Software Installation (DL option) .................................................................... 11
8 Recording the Pressure Signal with an External Potentiostat ........................ 12
9 Disassembly and Cleaning ............................................................................ 13
10 Accessories and Spare Parts ......................................................................... 14
11 Cables and Connectors ................................................................................. 15
12 Technical Support .......................................................................................... 17
13 Warranty ........................................................................................................ 18
EL-CELL GmbH
Tempowerkring 8 D-21079 Hamburg - Germany phone:+49 (0)40 790 12 734 fax: +49 (0)40 790 12 736
info@el-cell.com www.el-cell.com
Page 3 of 18 Manual_ECC_PRESS_Release_3.6.docx / 12.06.2017
1 Product Description
The ECC-Press test cell is dedicated to the measurement of the pressure build-up during the electrochemical cycle. The low-dead-volume sampling port/valve serves to draw gas samples from the cell’s headspace. The cell hardware is based on the ECC-Std and ECC-Ref test cells and can be used with or without a reference electrode.
The ECC-Press comes with a controller box for easy interfacing with an external potentiostat. Analog outputs for both pressure and temperature are provided for seamless integration with external instruments. As an option, all relevant signals of the experiment – cell pressure, temperature, cell potentials and current – may be recorded with an integrated USB data logger (DL option).
This manual covers only the special features of the ECC-Press and ECC-Press-DL. It is assumed that the reader is already familiar with the assembly and operation of both the ECC-Ref and the ECC-Std test cell.
Pressure sensor 0 to 2.5 bar abs.
Sensor dead volume spacer (2.424 cm3)
Stand (Wall mountable)
Bypass port with PTFE plug Transfer line
Dead volume spacer cell (0.723 cm3)
Lid with two ports (1/4“-28) and 2mm banana jacks
Sampling valve with septum port
Cell base with reference pin assembly
Lid seal
Page 4 of 18 Manual_ECC_PRESS_Release_3.6.docx / 12.06.2017
2 Features and Specification
Precise pressure sensor attached to the electrochemical cell for monitoring
pressure changes during charge / discharge
Sampling valve with septum port for drawing gas samples by means of a
syringe
Default pressure range of 0 to 2.5 bar (absolute) Leakage tested; after 24h the pressure changes by less than 0.2 mbar/h
(tested with PEEK lid seal, air, 2 bar, 50°C, both spacers installed)
Conditioning electronics with analog outputs for pressure and temperature Optional integrated USB data logger for recording of cell pressure,
temperature, cell potentials and current
Adjustable dead volume between 3.024 cm3 and 6.172 cm3.
Dead volume in cm3
1. Both spacers installed*
3.024
2. Only ECC1-00-0202-A (sensor spacer) installed
3.747
3. No spacer installed**
6.172
4. Only ECC1-00-0103-B (cell spacer) installed**
5.449
* delivery condition
**Requires de-mounting of the sensor and must be done at factory.
Visualised dead volume (both spacers installed = delivery condition = 3.024 cm
3
)
Fully equipped for operation with and without a reference electrode (contains
all parts of ECC-Std and ECC-Ref test cells)
Page 5 of 18 Manual_ECC_PRESS_Release_3.6.docx / 12.06.2017
High precision 18 mm diameter sandwich geometry with <0.1 mm electrode
concentricity
Operating temperature range -20 to +70°C (cell). We recommend using PEEK
lid seals above 50°C.
Easy and reliable electrolyte filling upon assembly Fast assembly and dismantling, and easy cleaning of cell components Electrodes are easily accessible for post-mortem analysis Reusable cell components except for lid seals Small and defined electrolyte volume down to 0.05 cm3  Adjustable, reproducible and homogeneous mechanical pressure on
electrodes
Materials in media contact are stainless steel 1.4404 and PEEK (other
materials on request)
Modular cell construction with many interchangeable components. Size (including stand): 90 mm x 80 mm x 240 mm (height x width x depth) Weight: 900 g Dedicated tools available to ease cell assembly and operation
http://el-cell.com/products/accessories-tools
Page 6 of 18 Manual_ECC_PRESS_Release_3.6.docx / 12.06.2017
Dimensions (Cell)
Electrical Specifications (Controller)
- Input: 24V (0.6A) supplied by PSA15R-240P mains adapter
- Analog output voltages: 0 to 10V (1 kOhm / 100 nF) for pressure and temperature
- Resolution of pressure analog output: <0.05% of measurement range
- Error of pressure analog output: <0.5% of measurement range (comprising non-linearity including hysteresis and non-reproducibility)
- Resolution of temperature analog output: <0.01°C
- Error of temperature output: < 1°C
- Optional USB data logger powered by host PC with galvanic isolation
- Operating temperature range (controller): 0 to 40°C